Littelfuse AN9773 Application Note

Varistor Testing
Application Note January 1998
Introduction
This note details the common tests of varistor parameters and describes suitable test methods using simplied test circuits.
All tests are performed at 25 The test circuits and methods given herein are intended as a general guide. Since the tests frequently entail high voltages and currents, the user must exercise appropriate safety precautions.
Engineering Evaluation
It is important to focus on the key characteristics and ratings to determine if the component can perform as expected. Typically, for a varistor, its nominal voltage, clamping voltage, standby current, insulation resistance, and capacitance are measured. The surge current, or energy, and waveshape available in the circuit together with its frequency of occurrence should be measured or computed. The characteristics of these expected transients should then be checked against the pulse ratings and the power dissipation ratings of the selected varistor type. Where suitable equipment is available, these ratings may be veried.
Product Qualication
A product qualication plan often will be used to detail the electrical and environmental tests to which sample components may be subjected. The suggested electrical characteristics tests could include (with appropriate conditions and limits): nominal varistor voltage, V maximum clamping voltage, (optional, especially for AC applications); insulation resistance; and capacitance. A test to ensure surge current withstand capability may be included in the qualication plan. This test must be carefully performed and specied (by using either 8/20µs or 10/1000µs waveshapes) consistent with the pulse lifetime rating chart of the varistor selected. Other qualication tests may be used to ensure mechanical integrity, humidity resistance, solderability, and terminal/lead strength.
Incoming Inspection
The equipment maker may wish to verify that shipments received consist of correct parts at the expected quality level. For incoming inspection of Littelfuse Varistors, it is recommended that sample testing include nominal varistor voltage (V voltages specied on the purchase drawing/specication. Other electrical sampling tests frequently performed can include insulation resistance and capacitance. Tests such as maximum clamping voltage, V are usually checked only on a periodic audit basis.
) tested against the minimum and maximum
N
C, unless otherwise specied.
;
N
VC; DC standby current, ID
, and DC standby current, ID,
C
AN9773
Field Maintenance
Field maintenance testing is done to verify that the varistor is still providing the intended protection function.
The nominal varistor voltage should be tested against the minimum limits for the model using the method described in the Nominal Varistor Voltage V
section. If the varistor is
N
open, short, or more than 10% outside either limit, it should be replaced. The DC standby current may also be measured.
Measurement of Varistor Characteristics [1]
Nominal Varistor Voltage V
This is measured at a DC test current, I models. A simplied circuit for instrumenting this test, shown in Figure 1, is suitable for varistors up through a rating of 300V voltage will be needed. Resistor R1 has a dual purpose. In conjunction with the variable voltage supply, E1, it forms a quasi-current source providing up to 6mA when switch S1 is closed. Also, R1 is used as a current sensor to measure current owing through the varistor-under-test. To use the circuit, the operator places switch S2 in position I and S3 into position V S1 is closed. E1 is then adjusted to obtain a reading of 100V ±5V on the digital voltmeter. Approximately 1mA of current will be owing in R1. When switch S2 is placed in position V, the varistor voltage will be indicated on the voltmeter. The values of R1 and E1 supply voltage can be scaled appropriately for other voltage-current test points.
If the varistor voltage test is implemented on automatic test equipment, a “soak” time of 20ms minimum should be allowed after application of test current before voltage measurement. This is necessary to allow varistor voltage to settle toward a steady-state value. Figure 2 illustrates the time response of a specimen varistor with a constant 1.0mA current applied. As can be seen, the varistor voltage initially may rise to a value up to 6% greater than nal. With a 20ms or greater soak time, the measured value will differ by less than 2% from the steady-state value.
. Above the 300V
RMS
. A test device is then inserted into the socket and
N
S1
R1
S3
V
0V-600V
E1
FIGURE 1. SIMPLIFIED CIRCUIT FOR VARISTOR VOLTAGE
AND DC STANDBY CURRENT TESTS
N
N
of 1mA for product
N
rating, a higher supply
RMS
R2 I
D
+
I
DVM
S2
-
V
R1 = 100kΩ, 1%, 1W(V R2 = 1kΩ, 1%, 1/2W(I
TEST)
N
TEST)
D
10-141
1-800-999-9445 or 1-847-824-1188 | Copyright © Littelfuse, Inc. 1998
C
Application Note 9773
V (T) 5V/DIV
250V
240V
230V
FIGURE 2. VOLTAGE-TIME V(T) CHARACTERISTICS OF A
LITTELFUSE VARISTOR (V130LA10A) OPERATING AT A CONSTANT DC CURRENT OF
0.1ms/DIV
1ms/DIV 10ms/DIV
100ms/DIV 1000ms/DIV
T
For varistor models that are commonly used on 60Hz power lines, the V
limits may be specified for a 1.0mA peak AC
N
current applied. If an AC test is preferred by the user, a schematic approach similar to that shown in Figure 1 is used, except an AC VARIAC™ is substituted for the DC power supply, and an oscilloscope is substituted for the voltmeter. This circuit is equivalent to that of a typical curve tracer instrument.
To avoid unnecessary concern over minor measurement anomalies, three behavioral phenomena of metal-oxide varistors should be noted. First, it is normal for the peak varistor voltage measured with AC current to be about 2% to 5% higher than the DC value, as illustrated by Figure 3. This “AC-DC difference” is to be expected, since the one-quarter cycle period of a 60Hz wave is much less than the 20ms minimum settling time required for DC readout.
Second, it is normal for the varistor voltage to increase slightly when first subjected to electrical current, as shown in Figure 4. This might be considered a “break-in” stabilization of the varistor characteristics. During normal measurement
the voltage shift typically is less than 1%. This voltage shift is of little consequence for most measurement purposes but might be noticeable when viewing a DVM as in the test method of Figure 1. The visual DVM observation should be made shortly after power is applied, with measurement to not more than three significant figures.
Third, it is normal for the varistor voltage-current characteristic to become slightly asymmetrical in polarity under application of DC electrical stress over time. The varistor voltage will increase in the same direction as the polarity of stress, while it will be constant or will decrease in the opposite polarity. This effect will be most noticeable for a varistor that has been subjected to unipolar pulse stresses or accelerated DC life tests. Therefore, to obtain consistent results during unipolar pulse or operating life tests, it is essential to provide a polarity identification for the test specimens. However, for initial readout purposes, this effect usually is insignificant.
Maximum Clamping Voltage, V
Two typical current impulses that may be used to define the varistor clamping voltage are the 8/20 µ s and the 10/1000 µ s pulses. Figure 5 shows typical varistor test waveforms for these two impulses.
The clamping voltage of a given model varistor at a defined current is related by a factor of the varistor voltage. Therefore, a test of the nominal varistor voltage against specifications may be sufficient to provide reasonable assurance that the maximum clamping voltage specification is also satisfied. When it is necessary to perform the V special surge generators are required. For shorter impulses than 8/20 µ s, precautions must be observed to avoid an erroneous “overshoot” in the measurement of the clamping voltage. The Equipment for Varistor Electrical Testing section gives general information on surge generators; a brief description of the “overshoot” effect follows.
test,
C
V(T)
5V/DIV
DC
100
VOLTAGE (V)
10
10
FIGURE 3. AC AND DC CHARACTERISTIC CURVES FIGURE 4. V130LA10A) VARISTOR VOLTAGE FOR THE
-6
-7
10
-5
10
I, CURRENT (A)
10-142
AC 60Hz
130V
RMS
PRODUCT MEDIUM VOLTAGE MATERIAL
-4
10
10-310-210
RATED
-1
INITIAL CYCLES OF 60Hz OPERATION AT A PEAK CURRENT OF 1.0mA
VARIAC™ is a trademark of Glen Rad, Inc.
T, 50ms/DIV
Application Note 9773
10A/DIV
100V/DIV
0
FIGURE 5A. 8/20 µ s, WAVE I
OUTPUT LEAD
FROM
TRANSIENT
GENERATOR
AREA
0.5cm
10µs/DIV
= 50A, V
P
= 315V
P
FIGURE 5B. 10/1000 µ s, WAVE I
FIGURE 5. TYPICAL CLAMPING VOLTAGE TEST WAVEFORMS (LITTELFUSE
FIGURE 6. VARISTOR TYPE V130LA10A)
VOLTAGE PROBE
2
COPPER TUBE
SURROUNDING
VOLTAGE PROBE
CURRENT
PATH
AREA
22cm
2
1ms/DIV
= 50A, V
P
= 315V
C
10A/DIV
100V/DIV
0
CURRENT
PAT H
VARISTOR
FIGURE 7A. MINIMAL LOOP AREA
GROUND
VARISTOR
FIGURE 7B. EXCESSIVE LOOP AREA TYPICAL
“OVERSHOOT” OF LEAD-MOUNTED VARISTOR WITH STEEP CURRENT IMPULSES
FIGURE 7C. CURRENT RISE OF 8 µ s FIGURE 7D. CURRENT RISE OF 0.5 µ s
FIGURE 7. EFFECT OF LEAD LENGTH ON “OVERSHOOT”
10-143
Loading...
+ 5 hidden pages