Lindab PC 410 User Manual

Page 1
lindab | we simplify construction
PC 410
Manual Flow and Pressure Meter
Page 2
2
Content
6.7 Option “Heating Check” (Energy Utilization Check) 37
7.3 Data Transfer with the Computer or Notebook 52
Page 3
3
1. General information
1.1 Operation Manual Information This operation manual allows you to safely work with the PC
410. Please keep this manual for your information.
The PC 410 Flow and Pressure Meter should be employed by professionals for its in-tended use only.
Liability is void for any damages caused by not following this manual.
1.2 Notes
WARNING!
Not following this warning can cause injury or death.
ATTENTION!
Not following this note can cause permanent damage to the device.
NOTE!
Useful information
1.3 Intended Use The meter can be used for the measurement of differential pressure, airflow, flow speed and temperature as well as for the registration of humidity and ambient pressure. It is especi­ally designed for the following purposes:
• Acceptance measurements at ventilation systems accord-
ing to DIN EN 12599.
• 4 PA Pressure Test to confirm the combustion air availabil-
ity.
• Measurement of the ventilation loss and the surface loss.
• Flow pressure, installation pressure, standing pressure, jet
pressure.
• Flow measurement in air and flue gas ducts.
• Data logger.
Page 4
4
1.4 Basic equipment
Device Components
PC 410 Flow and Pressure Meter with bat-
teries Hand strap Silicone hose, 6 m Rubber hose, 1,5 m Plastic case
1.2 Transport
ATTENTION!
Improper transport can harm the instrument.
Always transport the instrument in the provided carrying case in order to prevent damage. The pressure connectors have to be protected by the protection cap.
The case is included in the sets and also can be bought separately.
1.6 Information on disposal Electronic equipment does not belong into domestic waste, but must be disposed in accordance with the applicable statutory provisions.
You may hand in any defective batteries taken out of the unit to our company as well as to recycling places of public dis­posal systems or to selling points of new batteries or storage batteries.
1.7 Contact Please contact your local Lindab distributor
or send for technical questions an E-Mail:
Page 5
5
2. Specifications
2.1 Measured values
Description Data
Range ±100 hPa Accuracy < 3 % of the reading, if the value is
< ± 10 Pa, better than ± 0.3 Pa
Resolution 0.1 Pa at -1100 hPa to +1100 hPa,
otherwise 1 Pa
Measurement of very small pressures
Description Data
Range ±100 hPa Accuracy < 3 % of the reading, if the value is
< ± 10 Pa, better than ± 0.3 Pa
Resolution ± 0,01 Pa
Internal temperature measurement
Description Data
Range -20 °C to 60°C Accuracy < ± 1 °C Resolution 0.1°C
External temperature measurement
(optional, e.g. with Measuring Pliers for Temperature Measurement or Special Wall Temperature Probe)
Description Data
Range 2 canals, -20.0 °C to +800.0 °C Accuracy < ± 2 °C at 0°C to 133°C, otherwise
1.5 % of the value, according to the directive EN 50379-2
Resolution 0.1°C
Humidity measurement (only AF version)
(integrated humidity sensor)
Description Data
Range 0 % to 100 % RH (relative
humidity), non condensing
Accuracy < ± 2 % RH, in range 0 to 90 % RH,
otherwise < 3 % RH
Resolution 0.1 % RH
Ambient Pressure measure­ment (only AF version with humidity sensor)
Description Data
Range ±300 hPa, else 1100 Pa Accuracy < ± 1.5 % hPa of reading Resolution 0.1 hPa
Page 6
6
2.2 Calculated values
Variable Calculation
Pressure units hPa, Pa, mmH
2
O, PSI, inwc, bar, mbar, conversion according to the general conversion rules.
Temperature units °C, °F,
conversion according to the general conversion rules.
Air speed
Description Data
Range 0.3 m/s to 120 m/s Resolution 0.01 m/s Continuous density
correction
automatic, with temperature signal.
2.3 Logger Description Data
Extent 9999 measurements, each with the
measured value of pressure and humidity (optional) and three tem­perature values (if the external tem­perature sensors are connected) can be stored in the internal memory. The user can set the log rate from 1 second to 24 hours.
USB interface for data transfer to a PC
Permanent online data, already during the registration.
User adjustable log rate
Sampling interval of 1 second to 24 hour can be set.
NOTE!
Ensure that the meter is connected to the power supply or that the batteries are charged.
Page 7
7
2.4 Technical Data Description Data
Power Consumption
4 AA rechargeable batteries or 4 AA disposable batteries 1.5 volt
- working mode: ca. 60 mA,
- off-mode and log mode: ca. 45 µA for clock and processor
Interface USB- (COM-Port),
data transfer to the PC.
It is possible to transfer the data directly from the meter to the TD 600 thermal fast printer.
Storage temperature
-20 °C to +60 °C
Operating temperature
5 °C to +40 °C
Mass ca. 365 g, with batteries and
integrated magnet Dimension 80 × 550 × 60 mm Date and time Date and time are included in the
measurement report Internal Memory 2 MB
Page 8
8
11
12
2
9
10
1
8
4
5
6
7
3
3. Component explanation
3.1 Basic unit
Fig. 1: Overview
Page 9
9
Number Part
1 Mini-USB-Port 2 Connection for protection cap 3 Power jack 4 Color display 5 ESC key context-sensitive
Escape
NOTE!
Keeping this key depressed will always call up the main menu.
6 Arrow keys The function changes according to the context.
Scroll up and down.
7 ENTER
and On-/Off key
a) Confirm the setting.
b) The function of the key changes according to the context.
c) Press and hold the key for 3 seconds to turn off the unit. 8 Infrared interface Thermal fast printer to print out the measurement report 9 Overpressure socket (+) Main connection 10 Underpressure socket (--) Reference socket for the measurement of the differential
pressure and static socket for the S-tube probe 11, 12
14
Standard jacks for temperature probes NiCr-Ni
14 Battery cover (in Fig. 1 on the rear side)
There are diffusion openings for the internal registration of the
humidity and temperature on the rear side of the Lindab
PC 410. The internal temperature measurement is necessary
for the temperature compensation of the pressure sensor.
Page 10
10
3.2 Probes and Components
S-tube probe for measuring the flow speed (see chapter 6.5) and the airflow (see chapter 6.2).
• Insert the plug of the S-tube probe into jack 11 (Fig. 1) of the PC 410.
• Connect the overpressure hose (turned against the flow) of the S-tube probe to the overpressure socket (+) (Fig. 1, part
9) and the underpressure hose to the underpressure socket (-) (fig. 1, part 10).
Fig. 2: S-tube probe
Flexible capillary
For the 4 PA test two flexible capillaries are needed.
Fig. 3: Capillary tube
Page 11
11
Two temperature measuring pliers can be connected at the same time.
• Insert the plug of the temperature measuring pliers into the jacks 11 and 12 (Fig. 1) of the PC 410.
The temperature value measured by the meter connected to jack 11 will be indicated in the display as T1 and the one measured by the meter connected to jack 12 will be shown as T2.
Fig. 4: Connecting temperature measuring pliers
Connection of the Special Wall Temperature Probe to measure the heat transmission coefficient (See chapter 6.9) Two Special Wall Temperature Probes can be connected at the same time.
• Insert the plug of the temperature measuring pliers into the jacks 11 and 12 (Fig. 1) of the PC 410.
Fig. 5: Connection of the Special Wall Temperature Probe for measuring the heat transmission coefficient.
NOTE!
Any thermocouple type K may be connected to the jacks 11 and 12.
Page 12
12
3.3 Function
Fig. 6: Interior structure of the meter
Pressure sensor
Pressure socket (+)
Pressure socket (-)
Ambient pressure and humidity sensor
Page 13
13
3.4 Operating Mode
The Flow and Pressure Meter PC 410 is a high precision mul­timeter for the measurement of differential pressure, volume flow and temperature as well as the registration of humidity and ambient pressure. (These two functions are only available in the version AF). Because of its extremely high precision, the meter measures even very small pressures. Therefore, it is ideal to determinate air speed, gas pressure, volume, tension and also the flue gas temperature. A maximum measuring range up to 100 hPa and a burst pressure of 0.75 bar provide a sufficient degree of security even for high pressure ranges.
3.5 Display and key setup
1. The PC 410 has a color 2.4' display.
2. The OLED-technology allows reading the display from
almost any angle of view.
3. Four keys are available below the display to operate the
PC 410.
Fig. 7: Display and keys
Escape On/Off
Scroll/Zoom
Page 14
14
Fig. 8: Display details
The display is divided into a status, a menu and a readings segment.
The currently selected task is shown in the left status seg­ment.
– Date and time, customer (if selected), system diagnosis status and battery level are shown in the status segment on the right.
The readings segment displays the readings or the menu items.
The menu segment is situated at the bottom of the display. It consists of three soft keys.
Version AF (with internal ambient pressure and humidity sensor): The values are indicated in the "pressure measure­ment" menu.
Page 15
15
4. Getting started
4.1 Check battery status
When the device is switched on, the battery indication is displayed in the upper right hand corner of the screen. A fully charged set of batteries is shown as a solid green battery symbol. An empty red symbol indicates discharged or empty batteries. If the batteries are almost empty the back light will dim out. The warning "charge batteries" will appear. In this case, finish the measuring process as soon as possible.
NOTE!
You can still go on measuring while battery charging is in progress.
WARNING!
Incorrect use of batteries can cause injury!
Never dispose the battery to fire or high temperatures. Danger of explosion!
Liquids can leak out of the batteries due to mistreatment. Never touch the liquid. If you get in contact to the liquid remove it with water. If the liquid got into your eyes, remove it with water and see your doctor as soon as possible.
WARNING!
Risk of electrical shock!
Never touch the power supply with wet hands!
Protect the power supply against water and moisture!
Do not unplug the recharger by pulling the cable!
Do not use the power supply when the voltage requirements of the recharger and the supply do not match!
The batteries can be recharged even while they are located in the analyzer.
Fig. 9: Position of the batteries in the PC 410
Page 16
16
How to charge the batteries:
• Before you plug the recharger into the outlet, connect it with the charging connection of the PC 410 (Fig. 1, part 3).
Standard AA rechargers can be used for external charging when the batteries are not located in the analyzer.
• To remove the batteries from the meter open the battery compartment cover by pressing the clip and pushing the battery compartment cover at the same time.
Recharging empty batteries can take up to 3 hours.
ATTENTION!
– Ensure rechargeable batteries are equipped in the meter before charging. Never try to charge disposable batteries.
Use 4 AA rechargeable batteries only. – Only use the charger delivered with the device.
NOTE!
To ensure maximum capacity use batteries of same age and condition only. Always replace a full set of batteries when necessary.
Fig. 10: Open the battery compartment.
Page 17
17
5.2 Measuring After that the meter will automatically switch to the Air flow measuring mode.
1. Al readings (measured and computed values) will be
shown continuously in the readings segment. The following functions can be selected from the menu segment
• Submenu „PD=0“: The pressure sensor will be zeroed once
again.
• Change the k-value
• Stop-key to stop the measurement
5. Operation
5.1 Turning the Meter on and Self
Check
• Turning on the Meter Press on/off-key (on the right) The Meter will automatically start with a 10 second self test
and zeroing.
ATTENTION!
Before the Meter is used, a visual check must ensure each and every time that all functions work properly.
ATTENTION!
Do not connect any hose while the Meter is performing the self check and do not expose it to any differential pressure, because the pressure sensor will automatically be zeroed during the self test.
Fig. 11: Key pad with on/off-key
Fig. 12: Zeroing
Fig. 13: Measuring mode
Page 18
18
5.3 Using the Pressure computer
• Holding the left button depressed (ESCAPE) will always
stop the current menu and call up the main menu.
• In the pressure measuring mode, press the up or down key
to enter the zoom-mode. When the readings are zoomed in, they can be read even from a distance. In the zoom mode three values are shown in the display. Press the up or down key shortly to scroll the readings.
• Hold the up or down key depressed to zoom in or out.
Fig. 15: Zoom feature
Fig. 14: Display and keys
Escape On/Off
Scroll/Zoom
Page 19
19
Fig. 16: PC 410 with supporting loop
• The PC 410 is equipped with integrated magnetic holders
on the rear side. The magnetic holders will fix the analyzer on any magnetic material surface.
• It is also possible to fix the PC 410 with the hand strap
during the measurement.
While measuring either
• hold the Pressure Computer in one hand or
• place it with the magnetic holder to any appropriate surface or
• fix it with the hand strap.
ATTENTION!
Do NOT place the Pressure Computer unsecured, it can fall down and get damaged.
Page 20
20
6. Measurement modes in the main menu
6.1 Selecting the main menu
• Top the current measurement.
• Press "menu" to activate the main menu. In the main menu,
all submenus and applications can be activated. The fol­lowing submenus can be selected:
Fig. 17: Main menu
Page 21
21
6.2 Option “Air flow (k value)” Select the submenu "Air flow (k-value)" to measure the air flow at apertures or similar measuring openings of air pipes or vents. This method is simple and precise and therefore suitable for the adjustment of air pipe systems and vents.
6.2.1 Theoretical background In this menu, the air flow measurement is based on the k value. The k value is a factor which describes the relation between the decrease in pressure and the air flow.
The air flow of measuring equipments mentioned in chapter
6.2 is proportional to the term
r
Dr
.
The different manufacturers indicate the proportionality factor in a different way (see table 1). In the PC 410 the measure­ment type is indicated with the corresponding short descrip­tion (see table 1). The user can choose the unit of the volume flow in the setup-menu (l/s, m³/h, m³/s, cfm). The factory setting is l/s and m3/h respectively.
NOTE!
If you select a certain measurement type in the setup-menu, the displayed unit will not change automatically.
The PC 410 will always calculate with the unit l/s, even when another unit is displayed. Table 1 shows the formula used for the calculation.
For all calculations, the PC 410 will consider the standard air density r, if no external temperature probe is connected. When a temperature probe is connected, the air density will be calculated automatically.
Page 22
22
Measurment type Short description
Manufacturer specific formula and unit
Components Formula to determine l/s
considering the density
l/s
∆pq = k · [l/s]
e.g. small fans and air passages, iris and calipers
1,2 · V = K
[l/s]
·
K
min
= 0,1 K
max
= 1000
ρ
∆p
E
m³/h
∆pq = k · [m³/h]
e.g. fans and air passages
1,2
·
3,6
V = K
[m³/h]
·
K
min
= 1 K
max
= 5000
ρ
∆p
E
1/(m³/s)
∆pq = · [m³/s]
1 k
e.g. Scandinavian fans
1,2 · 1000
·
V =
K
[1/(m³/s)]
ρ
∆p
E
K
min
= 0,01 K
max
= 100
(m³/s) × 1,3
q = K · [m³/h]
ρ
2
× ∆p
only used by few German fan produc­ers
1,2
·
3,6
V = K
[m³/h·1,3]
ρ
∆p
E
K
min
= 1 K
max
= 5000
Table 1
Explanation: V = air flow k = k value (Take the k-value from the data sheet and enter it Into the meter).
Δp= measured pressure
r = air density, calculated on the basis of temperature and the pressure
Units V l/s
Dp in Pa
p in kg/m³
E 0,50 – 1,00
• In the Setup-Menu, select the formula that corresponds to the manufacturer according to table 1
(Setup>k-Value).
By default, the exponent will be 0.5 as indicated in the formula. Some fan manufacturers inform explicitly in the data sheet that a special exponent has to be used for the airflow calculation.
• In this case, enter the exponent indicated in the data-sheet, in the setup-menu of the PC 410.
(see chapter 6.12, „exponent“).
Page 23
23
6.2.2 Connecting the meter to the vent
• Connect the Wöhler meter with two measuring hoses to the
vent.
The overpressure socket (+) of the meter has to be connected to the aperture with the higher pressure and the underpres­sure socket (-) to the aperture with the lower pressure.
NOTE!
If the hoses are wrongly connected, the reading will be negative. In this case inverse the hoses.
• During the measurement, the Wöhler meter can be fixed to
the protective grid with a magnet, so that the results can easily be read from the display.
• When the aperture is under the ceiling, the meter can be
fixed with the hand strap.
Fig. 18: Example - Connection of the measuring hoses
Fig. 19: Example - Fixing with a magnet
Fig. 20: Example - Fixing with the hand strap
Page 24
24
6.2.3 Performing the Test
After the measuring mode has been activated, in the upper part of the display the air flow will be indicated in the unit selected in the setup-menu.
Below, the airflow will be indicated in the other unit.
If temperature sensors are connected, T1 is the reading of the temperature measured by the sensor connected left and T2 is the reading of the temperature measured by the sensor connected right. If no temperature sensor is connected, the meter will calculate with the standard density
value r = 1,200 kg/m³.
The following values will also be indicated:
Differential pressure P
D
Air density r
k value Below, the k value that has been entered by the user is indi-
cated.
Fig. 21: Menu "Air flow (k value)"
Enter the k-value as follows.
• Select K value, see Fig. 21: Menu "Air flow (k value)"
• Increase or reduce the value with the up and down keys.
• Press OK to confirm.
NOTE!
The k value is specified in the data sheet of the air vent. The user must enter this k value into the Lindab PC 410 before starting the measurement.
Fig. 22: Entering the K-value
NOTE!
Always enter the k value related to l/s (even if you have changed the air flow unit from l/s to m3/h in the setup­menu).
Page 25
25
PD = 0 Press PD = 0 (menu bar, see Fig. 21: Menu "Air flow (k value)"
to zero the pressure sensor.
NOTE!
We recommend to zero the pressure sensor, if the pressure indicated in the display is not PD = 0 after the measuring hoses have been removed.
• When measuring, wait until the value has stabilized and
then press "Stop".
• Press "Accept" to capture and save the value. In the main
menu a tick mark beside the menu "air flow (k value)" will indicate that the measurement has already been done.
• Press "Escape" to exit the menu without saving. Press "Proceed" to return to the menu "air flow (k value)" for
further measurements.
NOTE!
Select "Print" to print out a report.
NOTE!
If the airflow reading is not stable or if the resolution is not high enough, go to the Setup-menu and select Attenuation > fine (see chapter 6.12)
Fig. 23: measuring value that has not been accepted yet.
Page 26
26
6.3 Option “K-Value Matching”
This mode allows to adjust the k-value of iris apertures, air passages and terminal boxes.
If the component is equipped with an adjustment scale for the k-value, the airflow can be adjusted without a calculator or a diagram.
Go to the main menu and open the submenu "k-Value Match­ing".
The menu shown in the figure will appear. The k value is indi­cated in the upper part of the display.
Below the k value, the differential pressure PD and the air density r will appear.
If temperature sensors are connected, T1 is the reading of the temperature measured by the sensor connected left and T2 is the reading of the temperature measured by the sensor connected right.
Target Flow Below the target flow is indicated. The target flow has to be
entered by the user.
Enter the target flow as follows:
• Select the "Tar. flow"-key from the menu bar (see Fig. 24).
• Increase or reduce the value with the up and down keys.
• If necessary, select m³/h or l/s and change the unit.
• Press OK to confirm.
Fig. 24: Menu K-Value Matching
Fig. 25: Entering the Target Flow
Page 27
27
PD = 0 Press PD = 0 (menu bar, see Fig. 24 ) to zero the pressure
sensor.
NOTE!
We recommend to zero the pressure sensor, if the pressure indicated in the display is not PD = 0 after the measuring hoses have been removed.
• Modify the k-value of the component until the k value of
the component matches with the k-value indicated by the meter.
• Press OK to confirm. The air flow mode (k-value) will
appear again.
• Now the user can perform another measure-ment to con-
firm the result and save the value afterwards.
• If necessary, select the "k-value matching" mode from the
main menu again.
Fig. 26: Air Flow Mode
Page 28
28
6.4 Option “Airflow (Pitot)” In this menu, the airflow (volume flow) is measured in l/s or in m3/h. Select the unit in the setup menu. (see chapter 6.12)
1. With the Lindab PC 410, grid point measurements can be
done, so that the information provided by the measure­ment is significant (see chapter 6.4.2).
6.4.1 Measuring the airflow (Pitot) Connect the S-tube probe to the Lindab PC 410 and insert it into the measurement opening. Proceed as described in chapter 0
After the Menu "Airflow (Pitot)" has been selected, the current airflow value V will be indicated.
The airflow is calculated from the surface area and the air velocity.
V = A · v Formula 3
The Airflow v is indicated in the unit which has been selected in the setup-menu.
Below the airflow is indicated in the other possible unit.
If temperature sensors are connected, T1 is the reading of the temperature measured by the sensor connected left and T2 is the reading of the temperature measured by the sensor connected right. If no temperature sensor is connected, the meter will calculate with the standard density
value r = 1,200 kg/m³.
The following values will also be indicated.
Air speed v.
Air pressure Pamb
Differential pressure P
D
Air density r
Pitot factor of the S-tube probe (Pf: can be adjusted in the setup menu, 0.93 for the S-tube probe) cross section area A of the duct (The user has to enter the cross section area A as follows:)
Fig. 27: Menu Airflow (Pitot)
Page 29
29
Entering the cross section area:
• Press "Stop" and then "Area" to bring up menu of the cross
section area.
• Go to the parameter "Profile" with the up and down keys.
• Use the right arrow to select between "round, rectangle,
arbitrary"
• Select by using the up and down keys.
• Change the other parameters in the same way.
The Lindab PC 410 will calculate the cross section area automatically.
Press "Return" and then “Proceed” to start the measuring with the correct area value.
6.4.2 Grid-point measurements The readings shown in the display are instantaneous values.
With the Lindab PC 410 grid-point measurements can be done. This means that the air flow is measured at up to 12 different points of the tube and the meter will calculate the average of the readings.
Fig. 28: Entering the cross section area
NOTE!
The average value determined by the grid-point measure­ment is much more informative than a value measured in one point only, because possible fluctuations during the measurement will be compensated.
Measurement 1
• Enter the area-value and insert the S-tube into the
measuring opening as described before. Then start measuring.
• When the value has stabilized, press "Stop".
Page 30
30
The readings of the first measurement are shown in the display.
• Press the Evaluation-key to display all readings.
The result of the first measurement is not shown yet, because it has not been captured yet.
• Press "Accept" to capture the value as the velocity 1.
The air flow is indicated in the upper part of the display as the result of the measurement 1.
• Press "Return" to bring up the measurement menu.
• Press "Proceed" to start a new measurement.
Measurements 2 - 12 The current readings will be indicated in the display again.
• Press "Stop", when the values are stable.
• Press "Evaluation" to bring up a summary of all measure-
ments.
• Press "Accept" to capture the recent value.
• Perform the measurements 2-12 the same way.
Fig. 29: Standard report, when no reading has been captured yet.
Fig. 30: Standard report after the readings have been captured.
Page 31
31
In the display, the mean speed is shown above the readings.
The air flow is indicated in the upper part of the display as the average of the results of all mea-surements performed until now.
Fig. 31: Average value
NOTE!
Select "Print" to print out a report.
• In the Evaluation menu select "Finish mea-surement" (top
of the display) to finish the measurement.
Fig. 32: Finishing the measurement
NOTE!
If the airflow reading is not stable or if the resolu-tion is not high enough, go to the Setup-menu and select Attenuation > fine (see chapter 6.12)
In the main menu, a tick mark beside the menu "Air flow (Pitot)" will indicate that the air flow measurement has already been done.
Select "Print" to print out a report.
Page 32
32
6.5 Option “Air speed” In the air speed mode, the air speed (v) of the gas is meas- ured in m/s when the gas temperature is 800 °C or less.
Fig. 34: Connection of the S-tube probe to the PC 410
NOTE!
An S-tube probe (see accessories) is needed for this meas­urement.
The wanted signal of the Wöhler S-tube probe is better than that of a normal pitot tube, if the air speed is equal. In the setup-menu the pitot factor 0,93 must be set once (see chapter 6.12).
6.5.1 Measuring the air speed
1. Connect the pressure connector of the S-tube to the (+)
overpressure socket of the Lindab PC 410.
2. Connect the static pressure connector of the S-tube to
the (-) underpressure socket of the Lindab PC 410.
3. In calm air, switch on the Lindab PC 410 and wait until
the zeroing procedure of the sensor has finished.
4. When closed, the S-tube can be inserted through the
measurement opening into the gas flow.
5. Fix the S-tube with a cone, open it and direct the sensors
as shown in the figure.
6. If necessary, adapt the S-tube to the cross section of the
gas line with the help of the two knurled screws.
NOTE!
The S-tube must be adjusted before the measurement pro­cedure starts, since it is required to check and/or adjust the oppositely running orientation of the measuring tubes after loosening the knurled screws and adjusting the immersion depth.
Static pressure Temperature
Immersion depth
Flow
Knurled screws
Fig. 35: Measuring the air speed with the S-tube
7. In the main menu, select "Air speed" and confirm with
OK.
The air speed v is indicated in the upper part of the display. The temperature, measured by the sensor or the S-tube, and the ambient pressure Pamb, measured by the integrated ambient pressure sensor, are indicated in the lower part of the display.
Also the differential pressure P
D
and the density r are indi-
cated.
Fig. 36: Measurement of the air speed
Page 33
33
• Press the left key to bring up the menu again.
• Press the right key to zero the pressure sensor.
NOTE!
The pressure sensor must be zeroed, if the meter does not indicate V=0 after the tubes have been removed.
6.5.2 Theoretical background
The PC 410 will calculate the air speed automatically according with the following formula:
v = S
p
2 ∆p
Formula 1
The air density r is determined as follows:
ρ =1,29 ·
kg
m³
273 °C .P
akt [hPa]
(ϑakt [°C]ϕ+273 °C) · 1013 hPa
Formula 2
Explanation: v: air velocity in m/s
r: air density, ca. 1.29 kg/m³ (depends on temperature and air pressure)
∆p: Differential pressure in Pa, measured with the S-tube
P
akt
: ambient air pressure, measured by the PC 410
S: Pitot factor 0,93 for S-tube probe
J
akt
: Measured Temperature
NOTE!
If the airflow reading is not stable or if the resolution is not high enough, go to the Setup-menu and select Attenuation > fine (see chapter 6.12)
Page 34
34
6.6 Option “4 PA Test” The 4 Pa Pressure Test is a simple control of the underpres- sure limit 4 Pa for the confirmation of a sufficient combustion air supply.
In a closed room or in connected rooms, simultaneous opera­tion of non-roomsealed heating appliances and ventilation and air conditioning systems can cause dangerous under pressure.
With the 4 Pa Pressure Test the underpressure value can be controlled. The readings are plotted in a graph during about 4 minutes, and can be saved and print out afterwards.
Fig. 37: PC 410 with capillary hoses
NOTE!
For the 4 Pa Test two capillary hoses must be connected to the PC 410.
The 4-Pa Test measures the pressure difference between the room where the fireplace is installed and another room which is not connected (e.g. outside air, stairway etc.). Before starting the measurement, two flexible capillary hoses must be connected (see accessories). The pressure sensor of the meter will measure the pressure difference.
Before starting the measurement:
1. Select 4 Pa Test in the main menu of the PC 410. The
current differential pressure reading will be displayed.
2. Connect the capillary hose leading to the reference point
(stairway or outside air) to the negative pressure socket (Fig. 1, part 10).
3. Connect the second capillary hose to the positive pres-
sure socket (Fig. 1, part 9). The second hose stays in the room where the fireplace is installed. With this hose the march of pressure will be recorded.
Underpressure in the fireplace room in relation to the pressure at the reference point will lead to a negative differential pres­sure.
Fig. 38: Initial display of the 4 Pa test
Capillary hose inside pressure
Capillary hose outside pressure
Page 35
35
How to perform the 4 Pa Pressure Test:
4. Switch on appliance and all air conditioning (fan, dryer)
with maximum power.
5. Open an outside window or a door to the reference room
and test proper operation of the appliance, ensure that there are no backdraft conditions.
6. Position the reference capillary hose: Lead the hose out-
side through a window seal or into the stairways through the door rebate or the keyhole.
Fig. 39: 4-Pa Test
NOTE!
Especially on stormy days, the stairways may be a stable reference room.
If the stairways are used as reference room, all windows, doors, cellar doors and trap doors have to be closed.
The second capillary hose will stay uncut in the fireplace room.
7. Press „PD =0“ to zero the pressure sensor.
8. Press "Start" to start the 4-PA Test.
The meter will now record the pressure profile for 4 minutes.
9. Open the window/door for about 30 seconds, so that the
zero line can be registered.
10. Close the window/door for about 30 seconds, control
underpressure.
11. Open the window/door for about 30 seconds, the zero
line should be reached again.
12. Close the window/door for about 30 seconds, control
underpressure.
13. Open the window/door for about 30 seconds, the zero
line should be reached again.
14. Close the window/door for about 30 seconds, control
underpressure.
For a better orientation, every 30 seconds there are auxil­iary lines in the diagram. After a maximum of 4 minutes, the measurement will stop automatically.
Fig. 40: Pressure profile during the 4 Pa Test
NOTE!
Press "Stop" to stop the 4 PA Test earlier.
Page 36
36
Normally a diagram as shown in Fig. 41. will appear. The pressure peaks in the diagram are caused by the rapid move­ment of the window or the door and therefore they are not relevant for the interpretation of the diagram.
In Fig. 41 the pressure drop is about 2,0 Pa.
The reliability of the system is considered suffi-cient, when the pressure drop caused by opening and closing the window is less or equal to 4 Pa (8 Pa in the case of wood burning without surrounding air).
After the 4 Pa test is completed, "4 Pa Test" will be marked by a tick mark in the main menu. Se-lect "Print" from the main menu to print out the readings and select "Save measure­ments" (see chapter 0) to save them.
For the detailed procedure according to the Technical Rules for Gas Installations (TRGI), working sheet G 625 (2009), it is necessary to replace the fireplace by the Wöhler DP 23 (see manual Wöhler DP 23).
Fig. 41: Diagram 4 Pa Test
Fig. 42: 4 Pa Test with Wöhler DP 23
Page 37
37
6.7 Option “Heating Check” (Energy Utilization Check)
The Energy Utilization Check provides a checklist to deter­mine the potential efficiency improvements.
Select the supmenu "Heating Check" from the main menu. In the Display the following options will appear:
• Flue gas loss
• Ventilation loss
• Surface loss
• Evaluation How to activate a submenu:
• Select the option with the up or down key.
• Confirm with "OK".
Potential Improvement Points
Flue gas loss
In the submenu "Flue gas loss" the most recently measured flue gas loss will be converted into potential improvement points according to DIN EN 15378, national annex.
How to convert the flue gas loss into potential improvement points:
• In the submenu "Heating check" select "flue gas loss".
Enter the measured flue gas loss.
The meter will convert the flue gas loss automatically accord­ing to the standard curve in Fig. 44.
After the conversion, "Flue gas loss" will be marked by a tick mark in the "Heating check" screen.
NOTE!
It is necessary to connect the S-tube probe (see acces­sories).
Fig. 44: Standard characteristic curve
Fig. 45: Entering the flue gas loss
Fig. 43: Energy Utilization Check
Page 38
38
Measuring the Ventilation Loss The ventilation loss has to be measured 30 seconds after
burn-out. It is based on the measurement of the flow speed and of the temperature of the residual core flow. The test fulfills the requirements of DIN EN 15378, national annex, concerning the evaluation of the ventilation loss of boilers.
How to perform the measurement:
Fig. 46: S-tube probe
NOTE!
A room temperature probe can be connected. Connect the Special Wall Temperature Probe (see accessories) to jack 12, fig. 1.
If you do not connect any room temperature probe, the tem­perature measured by the internal probe will be adopted.
• In the submenu "Heating check", select "ventilation loss".
Confirm with OK.
• Enter or control the parameters. (Nominal power, diameter
of the flue gas pipe, outside temperature and ambient pres­sure).
Press "Procced". A message will appear: "Zeroing pressure sensor. Remove hoses."
• Confirm with "OK". The pressure sensor will be zeroed. It is
important that the S-tube probe is not connected yet.
NOTE!
If the Lindab PC 410 is equipped with an internal ambient pressure probe, the meter will determine the ambient pres­sure automatically.
• Connect the S-tube probe to the Lindab PC 410.
• Insert the closed S-tube through the measurement opening
into the gas line, while the burner is working.
• Fix the S-tube with a cone, open it as shown in Fig. 47 and
direct the sensors as shown in the figure.
• If necessary, adjust the S-tube with the knurled screws, so
that it fits into the gas line.
NOTE!
Do not change the position of the meter after the pressure sensor has been zeroed. The sensor is extremely sensitive.
Fig. 48: Entering the parameters for the ventilation loss measurement
Fig. 47: Measuring the air speed with the S-tube
Static pressure
Temperature
Immersion depth
Knurled screws
Flow
Page 39
39
q
LS_Norm
=
·
Q
burner
Av · v(t) · ρ
AIR
· cpl · (ϑL(t) — ϑU)
273 + ϑ
outside ACTUAL
273 + ϑ
outside REF
The measurement will be done automatically.
• When the burner is turned off, press "Start" to start the
measurement. After 30 seconds, the data recording will start automati­cally. The captured values are shown in the display.
• Press "OK" to capture the calculated ventilation loss.
NOTE!
The S-tube must be adjusted before the measurement pro­cedure starts, since it is required to check and/or adjust the oppositely running orientation of the measuring tubes after loosening the knurled screws and adjusting the immersion depth
Fig. 49: Measuring the ventilation loss
NOTE!
The wanted signal of the S-tube probe is better than that of a normal pitot tube, if the flow speed is equal. In the setup­menu the pitot factor 0.93 must be set once.
The measured values are converted according to the follow­ing formula:
Description Data
q
LS_Norm
Ventilation loss in %
A
v
Area cross section of the flue gas pipe in m²
v(t) Air velocity in the flue gas pipe in m/s
r
AIR
air density, temperature compensated, e.g. 20 °C 1,2 kg/m³
c
pl
specific heat capacity, norm: 0,279 Wh/kg °C
J
L
air temperature in the exhaust stack
J
U
air temperature in ° in the fireplace room
J
outside ACTUAL
Outside temperature, actual value in °C
J
outside REF
Outside reference tem-perature, 15 °C
Q
burner
Nominal power of the heat generator
The conversion rule for the determination of the flue gas loss is integrated in the meter's program.
Page 40
40
After the measuring of the ventilation loss has been com­pleted, "ventilation loss" will be marked by a tick mark in the "Heating check" screen.
How to print out a report:
• Select the 'PRINT' option from the main menu. How TO save the results:
• Select the 'Save' option from the main menu.
Measuring the surface loss The surface loss is the temperature that the burner emits via
the surface.
NOTE!
The Special Wall Temperature Probe is needed to measure the surface loss (see accessories).
Before measuring the surface loss, it is necessary to enter certain data. For that purpose all partial areas of the burner must be considered that do not adjoin directly to other instal­lations and/or walls and floors
Determining Denomination A partial area = measured
value/manufacturer information
Partial area in m²
T
S
partial area = measured
value
Temperature of the partial area in °C
T
R
Room = measured value air temperature in ° in the
fireplace room
a = value determined according to EN 304
Heat transfer coefficient in W/m²K
Q
burner
= value indicated in the instruction plate of the burner
Nominal power of the heat generator in kW
Page 41
41
How to measure the surface loss:
• Enter the nominal power of the burner.
• Press "Proceed"
• Enter the surface temperature TS of the single areas and their dimensions (width x height).
• Measure the average surface temperatures with the Special Wall Temperature Probe connected to the Lindab PC 410 and accept it by pressing the “Accept” key.
• Proceed the same way for each partial surface.
The meter will calculate the surface loss qr (r: radiation). The surface loss will be referred to the heat emission and then converted into improvement potential points accord­ing to the DIN EN 15378, national appendix. The conversion into improvement potential points is done according to the graphic in Fig. 51: Converting the surface loss in potential improvement points.
After the measurement of the surface loss has been com­pleted, "Surface loss" will be marked by a tick mark in the "Heating check" screen.
Evaluation The menu item "Evaluation" presents all results and the corre-
sponding improvement potential points according to DIN EN
15378. These have to be registered in the Energy Utilization
Check certificate.
Fig. 51: Converting the surface loss in potential improvement points
Fig. 52: Evaluation of the heating check
Fig. 50: Measuring the surface loss
Page 42
42
6.8 Option „Volume“ In this submenu, the content of a closed hermetic room is determined (e.g. tank, bottle or pipe) with a volume up to 6000 l.
NOTE!
For the volume measurement, a a medical sy-ringe (up to 100 ml), a Wöhler Soot Test Pump, a cross-union and connection hoses are needed. We recommend the Wöhler Leakeage Test Set.
Measuring principle of the determination of the volume
After a sample V
sample
has been drawn with the syringe (up to 100 ml) or the soot test pump, the pressure in the tube will change. Based on this pressure change, the total volume V
tube
can be determined.
The Volume V
tube
is determined with the following equation in
accordance with the principle of the Boyle-Mariott's Law:
V
Tube
= V
Sample
· ∆p
P
act
-1
|Temp. = const.
V
tube
Tube volume to be determined, max. 6000 l
V
sample
Volume drawn as a sample
Dp
Maximum pressure difference that results when the sample is drawn
p
act
ambient pressure, has to be entered in the setup menu by the user (Default: 1013 hPa)
The Lindab PC 410 measures the pressure difference ∆p. For an exact measuring result, the pressure difference ∆p should at least amount to 200 Pa so that an exact result can be reached. Therefore, the sample V
sample
that the user draws with the syringe or the soot test pump should amount to at least 1/500 of the tube volume (see table).
Guidelines for the selection of the sample volume that has to be drawn with the soot test pump.
Pump volume Max. tube volume
163 ml (1 strike with the soot test pump)
80 l
489 ml (3 strikes with the soot test pump)
240 l
Page 43
43
Performing the test
Fig. 53: Volume measurement with the Lindab PC 410 and a syringe (up to 100 ml tube volume).
NOTE!
If the volume is smaller or equal 100 ml, a sy-ringe has to be used. If the volume exceeds 100 ml, a Soot Test Pump will be needed. The vo-lume of the Soot Test Pump is 163 ml per strike.
• Close the gas line and seal it with a suitable test plug.
• Switch the Lindab PC 410 on before connect-ing the over-
pressure socket (+) via a hose to the test plug.
• Connect the soot test pump or the syringe with a second
hose and the cross piece.
• Go to the main menu, select "Volume" and confirm with
OK.
• Enter the test volume into the meter according to the
sample volume that has to be drawn with the soot test pump or the syringe, e.g. 163 ml, that is one strike with the soot test pump.
WARNING!
Please observe the valid accident prevention regulations!
Fig. 54: Volume measurement with the Lindab PC 410 and a syringe (up to 100 ml tube volume).
Page 44
44
Fig. 55: Presettings (Volume test)
Fig. 56: Volume test result
• Press "Start".
In the display the request to add/remove the sample volume will appear.
• Take the sample volume with the syringe or the soot test
pump.
• When the value V keeps stable, press "OK to save the
data.
In the main menu a tick mark beside "Volume" will indicate that the volume test already has been done.
V
TST
= Volume
Test
PT = Pressure
Tube
P
amb
= Ambient Pressure
Page 45
45
6.9 Option “U-value” The heat transfer coefficient (U value) describes the heat flow W/m²k that flows through a building element.
Fig. 57: Application example with Special Wall Temperature Probe
Fig. 58: Display during the measurement of the U value
NOTE!
A Special Wall Temperature Probe is needed to measure the heat transfer coefficient (see accessories). All temperatures (also the inside air temperature tai and the outside air tem­perature tao should be measured with the same sensor.
The temperature values measured by the Lindab PC 410 and the Special Wall Temperature Probe are necessary for the determination of the heat transfer coefficient. The u-value is determined with the following equation according to the DIN standard 4108.
αi · [tLi – tWi]
[t
Li
– tLa]
U
ist
=
U
current
U-value in W/(m²k), current value
t
AI
interior air temperature
t
WI
interior wall temperature
a
i
fix value, 7,69 W/(m²k)
All temperatures (also the interior air temperature tai and the outside air temperature t
ao
should be measured with the
same sensor.
Page 46
46
6.10 Option “Graph / Record” The "graph / record" mode is used to start a permanent data recording. A graph of the records will appear on screen.
The user can modify the log interval, the number of measure­ments, the stop date and the stop time as follows:
• • Go to the next digit by pressing the left-/right-key
and increase or reduce the value by pressing the up- and down-keys.
The display can be switched off during the data recording.
Fig. 59: Submenu Logging
NOTE!
The display only will switch off between the measuring cycles, if a record interval of at least 20 seconds has been set.
• Select "Start recording" and press "Start"
A graph which presents the measured values appears in the display.
• Press "Stop" to save the logged data.
• Press "Print" to print a graph.
• Press "Proceed". The request " Accept measurements?"
will appear in the display.
• Press "No" to return to the main menu without saving the
measured values. In this case there will be no tick mark beside the menu "Graph / logging".
• Press "Yes" to return to the main menu after saving
the measured values. Beside the menu "Graph / Log­ging" there will appear a tick mark that indicates that the recorded data have been saved.
After the logging process has stopped select the 'save meas­urements' option in the main menu screen to save the logged data as a customer record.
After the logging process has stopped select the 'save meas­urements' option in the main menu screen to save the logged data as a customer record.
The data can be transfered to the PC with the Lindab Soft­ware PC 410 (see accessories).
NOTE!
Ensure that the meter is connected to the power supply or that the batteries are charged.
Page 47
47
6.11 Option „Print“ In the main menu, select "PRINT" to start a data printout of all measured values.
First, a print preview will appear on screen. In the Setupmenu, the user can select between printing with the thermal fast printer TD 600 or another printer.
The menu options in the preview screen are:
• Option "ESC" (left key): Click the left key to exit and return
to the main menu screen.
• Option „“ (center keys): Click either up or down key to
scroll up or down in the preview. Keep the key pressed to scroll various lines.
• Option "Print" (right key): Click the right key to start the
print out.
Fig. 60: Transmission of data records from the PC 410 to the thermal fast printer TD
600.
NOTE!
The menu item "Print" can only be selected, when meas­ured values have been accepted and a submenu is marked with a tick mark.
When printing is in progress, a status bar will display the cur­rent process state. Press "ESC" to stop the printout.
6.12 Option "Setup" Select the option "Setup" to enter the main setup menu.
• Use the arrow keys to select and change the parameters. The parameter that can be changed appears in red.
• Press "Return" to return to the main menu without saving
the modification or press the right arrow key to confirm the modification.
Fig. 61: Option "SETUP"
Option "Time" Change the current time of the internal clock (format 00:00)
Option "Date" Change the current date of the internal calendar (format
01.01.2011)
Option "Ambient Pressure" It is not necessary to enter an ambient pressure value,
because the meter is provided with an ambient pressure probe (only version AF).
Page 48
48
Air density Standard air density in kg/m3. This value will be consid-
ered for all calculations, if no external temperature probe is connected. When a temperature probe is connected, the air density will be calculated automatically.
Pressure unit Select the pressure unit hPa, Pa, mm/H2O, PSI, inwc, bar or
mbar. The preselected unit is hPa.
Pressure measurement Select “fine pressure measurement” (1 reading per second,
the attenuation is adjustable) "normal pressure measurement" (1 reading per second) and "fast pressure measurement" (4 readings per second).
When measuring the pressure and the flow the user can select “fine”, so that the resolution will be 0,01 PA (range: ± 100 Pa).
Attenuation If the user increases the attenuation, the signal that repre-
sents the measured value will be more stable. At the same time the readings display will become slower. The PC 410 will only consider the attenuation, if “fine pressure measurement” is selected in the setup-menu. (Setup>attenuation>fine pres­sure measurement.
Temperature unit Select the temperature unit °C or °F.
Air flow unit Select between l/s, m3/h, m3/s and cfm. The air flow value
will be indicated in the center of the display in the unit that is selected first (here: l/s) and below, smaller, in the unit that is selected afterwards (here: m3/h).
Fig. 62: Selecting the airflow unit
K-value Depending on the manufacturer of the measurement equip-
ment, the user can choose between different formula for the determination of the volume flow (see chapter 6.2.1).
K Value Input Select between "Digits" (the user selects the digits that he
wants to change) and "Fast" (the user changes the k-value by selecting it and simply pressing the up or down keys)
Exponent Here the exponent of the formula of chapter 6.2 (Air flow (k
value)) is entered. Per default the exponent is 0.50. The expo­nent can be risen from 0.50 to 1.00. It will only be necessary to change the exponent, if another exponent is specified in the data sheet of the air vent.
Brightness Adjust the brightness of the display from 20 % to 90 %.
Printer Here the user can select between printing with the thermal
fast printer TD 600 or another printer.
Page 49
49
Print graph Option "Yes": A graph will be printed with the diagram.
Option "No": No graph will be printed. The preselected option is "Yes".
Search by Select between "customer name" and "customer number".
Depending on this selection, the name or the number of the customer can be selected in the customer menu.
Pitot factor The pitot factor is important for the measurement of the flow
speed. It depends on the cross section of the utilized s-tube. The pitot factor of the Wöhler S-tube probe is 0.93. Therefore the preselected value is 0.93.
If a “normal” S-tube is connected, the pitot factor 1.0 has to be entered.
Thermal capacity The default setting of the specific heat capacity is 0,2790
Wh/kg °C. The value is important for the measurement of the ventilation loss (heating check)
Reference temperature The reference temperature can be adjusted from - 30 °C to +
70 °C.
Alpha Heat transmission coefficient for the u-value measurement
Language Depending on the version, the languages English, German or
other languages can be selected.
Printer logo Enter a custom corporate logo (6 lines) that will appear in
every printout.
Default Reset the PC 410 to its factory default setting (except the
calibration).
6.13 Option "Save measurements" To save all records which are marked by a tick mark in the main menu, select the "Save measurement" option.
Page 50
50
7. Option 'Data management'
When working with the Lindab PC 410 it is possible to save and manage customized data which can be assigned to different ducts. The user can create a folder for every customer and assign up to 128 ducts to each one.
7.1.1 Save Customer Records When different measurements have been per-formed at one installation, they can be assigned to a customer as follows:
• Select the 'Save measurements' option from the main
menu.
• Use the up- and down keys to navigate through the menu.
If there exists no folder of the costumer, whose data shall be saved, it can be created in the "new customer" menu (see chapter 7.1.2).
Fig. 63: Customer selection
Fig. 64: Duct selection
NOTE!
Keep the up or down key depressed to scroll faster.
• Confirm with "OK". The duct list will appear. Select "New duct" to assign a new duct to this client.
• Confirm with "OK".
NOTE!
The records marked with a tick mark in the main menu are saved.
After the records have been successfully saved, a lock will appear instead of the tick mark.
NOTE!
During the saving process, the data that has al-ready been saved before under this duct, will be deleted.
Page 51
51
How to add measurement data to the data that has already been saved under a duct:
• Press "Customer" directly after switching on the meter,
during the zeroing-phase. Then select the duct, where you want to add the new data.
In the main menu, the saved data will be marked by a lock and the new measurements that have not been saved yet, will be marked by a tick mark.
Save the data as described.
7.1.2 Creating a New Cus-tomer Folder
The user can create new customer folders or new ducts in the customer menu (see Fehler! Ver-weisquelle konnte nicht gefunden werden.) or when saving the data. Enter the name, a customer number and the name of the duct.
Fig. 65: customer option
Fig. 66: Creating a new customer folder
NOTE!
The user can save all together 128 ducts in the meter and assign them to different customers. The number of ducts assigned to a customer is arbitrary.
7.2 Option "Data management" In the "Data management" display, the current number of customer folders and ducts is indi-cated.
Select "Print protocol" to print a report of any saved meas­urement.
Select "Delete duct" to delete a single duct.
Fig. 67: USB data transfer
Select "Delete customer" to delete the complete customer folder with all ducts.
Select "Delete all customers" to delete all customer folders.
Select "USB data exchange" to communicate with the com­puter, see chapter 7.3.
NOTE!
When there is only one duct saved in a customer folder, the customer folder will be deleted when the duct is deleted.
Page 52
52
7.3 Data Transfer with the Computer or Notebook
The user can easily transfer data from the PC 410 to the Computer or notebook via an USB cable. How to transfer data to the Computer:
• Connect the Lindab PC 410 to the Computer via the USB
cable included in delivery.
• Select Main menu < Data management > USB data
exchange
The USB mode of the Lindab PC 410 is activated now.
• Open the Data Management Programm PC 410 on the
computer. Follow the instructions of the PC 410 Software.
Fig. 68: USB mode
7.4 Data Logging With the Lindab PC 410 Software data logging is possible.
8. Troubleshooting
Error message Possible reason Solution
Batteries low Disposable or rechargeable
batteries are empty.
Replace or recharge batteries.
ATTENTION! Overpressure High pressure Deflate pressure
Page 53
53
9. Maintenance
Proper operation of the Lindab PC 410 requires regular maintenance.
9.1 Maintenance work
Interval Maintenance work
If required (user) Clean the device with a wet
cloth and a smooth cleanser.
Once a year (factory) Send the device to one of our
service points for checking and calibration.
10. Warranty and Service
10.1 Warranty Each Lindab PC 410 probe will be tested in all functions and will leave our factory only after extensive quality control tes­ting. The final control will be recorded in detail in a test report and delivered with any unit.
If used properly, the warranty period for the Lin-dab PC 410 will be twelve month from the date of sale. Not covered by the warranty are batteries and defects of the pressure sensor caused by overpressure.
This warranty does not cover the freight and packing costs when the device is sent to the factory for repair.
Service by non authorized personnel or making modifications to the analyzer voids any warranty.
Page 54
54
11. Accessories
Probes and sensors
Prandl tube, 1000 mm PCPRA
Capillary tubes 4 Pa Test PCKAP
S-tube probe PCSRS
Measuring pliers for temperature measurement PCZA500
Special Wall Temperature Probe PCHC
Wall Temperature Probe PCTF1
Penetration Temperature Probe PCTF4
Combustion and Temperature Probe PCTF5
Extras
Rechargeable batteries NiMH, 2 Ah PCAKKU
Charger PCLGA400
Thermal Fast Printer TD 600 TD 600
Thermal Paper, 10 rolls LTPP
Software
PC 410 Software with USB cable. PC410DAT
Page 55
55
12. Declaration of conformity
The manufacturer:
Wöhler Messgeräte Kehrgeräte GmbH
Schützenstr. 41, D-33181 Bad Wünnenberg
declares that the product
product name: Flow and Pressure Meter
model number: Lindab PC 410
complies with the following protective requirements:
• requirements for electromagnetic compatibility in accordance with EN 61326-1 1997+
A1: 1998+A2:2001
• requirements according to DIN EN 61010-1:2002
• the essential requirements for the protective system in accordance with EN 60529:1991
certified by TÜV SÜD
• Directive "Richtlinie für die Eignungsprüfung von Differenzdruckmessgeräten zur Messung von
Unterdrücken in Aufstellungsräumen von Feuerstätten“ (ZIV, Dez. 2007)"
• Requirements "Mindestanforderungen an geeignete Messgeräte zur Messung des Abgas-Ventila-
tions- und Strahlungsverlustes im Rahmen der energetischen Inspektion, Entwurf, ZIV, Stand 14. Januar 2010"
This declaration is submitted for the manufactorer by:
Dr. Stephan Ester, managing director
Bad Wünnenberg, 22nd of october 2010
13. Lindab local
Contact your local Lindab distributor or find our representatives on www.lindab.com
Page 56
1072 EN 2011-11-04
Loading...