Linak MJB User Manual

Page 1
USER MANUAL
MJB - Gateway
To learn more about LINAK please visit:
WWW.LINAK.COM
Page 2
2
Page 3
Contents
Preface .................................................................................................................................................................. 4
Important information ........................................................................................................................................ 5
Safety instructions ............................................................................................................................................... 5
Residual risk ......................................................................................................................................................... 6
Before installation, reinstallation or troubleshooting ...................................................................................... 7
Before start-up ..................................................................................................................................................... 7
During operation ................................................................................................................................................. 7
Information on the MJB ...................................................................................................................................... 7
Repairs .................................................................................................................................................................. 8
Misc. on the MJB system .................................................................................................................................... 9
Maintenance ................................................................................................................................................... 9
MJB - GATEWAY variant .................................................................................................................................... 10
Standard functionality .................................................................................................................................... 10
Relay 1 and 2 ................................................................................................................................................ 10
Features ......................................................................................................................................................... 11
Usage ............................................................................................................................................................ 11
Precautions ......................................................................................................................................................... 13
IP Protection degree .......................................................................................................................................... 15
Keys to symbols ................................................................................................................................................. 18
Drawing appendix ............................................................................................................................................. 20
Addresses ........................................................................................................................................................ 24
3
Page 4
Preface
We are delighted that you have chosen a product from LINAK®.
LINAK systems are high-tech products based on many years of experience in the manufacture and development of actuators, electric control boxes, controls and chargers.
This User Manual will tell you how to install, use and maintain your LINAK equipment.
We are sure that your LINAK system will give you many years of problem-free operation. Before our products leave the factory they undergo full function and quality testing. Should you nevertheless experience problems with your systems, you are always welcome to contact your local dealer.
LINAK subsidiaries and some distributors situated all over the world have authorised service centres, which are always ready to help you.
LINAK provides a warranty on all its products. This warranty, however, is subject to correct use in accordance with the specifi cations, maintenance being done correctly and any repairs being carried out at a service centre, which is authorised to repair LINAK products.
Changes in installation and use of LINAK systems can affect their operation and durability. The products are not to be opened by unauthorised personnel. The User Manual has been written on the basis of our present technical knowledge. We are constantly working on updating the information and we therefore reserve the right to carry out technical modifi cations.
LINAK A/S
4
Page 5
Important information
Description of the various signs used in this manual. Warning!
Failure to comply with these instructions may result in accidents involving serious personal injury.
Failing to follow these instructions can result in the product being damaged or destroyed.
Safety instructions
Safe use of the system is possible only when the operating instructions are read completely and the instructions contained are strictly observed. Failure to comply with instructions marked with the”NOTE” symbol may result in serious damage to the system or one of its components. Persons who do not have the necessary experience or knowledge of the product/ products must not use the product/ products. Besides, persons with reduced physical or mental abilities must not use the product/products, unless they are under surveillance or they have been thoroughly instructed in the use of the apparatus by a person who is responsible for the safety of these persons. Moreover, children must be under surveillance to ensure that they do not play with the product.
It is important for everyone who is to connect, install, or use the systems to have the necessary information and access to this User Manual.
If there is visible damage to the product it must not be installed.
The appliance is not intended for use by young children or infi rm persons without supervision.
Young children should be supervised to ensure that they do not play with the appliance.
5
Page 6
Residual risk
Some of the products contains software based components. LINAK has made various possible efforts to assure that the software is free of errors and that the software has been developed according to the rules of IEC 60601-1-4 (software in Medical products). That involves Risk Analysis which shows a small residual risk for unwanted/unintended movement of actuators under specifi c conditions. According to the above rules it must be informed and if necessary considered in the Risk analysis of the fi nal application - More details to residual risk can be provided by LINAK if necessary.
Warning!
The plastic parts in the system cannot tolerate cutting oil.
Warning!
LINAK’s actuators and electronics are not constructed for use within the following fi elds:
• Planes and other aircrafts
• Explosive environments
• Nuclear power generation
Warning!
If faults are observed, the products must be replaced.
Warning!
Never spray directly on the products with a high-pressure cleaner.
Warning!
Always insure that the patient is accessible in case of emergency.
Warning!
Cables must be secured safely so that persons cannot get trapped in it.
Warning!
Always use locking mechanism and O-ring
Warning!
Sockets not used must be fi tted with blind plugs to ensure IP degree
The LINAK products cannot tolerate the infl uence of strong solvents, basic or alkaline liquids.
6
Page 7
Before installation, re-installation, or troubleshooting
• Stop the application
• Switch off the power supply and pull out the mains plug.
• Relieve the application of any loads, which may be released during the work.
Before start-up
• Make sure that the system has been installed as instructed in the User Manual.
• Make sure that the voltage of the control box is correct before the system is connected to the mains.
• System connection. The individual parts must be connected before the control box is connected to the mains.
During operation
• Ensure that the cables are not damaged.
• Unplug the mains cable on mobile equipment before it is moved.
Information on the MJB
The MJB Modular Junction Box is designed for use together with OpenBus™ control boxes. The MJB makes it possible to connect multiple handsets, attendant controls or it makes it possible to use the MJB as a control unit for 3rd party products as Out of Bed detection, USB charger, Under Bed Light etc.
LINAK A/S delivers only the OpenBus system and is not responsible for any products other than LINAK products (i.e. products from 3rd party suppliers) or the compliance of such products with the LINAK OpenBusTM system.
7
Page 8
Repairs
In order to avoid the risk of malfunction, all repairs must only be carried out by authorised LINAK workshops or repairers, as special tools must be used and special gaskets must be fi tted. Products under warranty must also be returned to authorised LINAK workshops.
Warning!
If any of the LINAK products are opened, there will be a risk of subsequent malfunction.
Disposal of LINAK’s products
LINAK’s products may be disposed of, possibly by dividing them into different waste groups for recycling or combustion.
We recommend that our product is disassembled as much as possible at the disposal and that you try to recycle it. As an example of main groups within sorting of waste we can mention the following:
Metal, plastic, cable scrap, combustible material and collection for recoverable resources.
Some of these main groups can be sub-divided into groups e.g. metal can be divided into iron, stainless steel and aluminium and alloy steel. Plastic can e.g. be divided into ABS, PA, PE and PP.
The MJB can be disassembled and sorted into the following recycling groups:
Component Recycling group
Plastic housing – Plastic recycling or combustion
MJB
By now all our casted plastic units are provided with an interior code for plastic types and fi bre contents, if any.
Cable – Cable scrap or combustion
PC-board – Electronics scrap
8
Page 9
Misc. on the MJB system
Warranty and service life
The LINAK warranty covers manufacturing defects in the products, starting from the date of manufacture. There is 18 months’ warranty for MEDLINE and CARELINE products. The warranty is limited to the value of the LINAK product.
LINAK’s guarantee is only valid so far as the products have been used and maintained correctly and has not been tampered with. Furthermore, the products must not be exposed to violent treatment. In the event of this, the warranty will be ineffective / invalid. LINAK’s warranty is only valid if the system is unopened and has been used correctly.
All LINAK products are designed to have an optimum service life as a matter of course, but the expected service life in a specifi c application is very dependent on how the products are used.
Maintenance
• The LINAK products must be cleaned at regular intervals to remove dust and dirt and inspected for mechanical damage, wear and breaks.
• The LINAK MJB is a closed unit and requires no internal maintenance.
• IPX6 is waterproof.
• Make sure that the plugs in the MJB are correctly fi tted with O-rings before washing.
• Electronics must be inspected at attachment points, wires, cabinet, and plugs.
• Inspect the connections, cables, cabinet and plugs, and check for correct functioning.
9
Page 10
MJB - GATEWAY variant, 504-01 & 504-02 :
The MJB is a simple gateway interface, which can connect the bed to the infrastructure of the hospital or nursing homes. The Gateway can be used in combination with an Out of Bed Sensor, which makes it possible to indicate if the patient leaves the bed. The Gateway can also be used as a service/nurse call system. Please note it is only to be used as a guided system, not for emergency issues.
Port 5: Relay connetions
Relay 1:
Port 5
Relay 2:
Port 1
OpenBus connetion
Standard functionality:
The MJB - Gateway consist of two relays, which are connected through Port 5 on the MJB. This can be made with the special cable 0964140 (Modular plug - Open end) There are as standard 2 confi gurations, 504-01 and 504-02, see description on the next pages.
- Pin 1
- Pin 2
- Pin 3
- Pin 4
- Pin 5
- Pin 6
NC
NO
NC
NO
Relay1: NC (normally closed) = Pin 2 + Pin 1 / NO (normally open) = Pin 2 + Pin 3 Relay2: NC (normally closed) = Pin 5 + Pin 4 / NO (normally open) = Pin 5 + Pin 6
Relay 1 is NO when connected to the mains and NC when no mains, this means the relay is “active”, when connected to the mains (closed loop). The Closed Loop principal is to ensure that an alarm is sent if power is missing on the bed. When the mains is disconnected (power is missing), the relay will go from NO to NC, which will automatically result in an alarm (status indicator) on the OpenBus™. Relay 1 can be activated via the patient control (Key1 / Key4) or the attendant control (Key2 / Key3). When a key is activated, the relay will switch state from NO to NC for 2 seconds. After 2 seconds the relay will automatically change state from NC to NO. The status of the relay is indicated on the OpenBus™ and can be used for switching on an LED.
Relay 2 is as default NC, with or without mains, (Open Loop) Relay 2 can be activated via the patient control (Key 1) or the attendant control (Key 2 / Key 3). When the relay is activated, it will switch from NC to NO for 2 seconds. After 2 seconds the relay will automatically change state from NO to NC. The status of the relay is indicated on the OpenBus™ and can be used for switching on an LED.
10
Page 11
By using the variant 504-02 relay 2 will be controlled by the MJB variant 505/506 with switch input. Alarm 1 = Switch input S2 (OoBD alarm) / Alarm 2 = Switch input S1 When using this combination, is it important to have the MJB 505/506 connected to the system all the time. If it is not connected, the Gateway MJB will see it as an alarm and the relay will be activated. The Gateway MJB is “scanning” the OpenBus™ system every minute for an alarm.
Features:
• Add-on value for hospital & care bed
• Compact and functional design which allows easy mounting
• 2 connections, Port1 OpenBus™ connection / Port 5 Gateway
• Colour: Light grey RAL 7035
• Protection class IPX6
• Connection to OpenBus™ control box via modular plug cable, 6 wires
• Modular plug to Open end, see precautions.
Options (special article):
• Customised SW:
It is possible to defi ne the OpenBus™ signals to be used for activating the relays and / or change the alarm indicator, so that it fi ts to the customers system. Please contact your Sales Engineer at LINAK A/S for further information.
Usage:
• Compatible with all LINAK OpenBus™ Control Boxes, light & full Contact rating :
• Max continuous current : = 1A
• Maximum switching voltage : = 125V (DC) / 125V (AC)
• Maximum switching capacity : = 24W (DC) / 62,5VA (AC)
• Contact resistance : < 100 m
Current consumption:
• V bus 8V : = 9 mA
• V permanent 40V : = 14 mA
• Approvals: LINAK and UL International DEMKO A/S are currently in the process of certifying the MJBX0X, re. UL60601 and EN60601-1
11
Page 12
Relay 1
CB power mode Alarm level Relay state
Mains unplugged, CB is in power down or the OpenBus™ is not
Alarm NC\*
running (Clock/ data is missing)
On mains or battery at “wake up” Bed Alarm\** NC
No Alarm NO
Relay 2
CB power mode Alarm level Relay state
Mains unplugged or CB is in power down No Alarm NC
On mains or battery at “wake up” Bed Alarm\** NO
No Alarm NC
\* When the mains is unplugged or CB is in power down will the relay shift to NC state and generate alarm. \** Alarm generated by nurse call or bed alarms
Example of closed loop:
3rd. party site
In this state no Alarm
Example of Open Loop:
3rd. party site
In this state no Alarm
12
Page 13
Precautions:
• Always use locking mechanism and O-ring
• Sockets not used must be fi tted with blind plugs to ensure IP Degree
• When using the modular plug with open end, is it up to the customer to maintain the IP degree.
• If the MJB – Gateway is used as open loop, there is a risk of not sending an alarm, when no mains are connected. The MJB – Gateway will only send an alarm if mains is missing, when using a Closed Loop (see functionality description.)
• When the Gateway is used on a system with battery, the gateway functionality will follow the power down mode of the control box, see table
CB Power mode Gateway
On mains Function ok
Power down On battery “Wake Up”
No function
Function ok
• The MJB – Gateway is not to be used as safety, it is only to be used as a guided system.
• It is important to test the specifi ed alarms is working in the correct way, before sending the system to the end user.
• Do not use 2 MJB variants with same device ID on the OpenBus™. This will cause confl icts and the SDT is not able to identify the different products attached.
• Before the fi nal functional test in the production, it is important that the system is re-powered. This is to ensure that all items have been detected on the OpenBus™.
• We recommend the end user to make a regular test procedure in order to prevent failures and hazardous situations on the system, e.g. defective OOB
rd
sensor, squeezed cables. The MJB is not able to detect defective 3
. party products.
• LINAK only takes responsibility for LINAK products, not 3rd. party products. Please pay attention to the “Patient Environment” Clause 3.79 - EN60601-1 3rd edition: Any volume in which contact can occur between a patient and parts of the medical equipment or between a patient and other persons touching parts of the equipment. Justifi ed practice here is 1.5 m. This volume must be subject to the Risk Analysis. It is important to inform the customer about this.
• There can be a risk of confl ict with other OpenBus™ accessories, such as HB, ACP,
etc. It is therefore recommended to make a system/bit overview.
13
Page 14
Cable ordering nos.:
Modular plug – open end: 0964140-3000
Please note, the slit is placed on the opposite side of the modular plug. This is to ensure that the cable is only being used for the Gateway MJB.
14
Page 15
IP Protection degree:
All LINAK products can be cleaned as follow according to their IP protection, which is stated on the product label. The IP code specifi es the degrees of protection provided by the enclosures. Only the protection against ingress of water (second characteristic numeral) is specifi ed, ingress of solid foreign objects (fi rst characteristic numeral) is not specifi ed and therefore replaced by the letter X in the code.
IP Protection Cleaning instrucions
IPX0 Clean with a damp cloth
IPX1 Clean with a damp cloth
IPX2 Clean with a damp cloth
IPX3 Clean with a damp cloth
IPX4 Clean with a damp cloth
IPX5 Wash with a brush and water, but not water under pressure
IPX6 X Wash with a brush and water. The water can be under pressure, but the system
IPX6 Washable Clean by the use of water tunnels
The materials are resistant to the majority of cleaners and disinfectants used in the hospital and nursing home sector.
Cleaning with a steam cleaner, for example, is not permitted, as it will not be possible to maintain a minimum distance of 300 mm from the electrical parts.
must not be hosed down directly with a high-pressure cleaner. Max. 20oC
The following directions regarding cleaners and disinfectants must be complied with:
• They must not be highly alkaline or acidic (pH value 6-8).
• They must not contain caustic agents.
• Their contents must not be able to change the structure of the surface or
adhesion of the plastic.
• They must not break down grease.
The following disinfectants/detergents are recommended:
• Neodischer Dekonta, dosage 1-3%, made by CHEMISCHE FABRIK Dr. WEIGERT,
• CCOTRADE RW, dosage 0.5%, made by CCOTRADE GmbH.
15
Page 16
Warning!
The systems must not be sprayed directly with a high-pressure cleaner.
Warning!
Interconnecting cables must remain plugged in during cleaning to prevent the ingress of water.
Precaution!
When cleaned, water must never be sprayed upwards from underneath the bed, because it may enter the cable outlet.
Environmental conditions:
Environmental conditions
Operating:
Temperature Relative humidity Atmospheric pressure
Storage:
Temperature Relative humidity Atmospheric pressure
Valid where nothing otherwise is stated under the specifi c products in a later section.
5°C to 40°C 20% to 90% @ 30°C – not condensing 700 to 1060 hPa
-10°C to +50°C 20% to 90% @ 30°C – not condensing 700 to 1060 hPa
16
Page 17
Modular plug with wide slit
To be used with LINAK products.
Can be connected to all ports in the MJB,
both ports with narrow and wide slit.
Modular plug with narrow slit
Can only be connected to ports in the MJB with
narrow slit. This is to prevent 3rd party products
to interfere with the OpenBusTM connections.
To be used with 3rd party products.
17
Page 18
Key to symbols
The following symbols are used on the label on the LINAK products.
Type B equipment, as per EN 60601-1
Protection against contact/foreign matter (first character) and water
IPXX
(second character) as per EN60529
Class 2 equipment
For indoor use
Double insulated transformer
Protective earth
Alternating Current
Direct current
Attention, consult accompanying documents
Lock function
Release function
Charge indicator
Safety switch/enable button
18
Page 19
Demko approval
UL fi le number
Fimko approval
Australian approval mark
Australian approval mark
Recognised - Component Mark
Canadian Recognised - Component Mark
Recognised Component Mark for Canada and the United States
T-Mark
RW-Tüv approval
Tüv.Produkt Service
TÜVRheinland
UL fi le number
UL fi le number
CSA
PSE-Mark
Product with a thermofuse
For indoor use (House).
Safety isolating transformer.
Electronics scrap
Equipment Kl.2 (Double square)
Patient part of type B (Mand)
Patient part of type BF
ETL
C-ETL
LGA
UL Listing Mark
C-UL Listing Mark
C-UL US Listing Mark
UL Listing Mark
Earth protection
Equipment class1.
Earth
CE Mark
C-TICK
19
Page 20
DRAWING APPENDIX
MJB - Modular Junction Box
MJB
20
Page 21
LINAK APPLICATION POLICY
The purpose of the application policy is to defi ne areas of responsibilities in relation to applying a LINAK product defi ned as hardware, software, technical advice, etc. related to an existing or new customer application.
LINAK products as defi ned above are applicable for a wide range of applications within the Care and Health, Comfort furniture, Desk and Industry areas. Yet, LINAK cannot know all the conditions under which LINAK products will be installed, used, and operated, as each individual application is unique.
The suitability and functionality of the LINAK product and its performance under varying conditions (application, vibration, load, humidity, temperature, frequency, etc.) can only be verifi ed by testing, and shall ultimately be the responsibility of the LINAK customer using any LINAK product.
It is also the responsibility of the LINAK customer to make and supply a comprehensive user manual of the application.
LINAK shall be responsible solely that the LINAK products comply with the specifi cations set out by LINAK and it shall be the responsibility of the LINAK customer to ensure that the specifi c LINAK product can be used for the application in question.
21
Page 22
22
Page 23
23
Page 24
FACTORIES
CHINA
LINAK (Shenzhen) Actuator Systems, Ltd. Phone: +86 755 8610 6656 . Fax: +86 755 8610 6990 E-mail: [email protected] . www.linak.cn
DENMARK
LINAK A/S . Group Headquarters Guderup Phone: +45 73 15 15 15 . Fax: +45 74 45 80 48 Fax (Sales): +45 73 15 16 13 E-mail: [email protected] . www.linak.com
SLOVAKIA
LINAK Slovakia s.r.o. Phone: +421 517563 414 . Fax: +421 517563 410 E-mail: [email protected] . www.linak.com
USA
LINAK U.S. Inc. North and South American Headquarters Phone: +1 502 253 5595 . Fax: +1 502 253 5596 E-mail: [email protected] . www.linak-us.com
SUBSIDIARIES
AUSTRALIA
LINAK Australia Pty. Ltd Phone: +61 3 8796 9777 . Fax: +61 3 8796 9778 E-mail: [email protected] . www.linak.com.au
AUSTRIA
LINAK GmbH - Austria Phone: +43 (0) 2266 61 9 62 . Fax: +43 (0) 2266 63 2 99 E-mail: [email protected] . www.linak.at
BELGIUM & LUXEMBOURG
LINAK Actuator-Systems NV/SA Phone: +32 (0)9 230 01 09 . Fax: +32 (0)9 230 88 80 E-mail: [email protected] . www.linak.be
BRAZIL
LINAK DO BRASIL COMÉRCIO DE ATUADORES LTDA. Phone: +55 (11) 2832-7070 . Fax: +55 (11) 2832-7060 E-mail: [email protected] . www.linak.com.br
CANADA
LINAK Canada Inc. Phone: +1 502 253 5595 . Fax: +1 905 821 4281 E-mail: [email protected] . www.linak.ca
CZECH REPUBLIC
LINAK C&S S.R.O. Phone: +420581741814 . Fax: +420581702452 E-mail: [email protected] . www.linak.cz
DENMARK
LINAK DANMARK A/S Phone: +45 86 80 36 11 . Fax: +45 86 82 90 51 E-mail: [email protected] . www.linak.dk
FINLAND
LINAK OY Phone: +358 10 841 8700 . Fax: +358 10 841 8729 E-mail: [email protected] . www.linak.fi
FRANCE
LINAK FRANCE E.U.R.L Phone: +33 (0)2 4136 3434 . Fax: +33 (0)2 4136 3500 E-mail: [email protected] . www.linak.fr
GERMANY
LINAK GmbH Phone: +49 6043 9655 0 . Fax: +49 6043 9655 60 E-mail: [email protected] . www.linak.de
INDIA
LINAK A/S Asian Representative Office Phone: +91 120 4797013 E-mail: [email protected] . www.linak.in
IRELAND
LINAK UK Limited Phone: +44(0)121 544 2211 . Fax: +44(0)121 544 2552 E-mail: [email protected] . www.linak.co.uk
ITALY
LINAK ITALIA S.r.l. Phone: +39 02 48 46 33 66 . Fax: +39 02 48 46 82 52 E-mail: [email protected] . www.linak.it
JAPAN
LINAK K.K. Phone: +81 45 533 0802 . Fax: +81 45 533 0803 E-mail: [email protected] . www.linak.jp
MALAYSIA
LINAK A/S Asian Representative Office Phone: +60 4 210 6500 . Fax: +60 4 226 8901 E-mail: [email protected]
.
www.linak-asia.com
NETHERLANDS
LINAK Actuator-Systems B.V. Phone: +31 76 5 42 44 40 . Fax: +31 76 5 42 61 10 E-mail: [email protected] . www.linak.nl
NEW ZEALAND
LINAK New Zealand Ltd. Phone: +64 9580 2071 . Fax: +64 9580 2072 E-mail: [email protected] . www.linak.co.nz
NORWAY
LINAK Norge AS Phone: +47 32 82 90 90 . Fax: +47 32 82 90 98 E-mail: [email protected] . www.linak.no
POLAND
LINAK Polska Phone: +48 (22) 500 28 74 . Fax: +48 (22) 500 28 75 E-mail: [email protected] . www.linak.pl
SPAIN
LINAK Actuadores, S.L.u Phone: +34 93 588 27 77 . Fax: +34 93 588 27 85 E-mail: [email protected] . www.linak.es
SWEDEN
LINAK Scandinavia AB Phone: +46 8 732 20 00 . Fax: +46 8 732 20 50 E-mail: [email protected] . www.linak.se
SWITZERLAND
LINAK AG Phone: +41 43 388 31 88 . Fax: +41 43 388 31 87 E-mail: [email protected] . www.linak.ch
TAIWAN
LINAK A/S Taiwan Representative Office Phone: +886 2 250 80296 . Fax: +886 2 2508 3604 E-mail: [email protected] . www.linak.com.tw
TURKEY
LINAK A/S Turkey Representative Office Phone: +90 312 4726338-59 . Fax: +90 312 4726635 E-mail: [email protected] . www.linak.com.tr
UNITED KINGDOM
LINAK UK Limited Phone: +44(0)121 544 2211 . Fax: +44(0)121 544 2552 E-mail: [email protected] . www.linak.co.uk
DISTRIBUTORS
ARGENTINA
NOVOTEC ARGENTINA SRL Phone: +[54] (11) 4303-8900/89 E-mail: [email protected] www.novotecargentina.com
COLOMBIA
MEM Ltda Phone: +[57] (1) 334-7666 E-mail: [email protected] www.memltda.com.co
INDONESIA
PT. HIMALAYA EVEREST JAYA Phone: +6 221 544 8956/65 E-mail: [email protected] . www.hej.co.id
.
Fax: +[54] (11) 4032-0184
.
Fax: +[57] (1) 282-1684
.
Fax: +6 221 619 4658/1925
IRAN
Bod Inc. Phone: +98 2188998635 - 6 E-mail: [email protected]
MEXICO
ILSA S.A. de C.V. Phone: +[52] (55) 5388-3960 E-mail: [email protected]
PERU
PERCY MARTIN DEL AGUILA UBILLUS Phone: +51 99- 883-9879
RUSSIAN FEDERATION
000 FAM Phone: +7 812 3319333 E-mail: [email protected]
SINGAPORE
SERVO DYNAMICS PTE. Ltd. Phone: +65 6844 0288 E-mail: [email protected]
SOUTH AFRICA
Industrial Specialised Applications CC Phone: +27 11 312 2292 or +27 11 2077600 Fax: +27 11 315 6999 E-mail: [email protected]
SOUTH KOREA
UNITEK ENG. Phone: +82 2 567 0888 E-mail: [email protected]
UNITED ARAB EMIRATES
Mechatronics Phone.: +971 4 267 4311 E-mail: [email protected] www.mechatronics.ae
For contact details on other countries please visit www.linak.com or contact:
LINAK INTERNATIONAL
Fax: +45 74 45 90 10 E-mail: [email protected]
Terms of use
The user is responsible for determining the suitability of LINAK products for specifi c application. LINAK takes great care in providing accurate and up-to-date information on its products. However, due to continuous development in order to improve its products, LINAK products are subject to frequent modifi cations and changes without prior notice. Therefore, LINAK cannot guarantee the correct and actual status of said information on its products. While LINAK uses its best efforts to fulfi l orders, LINAK cannot, for the same reasons as mentioned above, guarantee the availability of any particular product. Therefore, LINAK reserves the right to discontinue the sale of any product displayed on its website or listed in its catalogues or other written material drawn up by LINAK. All sales are subject to the Standard Terms of Sale and Delivery for LINAK. For a copy hereof, please contact LINAK.
.
Fax: +98 2188954481
.
www.bod.ir
.
Fax: +[52] (55) 5388-3966
.
www.ilsamexico.com
.
Fax: +7 812 3271454
.
.
Fax: +65 6844 0070
.
www.isaza.co.za
.
Fax: +82 2 3453 1177
.
Fax: +971 4 267 4312
.
www.linak.com
www.fam-drive.ru
.
www.servo.com.sg
.
www.unitekeng.co.kr
LINAK 2010.10 MA-M9-02-348 LINAK A/S reserve the right to make technical alterations
©
Copyright
Loading...