LivaNova VNS Therapy 250 Physician's Manual

Physician’s Manual
For Healthcare Professionals
VNS Therapy® Programming Software
Model 250 Version 8.1 Handheld Computer
June 2017
Worldwide Version
Caution: U.S. federal law restricts this device to sale by or on the order of a physician.
Note: This manual contains information on the use of the LivaNova VNS Therapy Programming Software, Model 250 Version 8.1. Physicians should refer to the VNS Therapy Pulse Generator physician’s manuals for additional important prescribing and safety information. A copy of this VNS Therapy manual is posted at www.livanova.com.
0344
2011
26-0008-1100/5 (Worldwide) — 1 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
© Copyright 2012-2017 LivaNova, PLC, London, UK All rights reserved.
LivaNova is a registered United States trademark of LivaNova, PLC. NCP, Demipulse, Demipulse Duo, Perennia, VNS Therapy, AspireHC, PerenniaFLEX, PerenniaDURA and AspireSR are registered trademarks of LivaNova USA, Inc. Pulse and Pulse Duo are trademarks of LivaNova USA, Inc. Corresponding foreign trademarks may also be registered or pending.
2 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Table of Contents
1. BRIEF DEVICE DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.1. General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2. Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3. Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.4. System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.5. Conventions Used in this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.6. Communication Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2. H
OW TO GET STARTED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3. H
OW TO USE THE SOFTWARE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1. MAIN Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.1. Interrogate Device overview . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.2. View Database overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.3. View Last Parameters overview . . . . . . . . . . . . . . . . . . . . . . 13
3.1.4. User Preferences overview . . . . . . . . . . . . . . . . . . . . . . . . . 13
4. I
NTERROGATE DEVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.1. How to Program New Parameter Settings . . . . . . . . . . . . . . . . . . . . 21
4.1.1. Partial Programming Due to Interrupted Programming
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.1.2. Cross-programming (applicable to Model 100, 101, and
102 Pulse Generators ONLY) . . . . . . . . . . . . . . . . . . . . . . . 26
4.2. PARAMETER Screen Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.2.1. Interrogate Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.2.2. Device Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.2.2.1. Interrupted System and Generator Diagnostics . . . 31
4.2.2.2. Diagnostic test parameters . . . . . . . . . . . . . . . . . . . 34
4.2.2.3. System Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . 39
4.2.2.4. Normal Mode Diagnostics . . . . . . . . . . . . . . . . . . . 40
4.2.2.5. Magnet Mode Diagnostics . . . . . . . . . . . . . . . . . . . 41
4.2.2.6. Generator Diagnostics . . . . . . . . . . . . . . . . . . . . . . 42
4.2.3. Program Patient Data screen . . . . . . . . . . . . . . . . . . . . . . . 44
4.2.4. Display Device History screen . . . . . . . . . . . . . . . . . . . . . . . 45
4.2.5. View Database screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4.2.6. Close Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4.2.7. Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4.2.8. Handheld computer battery status . . . . . . . . . . . . . . . . . . . . 46
5. V
IEW DATABASE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1. Specify Search Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1.1. Sort by Date, Patient ID, or Serial Number . . . . . . . . . . . . . 48
5.1.2. View All Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.1.3. Earliest Date/Latest Date . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.1.4. Patient ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.1.5. Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.2. View the Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.2.1. View Parameter History . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.2.2. View Diagnostic History . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.2.3. View Magnet History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
26-0008-1100/5 (Worldwide) — 3 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
6. VIEW LAST PARAMETERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7. U
SER PREFERENCES SCREEN. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.1. Set Timeout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.2. Set Maximum Number of Records to View . . . . . . . . . . . . . . . . . . 55
7.3. Set the Font Size for Database Displays . . . . . . . . . . . . . . . . . . . . . 55
7.4. Set Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.5. Backup to FlashCard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.6. Restore from FlashCard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.7. Export Database to Text File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.8. Language Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.9. Handheld Computer Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . 57
8. M
AINTENANCE, HANDLING, AND STORAGE . . . . . . . . . . . . . . . . . . . . . . . . 58
9. P
RECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10. T
ROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
10.1. General Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
10.1.1. Non-responsive Handheld Computer . . . . . . . . . . . . . . . . . . 61
10.2. Hardware Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
10.2.1. Restore and reinstall software . . . . . . . . . . . . . . . . . . . . . . . 62
10.2.2. Handheld screen alignment . . . . . . . . . . . . . . . . . . . . . . . . . 62
10.2.2.1.Screen alignment for the Dell Axim X5 handheld
computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
10.2.2.2.Screen alignment for the Dell Axim X50 handheld
computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10.2.3. Other hardware issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
10.3. Troubleshooting in the OR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
10.3.1. Communication problems in OR . . . . . . . . . . . . . . . . . . . . . 64
10.3.2. High Lead impedance on System Diagnostics in OR –
initial implant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
10.3.3. High Lead impedance on System Diagnostics in OR –
Pulse Generator replacement . . . . . . . . . . . . . . . . . . . . . . . . 69
10.3.4. Low Lead impedance on System Diagnostics in OR –
initial implant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
10.3.5. Low Lead impedance on System Diagnostics in OR –
Pulse Generator replacement . . . . . . . . . . . . . . . . . . . . . . . . 74
10.3.6. Low Battery/End of Service indications prior to surgery, outside of the sterile field – initial implant or Pulse Generator
replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
10.3.7. Low Battery/End of Service indications during surgery, inside the sterile field – initial implant or Pulse Generator
replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
10.4. Troubleshooting at Follow-up Visits . . . . . . . . . . . . . . . . . . . . . . . . 78
10.4.1. Communication problems at follow-up visits . . . . . . . . . . . . 78
10.4.2. High Lead Impedance on a Diagnostic Test at follow-up
visit for Models 100, 101, and 102 Pulse Generators . . . . . 79
10.4.3. High Lead Impedance, Low Lead Impedance, or Low Output Current on a Diagnostic Test at follow-up visit
for Model 103, 104, and 105 Pulse Generators . . . . . . . . . . 83
10.4.4. Pulse Generator disabled due to Vbat < EOS found at first interrogation after implantation or other
surgical procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
4 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
10.4.5. Sudden decrease in estimated device longevity (% battery
power remaining) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
11. G
LOSSARY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
12. L
IST OF APPROVED COMPUTERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
13. P
RODUCT INFORMATION AND SUPPORT . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
List of Tables
TABLE 1PULSE GENERATOR BATTERY LEVEL INDICATIONS AND
R
ECOMMENDATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
T
ABLE 2PROGRAMMABLE PARAMETERS FOR THE PULSE GENERATOR . . 21
T
ABLE 3CROSS-PROGRAMMING EXAMPLE . . . . . . . . . . . . . . . . . . . . . . . . 27
T
ABLE 4DEVICE DIAGNOSTIC SCREEN STATUS READINGS FOR
C
OMMUNICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
T
ABLE 5DEVICE DIAGNOSTIC SCREEN STATUS READINGS FOR OUTPUT
S
TATUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
T
ABLE 6DEVICE DIAGNOSTIC SCREEN STATUS READINGS FOR OUTPUT
C
URRENT FOR MODELS 100, 101, AND 102 . . . . . . . . . . . . . . . . 36
T
ABLE 7DEVICE DIAGNOSTIC SCREEN STATUS READINGS FOR LEAD
I
MPEDANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
T
ABLE 8DC-DC CONVERTER CODES AND LEAD IMPEDANCE (MODEL
100, SN ³10,000) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
T
ABLE 9DC-DC CONVERTER CODES AND LEAD IMPEDANCE
(M
ODEL 101) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
T
ABLE 10 DC-DC CONVERTER CODES AND LEAD IMPEDANCE
(M
ODEL 102) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
T
ABLE 11 MINIMUM NORMAL MODE AND MAGNET MODE DIAGNOSTIC
S
ETTINGS (MODELS 100, 101, AND 102) . . . . . . . . . . . . . . . . . . . 42
26-0008-1100/5 (Worldwide) — 5 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
List of Figures
FIGURE 1 MAIN MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
F
IGURE 2 START INTERROGATION SCREEN . . . . . . . . . . . . . . . . . . . . . 14
F
IGURE 3 PARAMETER SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
F
IGURE 4 TIMEOUT WARNING SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . 16
F
IGURE 5 PARAMETER SCREEN WITH ACTIVE BATTERY STATUS
INDICATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
F
IGURE 6 BATTERY STATUS INDICATOR - NEAR EOS WARNING
S
CREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
F
IGURE 7 BATTERY STATUS INDICATOR - IFI WARNING
S
CREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
F
IGURE 8 BATTERY STATUS INDICATOR - N EOS WARNING
S
CREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
F
IGURE 9 PULSE DISABLED AND BATTERY STATUS INDICATOR
- EOS WARNING S
F
IGURE 10 PARAMETER SETTINGS POP-UP MENU . . . . . . . . . . . . . . . . 22
CREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
F
IGURE 11 OUTPUT WARNING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
F
IGURE 12 START PROGRAMMING SCREEN . . . . . . . . . . . . . . . . . . . . . . 23
F
IGURE 13 PROGRAMMING FAILED WARNING SCREEN . . . . . . . . . . . 24
F
IGURE 14 CHANGED SETTINGS PROGRAMMING WARNING
S
CREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
F
IGURE 15 PROGRAMMING FAILED WARNING SCREEN . . . . . . . . . . . 26
F
IGURE 16 PARAMETER SCREEN MENU . . . . . . . . . . . . . . . . . . . . . . . . . . 28
F
IGURE 17 DEVICE DIAGNOSTICS MENU . . . . . . . . . . . . . . . . . . . . . . . . 29
F
IGURE 18 DEVICE DIAGNOSTICS SUB-MENU . . . . . . . . . . . . . . . . . . . . 30
F
IGURE 19 DIAGNOSTIC - SINGLE FAILURE WARNING SCREEN . . . . 32
F
IGURE 20 DIAGNOSTIC - MULTIPLE FAILURE WARNING SCREEN . 32
F
IGURE 21 CHANGED SETTINGS - INTERRUPTED DIAGNOSTIC
WARNING S
F
IGURE 22 PARAMETER SCREEN WITH UNKNOWN SETTINGS . . . . . . 33
F
IGURE 23 SYSTEM DIAGNOSTICS SCREEN . . . . . . . . . . . . . . . . . . . . . . 40
F
IGURE 24 NORMAL MODE DIAGNOSTICS SCREEN . . . . . . . . . . . . . . . 41
F
IGURE 25 MAGNET MODE DIAGNOSTICS SCREEN . . . . . . . . . . . . . . . 42
F
IGURE 26 GENERATOR DIAGNOSTICS SCREEN . . . . . . . . . . . . . . . . . . 43
F
IGURE 27 PROGRAM PATIENT DATA SCREEN . . . . . . . . . . . . . . . . . . . 44
CREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
F
IGURE 28 DEVICE HISTORY SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
F
IGURE 29 VIEW DATABASE SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
F
IGURE 30 VIEW PARAMETER HISTORY SCREEN . . . . . . . . . . . . . . . . . 50
F
IGURE 31 VIEW DIAGNOSTIC HISTORY SCREEN . . . . . . . . . . . . . . . . . 51
6 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
FIGURE 32 VIEW MAGNET HISTORY SCREEN . . . . . . . . . . . . . . . . . . . . 52
F
IGURE 33 VIEW LAST PARAMETERS SCREEN . . . . . . . . . . . . . . . . . . . 53
F
IGURE 34 USER PREFERENCES SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . 54
F
IGURE 35 LANGUAGE SELECTION SCREEN . . . . . . . . . . . . . . . . . . . . . 57
F
IGURE 36 CORRECT FLASHCARD ORIENTATION . . . . . . . . . . . . . . . . . . . . . 60
F
IGURE 37 DELL AXIM X5 HANDHELD COMPUTER . . . . . . . . . . . . . . . . . . . 63
F
IGURE 38 DELL AXIM X50 HANDHELD COMPUTER . . . . . . . . . . . . . . . . . . 64
F
IGURE 39 COMMUNICATION PROBLEMS IN THE OR AND AT FOLLOW-UP
V
ISITS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
F
IGURE 40 HIGH LEAD IMPEDANCE ON SYSTEM DIAGNOSTIC TEST IN OR -
I
NITIAL IMPLANT OR GENERATOR REPLACEMENT . . . . . . . . . . . 71
F
IGURE 41 LOW LEAD IMPEDANCE ON SYSTEM DIAGNOSTICS IN OR -
I
NITIAL IMPLANT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
F
IGURE 42 LOW LEAD IMPEDANCE ON SYSTEM DIAGNOSTICS IN OR -
P
ULSE GENERATOR REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . 74
F
IGURE 43 LOW BATTERY/END OF SERVICE INDICATIONS PRIOR TO
S
URGERY, OUTSIDE OF THE STERILE FIELD - INITIAL
I
MPLANT OR PULSE GENERATOR REPLACEMENT . . . . . . . . . . . . 76
F
IGURE 44 HIGH LEAD IMPEDANCE ON A DIAGNOSTIC TEST AT
F
OLLOW-UP VISIT FOR MODELS 100, 101, AND 102 PULSE
G
ENERATORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
F
IGURE 45 HIGH LEAD IMPEDANCE, LOW LEAD IMPEDANCE, OR
L
OW OUTPUT CURRENT ON AN INTERROGATION OR
D
IAGNOSTIC TEST AT FOLLOW-UP VISIT FOR MODELS
103, 104,
AND 105 PULSE GENERATORS . . . . . . . . . . . . . . . . . . . 84
26-0008-1100/5 (Worldwide) — 7 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
8 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
1. BRIEF DEVICE DESCRIPTION ___________

1.1. General Description

The LivaNova® VNS Therapy® Programming Software, Model 250 Version 8.1, enables you to interrogate and program LivaNova’ Pulse Generator using the Programming Wand.
Software capabilities include:
Use of the software with the Programming Wand to interrogate and
program a Pulse Generator’s parameters and assess Pulse Generator function.
Use of the software without the Programming Wand to display the
operating histories of all the Pulse Generators interrogated with your handheld computer.
Use of screen displays that provide prompts and messages to aid in
interrogating and programming.

1.2. Symbols

Symbols and definitions used for the VNS Therapy System include the following:
Notice for reader to pay special attention to details that follow
Serial Number
Storage
Batch Code
Date of Manufacture
Information
Sidebar Note (cross-references and other useful information)
MRI Unsafe
26-0008-1100/5 (Worldwide) — 9 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Humidity limitations—Indicates the range of humidity to which the medical device can be safely exposed

1.3. Intended Use

The Model 250 VNS Therapy Programming Software is intended for use only with the LivaNova Pulse Generators and Programming Wand and is subject to the same indications for use.

1.4. System Requirements

The computer must have the following minimum specifications:
Microsoft
Pocket-PC (2002, 2003 Operating System) or Windows
Mobile (2003 2nd Edition)
At least 16 MB of memory
Compact FlashCard
Serial communication capability
The communication system of the VNS Therapy Programming Software Model 250 Version 8.1 is designed to minimize the possibility of misprogramming or “phantom” programming. Phantom programming is the inadvertent programming via environmental sources of electromagnetic interference.
For Models 100, 101, and 102 Pulse Generators, each parameter is programmed and verified individually during a programming event. For the Models 103, 104, and 105 Pulse Generators the parameters are programmed and verified as a group during a programming event.

1.5. Conventions Used in this Manual

Caution: LivaNova recommends that the software be loaded on a dedicated computer used only to program the Pulse Generator.
Note: See “List of Approved Computers” on page 91 for a list of computers that have been qualified for use with the VNS Therapy Programming Software Model 250 Version 8.1.
Note: See the physician’s manual for the VNS Therapy Pulse Generators for a complete description of the Pulse Generator, its indications for use, and its operation.
The following conventions are used throughout this manual:
The VNS Therapy Programming Software Model 250 Version 8.1
operates in a Windows environment with drop-down menus. You will use a stylus to navigate through a Windows environment on the handheld computer. The stylus resembles a small pen and is kept with handheld computer.
Screen titles appear in all capital letters. The names of selectable
buttons (or fields) and drop-down menus appear in bold typeface. Each of these buttons or fields represents an operation, parameter, or parameter value.
10 of 92 — 26-0008-1100/5 (Worldwide)
Note: See the physician’s manual for the Programming Wand for a description of the Programming Wand.
Physician’s Manual — VNS Therapy Programming Software (8.1)
To select a button or field, tap it with the stylus.
Some information can be entered with a “soft” or “on-screen”
keyboard (an example of the soft keyboard is shown at the bottom of Figure 27 on page 44). To maximize (enable) or minimize (disable) the soft keyboard, tap the keyboard icon at the lower right-hand corner of the screen. You can type letters or numbers by tapping them with the stylus.
When a selection requires communication with the Pulse Generator, a
message will appear prompting you to hold the Programming Wand over the Pulse Generator.
When a button is selected, a border will appear around the button (as
on the Interrogate Device button in Figure 1 on page 12).
To return to a previous screen, tap Menu or Exit.
A list of symbols is found in the front of this manual.
A glossary is found in the back of this manual.
The words ON and OFF appear in all capital letters only when they
refer to VNS stimulation.
References to Model 102 include both Models 102 and 102R.
For more information, see “Troubleshooting ” on page 61.

1.6. Communication Signals

Four rapid, ascending musical tones and a screen message signal the successful completion of an interrogation or programming operation.
Four rapid, descending musical tones and a screen message signal a warning or a failed operation.
The DATA/RCVD indicator on the Programming Wand lights up during a successful programming operation or interrogation.
Caution: If the Procedure Failed message appears on the screen, check both the connection between the handheld computer and the Programming Wand and the positioning of the Programming Wand in relation to the Pulse Generator (see the physician’s manual for the Programming Wand). If you continue to have communication problems, change the battery in the Programming Wand.
Caution: The DATA/RCVD indicator on the Programming Wand may also light up or flicker in the presence of electromagnetic interference (EMI) or noise. Information on how to detect EMI is provided in the physician’s manual for the Programming Wand, sections entitled Understanding the Indicator Lights and Troubleshooting. If the presence of EMI or noise is suspected, you must verify that the programming or interrogation occurred successfully by reviewing the PARAMETER Screen shown in Figure 3 on page 15.
26-0008-1100/5 (Worldwide) — 11 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
2. HOW TO GET STARTED _______________
The VNS Therapy Programming Software Model 250 Version 8.1 has been pre-installed on the handheld computer. Before proceeding to the next section, be sure you have arrived at the MAIN Menu as shown in Figure 1. If for any reason the software must be reinstalled, directions for doing so are provided in the Troubleshooting section of this manual.
1. Turn on the handheld computer by pressing the on-off button. The handheld computer will display the same screen it was displaying when it was last turned off (unless it was reset since the last use).
2. If the computer doesn’t display the MAIN Menu, select Exit or Menu until the MAIN Menu is displayed.
3. If you haven’t been able to access the MAIN Menu after following Steps 1 and 2 above, see “Precautions ” on page 59 and “Troubleshooting ” on page 61 before proceeding.
Figure 1. MAIN Menu
Caution: When in use,
LivaNova recommends the handheld computer to be operated on battery power only (unplugged from AC power). Having the handheld computer plugged into AC power during use may contribute to communication difficulties. When not in use, the handheld computer should be plugged into AC power to maintain charge.
12 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
3. HOW TO USE THE SOFTWARE __________
It is LivaNova’s intent to make the software as intuitive and “user friendly” as possible. Messages and prompts will guide you through this software.
There are four menus in this software. The first is the MAIN Menu.

3.1. MAIN Menu

The following sections provide an overview of each of the MAIN Menu’s four primary functions, as shown in Figure 1.

3.1.1. Interrogate Device overview

Note: See “Interrogate Device” on page 14 for more information.
Note: See “View Database” on page 47 for more information.
Note: See “View Last Parameters” on page 53 for more information.
Note: See “User Preferences Screen ” on page 54 for more information.
The first option, Interrogate Device, is used to establish communication with the implantable device (Pulse Generator). The other three options work with data stored on the handheld computer itself and require no link to the device via the Programming Wand.

3.1.2. View Database overview

View Database is used to view information from all Pulse Generator interrogations and programming events performed by your handheld computer and the Programming Software.

3.1.3. View Last Parameters overview

View Last Parameters is used to view the most recent parameters selected for a specific patient or Pulse Generator.

3.1.4. User Preferences overview

User Preferences allows you to select values for five user settings:
The number of minutes that the computer may be idle before a new
interrogation becomes necessary (applicable for Models 100C, 101, and 102 only).
The maximum number of records viewable at a time
The size of type font for the diagnostics screens
The current time and date
Language selection, when available
26-0008-1100/5 (Worldwide) — 13 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
4. INTERROGATE DEVICE________________

To retrieve information from the Pulse Generator or to program the Pulse Generator, you must first select the Interrogate Device button from the MAIN Menu. The START INTERROGATION Screen will be displayed (see Figure 2).

Figure 2. START INTERROGATION Screen
To interrogate the Pulse Generator, place the Programming Wand over the Pulse Generator and tap Start Interrogation. Tap Cancel to return to the previous screen.
Upon completion of a successful interrogation, the Programming Software automatically displays the PARAMETER screen (see Figure 3 on page 15) unless an error condition (i.e., Generator End of Service or High Lead Impedance) is detected with the Pulse Generator. If an error condition is detected with the Pulse Generator, the associated warning message is displayed and must be acknowledged by the user before proceeding to the PARAMETER screen. The PARAMETER screen displays the operating parameters of the Pulse Generator with their current settings.
There is one line for each of the eight parameters. Each line contains three types of information:
1. The name of the parameter and the units used to express its values
2. The present setting for that particular parameter (***** is shown for any unknown or out of range parameter values)
14 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Models 100, 101, 102 Models 103, 104, 105
3. The New button, when tapped, displays the range of possible settings for that parameter
Figure 3. PARAMETER Screen
Note: The inactivity
timeout function is not used in Models 103, 104, and 105.
Inactivity Timeout Indicator (Progress Bar)
For the Models 100, 101, 102 Pulse Generators, the progress bar across the bottom of the screen is an “inactivity timeout” indicator. When the bar reaches the right margin, the most recent interrogation will have expired and the warning shown in Figure 4 on page 16 will be displayed. Also, it will be necessary to reinterrogate the Pulse Generator before programming. The inactivity timeout period can be adjusted on the USER PREFERENCES Screen. See See “User Preferences Screen ” on page 54.
26-0008-1100/5 (Worldwide) — 15 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 4. TIMEOUT WARNING Screen
Note that the timeout bar has advanced to the right margin.
In addition to the eight programmable parameters, the PARAMETER screen will also indicate battery status for the Pulse Generator. The Programming Software has one battery status indicator, Near End of Service (Near EOS), for Models 100, 101, and 102 Pulse Generators. The Programming Software has three indicators, Intensified Follow-up Indicator (IFI), Near End of Service (N EOS), and End of Service (EOS), for Models 103, 104, and 105 Pulse Generators. On the PARAMETER screen, the Near EOS status for Models 100, 101, and 102 Pulse Generators is displayed on the lower portion of the screen (see Figure 3 on page 15). For Model 103, 104, and 105 Pulse Generators, the remaining generator battery power is indicated by the battery icon located in the top left portion of the PARAMETER screen (see Figure 3 on page 15). The battery icon is shaded and colored to indicate various levels of remaining generator battery power. As the battery level decreases and the Pulse Generator enter states of IFI = Yes, N EOS = Yes or EOS = Yes, an additional indicator is displayed at the bottom of the PARAMETER screen (see Figure 5 on page 17).
Note: For more information on the timeout function, see “User Preferences Screen ” on page 54.
Caution: LivaNova recommends that you reinterrogate the Pulse Generator after each programming event to verify that the programming occurred as planned.
16 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 5. PARAMETER Screen with ACTIVE BATTERY
STATUS INDICATOR
Models 103, 104, 105
Upon completion of an interrogation or Diagnostics testing, the Programming Software will display the associated warning message should the Pulse Generator battery reach Near EOS for Models 100, 101, and 102 (see Figure 6 on page 17) and IFI, N EOS, or EOS for Models 103, 104, and 105 (see Figure 7, Figure 8, and Figure 9 on pages 18 and
19).
Figure 6. BATTERY STATUS INDICATOR - NEAR EOS
WARNING Screen
Models 100, 101, 102
26-0008-1100/5 (Worldwide) — 17 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 7. BATTERY STATUS INDICATOR - IFI
WARNING Screen
Models 103, 104, 105
Figure 8. BATTERY STATUS INDICATOR - N EOS
WARNING Screen
Models 103, 104, 105
18 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 9. PULSE DISABLED and BATTERY STATUS
INDICATOR - EOS WARNING Screen
Table 1 shows the various indications of Pulse Generator battery level and the associated recommendations.
Table 1. Pulse Generator Battery Level Indications and
Pulse
Generator
Model #
100/101/102 Battery
100/101/102 Battery
Estimated %
Battery Power
Remaining
measurement not supported by Generator
measurement not supported by Generator
Models 103, 104, 105
Recommendations
Battery
Icon
Displayed
None displayed
None displayed
Pulse Generator Battery Status /
Near EOS = No
The Pulse Generator battery level is good. No warning message is displayed from the Programming Software and no special attention is required.
Near EOS = Yes
The Pulse Generator is nearing End of Service (EOS). The Programming Software will display a warning message indicating this status upon completion of interrogation or Diagnostics testing. It is recommended that the Pulse Generator be replaced as soon as possible.
Recommendations
26-0008-1100/5 (Worldwide) — 19 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Pulse
Generator
Model #
103/104/105 >75% to 100% IFI = No
103/104/105 >50% to 75%
103/104/105 >25% to 50%
103/104 >18% to 25%
105 >11% to 25%
103/104 >8% to 18% IFI = Yes
105 >5% to 11%
Estimated %
Battery Power
Remaining
Battery
Icon
Displayed
Pulse Generator Battery Status /
The Intensified Follow-up Indicator (IFI) has not been set. The Pulse Generator battery level is good. No warning message is displayed from the Programming Software and no special attention is required.
The Intensified Follow-up Indicator (IFI) has been set for the Pulse Generator and a warning message will be displayed by the Programming Software to indicate this status upon completion of interrogation or Diagnostics testing. The Pulse Generator battery has depleted to a level where more frequent clinical monitoring is recommended.
Recommendations
103/104 >0% to 8% N EOS = Yes
The Pulse Generator is near End of Service (N EOS). The Programming Software will
105 >0% to 5%
103/104/105 0% EOS = Yes
display a warning message indicating this status upon completion of interrogation or Diagnostics testing. It is recommended that the Pulse Generator be replaced as soon as possible.
The Pulse Generator has reached End of Service (EOS) and is NOT supplying stimulation. The Programming Software will display a warning message indicating this status upon completion of interrogation or Diagnostics testing. Immediate replacement of the Pulse Generator is recommended. If the Pulse Generator is not replaced, it will eventually lose the ability to communicate with the Programming Software.
20 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)

4.1. How to Program New Parameter Settings

The PARAMETER Screen shown in Figure 3 on page 15 is used to change the Pulse Generator’s parameter settings. Table 2 shows the range of possible settings for all parameters.
Table 2. Programmable Parameters for the Pulse
Generator
Programmable
Parameters
Output Current
Frequency
Pulse Width
Signal ON Time
Signal OFF
Time
0.0-3.5 milliamperes (in 0.25-mA steps)
1, 2, 5, 10, 15, 20, 25, 30 Hz
130, 250, 500, 750, 1,000 μsec
7, 14, 21, 30, 60 sec
0.2, 0.3, 0.5, 0.8, 1.1, 1.8, and 3 min; 5 to 60 in 5-min steps; 60 to 180 in 30-min steps
Settings
To change a parameter setting, follow these steps from the PARAMETER Screen (Figure 3 on page 15):
1. Tap the New button for the parameter you want to change. A pop-up menu displays the range of possible values, with the current value setting appearing in boldface type (see Figure 10). If there are values greater than or less than those shown on the screen you can view them by tapping the appropriate set of the double arrows. Look at the left half of the screen and note that the parameter for which the value is being changed is enabled (clearly visible) and the other parameters are disabled (“grayed-out”). In the example in Figure 10, Output Current is enabled and the other parameters are disabled. The value of 1.00 mA has been selected for the Output Current parameter.
26-0008-1100/5 (Worldwide) — 21 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 10. PARAMETER SETTINGS POP-UP Menu
2. Select the new target value for the Output Current by tapping the desired value on the screen. If the target value selected is greater than
0.25 mA above the currently programmed value setting in the Pulse Generator, an Output Warning will appear (see Figure 11 on page 22).
Note: LivaNova recommends that during the initial programming, the output current be programmed at 0 mA and then slowly increased by 0.25 mA increments until the patient feels the stimulation at a comfortable level. Even patients who are receiving replacement generators should be started at 0 mA Output Current, followed by incremental increases of
0.25 mA to allow for reaccommodation to the therapy.
Figure 11. OUTPUT WARNING
22 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
3. Tap the Program button, which turns green when a new parameter value is selected, at the bottom of the PARAMETER Screen to access the START PROGRAMMING Screen (see Figure 12).
Figure 12. START PROGRAMMING Screen
4. Place the Programming Wand over the Pulse Generator.
5. From the “Start Programming” screen, tap Start Programming to program the new value or Cancel to return to the PARAMETER screen menu.
Caution: For the Models 100, 101, and 102 Pulse Generators, do not use frequencies of 5 Hz or less for long-term stimulation. These frequencies always generate an electromagnetic trigger signal which results in excessive battery depletion of the implanted Pulse Generator; therefore, use these low frequencies for short periods of time only.
Caution: Stimulation at a combination of high frequency (≥ 50 Hz) and ON time OFF time has resulted in degenerative nerve damage in laboratory animals. ON time OFF time can be simulated by very frequent magnet activation. LivaNova recommends that stimulation at these combinations of ranges be avoided.
4.1.1. Partial Programming Due to Interrupted
Programming Operation
For Models 100, 101, and 102 Generators, each parameter is programmed and verified individually during a programming event. Therefore, these Pulse Generators can be partially programmed to unintended settings if the communication between the Wand and the Pulse Generator is interrupted during programming. If this occurs, the Programming Software displays a
26-0008-1100/5 (Worldwide) — 23 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
warning message indicating that the procedure failed and device settings were altered due to the interrupted programming attempt (see Figure 13 on page 24). The warning message also indicates some troubleshooting options and allows the user to retry or cancel the programming operation. If the user selects Retry, the Programming Software will re-program the Pulse Generator. If the user selects Cancel, the Programming Software will default back to the START INTERROGATION screen (see Figure 2). It is recommended that the Pulse Generator be interrogated immediately, as this will verify the currently programmed settings. After completion of the interrogation, the Programming Software displays a warning message stating that device settings were altered due to the interruption (see Figure 14) and allows the user to re-program the Pulse Generator to desired settings upon selecting Proceed.
Figure 13. PROGRAMMING FAILED WARNING Screen
Models 100, 101, 102
24 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 14. CHANGED SETTINGS PROGRAMMING
WARNING Screen
Models 100, 101, 102
For Models 103, 104, and 105 Generators, the device parameters are programmed and verified as a group during a programming event; therefore, these Pulse Generators are not susceptible to partial programming. If an interruption occurs during a programming step for these Pulse Generators, the Programming Software displays a warning message indicating that the procedure failed and allows the user to retry or cancel the programming operation (see Figure 15 on page 26). If the user selects Retry, the Programming Software will re-program the Pulse Generator. If the user selects Cancel, the Programming Software will return to the PARAMETER screen where the user can re-program the Pulse Generator to desired settings (see Figure 3 on page 15).
26-0008-1100/5 (Worldwide) — 25 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 15. PROGRAMMING FAILED WARNING Screen
Models 103, 104, 105

4.1.2. Cross-programming (applicable to Model 100, 101, and 102 Pulse Generators ONLY)

Models 100, 101, and 102 Pulse Generators are susceptible to an event known as cross-programming, where parameter settings from one patient's Pulse Generator are inadvertently programmed to another patient's Pulse Generator. Cross-programming occurs when all the following conditions are met:
1. Patient A is implanted with the same model Pulse Generator as Patient
B
2. Patient A's Pulse Generator was interrogated/programmed with the
Programming Software and the Programming Software is left on the PARAMETER screen
3. Inactivity timeout has not expired, or is disabled
4. Patient B's Pulse Generator is programmed at the PARAMETER
screen in step 2 without first performing an interrogation
The result of the above scenario is that Patient B's output current and magnet output current settings are programmed to Patient A's output current and magnet output current settings, unless those parameters values were changed as part of step 4. In addition, the Programming Software database will incorrectly indicate Patient A to be programmed to setting changes made in step 4. Table 3 gives an example of cross-programming
26 of 92 — 26-0008-1100/5 (Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
and its effect on device parameters. The parameters marked with asterisks indicate unintentionally altered or incorrectly displayed values.
Table 3. Cross-programming Example
After Step 2 At Step 4 After Step 4
Parameter
Output Current (mA)
Frequency (Hz)
Pulse Width (µs)
On Time (s) 30 7 No change 7 30
Off Time (min) 5 1.8 3 3 3*
Mag. Output Current (mA)
Mag. On Time (s)
Mag. Pulse Width (µs)
Patient A Final Settings
0.25 0.75 No Change 0.25* 0.25
30 20 15 15 15*
250 500 No Change 500 250
0.5 1.00 No Change 0.5* 0.5
30 30 60 60 60*
500 500 250 250 250*
*Unintentionally altered or incorrectly displayed values.
Patient B Initial Settings
Patient B Intended Final Settings
Patient B Actual Final Settings
Patient A Final Settings in Database
Due to the susceptibility of Model 100, 101, and 102 Pulse Generators to cross-programming, LivaNova recommends that an initial and a final interrogation be performed at each office visit for all VNS patients in order to verify parameter settings.

4.2. PARAMETER Screen Menu

Tapping the Menu button at the right top of the PARAMETER Screen causes the PARAMETER Screen Menu to be displayed, as shown in Figure 16 on page 28.
26-0008-1100/5 (Worldwide) — 27 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 16. PARAMETER Screen Menu
The functions available via this menu are explained in the following paragraphs.

4.2.1. Interrogate Device

Interrogation of the Pulse Generator is required to reach the menu shown in Figure 16. An Interrogate Device button is available here so you can interrogate the device after each programming event. A post-programming interrogation should be performed to ensure that the Pulse Generator was programmed as desired.

4.2.2. Device Diagnostics

Tap the Device Diagnostics button on the PARAMETER Screen Menu to display the DEVICE DIAGNOSTICS Menu (see Figure 17 on page 29).
28 of 92 — 26-0008-1100/5 (Worldwide)
Loading...
+ 64 hidden pages