All rights reserved, especially the right to reproduction and distribution as well as translation. No part of this
document may be reproduced – by photocopy, microfilm or any other process – nor may any part of it be
stored, edited, duplicated or distributed by electronic means without the written permission of
LivaNova Deutschland GmbH.
Trademarks used in these operating instructions:
The name “Stöckert®” is a registered trademark.
Minncare® is a registered trademark of the Minntech Corporation, a Cantel Medical Company.
Clorox® is a registered trademark of the Clorox Company.
Pall-Aquasafe™ is a registered trademark of the Pall Corporation.
CaviWipesTM and CaviCideTM are registered trademarks of Metrex Research, LLC.
Indications for use
The Heater-Cooler System 3T is used to circulate water through heat exchangers to warm or cool a patient
during cardiopulmonary bypass procedures lasting 6 hours or less.
Caution: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
Distributed in the U.S.A. by:
LivaNova USA, Inc.
14401 West 65th Way
Arvada, CO 80004
8.2Troubleshooting or returning the 3T Aerosol Collection Set . . . . . . . . . . . . . . 185
4CP_IFU_16-XX-XX_USA_021
1Introduction
This chapter provides an overview of
the device, including indications for
use and contraindications. This
chapter also provides an
explanation of each chapter,
technical terms and symbols used in
this operating manual, and a
definition of the filtered water
required for use in the heater-cooler.
CP_IFU_16-XX-XX_USA_0215
Heater-Cooler System 3T • Introduction
Introduction
1.1Introduction
1.1.1 Device summary
The Heater-Cooler System 3T is an independent 3-circuit-heating/cooling system
which is suitable for continuous use.
The heater-cooler has three circuits that you control from the control display
panel:
● Cardioplegia circuit, whose temperature can be controlled from independent
heating and cooling tanks.
● Patient 1 circuit and patient 2 circuit, which can be used with the oxygenator
and/or the single-use heating/cooling blanket. The temperature for these two
circuits can be controlled from one heating/cooling tank.
The heater-cooler is a non-sterile device. No part of the heater-cooler is provided
sterile or can be sterilized.
Control panel
Cardioplegia circuit
Patient 1 circuit
C°C°C°
1
2
C°C°C°
pascals
40
20
60
0
80
TANK VACUUM
100
CALIBRATED FOR
VERTICAL POSITION
MAX
PRESS
URE
100
kPa
C°C°
°
C
1
C°
2
C°
°
C
Patient 2 circuit
6CP_IFU_16-XX-XX_USA_021
1.1.2 Indications for use
The Heater-Cooler System 3T is used to circulate water through heat exchangers to
warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours
or less.
1.1.3 Contraindications
There are no known contraindications for the Heater-Cooler System 3T. The
attending physician is solely responsible for the use of the system.
1.1.4 Legal disclaimer
LivaNova Deutschland GmbH will not assume any liability for any injuries and/or
damage to property caused by failure to observe the safety or operating
instructions or by failure to exercise due care. This also applies even if the duty to
exercise due care has not been expressly stated.
Heater-Cooler System 3T • Introduction
Introduction
1.1.5 About these operating instructions
● These operating instructions are intended for clinical and maintenance
personnel, and provide instructions for using, operating, and maintaining the
Heater-Cooler System 3T.
● To ensure the safety of patients and the operators, read these operating
instructions thoroughly before using the heater-cooler for the first time.
● These operating instructions provide valuable information. They describe
operational steps and also contain information on how to avoid dangerous
situations and errors. These operating instructions also provide
troubleshooting instructions.
CP_IFU_16-XX-XX_USA_021
7
Heater-Cooler System 3T • Introduction
Introduction
1.1.6 Symbols used in these operating instructions
WARNING
Indicates a hazardous situation that, if not avoided, could result in death or
serious injury.
CAUTION
Indicates a hazardous situation that, if not avoided, could result in minor or
moderate injury.
NOTICE
Indicates a hazardous situation that, if not avoided, may result in damage to the
device or other property.
Important
i
Indicates information considered important but not hazard related.
NOTE: Provides additional information that can help you better understand a
specific instruction of function.
1 Checkmark: Used to indicate an item to check before or after completing a task.
These items are for reference and do not replace detailed instructions.
8CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Introduction
Introduction
1.1.7 The chapters in these operating instructions
1Introduction
This chapter provides an overview of the device, including indications for use,
contraindications. This chapter also provides an explanation of each chapter,
technical terms and symbols used in this operating manual, and a definition of the
filtered water required for use in the heater-cooler.
2Safety
This chapter provides general safety information that you must follow to ensure
safe and effective use. Additional safety information, such as Warnings and
Cautions, are included throughout the instructions in this manual.
3System description
This chapter describes the system’s features and components, and the controls
that you will use to control the system’s circuits.
4Preparing the heater-cooler for a procedure
This chapter provides instructions for preparing the heater-cooler for use.
CP_IFU_16-XX-XX_USA_021
9
Heater-Cooler System 3T • Introduction
Introduction
5Using the heater-cooler during a procedure
This chapter provides instructions for positioning the heater-cooler in the OR and
using the heater-cooler’s controls during a procedure. This chapter also provides
the basic steps you will conduct after a procedure.
6Maintaining the heater-cooler
This chapter provides instructions for the routine maintenance that is part of the
operating conditions for the heater-cooler. This applies to the routine
maintenance performed by the responsible organization as defined in IEC 60601-1
or the equivalent national standard, as well as to the preventive maintenance and
safety checks performed by LivaNova representatives or competent service
personnel.
7Technical specifications
This chapter provides technical specifications including information about the
heater-cooler’s physical characteristics, labeling, part numbers, accessories, and
electromagnetic compatibility. This chapter also includes warranty information.
8Troubleshooting
This chapter provides troubleshooting for warnings and error codes you might see
on the control panel, and troubleshooting related to the aerosol collection set.
10CP_IFU_16-XX-XX_USA_021
1.1.8 Terminology and abbreviations
Terminology
Abbreviated termFull term
Heater-coolerStöckert Heater-Cooler System 3T
S5 SystemStöckert S5 System, modular heart-lung
machine
C5 SystemSorin Compact 5 System, compact heart-
lung machine
Aerosol collection set 3T Aerosol Collection Set
Abbreviations
Heater-Cooler System 3T • Introduction
Introduction
Abbreviated termFull term
ACSAerosol Collection Set 3T
ALAllowable Limit
CAN Controller Area Network (CAN bus)
3
ccCubic centimeter (cm
CFUColony Forming Unit
EMCElectromagnetic Compatibility
EPA Reg. No.Registration number of the United States
Environmental Protection Agency (applies
to the United States only)
HC3THeater-Cooler System 3T
HLMHeart-Lung Machine
LLiters
LPM Liters per minute
mLMilliliters
NTMNon-tuberculous Mycobacteria
)
OROperating Room
PaPascal
kPaKilopascal
ppmParts per million
RFRadio Frequency
UDIUnique Device Identifier
CP_IFU_16-XX-XX_USA_021
11
Heater-Cooler System 3T • Introduction
Introduction
1.1.9 Requirement for filtered tap water
Tap water used to fill the heater-cooler must be filtered using the following filter
and specifications:
Disposable Pall-Aquasafe water filter with an 0.2 μm membrane (Pall part
reference in the U.S.: “AQINA”; “AQIN” in other countries) or a filter of equivalent
performance that meets the requirements for bacterial retention of
Brevundimonas diminuta to ≥ 10
For instructions on filter management and disposal after the specified use period
please refer to the manufacturer’s instructions for use.
7
CFU/cm2 of effective filtration area.1
1 American Standard Test Method F838-15ae1 “Determining Bacterial Retention of Membrane
Filters Utilized for Liquid Filtration”
12CP_IFU_16-XX-XX_USA_021
2Safety
This chapter provides general safety
information that you must follow to
ensure safe and effective use.
Additional safety information, such
as Warnings and Cautions, are
included throughout the
instructions in this manual.
CP_IFU_16-XX-XX_USA_02113
Heater-Cooler System 3T • Safety
Safety information
2.1Safety information
2.1.1 General safety
● Read and understand the instructions for use before operating the heater-
cooler.
● Disinfect the external surfaces and water circuits per chapter 6 prior to first use.
● Do not place any objects (including the portable vacuum source) on top of the
heater-cooler.
● Maintain at least 70 cm clearance from walls and other devices while operating
the heater-cooler to prevent the heater-cooler from overheating.
● Do not operate the heater-cooler in the presence of explosive substances.
● Do not modify any of the mechanical or electrical systems of the heater-cooler.
● Do not remove the side or rear panels of the heater-cooler. Contact with the
internal components can cause serious injury or death.
● A list of approved accessory devices is contained in chapter 7. Do not use any
unapproved accessory devices with the heater-cooler.
● Do not use self-closing tubing connectors or connectors that incorporate a
valve. These connectors can reduce water flow to the external circuits.
● Drain heater-cooler water tanks outside the OR.
● Do not replace the aerosol collection canister in the OR environment.
● Avoid splashing heater-cooler water in the OR environment during use and
when making water circuit connections.
● Remove water spills from the OR floor using suitable disinfectants as
recommended by hospital procedures.
● Do not perform any maintenance activities while the heater-cooler is in use or
connected to an oxygenator or other heat exchanger device.
14CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Safety
Safety information
2.1.2 Information on diffusion of hydrogen peroxide
through the oxygenator heat exchanger
The water in the heater-cooler contains approximately 330 ppm of hydrogen
peroxide when prepared per these operating instructions. Contact the
manufacturer of the disposable oxygenator for information specific to the
hydrogen peroxide permeability of the heat exchanger for concentrations at or
below this level.
Do not use the heater-cooler with disposable oxygenators that have a heat
exchanger with a permeability rate that exceeds the daily allowable limit (AL) as
defined in the following table:
Allowable mass of hydrogen peroxide
) in mg/day the can be transferred
(H
202
in patient blood for:
Allowable limit (AL)
in mg/day per kg of
Duration of
exposure to H
Up to 6 hours0.2112.600.63
202
patient body
weight
An adult
(typical body
weight is 60 kg)
A neonate
(typical body
weight is 3 kg)
Over 6 hours0.0342.020.10
Although allowable limits (AL) for hydrogen peroxide diffusion are provided, actual
performance may vary in individual cases. Monitor and manage all patients.
CP_IFU_16-XX-XX_USA_021
15
Heater-Cooler System 3T • Safety
Safety information
2.1.3 Procedural safety
● The heater-cooler does not monitor the temperature of the patient’s blood or
the cardioplegia solution.
● Do not empty the external circuits until the patient has been discharged from
the OR.
● LivaNova recommends always having a replacement heater-cooler available in
case the heater-cooler in use is no longer functional (e.g., due to a total system
failure). The replacement heater-cooler must have compatible connectors.
2.1.4 Electrical safety
● Do not modify any of the electrical systems of the heater-cooler.
● The heater-cooler should only be connected to a hospital-quality power source
with an integrated protective conductor (ground).
● To avoid the risk of electric shock, the heater-cooler must only be connected to
a supply mains with protective earth.
● Check the condition of the power cable and plug regularly and do not use the
heater-cooler if the cable or plug shows signs of wear or damage.
● Do not position the heater-cooler where it will block access to the mains power
plug. Removing the mains plug from the receptacle isolates the heater-cooler
from mains voltage and the plug must remain accessible at all times.
2.1.5 Integrating the HC3T into your facility
As part of the integration of the Heater-Cooler System 3T into your facility,
LivaNova strongly recommends the following:
● Ensure that all operators and maintainers of the heater-cooler read the
instructions and familiarize themselves with the use and maintenance of the
device (chapters 4, 5 and 6), with special emphasis on the critical tasks listed
in the table below.
● Integrate all maintenance activities listed in chapter 6.2.1 into your facility’s
preventative maintenance scheduling system. This includes the cleaning and
disinfection of the heater-cooler before first use. Failure to perform the
maintenance tasks as indicated can lead to bacterial growth inside the device
and an increased risk of exposing the patient to harmful bacteria.
● Implement regular bacterial sampling of the water in the heater-cooler into your
preventative maintenance scheduling system. This testing is critical to
understanding if the disinfection and water preservation activities are being
completed effectively. This testing can also indicate if an external source of
contamination (e.g. a contaminated water hose) is present.
16CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Safety
Safety information
If you have any questions about the integration of the Heater-Cooler System 3T
into your facility, please contact your LivaNova sales representative.
Human Factors Testing and clinical experience have shown that operators are
most likely to have issues understanding or performing the critical tasks in the
following table. This table provides the task description, where to locate the
instructions related to this task, and the consequences of failing to perform the
task correctly.
TaskLocation of
instructions
Do not use the HC3T without
Chapter 4.5Failure to use the ACS may result in the
the ACS attached.
Cap unused ports on the ACS.Chapter 4.5
Task 3, 3
Write the installation date on
the ACS.
Replace the ACS after the
disinfection procedure.
Clean and disinfect the HC3T
Chapter 4.5
Task 4, 7
Chapter 6.4.2
Task 11
Chapter 6.2.1Bacteria may be present on or in the heater-
prior to first use.
Potential consequences of incorrect task
performance or omission
emission of aerosol.
Failure to cap the tandem and pour ports
may lead to a loss of vacuum in the ACS,
potentially reducing the effectiveness in
preventing the emission of aerosol.
Failure to write down the installation date
can lead to use of the ACS beyond 7 days,
potentially clogging the filter and reducing
the effectiveness in preventing the
emission of aerosol.
Failure to replace the ACS after the
disinfection procedure could lead to
disinfectant vapor entering the vacuum
system.
cooler if the packaging was damaged
during transport or storage.
Disinfect the connectors and
tubing fittings prior to each
connection.
Drain water tanks prior to
adding disinfectant.
During disinfection, fill the
water tanks to the Patient level
second green LED after adding
disinfectant.
Circulate disinfectant by
pressing the WARM CP,
Patient 1, and Patient 2
buttons.
Chapter 6.3.2
Chapter 6.3.3
Chapter 6.4.2
Task 2
Chapter 6.4.2
Task 4, 4
Chapter 6.4.2
Task 8, 1
Failure to disinfect the connectors and/or
fittings could lead to bacterial growth that
can be transferred to other parts of the
heater-cooler or the OR.
Failure to drain the tanks prior to adding
disinfectant could reduce the
disinfectant’s effectiveness.
Failure to completely fill the tanks during
disinfection could lead to a low level in one
tank, potentially resulting in incomplete
disinfection.
Failure to circulate disinfectant using
Patient 1, Patient 2, and the WARM CP
buttons could lead to incomplete
disinfection of one or more tanks.
CP_IFU_16-XX-XX_USA_021
17
Heater-Cooler System 3T • Safety
Safety information
TaskLocation of
instructions
Do not use disinfectant during
Chapter 6.4.1Use of disinfectant in the water during a
a procedure.
Mix chemicals (disinfectant or
hydrogen peroxide) by
connecting a tube between the
INLET of the CP circuit and the
INLET of the Patient 1 circuit.
Chapter 4.7
Task 3 & 4
Chapter 6.4.2
Task 6
Chapter 6.6
Task 4 & 5
Mix chemicals (disinfectant or
hydrogen peroxide) using the
COLD CP button.
Change the water in the heatercooler every 7 days.
Chapter 6.4.2
Task 6, 1 and
Task 6, 2
Chapter 6.2.1
Section 6.6
Potential consequences of incorrect task
performance or omission
procedure could lead to failure of the
heater-cooler and the attached accessory
devices.
Failure to properly connect the tubing prior
to mixing could lead to one or more tanks
not being exposed to the correct amount of
the chemical, potentially reducing the
effectiveness of the water preservation
process.
Failure to use the COLD CP button to mix
disinfectant could result in insufficient
exposure of one or more tanks to the
required quantity of disinfectant.
Failure to change the water can lead to
ineffective prevention of bacterial growth.
Add 150 ml of 3% medicalgrade hydrogen peroxide when
filling the heater-cooler.
Performing the monthly
bacterial monitoring
procedure.
Chapter 4.7
Task 2, 3
Chapter 6.6
Failure to add hydrogen peroxide when
filling the water tanks can lead to
ineffective prevention of bacterial growth.
Task 3, 3
Chapter 6.5.2Failure to properly sample the water in the
heater-cooler can lead to an inability to
identify potentially harmful levels of
bacterial growth in the device.
18CP_IFU_16-XX-XX_USA_021
3System description
This chapter describes the system’s
features and components, and the
controls that you will use to control
the system’s circuits.
CP_IFU_16-XX-XX_USA_02119
Heater-Cooler System 3T • System description
General description
3.1General description
3.1.1 Device summary
The Heater-Cooler System 3T is an independent 3-circuit-heating/cooling system
which is suitable for continuous use.
The heater-cooler has three circuits that you control from the control display
panel:
● Cardioplegia circuit, whose temperature can be controlled from independent
heating and cooling tanks.
● Patient 1 circuit and patient 2 circuit, which can be used with the oxygenator
and/or the single-use heating/cooling blanket. The temperature for these two
circuits can be controlled from one heating/cooling tank.
Control panel
Cardioplegia circuit
Patient 1 circuit
C°C°C°
1
2
C°C°C°
pascals
40
20
60
0
80
TANK VACUUM
100
CALIBRATED FOR
VERTICAL POSITION
MAX
PRESS
URE
100kPa
C°C°
°
C
1
C°
2
C°
°
C
Patient 2 circuit
20CP_IFU_16-XX-XX_USA_021
3.1.2 System components
Below are lists of standard components, mandatory components, and optional
components that can be used with the heater-cooler. To obtain additional
components, contact LivaNova Deutschland GmbH or your local LivaNova
distributor.
Standard components (included in delivery)
● Heater-Cooler System 3T
● Aerosol collection container holder
● 1/2" tubing connectors, straight (3 pieces)
● 1/2" tubing connectors, 90° angle (3 pieces)
● CAN cable for connection to the S5/C5 System
● Potential equalization cable
Heater-Cooler System 3T • System description
General description
● Screwdriver for zeroing the vacuum gauge
● Operating instructions
Mandatory components (not included in delivery)
● 3T Aerosol Collection Set
Optional components (not available from LivaNova)
● Tubing sets and/or adapters for connecting single-use heating/cooling
blankets
● Single-use heating/cooling blankets for adults (55 x 150 cm)
CP_IFU_16-XX-XX_USA_021
21
Heater-Cooler System 3T • System description
General description
3.1.3 Explanation of the heater-cooler circuits and
tanks
The heater-cooler has three circuits: the cardioplegia circuit, the
patient 1 circuit, and the patient 2 circuit.
See chapter 3.2.2 for more details on the control panel which you will use to
control each circuit.
The cardioplegia circuit uses two tanks
● There is one cardioplegia circuit, accessed from the inlet and outlet valves on
the back of the heater-cooler.
● This circuit can be switched between either the heating tank or the cooling tank
inside the heater-cooler. The heating and cooling tanks are separate so that
water can be kept at desired temperatures even when not being accessed:
Cold cardioplegia tank
Temperature range: 2 °C to 10 °C
Tank volume: 3 L
Warm cardioplegia tank
Temperature range: 15 °C to 41 °C
Tank volume: 3 L
Cardioplegia circuit
(inlet and outlet)
● The cardioplegia circuit can only be circulating from one of these tanks at a
time. For example, if the warm cardioplegia pump is running and you power on
the cold cardioplegia pump, the heater-cooler will circulate water from the cold
cardioplegia tank and stop circulating water from the warm cardioplegia tank.
22CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • System description
General description
The two patient circuits use one tank
● There are two patient circuits, accessed from the patient 1 and patient 2 inlet
and outlet valves on the back of the heater-cooler. These circuits can be used
with the oxygenator and/or the single-use heating/cooling blanket.
● The patient 1 and patient 2 circuits can be run at the same time or one at a time.
● Both patient circuits access the same tank (i.e., one tank) inside the heater-
cooler. As such, both patient circuits will always be set to the same
temperature.
Patient tank
Temperature range: 2°C to 41°C
Tank volume: 6 L
Patient 2 circuit
(inlet and outlet)
Patient 1 circuit
(inlet and outlet)
● The circuits for the patient and the circuit for cardioplegia can be switched off
separately to increase the heating and cooling performance of active circuit(s).
See chapter 5.6 for details.
CP_IFU_16-XX-XX_USA_021
23
Heater-Cooler System 3T • System description
General description
How the heater-cooler’s three tanks fill
As you pour water into the single filler neck, all three tanks will fill in the following
order:
1. Cold cardioplegia tank
2. Warm cardioplegia tank
3. Patient tank
Additional information about water tank levels:
● The cardioplegia tanks must fill completely before the patient tank begins to fill.
● After filling, and during operation, the system will equalize the water levels
between the two cardioplegia tanks. Therefore, running one pump (i.e., the
warm or cold pump) will reduce the water level in both tanks.
24CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • System description
3.2Structure of the heater-cooler
3.2.1 Heater-cooler overview
Structure of the heater-cooler
13
2
pascals
40
20
60
0
80
TANK VACUUM
100
CALIBRAT
ED FOR
VERTICAL POS
ITION
MAX
PRESS
URE
100
kPa
C°C°
°
C
C°
1
2
C°
°
C
1720
4
5
16
6
789 10
11
12
15
18
19
25
26
27
22
23
24
28
14
21
13
CP_IFU_16-XX-XX_USA_021
25
Heater-Cooler System 3T • System description
Structure of the heater-cooler
Items depicted in heater-cooler overview
ItemNameFunction
1Mains power switch● For powering the heater-cooler on/off
● Integrated automatic circuit breaker
2Control panelFor separated operation and configuration
of the three circuits
● Two patient circuits
● One cardioplegia circuit
3Vacuum gauge with scaleIndicates the vacuum level inside the tanks
4Vacuum gauge service portFor removing potential obstructions within
the internal pressure monitoring line
5Filler neck with capFor filling all tanks with filtered tap water
6CAN jack with coverFor connecting the heater-cooler to the S5/
C5 System
7Patient 2 circuit outlet connector with
with valve lever
13Patient tank drain valveFor emptying the tanks
14Cardioplegia tank drain valve
15Overflow outletFor draining excessive water from the tanks
and for connecting the aerosol collection
container
16FanFor ventilation of the heater-cooler
17Ventilation grill
18Potential equalization pointFor connecting the potential equalization
cable
19Power cablePower supply of the heater-cooler
20CastorsCastors with brakes
21BumperProtects the valves, drains, etc. against
accidental crashes
22Aerosol collection set (with tubing) For collecting aerosol
26CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • System description
Structure of the heater-cooler
Items depicted in heater-cooler overview (continued)
ItemNameFunction
23Connection line (short line)For connecting the aerosol collection
container to the heater-cooler
24Vacuum port (V)Connection point on the aerosol collection
container for the vacuum source line
25Vacuum source line (long line)For connecting the aerosol collection
container to the external vacuum source
26Patient port (P)Connection point on the aerosol collection
container for the connection line
27Tandem port (T)Unused (capped at all times)
28Pour spout (S)Port for emptying the aerosol collection
container (capped during normal use)
CP_IFU_16-XX-XX_USA_021
27
Heater-Cooler System 3T • System description
Structure of the heater-cooler
3.2.2 Heater-cooler control panel overview
The control panel manages all control and monitoring functions of the heatercooler. Specifically, the control panel:
● Is used to adjust the temperature set values
● Is used to start and stop the circuit pumps
● Displays all set and actual temperatures
● Displays water level in the three tanks
The control panel is divided into two sections:
Cardioplegia circuit controls, which include the warm tank and the cold tank.
Patient 1 and patient 2 circuit controls, which includes one tank for the two
patient circuits.
Cardioplegia circuit
Cold
Warm
Patient 1 circuit,
Patient 2 circuit
1
2
C°C°C°
C°C°C°
28CP_IFU_16-XX-XX_USA_021
Cardioplegia circuit controls
ColdWarm
18189
Heater-Cooler System 3T • System description
Structure of the heater-cooler
C°C°C°
2
3
544656710
1
2
C°C°C°
ItemNameFunction
1Actual temperature display Display the actual tank temperatures
2Cold cardioplegia circuit Stop/Stop
button
3Warm cardioplegia circuit Stop/Stop
button
For starting and stopping the circuit and for
switching from the warm to the cold tank
For starting and stopping the circuit and for
switching from the cold to warm tank
4Set temperature displayDisplays the tank set temperatures
5Set value down buttonFor decreasing the set temperature
6Set value up buttonFor increasing the set temperature
7Standby buttonFor switching the cardioplegia circuit on
and off
8High temperature indicatorIlluminates red if high tank temperature
detected
9Water level displayDisplays the water level in the tanks
10Pause audio button (global control)Pauses the audible tone for a maximum of
2 minutes. Pressing the button again
switches the audible tone back on.
CP_IFU_16-XX-XX_USA_021
29
Heater-Cooler System 3T • System description
Structure of the heater-cooler
Patient 1 and patient 2 circuit controls
189
C°C°C°
1
456710
2
C°C°C°
ItemNameFunction
1Actual temperature display Display the actual tank temperatures
2Patient 1 circuit Start/Stop buttonFor starting and stopping the patient 1
circuit
3Patient 2 circuit Start/Stop buttonFor starting and stopping the patient 2
circuit
4Set temperature displayDisplays the set temperatures
5Set value down buttonFor decreasing the set temperature
3
2
6Set value up buttonFor increasing the set temperature
7Standby buttonFor switching the patient 1 and patient 2
circuits on and off.
8High temperature indicatorIlluminates red if high tank temperature
detected
9Water level displayDisplays the water level in the tanks
10Pause audio button (global control)Pauses the audible tone for a maximum of
2 minutes. Pressing the button again
switches the audible tone back on.
30CP_IFU_16-XX-XX_USA_021
Explanation of the water level display
Heater-Cooler System 3T • System description
Structure of the heater-cooler
Water level displays
C°C°C°
1
DisplaySegment colorWater level meaning
Second greenMaximum capacity
First greenFull
OrangeSolid: Low level
Blinking: Low level warning. Refill
recommended.
RedMust be refilled. Pump(s) and heating/
cooling functions stop until refilled. Circuit will not operate.
2
CP_IFU_16-XX-XX_USA_021
31
Heater-Cooler System 3T • System description
Structure of the heater-cooler
Explanation of special Set value button operations
● For the cardioplegia Set values, pressing and holding a Set value button
changes the temperature value by 1°C. Continuing to hold the Set value button
will increase or decrease the set value more rapidly.
● For the patient set value, pressing and holding a Set value button initially
changes the temperature value by 0.1 °C, then changes the temperature value
by 1 °C. If you then release the Set value button and immediately press it once,
the temperature value will still change by 1 °C.
C°C°C°
Set value buttons
● You can increase the temperature value change increment by pressing and
holding one Set value button and pressing its counterpart (i.e., up or down)
once.
3.2.3 Control panel S5/C5 System
The CAN port allows the heater-cooler to be connected to an S5 or C5 HLM.
Operating the heater-cooler with the S5/C5 System does not require additional
remote control modules. You do not need to make additional connections or
perform additional steps besides connecting the CAN cable between the heatercooler and the S5/C5 System.
You may control the heater-cooler with the S5/C5 System menu or directly on the
base unit of the heater-cooler.
NOTE: Do not connect the heater cooler’s CAN port to any other device. The heatercooler is only compatible with the S5/C5 System.
Refer to the S5/C5 System’s operating instructions for information about the
heater-cooler’s displays when in use with the S5/C5 System.
Important
i
The length of water circuit tubing between the heater-cooler and accessory
devices must not exceed 5 m.
32CP_IFU_16-XX-XX_USA_021
4Preparing the
heater-cooler for a
procedure
Read the following chapter
thoroughly
heater-cooler for the first time.
before operating the
This chapter provides instructions for
preparing the heater-cooler for use.
See chapter 5 for detailed
instructions for positioning and
using the heater-cooler during a
procedure.
See chapter 6 for detailed
instructions for conducting
scheduled maintenance activities.
CP_IFU_16-XX-XX_USA_02133
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
General technical requirements
4.1General technical requirements
4.1.1 General use requirements
● Operate the heater-cooler only at ambient temperatures between 10 °C and
30 °C.
● Operate the heater-cooler in a spacious area. The heater-cooler can overheat if
operated in a confined space. Maintain a separation of at least 70 cm between
the ventilation grills and walls or other devices.
● Do not position the heater-cooler where it will block access to the mains power
plug. Removing the mains plug from the receptacle isolates the machine from
mains voltage and the plug must remain accessible at all times.
● The heater-cooler must be connected to a potential equalization point for the
safety of operators and patients.
● Electrical installations must abide by the requirements stated in
IEC 60364-7-710, or the corresponding equivalent local regulations.
● A central hospital vacuum source or a portable vacuum source must be
available for operating the 3T Aerosol Collection Set.
○ Operate the heater-cooler with an applied vacuum source. Aerosol will not be
collected from the water tanks without a vacuum source.
○ The vacuum source must be capable of a flow of 20 LPM or greater
(ISO 10079-1 for Medical Suction Equipment).
○ Portable vacuum sources must be specified for medical applications and
meet the requirements of IEC 60601-1 and IEC 60601-1-2.
○ The power supply for the vacuum source must be connected to a back-up
power generator to avoid power loss during use.
34CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
General technical requirements
4.1.2 Materials you will need (not provided with the
heater-cooler)
● Tubing sets compatible for connecting:
○ the oxygenator
○ the cardioplegia heat exchanger
○ a single-use heating/cooling blanket, if used
● 3T Aerosol Collection Set (part number 050900100)
Important
i
LivaNova recommends always having a replacement heater-cooler available in
case the heater-cooler in use is no longer functional (e.g., due to a total system
failure). The replacement heater-cooler must have compatible connectors.
CP_IFU_16-XX-XX_USA_021
35
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Conducting any required disinfection and maintenance
4.2Conducting any required disinfection
and maintenance
1
2
Before first-time use of the heater-cooler
Complete the following procedures per the maintenance schedule in chapter 6.2.1
“Schedule: Disinfection and maintenance”:
1 Disinfect the external surfaces
1 Disinfect the water circuits
Before each use of the heater-cooler, verify that device maintenance
is up to date
Verify that, if necessary, the following maintenance procedures have been
completed per the maintenance schedule in chapter 6.2.1 “Schedule: Disinfection
and maintenance”:
1 Disinfect the external surfaces
1 Monitor the hydrogen peroxide concentration
1 Replace the aerosol collection set
1 Change the water and add hydrogen peroxide
1 Disinfect the water circuits
1 Monitor the water quality for total bacteria count
1 Monitor the water quality for NTM
1 Replace the tubings used with the heater-cooler
36CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting the procedural tubing
4.3Connecting the procedural tubing
Important
i
● Follow your hospital’s policies regarding the use of personal protective
equipment (PPE).
● Before connecting any tubings to the heater-cooler:
○ Disinfect all heater-cooler connectors, tubing fittings, and tubing
connectors.
○ Disinfect all connectors and fittings on the tubings as well as the
connectors for the water circuits.
● Do not use self-closing connectors. They will reduce water flow.
● Check all tubings, connectors, and accessories to ensure all water
connections are properly sealed. Replace damaged components
immediately.
● Check that all tubings are free of bends and kinks.
● Close inactive circuits with short-circuit tubing.
CP_IFU_16-XX-XX_USA_021
37
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting the procedural tubing
1
Connect tubing to the water circuits
1 Before connecting any tubing to
the heater-cooler, disinfect the
following items:
○ all heater-cooler connectors
○ all tubing fittings and connectors
Important
i
Disinfect connectors and fittings every time you make a connection according to
chapter 6.3.2 and chapter 6.3.3.
Continue step on next page ▸
38CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Ci
Gentle t
Connecting the procedural tubing
2 Connect the procedural tubing to
the circuit inlet and outlet ports:
○ Push and hold the quick release
ring back.
○ Insert the tubing fitting into the
quick release ring.
○ Release the quick release ring
which will click back into place
and lock the tubing.
rcuit inlet and outlet ports
Quick release ring
Insert tubing fitting
Release to lock
3 Check that the connection is
secure by gently pulling on the
tubing fitting.
ug
CP_IFU_16-XX-XX_USA_021
39
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting the potential equalization cable and the power supply
4.4Connecting the potential equalization
cable and the power supply
Summary of steps you will complete in this section
1Connect the potential equalization cable
2Connect the power cable
1
Connect the potential equalization cable
1 Connect the potential equalization
cable to the heater-cooler’s
potential equalization point.
2 Connect the potential equalization
cable’s other end to the potential
equalization point at the OR's
central ground.
Potential
equalization point
40CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting the potential equalization cable and the power supply
2
Connect the power cable
1 Connect the AC power cable
to an appropriate outlet.
Power cable
WARNING
Do not connect the heater-cooler to a heart-lung machine’s auxiliary outlet. The
heater-cooler should be connected to a separate dedicated AC line.
To avoid the risk of electric shock, the heater-cooler must only be connected to
a supply mains with protective earth.
CP_IFU_16-XX-XX_USA_021
41
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting the aerosol collection set
4.5Connecting the aerosol collection set
WARNING
Operation of the heater-cooler without an applied vacuum source will stop the
collection of aerosol from the water tanks.
Summary of steps you will complete in this section
1Dispose of the aerosol collection set after 7 days
2Prepare the heater-cooler
3Prepare the aerosol collection canister
4Connect the aerosol collection set to the heater-cooler
42CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting the aerosol collection set
1
Dispose of the aerosol collection set after 7 days
If an aerosol collection set is already attached to the heater-cooler (e.g., from a
previous procedure), check the installation date written on the aerosol collection
canister lid:
● If the aerosol collection set is more
than 7 days old, bring the aerosol
collection set to the disposal area,
and dispose of in accordance with
your hospital’s policies. Be sure to
dispose of all components:
○ Canister
○ Canister lid
○ Heater-cooler connection line
○ Vacuum source line
○ Vacuum source extension line (if
applicable)
Installation date
● If the aerosol collection set is less
than 7 days old, you may empty the
canister according to chapter 5.7.2
and reconnect the emptied canister
following the applicable steps in this
chapter. Or, you may replace the
entire aerosol collection set.
CP_IFU_16-XX-XX_USA_021
43
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting the aerosol collection set
2
Prepare the heater-cooler
1 Make sure that the heater-cooler
is powered off.
2 Verify that the vacuum gauge
reads 0 Pa.
If the vacuum gauge does not read
0 Pa, use the provided screwdriver
to turn the adjustment screw near
the bottom of the gauge. Turn the
adjustment screw until the
indicator needle is set to 0 Pa.
Off
C°C°
C°
C°
C°C°
C°
C°
Vacuum = 0 Pa
pascals
40
20
0
TANK VACUUM
CALIBRATED FOR
VERTICAL POSITION
M
AX PRESSU
RE 1
°
C
1
2
°
C
°
C
1
2
°
C
60
80
100
00
kPa
3 Attach the aerosol collection
canister holder to the heatercooler by inserting the two top
tabs of the holder into the vent
slots on the heater-cooler. Release
the holder and allow it to rest in a
horizontal position.
NOTE: Make sure that the holder is
positioned such that the aerosol
collection canister does not touch
the floor or any part of the heatercooler.
44CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
P
t
Connecting the aerosol collection set
3
Prepare the aerosol collection canister
1 Gather a new aerosol collection
set, and remove the components
from the packaging:
○ Aerosol collection canister
○ Canister lid
○ Vacuum source line (with right-
angle connector)
○ Heater-cooler connection line
CAUTION
Prior to use, visually inspect the 3T Aerosol Collection Set to ensure that
connections, seals, and components were not damaged during shipment. Do
not use the 3T Aerosol Collection Set if any defects are found.
2 Place the lid on the aerosol
collection canister, and secure the
lid around the entire perimeter.
3 Apply the following caps over their
respective openings:
○ Pour spout
○ Tandem port
4 Place the aerosol collection
canister into the holder.
our spou
Tandem port
CP_IFU_16-XX-XX_USA_021
45
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
V
Right
r
V
Connecting the aerosol collection set
4
Connect the aerosol collection set to the heater-cooler
1 Connect the heater-cooler
connection line (the short line) to
the heater-cooler overflow outlet.
CAUTION
Do not connect the heater-cooler connection line to the vacuum service port or
the drain valve. This will limit the vacuum flow to the aerosol collection canister.
2 Connect the other end of the
heater-cooler connection line (the
short line) to aerosol collection
canister lid’s PATIENT port (P) .
Heater-cooler
overflow outlet
Heater-cooler
connection line
PATIENT port (P)
Important
i
Pushing the connection line too far onto port will make the line very difficult to
remove later.
3 Connect the vacuum source line’s
(the long line) right angle
connector to the aerosol collection
canister lid’s VACUUM port (V).
4 Check that the vacuum source line
acuum source line
is free of kinks.
ACUUM port
(V)
Continue step on next page ▸
angle connecto
46CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting the aerosol collection set
5 Connect the vacuum source line’s
(the long line) other end to the
external vacuum source.
NOTE: LivaNova recommends
connecting an overflow safety trap
to the external vacuum source.
This will avoid liquid introduction
into the external vacuum source’s
system.
NOTE: If additional length is
needed to access the external
vacuum source, attach a vacuum
extension line to the vacuum
source line. Order this part
separately (part number
050900111).
CAUTION
External vacuum source
Vacuum source line
You must connect the vacuum source line to a vacuum regulator. This will
control and adjust the flow to the aerosol collection canister.
Continue step on next page ▸
CP_IFU_16-XX-XX_USA_021
47
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
T
lid
Connecting the aerosol collection set
6 Check that all lid caps and
connections are secure:
1 Vacuum source line (a)
1 Heater-cooler connection line (b)
1 Tandem port cap (c)
1 Pour spout cap (d)
1 Vacuum source extension line (if
applicable)
7 Write the date of installation on
the lid’s label.
8 The aerosol collection set is now
ready for use.
op view of
a
c
b
d
48CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Connecting to the S5/C5 System (if applicable)
4.6Connecting to the S5/C5 System (if
applicable)
1
Connect the heater-cooler to the S5/C5 System
Connect the heater-cooler to the S5/C5 System according to the separate S5/C5
System’s operating instructions.
CP_IFU_16-XX-XX_USA_021
49
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Filling and mixing water tanks
4.7Filling and mixing water tanks
Important
i
● Follow your hospital’s policies regarding the use of personal protective
equipment (PPE).
● Do not use de-ionized or reverse osmosis processed water. These types of
water can cause deterioration of the refrigeration system.
Summary of steps you will complete in this section
1Prepare for filling
2Fill the water tanks
3Prepare for mixing
4Mix the tank contents
50CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
C
C
Filling and mixing water tanks
1
Prepare for filling
1 Make sure that you have access to filtered tap water that has been filtered
using the following filter and specifications:
Important
i
Disposable Pall-Aquasafe water filter with an 0.2 μm membrane (Pall part
reference in the U.S.: “AQINA”; “AQIN” in other countries) or a filter of
equivalent performance that meets the requirements for bacterial retention of
Brevundimonas diminuta to ≥ 10
(*American Standard Test Method F838-15ae1 “Determining Bacterial Retention of Membrane
Filters Utilized for Liquid Filtration”)
2 Gather the following materials:
○ 150 mL of medical grade 3% hydrogen peroxide solution
○ Measuring cup
3 Check to ensure that hydrogen peroxide solution is not expired.
7
CFU/cm2 of effective filtration area.*
4 If the water tanks have been
drained (i.e., are empty), the red
segment on both water level
displays will be illuminated.
NOTE: Confirm that the water level
displays function correctly after
performing a water change or
disinfection.
5 Make sure that the aerosol
collection canister is empty. Refer
to chapter 5.7.2 for instructions on
emptying the aerosol collection
canister.
1
Empty
C°
2
C°
Continue step on next page ▸
CP_IFU_16-XX-XX_USA_021
51
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
0
20
40
TANK VACUUM
CALIBRATED FOR
VERTICAL POSITION
M
AX P
RES
SURE
100
k
Pa
pascals
60
80
100
Filling and mixing water tanks
6 Make sure that both drain valves
are fully closed so that water does
not flow out when filling the water
tanks.
7 Press the mains power switch to
power on the heater-cooler.
NOTE: A tone will sound indicating
that the heater-cooler tanks are
empty. You can press the Pause
Audio button to pause the tone.
On
Pause audio
C°C°
°
C
1
2
C°C°
°
C
8 Turn the filler neck cap
counterclockwise to remove.
9 Apply four sprays of disinfectant
on the cap until all surfaces are
wet, including the cap’s underside
and sealing surface.
10Wait for 5 minutes. Ensure that all
surfaces remain wet with
disinfectant during this time.
Apply additional disinfectant if
necessary.
Wipe all surfaces with a
disinfectant wipe for 2 minutes to
remove excess disinfectant.
Wait for 5 minutes, then wipe
off excess for 2 minutes
52CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
filt
20
40
TANK VACUUM
CALIBRAT
ED FOR
VERTICAL POSITION
MAX
PRES
SUR
E
10
0kPa
pasc
als
60
80
100
100
Filling and mixing water tanks
2
Fill the water tanks
Important
i
As you pour water into the single filler neck, all three tanks will fill in the
following order:
● Cold cardioplegia tank
● Warm cardioplegia tank
● Patient tank
If the water tanks are not empty:
● Prepare a mixture of hydrogen
peroxide and filtered tap water at a
ratio of 1:91. For example: 10 mL of
hydrogen peroxide and 910 mL of
filtered tap water
● Fill the water tanks with the mixture
until the second green segments of
both circuit water level displays light
up.
1:91 mixture hydrogen peroxide to
tap water
5
37
7
C°
38
°
36.8
ered
pasc
al
s
40
20
60
0
80
TANK VACUUM
100
C
ALIBRA
TE
D F
V
E
OR
RT
I
C
AL
P
OSITION
M
AX
P
RE
SS
UR
E
10
0k
P
a
36.8
C°C°C°
1
2
C°C
● Continue to step 3, Prepare for
mixing.
If the water tanks
are empty, complete
the rest of step 2:
1 Begin filling the water tanks with
filtered tap water.
NOTE: The use of any other liquid is not
recommended.
Continue step on next page ▸
CP_IFU_16-XX-XX_USA_021
53
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
C
C
150 mL hyd
C
C
0
20
40
TANK VACUUM
CALIBRAT
ED FOR
V
ERTICAL POSITION
M
A
X PRESS
URE
1
00kP
a
pascals
60
80
100
Filling and mixing water tanks
2 Pause filling when the orange
segment on the patient circuit
water level display lights up.
3 Pour 150 mL of medical grade 3%
hydrogen peroxide solution into
the tank.
4 Continue filling the tanks with
filtered tap water until the second
green segment of the patient
circuit water level display lights
up.
C°
1
2
C°
rogen peroxide
5
37
7
C°
1
36.8
1
38
2
°
36.8
C°C
C°
2
C°
C°C°C°
Patient circuit
pas
c
al
s
40
20
60
0
80
TANK
100
VACUUM
C
ALIBRA
TE
D
V
F
E
OR
RT
I
C
AL
P
O
SITION
M
AX
PRE
SS
UR
E
10
0k
P
a
Patient circuit
5 Attach the filler neck cap, and turn
clockwise to secure.
54CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
Cl
Patient 1C
Filling and mixing water tanks
3
Prepare for mixing
1 On the back of the heater-cooler,
make sure that the three red valve
levers are closed. Turn clockwise
to close.
2 Disinfect the heater-cooler
connectors and short-circuit
tubing fittings.
Important
i
Disinfect connectors and fittings every time you make a connection according to
chapter 6.3.2 and chapter 6.3.3.
3 Connect short-circuit tubing
between the cardioplegia circuit
inlet and the patient 1 circuit inlet.
ose all valves
ardioplegia
4 To avoid triggering a high temperature warning while mixing:
Set the cold cardioplegia
circuit to 10 °C.
CP_IFU_16-XX-XX_USA_021
Set the warm cardioplegia
circuit to 20 °C.
1
Set the patient circuit to
20.0 °C.
C°C°C°
2
C°C°C°
55
Heater-Cooler System 3T • Preparing the heater-cooler for a procedure
P
t
Af
Filling and mixing water tanks
4
Mix the tank contents
1 Press the cold cardioplegia circuit
Start/Stop button to start mixing.
NOTE: The green LEDs will flash
alternately to indicate the circuit is
running.
CAUTION
Operate ONLY the cold cardioplegia circuit to ensure proper mixing.
2 After 5 minutes, press the cold
cardioplegia circuit Start/Stop
button again to stop mixing.
ress cold cardioplegiato star
ter 5 minutes, press to stop
C°C°
The mixing process is now
complete.
C°C°
3 Disconnect the short-circuit
tubing.
Tank filling is now complete.
56CP_IFU_16-XX-XX_USA_021
5Using the
heater-cooler
during a procedure
Read the following chapter
thoroughly
heater-cooler for the first time.
before operating the
This chapter provides instructions for
positioning the heater-cooler in the
OR and using the heater-cooler’s
controls during a procedure.
See chapter 4 for detailed
instructions for preparing the heatercooler for use.
This chapter also provides the basic
steps you will conduct after a
procedure.
See chapter 6 for detailed
instructions for conducting
scheduled maintenance activities.
CP_IFU_16-XX-XX_USA_02157
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Positioning the heater-cooler in the OR
5.1Positioning the heater-cooler in the OR
Important
i
● Do not cover the ventilation grills.
● Maintain at least 70 cm clearance from walls and other devices to prevent the
heater-cooler from overheating.
● The length of water circuit tubing between the heater-cooler and accessory
devices must not exceed 5 m.
Summary of steps you will complete in this section
1Bring the heater-cooler into the OR
1
2Place the heater-cooler in a safe position
3Connect a portable vacuum source, if needed
Bring the heater-cooler into the OR
1 If the heater-cooler is not already
in the OR, bring the heater-cooler
to the OR.
2 Place the heater-cooler on a level,
even floor.
Even floor
58CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Positioning the heater-cooler in the OR
2
Place the heater-cooler in a safe position
pascals
40
20
60
0
1 Place the heater-cooler at a
sufficient distance (at least 70 cm)
from walls and other devices.
C°C°
Make sure that the ventilation
grills are not covered.
2 Position the heater-cooler’s and portable vacuum source’s (if applicable)
exhaust flow away from the operating field and toward the exhaust vent
system.
Exhaust (on back)
80
TANK VACUUM
100
CAL
IBR
ATED FOR
VERTIC
AL POSIT
I
ON
M
A
X PR
ESS
URE 100kPa
°C°C°
C
1
2
°
C
Ventilation grills
WARNING
Do not position the heater-cooler’s and portable vacuum source’s (if applicable)
exhaust flow toward the operating field. Position the exhaust flow away from the
operating field and toward the exhaust vent system.
Continue step on next page ▸
CP_IFU_16-XX-XX_USA_021
59
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Positioning the heater-cooler in the OR
3 Step down on the brakes and lock the castors.
Lock brakes
pasca
l
s
40
20
60
0
80
TANK VACUUM
100
CAL
I
BR
ATED FOR
VERTIC
A
L POSIT
I
O
N
M
AX PR
ESSU
R
E 100
kP
a
C°C°
°
C
1
°
C
2
C°
°
C
3
Connect a portable vacuum source, if needed
If you are using a portable vacuum source, mount and operate the device
according to its instructions for use. Pay special attention to the instructions
related to use of the vacuum source in the OR.
60CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Connecting the procedural tubing to external devices
5.2Connecting the procedural tubing to
external devices
Summary of steps you will complete in this section
1Connect the procedural tubing to the external devices
2Check the procedural tubing and connections to all devices
1
Connect the procedural tubing to the external devices
1 Before making any connections,
disinfect the connectors on the
procedural tubing.
Important
i
● Disinfect the connectors and fittings every time you make a connection
according to chapter 6.3.2 and chapter 6.3.3.
● Do not apply disinfectant to the fittings on the external devices.
2 Connect the procedural tubing to the external devices.
2
CP_IFU_16-XX-XX_USA_021
Check the procedural tubing and connections to all devices
Inspect the heater-cooler procedural tubing and make sure:
● Tubing is free of kinks
● Tubing is not damaged
● Connections to devices in the OR are secure
61
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Connecting the procedural tubing to external devices
Overview of a typical circuit
Cardioplegic
solution
Heating/cooling blankets
(single-use)
Oxygenator
Oxygenator
heat exchanger
Patient
Patient
Cardioplegia
heat exchanger
Connection
patient circuit 1
Connection
patient circuit 2
Rear view
Heater-Cooler System 3T
Connection
cardioplegia circuit
62CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Powering on and checking the panel
5.3Powering on and checking the panel
WARNING
Operation of the heater-cooler without an applied vacuum source will stop the
collection of aerosol from the water tanks.
Summary of steps you will complete in this section
1Connect the vacuum source
2Power on the heater-cooler and observe the self-test
1
3Check the control panel to confirm the heater-cooler is ready for
use
Connect the vacuum source
1 Verify that the vacuum gauge
reads 0 Pa.
If the vacuum gauge does not read
0 Pa, use the provided screwdriver
to turn the adjustment screw near
the bottom of the gauge. Turn the
adjustment screw until the
indicator needle is set to 0 Pa.
Continue step on next page ▸
C°C°
C°
C°
Vacuum = 0 Pa
°
C
1
2
°
C
pascals
40
20
60
0
80
TANK VACUUM
100
CALIBRATED FOR
VERTICAL POSITION
M
AX PR
ESSU
RE 100
kPa
CP_IFU_16-XX-XX_USA_021
63
Heater-Cooler System 3T • Using the heater-cooler during a procedure
N
ld
Powering on and checking the panel
2 Check the installation date on the
aerosol collection canister.
If more than 7 days have passed
since the installation date,
dispose of the aerosol collection
canister and install a new one
according to chapter 4.5.
3 Power on the external vacuum
source and set the vacuum
regulator to full open flow.
o more than 7 days o
Full open flow
4 Make sure that the heater-cooler’s
vacuum gauge reads greater than
50 Pa.
NOTE: With full open vacuum flow,
the heater-cooler’s vacuum gauge
might read greater than 100 Pa.
The vacuum gauge can accept full
vacuum flow.
Vacuum > 50 Pa
C°C°
°
C
1
2
C°C°
°
C
pascals
40
20
60
0
80
TANK VACUUM
100
CALIBRATED FOR
VERTIC
AL POSITION
M
AX PR
ESSURE 100
kPa
64CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Powering on and checking the panel
2
Power on the heater-cooler and observe the self-test
1 Press the mains power switch to
power on the heater-cooler.
On
C°C°
C°
2 The heater-cooler will run a self-test while powering up. All of the control
panel’s LEDs and 7-segment displays will illuminate for 2 seconds, and a
tone will sound.
Check that all LEDs and 7-segment displays are properly illuminated before
use.
1
C°
°
C
2
°
C
3 After the control panel lights up,
the device’s software version
number (EPROM) will display for
1 second in the patient circuit
display.
4 The heater-cooler is now finished
powering on and will display the
last set temperature values. The
heater-cooler is now actively
controlling the tank temperatures,
but the circuit pumps are not
running.
C°C°C°
1
2
C°C°C°
1
2
C°
Version number appears in patient circuit
display
CP_IFU_16-XX-XX_USA_021
65
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Powering on and checking the panel
3
Check the control panel to confirm the heater-cooler is ready for use
Important
i
Do not use the heater-cooler if the elements on the control panel are flashing. If
the control panel is flashing, the heater-cooler is in service mode. Power off the
heater-cooler and power it on again to exit service mode.
1 Once the self-test is complete, confirm the following:
○ The green LEDs of the Standby buttons are not illuminated (i.e., the circuits
are not in standby mode).
○ The control panel displays the last set temperature values.
○ Both circuit water level displays are illuminated to the second green
segment.
2 The heater-cooler is ready for use.
C°C°C°
1
2
C°C°C°
66CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
If errors EE(E) appear on control panel
If a circuit’s set point display flashes
between “EE(E)” and a number, this is
an error code:
● Error code E08, E19, or E23 might
display during priming, but should
clear when priming completes (all air
is removed from the circuit).
● Do not use a circuit if an error code
remains after priming. Refer to
chapter 8.1.4 for error code
definitions, and notify your
authorized service technician as soon
as possible.
NOTE: You may use the unaffected
circuits even if one circuit displays an
error code.
Powering on and checking the panel
Flashes between error and code number
CP_IFU_16-XX-XX_USA_021
67
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Performing a functional check prior to operation
5.4Performing a functional check prior to
operation
Important
i
● Perform the steps outlined in this chapter before every time you use the
heater-cooler.
● Address any error codes immediately. Depending on the error code, you
might not be able to use a particular circuit, or you might not be able to use
the heater-cooler at all. Refer to chapter 8.1.4 for error code descriptions.
● The heater-cooler is ready for operation after you have performed the
functional check.
Summary of steps you will complete in this section
1Check the setup
2Check when first switching power on
68CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Performing a functional check prior to operation
1
Check the setup
1 Confirm the heater-cooler’s and portable vacuum source’s (if applicable)
exhaust is pointed away from the operating field.
WARNING
Do not position the heater-cooler’s and portable vacuum source’s (if applicable)
exhaust flow directed toward the operating field. Position the exhaust flow away
from the operating field and toward the exhaust vent system.
1 Confirm that the heater-cooler is positioned on a level, even floor.
1 Confirm that the castor brakes are locked.
1 Confirm that the power cable and potential equalization cable have been
connected according to applicable regulations.
1 Confirm that the mains plug is completely dry.
1 Confirm that the power cable is free of damage.
1 Confirm that all tubing connectors are secure and locked.
1 Confirm that all procedural tubings are free of damage and leaks.
1 Confirm that the aerosol collection set is less than 7 days old.
1 Confirm that the aerosol collection canister is holding less than 0.5 L of fluid.
1 Confirm that the heater-cooler’s vacuum gauge reads greater than 50 Pa with
full open vacuum flow applied.
1 Confirm that a replacement heater-cooler with compatible connectors is
available.
1 Confirm that the CAN cable is properly connected (if you are operating the
heater-cooler with a S5/C5 System).
CP_IFU_16-XX-XX_USA_021
69
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Performing a functional check prior to operation
2
Check when first switching power on
1 Confirm that all LEDs and temperature displays on the control panel are
illuminated.
1 Confirm that a tone sounded.
1 Confirm that the control panel is reflective of a functioning heater-cooler (i.e.,
the heater-cooler is not in service mode and does not display errors). If the
heater-cooler displays an error, refer to chapter 8.1.4 for error code
descriptions.
1 Confirm that the water level indicators on the patient and cardioplegia circuit
displays are not blinking orange or solid red.
70CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Up butt
5.5Priming the complete circuit
Summary of steps you will complete in this section
1Adjust the set values as needed
2Place unused circuits in standby mode
3Prime the external circuits
Priming the complete circuit
1
Adjust the set values as needed
Important
i
The heater-cooler will begin adjusting temperatures to the set values 2 seconds
after the last Set value button press.
To adjust the set values:
○ Press the Set value up button to
raise the circuit temperature.
C°C°C°
○ Press the Set value down button to
lower the circuit temperature.
Down buttons
ons
2
CP_IFU_16-XX-XX_USA_021
Place unused circuits in standby mode
If not being used, press the
circuit’s Standby button to place
the circuit on standby.
NOTE: The other circuit will
continue to function and the
heating/cooling performance will
be increased.
Standby buttons
71
Heater-Cooler System 3T • Using the heater-cooler during a procedure
O
U
Priming the complete circuit
3
Prime the external circuits
Important
i
Prime the circuits’ tubings and heat exchanger before using the heater-cooler.
● When you power on a circuit for the first time, an error code might appear
indicating that air is present in the tubings.
● Do not use the heater-cooler until you have resolved the errors by completely
priming the external circuits.
1 On the back of the heater-cooler,
for any circuits with tubing
connected, turn the red levers
counterclockwise to open the
valves.
pen valves connected to tubing
2 Prime the cardioplegia circuit
using the cold cardioplegia tank.
NOTE: Do not use the warm
cardioplegia tank to prime. This
will trigger an error code and stop
the circuit pump.
Continue step on next page ▸
se cold cardioplegia circuit to prime
C°C°
72CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Pri
A
Priming the complete circuit
3 Prime the patient circuit(s) using
the associated pump(s).
NOTE: Do not run a pump if there
is no tubing connected.
4 Depending on the volume of the
circuit tubing, you might need to
refill the water tanks after priming.
Check the water level displays:
○ Make sure the first or second
green segments on both water
level displays are illuminated.
○ If not, refill the water tanks before
use according to chapter 4.7, step
1 and step 2.
me patient 1 and patient 2 circuits
C°
1
cceptable volume
Cardioplegia or patient circuits
Unacceptable volume
Cardioplegia or patient circuits
2
CP_IFU_16-XX-XX_USA_021
73
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Using the device controls during a procedure
5.6Using the device controls during a
procedure
5.6.1 Checking the vacuum
Checking the vacuum
Adjust the external vacuum source to full
open vacuum flow to check that the
Vacuum > 50 Pa
heater-cooler operates at vacuum levels
pascals
40
greater than 50 Pa.
C°C°
°
C
1
2
C°C°
°
C
20
60
0
80
TANK VACUUM
100
CALIBRATED FOR
VERTIC
AL POSITION
M
AX PRESSU
RE 1
00
kPa
WARNING
Operating the heater-cooler with the vacuum level indicated on the tank vacuum
gauge in the red solid zone limits the aerosol collection set’s ability to collect
aerosol from the water tanks. Refer to chapter 8.2.1 for possible remedies.
74CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
O
Cl
Using the device controls during a procedure
5.6.2 Opening and closing the circuit valves
Important
i
● To fill the tubing and to run water through the circuits, the valves must be
open.
● To drain tubing, the valves must be closed.
Opening a valve
On the back of the heater-cooler, for any
circuits with tubing connected, turn the
red levers counterclockwise to open the
valves.
Closing a valve
On the back of the heater-cooler, turn
the red valve lever clockwise to close the
valve.
pen valves connected to tubing
ose valves
CP_IFU_16-XX-XX_USA_021
75
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Up b
Using the device controls during a procedure
5.6.3 Using the circuit controls
Adjusting the set values
Important
i
The heater-cooler will begin adjusting temperatures to the set values 2 seconds
after the last Set value button press.
To adjust the set values:
● Press the Set value up button to raise
the circuit temperature.
● Press the Set value down button to
C°C°C°
uttons
lower the circuit temperature.
Down buttons
Special Set value button operations:
Cardioplegia: Pressing and holding a Set value button changes the temperature
value by 1 °C. Continuing to hold the Set value button will increase or decrease
the set value more rapidly.
Patient: Pressing and holding a Set value button initially changes the temperature
value by 0.1 °C, then changes the temperature value by 1 °C. If you then release
the Set value button and immediately press it once, the temperature value will still
change by 1 °C.
You can increase the temperature value change increment by pressing and holding
one Set value button and pressing its counterpart (i.e., up or down) once.
76CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Operating the cardioplegia circuit
Using the device controls during a procedure
How the cardioplegia circuit works:
The cardioplegia circuit has two tanks: a
warm tank and a cold tank. The warm
and cold cardioplegia tanks operate
using a single circuit. Only one
cardioplegia circuit (warm or cold) can
run at a time.
You can control each tank’s temperature
independently so that warm and cold
water sources are available on demand
during a procedure.
The temperature range for the cold
cardioplegia tank is: 2 °C to 10 °C.
The temperature range for the warm
cardioplegia tank is: 15 °C to 41 °C.
Starting the cardioplegia circuit:
Press the Start/Stop button of your
desired cardioplegia circuit to run the
circuit.
One circuit
Cold cardioplegiaWarm
cardioplegia
C°C°
C°C°
Cold cardioplegiaWarm
cardioplegia
The water from the selected cardioplegia
circuit’s tank (warm or cold) will fill the
cardioplegia circuit.
NOTE: The green LEDs will flash
alternately to indicate the circuit is
running.
ON
LEDs flash
ON
LEDs flash
CP_IFU_16-XX-XX_USA_021
77
Heater-Cooler System 3T • Using the heater-cooler during a procedure
S
ill
Using the device controls during a procedure
Switching between cardioplegia tanks
(warm or cold):
Press the Start/Stop button of the
inactive cardioplegia tank to run the
circuit.
Within 5 seconds, the active
cardioplegia circuit will stop running,
and the water from the selected
cardioplegia circuit’s tank (warm or
cold) will fill the cardioplegia circuit.
NOTE: The green LEDs will flash
alternately to indicate the circuit is
running.
Stopping the cardioplegia circuit:
Press the Start/Stop button of the active
cardioplegia circuit to stop the circuit.
The cardioplegia circuit will stop
running.
NOTE: The green LEDs will go out when
the cardioplegia circuit stops running.
tart one cardioplegia circuit and the other w
stop
C°C°
Cold cardioplegiaWarm
cardioplegia
OFF
LEDs off
OFF
LEDs off
Important
i
Put the cardioplegia circuit in standby when it is
not in use. Leaving the cardioplegia circuit active
Cardioplegia
Standby button
reduces the patient circuits’ heating/cooling
performance.
78CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Pati
Pati
2
Operating the patient circuits
Using the device controls during a procedure
How the patient circuits work:
The patient 1 and patient 2 circuits
operate using separate circuit valves
and pumps. Both patient circuits can run
at the same time.
The patient 1 and patient 2 circuits use
the same water tank. The temperature of
both patient circuits will always be the
same.
The temperature range for the patient
tank is 2 °C to 41 °C.
Starting the patient circuit(s):
Press the Start/Stop button of your
desired patient circuit to run the circuit.
NOTE: The green LEDs on the patient(s)
circuit(s) will flash alternately to
indicate the circuit is running.
ent 1
1
Patient 1Patient 2
1
ON
LEDs flash
ent
C°
2
C°
2
ON
LEDs flash
Stopping the patient circuit(s):
Press the Start/Stop button of the active
patient circuit(s) to stop the circuit(s).
NOTE: The green LEDs will go out when
the patient(s) circuit(s) stops running.
Important
i
Put the patient circuits in standby when they are
not in use. Leaving the patient circuits active
reduces the cardioplegia circuits’ heating/cooling
performance.
Patient 1Patient 2
1
OFF
LEDs off
Patient Standby
button
2
OFF
LEDs off
CP_IFU_16-XX-XX_USA_021
79
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Completing a procedure
5.7Completing a procedure
WARNING
Do not empty the external circuits until the patient has been discharged from
the OR environment to reduce the potential of exposure to aerosol.
5.7.1 Drain and disconnect tubing
Summary of steps you will complete in this section
1Confirm the patient is not in the room
2Prepare the heater-cooler for draining
3Drain the external circuits
4Disconnect the tubing from the external devices
5Power off the heater-cooler
80CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Completing a procedure
1
Confirm the patient is not in the room
Confirm that the patient has been discharged from the OR before
conducting any further steps.
Patient must NOT be in
the room
CP_IFU_16-XX-XX_USA_021
81
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Completing a procedure
2
Prepare the heater-cooler for draining
1 Make sure that the heater-cooler
is powered on when draining. If
needed, press the mains power
switch to power on the heatercooler.
2 Make sure that the aerosol
collection set is connected and
the vacuum source is connected
and set to maximum.
On
Vacuum source
connected
C°C°
°
C
C°
1
2
C°
°
C
WARNING
Do not shut off the applied vacuum source until you have drained the circuits
and powered off the heater-cooler. Operate the heater-cooler with an applied
vacuum source to ensure that aerosol continues to be collected from the water
tanks.
Important
i
If you are draining longer tubes with high volumes, there is a potential for
overflowing the aerosol container. If the volume in the container exceeds 0.5 L,
pause the draining of the circuit and empty the aerosol container according to
chapter 5.7.2, then continue draining.
Continue step on next page ▸
82CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
E
t/St
3 To prevent excess overflow, drain
water from the patient circuit tank
until the orange segment on the
patient circuit water level display
flashes.
Close the drain valve after draining
this amount from the tank.
Drain patient circuit
until orange segment
displays
C°
1
2
C°
Completing a procedure
3
Drain the external circuits
WARNING
Do not close all valves at once when draining water circuits. Always close one
circuit valve at a time. Failure to close the external water circuit valves in the
order specified below may result in increased levels of aerosol emission.
1 Make sure that the circuit pumps
are running on any circuits that are
connected. Press the Start/Stop
buttons if needed.
NOTE: The green LEDs will flash
alternately to indicate the circuit is
running.
Continue step on next page ▸
xample of Star
Warm
cardioplegia
ON
LEDs flash
op buttons:
Patient 1
1
ON
LEDs flash
CP_IFU_16-XX-XX_USA_021
83
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Cl
Cl
Cl
E
t/St
Completing a procedure
2 With either cardioplegia pump
running, turn the cardioplegia circuit’s red valve lever clockwise
to close the valve.
Allow the external water circuit
volume to enter the tank before
closing the next valve.
3 Turn the patient 1 circuit’s red
valve lever clockwise to close the
valve.
Allow the external water circuit
volume to enter the tank before
closing the next valve.
ose cardioplegia valve, wait to drain
ose patient 1 valve, wait to drain
4 If in use, turn the patient 2
circuit’s red valve lever clockwise
to close the valve.
The water from the external circuit
will return to the patient circuit
tank.
5 Press the Start/Stop button of the
active circuit(s) to stop the
circuit(s).
NOTE: The green LEDs will go out
when the circuit(s) stops running.
ose patient 2 valve, wait to drain
xample of Star
Warm
cardioplegia
OFF
LEDs off
op buttons:
Patient 1
1
OFF
LEDs off
84CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Completing a procedure
4
Disconnect the tubing from the external devices
1 Make sure that the circuits are
drained before disconnecting any
tubing.
2 Disconnect all procedural tubing
from external devices.
3 Disconnect the procedural tubing
from the heater-cooler
OR
Disinfect the tubing connectors
and suitable short-circuit
connectors (part number
73-300-160). Then, short-circuit
the procedural tubing.
Important
i
Disinfect the connectors and fittings every time you make a connection
according to chapter 6.3.2 and chapter 6.3.3.
Connect tubing
5
Power off the heater-cooler
Press the mains power switch to
power off the heater-cooler. The
vacuum source may now be shut
off and/or disconnected.
Off
C°C°
°
C
C°
1
2
C°
°
C
CP_IFU_16-XX-XX_USA_021
85
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Completing a procedure
5.7.2 Emptying and/or disposing of the aerosol
collection set
Summary of steps you will complete in this section
1Empty the aerosol collection canister
2Dispose of and replace, or reconnect the aerosol collection set
based on the date it was installed
1
Empty the aerosol collection canister
WARNING
The aerosol collection canister contents are potentially hazardous due to
collection of bacteria. Use appropriate Personal Protection Equipment (PPE),
and handle the contents accordingly.
CAUTION
Do not expose the aerosol collection canister’s hydrophobic lid filter to the
water inside the aerosol collection canister. This will seal the filter, which will
stop vacuum flow and limit the aerosol collection canister’s ability to collect
aerosol from the heater-cooler’s water tanks.
If the hydrophobic lid filter has been exposed to water, replace the aerosol
collection set according to the steps in this chapter.
1 Make sure that the heater-cooler
is powered OFF. If needed, press
the mains power switch to power
Off
off the heater-cooler.
C°C°
°
C
C°
1
2
C°
°
C
Continue step on next page ▸
86CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
V
H
DO NOT lift by lid
Completing a procedure
CAUTION
Do not disconnect the vacuum source if the heater-cooler is powered on.
2 Disconnect the vacuum source
line (the long line) from the
vacuum regulator.
Disconnect
vacuum
3 Disconnect the heater-cooler
connection line and vacuum
source line from the canister lid.
4 If you are disposing of the aerosol
collection set (instead of only
emptying), also disconnect the
heater-cooler connection line from
the back of the heater-cooler.
5 Lifting from the canister body,
carefully remove the aerosol
collection canister from the
holder.
NOTE: Do not lift the aerosol
collection canister by the lid, as
the weight of the contents might
cause the lid to separate from the
aerosol collection canister.
eater-cooler connection
acuum source line
.
Lift canister body
Continue step on next page ▸
CP_IFU_16-XX-XX_USA_021
87
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Completing a procedure
6 Uncap the pour spout on the
Uncap pour spot
aerosol collection canister.
CAUTION
Do not expose hydrophobic filter in Canister Lid to water when emptying the
canister. Exposure to water will block filter and cause it to not function.
7 Empty the aerosol collection
canister in accordance with your
hospital’s policies.
NOTE: Empty the aerosol
collection canister outside of the
OR, if possible.
8 Recap the pour spout on the
aerosol collection canister.
DD/MM/YYYY
Recap pour spot
Empty
88CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Completing a procedure
2
Dispose of and replace, or reconnect the aerosol collection set based
on the date it was installed
1 Check the installation date written
on the aerosol collection canister
lid:
○ If the aerosol collection set is less
than 7 days old, you may
reconnect the emptied canister
using the same heater-cooler
connection line and vacuum
source line. Or, you may replace
the entire aerosol collection set.
○ If the aerosol collection set is
more than 7 days old, replace the
entire aerosol collection set
including all its components and
lines.
2 Follow the applicable steps in
chapter 4.5 to connect a new or
emptied aerosol collection set to
the heater-cooler.
Installation date
CP_IFU_16-XX-XX_USA_021
89
Heater-Cooler System 3T • Using the heater-cooler during a procedure
Completing a procedure
5.7.3 Conducting any required maintenance according
to schedule
1
2
Clean and disinfect the heater-cooler surfaces after every use
Clean and disinfect the heater-
cooler surfaces after every use
according to the instructions in
chapter 6.3.
C°C°
°
C
C°
pascals
40
20
60
0
80
TANK VACUUM
100
CALI
BRATED FOR
VER
TIC
AL P
O
SI
TIO
N
M
AX PRE
SSU
RE 1
00
k
Pa
°
C
1
2
°
C
Review the schedule for disinfection and maintenance
Review the schedule for disinfection and maintenance in chapter 6.2.1 to
determine if any additional maintenance, cleaning, and/or disinfection tasks
are required.
90CP_IFU_16-XX-XX_USA_021
6Maintaining the
heater-cooler
Regular maintenance is an important
factor for the operation of the
heater-cooler as it results in
operational safety, reliability and
increased service life.
This chapter provides instructions for
the routine maintenance that is part
of the operating conditions for the
heater-cooler. This applies to the
routine maintenance performed by
the responsible organization as
defined in IEC 60601-1 or the
equivalent national standard, as well
as to the preventive maintenance
and safety checks performed by
LivaNova representatives or
competent service personnel.
CP_IFU_16-XX-XX_USA_02191
Heater-Cooler System 3T • Maintaining the heater-cooler
General maintenance precautions
6.1General maintenance precautions
6.1.1 Safety instructions for routine maintenance
WARNING
It is important to follow the cleaning and disinfection process described in
these operating instructions to help reduce the risk of contamination.
Personnel and preparation:
● Only individuals trained to use the heater-cooler should perform
maintenance on the heater-cooler.
● Only service technicians authorized by LivaNova may repair the heater-cooler
system.
● Only refrigerant systems experts are authorized to repair the cooling circuit.
● Do not service the internal components of the heater-cooler. Only LivaNova
representatives — or other service personnel deemed qualified by LivaNova —
are authorized to maintain, service, and repair internal components of the
heater-cooler.
Materials and protective equipment:
● When performing any maintenance activities on the heater-cooler, follow
your hospital’s personal protection equipment (PPE) policy to protect against
exposure to hazardous materials, such as bacteria and/or disinfectant
chemicals.
92CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Maintaining the heater-cooler
General maintenance precautions
CAUTION
Materials and protective equipment:
● Use only cleaning agents and disinfectants that are approved for use in the
concentrations specified in chapter 6.3 and 6.4.1. Deviating from the
labeling could cause damage to the heater-cooler.
Timing:
● Conduct regular maintenance of the heater-cooler. Doing so will result in
operational safety, reliability, and increased service life.
● Perform all cleaning and disinfecting procedures according to the
maintenance schedule provided in chapter 6.2.
● Disinfect the heater-cooler before using it for the first time. Disinfect the
heater-cooler before storage.
NOTICE
● Adhere to local regulations related to documenting all maintenance
procedures, as required by European Directive 93/42, and any operating
failures in a medical device log. The medical device log should be presented
to authorized service personnel upon request.
NOTE: LivaNova recommends documenting this information regardless of local
regulations. Providing this information to LivaNova will assist with maintenance
and support activities.
● Exposing the heater-cooler to disinfectant for longer than its intended
exposure time will increase the wear on the device and might lead to
premature failure of the heater-cooler.
● All ventilation grills must be cleaned at regular intervals to avoid
contamination. Detailed information on this subject is provided in
chapter 6.8.
CP_IFU_16-XX-XX_USA_021
93
Heater-Cooler System 3T • Maintaining the heater-cooler
Maintenance schedule and checklists
6.2Maintenance schedule and checklists
This chapter includes a schedule of maintenance tasks and several checklists you
should use to make sure that the heater-cooler is properly maintained,
disinfected, and functioning.
6.2.1 Schedule: Disinfection and maintenance
TimingTaskChapter
Before first-time use,
before storing
Upon every connectionDisinfect connections and fittings6.3.3
After every useClean and disinfect external surfaces6.3.1
Daily, before use
Daily, when in storage
Every 7 daysReplace the aerosol collection set5.7.2
Every 7 daysChange the water and add hydrogen peroxide6.6
Every 14 days (including
when in storage)
Once a month● Monitor the water quality for total bacteria count
Once a yearReplace the tubings used with the heater-cooler6.9
Clean and disinfect external surfaces, connectors,
and fittings
Disinfect the water circuits6.4
Monitor the hydrogen peroxide concentration6.5.1
Disinfect the water circuits6.4
● Monitor the water quality for NTM
NOTE: Conduct before and after disinfecting the
water circuits
6.3
6.5.2
Once a year or 1,000 hours
(which ever comes first)
94CP_IFU_16-XX-XX_USA_021
Schedule a maintenance check after every 1,000
operating hours or at least once every 12 months,
whichever comes first.
6.2.4
Heater-Cooler System 3T • Maintaining the heater-cooler
Maintenance schedule and checklists
6.2.2 Checklist: Daily or with every use
1
Perform these checks daily or every time you use the heater-cooler
1 Check the maintenance schedule in 6.2.1 to determine whether any
maintenance activities are due. Perform the required maintenance activities
according to their corresponding instructions.
1 Make sure that the daily monitoring of the hydrogen peroxide concentration has
been completed per 6.5.1.
1 Dispose of used single-use items according to local environmental regulations
and the manufacturer’s instructions.
1 Check the installation date on the aerosol collection canister.
○ If more than 7 days have passed, replace the aerosol collection canister.
Refer to chapter 5.7.2 for instructions on replacing the aerosol collection set.
○ If 7 days or fewer have passed, check the fill level on the aerosol collection
set and empty it if needed, according to instructions in chapter 5.7.2.
1 Contact LivaNova if the heater-cooler or any of its components are defective or
damaged.
CP_IFU_16-XX-XX_USA_021
95
Heater-Cooler System 3T • Maintaining the heater-cooler
Maintenance schedule and checklists
6.2.3 Checklist: Mechanical checks
● Perform regular checks to ensure that the heater-cooler is in proper working
condition, and to reduce or eliminate the chance of malfunction.
● Do not use the heater-cooler if the heater-cooler does not meet all the
conditions listed in this chapter.
● Contact LivaNova if the heater-cooler or any of its components are damaged or
require service.
1
2
Check the heater-cooler’s main power, potential equalization, and
CAN cable
Make sure:
1 The cables are free of damage (e.g., cracks, cuts, kinks) along the entire length
of the cable
1 All contacts are clean
1 The plugs are free of mechanical damage (e.g., no damage to housing and no
bent pins)
1 The mains power connection to the heater-cooler is secure with no movement
between cable and strain relief
Check the water circuits
Make sure:
1 Water tubing is free of damage and leaks
1 Water circuit connections are free of damage and leaks
1 Water circuit connectors lock and unlock smoothly
96CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Maintaining the heater-cooler
Maintenance schedule and checklists
6.2.4 Checklist: Regular maintenance from LivaNova
You/your organization is responsible for scheduling maintenance checks from a
LivaNova service technician.
Only LivaNova representatives — or other service personnel deemed qualified by
LivaNova — are authorized to perform maintenance checks of the heater-cooler.
Maintenance checks by LivaNova service technicians include, but are not limited
to, the following:
● Status checks
● Electrical safety checks
● Operational safety checks
● Status of the water circuits/tanks for calcification and biofilm
● Status of cooling coils for corrosion
● Decalcification and/or disinfection
Important
i
As a preventive measure, contact LivaNova service personnel to schedule a
maintenance check of the water tank cooling coils if you observe either of the
following:
● Dark gray liquid is observed during the water circuit disinfection process.
● Excessive hydrogen peroxide consumption is observed during the daily
monitoring of hydrogen peroxide concentration.
Note: The rate of decrease in hydrogen peroxide concentration is expected to
vary by device, and depends on several factors such as age or overall condition
of the device, past maintenance practices, and local water conditions.
Important
i
The expected service life of the Heater-Cooler System 3T is 10 years. Use of the
device beyond the expected service life, improper maintenance practices, or the
incorrect use of disinfection and water preservation chemicals may lead to
excessive corrosion damage of the water circuit components.
CP_IFU_16-XX-XX_USA_021
97
Heater-Cooler System 3T • Maintaining the heater-cooler
Maintenance schedule and checklists
1
2
Record or estimate the operating hours
1 Where required by local regulations, record the operating hours of the heater-
cooler.
1 If local regulations do not require tracking of operating hours, LivaNova
recommends estimating the operating hours.
1
Schedule a maintenance check
Schedule a maintenance check with a LivaNova service technician every
1,000 operating hours or every 12 months, whichever comes first.
6.2.5 Checklist: Proper disposal
1
Dispose of the heater-cooler and/or accessories according to your
local environmental regulations and hospital policy
1 Dispose of used single-use items according to local environmental regulations
and the manufacturer’s instructions. Wear appropriate personal protection
equipment (PPE) while disposing of these items.
1 The heater-cooler must not be disposed of in accordance with Directive
2012/19/EU or its national transpositions as the system is to be regarded as
potentially infectious at the end of its service life.
1 Dispose of all heater-cooler accessories according to your hospital’s policy,
local environmental regulations, and the accessories’ instructions.
1 Only refrigerant systems experts are authorized to dispose of the cooling
circuit’s refrigerant (CFC-free HFC-[hydrofluorocarbons]) refrigerant. The
refrigerant system service information is contained on the refrigerant system
label.
1 As required by European Directive 93/42 and/or local regulation
98CP_IFU_16-XX-XX_USA_021
Heater-Cooler System 3T • Maintaining the heater-cooler
Cleaning and disinfecting external surfaces, connectors, and fittings
6.3Cleaning and disinfecting external
surfaces, connectors, and fittings
Cleaning and disinfecting schedule
● After every use, clean and disinfect the device surfaces
● Upon every connection, disinfect the tubing connectors on the heater-cooler
● Upon every connection, disinfect the fittings and connectors on the tubing
Important
i
● Make sure the CAN jack cover is closed before cleaning and disinfecting.
● Use only disinfectant wipes and sprays that are approved for use with the
heater-cooler. See the table below for a list of approved cleaning supplies.
Approved pre-soaked disinfectant wipe
Product nameManufacturer
CaviWipesMetrex
Approved disinfectant spray
Product nameManufacturer
CaviCideMetrex
CP_IFU_16-XX-XX_USA_021
99
Heater-Cooler System 3T • Maintaining the heater-cooler
Cl
Disinfect
Cleaning and disinfecting external surfaces, connectors, and fittings
6.3.1 Cleaning and disinfecting heater-cooler external
surfaces
Summary of steps you will complete in this section:
1Clean and disinfect the vacuum gauge
2Clean all accessible device surfaces
3Disinfect all accessible device surfaces
1
Clean and disinfect the vacuum gauge
1 Using only approved disinfectant
wipes, clean the vacuum gauge,
removing any surface
contamination and liquid spills.
2 Using only approved disinfectant
wipes, disinfect the vacuum
gauge.
Important
i
Using other disinfectants may result in damage or a shortened lifetime of the
gauge face.
ean
pascals
40
20
60
0
80
TANK VACUUM
100
CAL
IBRATED FOR
VERTI
C
AL POSITION
M
AX PRESSURE 1
00kPa
pascals
40
20
60
0
80
TANK VACUUM
100
CAL
IBRATED FOR
VERTIC
AL POSITION
M
AX PRESSURE 1
00kPa
100CP_IFU_16-XX-XX_USA_021
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.