The Safety and Warranty Information section provides details about cautionary
symbols used in the manual, safety markings used on the instrument, and
information about the Warranty including Customer Service contact information.
Safety Information and the Manual
Throughout this manual, you will see the words Caution and Warning indicating
potentially dangerous or hazardous situations that, if not avoided, could result in
death, serious or minor injury, or damage to the product. Specifically:
Caution indicates a potentially hazardous situation that can result in minor or moderate
injury or damage to the product or equipment.
Warning indicates a potentially dangerous situation that can result in serious injury or
death.
WARNING
Visible and/or invisible laser radiation. Avoid direct exposure to the beam.
General Safety Considerations
If any of the following conditions exist, or are even suspected, do not use the
instrument until safe operation can be verified by trained service personnel:
• Visible damage
• Severe transport stress
• Prolonged storage under adverse conditions
• Failure to perform intended measurements or functions
If necessary, return the instrument to ILX Lightwave, or authorized local ILX
Lightwave distributor, for service or repair to ensure that safety features are
maintained (see the contact information on page xiii).
04_06LDX-3232 ix
SAFETY SYMBOLS
All instruments returned to ILX Lightwave are required to have a Return
Authorization Number assigned by an official representative of ILX Lightwave
Corporation. See Returning an Instrument on page xi for more information.
SAFETY SYMBOLS
This section describes the safety symbols and classifications.
Technical specifications including electrical ratings and weight are included within
the manual. See the Table of Contents to locate the specifications and other
product information. The following classifications are standard across all ILX
Lightwave products:
• Indoor use only
• Ordinary Protection: This product is NOT protected against the harmful ingress of moisture.
• Class I Equipment (grounded type)
• Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
• Pollution Degree II
• Installation (overvoltage) Category II for transient overvoltages
• Maximum Relative Humidity: <80% RH, non-condensing
• Operating temperature range of 0°C to 40°C
• Storage and transportation temperature of -40°C to 70°C
• Maximum altitude: 3000 m (9843 ft.)
• This equipment is suitable for continuous operation.
x LDX-3232
Safety Marking Symbols
This section provides a description of the safety marking symbols that appear on
the instrument. These symbols provide information about potentially dangerous
situations that can result in death, injury, or damage to the instrument and other
components.
Caution,
refer to
manual
Earth
ground
Te r mi n a l
Alternating
current
Visible and/or
invisible laser
radiation
WAR RA NT Y
Caution, risk
of electric
shock
On: In position of a bistable push control.
The slash (I) only denotes that mains are on.
or
(I)
WARRANTY
ILX LIGHTWAVE CORPORATION warrants this instrument to be free from
defects in material and workmanship for a period of one year from date of
shipment. During the warranty period, ILX will repair or replace the unit, at our
option, without charge.
Limitations
This warranty does not apply to fuses, lamps, defects caused by abuse,
modifications, or to use of the product for that it was not intended.
Protective
Conductor
Te r mi n a l
or
(O)
Caution, hot
surface
Off: Out position of a bistable push control.
The circle (O) only denotes that mains are off.
Frame or
chassis
Te r mi n a l
This warranty is in lieu of all other warranties, expressed or implied, including any
implied warranty of merchantability or fitness for any particular purpose. ILX
Lightwave Corporation shall not be liable for any incidental, special, or
consequential damages.
If a problem occurs, please contact ILX Lightwave Corporation with the
instrument's serial number, and thoroughly describe the nature of the problem.
Returning an Instrument
If an instrument is to be shipped to ILX Lightwave for repair or service, be sure to:
1Obtain a Return Authorization number (RA) from ILX Customer Service.
2Attach a tag to the instrument identifying the owner and indicating the required service or
repair. Include the instrument serial number from the rear panel of the instrument.
3Attach the anti-static protective caps that were shipped with the instrument and place the
instrument in a protective anti-static bag.
4Place the instrument in the original packing container with at least 3 inches (7.5 cm) of
compressible packaging material. Shipping damage is not covered by this warranty.
5Secure the packing box with fiber reinforced strapping tape or metal bands.
04_06LDX-3232 xi
WAR RA NT Y
6Send the instrument, transportation pre-paid, to ILX Lightwave. Clearly write the return
authorization number on the outside of the box and on the shipping paperwork. ILX
Lightwave recommends you insure the shipment.
If the original shipping container is not available, place your instrument in a
container with at least 3 inches (7.5 cm) of compressible packaging material on all
sides.
Repairs are made and the instrument returned transportation pre-paid. Repairs
are warranted for the remainder of the original warranty or for 90 days, thatever is
greater.
Claims for Shipping Damage
When you receive the instrument, inspect it immediately for any damage or
shortages on the packing list. If the instrument is damaged, file a claim with the
carrier. The factory will supply you with a quotation for estimated costs of repair.
You must negotiate and settle with the carrier for the amount of damage.
xii LDX-3232
WAR RA NT Y
Comments, Suggestions, and Problems
To ensure that you get the most out of your ILX Lightwave product, we ask that
you direct any product operation or service related questions or comments to ILX
Lightwave Customer Support. You may contact us in whatever way is most
convenient:
P. O. Box 6310
Bozeman, Montana, U.S.A 59771
www.ilxlightwave.com
When you contact us, please have the following information:
Model Number:
Serial Number:
End-User Name:
Company:
Phone:
Fax:
Description of what is
connected to the ILX
Lightwave instrument:
Description of the problem:
If ILX Lightwave determines that a return to the factory is necessary, you are
issued a Return Authorization (RA) number. Please mark this number on the
outside of the shipping box.
You or your shipping service are responsible for any shipping damage when
returning the instrument to ILX Lightwave; ILX recommends you insure the
shipment. If the original shipping container is not available, place your instrument
04_06LDX-3232 xiii
WAR RA NT Y
in a container with at least 3 inches (7.5 cm) of compressible packaging material
on all sides.
We look forward to serving you even better in the future!
xiv LDX-3232
INTRODUCTION AND SPECIFICATIONS
Product Overview
CHAPTER 1
CHAPTER 1
INTRODUCTIONAND SPECIFICATIONS
This chapter is an introduction to the LDX-3232 High Compliance Curren t Source.
This chapter contains first time setup information, important safet y considerations,
maintenance information, instrument specifications, and general LDX-3232
information.
If any of the following symptoms exist, or are even suspected, remove the LDX-3232
from service. Do not use the LDX-3232 until trained service personnel can verify safe
operation.
Visible damage
Severe transport stress
Prolonged storage under adverse conditions
Failure to perform intended measurements or functions
If necessary, return the LDX-3232 to ILX Lightwave for service and repair to ensure that
safety features are maintained.
Product Overview
The LDX-3232 instrument is a high compliance laser diode current source
intended for use with quantum cascade laser diodes and other devices requiring a
high compliance output. The LDX-3232 current source provides a high stability
output with multiple laser protection features and a fully redundant current limit.
The LDX-3232 offers the added benefits of dual laser interlocks and a GPIB
interface to allow full automation of testing.
04_06LDX-3232 1
CHAPTER 1
INTRODUCTION AND SPECIFICATIONS
Initial Inspection
Initial Inspection
When you receive your LDX-3232 instrument, verify that the following items were
shipped with the instrument:
• LDX-3232 High Compliance Current Source
• LDX-3232 Series Instruction Manual
• Power Cord
• Shipping Kit
Installing Your LDX-3232 High Compliance Current Source
Grounding Requirements
The LDX-3232 High Compliance Current Source comes with a three conductor
AC power cable. The power cable must either be plugged into an approved threecontact electrical outlet or used with a three-contact or two-contact adapter with
the grounding wire connected to an electrical ground (safety ground). The
LDX-3232's power jack and supplied power cable meet IEC safety standards.
AC Line Power Requirements
You can operate the LDX-3232 instrument from a single phase power source
delivering nominal line voltages of 100, 120, 220-240 VAC (all values RMS), at
50/60 Hz. The line power voltage can vary ±10% but cannot exceed 250 VAC.
Maximum power consumption is 60 VA (Volt-Amps). The instrument's operational
voltage is factory configurable and need not be changed before operating the
instrument. However, check to be sure that the voltage printed on the back panel
of the instrument matches the power-line voltage in your area.
Before connecting the LDX-3232 High Compliance Current Source to a power source,
verify that the AC power source matches the setting of the LDX-3232's voltage printed on
the rear panel of the instrument.
T o avoid electrical shock hazard, connect the instrument to properly earth-grounded,
3-prong receptacles only. Failure to observe this precaution can result in severe injury or
death.
2 LDX-3232
INTRODUCTION AND SPECIFICATIONS
Installing Your LDX-3232 High Compliance Current Source
CHAPTER 1
GPIB Connector
The IEEE 488 GPIB interface connector is located on the rear panel, directly
above the power input module and fuse. See Figure 1.2, LDX-3232 Rear View.
Attach the GPIB cable to the 24-pin connector located on the rear panel. The
connector is tapered to ensure proper orientation. Using your fingers, tighten the
two screws on the cable connector.
A total of 15 devices can be connected together on the same GPIB interface bus.
The cables have single male/female connectors on each end so that several
cables can be stacked. This allows more than one cable to be attached to any
one device. However, the maximum length of the GPIB cables must not exceed
20 meters (65 feet) total or 2 meters (6.5 feet) per device.
The GPIB Address
The talk and listen addresses on the LDX-3232 High Compliance Current Source
are identical. This GPIB address is read locally by pressing the (GPIB) LOCAL
switch with the address displayed on the LED display . The instrument comes from
the factory configured with the GPIB address set to 1. You can change the
LDX-3232's GPIB address locally (via front panel). A procedure for changing the
address can be found in the section "Changing the GPIB Address" in Chapter 3.
Tilt-Foot Adjustment
The LDX-3232 High Compliance Current Source comes standard with folding
front legs and two rear feet for use as a bench top instrumen t. Extendin g the front
feet so that the instrument front panel sits up makes it easier to view the LED
displays. To use them, place the unit on a stable base and rotate the front legs
downward until they lock into position.
Operating the LDX-3232 Precision Current Source
Now that the LDX-3232 High Compliance Current Source is installed and ready
for use, you can begin to learn about its operation. The follo wing photos show the
functional keypad groupings, back panel connectors, and so on. Use these figures
to familiarize yourself with the LDX-3232. After that, use Chapter 2 for
fundamentals of operating your instrument.
04_06LDX-3232 3
Display
Power
Switch
CHAPTER 1
INTRODUCTION AND SPECIFICATIONS
Installing Your LDX-3232 High Compliance Current Source
Adjust
Section
GPIB Section
Fan
Parameter
Selection
PD Bias Adjust
Laser Mode
Selection
Figure 1.1 LDX-3232 Front View
LD Connector
Trigger Output
Connector
Interlock Input
Connector
Current Range
Selection
GPIB Connector
Modulation Input
Connector
Power Cord
Connector
4 LDX-3232
Figure 1.2 LDX-3232 Rear View
INTRODUCTION AND SPECIFICATIONS
Maintenance
Maintenance
You may have to return your instrument to the ILX Lightwave facility at some time
for repair or service whether or not it is under warranty. There is a charge for
repairs after the warranty period has expired. Contact an ILX Lightwave service
representative for shipping instructions prior to returning the in strument. Have th e
model number, instrument serial number, and a description of the problem or
request at the time of the call. All ILX Lightwave instruments are identified by a
serial number located on the rear panel. The first four or five digits are the model
number; the last four digits are unique to the instrument and identify your
instrument specifically. A return authorization number will be given to you at the
time of your request for repair or service. Please use this number in all
communications concerning your instrument.
If you need to ship your LDX-3232 High Compliance Current Source back to the
factory for repair, be sure that the LDX-3232 is packaged in an enclosure with
cushioning material to prevent damage to the instrument during shipment (use the
original shipping containers and accessories if possible). Re-install the ESD
protective caps on the rear and front panels over the connectors (9 pin D, BNC,
and GPIB). Shipping damage is not covered under warranty.
CHAPTER 1
Attach a tag to the instrument identifying the owner and indicating the service or
repair needed. Include the model number and serial number. We suggest that you
insure the shipment.
In North America, contact ILX Lightwave for return authorization and shipping
instructions. Outside of North America, contact your ILX distributor for shipping
information.
For more information on how to obtain repair services, see the Safety and
Warranty section of this manual.
If cleaning is required, use a clean dry cloth. Do not use solvents.
Potentially lethal voltages exist within the LDX-3232 High Compliance Current Source. To
avoid electric shock, do not perform any of the maintenance on the instrument unless
you are qualified to do so. Qualified service personnel are required to wear protective
eyeglasses and anti-static wristbands while working on the LDX-3232 High Compliance
Current Source circuit boards. High voltages are present on and around the instrument’s
printed circuit boards.
04_06LDX-3232 5
CHAPTER 1
INTRODUCTION AND SPECIFICATIONS
Specifications
Specifications
Drive Current Output
Output Current Range0 to 2000 mA0 to 4000 mA
Setpoint Resolution
Setpoint Accuracy (% of FS)±0.05%±0.05%
Compliance Voltage0 - 15 V, adjustable0 - 15 V, adjustable
Temperature Coefficient<100 ppm/°C<100 ppm/°C
Short Term Stability (1 hour)
Long Term Stability (24 hours)
Noise and Ripple (µA rms)
Range1 to 2020 mA1 to 4040 mA
Resolution10 mA20 mA
Accuracy±20 mA±40 mA
Photodiode Feedback
TypeDifferentialDifferential
PD Reverse Bias0 - 5V, adjustable0 - 5V, adjustable
PD Current Range5 to 10,000 µA5 to 10,000 µA
Output Stability
Accuracy, setpoint (% of FS)±0.05%±0.05%
External Analog Modulation
Input0 - 10 V, 1 kΩ0 - 10 V, 1 kΩ
Transfer Function200 mA / V400 mA / V
8
±0.02%±0.02%
6 LDX-3232
INTRODUCTION AND SPECIFICATIONS
Specifications
LDX-3232
Bandwidth (3 dB)
High Bandwidth
Low Bandwidth
Trigge r Output
TypeTTLTTL
Pulse Width13 µs13 µs
9
DC to 250 kHz
DC to 10 kHz
DC to 250 kHz
DC to 10 kHz
CHAPTER 1
Delay12 ms12 ms
Measurement (Display)
Output Current Range0 to 2000.0 mA0 to 4000.0 mA
Output Current Resolution
Photodiode Current Range0 to 10,000 µA0 to 10,000 µA
Photodiode Current Resolution1 µA1 µA
Photodiode Current Accuracy±4 µA±4 µA
PD Responsivity Range
PD Responsivity Resolution0.01 µA/mW0.01 µA/mW
Optical Power Range0.00 to 5000.0 mW0.00 to 5000.0 mW
Optical Power Resolution0.1 mW0.1 mW
Forward Voltage Range0.000 to 10.000 V0.000 to 10.000 V
Forward Voltage Resolution1 mV1 mV
Forward Voltage Accuracy
General
Size (H x W x D)4” x 8.5” x 13.4”; 102 mm x 216 mm x 340 mm
Weight~ 20 pounds (9 kg)
Power100/120/230-240 V~; 2/2/1A 50/60 Hz
10
11
12
0.1 mA0.1 mA
0.00 to 1000.00 µA/mW0.00 to 1000.00 µA/mW
±2 mV±2 mV
Temperature0°C – 40°C operating; -40°C – 70°C storage
Humidity< 90% relative humidity, non-condensing
Laser Safety FeaturesKey switch, interlock and output dela y (meets CDRH US
Display Type5-digit, green LED
Notes
1. All values relate to a one-hour warm-up period at room temperature, 25°C.
2. Based on resolution of digital-to-analog converters used in the design.
3. Over any 1-hour period, half-scale output.
4. Over any 24-hour period, half-scale output.
5. Measured electrically with a 1 ohm load at half-scale output; evaluating spectral noise density over a 150kHz bandwidth from
125kHz to 275 kHz.
6. Maximum output current transient from normal operational situations (e.g., power on-off, current on-off), as well as accident al
situations (e.g., power line plug removal).
7. Maximum output current transient resulting from a 1000 V power-line transient spike. Tested to ILX Lightwave Technical Standard
#LDX-00196.
04_06LDX-3232 7
21 CFR 1040.10)
CHAPTER 1
Available Options and Accessories
INTRODUCTION AND SPECIFICATIONS
Specifications
8. Maximum monitor photodiode current drift over any 30 minute period. Assumes zero drift in responsivity of photodiode.
9. 50% modulation at mid-scale output. Higher bandwidth is possible with smaller magnitude modulation signal.
10.Similar resolution available over GPIB.
11.Responsivity value is user-defined and is used to calculate the optical power.
12.Four wire voltage measured at the load. Voltage measurement accuracy while driving calibration load. Accuracy is dependent
upon load and cable used.
Options and accessories available for the LDX-3232 High Compliance Current
Sources include the following:
DESCRIPTIONMODEL NUMBER
Noise FilterLNF-320
Current Source Interconnect Cable (terminated)CC-305S
Current Source Interconnect Cable (unterminated)CC-306S
DIL Laser Diode MountLDM-4982
Butterfly Laser Diode MountLDM-4894
High Power Laser Diode MountLDM-4442
Single Rack Mount KitRM-139
Dual Rack Mount KitRM-140
Other laser diode mounts are available. Please contact ILX Lightwave for information on
additional options for your application.
Our goal is to make the best laser diode instrumentation available anywhere. To
achieve this, we need your ideas and comments on ways we can improve out
products. We invite you to contact us at any time with your suggestions.
8 LDX-3232
OPERATION
Applying Power to Your LDX-3232
CHAPTER 2
CHAPTER 2
OPERATION
This chapter introduces you to the operation of the LDX-3232 High Compliance
Current Source. It offers instructions for connecting your laser to the current
source and describes powering up the instrument. This chapter also cont ains step
by step procedures that teach you how to operate your current source in Const ant
Current Mode and Constant Power Mode. We recommend that you review the
contents of this chapter at a minimum before operating the LDX-3232 Current
Source.
Applying Power to Your LDX-3232
To turn on the LDX-3232, rotate the key in the "POWER"
section of the front panel from "O" to "I". This action will initiate
the power on sequence. If the LDX-3232 does not appear to
turn on, verify that it is connected to line power. If line power is
not the problem, remove the power cord and check the voltage
selection block in addition to the line power fuse.
The Power On Sequence
During the power-up sequence, the following takes place. For about three
seconds all indicators light up, and all of the 7-segment displays indicate "8". Then
all lamps are turned off for three seconds. Next, the firmware version will be
displayed. Then, a self-test is performed to ensure that the instrument' s hardware
and processor are communicating. After this test, the instrument is ready to
operate and is configured to the state it was in when the power was last shut off.
04_06LDX-3232 9
CHAPTER 2
The Power On State
OPERATION
Connecting to the Laser
You can choose to "clear" the last saved state by recalling the default values.
These values are stored in BINS to that you can save or recall instrument
configurations (described in more detail in Chapter 5). "BIN 0" contains all of the
default configurations listed in Table 2.1.
Note: If the instrument cannot successfully complete the power on test, an error message
of E-512 or E-513 will be displayed on the Display. See Chapter 3 for a list of error codes
and actions to take when your instrument indicates these errors.
Table 2.1 LDX-3232 Default Settings
GPIB mode in LOCAL via front panel, or in REMOTE via GPIB
PARAMETERS not selected
CAL PD = 0 µA / mW
LIM I (high range) = 1000 mA
LIM I (low range) = 500 mA
LIM V = 15 volts
LIM P = 5000 mW
Output off
DISPLAY enabled, in I mode
Constant I, low bandwidth mode selected
I setpoint = 0 mA
setpoint = 0 µA
I
PD
setpoint = 0 mW
P
PD
RECALL BIN number = 0
Connecting to the Laser
When connecting your laser to the LDX-3232, we recommend that the instrument
be powered up with the output off. In this condition, a low impedance shunt is
active across the output terminals. When disconnecting devices, it is only
necessary to turn the current source output off.
It is also recommended that the connections to the LDX-3232 output be made
using twisted wire pairs with an earth-grounded shield (see Figures 2.1 - 2.4). The
output terminals of the instrument are left floating relative to earth ground to
suppress AC power-on/power-off transients that may occur through an earthground path. If the output circuit is earth-grounded at some point (such as through
the laser package and mount), the user must be careful to avoid multiple earth
10 LDX-3232
OPERATION
Connecting to the Laser
CHAPTER 2
grounds in the circuit. Multiple earth grounds may provide circuit paths that induce
spurious currents in the photodiode feedback circuit and output leads.
Note: Cable connections to the laser must be secure! Loose connections can cause
momentary open circuits that can damage the laser.
Figures 2.1 - 2.4 show the possible configurations of connecting laser diodes and
photodiodes with the LDX-3232 High Compliance Laser Current Source.
OUTPUT
LDX-3232 Current Source
7
Bias
+
6
+
-
9
5
P. D. L. D.
3
Earth Ground
Figure 2.1 Common Laser Cathode - Photodiode Cathode
P. D. L. D.
OUTPUT
LDX-3232 Current Source
7
Bias
+
6
9
5
3
+
-
Earth Ground
04_06LDX-3232 11
Figure 2.2 Common Laser Cathode - Photodiode Anode
P. D. L. D.
OUTPUT
LDX-3232 Current Source
7
Bias
+
6
9
5
3
+
-
Earth Ground
Figure 2.3 Common Laser Anode - Photodiode Cathode
CHAPTER 2
OPERATION
Connecting to the Laser
P. D. L. D.
OUTPUT
LDX-3232 Current Source
7
Bias
+
6
9
5
3
+
-
Earth Ground
Figure 2.4 Common Laser Anode - Photodiode Anode
The 9-pin connector marked CURRENT SOURCE on the back panel is used to
connect your laser diode to the LDX-3232. There are connections provided for
laser cathode and anode, photodiode cathode and anode, chassis ground and
interlock. The pinout diagram for this connector is shown in Figure 2.5.
1, 2 Interlock
3 Chass is Groun d
4 Cathode Vo lta g e S ense
5 Laser Cathod e
6 PD Cathode ( + )
7 PD A n o d e ( -)
8 Anod e Voltage Sense
9 Laser Anode
1
6
2
7
3
8
4
9
5
12 LDX-3232
Figure 2.5 Back Panel LD Connector
Interlock Connections
In order for the laser output to be enabled a short circuit must exist between the
Interlock Pins (Pins 1 and 2) of the connector. The short can be a direct short
across the pins or a switch to prevent laser operation until the switch is closed. If a
short does not exist between these two pins, the INTERLOCK LED will illuminate
on the front panel and the laser output will be disabled.
Note: The BNC Interlock Connection from the BNC conne ctor on the rear panel will also
make the INTERLOCK LED illuminate. Please see the notes in chapter five on use of the
BNC interlock connection to learn more about its operation.
The interlock terminals on the LASER connector, pins 1 and 2, must be kept isolated
from all other connections including earth g round.
OPERATION
Connecting to the Laser
CHAPTER 2
Photodiode Connections
Many laser diode modules contain an internal photodiode that monitors the backfacet emission of the laser. Usually, this photodiode is internally connected to
either the laser anode or cathode.
The photodiode and laser connections to the LDX-3232 are electrically isolated
from ground and each other. So, if a 4 pin connection is made (no common
connections), no additional jumpers are required. Figures 2.1 - 2.4 show the
recommended connections and shielding for 3-pin lasers (where the common
connection is internal to the device). A 4-pin laser should be connected with the
same shielding as shown in Figure 2.1, but the common connection (between the
photodiode and the laser) is optional.
Setting the PD Bias
The LDX-3232 provides an adjustable reverse bias of 0-5 VDC for the photodiode.
To set the photodiode bias to 5 V reverse bias, turn the back panel PD BIAS
ADJUST fully clockwise. To set the photodiode bias to 0 V reverse bias, turn the
back panel PD BIAS ADJUST fully counter-clockwise.
Grounding Considerations
The LASER outputs of the LDX-3232 High Compliance Current Source are
isolated from chassis ground, allowing either output terminal to be grounded at the
user's option. Figures 2.1 - 2.4 show the proper earth-ground shielding for laser
diode/photodiode connections.
04_06LDX-3232 13
CHAPTER 2
OPERATION
Front Pan el Operation
Front Panel Operation
This section describes fundamentals of operation for your LDX-3232 in two
operating modes: Constant Current (I) and Constant Power (P).
The Display
The digital display is used to show control parameters such as laser drive current
(mA), laser current limit (mA), laser forward voltage (V), voltage limit (V), monitor
photodiode current (µA), and laser optical power (mW) in both setpoint and
measurement modes. It will also display error codes that relate to LASER
operation. Error indicators underneath the display indicate control errors and are
explained in the section titled “ERROR INDICATORS”. All laser control
parameters are displayed on this digital display with the appropriate annunciator
illuminated.
Operating a Laser in Constant Current (I) Mode
Suppose you have an LDX-3232 and you want to run the instrument in Constant
Current, low bandwidth mode, with a setpoint of 1500 mA and a current limit of
1750 mA. This particular mode uses a low-pass filter on the laser drive current
output to significantly reduce noise. In this mode a modulation input of up to
10 kHz may be used via the front panel MOD EXTERNAL connector. Follow the
instructions below in the sequence presented.
14 LDX-3232
Operating a Laser in Constant Current (I) Mode
Setting Up the Precision Current Source
Set the LDX-3232 in Constant Current, low bandwid th mode
by selecting I in the MODE section of the front panel. Do this
by toggling the SELECT push button until the I indicator is
illuminated. Repeatedly pressing the (MODE) SELECT
switch cycles through the current (I), light power (P), or
high-bandwidth current (I
Next, in the RANGE section set the current source range by
selecting the 2000 mA range indicated by the accompanying
LED. You will need to press the SELECT push button twice
to change the range. It must be pressed twice quickly (within
less than 1 second) to switch between ranges, and the
LASER output must be off (LASER MODE ON indicator
unlit) in order to change ranges.
HBW
OPERATION
) control modes.
CHAPTER 2
Configure the display so that it reads laser current in mA. Press and release the
SELECT switch in the DISPLA Y section of the front panel until the mA annunciator
is lit. The (DISPLAY) SELECT switch is also used to select the I
µA), P
display modes. When the setpoint mode is active, the corresponding annunciator
will flash to indicate setpoint rather than measurement display.
To only view the setpoint, press and hold down the display SELECT switch for two
or more seconds. When the SELECT switch is released the previous
measurement mode will be restored.
At this point the LDX-3232 is in Constant Current Mode, (I), in the 2000 mA rang e,
and the display is reading in mA. Next, we need to adjust the setpoint of the laser
current source.
To adjust the setpoint of the current source to 1500 mA, turn the adjust knob to
vary the setpoint. When the adjust knob is turned, the LDX-3232 automatically
changes to setpoint mode indicated by the flashing annunciator. Adjust the
setpoint to 1500 mA. After an adjustment has been made, and the adjust knob is
released, the setpoint mode will time out in three seconds. The display will revert
to the original measurement display mode (I, I
The instrument is still not ready to turn on the laser. One of the laser protection
features our precision current sources offer is a fully independent hardware
current limit. The current limit protects your laser by never allowing the laser drive
current to exceed the limit independent of the setpoint and the output control
mode (Constant Current or Constant Power).
(laser power, mW), or V (forward voltage, V) measurement or setpoint
PD
, V, or PPD).
PD
(PD current,
PD
04_06LDX-3232 15
CHAPTER 2
OPERATION
Operating a Laser in Constant Current (I) Mode
To set the current source limit, use the SELECT
push button in the P ARAMETER se ction of the front
panel to select LIM I. Press and hold the
(PARAMETER) SET push button and rotate the
adjust knob to vary the limit setpoint value. Set the
limit to 1750 mA. Once the (PARAMETER) SET
push button is released, the new parameter value is
stored in non-volatile memory.
When the LASER current limit is reached, the CURRENT LIMIT error indicator
flashes. The current limit setting is independent of the voltage drop of the device
connected to the LASER output, and therefore, no dummy load is required for
precise adjustment of the current limit. Furthermore, since the current limit
circuitry is fully independent of the main current control, the current limit can be
adjusted safely, even while the LASER output is active.
An additional feature of the LDX-3232 instrument is an adjustable voltage limit.
This allows an extra level of laser protection in the event of an intermittent open
circuit. For maximum protection via the fastest open-circuit detection, the voltage
limit should be set to a value slightly greater than the intended operating voltage.
To set the LASER voltage limit, use the SELECT push
button in the PARAMETER section of the front panel to
select COMP V and then press and hold the (P ARAMETER)
SET push button. While holding the (PARAMETER) SET
push button, adjust the voltage limit using the adjust knob.
Now you are ready to enable the current source output for
your laser diode.
In the MODE section of the front panel, press the push button labeled OUTPUT.
The accompanying LED labeled ON will illuminate, indicating that the laser current
source's output is enabled. The instrument will drive the current source to the
value set by the corresponding MODE setpoint. When the LASER output is off, an
internal short is placed across the LASER output. The OUTPUT switch has a
toggling action. Pressing the switch again will disable the current source output
and the ON LED will not be illuminated. The default condition for the current
source output is off when the instrument is first powered up. With the laser output
enabled, the LASER display should be indicating measured laser current. If the
output won't function, check the following conditions:
16 LDX-3232
OPERATION
Operating a Laser in Constant Power (P) Mode
Conditions That Will Automatically Shut Off the LASER OUTPUT
The following conditions will automatically cause the instrument to disable the
current source output:
• LASER High Power Limit
• LASER Interlock / Key Lock Turned Off
• LASER Open Circuit or Voltage Limit
Operating a Laser in Constant Power (P) Mode
Suppose you want to operate a 3 mW laser at constant 1mW level. The LDX-3232
allows you to operate the laser current source driver in a Constant Power mode.
In this mode, the LDX-3232 drives current to the laser to reach a setpoint power
value (in mW). The control loop feedback parameter is photodiode current that the
LDX-3232 converts to optical power via a user-defined photodiode responsivity
number. If laser power changes due to internal or environmental conditions, the
LDX-3232 will increase/decrease the current to the laser to maintain the power
setpoint. The following paragraphs detail the instrument configuration in Constant
Power Mode.
CHAPTER 2
To set the LDX-3232 in Constant Power mode, select P in the MODE section of
the front panel by toggling the SELECT push button. Repeatedly pressing the
(MODE) SELECT switch cycles through the current (I), light power (P), or high
bandwidth current (I
mode.
) control modes. The LED indicator shows the selected
HBW
Entering Photodiode Responsivity Values
The photodiode responsivity, CAL photodiode, is
used to convert between photodiode current and
optical power of the laser diode. To enter the CAL
photodiode parameter, select CAL PD with the
(PARAMETER) SELECT push button by toggling
down past LIM P. The indicator becomes lit when
CAL PD is selected and the display indicates the
responsivity parameter in µA/mW (not annunciated). Rotate the adjust knob while holding the
(PARAMETER) SET push button to enter a new
value for responsivity. When the correct value is
displayed, store the number by releasing the
(PARAMETER) SET pushbutton.
04_06LDX-3232 17
CHAPTER 2
OPERATION
Operating a Laser in Constant Power (P) Mode
A method to calculate the responsivity is suggested below.
1Measure (with a calibrated detector) the output power of the device.
2Measure the corresponding photodiode current.
3Calculate the responsivity by dividing the photodiode current by the optical power noting
the units required are µA/mW.
The LDX-3232 High Compliance Current Source can be put
into a special mode with the CAL PD parameter set to zero.
When the CAL PD value is zero, the LASER output will be
controlled to the I
instrument is in Constant I
"-.-" for the power measurement.
Next, configure the display so that it reads in mW by pressing and releasing the
(DISPLAY) SELECT switch until the mW annunciator is lit.
setpoint value. Therefore, the
PD
mode. The display will indicate
PD
Adjust the Constant Power setpoint of the LDX-3232 to 1.00 mW. To do this, turn
the adjust knob to vary the setpoint. When the adjust knob is turned, the
LDX-3232 automatically changes to setpoint mode as indicated by the flashing
annunciator. Adjust the setpoint to 1.00 mW. After an adjustment has been made,
and the adjust knob is released, the setpoint mode will time out in three seconds.
The display will then return to the measurement display mode (I, I
Before the laser output is enabled, the power limit for the laser under test should
be set. In Constant Power mode, this is accomplished by setting a Power limit,
LIM P value. The LDX-3232 uses the responsivity value entered to calculate a
power limit as a function of monitor current (I
To set the Constant Power limit, use the SELECT push button in the
PARAMETER section to select LIM P. When the LIM P is indicated by the
accompanying LED, the display will indicate the limit setpoint in mW. Press and
hold the (PARAMETER) SET push button and rotate the adjust knob to set the
limit to 1.50 mW. When the desired value is reached, release the (PARAMETER)
SET push button to store the new limit value.
The LIM P is the absolute limit of optical power, as detected by the unit via the
monitor photodiode feedback loop. In order for this limit to be meaningful, the
monitor photodiode responsivity (CAL PD) must be adjusted correctly. This limit is
a software limit only. The LASER output is normally turned off if this limit is
reached (See LASer:ENABle:OUTOFF command in Chapter 4 for instructions on
disabling this function). The maximum LIM P value is 5000.0 mW.
). This is a software limit only.
PD
, V, or PPD).
PD
Also, in constant power mode the LASER current limit and voltage limit values are
in effect. Setting of these limits is described later in this chapter.
18 LDX-3232
OPERATION
Error Indicators
Now the instrument is configured to operate in Constant
Power Mode. In the MODE section, enable the output of the
laser current source by depressing the OUTPUT push
button once. The accompanying LED labeled ON will
illuminate indicating that the laser output is on and current is
being driven to the laser. The LDX-3232 current source will
drive the laser to the Constant Power setpoint and maintain
closed loop control with the photodiode current
measurement.
Error Indicators
The LDX-3232 High Compliance Current Sources indicate general operational
error conditions. Each error condition results in an action as shown in the table
below.
Table 2.2 LDX-3232 Error Indicators
CHAPTER 2
Error ConditionAction
InterlockOutput off; INTERLOCK light flashes at 1 Hz
Open CircuitOutput off; OPEN CIRCUIT light flashes at 1 Hz; if this condition was caused by
Current LimitCURRENT LIMIT light flashes at 1 Hz
Voltage LimitVOLTAGE LIMIT light flashes at 1 Hz; this indicator warns that the output is within
Optical Output
Power Limit
an excessively high compliance voltage, the VOLTAGE LIMIT light will also be
turned on.
0.25 V of LIM V setting; if the voltage limit (LIM V) is reached, the LASER output
will shut off.
Output off; POWER LIMIT light flashes at 1 Hz
04_06LDX-3232 19
CHAPTER 2
OPERATION
Error Indicators
20 LDX-3232
REMOTE OPERATION
CHAPTER 3
CHAPTER 3
REMOTE OPERATION
Everything you learned to do from the front panel can also be done remotely and
in some cases, with more flexibility. For instance, in remote mode, you have
access to commands for functions not found on the front panel, such as
commands that will increment (INC) and decrement (DEC) the current setpoint by
a pre-defined step value. The following sections show you the fundamentals of
operating your LDX-3232 High Compliance Current Source remotely through the
GPIB interface.
Reading the GPIB Address
Before you can operate the LDX-3232 instrument from a remote location, you
need to know its GPIB address. Simply press the LOCAL switch in the section
labeled GPIB on the front panel. The instrument will display the address. The
default address from the factory is address "Adr 01".
Changing the GPIB Address
Every device on the GPIB bus must have a unique address. If it is necessary to
change the address, press and hold in the (PARAMETER) SET switch while
displaying the GPIB address. Turn the ADJUST knob until the desired address
value is displayed, then release the SET switch. The new GPIB address will then
be stored in non-volatile memory . The allowable address range is 0-30 for primary
GPIB addressing. Extended GPIB addressing is not implemented on the
LDX-3232 at this time.
04_06LDX-3232 21
CHAPTER 3
Changing Operation from Local to Remote
REMOTE OPERATION
LDX-3232 Current Source Command Set
Sending a command over the GPIB will automatically put the
instrument in REMOTE mode. The REMOTE indicator will
identify when the Current Source is in remote operation mode.
When the instrument is in REMOTE mode, pressing the LOCAL
switch returns the instrument to LOCAL control mode unless the
Local Lockout state has been activated by the host computer.
Local Lockout disables all LDX-3232 High Compliance Current
Source front panel switches until this condition is changed by the
host computer. When the instrument is put in Local Lockout
Mode by the host computer, the REMOTE indicato r will flash at a
1Hz rate to indicate that the front panel is completely disabled by
Local Lockout.
The TALK/LISTEN indicator is illuminated when the instrument is communicating
over the GPIB bus.
LDX-3232 Current Source Command Set
For the most efficient and effective remote control of your LDX-3232 High
Compliance Current Source, we recommend you study the following sections. You
will learn about the LDX-3232 command sets, both GPIB-IEEE488 Command
Commands and the most used device specific commands, and command synt ax.
Later in this chapter you will be introduced to the error and status registers.
Command Syntax
Any command or command query must contain all of the letters tha t a re shown in
the LDX-3232 Command Path Structure, Figure 3.2. However, the Current
Source's command parser is NOT CASE SENSITIVE so upper or lower case may
be used in any combination. The lower case letters shown with the Device
Dependent commands in Chapter 4 are optional letters and may be used for
clarity. For example, the following commands are equal:
• LAS:DIS 1
• LAS:DisP 1
• LAS:DISPLAY 1
In this example, only the first three letters "DIS" are required, while the other
letters, "play" are optional. These optional letters must be used in the proper
sequence as shown.
22 LDX-3232
REMOTE OPERATION
LDX-3232 Current Source Command Set
CHAPTER 3
The syntax of the LDX-3232 High Compliance Current Source commands fo llows
the rules laid out in the GPIB-IEEE488 standard. Colons (:) indicate the start of a
new command path, while semicolons (;) indicate a separation of commands
within a command string. A leading colon on a command may be used to return
the LDX-3232 High Compliance Current Source command parser to the
command path root (see Figure 3.1).
Sp aces or wh ite sp ace may be used to sep arate commands and /or dat a (af ter the
command header or query question mark). Spaces or white space must be used
to separate the command (header) from the first p arameter (or program data). The
following examples show valid syntax for commands with the LDX-3232 High
Compliance Current Source:
• LAS:display:ldi
• Laser:limit:ldi 400
• LAS:DIS 1; las:set:ldi?
• Las:MODE:Mdp; LAS:OUT 1
The following are examples of invalid syntax for the LDX-3232 High Compliance
Current Source. These command strings would produce an erroneous result, as
explained:
LAS:MODE MDPMissing colon, MODE? expected
LAS:MODE:ILBW DECMissing semicolon, DEC command generates an error
LAS:DIS ?Space not allowed before question mark, DIS command
Las:LDI33;dis?Space missing between LDI command and the parameter
expected
value, 33
For further clarity in programming, the (Boolean) values of one (1) and zero (0)
may be used or their names as indicated in the following table:
If multiple parameters are expected, they should be separated with commas:
LAS:INC 100,50
A query has no space between the mnemonic and the question mark:
LAS:LDI?
The LDX-3232 High Compliance Current Source uses a terminator of
<NL><^END> (new line with EOI). For users whose GPIB driver defaults expect a
carriage return in the terminator , <CR><NL><^END>, the TERM command may be
used for convenience (see TERM command in Chapter 4). For more information,
refer to your GPIB driver configuration manual.
The command structure is illustrated in Figure 3.1. Table 4.1 lists all of the
LDX-3232 High Compliance Current Source’s device-dependent commands, with
the full path shown for each command and a brief explanation of its usage.
24 LDX-3232
REMOTE OPERATION
Command Paths
Command Paths
The LDX-3232 High Compliance Current Source device-dependent commands
are structured into a tree format (see Figure 3.1). Each of the legal paths is listed
below, followed by its list of path options, each of that is followed by the
commands themselves. It is recommended that the first-time user begin learning
the commands by using the full path notation. Once you are familiar with the
commands, you may wish to take advantage of the shortcuts allowed for
command paths.
The IEEE488.2 Common Commands and Queries are distinguished by the "*"
that begins each mnemonic. The diagrams below show the syntax structure for
common commands, common commands with numeric data required, and
command command queries.
:
*
Comm on Com m a nd
Numeric data is required with *PSC (1=on, 0=off), *RCL (0 to 10, see front panel
RECALL function), *SAV (1 to 10, see front panel SAVE function), *ESE (0 to 255,
see Figure 3.2), *SRE (0 to 255, see Figure 3.2), and *PUD (used at factory only).
A list of all of the GPIB-IEEE488 Common Commands supported by the
LDX-3232 follows:
For more information on these commands, refer to an ANSI/IEEE 488-1987
standards reference.
Note: Care should be taken to set the GPIB time-out appropriately for use with the *WAI,
26 LDX-3232
DELAY, or *OPC? Commands. If the time for a response after one of these commands
exceed the GPIB time-out period, a bus (timeout) error will occur. Usually, after this timeout
error, the LDX-3232 Series Instrument will generate a query error (E302). This error code is
reported via the ERR? Query.
REMOTE OPERATION
LDX-3232 Frequently Used Commands
LDX-3232 Frequently Used Commands
The LDX-3232 High Compliance Current Source’s complete command set
contains over 60 commands that allow you to operate the current source for a
variety of applications. Within the command set, however, is a smaller subset of
commands that will meet most of your needs. These commands are listed in
Table 3.2.
Table 3.2 LDX-3232 Often Used Commands
CHAPTER 3
NameParametersFunction
ERR?NONEReturns errors generated since the last query
LAS:CALMD1Used to set the CAL PD (monitor responsivity) constant
LAS:LDI1Used to set the current source setpoint value
LAS:LDI?NONEUsed to return the constant current source measured value
LAS:LDV?NONEUsed to return the measured laser voltage value
LAS:LIM:I21Used to set the low range current limit
LAS:LIM:I41Used to set the high range current limit
LAS:LIM:MDP1Used to set the constant optical power (from monitor
photodiode) limit value
LAS:LIM:V1Used to set the current source compliance voltage limit value
(LIM V)
LAS:MDP1Used to set the constant optical power setpoint
LAS:MDP?NONEReturns the actual monitor photodiode power value
LAS:MODE?NONEReturns the mode, ILBW, IHBW, or MDP (optical power)
LAS:MODE:IHBWNONESets the mode to constant current, high bandwidth mode
LAS:MODE:ILBWNONESets the mode to constant current, low bandwidth mode
LAS:MODE:MDPNONESets the mode to constant optical power mode
LAS:OUT1Used to enable/disable the current source output
LAS:OUT?NONEReturns the current source output status
LAS:RAN1Sets the laser current output range
04_06LDX-3232 27
CHAPTER 3
REMOTE OPERATION
Status Reporting
Status Reporting
This section contains information that is relevant for understanding instrument
error and statu reporting. It also contains information regarding the use of the
instrument status for generating interrupts for interrupt driven programs or
subroutines. Understanding the Operation Complete definition for the instrument
is useful for programming synchronization. The Output Off Register section also
contains iformation on setting some of the conditions that will force the laser
current source output off.
The following sections describe the Event and Condition registers, Operation
Complete Flag, Output Off registers, and Error Messages.
Event and Condition Registers
In addition to the required GPIB-IEEE488 status reporting structures, the
LDX-3232 High Compliance Current Source remote interface provides Event and
Condition Registers for LASER operations. The Event Registers are used to
report events that occur during the operation of the LDX-3232 High Compliance
Current Source. Events differ from conditions in that events signal an occurrence
once, and are not reset until the Event Register is queried, the status cleared by
the *CLS command, or the instrument is powered off. Conditions reflect the
current state of the device and therefore may change many times during
operation. Querying a Condition Register does not change its contents.
Figure 3.3 shows the status reporting scheme of the LDX-3232 High Compliance
Current Source. Each of the registers that may be accessed by a command or
query has the appropriate command or query written above or below the register
representation. For example, the LASER Condition Register may be queried via
the LASer:COND? query.
The condition or event registers are logically ANDed with their respective enable
registers. These bits are then logically ORed to form a summary message in the
status byte for that particular register.
Operation Complete Definition
Note that Bit 0 of the Standard Event Status Register contains the status of the
Operation Complete flag. Enabling this bit via the *ESE command allows the user
to update bit 5 of the status byte. Then, if the SRE ma sk has bit 5 set, and the user
issues an *OPC command, the SRQ (bus) signal will be generated upon
completion of the currently processed commands. This bus signal (SRQ) may be
used to initiate service request routines, that depend on the completion of all
previous commands.
28 LDX-3232
REMOTE OPERATION
L
:
L
e
L
1
L
t
Status Reporting
CHAPTER 3
For example, the user may turn the current source on (LAS:OUT 1), enable an
SRQ on Operation Complete (set *ESE 1 and *SRE 5), and have an SRQ
handling routine in the user's software, that reads a new measurement after the
output on state has been reached. This allows the use of the operation complete
features of the LDX-3232, without the need for program loo ping or polling that can
tie up the GPIB. Operation Complete on the LDX-3232 is defined as:
1No operations to the LASER current source hardware are pending.
2No EEPROM (non-volatile) memory write cycles are in progress.
Note: If the GPIB times out while waiting for a response, either set the GPIB time-our
longer or use SRQ generated interrupts in your program. See your GPIB manual for
time-out configuration or SRQ programming setup.
The *OPC, *OPC? and *WAI commands should not be used inside a calibration routine.
04_06LDX-3232 29
Figure 3.2 LDX-3232 Status Reporting Schematic Diagram
CHAPTER 3
Output Off Register
REMOTE OPERATION
Status Reporting
The Output Off Enable Register allows you to determine that conditions and
events in the LDX-3232 High Compliance Current Source can cause the current
source output to be turned off. This reigster is configured in a manner similar to
the status reporting registers. However, the register output is not reported in the
Status Byte Register. Rather, it sets the hardware that controls the output
switching. The events and conditions that may be set to cause the current source
output to be turned off are shown in Figure 3.3.
The default (factory) settings for this register are shown in Table 3.3. These
settings are not affected by the *PSC (Power-On Status Clear) command.
Table 3.3 LDX-3232 Series Default Settings for Output Off Registers
0 - LASER Current Limit8 - N/A
1 - LASER Voltage Limit9 - LASER Output Changed to be Out of Tolerance
2 - N/A10 - N/A
3 - LASER Power Limit (output on)11 - N/A
4 - N/A12 - N/A
5 - N/A13 - N/A
6 - N/A14 - N/A
7 - N/A15 - N/A
04_06LDX-3232 31
CHAPTER 3
REMOTE OPERATION
Command Timing and Completion
Command Timing and Completion
This section describes, for each device-dependent command, whether that
command is performed in an overlapped or sequential manner. In other words, it
states whether the next command may begin while the current command is being
executed, or if the next command must wait until the current command is
completed before its execution begins. The conditions for setting the operation
complete flag are given in the Chapter 3 section titled Operation Complete
Definition.
All LDX-3232 device-dependent commands are executed in an overlapped
manner, except the DELAY command that is sequential. The operation complete
flag is set after the conditions outlined in the Operation Complete Definition have
been satisfied.
The *WAI (common command) is an example of a sequential command that
forces the next command to wait until the no-operation-pending flag is true. This is
essentially the same as waiting for the OPC flag to become true, because the nooperations-pending flag is used to set the OPC flag (bit 0 of the Standard Event
Status Register).
Commands that change the status of the instrument limits or change its mode or
current range, step value, or status enable registers will not have their OPC flag
set until all current writing to non-volatile memory has been completed. This is
done to ensure that the OPC flag is never set prematurely.
Whenever there is any output (response) data in the Output Queue, bit 4 is set in
the Status Byte Register. Whenever there is any error message in the Error
Queue, bit 7 is set in the Status Byte Register.
Error Messages
In the event of a hardware error condition, error messages will be displayed on the
display. In most cases, the error message will appear for three seconds. In some
cases the error code display will remain until the user changes display modes. In
the case of multiple error messages, the display may sequentially show each
message for three seconds. In addition to the hardware errors, GPIB errors may
be read via the ERR? query. Table 3.4 lists the numerical error ranges by function.
Table 3.5 contains all of the error messages that may be generated. Not all of
these messages may appear on the front panel displays. Some refer to GPIB
activities only.
In remote operation, the errors can be read by issuing the ERR? query. When this
is done, all of the error messages that are resident in the error queue are returned
32 LDX-3232
REMOTE OPERATION
Error Messages
CHAPTER 3
(up to 10 may be stored). Reading the error queue via GPIB clears the error
queue.
Table 3.4 Error Code Classifications
Error Code RangeArea of Operation
E-001 to E-099Internal Program Errors
E-100 to E-199Parser Errors
E-200 to E-299Execution Control Errors
E-300 to E-399GPIB Errors
E-500 to E-599LASER Control Errors
Table 3.5 LDX-3232 Series Error Codes
Error CodeExplanation
E-001Memory allocation failure
E-103<DEFINITE LENGTH ARBITRARY BLOCK PROGRAM DATA> length too long
E-104<NON-DECIMAL NUMERIC PROGRAM DATA> type not defined
E-105<DECIMAL PROGRAM DATA> exponent not valid
E-106<DECIMAL PROGRAM DATA> digit expected
E-123Command not found
E-124<program mnemonic> Lookup, failed because query/command type match failed
E-126Too few or too many program data elements
E-201<PROGRAM DATA> value out of range
E-202<PROGRAM DATA> will not convert to valid type
E-203Security violation, command is not available without clearance
E-205<PROGRAM DATA> is not a Boolean value or word
E-207<PROGRAM DATA> will not convert to an unsigned 16-b it value
E-209<PROGRAM DATA> will not convert to an unsigned 32-b it value
E-210<PROGRAM DATA> will not convert to a floating point value
E-211<PROGRAM DATA> will not convert to a character value
E-213<PROGRAM DATA> is incorrect block data length
E-214<PROGRAM DATA> length exceeds maximum
E-302Query error; device was addressed to talk, but GPIB controller failed to read all of the
<RESPONSE MESSAGE>
04_06LDX-3232 33
CHAPTER 3
REMOTE OPERATION
Error Messages
Error CodeExplanation
E-501LASER interlock disabled output
E-503LASER open circuit disabled output
E-504LASER current limit disabled output
E-505LASER voltage limit disabled output
E-507LASER power limit disabled output
E-510LASER out of tolerance disabled output
E-511LASER control error disabled output
E-512Analog section status is all 1’s or all 0’s (power down)
E-513Serial EEPROM checksum error
E-515Laser output must be off to change ranges
E-516Incorrect configuration for calibration sequence to start
E-519Setting a measurement is only valid during the calibration phase for that measurement;
user has tried to calibrate a measurement without first entering the required calibration
mode.
34 LDX-3232
COMMAND REFERENCE
CHAPTER 4
CHAPTER 4
COMMAND REFERENCE
This chapter is a guide to all of the device-dependent commands for the
LDX-3232 High Compliance Current Source. This chapter is divided into two
parts. The first part contains an overview of the remote commands used by the
LDX-3232 High Compliance Current Source. The second part contains all of the
LDX-3232 commands in alphabetical order. Figure 4.1 shows the format for the
device command descriptions in this chapter. The commands that emulate local
(front panel) operation are denoted by the solid box next to the Local label in the
upper left corner of the command description.
Table 4.1 LDX-3232 Series Device-Dependent Comm a nd s
NameParametersFunction
DELAY1Used to create a delay in the execution of further
ERR?NONEReturns errors generated since the last query
LAS:CALMD1Used to set the CAL PD (monitor responsivity) constant
LAS:CALMD?NONERe turns the CAL PD (monitor responsivity) constant
LAS:CAL:CONST?1Factory use only
LAS:CAL:LDINONEUsed to enter the laser current calibration mode
LAS:CAL:LDI?NONEReturns the ready state for entering the laser current
LAS:CAL:LDVNONEUsed to enter the laser voltage calibration mode
LAS:CAL:LDV?NONEReturns the ready state for entering the laser voltage
LAS:CAL:MDINONEUsed to enter the monitor photodiode current calibration
04_06LDX-3232 35
commands
calibration value
calibration mode
mode
CHAPTER 4
COMMAND REFERENCE
NameParametersFunction
LAS:CAL:MDI?NONERe turns the ready state for entering the monitor
photodiode current calibration value
LAS:COND?NONEReturns the value of the LASER condition register
LAS:DEC0, 1 or 2Used to decreate the laser current setpoint value by one
or more steps
LAS:DIS1Turns the LASER display on or off
LAS:DIS?NONEReturns the LASER display value
LAS:DIS:LDINONETurn on/off LASER display to show laser output current
LAS:DIS:LDI?NONEReturns status of laser output current display, on/off
LAS:DIS:LDVNONETurn on/off LASER display to show laser output voltage
LAS:DIS:LDV?NONEReturns status of laser output voltage display, on/off
LAS:DIS:MDINONETurn on/off LASER display to show photodiode monitor
current
LAS:DIS:MDI?NONEReturns status of photo diode monitor current display,
on/off
LAS:DIS:MDPNONETurns on/off LASER display to show photodiode monitor
power
LAS:DIS:MDP?NONERe turns status of photodiode monitor power display,
on/off
LAS:DIS:PARAMNONEEnables LASER display to show parameter values
LAS:DIS:SETNONEShows the LASER display setpoint corresponding to
mode
LAS:DIS:SET?NONEReturns status of LASER setpoint display, on/off
LAS:ENAB:COND1Sets the enable register for LASER conditions
LAS:ENAB:COND?NONEReturns the value of the LAS conditions enable register
LAS:ENAB:EVE1Sets the enable register for LAS events
LAS:ENAB:EVE?NONEReturns the value of the LAS event enable register
LAS:ENAB:OUTOFF1Sets the enable register for LAS conditions, that turn the
LAS output off
LAS:ENAB:OUTOFF?NONEReturns the value of the LAS outoff enable register
LAS:EVENT?NONERe turns the value of the LAS event register
LAS:LDI1Used to set the LASER constant current source setpoint
value
LAS:LDI?NONEUsed to return the constant current source measured
LAS:INC0, 1, or 2Used to increment the LASER setpoint value by one or
LAS:LDV1Used to set the laser voltage value for calibration
LAS:LDV?NONEUsed to return the measured laser voltage value
36 LDX-3232
value
more steps
COMMAND REFERENCE
NameParametersFunction
LAS:LIM:I21Used to set the LASER low range current source limit
LAS:LIM:I2?NONEUsed to return the LASER low range current source limit
LAS:LIM:I41Used to set the LASER high range current source limit
LAS:LIM:I4?NONEUsed to return the LASER high range current source
limit
LAS:LIM:MDP1Used to set the constant optical power (from monitor
photodiode) limit value
LAS:LIM:MDP?NONEUsed to return the optical power (from monitor
photodiode) limit value
LAS:LIM:V1Used to set the LASER compliance voltage limit value
LAS:LIM:V?NONEUsed to return the LASER compliance voltage limit
value
CHAPTER 4
LAS:MDI1Used to set the constant optical power setpoint, if
LAS:MDI?NONEUsed to return the monitor photodiode current
LAS:MDLN1Used to enable/disable the front panel modulation input
LAS:MDLN?NONEUsed to return the status of the front panel modulation
LAS:MDP1Used to set the constant optical power setpoint
LAS:MDP?NONEReturns the actual monitor photodiode power value
LAS:MODE?NONEReturns the mode, I (current), IHBW (high bandwidth),
LAS:MODE:IHBWNONESets the mode to constant current, high bandwidth
LAS:MODE:ILBWNONESets the mode to constant current, low bandwidth
LAS:MODE:MDPNONESets the mode to constant optical power mode
LAS:OUT1Used to enable/disable the current source outpu t
LAS:OUT?NONEReturns the current source output status
LAS:RAN1Sets the laser current output range
LAS:RAN?NONEReturns the laser current output range
LAS:SET:LDI?NONEReturns the laser constant I (current) setpoint
photodiode responsivity is 0
measured value
input
or P (optical power)
LAS:SET:MDI?NONEReturns the laser constant optical power setpoint
LAS:SET:MDP?NONEReturns the constant P (optical power) setpoint
LAS:STEP1Used to set the LASER setpoint step value
LAS:STEP?NONEReturns the LASER setpoint step value
LAS:TOL2Used to set the LASER setpoint tolerance value and
04_06LDX-3232 37
time period value
CHAPTER 4
COMMAND REFERENCE
NameParametersFunction
LAS:TOL?NONEUsed to return the LASER setpoint tolerance value and
time period value
MES1Used to enter a string message of up to 16 bits
MES?NONEReturns a previously stored ASCII message
RAD1Sets the radix type for numerical data; decimal, binary,
RAD?NONEUsed to return the radix type for numerical data
TERM1Used to add a carriage return to the device terminator
TERM?NONEReturns the status of the TERM command
TIME?NONEReturns the elapsed time since the instrument was last
TIMER?NONEReturns the elapsed time since the timer was last reset
octal or hexadecimal
powered up
38 LDX-3232
COMMAND REFERENCE
q
pp
p
p
p
y
(
p
p
(
p
y
y
p
p
p
yp
LDX-3232 Device-Dependent Commands
LDX-3232 Device-Dependent Commands
This section contains all of the device-dependent commands for the LDX-3232
High Compliance Current Source, listed in alphabetical order. Sub-sections for
each path are presented, listing the commands that are legal for that path. See
Figure 3.1 for command path tree structure.
Command Path:Na m e
uired letters are shown in
Re
u
er case; lower case letters
tional.
are o
CHAPTER 4
Front Panel/
Remote Execution
A solid box means the command
can be executed in that mode.
Comm and mnemonics in front
anel (local) mode may vary.
Command Descri
S
ntax Diagram
tion
For Remote Commands)
Parameters
Tells what
ex
ected, and their type.
An
arameters are
<nrf value>
refers to a
flexible numeric notation
IEEE-488.2). Integer,
onential, or floating point
ex
ma
be used with the decimal
radix. Alternate radices ma
selecte d in remote o
be
eration.
Points of Interest
ecial information
Has s
about the comm an d.
Front Panel
Remote
The LASer:LIMit:I2 command sets the laser current
limit value.
SYNTAX DIAGRAM
:
PARAMETERS
POINTS OF
INTEREST
LAS
:
I2
An <nrf value> that represents the LASER limit
current, in mA.
The current limit is in effect in all modes of
operation of the laser output.
In local operation, the limit current is entered
by selecting the LIM I parameter, pressing and
holdi ng in the ( PARAMETER) S ET switch,
adjusting the ADJ US T knob until the desire d v alue
is displayed, and then releasing the SET switch.
LASer:LIMit:I
:
<white
space>
LIM
< nrf
value >
1
les
Exam
Shows t
ical statements
using the command.
04_06LDX-3232 39
EXAMPLES
"Las:Lim:I2 100" - action: the LASER current limit
is set to 100.0 mA.
"LAS : LIM:I 4 3000" - action: the LASER current limit
is set to 3000.0 mA.
Figure 4.1 Command Description Format
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LDX-3232 Device-Dependent Command Reference
The following pages contain a reference for the device-dependent commands of
the LDX-3232 High Compliance Current Source. This reference contains useful
information for both local and remote operation of the LDX-3232.
References to the front panel labels are capitalized in the following reference
pages (as is done throughout this manual). Therefore, a reference to the LOCAL
switch is meant to signify the actual switch labeled LOCAL on the front panel of
the LDX-3232.
In some reference, parentheses are used to signify the labeled are for a particular
switch or LED indicator on the front panel. For example, (PARAMETER) SET
refers to the switch labeled “SET” in the PARAMETER area of the front panel.
DELAY
The DELAY command causes the execution of commands to be delayed by a userdefined time. This command is similar to the *WAI common command, except that
execution resumes after the specified number of miliseconds, instead of waiting for the
Operation-Complete flag to be set.
Syntax Diagram
:
ParametersAn <nrf value> that represents the delay time in milliseconds.
NotesThe Operation-Complete flag is held false until the delay period elapses, but the *OPC?
query will not execute until the delay period has elapsed.
This command is useful for creating delays that don't require a lot of program code and
don't tie up the GPIB during execution.
The delay time is approximate, with an error of about ±10%.
Examples"DELAY 500" -action: Further commands and queries are not executed until about 0.5
seconds have elapsed from the time this command is executed.
“LAS:LDI 22;Delay 2000;LAS:LDI?” -actions: The current is set to 22.0 mA, then the
LDX-3232 High Compliance Current Source waits for about 2.0 seconds before returning
the measured current.
Note: Care should be taken to set the GPIB time-out appropriately for use with the Delay
command.
DELAY
<white
space>
<nrf
value>
FRONT PANEL
REMOTE
40 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
ERRors?
The ERRors? query returns a list of command and device errors that have occurred
since the last query. These errors are notated by a number that corresponds to the type
of error that occurred. See Chapter 3 for information regarding error handling.
Syntax Diagram
:
ParametersNone.
NotesThe response data will be a list of the current errors. The errors are represented by
numbers and are separated by commas. A response of 0 indicates that no errors were
reported. The response data is sent as character data. Up to 10 error codes are stored
between error queries.
Examples"ERR?" -response: 0, means no errors reported.
"Errors?" -response: 201, means that the <PROGRAM DATA> (parameter) value out of
range error was reported since the last ERR? query.
ERR
?
LASer:CALMD
The LASer:CALMD command sets the laser's photodiode feedback responsivity
(the CAL PD parameter).
Syntax Diagram
FRONT PANEL
REMOTE
FRONT PANEL
REMOTE
CALMD
:
ParametersAn <nrf value>, in mA/mW. The range is 0 to 1000.
NotesIf the parameter is set to 0, the LDX-3232 will operate in a constant I
Constant P (MDP) mode is selected.
The parameter should be set to 0 for I
parameter is used to convert between I
mA/mW.
Examples"LAS:CALMD 0" -action: sets the CAL PD parameter to 0. This enables the constant I
mode of operation.
"Laser:Calmd 1" -action: sets the CAL PD parameter to 1.00 mA/mW. Therefore, a
photodiode feedback current of 1 mA will cause the P
04_06LDX-3232 41
LAS
:
CALPD
operation modes. Otherwise, the value of this
PD
and PPD values. The units of this parameter are
PD
<white
space>
PD
display to read 1 mW.
PD
<nrf
value>
mode, when
PD
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:CALMD?
The LASer:CALMD? query returns the value of the laser's photodiode feedback
responsivity (CAL PD parameter) setting.
Syntax Diagram
CALMD
:
NotesIf this value is 0, the LDX-3232 will be set to operate in constant I
setpoint value will be in effect. If this value is non-zero, the LDX-3232 will be set to
operate in constant P
If this value is 0, the front panel P
Examples"LASER:CALMD?" -response: 0, means the LDX-3232 is set for operation in constant
mode (if P mode is also selected).
I
PD
"Las:Calmd?" -response: 1.1, means the LDX-3232 is set for operation in constant PPD
mode (if P mode is also selected) and the responsivity is set to 1.1 µA/mW. 1.1 µA of
photodiode feedback current represents 1 mW of optical power.
LAS
mode, and the PPD setpoint value will be in effect.
PD
:
CALPD
will display "-.-” and no PPD value can be calculated.
PD
FRONT PANEL
REMOTE
?
mode, and the IPD
PD
LAS:CAL:CONST?
The LAS:CAL:CONST? query returns a stored calibration value. It is intended for factory
use only.
FRONT PANEL
REMOTE
42 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:CAL:LDI
The LASer:CAL:LDI command is used to enter the LASER current setpoint,
measurement, and limit (in low bandwidth mode) calibration mode.
Syntax Diagram
:
ParametersNone
NotesSince the limit circuit is the same for both high and low bandwidth modes, it is only
calibrated when low bandwidth mode is selected.
After this command is issued, the LDX-3232 High Compliance Current Source will allow
calibration of the current setpoint, measurement, and limit (if low bandwidth mode is
selected). This procedure is outlined in Chapter 6.
The calibration defaults to the selected bandwidth setting. Therefore, it is necessary to
select the desired bandwidth (I or I
the calibration for that bandwidth.
Calibration is performed at the current setpoint, wherever it is set. If the LASER output is
not on, E516 is displayed.
In remote operation, the LASer:CAL:LDI? query may be used to determine if the
LDX-3232 High Compliance Current Source is ready for the user to enter a value.
Examples"Las:CAL:LdI" -action: the LDX-3232 enters calibration mode for LASER current.
"Las:Cal:LDI" -action: the LDX-3232 enters calibration mode for LASER current.
LAS
::
HBW
CAL
) and turn the LASER output on before performing
FRONT PANEL
REMOTE
LDI
I
04_06LDX-3232 43
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:CAL:LDI?
The LASer:CAL:LDI? query is used to determine that the LDX-3232 is ready for a value
to be entered during the calibration cycle of the LASer:CAL:LDI mode.
Syntax Diagram
:
NotesAfter this query is issued and a response of 1 is received, the LDX-3232 will be ready for
Examples"LASer:CAL:LDI?" -response: 1, means the LDX-3232 is ready for the user to enter a
LAS
the user to enter a current value via the LASer:LDI command (see Chapter 6).
current value via the LASer:LDI command.
“LASer:Cal:LdI?” -response: 0, means the LDX-3232 is not yet ready for the user to enter
a LASER current value
:
CAL
:
LASer:CAL:LDV
The LASer:CAL:LDV command is used to enter the LASER voltage measurement
calibration mode.
Syntax Diagram
FRONT PANEL
REMOTE
LDI
?
I
FRONT PANEL
REMOTE
ParametersNone
NotesAfter this command is issued, the LDX-3232 will allow calibration of the laser voltage
Examples"Las:CAL:Ldv" -action: the LDX-3232 enters calibration mode for LASER voltage.
44 LDX-3232
:
measurement. This procedure is outlined in Chapter 6.
The LASer:CAL:LDV? query may be used to determine if the LDX-3232 is ready for the
user to enter a value.
"Las:Cal:LDV" -action: the LDX-3232 enters calibration mode for LASER voltage.
LAS
::
CAL
LDV
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
:
FRONT PANEL
REMOTE
LDV
FRONT PANEL
REMOTE
?
LASer:CAL:LDV?
The LASer:CAL:LDV? query is used to determine that the LDX-3232 is ready for a value
to be entered during the calibration cycle of the LASer:CAL:LDV mode.
Syntax Diagram
:
NotesAfter this query is issued and a response of 1 is received, the LDX-3232 will be ready for
the user to enter a voltage value via the LASer:LDV command (see Chapter 6).
Examples"LASer:CAL:LDV?" -response: 1, means the LDX-3232 is ready for the user to enter a
voltage value via the LASer:LDV command.
"LASer:Cal:LdV?" -response: 0, means the LDX-3232 is not yet ready for the user to
enter a LASER voltage value.
LAS
:
CAL
LASer:CAL:MDI
The LASer:CAL:MDI command is used to enter the LASER photodiode current
calibration mode.
Syntax Diagram
IPD
:
ParametersNone
NotesAfter this command is issued, the LDX-3232 will automatically enter the LASER
photodiode current calibration mode. When the LDX-3232 is ready, the user should enter
the true measured value. This procedure is outlined in Chapter 6.
If the LASER output is not ON or if the P mode is not selected, error code E-516 will be
generated.
In remote operation, the LASer:CAL:MDI? query may be used to determine if the
LDX-3232 is ready for the user to enter a value via the LASer:MDI command.
Examples"Las:CAL:MDI" -action: the LDX-3232 enters the LASER photodiode current calibration
mode.
"LAS:Cal:MDI" -action: the LDX-3232 enters the LASER photodiode current calibration
mode.
LAS
::
CAL
MDI
04_06LDX-3232 45
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:CAL:MDI?
The LASer:CAL:MDI? query is used to determine that the LDX-3232 is ready for a value
to be entered during the calibration cycle of the LASer:CAL:MDI mode.
Syntax Diagram
:
NotesAfter this query is issued and a response of 1 is received, the LDX-3232 be ready for the
Examples"LAS:CAL:MDI?" -response: 1, means the LDX-3232 is ready for the user to enter a
LAS
user to enter a photodiode current value via the LASer:MDI command (see Chapter 6).
photodiode current value via the LASer:MDI command.
:
CAL
:
LASer:COND?
The LASer:COND? query returns the value of the status condition register of the LASER
operations.
Syntax Diagram
FRONT PANEL
REMOTE
MDI
?
IPD
FRONT PANEL
REMOTE
46 LDX-3232
:
-where the response is the sum of the following:
1LASER Current Limit256Output is shorted
2LASER Voltage Limit512Output is outside tolerance limit
4N/A1024Output on/off state
8Power limit2048Ready for calibration data state
16Interlock disabled4096N/A
32N/A8192N/A
64Compliance Voltage Limit16384N/A
128Open circuit32768N/A
NotesThe LASER conditions that are reported to the status byte are set via the
LASer:ENABle:COND command.
The Open circuit condition is only present while the LASER output is on, and when the
hardware detects this condition, it will turn the LASER output off. Therefore, the Open
LAS
:
COND
?
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
Circuit condition is fleeting and may be missed via the LAS:COND? query . Therefore, the
user should test for the Open Circuit Event via the LAS:EVEnt? query.
The LASER condition status is constantly changing, while the event status is only
cleared when the event status is read or the *CLS command is issued.
Examples"LAS:COND?" -response: 513, means that the LASER limit current and out of tolerance
LASER conditions currently exist.
"Radix Hex; Laser:Cond?" -response: #H108, means that the LASER Output shorted
and Power limit conditions currently exist.
CHAPTER 4
<nrf
value>
FRONT PANEL
REMOTE
,
<nrf
value>
LASer:DEC
The LASer:DEC command decrements the selected laser control mode setpoint by one
or more steps. Optional parameters allow multiple steps to be decremented and the time
(in milliseconds) between decrements to be set, respectively. This is an overlapped
command. Other commands may be executed while this function is in progress.
Syntax Diagram
:
Parameters0, 1, or 2. The first optional parameter is the number of steps, and the second optional
NotesThe decremental default amount is one step. The step size can be edited via the
Examples"LAS:MODE:I; LAS:STEP 3; LAS:DEC" -action: The laser source current setpoint is
LAS
parameter is the number of milliseconds between steps.
LAS:STEP command. LDX-3232 default values are 0. 1 mA, 0.1 mW, or 1 µA
(if CALMD = 0).
If the first optional parameter is used, but not the second, the user may decrement the
LASER setpoint by a multiple of the LAS:STEP size, without changing the LAS:STEP
size.
If the both optional parameters are used, the user may create an automated stepping
ramp function for the LASER output.
If the first optional parameter is entered as zero, "LAS:DEC 0", the command will do
nothing.
The minimum time to complete one decrement is about 20 ms. Therefore, values for the
second optional parameter (time between decrements) have a practical minimum of 20.
decremented by 0.3 mA.
"LAS:MODE:I; LAS:STEP 3; LAS:DEC 3" -action: The laser source current setpoint is
decremented by three times the amount described in the first example.
"LAS:MODE:I; LAS:STEP 3; LAS:DEC 3,5000" -action: The laser source current
setpoint is decremented by the amount described in the first example, three times, with
five seconds between decremental steps.
:
DEC
<white
space>
04_06LDX-3232 47
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
FRONT PANEL
REMOTE
LASer:DISplay
The LASer:DISplay command enables or disables (turns off) the display and some LEDs
Syntax Diagram
:
ParametersAn <nrf value>; -where 1 = on, 0 = off.
NotesT urning the display and LEDs off means that a message of all blank spaces is sent to the
display, and most of the LEDs will be turned off.
Examples"las:dis 1" -action: turns the LASER display on and enables the LASER indicator LEDs.
"Laser:dis Off" -action: turns the LASER display and disables the LASER indicator
LEDs.
LASDIS
:
LASer:DISplay?
The LASer:DISplay? query returns the value shown on the display.
Syntax Diagram
:
NotesReturns the actual (6-character) string from the output buffer to the display. If the display
is disabled, it returns " .".
Examples"LAS:DIS?" -response: "- 99.9", means the LASER display shows "- 99.9".
"Laser:DISp?" -response: " 0.6", means the LASER display shows " 0.6".
LAS DIS
:
?
48 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:DISplay:LDI
The LASer:DISplay:LDI command sets the laser display to show the constant current
measurement.
Syntax Diagram
:
ParametersNone.
NotesThe LDI measurement display mode is turned off automatically when another display
selection is enabled.
Examples"LAS:DIS:LDI" -action: enables the display for current values.
LAS
::
DIS
LASer:DISplay:LDI?
The LASer:DISplay:LDI? query returns the status of the LDI measurement display mode.
Syntax Diagram
FRONT PANEL
REMOTE
I
LDI
FRONT PANEL
REMOTE
LDI
:
NotesIn local operation, the status of the LDI measurement display mode is indicated by the
Examples"LAS:DIS:LDI?" -response: 0, means that the LDI measurement is not displayed.
LAS
- where the response is 0 or 1.
enunciator next to the display (mA will be illuminated and not blinking when the display is
in LDI measurement mode).
"Las:dis:ldI?" -response: 1, means that the LDI measurement is displayed.
:
DIS
:
I
?
04_06LDX-3232 49
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:DISplay:LDV
The LASer:DISplay:LDV command sets the laser display to show the laser forward
voltage measurement.
Syntax Diagram
:
ParametersNone
NotesThe LDV measurement mode is turned off automatically when another display selection
is enabled.
Examples"LAS:DIS:LDV" -action: enables the display for LDV measurement mode.
LAS
::
DIS
LASer:DISplay:LDV?
The LASer:DISplay:LDV? query returns the status of the LDV (laser diode voltage)
measurement display mode.
Syntax Diagram
FRONT PANEL
REMOTE
V
LDV
FRONT PANEL
REMOTE
50 LDX-3232
LDV
:
NotesIn local operation, the status of the LDV measurement mode is indicated by the
Examples"LAS:DIS:LDV?" -response: 0, means that laser voltage measurement is not displayed.
LAS
- where the response is 0 or 1.
enunciator next to the display (V will be illuminated when the display is in LDV
measurement mode).
"LAS: DIS:LDV?" -response: 1, means that the laser voltage measurement is displayed.
:
DIS
:
V
?
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:DISplay:MDI
The LASer:DISplay:MDI command sets the laser display to show the monitor photodiode
current measurement.
Syntax Diagram
:
ParametersNone
NotesThe MDI measurement display is turned off automatically when another display selection
is enabled.
Examples"LAS:DIS:MDI" -action: enables the display for photodiode current measurement values.
Laser:Disp:Mdi" -action: enables the LASER display for photodiode current
measurement values.
LAS
::
DIS
LASer:DISplay:MDI?
The LASer:DISplay:MDI? query returns the status of the (LASER DISPLAY) I
Syntax Diagram
FRONT PANEL
REMOTE
IPD
MDI
FRONT PANEL
REMOTE
switch.
PD
MDI
:
NotesIn local operation, the status of the MDI measurement mode is indicated by the
Examples"LAS:DIS:MDI?" -response: 0, means that the laser photodiode monitor current
04_06LDX-3232 51
LAS
- where the response is 0 or 1
enunciator next to the display (µA will be illuminated and not blinking when the display is
in MDI measurement mode).
measurement is not displayed.
"Las:dis:MDI?" -response: 1, means that the monitor photodiode current measurement
is displayed.
:
DIS
:
IPD
?
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:DISplay:MDP
The LASer:DISplay:MDP command sets the display to show the monitor photodiode
power measurement.
Syntax Diagram
:
ParametersNone
NotesThe MDP measurement display is turned off automatically when another display
selection is enabled.
Examples"Las:Display:MDp" -action: enables the display for photodiode power measurement
values.
"Laser:Disp:MDP" -action: enables the display for photodiode power measurement
values.
LAS
::
DIS
LASer:DISplay:MDP?
The LASer:DISplay:MDP? query returns the status of the MDP (monitor diode power)
measurement display mode.
Syntax Diagram
FRONT PANEL
REMOTE
PPD
MDP
FRONT PANEL
REMOTE
52 LDX-3232
MDP
:
NotesIn local operation, the status of the MDP measurement mode is indicated by the
Examples"LAS:DIS p:MDp?" -response: 0, means that the MDP measurement is not displayed.
LAS
- where the response is 0 or 1.
enunciator next to the display (mW will be illuminated and not blinking when the display
is in MDP measurement mode).
"Las:dis:MDP?" -response: 1, means that the MDP measurement is displayed.
:
DIS
:
PPD
?
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:DISplay:PARAM
The LASer:DISplay:PARAM command enables the display to show the parameter
values.
Syntax Diagram
:
ParametersNone
NotesThis command has the same effect as pressing the (PARAMETER) SELECT switch
while in LASER mode. The selected parameter will be displayed for three seconds. The
actual LASER PARAM display is turned off automatically when a display selection is
enabled.
Each time the command is issued, the next LASER parameter will be selected, see
Chapter 2 for more information.
Examples"Laser:Display:Param" -action: selects a parameter and displays its value.
"LAS:DIS:PARAM" -action: selects a parameter and displays its value.
LAS
::
DIS
LASer:DISplay:SET
The LASer:DISplay:SET command sets the display to show the setpoint of the selected
display mode.
Syntax Diagram
FRONT PANEL
REMOTE
PARAM
FRONT PANEL
REMOTE
:
ParametersNone
NotesUsing this command has the same ef fect as physically holding the (DISPLAY) SELECT
switch down continuously.
Examples"Las:Dis:Set" -action: enables the display for the setpoint of the selected mode: LDI,
MDI (I
"LAS:Dis:Set" -action: enables the display for the setpoint of the selected display mode.
04_06LDX-3232 53
) or MDP (PPD)
PD
LAS
::
DIS
SET
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
:
FRONT PANEL
REMOTE
SET
FRONT PANEL
REMOTE
?
LASer:DISplay:SET?
The LASer:DISplay:SET? query returns the status of the setpoint display mode.
Syntax Diagram
:
- where the response is 0 or 1.
NotesThe setpoint display will not time out when REMOTE operation is used. (It will be
continuously displayed.)
Examples"LAS:DISPLAY:SET?" -response: 0, means the setpoint is not shown on the display.
"Las:Dis:Set?" -response: 1, means the setpoint value is shown on the display.
LAS
:
DIS
LASer:ENABle:COND
The LASer:ENABle:COND command sets the condition status enable register of the
LASER operations for summary (in bit 3 of the status byte) and generation of service
requests.
Syntax Diagram
:
ParametersAn <nrf value> whose sum represents the enabled bits:
1LASER Current Limit256Output is shorted
2LASER Voltage Limit512Output is outside tolerance limit
4N/A1024Output on/off state
8Power limit2048Ready for calibration data state
16Interlock disabled4096N/A
32N/A8192N/A
64Compliance Voltage Limit16384N/A
128Open circuit32768N/A
NotesThe enabled or disabled LASER conditions can be read by using the
LAS
LASer:ENABle:COND? query.
The LASER condition status can be monitored by the LASer:COND? query. If any of the
enabled LASER conditions are true, bit 3 of the status byte register will be set.
The enable registers normally retain their values at power-up (as they were at power-
down) unless the power-on status clear flag is set true (for more information see the
*PSC definition, in the GPIB-IEEE488 specification).
:
ENAB
:
COND
<white
space>
<nrf
value>
54 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
Examples"LAS:ENAB:COND 129" -action: enables the LASER status condition register so that the
Open circuit and LASER current limit conditions will be summarized in the status byte
(bit 3).
Laser:Enable:Cond #H0F9B" - action: enables the LASER status condition register so
that any and all of the above conditions will be reported in the status byte register (bit 3).
CHAPTER 4
:
FRONT PANEL
REMOTE
COND
?
LASer:ENABle:COND?
The LASer:ENABle:COND? query returns the value of the status condition enable
register of the LASER operations.
Syntax Diagram
:
- where the response is the sum of the following:
1LASER Current Limit256Output is shorted
2LASER Voltage Limit512Output changed to be in/out of tolerance
4N/A1024Output on/off state
8Power limit2048Ready for calibration data state
16Interlock disabled4096N/A
32N/A8192N/A
64Compliance Voltage Limit16384N/A
128Open circuit32768N/A
NotesThe enabled LASER conditions can be set by using the LASer:ENABle:COND
command.
The LASER condition status can be monitored by the LASer:COND? query.
Examples"LAS:ENAB:COND?" -response: 17, means that the Laser Current Limit and Interlock
disabled LASER conditions will be reported (in summarized form) to the status byte
(bit 3).
"Radix Hex; Laser:Enable:Cond?" -response: #H0F9B, means that all of the above
conditions will be reported (in summarized form) to the status byte (bit 3).
LAS
:
ENAB
04_06LDX-3232 55
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:ENABle:EVEnt
The LASer:ENABle:EVEnt command sets the status event enable register of the LASER
operations. These events are summarized in bit 2 of the status byte register.
Syntax Diagram
:
ParametersAn <nrf value> whose sum represents the bits that are enabled:
1LASER Current Limit256Output is shorted
2LASER Voltage Limit512Output changed to be in/out of tolerance
4N/A1024Output on/off state changed
8Power limit2048New measurements taken
16Interlock changed state4096N/A
32N/A8192N/A
64Compliance Voltage Limit16384N/A
128Open circuit32768N/A
NotesThe enabled LASER events can be read by using the LASer:ENABle:EVEnt? query. The
Examples"LAS:ENAB:EVENT 136" -action: enables the LASER status event register so that the
LAS
LASER event status can be monitored by the LASer:EVEnt? query.
The enable registers normally retain their values at power-up (as they were at power-
down) unless the power-on status clear flag is set true (for more information see the
*PSC definition, in the GPIB-IEEE488 specification).
Open circuit and Power limit events will be reported (in summarized form) to the status
byte (bit 2).
"Laser:Enable:Event #H0F9B" -action: enables the LASER status event register so all of
the above events will be reported (in summarized form) to the status byte (bit 2).
:
ENAB
:EVE
FRONT PANEL
REMOTE
<white
space>
<nrf
value>
56 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
:
FRONT PANEL
REMOTE
EVE
?
LASer:ENABle:EVEnt?
The LASer:ENABle:EVEnt? query returns the value of the status event enable register of
the LASER operations.
Syntax Diagram
:
- where the response is the sum of the following:
1LASER Current Limit256Output is shorted
2LASER Voltage Limit512Output changed to be in/out of tolerance
4N/A1024Output on/off state
8Power limit2048New measurements taken
16Interlock state changed4096N/A
32N/A8192N/A
64Compliance Voltage Limit16384N/A
128Open circuit32768N/A
NotesThe enabled LASER events can be set by using the LASer:ENABle:EVEnt command.
The LASER event status can be monitored by the LASer:EVEnt? query.
Examples"LAS:ENAB:EVE?" -response: 1040, means that the Output on/off state change and
Interlock changed LASER events will be reported (in summarized form) to the status byte
register (bit 2).
"Radix Hex; Las:Enab:Eve?" -response: #HFDB, means that all of the above events will
be reported (in summarized form) to the status byte register (bit 2).
LAS
:
ENAB
04_06LDX-3232 57
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:ENABle:OUTOFF
The LASer:ENABle:OUTOFF command sets the status outoff enable register of the
LASER operations (things that will turn the LASER output off).
Syntax Diagram
:
ParametersAn <nrf value> whose sum represents the enabled bits:
1LASER Current Limit256N/A
2LASER Voltage Limit512Output is out of tolerance*
4N/A1024N/A
8Power limit (with output on)2048N/A
16N/A4096N/A
32N/A8192N/A
64N/A16384N/A
128N/A32768N/A
NotesThe enabled LASER outoff bits can be read by using the LASer:ENABle:OUTOFF?
Examples"LAS:ENAB:OUTOFF 9" -action: enables the LASER status outoff register so that Power
LAS
query.
The enable registers normally retain their values at power-up (as they were at power-
down) unless the power-on status clear flag is set true (for more information see the
*PSC definition, in the GPIB-IEEE488 specification).
The factory default value for this register is #B0000100000001000, #H808, or 2056
decimal.
limit and LASER current limit conditions will cause the LASER output to be turned off.
"Las:Enab:Outoff #HE0B" -action: enables the LASER status outoff register so that any
or all of the above conditions will cause the LASER output to be turned off.
:
ENAB
OUTOFF
:
FRONT PANEL
REMOTE
<white
space>
<nrf
value>
58 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
.
CHAPTER 4
:
FRONT PANEL
REMOTE
OUTOFF
?
LASer:ENABle:OUTOFF?
The LASer:ENABle:OUTOFF? query returns the value of the status outoff enable register
of the LASER operations.
Syntax Diagram
:
-where the response is the sum of the following:
1LASER Current Limit256N/A
2LASER Voltage Limit512Output if out of tolerance
4N/A1024N/A
8Power limit (with output on)2048N/A
16N/A4096N/A
32N/A8192N/A
64N/A16384N/A
128N/A32768N/A
NotesThe enabled LASER events can be set by using the LASer:ENABle:OUTOFF c ommand.
The LASER output status can be monitored by the LASer:EVEnt? query.
Examples"LAS:ENAB:OUTOFF?" -response: 1, means that Current Limit Condition will cause the
LASER output to be turned off.
"Radix Hex; Las:Enab:Eve?" -response: #H20B, means that all of the above conditions
will cause the LASER output to be turned off.
LAS
:
ENAB
04_06LDX-3232 59
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:EVEnt?
The LASer:EVEnt? query returns the value of the status event register of the LASER
operations.
Syntax Diagram
:
-where the response is the sum of the following:
1LASER Current Limit256Output is shorted
2LASER Voltage Limit512Output changed to be in/out of tolerance
4N/A1024Output on/off state changed
8Power limit2048Measurement ready
16Interlock disabled4096N/A
32N/A8192N/A
64Compliance Voltage Limit16384N/A
128Open circuit32768N/A
NotesThe LASER conditions that are reported in the status byte can be set by using the
LASer:ENABle:EVEnt command.
The LASER event status is only cleared when the event status is read or by the *CLS
command while the condition status is constantly changing.
Examples"LAS:EVE?" -response: 513, means that the LASER output tolerance changed and
current limit events have occurred since the last LASer:EVEnt? query.
"Radix Hex; Laser:Event?" -response: #H88, means that the LASER Power limit and
Open circuit events have occurred since the last LASer:EVEnt? query.
LAS EV E
:
FRONT PANEL
REMOTE
?
60 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
<nrf
value>
FRONT PANEL
REMOTE
,
<nrf
value>
LASer:INC
The LASer:INC command increments the selected laser control mode setpoint by one or
more steps. Optional parameters allow multiple steps to be incremented and the time (in
ms) between increments to be set, respectively. This is an overlapped command. Other
commands may be executed while this function is in progress.
Syntax Diagram
:
Parameters0, 1, or 2. The first optional parameter is the number of steps, and the second optional
NotesThe decremental default amount is one step. The step size can be edited via the
Examples“LAS:MODE:I; LAS:STEP 3; LAS:INC” -action: The laser source current setpoint is
LAS
parameter is the number of ms between steps.
LAS:STEP command. LDX-3232 default values are 0. 1 mA, 0.1 mW, or 1 µA.
If the first optional parameter is used, but not the second, the LASER setpoint is
incremented by a multiple of the LAS:STEP size without changing the LAS:STEP size.
The second optional parameter is the time, in ms, between steps.
If both optional parameters are used, the user may create an automated stepping ramp
function for the LASER output.
If the first optional parameter is entered as zero, “LAS:INC 0”, the command will do
nothing.
The minimum time to complete one increment is about 20 ms. Therefore, values for the
second optional parameter (time between increments) have a practical minimum of 20.
This command should not be overlapped with “LAS:DEC” or other “LAS:INC” commands.
incremented by 0.03 mA.
“LAS:MODE:I; LAS:STEP 3; LAS:INC 3” -action: The laser source current setpoint is
incremented by three times the amount described in the first example.
“LAS:MODE:I; LAS:STEP 3; LAS:INC 3,5000” -action: The laser source current setpoint
is incremented by the amount described in the first example, three ti mes, with
fiveseconds between incremental steps.
“LAS:STEP 1; LAS:Mode:P; Las:INC” -action: The power setpoint is incremented by
0.01 mW.
:
INC
<white
space>
04_06LDX-3232 61
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:LDI
The LASer:LDI command sets the laser control current.
Syntax Diagram
I
:
ParametersAn <nrf value> that represents the (laser) output current in mA.
Notessetpoint is the same for both low and high bandwidth output modes.
Examples"Las:LDI 400" -action: sets the laser output current to 400.00 mA.
"LAS:ldi 100" -action: sets the laser output current to 100.0 mA.
LAS
:
LDI
LASer:LDI?
The LASer:LDI? query returns the value of the measured laser current.
Syntax Diagram
FRONT PANEL
REMOTE
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
I
:
- where the response is the most recent laser current measurement.
NotesResponse is the measured laser output current for either low or high bandwidth modes.
This measurement is updated approximately once every 600 ms.
Examples"LAS:ldi?" -response: 30.0, means the measured laser output current is 30.0 mA.
"Laser:LDI?" -response: 100.0, means the measured laser output current is 100.0 mA.
LAS
:
LDI
?
62 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
FRONT PANEL
REMOTE
LASer:LDV
The LASer:LDV command sets the laser voltage for calibration of the laser voltage
measurement.
Syntax Diagram
:
ParametersAn <nrf value> that represents the (laser) output voltage, in volts.
Notessetpoint is the same for both low and high bandwidth output modes and is entered only
for LDV calibration.
Examples"Las:LDV 4" -action: enters a value of 4.000 V.
"LAS:ldv 1.025" -action: enters a value of 1.025 V.
LASLDV
:
LASer:LDV?
The LASer:LDV? query returns the value of the measured laser voltage.
Syntax Diagram
:
- where the response is the measured laser output voltage for either low or high
bandwidth modes.
NotesThis measurement is updated approximately once every 600 ms.
Examples"LAS:ldv?" -response: 3.03, means the measured laser output voltage is 3.03 V.
"Laser:LDV?" -response: 1.0, means the measured laser output voltage is 1.000 V.
LAS
:
LDV
?
04_06LDX-3232 63
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:LIMit:I2
The LASer:LIMit:I2 command sets the LASER current limit value for the 2000 mA range
on the LDX-3232.
Syntax Diagram
:
ParametersAn <nrf value> that represents the LASER limit current , in mA .
NotesThe current limit is in effect in all modes of operation (ILBW, IHBW , or P). The current
Examples"LAS:LIM:I2 180" -action: the LASER current limit is set to 180 mA.
LAS
limit in use depends on the LASER output range selection.
":Laser:Limit:I2 160" -action: the LASER current limit is set to 160 mA.
:
LIM
:
I2
LASer:LIMit:I2?
The LASer:LIMit:I2? query returns the value of the LASER current limit for the 2000 mA
range on the LDX-3232.
Syntax Diagram
FRONT PANEL
REMOTE
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
:
- where the response is the current limit value for the 2000 mA range.
NotesThe current limit is in effect in all modes of operation (ILBW, IHBW , or P). The current
limit in use depends on the LASER output range selection.
Examples"LAS:LIM:I2?" -response: 40, means the laser current limit is 40 mA.
"Laser:LIM:I2?" -response: 50, means the laser current limit is 50 mA .
LAS
:
LIM
:
I2
?
64 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:LIMit:I4
The LASer:LIMit:I4 command sets the LASER current limit value for the 4000 mA range
on the LDX-3232.
Syntax Diagram
:
ParametersAn <nrf value> that represents the LASER limit current , in mA .
NotesThe current limit is in effect in all modes of operation (ILBW, IHBW , or P). The current
Examples"LAS:LIM:I4 1800" -action: the LASER current limit is set to 1800 mA.
LAS
limit in use depends on the LASER output range selection.
In local operation, the limit current is entered by selecting LIM I parameter, pressing and
holding in the (PARAMETER) SET switch, adjusting the knob until the desired value is
displayed, and then releasing the SET switch.
":Laser:Limit:I4 160" -action: the LASER current limit is set to 160 mA.
:
LIM
:
I4
LASer:LIMit:I4?
The LASer:LIMit:I4? query returns the value of the LASER current limit for the 4000 mA
on the LDX-3232.
Syntax Diagram
FRONT PANEL
REMOTE
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
:
- where the response is the current limit value for the 4000 mA range.
NotesThe current limit is in effect in all modes of operation (ILBW, IHBW , or P). The current
limit in use depends on the LASER output range selection.
Examples"LAS:LIM:I4?" -response: 400, means the laser current limit is 400 mA.
"Laser:LIM:I4?" -response: 1500, means the laser current limit is 1500 mA.
LAS
:
LIM
:
I4
?
04_06LDX-3232 65
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:LIMit:MDP
The LASer:LIMit:MDP command sets the laser monitor photodiode power limit value.
Syntax Diagram
MDP
:
ParametersAn <nrf value> that represents the laser monitor photodiode power limit in mW.
NotesWhen constant MDP mode is used, the output level is limited only by the LIM Ix value.
Examples"LAS:LIM:MDP 10" -action: sets the laser output power limit to a value that corresponds
LAS
Exceeding the powere limit will either cause the output to shut off (default) or generate a
warning (see LAS:ENAB:OUTOFF command).
to producing 10.00 mW of photodiode feedback (optical) power.
"Las:Limit:MDp 5" -action: sets the laser output power limit to a value that corresponds to
producing 5.00 mW of photodiode feedback (optical) power .
::
LIM
P
LASer:LIMit:MDP?
The LASer:LIMit:MDP? query returns the value of the laser monitor photodiode power
limit.
Syntax Diagram
FRONT PANEL
REMOTE
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
66 LDX-3232
MDP
:
NotesThe (LASER) MDP limit is in effect for both laser output current ranges.
Examples"LAS:LIM:MDP?" -response: 3.0, means the monitor photodiode power limit is set to
LAS
- where the response is the LASER power limit setting.
The MDP limit is not in effect when CALMD=0. In this case, the monitor current
measurement is not converted to power.
3.0 mW.
":LAS:Limit:MDP?" -response: 10.0, means the monitor photodiode power limit is set to
10.0 mW.
:
LIM
:
P
?
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:LIMit:V
The LASer:LIMit:V command sets the LASER compliance voltage limit value.
Syntax Diagram
:
ParametersAn <nrf value> that represents the LASER limit voltage, in volts.
NotesThe voltage limit setting is useful for laser protection. When the maximum operating
Examples"LAS:LIM:V 5.0" -action: the LASER compliance voltage limit is set to 5.0 V.
LAS
voltage of a laser is known, the user may set the voltage limit to a value slightly higher
than the maximum operating voltage. Then, if the laser is accidentally disconnected, the
current source will quickly sense the over-voltage and shut off.
In local operation, the voltage limit is entered by selecting the COMP V parameter,
pressing and holding in the (PARAMETER) SET switch. Turn the adjust knob until the
desired value is displayed and then release the SET switch.
":Laser:Limit:V 4.60" -action: the LASER compliance voltage limit is set to 4.6 V.
:
LIM
:
V
LASer:LIMit:V?
The LASer:LIMit:V? query returns the value of the LASER compliance voltage limit
setting.
Syntax Diagram
FRONT PANEL
REMOTE
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
:
- where the response is the LASER limit voltage setpoint.
NotesThe voltage limit is valid for all modes of Laser operation.
In local operation, the voltage limit is seen by selecting the COMP V parameter , pressing
and holding in the (PARAMETER) SELECT switch.
Examples"LAS:LIM:V?" -response: 4.5, means the laser compliance voltage limit is 4.5 V.
"Laser:LIM:V?" -response: 10.0, means the laser compliance voltage limit is set to
10.0 V.
04_06LDX-3232 67
LAS
:
LIM
:
V
?
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:MDI
The LASer:MDI command sets the value of the optical power setpoint, in µA, if the
CALMD (CAL PD) responsivity is 0.
Syntax Diagram
MDI
:
ParametersAn <nrf value> that represents the photodiode feedback current, in µA.
NotesIf the CALMD (CAL PD) parameter is not set to 0, the LAS:MDI value will not be used. In
this case, the measured MDI would be converted to MDP (P
and the MDP (P
Examples"Las:Mdi 40" -action: The LASER output is controlled so that the photodiode feedback
current remains constant at 40 µA.
"Laser:MDI 200" -action: The LASER output is controlled so that the photodiode
feedback current remains constant at 200 µA.
LAS
) setpoint would be used.
PD
:
IPD
LASer:MDI?
The LASer:MDI? query returns the value of the laser photodetector current measurement
in µA.
Syntax Diagram
FRONT PANEL
REMOTE
<white
space>
, by the CAL PD factor),
PD
FRONT PANEL
REMOTE
<nrf
value>
NotesThe response is in µA. The response is valid, even when the unit is not in constant P
Examples"Las:Mode:MDP; Las:Calmd 0; Las:MDi?" -response: 100.0, means 100 µA of
68 LDX-3232
MDI
:
- where the response is the most recent LASER monitor current measurement.
mode.
This measurement is updated approximately once every 600 ms.
photodetector current. This feedback is controlling the laser current output.
"LAS:MODE:IHBW; LAS:MDI?" -response: 20.0, means 20 µA of photodetector current,
but photodiode monitor current is not controlling the laser output current.
LAS
:
IPD
?
COMMAND REFERENCE
p
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:MDLN
The LASer:MDLN command enables or disables (turns off) the front panel modulation
input (BNC).
Syntax Diagram
:
ParametersAn <nrf value> - where 1=on, 0=off.
NotesAfter turning the modulation off, it will remain off until turned back on via this command or
by resetting the instrument.
Examples“las:mdln 1” -action: turns the modulation input on and enables modulation via the front
panel BNC connector.
“Laser:mdln 0” -action: turns the modulation input off. Modulation signals at the front
panel BNC will not effect the LAS output.
LASMDLN
:
LASer:MDLN?
The LASer:MDLN? query returns that status of the LAS modulation control. When
disabled, a modulation signal at the front panel BNC will not effect the LAS output.
Syntax Diagram
FRONT PANEL
REMOTE
<white
s
ace>
FRONT PANEL
REMOTE
<nrf
value>
:
- where the response of 1=on, 0=off.
NotesThe LASER modulation control switches the front panel BNC in and out of the LAS
control circuit. At power up or after reset, modulation is enabled by default.
Examples“LAS:MDLN?” -response: 1, means that the LAS modulation input is enabled.
“las:mdln?” -response: 0, means that the LAS modulation input is disabled.
04_06LDX-3232 69
LAS
:
MDLN
?
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:MDP
The LASer:MDP command sets the value of the optical power setpoint in mW if the
CALMD (CAL PD) responsivity is greater than 0.
Syntax Diagram
MDP
:
ParametersAn <nrf value> that represents the photodiode feedback power in mW.
NotesIf the CALMD (CAL PD) parameter is set to 0, the LAS:MDP value will not be used. In
this case, the measured MDP would be invalid, and the MDI (I
Examples"Las:Mdp 40" -action: The LASER output is controlled so that the photodiode feedback
power remains constant at 40 mW.
"Laser:MDP 200" -action: The LASER output is controlled so that the photodiode
feedback power remains constant at 200 mW.
LAS
:
P
LASer:MDP?
The LASer:MDP? query returns the value of the laser photodetector power measurement
in mW.
FRONT PANEL
REMOTE
<white
space>
) setpoint would be used.
PD
FRONT PANEL
REMOTE
<nrf
value>
70 LDX-3232
MDP
:
- where the response is the most recent LASER power measurement.
NotesThe response is in mW. The response is valid, even when the unit is not in constant P
mode. This measurement is updated approximately once every 600 ms.
Examples"Las:Mode:MDP; Las:Calmd 460; Las:MDp?" -response: 100.0, means 100 mW of
photodetector power. This feedback is controlling the laser current output.
"LAS:MODE:IHBW; LAS:MDP?" -response: 20.0, means 20 mW of photodetector
power, but photodiode monitor current is not controlling the laser output current.
LAS
:
P
?
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:MODE?
The LASer:MODE? query returns the selected laser control mode.
:
NotesI
Examples"LAS:MODE?" -response: ILBW, means that constant I (current) mode is in effect for the
mode is the same as I mode (low bandwidth), except that the output low bandpass
HBW
filter is disabled in I
laser output.
":Las:Mode?" -response: MDP, means that constant P (power) mode is in effect for the
laser output, and CALMD>0.
"Las:Mode?" -response: MDI, means that constant P (power) mode is in effect for the
laser output, and CALMD=0.
"Laser:MODE?" -response: IHBW, means that constant I
mode is in effect for the laser output.
HBW
LAS
mode.
:
MODE
LASer:MODE:IHBW
The LASer:MODE:IHBW command selects laser high bandwidth constant current mode.
FRONT PANEL
REMOTE
?
(current, high bandwidth)
HBW
FRONT PANEL
REMOTE
:
ParametersNone
NotesThis mode of operation is constant I mode with the output lowpass filter disabled.
Examples"LAS:Mode:Ihbw" -action: enables the laser high bandwidth constant current mode.
04_06LDX-3232 71
LAS
::
MODE
IHBW
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:MODE:ILBW
The LASer:MODE:ILBW command selects laser constant current mode.
:
ParametersNone
NotesConstant I mode (low bandwidth) enables the output low bandpass filter.
Examples"LAS:MODE:ILBW" -action: sets the laser output for constant I mode (low bandwidth).
LAS
::
MODE
LASer:MODE:MDP
The LASer:MODE:MDP command selects laser constant power mode.
FRONT PANEL
REMOTE
I
ILBW
FRONT PANEL
REMOTE
P
:
ParametersNone
NotesThis mode of laser operation requires the laser's monitor photodiode feedback to
maintain constant optical power or constant monitor current.
In this mode, the displayed parameter will be either I
CALMD>0) in mW.
Examples"LAS:MODE:MDP" -action: sets the laser output mode of operation to constant optical
power mode.
LAS
::
MODE
(if CALMD=0) in µA, or PPD (if
PD
MDP
72 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
FRONT PANEL
REMOTE
LASer:OUTput
The LASer:OUTput command turns the laser output on or off.
:
ParametersAn <nrf value>; 1=on, 0=off.
NotesThere is a two-second delay after the output is turned on before the output shorting relay
is opened. Following this, the output current ramps up to the current setpoint slowly to
protect the laser. Therefore, the time to turn the output on and reach the setpoint current
varies from two to about three seconds.
After the output is turned on, it may be useful to wait until the output is stable (within
tolerance) before performing further operations, but it is not necessary . When the LASER
output is off, it is safe to connect or disconnect devices to the LASER output terminals.
When the LASER output is off, an internal short is placed across the output terminals.
Examples"LAS:I 20; LAS:OUT ON" -action: sets the laser output current to 20 mA and then turns
the output on.
LASOUT
:
LASer:OUTput?
The LASer:OUTput? query returns the status of the laser output switch.
:
- where the response is 0 or 1
NotesAlthough the status of the switch is on, the output may not have reached the setpoint
value.
Examples"Las:OUT?" -resp on se : 0, mea ns tha t the output switch is disabled, devices may be
safely disconnected or connected at the LASER output terminals.
"LAS:OUT?" -response: 1, means that the LASER output switch is enabled, LASER
output is present.
04_06LDX-3232 73
LAS
:
OUT
?
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
<white
space>
FRONT PANEL
REMOTE
<nrf
value>
FRONT PANEL
REMOTE
LASer:RANge
The LASer:RANge command selects the laser's drive current output range.
:
ParametersAn <nrf value> that represents the laser current output range.
2 = 2000 mA, and 4= 4000 mA range
NotesThis range setting effects the Laser Drive Current output current range only. The Laser
Drive Current output should be “off” when this command is issued. If the Laser Diode
Current output is “on” when this command is issued, the LDX-3232 will generate error
E515, and the range will not be changed.
Examples"LAS:RAN 4" -action: sets the laser output drive current range to 4000 mA.
"Laser:range 2" -action: sets the laser output drive current range to 2000 mA.
LASRAN
:
LASer:RANge?
The LASer:RANge? query returns the value of the Laser Drive Current range.
:
- where the response is in integer; 2 means 2000 mA range and 4 means 4000 mA
range.
NotesThe resolution and accuracy of the laser limit current are dependent on the Laser Drive
Current output range.
Examples"LAS:RAN?" -response: 2, means that the Laser Drive Current range is 2000mA.
"LAS:range?" -response: 4, means that the Laser Drive Current range is 4000mA.
LAS
:
RAN
?
74 LDX-3232
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:SET:LDI?
The LASer:SET:LDI? query returns the constant I value that is used for both output
ranges and both bandwidths.
:
- where the response is the constant I setpoint value, in mA.
NotesIn local operation, the constant I setpoint is read by selecting I or I
pressing and holding down the SELECT switch in the DISPLAY area of the front panel.
When the mA indicator is blinking, the display will show the LDI setpoint.
Examples"LAS:SET:LDI?" -response: 50.0, means the laser output current setp oi n t val u e i s
50.0 mA.
"Laser:set:ldi?" -response: 120.0 means the laser output current setpoint value is
120.0 mA.
LAS
LDI
:
SET
I
FRONT PANEL
REMOTE
:
?
mode, then
HBW
LASer:SET:MDI?
The LASer:SET:MDI? query returns the laser monitor photodiode current setpoint value
(when CALPD=0), in µA.
:
- where the response is the constand I
NotesIn local operation, the MDI setpoint is read by selecting P mode, then pressing and
holding down the SELECT switch in the DISPLAY area of the front panel. When the µA
indicator is blinking, the display will show the MDI setpoint.
The monitor photodiode current is directly proportional to the laser optical output power.
Therefore, the MDI (I
Examples"las:set:mdi?" -response: 30.0, means the laser monitor photodiode current is setpoint is
30 µA, for use in constant P mode with CALPD = 0.
LAS
MDI
) setpoint may be used to control optical output of the laser.
PD
:
SET
IPD
setpoint value, in µA.
PD
FRONT PANEL
REMOTE
:
?
04_06LDX-3232 75
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
"LAS:Set:MDI?" -response: 100.0 means the laser monitor photodiode current is
setpoint is 100 µA, for use in constant P mode with CALMD=0.
LASer:SET:MDP?
The LASer:SET:MDP? query returns the laser monitor photodiode power setpoint value
(when CALMD [CAL PD] is not zero), in mW.
:
- where the response is the constand P
NotesIn local operation, the MDP setpoint is read by selecting P mode, then pressing and
holding down the SELECT switch in the DISPLAY area of the front panel. When the mW
indicator is blinking, the display will show the MDP setpoint. This setpoint is used in
constand MDP mode only.
Examples"LAS:Set:MDP?" -response: 10.0, means the laser monitor photodiode feedback
setpoint is 10.0 mW (CALPD>0).
"Laser:Set:MDP?" -response: 25.0, means the laser monitor photodiode feedback
setpoint is 25.0 mW (CALPD>0).
LAS
MDP
:
setpoint, in mW.
PD
SET
P
FRONT PANEL
REMOTE
:
?
76 LDX-3232
LASer:STEP
The LASer:STEP command is used to increment or decrement the selected laser control
mode setpoint by the given amount, when used with the LASer:INC or LASer:DEC
command.
:
ParametersAn integer value of the step amount, in the range 1 to 9999.
NotesThe step of one corresponds to the smallest display resolution for the mode. For
example, a step of one means 0.1 mA, 0.1 mW, or 1 µA (if CALPD=0).
Examples"Las:Mode:i; Las:ldi 20; Las:Step 100; Las:Inc; Las:set:ldi?" -action: sets the step to
10.0 mA, so the Las:set:ldi? query will return a value of 21.0 mA.
"LAS:STEP 100" -action: sets the step size to 100; could mean 10.0 mA , 10.0 mW, or
100 µA.
LAS
:
STEP
FRONT PANEL
REMOTE
<white
space>
<nrf
value>
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
LASer:STEP?
The LASer:STEP? query is used to read back the LASer STEP value. This value is used
to increment or decrement the selected laser control mode setpoint by the given amount
when used with the LASer:INC or LASer:DEC command.
:
- where the response is an integer value.
Examples"Las:Mode:LDI; Las:Step?" -response: 1 means the step size is 0.1 mA, since Const I
mode is in effect.
"LAS:MODE:MDP; LAS:CALMD 1; LAS:STEP?" -response: 10 means the step size is
1.0 mW, since Const P mode is in effect.
LAS
:
STEP
LASer:TOLerance
The LASer:TOLerance command allows the programmer to determine the LASER
current tolerance and time window for it to occur, in order that the operation complete flag
be set after a "LASer:OUTput 1" command is issued or the LASER setpoint is changed.
::
LAS
TOL
<white
space>
<nrf
value>
FRONT PANEL
REMOTE
?
FRONT PANEL
REMOTE
,
<nrf
value>
ParametersTwo <nrf values>; the first represents the LASER current tolerance in mA with a range of
0.1 to 100.0 mA; and the second represents the time window, in seconds, with a range of
0.001 to 50.000 seconds.
NotesThe LDX-3232 defaults to a tolerance of 1.0 mA for three seconds, unless changed by
the LASer:TOLerance command.
If the LDX-3232 is operated in P mode, the current tolerance parameter is not used.
Instead a fixed value of 50 µA is used for the I
PPD power, and only the time window parameter may be adjusted.
Note: If the tolerance is set too tight it may never be achieved. This is due to the calibration
of the setpoint and measurement values.
Examples"Las:Tol 0.5,10" -action: the LDX-3232's LASER current source will be in tolerance when
the LASER current is within 0.5 mA for 10.000 seconds.
04_06LDX-3232 77
current. A fixed value of 1 mW is used for
PD
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
LASer:TOLerance?
The LASer:TOLerance? query allows the programmer to determine how the LASER
current tolerance is set.
:
- where the response consists of two values, the first is the current tolerance in mA and
the second is the time window, in seconds.
NotesThe tolerance of the LDX-3232 LASER current may be used to delay programming after
an "LASer:OUTput 1" command is issued or the setpoint is changed.
A change of the output into or out of tolerance flag sets a flag in the LASER status event
register and so entering or exiting LASER current tolerance may be used to generate
service requests.
Examples"Las:Tol?" -response: "0.2,5.0", means the LDX-3232 has a LASER current tolerance
setting of 0.2 mA with a time window of 5.000 seconds.
"LASER:TOL?" -response: "1.0,20.0", means the LDX-3232 has a LASER current
tolerance setting of 1.0 mA with a time window of 20.000 seconds.
LAS
:
TOL
MESsage
The MESsage command allows the user to enter an ASCII string of up to 16 non-NULL
characters. This command may be useful for storing messages that relate to a test or
configuration.
FRONT PANEL
REMOTE
?
FRONT PANEL
REMOTE
78 LDX-3232
:
ParametersAn ASCII string that is 1 - 16 bytes in length.
NotesThe message may contain any ASCII character, but will be terminated when a NULL
terminator character is received. If the message has less than 16 bytes, the software will
fill the remaining message space with the space character. After 16 bytes have been
entered, the software will null-terminate the string.
ExamplesMESSAGE "This is a test." -action: The string, "This is a test. " will be stored in
non-volatile memory.
Mes "Test 3" -action: The string, "Test 3 " will be stored in non-volatile memory.
MES
<white
space>
<ascii
string>
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
MESsage?
The MESsage? query returns the previously stored message. This message will always
be 16 bytes long and enclosed in quotes. The message is entered via the MESsage
command.
:
- where the response is a 16-byte long character string.
NotesThe response data will be a 16-byte long string. If there is no previously stored message,
the response will be " ", all spaces.
Examples"MES?" -response: "Test 3 ", means the previously stored message was "Test 3".
"Message?" -response: "This is a test. ", means the previously stored message was
"This is a test."
MES
?
RADix
The RADix command allows the programmer to select the radix type for status,
condition, and event query response data. Decimal, binary, hexadecimal, and octal are
allowed.
FRONT PANEL
REMOTE
FRONT PANEL
REMOTE
:
DEC
ParametersAs shown above, character program data is expected.
NotesDECimal is the default type. Only the first three letters of the words decimal,
hexadecimal, binary, or octal are required.
When the RADIX is selected, all status, condition, and event queries will return values in
the new radix.
In the cases where the radix is not DECimal, the flexible numeric type <nrf value> (as
shown in the Command Reference diagrams) will be replaced by HEX, BIN, or OCT
representation.
All of the above radixes may be used to enter program data at any time, without the need
for issuing the RADix command. The proper prefix must also be used with Hex (#H),
binary (#B), or octal (#Q).
This command may be useful for setting up status reporting blocks. The bit-wise status
representation may be more easily read in BIN, HEX, or OCT.
Examples"RAD dec" -action: the decimal radix is selected.
RAD
HEX
<white
space>
BIN
OCT
04_06LDX-3232 79
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
"rad hex; *ESR?" -action: the hexadecimal radix is selected; -response: #H80, means
power-on was detected.
RADix?
The RADix? query allows the programmer to determine that radix type for status,
condition, and event query response data is currently selected. Decimal, binary, octal,
and hexadecimal are allowed.
:
- where the character response data of DEC means decimal, BIN means binary, HEX
means hexadecimal, and OCT means octal.
NotesDEC is the default type. The LDX-3232 defaults to this radix at power-up.
The RADix command is used to select the desired radix. Once it is changed, the new
radix will remain in effect until the power is shut off or a new RADix command is issued.
Examples"RAD?" -response: Dec, means the selected radix is decimal.
"rad?" -response: Hex, means the selected radix is hexadecimal.
"RADIX?" -response: Oct, means the selected radix is octal.
RAD
?
SECURE
The SECURE command allows the service technician access to the protected user data
command, *PUD, that is a common command. This data is normally changed only at the
factor, and therefore the SECURE command is not needed by the user.
FRONT PANEL
REMOTE
FRONT PANEL
REMOTE
80 LDX-3232
TERM
The TERM command allows the programmer to change the default
(GPIB/IEEE488 standard) terminator to include the carriage return.
:
ParametersAn <nrf value>, 0=FALSE, non-z ero =TRUE
NotesAn altered terminator will be in the form <CR><NL><^END>. This technically takes the
LDX-3232 out of GPIB/IEEE488 specification, but may be done for convenience when
using non standard GPIB controllers. This termination will be sent with all output until the
"TERM 0" command is sent or the LDX-3232 is powered off.
Examples"Term 1" -temporarily sets <CR><NL><^END> as the output terminator.
"Term 0" -sets the GPIB/IEEE488 standard <NL><^END>terminator. This is the
LDX-3232 default value.
TERM
<white
space>
<nrf
value>
FRONT PANEL
REMOTE
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
CHAPTER 4
TERM?
The TERM? query allows the programmer to determine whether the default
(IEEE488.2 standard) terminator has been altered to include a carriage return.
:
- where the response is 0 or 1.
NotesAn altered terminator will be in the form <CR><NL><^END>. This termination will be
sent with all output until the "TERM 0" command is sent, or the LDX-3232 is powered off.
Examples"Term?" -response: 1, means that the <CR><NL> <^END> terminator is temporarily
being used.
"Term?" -response: 0, means that the <NL><^END> terminator (GPIB/IEEE488
standard) is being used.
?TERM
TIME?
The TIME? query allows the programmer to determine how much time has passed since
the LDX-3232 was last powered up.
FRONT PANEL
REMOTE
FRONT PANEL
REMOTE
:
- where the response is character data in the form: hours:minutes:seconds.
NotesThe TIME clock is independent of the TIMER clock.
Examples"Time?" -response: 1:02.36, means that one minute and 2.36 seconds have passed
since the LDX-3232 was powered up.
"TIME?" -response: 32:00:76, means that 32 minutes and 0.76 seconds have passed
since the LDX-3232 was powered up.
?TIME
04_06LDX-3232 81
CHAPTER 4
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
TIMER?
The TIMER? query allows the programmer to determine how much time has passed
since the last TIMER? query was issued.
:
- where the response is character data that represents hours:minutes:seconds.
NotesEach time the TIMER? query is issued, the timer is reset to 0 and the elapsed time since
the last TIMER? query is returned.
The timer counter is initially set at power-up, the same as the TIME? counter. So the first
time the TIMER? is issued its response will be the same as if a TIME? query's response.
Examples"Timer?" -response: 00:02:00.31, means the LDX-3232 has been on for two minutes and
0.31 seconds since the last TIMER? query was issued.
"TIMER?" -response: 00:00:12.03, means the LDX-3232 has been on for 12.03 seconds
since the last TIMER? query was issued.
?TIMER
FRONT PANEL
REMOTE
82 LDX-3232
FUNCTIONS AND FEATURES
Saving and Recalling from the Front Panel
CHAPTER 5
CHAPTER 5
FUNCTIONSAND FEATURES
This chapter introduces you to the LDX-3232 High Compliance Current Source
functions and operating features such as event triggering, analog modulation of
laser injection current, and Save and Recall.
Saving and Recalling from the Front Panel
For applications where you might need to alternate between pa rticu lar inst rumen t
configurations, the LDX-3232 Current Source offers a unique SAVE and RECALL
feature. The SA VE feature allows you to easily store all front panel settings for any
given instrument configuration. These settings, that are stored in one of ten
memory bins, can be retrieved at any time with the RECALL function. This saves
setup time, and it reduces the chance of setup error for tests that are periodically
repeated. For example, using the configuration described in Chapter 2 where the
Current Source is operating in Constant Current mode, the instrument parame ters
saved by using the SAVE function would be Constant Current Mode, 2000 mA
range, a limit setpoint of 1750 mA, a laser current setpoint of 1500 mA, and
Display Mode I. Once this setup is saved, the LDX-3232 Current Source may be
configured for another unique setup. You can recall the original setup for the first
experiment any time using the RECALL function.
To enter the SAVE/RECALL mode, press the SELECT switch in the PARAMETER
section until the SAVE indicator LED becomes lit and the unit displays the SAVE
"bin" on the display. If another bin number is desired, press and hold the SET
switch and rotate the adjust knob until the desired bin number is displayed.
Releasing the SET switch completes the SA VE operation and all of the instrument
parameters are saved to the indicated bin number as they appear on the front
panel.
04_06LDX-3232 83
CHAPTER 5
FUNCTIONS AND FEATURES
Saving and Recalling Under Remote Operation
To recall a previously saved instrument setup, press the SELECT switch until the
RECALL indicator led becomes lit. The instrument will display a RECALL "bin"
number on the display. If you desire to change the RECALL bin, press and hold
the SET switch and rotate the adjust knob until the desired bin number is
displayed. Releasing the SET switch completes the RECALL operation. The
instrument is restarted and the front panel parameters are reconfigured to th e new
parameters saved under that bin number.
The factory default values for front panel parame ters may be restored by recalling
"bin" zero (0). See Table 2.1 for the default settings.
Saving and Recalling Under Remote Operation
For saving under remote operation, use the command *SAVx, where x is the
desired "bin" number (1-10). For a recall under remote operation, use the
command
*RCL x, where x is the desired "bin" number (0 - 10).
Using the Laser Compliance Voltage Adjustment
For applications where you need to have critical protection of the laser, we
recommend the following. Using a 1
LASER voltage limit as described in Chapter 2.
Then with the LASER output on, slowly increase the LASER current and watch
the LASER voltage measurement. Press the (DISPLAY) SELECT push button to
see the voltage. At about 0.25 V below the LASER voltage limit value, the
VOLTAGE LIMIT warning indicator will become lit on the display. Continue to
adjust the current upwards until the voltage limit is reached and the output is shut
off. Record the voltage at that the LASER output shuts off. It should be within the
specification of the LASER voltage limit accuracy. However, there is typically
some offset error. For example, if the LASER voltage limit is set to 5.0 V, the
actual voltage where it shuts off may be 4.9 V. This error term of 0.1 V should be
considered when setting the LASER compliance voltage limit.
The closer the LASER voltage limit shut off point is to the operating voltage of
your laser, the faster the circuit will work in the event of an open circuit. Some
experimentation may be necessary for optimum results.
Ω resistor or other non-critical load, set the
84 LDX-3232
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.