L1125 User Guide
Troubleshooting
Symptom: L1125 appears to be dead. Power LED does not come on, display
remains blank
Possible causes: Rear power switch is off (power switch is adjacent to the rear
connectors, the switch on the battery bracket is not used with L1125)
Power lead is not connected
L1125 has been set to software upgrade mode and is awaiting new
software. To check for this, connect terminal software (e.g. Tera Term)
via the rear RS232 connection with settings of 9600 Baud, no parity,
8 data bits, 1 stop bit, XON/XOFF flow control and 20ms transmit
delay at end of line. Turn power off/on, if the L1125 is in
reprogramming mode it will send a message on the RS232
connection (refer to the Upgrading L1125 Software section for further
information).
Symptom: Camera does not react to Operators Control Panel control
adjustments
Possible causes: The Camera Address setting (in the Reverse Path menu) is not the
same as the OCP address (switch setting on side of OCP)
The Frequency setting (in the Reverse Path menu) is not the same as
the XPe Transmitter Interface.
The correct camera type has not been selected (in the Reverse Path
menu)
The wrong transmit filter is being used.
The camera requires an external control lead but the lead is not
fitted or not making proper connection
The camera supports selection of either internal or external control
links but the corresponding Cam I/F setting in the Reverse Path menu
has not been correctly configured
If the Data light on the OCP does not flicker when the iris control is
moved then the OCP is not communicating with the L1251 Interface,
check that the OCP address is not set to 0 and that the cable
between OCP and L1251 is properly connected and undamaged
If the L1251 status light and OCP Data light are working correctly,
remove the L1125 from the Camera, set the camera type (in the
Reverse Path menu) to Ikegami Handheld and check that the RX
status LED is on when the OCP iris control is moved. If the LED stays
off the L1125 is not receiving commands.
For further information, including the latest list of supported cameras and
details of any adaptor plate modifications or cables required, refer to the
Support Options section of the Link Research website www.linkres.co.uk or
contact the Support Line +44 1923 200909
Images are for illustration only and may differ from components supplied
Link Research Ltd, 23 Watford Metro Centre, Dwight Road, Watford, Herts, WD18 9XA, UK
Telephone +44 1923 200900 (general), +44 1923 200909 (Customer Support)
© Link Research Limited 2004, 2005 Page 25 of 25