Lighthouse Remote, Remote 2014P, Remote 3014P, Remote 5014P Operating Manual

Lighthouse Worldwide Solutions
REMOTE 2014P, 3014P, 5014P Airborne Particle Counter Gen C
Operating Manual
Copyright © 2011 by Lighthouse Worldwide Solutions. All rights reserved. No part of this document may be reproduced by any means except as permitted in writing by Lighthouse Worldwide Solutions.
The information contained herein constitutes valuable trade secrets of Lighthouse Worldwide Solutions. You are not permitted to disclose or allow to be disclosed such information except as permitted in writing by Lighthouse Worldwide Solutions.
The information contained herein is subject to change without notice. Lighthouse Worldwide Solutions is not responsible for any damages arising out of your use of the LMS program.
REMOTE 2014P™, REMOTE 3014P™, REMOTE 5014P™ and LMS™ are trademarks of Lighthouse Worldwide Solutions.
Microsoft®, Microsoft Windows™, and Excel™ are trademarks of Microsoft Corporation.
LWS PN 248083408-1 Rev 1
EU DECLARATION OF CONFORMITY
Manufacturer’s Name: Lighthouse Worldwide Solutions, Inc.
Manufacturer’s Address: Lighthouse Worldwide Solutions, Inc.
1221 Disk Drive Medford, OR 97501 USA
Declares that the product:
Product Name: Remote Airborne Particle Counter Model Number(s):
Conforms to the following Product Specifications:
REMOTE 2014P, 3014P, 5014P
SAFETY
LASER SAFETY
EMC
UL 61010A-1 - UL Standard for Safety Electrical Equipment for Laboratory Use; Part 1: General Requirements Replaces UL 3101-1
Supplementary information
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC amended by Directive 93/68/EEC and the EMC Directive 89/336/EEC amended by Directive 93/68/EEC and carries the CE marking accordingly.
EN61010-1:2001 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use Part 1: General Requirements IEC 61010-1:2000
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1: General Requirements
Guidance on Laser Products: Conforms to FDA 21 CFR Chapter 1 Subchapter 1
EN61326 Electrical Equipment for Measurement, Control and
Laboratory Use EMC Requirements Part 1: General Requirements Includes Amendment A1:1998; IEC 61326:1997 + A1:1998
Fremont, CA, May 15, 2007 William L. Shade – V.P. Engineering
00
Table of Contents
About this Manual
Text Conventions ........................................................................................................................... i
Additional Help .............................................................................................................................. i
Chapter 1 General Safety
Laser Safety Information ............................................................................................... 1-1
Chapter 2 Introduction
Overview ........................................................................................................................ 2-1
Description ..................................................................................................................... 2-1
Accessories .................................................................................................................... 2-2
Annual Calibration ......................................................................................................... 2-2
REMOTE 2014P Specifications .................................................................................... 2-3
REMOTE 3014P Specifications .................................................................................... 2-4
REMOTE 5014P Specifications .................................................................................... 2-5
Chapter 3 Unpacking, Inspecting and Installing
Initial Inspection ............................................................................................................ 3-1
Unpacking ...................................................................................................................... 3-1
Shipping Instructions ..................................................................................................... 3-2
Understanding the LEDs .................................................................................... 3-3
Connections ................................................................................................................... 3-4
Communication Ports ........................................................................................ 3-5
4-20mA Analog Inputs ...................................................................................... 3-6
Analog Sensors ...................................................................................... 3-6
DIP Switches ...................................................................................................... 3-7
Power ................................................................................................................. 3-7
Power Switch ..................................................................................................... 3-7
Remote Display Connector ................................................................................ 3-8
Peripheral Interface Connector .......................................................................... 3-8
USB Communications ....................................................................................... 3-8
Using the USB Port to connect to a PC ................................................. 3-9
Ethernet Configuration ................................................................................................ 3-10
Definitions of Terms Used ........................................................................................... 3-11
Preparing for Network Installation .............................................................................. 3-12
248083408-1 Rev 1 t-i
Lighthouse REMOTE 2014P, 3014P 5014P Operating Manual
Equipment Required: ....................................................................................... 3-12
Software Required: .......................................................................................... 3-12
Additional Requirements: ................................................................................ 3-12
Configure Device ......................................................................................................... 3-13
Ethernet Port Configuration ............................................................................. 3-13
Straight-through Cat5 Setup: ............................................................... 3-13
Cross-over Cat5 Setup: ........................................................................ 3-13
Program the Interface ................................................................................................... 3-14
Windows Telnet Programming: ....................................................................... 3-14
Connect REMOTE P to Ethernet LAN ........................................................................ 3-19
Connect Ethernet Cable to Instrument ............................................................. 3-19
Data .............................................................................................................................. 3-19
Chapter 4 Programming
General Information ....................................................................................................... 4-1
DIP Switches .................................................................................................................. 4-1
DIP Switch Settings ....................................................................................................... 4-2
GENERAL DEFINITIONS ............................................................................... 4-2
Communications Mode .......................................................................... 4-2
REMOTE Display Mode ....................................................................... 4-2
Addressing ......................................................................................................... 4-3
Communicating with the Instrument ............................................................................. 4-4
RS-485 Port ........................................................................................................ 4-5
RS-485 Communications ................................................................................... 4-5
RS-232 Communications ................................................................................... 4-6
Configuring with REMOTE Display ............................................................................. 4-6
Configuring with the MODBUS Protocol ..................................................................... 4-6
Setting the Real Time Clock .............................................................................. 4-7
Changing the Default Instrument Parameters .................................................... 4-7
Running the Instrument ................................................................................................. 4-9
MANUAL Counting Mode .............................................................................. 4-10
AUTOMATIC Counting Mode ....................................................................... 4-10
Chapter 5 Maintenance Procedures
Safety ............................................................................................................................. 5-1
Calibration ..................................................................................................................... 5-1
Cleaning ......................................................................................................................... 5-1
Purge Test ...................................................................................................................... 5-1
Appendix A MODBUS Register Map v1.48
COMM Settings ............................................................................................................ A-1
Supported MODBUS Commands ................................................................................. A-1
t-ii 248083408-1 Rev 1
Register Map ................................................................................................................. A-2
Sensor Settings Registers .................................................................................. A-2
Device Status .................................................................................................... A-7
Command Register ....................................................................................................... A-7
Data and Alarm Registers ............................................................................................. A-8
Data and Alarm Enable Registers ..................................................................... A-8
Enable Alarming for a Channel ...................................................................... A-10
Threshold Setup Registers .............................................................................. A-11
Setting the Alarm Threshold Value ................................................................ A-12
Data Registers ............................................................................................................. A-12
Device Status Word (30007 - 30008) ............................................................. A-14
Valid Data in Channels (30073 - 30076) ........................................................ A-15
Data Type Registers ........................................................................................ A-16
Data Units Registers ....................................................................................... A-17
Appendix B Zero Count Test
How to Run the Test ...................................................................................................... B-1
Table of Contents
Appendix C Limited Warranty
Limitation Of Warranties: .................................................................................. C-1
Warranty Of Repairs After Initial Two (2) Year Warranty: .............................. C-1
Index
248083408-1 Rev 1 t-iii
Lighthouse REMOTE 2014P, 3014P 5014P Operating Manual
t-iv 248083408-1 Rev 1
00
About this Manual
This manual describes the detailed operation and use of the Lighthouse REMOTE 2014P, 3014P and 5014P Airborne Particle Counters.
Text Conventions
Note: A note appears in
the sidebar to give extra information regarding a feature or suggestion
WARNING: A
warning appears in a paragraph like this and warns that doing something incorrectly could result in personal injury, damage to the instrument or loss and/or improper storage of data.
The following typefaces have the following meanings:
italics Represents information not to be typed
or interpreted literally. For example, file represents a file name. Manual titles are also displayed in italics.
boldface Introduces or emphasizes a term.
Courier font Indicates command syntax or text
displayed by the diagnostic terminal.
Bold Courier Indicates commands and information that
you type. You can use uppercase or lowercase letters; in this manual, commands are shown in uppercase.
Helvetica Italics Indicates a comment on a command or
text output.
Additional Help
For more information about Lighthouse REMOTE 2014P, 3014P or 5014P Airborne Particle Counter, contact Lighthouse Worldwide Solutions:
(800) 945-5905 Sales and Service (541) 770-5905 Outside of USA
techsupport@golighthouse.com www.golighthouse.com
248083408-1 Rev 1 i
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
ii 248083408-1 Rev 1
00
1 General Safety
Safety Considerations
Laser Safety Information
Warnings and cautions are used throughout this manual. Familiarize yourself with the meaning of a warning before operating the particle counter. All warnings will appear in the left margin of the page next to the subject or step to which it applies. Take extreme care when doing any procedures preceded by or containing a warning.
There are several classifications of Warnings defined as follows:
Laser - pertaining to exposure to visible or invisible laser radiation
Electrostatic - pertaining to electrostatic discharge
This product contains a laser-based sensor that is a Class 1 product (as defined by 21 CFR, Subchapter J of the Health and Safety Act of 1968) when used under normal operation and maintenance. Service procedures on the sensor can result in exposure to invisible radiation. Service should be performed only by factory-authorized personnel.
The particle counter has been evaluated and tested in accordance with EN 610109-1:1993, “Safety Requirements For Electrical Equipment for Measurement, Control, and Laboratory Use” and IEC 825-1:1993, “Safety of Laser Products”.
WARNING: The use
of controls, adjustments, or performance of procedures other than those specified within this manual may result in exposure to invisible (infrared) radiation that can quickly cause blindness.
For further technical assistance, contact our Technical Support Team at (800) 945-5905.
248083408-1 Rev 1 1-1
Figure 1-1 Warning label on unit
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
1-2 248083408-1 Rev 1
00
2 Introduction
Overview This operating manual introduces you to the Lighthouse REMOTE
2014P, 3014P or 5014P family of Airborne Particle Counters. Included in this manual are instructions for inspecting, using and maintaining the instrument.
Description All members of the ’P’ family of instruments have up to six particle
channel sizes and a flow rate of 0.1CFM. The model name denotes the starting particle size (sensitivity). The 2014P sensitivity is 0.2 microns, the 3014P sensitivity is 0.3 microns and the 5014P sensitivity is 0.5 microns.
Figure 2-1 REMOTE 3014P Airborne Particle Counter
The instrument uses a laser diode light source and laser beam shaping optics to illuminate a cross section of the air flow path with a laser beam. As particles move along the flow path, they cross the laser beam and scatter light. The light scattered is collected by an imaging optical system onto a photodiode. The photodiode converts this light into a current which is converted to a voltage and amplified.
248083408-1 Rev 1 2-1
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
The resulting pulse is measured for width and height. The width of the pulse is proportional to the time it takes the particle to cross the laser beam. The height or amplitude of the pulse is equal to the amount of light scattered by each particle and is proportional to its size.
These pulses are processed by additional electronics that analyze their height to determine the size of the corresponding particle. This allows counting of the particles and grouping based on their size.
The REMOTE 2014, 3014P and 5014P are effective in both ultra-clean areas (such as Class 1 or Class 10) and also in more traditional cleanzones rated as Class 100 or higher. They were designed for continuous 24/7 operation. Data is stored in a rotating buffer of 3000 records that can be downloaded real-time or manually as required.
Using an internal pump, the instrument can be installed where space is limited. The REMOTE 2014P, 3014P or 5014P integrate seamlessly into large facility monitoring/management systems and transfer up to 6 channels of simultaneous particle count data using RS-485/MODBUS output. Refer to Specifications in this chapter for additional instrument information.
The manufacturer recommends that your Lighthouse instrument be calibrated annually by a Certified Lighthouse Service Provider to ensure it continues to perform within specification.
Accessories You can order several accessories to tailor the instrument to your needs.
Please contact your Lighthouse Sales Engineer for detailed additional information.
Annual Calibration
The manufacturer recommends that your Lighthouse instrument be calibrated annually by a Certified Lighthouse Service Provider to ensure that it continues to perform within specification.
When Calibration comes due, the Service LED will illuminate and stay ON during sampling mode until the instrument has been calibrated.
2-2 248083408-1 Rev 1
REMOTE 2014P Specifications
Introduction
Size Range 0.2 - 2.0 Pm
4 Channel Thresholds Standard: 0.2, 0.3, 0.5, 1.0 μm
Standard: 0.2, 0.3, 0.5, 2.0 μm
Optional: 0.2, 0.25, 0.3, 0.5, 0.7, 1.0, 2.0 μm
6 Channel Thresholds Standard: 0.2, 0.3, 0.5, 0.7, 1.0, 2.0 μm
Optional: 0.2, 0.25, 0.3, 0.5, 0.7, 1.0, 2.0 μm
Flow Rate 0.1 CFM (2.83 LPM)
Counting Efficiency 50% (per ISO 21501-4)
LightSource Laser diode
Zero Count Level <1 count/5 minutes (per ISO 21501-4)
Vacuum Source Internal Vacuum Pump
Calibration NIST Traceable
Communication Modes RS-232 or RS-485 Modbus
Data Storage 3000 records, rotating buffer
Supporting Software Lighthouse Monitoring System, LMS
XChange, LMS Express, LMS ExpressRT
Environmental Sensors Optional: Temperature / Relative Humidity,
Air Velocity, Differential Pressure
Power Supply Input 100-240 VAC, 50-60Hz, 0.4A
Power Requirement +24V, 0.62A
Display Remote Display; color Touchscreen LCD
(optional)
Enclosure Stainless Steel
Dimensions 7.25"[l] x 3.75"(w) x 6"(h) [18.41 x 9.52 x
15.24 cm]
Weight 5.0 lbs (2.26 kg)
Operating Temp/RH 50° F to 104° F (10° C to 40° C) / 20% to 95%
non-condensing
Storage Temp/RH 14° F to 122° F (-10° C to 50° C) / Up to 98%
non-condensing
Table 2-1 REMOTE 2014P Specifications
248083408-1 Rev 1 2-3
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
REMOTE 3014P Specifications
Size Range 0.3 - 25.0 Pm
4 Channel Thresholds Standard: 0.3, 0.5, 1.0, 5.0 μm
Standard: 0.3, 0.5, 5.0, 10.0 μm
Optional: 0.3, 0.5, 0.7, 1.0, 2.0, 2.5, 3.0, 5.0,
10.0, 15.0, 20.0, 25.0 μm
6 Channel Thresholds Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm
Optional: 0.3, 0.5, 0.7, 1.0, 2.0, 2.5, 3.0, 5.0,
10.0, 15.0, 20.0, 25.0 μm
Flow Rate 0.1 CFM (2.83 LPM)
Counting Efficiency 50% (per ISO 21501-4)
Laser Source Laser diode
Zero Count Level <1 count/5 minutes (per ISO 21501-4)
Vacuum Source Internal Vacuum Pump
Calibration NIST Traceable
Communication Modes RS-232 or RS-485 Modbus
Data Storage 3000 records, rotating buffer
Supporting Software Lighthouse Monitoring System, LMS
XChange, LMS Express, LMS ExpressRT
Environmental Sensors Optional: Temperature / Relative Humidity,
Air Velocity, Differential Pressure
Power Supply Input 100-240 VAC, 50-60Hz, 0.4A
Power Requirement +24V, 0.62A
Display Remote Display; color Touchscreen LCD
(optional)
Enclosure Stainless Steel
Dimensions 7.25"[l] x 3.75"(w) x 6"(h) [18.41 x 9.52 x
15.24 cm]
Weight 5.0 lbs (2.26 kg)
Operating Temp/RH 50° F to 104° F (10° C to 40° C) / 20% to 95%
non-condensing
Storage Temp/RH 14° F to 122° F (-10° C to 50° C) / Up to 98%
non-condensing
Table 2-2 REMOTE 3014P Specifications
2-4 248083408-1 Rev 1
REMOTE 5014P Specifications
Introduction
Size Range 0.5 - 25.0 Pm
4 Channel Thresholds Standard: 0.5, 1.0, 5.0, 10.0 μm
Optional: 0.5, 0.7, 1.0, 2.0, 2.5, 3.0, 5.0, 10.0,
15.0, 20.0, 25.0 μm
6 Channel Thresholds Standard: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm
Optional: 0.5, 0.7, 1.0, 2.0, 2.5, 3.0, 5.0, 10.0,
15.0, 20.0, 25.0 μm
Flow Rate 0.1 CFM (2.83 LPM)
Counting Efficiency 50% (per ISO 21501-4)
Laser Source Laser diode
Zero Count Level <1 count/5 minutes (per ISO 21501-4)
Vacuum Source Internal Vacuum Pump
Calibration NIST Traceable
Communication Modes RS-232 or RS-485 Modbus
Data Storage 3000 records, rotating buffer
Supporting Software Lighthouse Monitoring System, LMS
XChange, LMS Express, LMS ExpressRT
Environmental Sensors Optional: Temperature / Relative Humidity,
Air Velocity, Differential Pressure
Power Supply Input 100-240 VAC, 50-60Hz, 0.4A
Power Requirement +24V, 0.62A
Display Remote Display; color Touchscreen LCD
(optional)
Enclosure Stainless Steel
Dimensions 7.25"[l] x 3.75"(w) x 6"(h) [18.41 x 9.52 x
15.24 cm]
Weight 5.0 lbs (2.26 kg)
Operating Temp/RH 50° F to 104° F (10° C to 40° C) / 20% to 95%
non-condensing
Storage Temp/RH 14° F to 122° F (-10° C to 50° C) / Up to 98%
non-condensing
Table 2-3 REMOTE 5014P Specifications
248083408-1 Rev 1 2-5
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
2-6 248083408-1 Rev 1
Unpacking, Inspecting and Installing
3 Unpacking, Inspecting
and Installing
Initial Inspection
The instrument is thoroughly inspected and tested at the factory and is ready for use upon receipt.
Unpacking It is presumed that when the shipment was received, the following took
place:
1. The shipping container was inspected for damage;
2. If the container was damaged, the shipper was notified
immediately.
3. The instrument was carefully inspected for broken parts,
scratches, dents and other damage before use, even if the container appeared to be undamaged, and
4. Any damages were reported to Lighthouse Technical Support at
+1-800-945-5905 (USA) or +1-541-770-5905 (Outside of USA) before proceeding.
Verify the contents of the package against the shipping list. If anything appears to be missing, please contact your sales representative at Lighthouse Worldwide Solutions immediately at 800 945 5905 or techsupport@golighthouse.com.
To maintain your warranty, keep the undamaged shipping container and all packing material for reshipment of the instrument for annual calibration. Order replacement containers and packing materials only from Lighthouse, directly, or from a Lighthouse-authorized distributor.
248083408-1 Rev 1 3-1
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Shipping Instructions
WARNING:
instrument is damaged during a return shipment due to inadequate user packing, the warranty may be voided and may result in additional repairs being billed to the customer.
If the
Should it become necessary to return the unit to the factory for any reason, contact Lighthouse Customer Service or visit our website, www.golighthouse.com/rma, and obtain a Return Merchandise Authorization (RMA) number. Reference this number on all shipping documentation and purchase orders. After receipt of the RMA number, follow the shipping instructions below:
1. Use the original container, nozzle caps and packing materials when­ever possible. Remove any instrument battery and package it to ship separately - refer to www.golighthouse.com/rma for detailed instructions. Remove attachments, such as TRH or Isokinetic probes, and package to prevent physical and ESD damage (TRH).
2. If the original container and packing materials are not available, wrap the unit in “bubble pack”, surround with 4 inch shock­absorbent material and place in a double-wall carton - the instrument should not rattle around when the carton is vigorously shaken. If the instrument is damaged during shipment due to inadequate user packing, the warranty may be voided and may result in additional repairs being billed to customer. You may contact Lighthouse to purchase a replacement shipping container and nozzle caps.
3. Seal container or carton securely. Mark “FRAGILE” and write the Return Merchandise Authorization (RMA) number on any unmarked corner.
4. Return the instrument to the address provided by your Lighthouse representative or the RMA website.
3-2 248083408-1 Rev 1
Operation Understanding the LEDs
The front-panel LEDs have special meanings when illuminated. The figure below shows location of the LEDs and gives a brief description of their meaning.
Unpacking, Inspecting and Installing
Figure 3-1 Front Panel LEDs
The green POWER LED turns on when the instrument is powered on.
The green FLOW LED turns on when the flow is correct.
The green FLOW LED will turn off if flow is insufficient.
The orange SERVICE LED will stay on steady if Laser power is low, sensor optics are dirty or the view chamber contains foreign objects.
The blue SAMPLING LED turns on when the instrument is counting.
Features The REMOTE 2014P, 3014P or 5014P instrument has the following
features:
1. Optional Remote Display connection (color Touchscreen interface)
2. Four optional 4-20mA output connections
3. DIP switch control for device addressing
248083408-1 Rev 1 3-3
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
4. MODBUS protocol (see “MODBUS Register Map v1.48” on page A-1).
5. Ethernet and RS-485 communications
6. 3000 record storage in a rotating buffer
7. Internal pump with auto-adjusting flow control
Connections The top of the instrument has one connection, the sampling inlet.
Figure 3-2 Connection on Top
The REMOTE 2014P, 3014P or 5014P can be used with a direct-mount
0.1 CFM isokinetic probe; the probe is attached via 1/8” ID tubing to
the 1/8” inlet barb.
1
6
2
7
3
8
4
5
Figure 3-3 Connections on the Back of Instrument
3-4 248083408-1 Rev 1
9
Unpacking, Inspecting and Installing
1. DIP Switches (under plate): for changing the addressing of the instrument and for selecting to use the Remote Display option.
2. Power Switch: for turning the instrument ON and OFF.
3. Power Connector: +24V power supply.
4. Peripheral Interface Connector: For attaching an external peripheral device such as an alarm buzzer, alarm light or a remote start/stop controller
5. REMOTE Display Connector: for attaching the Remote LCD Color Display with touchscreen interface.
6. RJ-45 Ethernet Connector: For MODBUS communications over an Ethernet LAN.
7. RS-232/485 IN Port : Communication ports.
8. Analog Input Ports: Four input ports for connecting 4-20mA analog sensors.
9. USB Port (Type B): For point-to-point communications via a PC.
Communication Ports
The RS-485 port is a dual purpose connection. It is a hybrid of the MODBUS over Serial standard incorporating both short distance single instrument RS-232 and long distance multi-point RS-485.
The RS-232 protocol is provided for quick simple cable connections from a single unit to a standard PC COM port.
The RS-485 MODBUS protocol is provided for industrial applications with multiple devices on the same bus. The RS-485 IN connector allows for easy connections using standard CAT 5 cable commonly used for office LANs. Attaching the instrument to an RS485 network requires special attention to topology (wiring scheme of network) before and after the design phase to make sure the REMOTE and other instruments perform as expected.
248083408-1 Rev 1 3-5
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
WARNING:
Lighthouse Technical Support for the correct instrument configuration BEFORE attempting to use RS485 COM mode with a PC. Failure to heed this warning can result in damage to PC, instrument or both.
Contact
For detailed description of the MODBUS registers and commands, please see “MODBUS Register Map v1.48” on page A-1.
4-20mA Analog Inputs
This REMOTE instrument has two 4-20mA analog inputs that can be fitted with up to four 4-20mA environmental sensor. Lighthouse provides optional environmental sensors: Temperature/Relative Humidity (T/RH), Air Velocity (AV), Differential Pressure (DP).
Analog Sensors
The Sensors are connected to the instrument with RJ-12 cables. The pinout of the cables is shown below:
Table 3-1 Analog Connector Pinout
Pin Number Function
15VDC
2 Analog #1 In
Note:
analog configuration will change historical environmental values in the data buffer.
Changes to the
3 24VDC (VCC)
4 GND (tied to Pin 6)
5 Analog #2 In
6 GND
Lighthouse sensors come complete with adapters and prewired cables.
The Lighthouse Temperature/Relative Humidity sensor has both functions wired to the same connector. Plug the connector into the port labeled “1”.
Each port can be used for one dual-channel or two single-channel 4­20mA sensors (when wired correctly), two devices total per port. When each port is connected to two single-channel analog sensors, a total of four analog sensors can be connected to the REMOTE. Otherwise, only one single-channel or one dual-channel device may be used at a time on each port. Please contact Lighthouse for additional information about converters and wiring.
3-6 248083408-1 Rev 1
Unpacking, Inspecting and Installing
WARNING: Wiring
the pins in reverse order may harm the instrument and/or the sensor and may void their respective warranties.
If you are wiring a cable to use with a sensor from another manufacturer, the connectors on telephone cables are mirror images of each other. For that reason, we have not included color coding in the table. The figure below shows the pin numbering on the back of the REMOTE - your cable must be wired accordingly.
Pin 1
Pin 6
Figure 3-4 Analog Sensor Ports, Showing Pin Numbering
WARNING: If the
REMOTE will be used with the Lighthouse MiniManifold, do NOT use the REMOTE 24VDC power supply - power will be supplied through the signal cable from the MiniManifold; otherwise, damage to either unit may occur.
Note that the ports are oriented with the locking tabs at the side.
DIP Switches
DIP switches 1-6 are used for addressing the instrument for RS-485 and setting the Communications Mode.
DIP switch 7 enables or disables the Remote LCD Display option. See Chapter 4 for more details.
Power
This REMOTE instrument uses an external power supply rated at +24VDC at 0.62A. The connector incorporates a threaded ring to lock it into position, preventing accidental disconnects.
Power Switch
The ON/OFF switch controls power to the instrument.
248083408-1 Rev 1 3-7
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Remote Display Connector
An optional REMOTE Display touchscreen can be connected to the 2014P, 3014P or 5014P instrument. This display is a color touchscreen interface. A longer cable can be attached to the Display so that the instrument can be running in a clean area while the Display is mounted on a wall bracket in a different area.
Please refer to the REMOTE Display Operating Manual for details.
Peripheral Interface Connector
An external peripheral device such as a light stick, buzzer or remote control device can be connected to the back or the REMOTE P via the Peripheral Interface Connector (PIC) .
The connector has the following pin-outs.
Table 3-2 PIC Connector Pin-Outs
Pin Connection
1 24 VDC
2GND
3Alarm+
4Alarm-
5 Start/Stop
6 Not Connected
7 Not Connected
8GND
USB Communications
The REMOTE P provides point-to-point communications via the USB Port.
If it hasn’t been done, yet, install the USB driver provided on the product CD. Once it has been installed and the instrument is connected, the driver will be used to establish communications using a virtual COM port on the PC.
3-8 248083408-1 Rev 1
Unpacking, Inspecting and Installing
Using the USB Port to connect to a PC
Note: Make sure the
USB driver has been installed on the PC before connecting and applying power to the instrument.
To connect the instrument to a computer using the USB port, make sure that the driver has been installed first. Connecting the instrument to the PC and applying power before the driver has been installed will be problematic. Refer to the following:
1. Remove power from the instrument.
2. Connect Side B of a standard “A to B” USB cable (See Figure 3-5) to the instrument (only Side B will fit).
Figure 3-5 “A to B” USB Cable
3. Connect Side A to any available USB port on the PC.
4. Apply power to the instrument.
5. Windows will recognize that a new hardware device has been attached and will finalize the installation as needed. This is a routine “registering” of components that should take only a moment. An additional COM port will be added to the PC.
248083408-1 Rev 1 3-9
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
6. Identify the computer port that has been added to the PC by using the “Properties” of My Computer and examining the COM ports. USB ports are generally added to the bottom of the list and named USB COM port ’x’. Refer to Figure 3-6 for screen examples (Windows XP).
Device Manager
Screen
Before port is added
After port is added
Ethernet Configuration
Figure 3-6 Example Device Manager Screens
Use LMS XChange or one of the LMS Express products to connect to the instrument and retrieve data from it. Use the port assignment determined above.
The REMOTE P comes pre-configured to use DHCP and receive an IP address automatically. If this is not a desired configuration, use this section to change these settings.
If you do not understand these terms or the possible impact of changing the REMOTE’s settings, do NOT proceed without contacting your facility’s IT personnel.
The following procedure requires pre­approval and planning by the IT department and a joint setup effort.
3-10 248083408-1 Rev 1
Unpacking, Inspecting and Installing
The Lighthouse Worldwide Solutions REMOTE P instruments incorporate MODBUS TCP/IP communication over Ethernet. Connecting a REMOTE P particle counter directly to an Ethernet network should only be done with the permission and guidance of the network administrator.
The primary advantage of using the REMOTE P on an Ethernet network is that a separate network does not have to be installed just for the instruments. Using the Ethernet Port allows the instruments to co­exist with computers, printers and servers on a LAN that is already in place, thereby reducing installation costs.
This document will explain how to program the Network Adapter through its TCP/IP or its RS232 interface. It will also list typical equipment required and provide some troubleshooting information. The end of the document includes examples of T568A and T568B wiring diagrams for straight-through and cross-over cables.
Definitions of Terms Used
This list is provided for the convenience of the user or technician.
ARP - Address Resolution Protocol, a program that can change
the ARP Cache of a host by adding or removing IP addresses. Each IP address is associated to a MAC address.
ARP Cache - List of IP addresses and their associated MAC
addresses for hosts whose addresses cannot be resolved normally.
DHCP - Dynamic Host Configuration Protocol, a program
running on a server that issues IP addresses to computers or devices (Hosts) on its LAN.
Gateway - a network device that controls traffic between two or
more networks.
Hardware address - a unique identifying code programmed by the
factory into a network device, such as a network PCB, comprised of six two-digit groups of letters A-F and numbers 0-9.
Host - a computer or device that allows access to itself via a LAN.
IP (Internet Protocol) Address - a unique logical address used to
identify a host on a TCP/IP network.
LAN - Local Area Network, a group of computers or hosts
connected together in a relatively small geographical area, such as a building or floor of a building.
MAC address - the same as hardware address.
Netmask - a logical hexadecimal number that prevents accessing
hosts outside of its range. A bit value of zero allows access and a non-zero blocks access.
248083408-1 Rev 1 3-11
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Subnet - a logical grouping of hosts based on their IP addresses.
TCP/IP - a communication protocol suite that is used for the
Internet and a large number of LANs that allows hosts to share data.
Telnet - a communication program used primarily to issue
commands directly to a TCP/IP-based host.
Preparing for Network Installation
Note:
where the instruments will operate must have an available IP address for each scheme, or subnet, must match that of the PC that will be used to retrieve the data from the instruments. Contact Lighthouse Technical Support for additional information.
Note: All values in
screen shots are examples only. Use only the values provided by the Network Administrator.
The network
instrument. The IP
Equipment Required:
Network enabled Personal Computer (IBM-compatible);
REMOTE 2014/3014/5014 P
Small 5-port hub or switch and two 3-foot long straight-through Cat5 Ethernet cables (For connection to network)
Or one 3-foot cross-over Cat5 Ethernet cable (For connection to a PC)
Software Required:
Contact the network administrator if any of the following are not installed or not functional:
Windows 2000, XP Professional, or Windows 7
Telnet
Additional Requirements:
1. Contact the network administrator and obtain administrator rights
on the PC to be used for this procedure.
2. For each unit to be attached, obtain an unused static IP address,
Gateway IP address and the appropriate Netmask for the LAN in which they will be installed.
3. Create a list and record the MAC address(es) for all REMOTEs to
be configured. It is suggested that the IP address that will be used for each REMOTE P be written next to the MAC address. This list can be provided to the network administrator for future reference.
4. The PC being used to program the REMOTE P must be using the
same subnet that the adapter will use. See the network administrator for assistance with this.
5. The PC that will monitor or retrieve data from the instrument(s)
must be using the same subnet and gateway programmed on the REMOTE(s) network interface.
3-12 248083408-1 Rev 1
Unpacking, Inspecting and Installing
Configure Device
Ethernet Port Configuration
This section is organized based on the cable used to perform the REMOTE programming.
Straight-through Cat5 requires a hub or switch and two straight-through Cat5 cables. Cross-over Cat5 connects the PC directly to the REMOTE P with no hub. Each will be explained separately. If troubleshooting is required, the cross-over cable technique is easier to use in the field because it requires fewer devices.
It cannot be stressed enough, however, that the instrument and PC be configured using the same IP scheme (IP range, Default Gateway and Netmask).
For troubleshooting outside of the LAN, it is suggested that the PC’s IP address be used as the Default Gateway for both the PC and the REMOTE P.
Frequently, it is necessary to change the instrument’s Ethernet parameters to allow for easier troubleshooting. When this is needed, make sure the instrument is reprogrammed to its previous network settings before reattaching to the LAN. Contact Lighthouse Technical Support or the network administrator for additional information.
Note: Screens shown
are examples only. Data displayed and command responses may differ.
Straight-through Cat5 Setup:
This section requires the PC, two straight-through Cat5 cables and the hub or switch.
1. Connect one end of a straight-through Cat5 cable to the
REMOTE’s Ethernet Port.
2. The other end of the Cat5 cable should plug into one port on the
hub or switch.
3. Attach another straight-through cable to the PC’s RJ45 receptacle
and an open port on the hub or switch.
4. Proceed to “Windows Telnet Programming:” on page 3-14.
Cross-over Cat5 Setup:
1. Attach one end of the cross-over cable to the RJ45 receptacle on
the PC.
2. Attach the other end of the cable to the REMOTE P Ethernet Port.
248083408-1 Rev 1 3-13
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
3. Apply power to the REMOTE. Apply power to the PC if it is not
already running. Observe the LEDs on the hub/switch RJ45 connector - blinking indicates network activity. Observe the network LEDs on the instrument’s RJ45 receptacle - they should blink indicating activity. Proceed to next section.
Program the Interface
WARNING:
only the steps or commands as provided in this guide. Failure to heed this warning can result in damage to equipment, personal injury or data loss and may void the equipment warranties.
Perform
Windows Telnet Programming:
1. Start Windows.
2. On the Taskbar, click on Start.
3. Select Run.
Figure 3-7 RUN Screen
Note: Typing commands
in the command console requires a space between the instruction and the command variables. For example, ’arp’ is followed by a space, then ’-d’ and another space, then the *.
4. In the Run window, type CMD and click OK.
Figure 3-8 Starting the Command Console
3-14 248083408-1 Rev 1
Unpacking, Inspecting and Installing
5. A command prompt window will open. Clear the ARP Cache by
typing arp -d * at the command line and press Enter. Ignore any error messages that indicate the address table doesn’t exist or has no entries.
Figure 3-9 Clear Address Table Command
6. The next step requires the unit’s assigned IP and its MAC
address. The MAC address is a group of six two-digit characters (0-9 and A-F) found on the back of the REMOTE that may be referred to as the HW address.
7. The next step adds the IP and MAC addresses to the ARP Cache
which allows direct communications with the instrument before an IP is programmed into its Ethernet interface.
8. Type arp –s xxx.xxx.xxx.xxx nn-nn-nn-nn-nn-nn and press Enter.
Replace the x’s with the desired IP address, such as 192.168.0.10 or 10.10.0.15, and the n’s with the desired MAC address, such as 00-20-4a-8a-26-31. The IP address will be four segments total with up to 3 digits per segment, each segment separated by a period. Note that the MAC address segments are separated by dashes (-), not periods.
Figure 3-10 ARP Command to Add New IP
9. Activate the REMOTE Ethernet interface by typing, telnet
xx.xx.xx.xx 1 (replace x’s with the IP address to be used for the
instrument) and press Enter. This command will cause a connect error but is required to establish communications.
Figure 3-11 Telnet 1
248083408-1 Rev 1 3-15
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
10. Type telnet xx.xx.xx.xx 9999 and press Enter. This command
accesses the telnet port of the instrument. The expected error message from Step 9. is shown in the example screen below.
Figure 3-12 Telnet 9999
You will have 4 seconds to respond in the next step. If you respond too slowly, you’ll have to repeat step 10.
11. Press Enter to start the REMOTE Ethernet interface setup
program. The example below displays some of the default settings - make no changes to these settings except as instructed.
Figure 3-13 Starting the Set Up Program
12. Type 1 to set the IP address. Type the desired IP address and press
Enter.
3-16 248083408-1 Rev 1
Unpacking, Inspecting and Installing
Figure 3-14 Assigning IP Address
13. Type Y to set the Gateway IP address. Type the Gateway IP
address and press Enter.
Figure 3-15 Assigning Gateway IP Address
14. Type Y to set the Netmask. type the desired Netmask value and
press Enter.
Figure 3-16 Assigning Netmask
248083408-1 Rev 1 3-17
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
15. Type N in response to “Change telnet config password”. The
instrument is shipped without a password. Changing the telnet password or providing one is discouraged, except for absolute security requirements. If a password is applied and forgotten, the instrument will have to be returned to Lighthouse to get the password cleared.
Figure 3-17 Telnet Config Password Screen
Note: When the
Ethernet restarts, it will lose connection to the PC, reported as, “Connection to host lost”.
16. Type S to Save and restart and press Enter to save the changes.
Figure 3-18 Saving the Settings
17. Perform Step 7. though Step 16. for each REMOTE P.
3-18 248083408-1 Rev 1
Unpacking, Inspecting and Installing
WARNING: Step 18 is
very important to prevent network address errors on the PC when the process is complete.
Connect REMOTE P to Ethernet LAN
18. Clear the ARP Cache by type arp –d * and press Enter. If this is
not done before quitting this process, the PC may create “ghost IPs” on the LAN and cause serious problems with the network.
19. Type Exit to quit the command console.
Connect Ethernet Cable to Instrument
Power the REMOTE P OFF before connecting the instrument to a network then attach the network cable to the REMOTE and the other end to the network receptacle. Apply power to the instrument and check the port LEDs to make sure the connection is working.
When all steps are completed, the REMOTE P will supply data to the LAN connection. After approximately 60 seconds, monitoring equipment and software should “see” the instrument and be able to retrieve data from it.
Contact Lighthouse Worldwide Solutions Technical Support for additional information or further assistance.
Data The REMOTE 2014P, 3014P or 5014P stores its data in a 3000-record
rotating data buffer.
The data can be downloaded real-time to LMS Express RT or to a Lighthouse Monitoring System; the instrument can run standalone until you decide to view the data. The data can then be downloaded to LMS Express or LMS XChange manually.
A real-time system will allow you to monitor the instrument counts on a minute-by-minute basis or for the length of the sample time. Downloading the data to LMS Express Real Time allows historical trending of the data.
Use the optional REMOTE Display to view the data as it is being collected.
With any of the data download transfer software, you can save the data to Excel, to a CSV file or to an HTML file.
Please refer to the respective data transfer software manuals for more information.
248083408-1 Rev 1 3-19
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
3-20 248083408-1 Rev 1
00
4 Programming
General Information
The REMOTE 2014P, 3014P or 5014P can be programmed using the MODBUS Protocol or by using the optional REMOTE Display. The full MODBUS protocol is detailed in Appendix A.
This chapter contains the information needed to program the basic configuration for the instrument to meet your needs.
DIP Switches The DIP switches are behind the cover plate illustrated below:
Figure 4-1 Panel Covering the DIP Switches
Remove the two Phillips head screws to expose the DIP switches.
Figure 4-2 Panel Removed, Switches Exposed
248083408-1 Rev 1 4-1
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
DIP Switch Settings
GENERAL DEFINITIONS
OFF (DOWN) = 0, ON (UP) = 1
Table 4-1 DIP Switch settings
Position# Description Setting
1 Binary Bit 0 OFF=0, ON=1
2 Binary Bit 1 OFF=0, ON=1
3 Binary Bit 2 OFF=0, ON=1
4 Binary Bit 3 OFF=0, ON=1
5 Binary Bit 4 OFF=0, ON=1
6 Binary Bit 5 OFF=0, ON=1
7 Remote Display OFF=0, ON=1
8 Communications
Mode
OFF: MODBUS
Note: The DIP Switches
must be set before the unit is powered on.
WARNING: Always
turn the instrument OFF before connecting or disconnecting the REMOTE Display.
Communications Mode
In order to communicate with the instrument using the MODBUS protocol, DIP switch 8 must be set to OFF.
REMOTE Display Mode
In order to use the REMOTE Display option, DIP switch 7 must be set to ON and the Display must be connected to the instrument before it is powered on.
If DIP switch 7 is ON but the REMOTE Display is not connected when the instrument is powered on, the unit will not start. If the Display is not needed, power the instrument OFF, change DIP switch 7 to OFF, then power the instrument ON. Otherwise, power the instrument OFF, connect the REMOTE Display then power the instrument ON.
4-2 248083408-1 Rev 1
Addressing
Programming
Note: Because Address
0 is reserved for broadcasting in RS-485 communications, Address 1 is set whenever all of the dip switches are OFF or when DIP switch1 is ON.
The following table details the addresses set by the binary DIP switches 1-6.
Table 4-2 DIP Switch Addressing
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
0 0 0 0 0 0 1 0 0 0 0 0 1 32
1 0 0 0 0 0 1 1 0 0 0 0 1 33
0 1 0 0 0 0 2 0 1 0 0 0 1 34
1 1 0 0 0 0 3 1 1 0 0 0 1 35
0 0 1 0 0 0 4 0 0 1 0 0 1 36
1 01 0 0 0 5 1 0 1 0 0 1 37
0 1 1 0 0 0 6 0 1 1 0 0 1 38
1 1 1 0 0 0 7 1 1 1 0 0 1 39
0 0 0 1 0 0 8 0 0 0 1 0 1 40
1 0 0 1 0 0 9 1 0 0 1 0 1 41
0 1 0 1 0 0 10 0 1 0 1 0 1 42
1 1 0 1 0 0 11 1 1 0 1 0 1 43
0 0 1 1 0 0 12 0 0 1 1 0 1 44
1 0 1 1 0 0 13 1 0 1 1 0 1 45
0 1 1 1 0 0 14 0 1 1 1 0 1 46
1 1 1 1 0 0 15 1 1 1 1 0 1 47
0 0 0 0 1 0 16 0 0 0 0 1 1 48
1 0 0 0 1 0 17 1 0 0 0 1 1 49
0 1 0 0 1 0 18 0 1 0 0 1 1 50
1 1 0 0 1 0 19 1 1 0 0 1 1 51
0 0 1 0 1 0 20 0 0 1 0 1 1 52
1 0 1 0 1 0 21 1 0 1 0 1 1 53
0 1 1 0 1 0 22 0 1 1 0 1 1 54
248083408-1 Rev 1 4-3
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Table 4-2 DIP Switch Addressing
Communicating with the Instrument
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
1 1 1 0 1 0 23 1 1 1 0 1 1 55
0 0 0 1 1 0 24 0 0 0 1 1 1 56
1 0 0 1 1 0 25 1 0 0 1 1 1 57
0 1 0 1 1 0 26 0 1 0 1 1 1 58
1 1 0 1 1 0 27 1 1 0 1 1 1 59
0 0 1 1 1 0 28 0 0 1 1 1 1 60
1 0 1 1 1 0 29 1 0 1 1 1 1 61
0 1 1 1 1 0 30 0 1 1 1 1 1 62
1 1 1 1 1 0 31 1 1 1 1 1 1 63
The RJ-45 connector on the instrument (marked “RS-485” in Figure 4-3) is used to connect the unit to a COM port on a desktop or laptop PC.
1
8
Figure 4-3 The Communications Ports, Showing Pin Numbers
4-4 248083408-1 Rev 1
Programming
RS-485 Port
The RJ-45 connector on the instrument’s RS-485 port has both RS-232 and RS-485 lines. The pinouts are as shown in Table 4-3:
Table 4-3 RS-232/485 IN Port Pinouts
RJ-45 Pin Signal Name
1 RS-232-TX
2 RS-232 RX
3 RESERVED for future use
4 RS-485B
5 RS-485A
6 RESERVED for future use
WARNING:
Lighthouse Technical Support for the correct instrument configuration BEFORE attempting to use RS485 COM mode with a PC. Failure to heed this warning can result in damage to PC, instrument or both.
Contact
7 RESERVED for future use
8 GND
A standard Cat5 cable is used to connect the instrument to an RS485 network.
RS-485 Communications
The instrument can use RS485 MODBUS as one of its communication modes. Attaching the instrument to an RS485 network requires special attention to topology (wiring scheme of network) before and after the design phase to make sure the REMOTE P and other instruments perform as expected.
The REMOTE P does not support a daisy-chain configuration and must be used in a "star" or hub-based configuration on the network. If the topology is unknown or the user is unsure how it should be installed, please contact Lighthouse before installation.
WARNING: If using
the REMOTE P RS-485 port, DO NOT connect the cable to a system Ethernet port.
A standard Cat5/Cat6 cable is used to connect the instrument to an RS485 network. The pinouts are shown inTable 4-3.
In order to use the RS-485 MODBUS protocol, ensure that DIP Switch 8 is set to 0 (OFF).
248083408-1 Rev 1 4-5
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
RS-232 Communications
The instrument can use RS232 as a communication mode by using the optional USB to RS232 Converter cable. The cable is connected to the RJ-45 Port on the Remote and to any USB port on a PC.
Windows will detect that a new hardware device has been attached and will finalize the installation as needed. This is a routine “registering” of components that should only take a moment. An additional COM port will be added to the PC.
Configuring with REMOTE Display
Configuring with the MODBUS Protocol
Note: For the full
MODBUS register map, please refer to Appendix A.
The instrument can be completely configured using the optional REMOTE Display touchscreen interface.
DIP Switch 7 must be ON before connecting the Display and powering on the instrument.
Please consult the REMOTE Display Operating Manual, LWS PN 248083285-1, for detailed information.
Alternatively, the instrument can be configured using the MODBUS protocol.
When you power up the instrument, the pump will automatically turn on after 5 seconds, and it will begin sampling using the default configuration:
Location = 0
Sample Time = 60 seconds
Hold Time = 0 seconds
Initial Delay = 0 seconds
Note: The automatic
starting of the sampling accommodates systems that do not send a START command, but just polls the instrument for its data.
4-6 248083408-1 Rev 1
To stop the pump and the sampling, send the command 12 to command register 40002.
Programming
Setting the Real Time Clock
The Real Time Clock (RTC) can be read in registers 40027 and 40028:
Table 4-4 Real Time Clock Registers
Register Data Type Description
40027 signed integer Real Time Clock (RTC) [high]. Works in
conjunction with 40028. Displays date and time, in number of seconds since midnight, 1/1/1970.
40028 signed integer Real Time Clock [low]
In order to change the RTC to the current local date/time, enter the high and low values as unsigned integers to registers 40035 and 40036 respectively, which are the Data Set registers.
Table 4-5 Data Set Registers
Register Data Type Description
40035 unsigned integer Data Set [high]. Works in conjunction
with 40036. Data entered here is applied to the device through the command register.
40036 unsigned integer Data Set [low]
Then write the command 13 to the command register 40002. This will write the values in the Data Set registers (40035 and 40036) to the RTC registers (40027 and 40028).
Changing the Default Instrument Parameters
The main instrument parameters involved with the operation of the REMOTE counter are Location, Sample Time, Hold Time and Initial Delay.
The Location is set by writing an unsigned integer to register 40026. The range of values is from 0 to 999.
Sample Time, Hold Time and Initial Delay all use 2 registers, a high word and a low word. If the desired value for any of these parameters is less than 9 hours 6 minutes 8 seconds, then only the low word register needs to be written (in seconds).
248083408-1 Rev 1 4-7
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
The low word register for Sample Time is 40034.
The low word register for Hold Time is 40032.
The low word register for Initial Delay is 40030.
Table 4-6 Instrument Parameters
Register Data Type Description
40026 unsigned integer Location number; Specifies location of
Particle Counter.
40029 unsigned integer Initial Delay [high]. Works in conjunction
with 40030. Number of seconds to wait before starting the first sample. Max value is 359,999, which equals 99h 59m 59s.
40030 unsigned integer Initial Delay [low]
40031 unsigned integer Hold Time [high]. Works in conjunction
with 40032. Number of seconds to wait between sample periods. Max value is 359,999, which equals 99h 59m 59s
40032 unsigned integer Hold Time [low]
40033 unsigned integer Sample Time [high]. Works in
conjunction with 40034. Number of seconds to sample. Max value is 86,399, which equals 23h 59m 59s.
40034 unsigned integer Sample Time [low]
4-8 248083408-1 Rev 1
Programming
Running the Instrument
The REMOTE can be run in several different ways. The applicable action commands are discussed here:
Table 4-7 Action Commands
Value Actio n
1 Saves all writable 4xxxx register values to the
EEPROM.
3 Clears the Data Buffer. Record count is set to zero.
4 Saves the instrument parameters in the 40xxx registers
to the EEPROM. Parameters include Sample Time, Hold Time, Initial Delay, and Location.
5 Enable Remote Control. Locks out the instrument’s
user interface. Can only change instrument parameters via MODBUS.
6 Enable Local Control. Unlocks the instrument’s user
interface. Instrument changes can be made at the device itself or through MODBUS.
7 Turns local pump on.
8 Stop pump.
9 External Start Counter. The instrument samples
continuously until it receives an External Stop Counter command. Does not turn on the pump. Ignores local timing parameters.
10 External Stop Counter. Records counts since External
Start.
11 Instrument Start. Uses defined Initial Delay, Hold
Time, Sample Interval and counting mode. Instrument executes samples and holds until an Instrument Stop command is issued. This command will start the pump.
12 Instrument Stop. Aborts current sample. Stops pump.
Stops data collection.
There are 2 basic modes of operation: Manual counting and Automatic counting. Write each of the described action commands to the command register (40002).
248083408-1 Rev 1 4-9
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
MANUAL Counting Mode
In Manual counting mode, the sample time is based on when the counter is instructed to stop counting. At that point, a data record is recorded and the sample time is the interval between the command to start counting and the command to stop counting.
7 Start Pump
9 Start Manual Count
10 Stop Manual Count; writes a data record, uses the time interval as
the sample time
The “hold” time is effectively the time between a STOP and a START command.
After the last desired sample is taken, send the following to stop the pump:
8 Stop Pump
AUTOMATIC Counting Mode
In Automatic counting mode, the instrument uses the configured sample time, hold time, initial sample delay to record samples.
The instrument will continue running samples until it receives a stop command. When the stop command is given, since the device will not have completed a complete sample cycle, the most current data will not record to the buffer.
After setting all the instrument parameters, run these commands:
11 Start Instrument; to start recording
12 Stop Instrument; to stop recording
4-10 248083408-1 Rev 1
Maintenance Procedures
5 Maintenance
Procedures
This Chapter provides the user with the procedures to perform in maintaining the REMOTE ’P’ instruments.
Safety Before performing any of the maintenance tasks described in this
chapter, review the safety warnings supplied throughout this manual.
Calibration To maintain optimum performance of this instrument, it should be
recalibrated annually by a Lighthouse Authorized Service Provider.
Cleaning This procedure may be superceded by customer requirements;
however, not under any circumstances apply Acetone to the REMOTE ’P’ instrument.
1. To keep the interior of the laser sensor clean, remove the
isokinetic probe from the air inlet and install the protective plastic cap supplied with the instrument.
2. Moisten a lint-free cloth with isopropyl alcohol. The cloth should
be moist, not wet.
3. Wipe down the exterior surfaces of the instrument.
Purge Test 1. Connect the Purge filter to the sample inlet. The purge filter
should be a 0.1 micron, 0.1 CFM filter.
2. Apply power to the instrument.
3. Configure the unit via the user interface to sample for 30 minutes.
4. Allow the instrument to sample through a 30 minute period. This
time allows the unit to warm up and purge any residual particles that might be inside it.
248083408-1 Rev 1 5-1
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
5. Configure the unit via the user interface to sample for 5 minutes
and set a 10 second hold.
6. Set Cycles to 10 so the instrument will take 10 five-minute
samples.
7. If an average of more than one count per five minute sample is
reported, reset the instrument to sample for 30 minutes again to purge it, then repeat the Purge Test again.
8. After the instrument has met the requirement of the Purge test,
return the instrument to its normal location and operating status.
If the instrument still fails the Purge Test, contact Lighthouse Tech Support for assistance.
5-2 248083408-1 Rev 1
MODBUS Register Map v1.48
A MODBUS Register
Map v1.48
COMM Settings
Supported MODBUS Commands
Lighthouse particle counters with MODBUS have the following communications settings:
Table A-1 MODBUS Communications Settings
Baud Rate 19200
Data Bits 8
Stop Bits 1
Parity None
Hardware Protocol RS-232C or RS-485 Standard
Software Protocol MODBUS ASCII (supports upper/lower case)
The MODBUS slave address is set on the particle counter. Valid addresses are 1-63. Address 0 is the broadcast address.
Table A-2 MODBUS Registers
Hex Command Description
03 Read Holding Registers
04 Read Input Registers
06 Write Single Holding Register
See www.modbus.org for documentation on how to use these commands.
248083408-1 Rev 1 A-1
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Register Map Sensor Settings Registers
Instrument settings are stored in holding registers (the 4xxxx series), which are mostly read/writable. Not all holding registers are writable. Table A-3 describes the content of these registers.
Table A-3 Sensor Settings Registers
Register Data Type Description
40001 unsigned integer MODBUS register map version. Matches the version number
of this document. Major version digits are hundreds. Minor version digits are tens and ones. For example, v1.35 = 135d = 0087h.
40002 unsigned integer Command register. Makes the counter execute a command.
See the description of this register in the table below.
40003 unsigned integer Device Status.
[bit 0=RUNNING, bit 1=SAMPLING, bit 2=NEW DATA, bit 3=DEVICE ERROR]
40004 unsigned integer Firmware version. Major version digits are hundreds. Minor
version digits are tens and ones. For example, 210 = v2.10.
40005 unsigned integer Serial Number [high]
40006 unsigned integer Serial Number [low]
40007 ASCII string Product Name char[0], char [1] (NULL terminated string)
40008 ASCII string Product Name char[2], char [3]
40009 ASCII string Product Name char[4], char [5]
40010 ASCII string Product Name char[6], char [7]
40011 ASCII string Product Name char[8], char [9]
40012 ASCII string Product Name char[10], char [11]
40013 ASCII string Product Name char[12], char [13]
40014 ASCII string Product Name char[14], char [15]
40015 ASCII string Model Name char[0], char [1] (NULL terminated string)
40016 ASCII string Model Name char[2], char [3]
40017 ASCII string Model Name char[4], char [5]
40018 ASCII string Model Name char[6], char [7]
40019 ASCII string Model Name char[8], char [9]
A-2 248083408-1 Rev 1
MODBUS Register Map v1.48
Table A-3 Sensor Settings Registers
Register Data Type Description
40020 ASCII string Model Name char[10], char [11]
40021 ASCII string Model Name char[12], char [13]
40022 ASCII string Model Name char[14], char [15]
40023 unsigned integer Flow Rate. Divide by 100 to get rate in CFM. For example,
100 = 1CFM.
40024 unsigned integer Record Count. Total number of records stored in the counter.
40025 unsigned integer Record Index. Zero based index to data in 30xxx register
series. Must be lower than the record count (register 40024). Set this index to expose a counter’s record in the 30xxx registers. Set to -1 to retrieve last record stored in the counter.
40026 unsigned integer Location number.
Particle Counters: Specifies location of Particle Counter. Must be 1 to 200 (maps to location names associated with registers 40200 - 40999). Manifold Controller: Specifies Manifold position. Values 1-32 for the Universal Manifold and values 1-6 for the MiniManifold Controller moves the arm to that position on the manifold. Value 0 moves arm to Home position.
40027 signed integer Real Time Clock (RTC) [high]. Displays instrument’s real-time
clock. Works in conjunction with 40028. Displays date and time, in number of seconds since midnight, 1/1/1970. Can be generated by ANSI C/C++ time() function.
40028 signed integer Real Time Clock [low]
40029 unsigned integer Initial Delay [high]. Works in conjunction with 40030.
Number of seconds to wait before starting the first sample. Max value is 359,999, which equals 99h 59m 59s.
40030 unsigned integer Initial Delay [low]
40031 unsigned integer Hold Time [high]. Works in conjunction with 40032. Number
of seconds to wait between sample periods. Max value is 359,999, which equals 99h 59m 59s.
40032 unsigned integer Hold Time [low]
40033 unsigned integer Sample Time [high]. Works in conjunction with 40034.
Number of seconds to sample. Max value is 86,399, which equals 23h 59m 59s.
248083408-1 Rev 1 A-3
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Table A-3 Sensor Settings Registers
Register Data Type Description
40034 unsigned integer Sample Time [low]
40035 unsigned integer Data Set [high]. Works in conjunction with 40036. Data entered
here is applied to the device through the command register.
40036 unsigned integer Data Set [low]
40037 unsigned integer Alarm Mode. Type of alarming performed
40038 unsigned integer Alarm Parameter. Control parameter for given alarm mode.
40039 unsigned integer Laser Reference Voltage (millivolts)
40040 unsigned integer View Volume. Divide by 100 to get percentage.
For example: 6550d = 65.50%
40041 ASCII string Flow Unit. Defines unit as cfm, lpm, mlpm.
40042 ASCII string Flow Unit. char[2], char[3]
40043 unsigned integer Calibration Reference Voltage (millivolts)
40047 signed integer Calibration Due Date [high]. Indicates when instrument is due
for calibration. this number can be generated by the ANSI C/ C++ time() function.
40048 signed integer Calibration Due Date [low].
40049 signed integer Printer Options
40050 unsigned integer Device Options
40051 signed integer Current record. Displays last record number recorded by
instrument.
40052 signed integer Record number. Set this value to desired record number to
display data in 3xxxx registers. Value corresponds to record number in instrument data buffer. Set to -1 to retrieve the last record stored by the counter.
...
40199 unsigned integer Number of available location names (0=not supported).
200 Locations Available (Register 40050 Bit 3 = 0)
40200 ASCII string Location_1_char[0], char[1] (NULL terminated string)
40201 ASCII string Location_1_char[2], char[3]
40202 ASCII string Location_1_char[4], char[5]
A-4 248083408-1 Rev 1
MODBUS Register Map v1.48
Table A-3 Sensor Settings Registers
Register Data Type Description
40203 ASCII string Location_1_char[6], char[7]
...
40996 ASCII string Location_200_char[0], char[1] (NULL terminated string)
40997 ASCII string Location_200_char[2], char[3]
40998 ASCII string Location_200_char[4], char[5]
40999 ASCII string Location_200_char[6], char[7]
400 Locations Available (Register 40050 Bit 3 = 1)
40200 ASCII string Location_1_char[0], char[1] (NULL terminated string)
40201 ASCII string Location_1_char[2], char[3]
...
40998 ASCII string Location_400_char[0], char[1]
40999 ASCII string Location_400_char[2], char[3]
Registers 40200-40999 are reserved for character names associated with location index values. If number of available locations is 200, Registers 40200-40999 used for 8 character names associated with location index. If number of available locations is 400, Registers 40200-40999 used for 4 character names.
Register 40199 indicates the number of location names supported on this device.
Alarm Mode (40037) defines the type of calculation performed to define an alarm condition. Alarm Mode = 0 corresponds to conventional threshold alarming; channel bit set if threshold exceeded for that given channel.
Alarm Parameter (40038) defines additional parameters that may be needed in defining an alarm mode.
248083408-1 Rev 1 A-5
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Printer Option (40049) displays the configuration of the printer function of the instrument.
Table A-4 Printer Options
Bit Description
0 Unused - Non-writable
1 Print on sample (1 - Enabled, 0 - Disabled)
2-15 Reserved
If Bit 1 of Register 40049 is set, the instrument will print the last recorded data at the end of each sample. This feature cannot be enabled if the One Second Data Update feature is enabled. It will be disabled if the One Second Data Update feature is enabled.
Device Options (40050) displays the device configuration of the instrument.
Table A-5 Device Options
Bit Description
3 Number of Locations (1-400 locations with 4 character
names, 0-200 locations with 8 character names)
If Bit 3 is of Register 40500 is set, the number of locations available (Register 40199) is set to 400 and Registers 40200-40999 are used for four character names associated with location index values. For example, the name for location 3 would be located at registers 40204-
40205.
If Bit 3 is of Register 40500 is 0, up to 200 locations can be specified. Registers 40200-40999 are reserved for eight character names associated with location index values. Thus the name for location=3 would be located at registers 40208-40211. Register 40199 indicates the number of location names supported on the device.
The Command Register (40002) is used to make the device perform an action. This register performs an action when an integer value is written to it. The action is completed when the device sends a MODBUS response. When this register is read, it always returns a zero.
A-6 248083408-1 Rev 1
MODBUS Register Map v1.48
Device Status
The Device Status register (40003) displays the current status of the device.
Table A-6 Device Status
Bit Description
0 RUNNING: Set when a start command is executed remotely via
Command 9 (manual start) or Command 11 (instrument start) or through the user interface. The flag will remained set until a stop command is executed.
1 SAMPLING: This is set only when the instrument is actually
sampling data that is to be recorded. Caution must be used in sending a command during this time that may invalidate current sample.
2 NEW DATA: Set to 1 to indicate that a new data record has been
recorded and it hasn't been read via modbus yet. When a data record has been read via modbus (registers 30001 to 30999), then this flag is reset to zero.
Command Register
3 DEVICE ERROR: In the event that there is a failure on the device,
this bit is set to indicate possible invalid data collected. An example of a device error could be a positioning error on a manifold device.
The Command Register (40002) is used to make the device perform an action. The register performs an action when an integer value is written to it. The action is completed when the device sends a MODBUS response. When this register is read, it always returns a zero.
Table A-7 Command Register
Value Action
1 Saves all writable 4xxxx register values to the device’s EEPROM.
2 Reserved for future use.
3 Clears the Data Buffer. Record count is set to zero.
4 Saves the instrument parameters in the 40xxx registers to the EEPROM.
Parameters include Sample Time, Hold Time, and Initial Delay.
5 Enable Remote Control. Locks out the instrument’s user interface. Can only
change instrument parameters via MODBUS.
6 Enable Local Control. Unlocks the instrument’s user interface. Instrument
changes can be made at the device itself or through MODBUS.
7 Turns local pump on, if applicable. Flow is regulated by an internal setpoint.
8 Stop pump, if applicable.
248083408-1 Rev 1 A-7
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Table A-7 Command Register
Value Action
9 Manual Start.
The instrument samples continuously until it receives a Manual Stop command. Ignores local timing parameters. Sets Sample Time for data record to equal the time interval between the Manual Start and Manual Stop command. If applicable to device, does not start pump.
10 Manual Stop.
Stops sampling. Records count since Manual Start.
11 Instrument Start (Automatic Counting).
Particle Counters counting mode. Instrument executes samples and holds until an Instrument Stop command is issued. For instruments with pumps, this command will start the pump. Manifold Controller changing positions when an Instrument Stop command is issued.
12 Instrument Stop. Aborts current sample. Stops pump, if applicable. Stops data
collection.
: Uses defined Initial Delay, Hold Time, Sample Interval and
: Uses defined Manifold Sequence. Stops counting and
Data and Alarm Registers
13 Set Real Time Clock.
Writes “Data Set” values (from Registers 40035 & 40036) to the local Real Time Clock. New time value is saved.
192 Change BAUD to 19200
576 Change BAUD to 57600
1152 Change BAUD to 115200
Data and Alarm Enable Registers
The Data and Alarm Enable input registers (43xxx series) are read/ write. All enable data items are 4 bytes long and are stored across 2 registers. Byte and word ordering is big-endian. Thus, data items are formed by placing the high bytes in front of the low bytes. For example:
<High Bytes><Low Bytes> = <4 Byte Data Item>
A-8 248083408-1 Rev 1
MODBUS Register Map v1.48
The 43xxx register series is used to determine which particle data channel is ENABLED and which are set to ALARM ENABLE. These registers supersede the older Data Enable Registers (31xxx) which have been obsoleted.
Table A-8 Enable/Disable Bits
Bit Description
0 DATA ENABLE (0=disable; 1=enable)
1 ALARM ENABLE (0=disable; 1=enable)
These registers run in parallel with the data registers (30xxx series). For example, data register 30010’s enable register would be 43010. Data register 30016’s enable register would be 43016.
Note: Alarm Enable
currently only works for Particle Channels.
The user can enable multiple particle channels for alarming at the same time.
Particle data registers for the Enable setting start at 43009 for the high word and 43010 for the low word for particle channel 1.
Table A-9 Alarm Enable Registers
Register Data Type Description
43009 unsigned int Enable for Particle Channel 1 [high] (smallest
particle size starts here)
43010 unsigned int Enable for Particle Channel 1 [low]
43011 unsigned int Enable for Particle Channel 2 [high]
43012 unsigned int Enable for Particle Channel 2 [low]
43013 unsigned int Enable for Particle Channel 3 [high]
43014 unsigned int Enable for Particle Channel 3 [low]
43015 unsigned int Enable for Particle Channel 4 [high]
43016 unsigned int Enable for Particle Channel 4 [low]
43017 unsigned int Enable for Particle Channel 5 [high]
43018 unsigned int Enable for Particle Channel 5 [low]
43019 unsigned int Enable for Particle Channel 6 [high]
43020 unsigned int Enable for Particle Channel 6 [low]
...
248083408-1 Rev 1 A-9
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Table A-9 Alarm Enable Registers
Register Data Type Description
43041 unsigned int Enable for Analog Channel 1 [high]
43042 unsigned int Enable for Analog Channel 1 [low]
43043 unsigned int Enable for Analog Channel 2 [high]
43044 unsigned int Enable for Analog Channel 2 [low]
43045 unsigned int Enable for Analog Channel 3 [high]
43046 unsigned int Enable for Analog Channel 3 [low]
43047 unsigned int Enable for Analog Channel 4 [high]
43048 unsigned int Enable for Analog Channel 4 [low]
Enable Alarming for a Channel
To enable alarming on the third particle channel, the user would enable Bit 1 for register 43014.
To disable alarming on the third channel and enable alarming on the second channel, disable Bit 1 for register 43014 and enable Bit 1 for register 43012.
To disable alarming completely, disable Bit 1 for register 43012. Now, no channels are enabled for alarms.
Table A-10 Example of Alarming on Channel 2
Registers
Particle
Channel
Bit 1
Enabled
43009 - 43010 1 0
43011 - 43012 2 1
43013 - 43014 3 0
43015 - 43016 4 0
43017 - 43018 5 0
43019 - 43020 6 0
Use the Threshold registers to set the alarm threshold value. This is described in the next section.
A-10 248083408-1 Rev 1
MODBUS Register Map v1.48
Threshold Setup Registers
Threshold data is stored in the input registers in the 45xxx series which are read/write. All threshold data items are 4 bytes long and are stored across 2 registers. Byte and word ordering is big-endian. Thus, data items are formed by placing the high bytes in front of the low bytes. For example:
<High Bytes><Low Bytes> = <4 Byte Data Item>
For particle channels, the threshold value is a 32-bit unsigned integer. If the data value exceeds the threshold value and the alarm is enabled for that channel, the threshold flag in the Data Status register (30007­30008, bit 4) is set.
Note: The table below
shows the registers for an 8 channel particle counter. Counters with less channels do not use the extra registers. The smallest particle channel starts at the xxx09 position.
The threshold registers (45xxx series) run in parallel with the data registers (30xxx series). For example, data register 30010’s corresponding threshold register would be 45010. Data register 30016’s threshold register would be 45016.
Table A-11 Alarm Threshold Registers
Register Data Type Description
45009 unsigned int Threshold for Particle Channel 1 [high]
(smallest particle size starts here)
45010 unsigned int Threshold for Particle Channel 1 [low]
45011 unsigned int Threshold for Particle Channel 2 [high]
45012 unsigned int Threshold for Particle Channel 2 [low]
45013 unsigned int Threshold for Particle Channel 3 [high]
45014 unsigned int Threshold for Particle Channel 3 [low]
45015 unsigned int Threshold for Particle Channel 4 [high]
45016 unsigned int Threshold for Particle Channel 4 [low]
45017 unsigned int Threshold for Particle Channel 5 [high]
45018 unsigned int Threshold for Particle Channel 5 [low]
45019 unsigned int Threshold for Particle Channel 6 [high]
45020 unsigned int Threshold for Particle Channel 6 [low]
248083408-1 Rev 1 A-11
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Setting the Alarm Threshold Value
The Alarm Threshold Value is set in the low register of the channels.
Table A-12 Alarm Threshold Registers set to
default value
Data Registers
Registers
45009 - 45010 1 1000
45011 - 45012 2 1000
45013 - 45014 3 1000
45015 - 45016 4 1000
45017 - 45018 5 1000
45019 - 45020 6 1000
Data is stored in the input registers (30xxx series), which are read-only. All data items are four bytes long and are stored across two registers. Byte and word order for particle data is big-endian. Thus, data items are formed by placing the high bytes in front of the low bytes.
Example:
<High Bytes><Low Bytes> = <4 Byte Data Item>
Particle
Channel
Threshold
Value
Analog data is little-endian. Thus, analog data items are formed by placing the low bytes in front of the high bytes.
Example:
<Low Bytes><High Bytes> = <4 Byte Data Item>
Not all particle and analog channels are necessarily active. Retrieving data from an inactive channel returns garbage. See the Data Enable Registers section of this document for details on how to record data from active channels.
This entire series of registers represents one data record in the device. The Record Index Register (40025) must be changed to index other records here.
A-12 248083408-1 Rev 1
MODBUS Register Map v1.48
The first record in the data buffer is located at Index=0. The most recently saved value is at Index=-1.
Table A-13 Data Registers
Register Data Type Description
30001 signed integer Timestamp [high] (# of seconds
since midnight, 1/1/1970)
30002 signed integer Timestamp [low]
30003 unsigned integer Sample Time [high] (In seconds)
30004 unsigned integer Sample Time [low]
30005 signed integer Location [high] (Place where data
was recorded)
30006 signed integer Location [low]
30007 unsigned integer Device Status [high]
30008 unsigned integer Device Status [low]
30009 unsigned integer Particle Channel 1 [high]
30010 unsigned integer Particle Channel 1 [low]
30011 unsigned integer Particle Channel 2 [high]
30012 unsigned integer Particle Channel 2 [low]
30013 unsigned integer Particle Channel 3 [high]
30014 unsigned integer Particle Channel 3 [low]
30015 unsigned integer Particle Channel 4 [high]
30016 unsigned integer Particle Channel 4 [low]
30017 unsigned integer Particle Channel 5 [high]
30018 unsigned integer Particle Channel 5 [low]
30019 unsigned integer Particle Channel 6 [high]
30020 unsigned integer Particle Channel 6 [low]
...
30041 IEEE Float Analog Channel 1 [high]
30042 IEEE Float Analog Channel 1 [low]
30043 IEEE Float Analog Channel 2 [high]
248083408-1 Rev 1 A-13
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Table A-13 Data Registers
Register Data Type Description
30044 IEEE Float Analog Channel 2 [low]
30045 IEEE Float Analog Channel 3 [high]
30046 IEEE Float Analog Channel 3 [low]
30047 IEEE Float Analog Channel 4 [high]
30048 IEEE Float Analog Channel 4 [low]
...
30073 unsigned int Valid analog channels [bit0=ch 1,
30074 unsigned int Valid particle channels
30075 unsigned int Alarm Flags - Analog Channels
..., bit15=ch16]
(bit 0 = channel 1 ...)
Note: Particle data is
always a cumulative raw count regardless of the instrument’s settings.
Note: Although
MODBUS sends 4 bytes of status information, Lighthouse instruments only use the first (least significant) byte.
30076 unsigned int Alarm Flags - Particle Channels
The timestamp field indicates when the data record was recorded. Timestamps are stored as the number of seconds since 1/1/1970, the Unix time epoch. This value can be written directly into a C/C++ time_t data type to be used by ANSI C time functions.
Device Status Word (30007 - 30008)
The registers used for the Device Status Word are 30007 and 30008.
The bit order of the Device Status Word is 7 to 0 (right to left), where bit 7 is the most significant bit and bit 0 is the least significant bit.
The bits within the Device Status Word are flagged to indicate particular conditions of the currently indexed data record.
A-14 248083408-1 Rev 1
MODBUS Register Map v1.48
If multiple states occur, the bits are added together. For example, a Flow Alert and a Particle Overflow would return a value of 6 in register 30008 (bits 1 and 2 are set TRUE).
Table A-14 Device Status Word
Bit Description
0 Laser Alert Status
0 = Laser is OK 1 = Laser Alert
1 Flow Alert Status
0 = Flow Rate is OK 1 = Flow Rate Alert
2 Particle Overflow Status
0 = No overflow 1 = Overflow occurred
3 Instrument Service Status
0 = Working correctly 1 = Instrument malfunction detected.
4 Threshold High Status
0 = Threshold not exceeded 1 = Threshold exceeded
5 Threshold Low Status
0 = Threshold not exceeded 1 = Threshold exceeded
6 Instrument Sampler Status
0 = Nominal Operation 1 = Sampler Error
Bits 7 to 31 are currently unused.
Valid Data in Channels (30073 - 30076)
Register 30073 represents the flag bits corresponding to valid data present in the analog register range. The mapping is such that bit 0 set to TRUE (=1) would correspond to valid data present in Analog Channel 1.
Register 30074 represents the flag bits corresponding to valid data present in the particle register range.
Register 30075 represents the flag bits corresponding to analog channels that have exceeded the threshold [Threshold High Registers (45xxx series)] based on alarm mode.
Register 30076 represents the flag bits corresponding to particle channels that have exceeded the threshold [Threshold High Registers (45xxx series)] based on alarm mode
248083408-1 Rev 1 A-15
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Data Type Registers
Note: All data records
have the same data types assigned to them. The user does not have to read the data type registers for every record.
The 41xxx register series is used to identify the type of data items in the 30xxx series. The Data Type registers run in parallel with the Data Registers. For example, Data Register 30041’s Data Type register is
41041.
Data Types are assigned 4 ASCII characters across 2 registers. If a Data Type string contains less than 4 characters, then the rest of the string is padded with NULL characters. Note that a Data Type using all four characters will not end with a NULL character.
Table A-15 Data Types
String Description
TIME Timestamp
STIM Sample Time
SVOL Sample Volume
LOC Location
STAT Status
TEMP Temperature
RH Relative Humidity
AIRV Air Velocity
DPRS Differential Pressure
ESD Electrostatic Discharge
FLOW Flow Rate
LASV Laser Voltage
VOLT Voltage
PRES Pressure
A-16 248083408-1 Rev 1
MODBUS Register Map v1.48
Note: Only Particle data
types have numbers in their strings.
Particle data items are typed specially. They contain numbers, sometimes a space and sometimes a period used as a decimal point. These entries are used to identify particle channel sizes and are always expressed in microns. These types represent raw counts only.
Table A-16 Examples of Particle Data Items
String Description
0.3 Particle type of size 0.3 micron
1.0 Particle type of size 1.0 micron
20.0 Particle type of size 20.0 micron
.015 Particle type of size 0.015 micron or 15 nanometer
Data Units Registers
The 42xxx register series identifies the units used by data items in the 30xxx series. These registers run in parallel with the Data Registers. For example, Data Register 30010’s Units Register is 42010.
Note: Not all data types
have units. LWS Particle Counters may use units not on the table.
Units are stored as 4 character ASCII strings across 2 registers. If the Units string contains less than 4 characters or no characters at all, the rest of the string is padded with NULLs.
The table below shows units that may be sent by the device. Some of these units are not currently used but are reserved for future use.
Table A-17 Data Units
Units Description Units Description
# Count (For Particles) ft/m Feet per minute
% Percent m/s Meters per second
s Seconds “H2O Inches of water
min Minutes “Hg Inches of mercury
hour Hours mmWa Millimeters of water
F Fahrenheit mmHg Millimeters of mercury
C Celsius cmHg Centimeters of mercury
K Kelvin Pa Pascals
ft Feet kPa Kilopascals
248083408-1 Rev 1 A-17
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Table A-17 Data Units
Units Description Units Description
m Meters Bar Bar
ft^2 Square feet mBar Milli-bar
m^2 Square meters V Volts
ft^3 Cubic feet mV Milli-volts
m^3 Cubic meters A Amperes
L Liters mA Milli-amps
CFM Cubic feet per minute Ohm Ohms
CMM Cubic meters per minute mOhm Milli-ohm
L/m Liters per minute p/f3 Particles per cubic foot
p/m3 Particles per cubic meter
NOTE: This register bank is obsolete and is maintained for backward compatibility.
A-18 248083408-1 Rev 1
B Zero Count Test
Zero Count Test
How to Run the Test
This is the Zero Count Test procedure. A Purge (Zero Count) Filter must be attached to the instrument and 6 five-minute samples must be taken. There should be no more than 1 count, average, per five-minute sample. The purge filter should be a 0.1 micron filter at 0.1 CFM.
1. Connect the instrument to the monitoring system.
2. Attach the Purge Filter to the sample inlet.
3. Apply power to the instrument.
4. Allow the instrument to sample through a 30 minute period. This
time allows the unit to warm up and purge any residual particles that might be inside it.
5. Configure the unit to sample for 5 minutes.
6. Allow the instrument to sample 6 five-minute periods.
7. If the average count for the total of the six sample periods is NOT
less than one per five minute period or if more than one count occurs in ANY five-minute sample period, allow the instrument to sample for an additional 30 minutes to purge it and repeat the test.
8. If the instrument still fails the Zero Count Test, call Lighthouse
Technical Support for assistance.
After the instrument meets the requirements of the Zero Count test, turn it off, remove the Purge Filter and return the instrument to its normal location and operating status.
248083408-1 Rev 1 B-1
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
B-2 248083408-1 Rev 1
00
C Limited Warranty
Limitation Of Warranties:
A. Lighthouse Worldwide Solutions (LWS) warrants that all
equipment shall be free from defects in material and workmanship under normal use for a period of two years from date of shipment to Buyer except that LWS does not warrant that operation of the software will be completely uninterrupted or error free or that all program errors will be corrected. Buyer shall be responsible for determining that the equipment is suitable for Buyer’s use and that such use complies with any applicable local, state, or federal law. Provided that Buyer notifies LWS in writing of any claimed defect in the equipment immediately upon discovery and any such equipment is returned to the original shipping point, transportation charges prepaid, within two years from date of shipment to Buyer and upon examination LWS determines to its satisfaction that such equipment is defective in material or workmanship, i.e. contains a defect arising out of the manufacture of the equipment and not a defect caused by other circumstances, including, but not limited to accident, misuse, unforeseeable use, neglect, alteration, improper installation, improper adjustment, improper repair, or improper testing, LWS shall, at its option, repair or replace the equipment, shipment to Buyer prepaid. LWS shall have reasonable time to make such repairs or to replace such equipment. Any repair or replacement of equipment shall not extend the period of warranty. If the Instrument is modified or in any way altered without the explicit written consent of LWS then the warranty is null and void. This warranty is limited to a period of two years, except as noted below, without regard to whether any claimed defects were discoverable or latent on the date of shipment. The length of warranty for pumps in the HANDHELD and REMOTE ’P’ particle counters is one (1) year. Batteries and accessories with all products are warranted for one (1) year. Fuses and purge filters carry no warranty. If a third party battery is used in the product, the product warranty is null and void. If the battery is charged by a third party battery charger the battery warranty is null and void.
B. If Buyer shall fail to pay when due any portion of the purchase
price or any other payment required from Buyer to LWS under this contract or otherwise, all warranties and remedies granted under this Section may, at LWS’s option, be terminated.
C. THE FOREGOING WARRANTY IS EXCLUSIVE AND IN
LIEU OF ALL OTHER REPRESENTATIONS, WARRANTIES AND COVENANTS, EXPRESS OR IMPLIED WITH RESPECT TO THE EQUIPMENT AND ANY DEFECTS THEREIN OF ANY NATURE WHATEVER, INCLUDING AND WITHOUT LIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. LWS SHALL NOT BE LIABLE FOR, AND BUYER ASSUMES ALL RISK OF, ANY ADVICE OR FAILURE TO PROVIDE ADVICE BY LWS TO BUYER REGARDING THE EQUIPMENT OR BUYERS USE OF THE SAME. UNDER NO CIRCUMSTANCES SHALL LWS BE
LIABLE TO BUYER UNDER ANY TORT, NEGLIGENCE, STRICT LIABILITY, OR PRODUCT LIABILITY CLAIM AND BUYER AGREES TO WAIVE SUCH CLAIMS. LWS’s SOLE AND EXCLUSIVE LIABILITY AND BUYERS SOLE AND EXCLUSIVE REMEDY, FOR ANY NONCONFORMITY OR DEFECT IN THE PRODUCTS OR ANYTHING DONE IN CONNECTION WITH THIS CONTRACT, IN TORT, (INCLUDING NEGLIGENCE), CONTRACT, OR OTHERWISE, SHALL BE AS SET FORTH IN THE SUBSECTION A HEREOF AS LIMITED BY SUBSECTION B HEREOF. THIS EXCLUSIVE REMEDY SHALL NOT HAVE FAILED OF ITS ESSENTIAL PURPOSE (AS THAT TERM IS USED IN THE UNIFORM COMMERCIAL CODE) PROVIDED THAT THE SELLER REMAINS WILLING TO REPAIR OR REPLACE DEFECTIVE EQUIPMENT (AS DEFINED IN SUBSECTION A) WITH A COMMERCIALLY REASONABLE TIME AFTER RECEIVING SUCH EQUIPMENT. BUYER SPECIFICALLY ACKNOWLEDGES THAT SELLER’S PRICE FOR THE EQUIPMENT IS BASED UPON THE LIMITATIONS OF LWS’S LIABILITY AS SET FORTH IN THIS CONTRACT.
Warranty Of Repairs After Initial Two (2) Year Warranty:
A. Upon expiration of the initial two-year warranty, all parts and
repairs completed by an authorized Lighthouse repair technician are subject to a six (6) month warranty.
B. Other than the above, LWS makes no warranty of any kind,
expressed or implied, except that the products manufactured and sold by LWS shall be free from defects in materials and workmanship and shall conform to LWS’s specifications; Buyer assumes all risk and liability resulting from use of the products whether used singly or in combination with other products. If instrument is modified or in any way altered without the explicit written consent of LWS, then the warranty is null and void.
C. WARRANTY REPAIRS SHALL BE COMPLETED AT THE
FACTORY, BY AN AUTHORIZED SERVICE LOCATION, BY AN AUTHORIZED SERVICE TECHNICIAN, OR ON SITE AT BUYER’S FACILITY BY A LIGHTHOUSE AUTHORIZED EMPLOYEE. BUYER PAYS FREIGHT TO FACTORY; SELLER WILL PAY STANDARD RETURN FREIGHT DURING THE WARRANTY PERIOD. BUYER MAY SELECT A FASTER METHOD OF SHIPMENT AT ITS OWN EXPENSE.
248083408-1 Rev 1 C-1
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
C-2 248083408-1 Rev 1
00
Index
Numerics
2014P instrument 2-1 3014P instrument 2-1 4-20mA Analog Inputs 3-6 5014P instrument 2-1 5-port hub 3-12
A
Accessories 2-2 Additional help 1-i Addressing 4-3 Administrator rights 3-12 Alarm Enable Registers A-8 Alarm Registers
Enable Alarming A-10 Analog Input Ports 3-5 Analog Sensors 3-6
4-20mA Analog Inputs 3-6
Connector Pinout 3-6
Ports 3-7 ARP 3-11 ARP Cache 3-11 ARP Command to Add New IP 3-15 arp -d * 3-15 arp –d * 3-19 arp –s 3-15 Assigning Gateway IP Address 3-17 Assigning Netmask 3-17 AUTOMATIC Counting Mode 4-10 Automatic Mode 4-9
B
Big-endian data A-12
C
Clear LAT 3-19 Clear the Data Buffer 4-9 Clearing the Address Table 3-15 Clearing the ARP Cache 3-19 CMD 3-14 Command Console 3-14 Command Register A-7 Communicating with the Instrument 4-4 Communication Mode 4-2 Communication Ports 3-5 Communication ports 3-5 Communications Settings A-1 Configuring with REMOTE Display 4-6 Configuring with the MODBUS Protocol 4-6 Connection to host lost 3-18 Connections 3-4
Analog Input Ports 3-5 Communication ports 3-5 Power 3-5
REMOTE Display Connector 3-5 Cross-over Cat5 3-12 Cross-over Cat5 Setup 3-13
D
Data 3-19 Data Registers A-12, A-15
Device Status Word A-14 Data Type Registers A-16 Data Units Registers A-17 Default Instrument Parameters
Changing 4-7 Definitions 3-11 Device Status A-7 Device Status Word A-14 DHCP 3-11 DIP Switches 3-7, 4-1
248083408-1 Rev 1 I-1
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Settings 4-2
Dip Switches 3-5
E
Enable Alarming A-10 Ethernet Configuration 3-13 External Start Counter 4-9 External Stop Counter 4-9
F
Features 3-3
4-20mA input 3-3 Auto-adjusting flow control 3-4 DIP switch control 3-3 Ethernet and RS-485 communications 3-4 Internal pump 3-4 MODBUS protocol 3-4 Remote Display 3-3 Rotating buffer 3-4
Front Panel LEDs 3-3
G
Gateway 3-11 Gateway IP 3-12 General Information 4-1
L
LAN 3-11 LAN topology 3-5, 4-5 Laser Source 2-5 Lighthouse Monitoring System 3-19 Limitation Of Warranties C-1 Limited Warranty C-1 Little-endian data A-12 LMS Express 3-19 LMS Express Real Time 3-19 LMS Express Real Time Plus 3-19 LMS XChange 3-19 Local Area Network 3-11 Local Control 4-9
M
MAC address 3-11, 3-12, 3-15 Maintenance 5-1 MANUAL Counting Mode 4-10 Manual Start 4-9 Manual Stop 4-9 MODBUS Protocol 4-1 Model number
Explanation 2-1
N
Netmask 3-11
H
Hardware address 3-11 Help 1-i Historical trending 3-19 Host 3-11
Network administrator 3-12
O
Operation 3-3
Understanding LEDs 3-3
I
P
Instrument Control 4-9 Instrument Start 4-9 IP 3-15 IP Address 3-11 IP address 3-12
I-2 248083408-1 Rev 1
Package Contents 3-1 Peripheral Interface Connector 3-5, 3-8 Personal Computer 3-12 Power 3-5, 3-7 Power Switch 3-5, 3-7
Index
Preventive Maintenance 5-1 Programming
DIP Switches 4-2
R
Real Time Clock
Setting 4-7 Real-time system 3-19 Register Map A-2 REMOTE 2014P
4 Channel Threshold 2-3
6 Channel Threshold 2-3
Calibration 2-3
Count Modes 2-3
Counting Efficiency 2-3
Data Storage 2-3
Dimensions 2-3
Display 2-3
Enclosure 2-3
Environmental Sensors 2-3
Flow Rate 2-3
Laser Source 2-3
Operating Temp/RH 2-3
Power Input 2-3
Power Requirement 2-3
Size Range 2-3
Storage Temp/RH 2-3
Vacuum Source 2-3
Weight 2-3
Zero Count Level 2-3 REMOTE 3014P
4 Channel Threshold 2-4
6 Channel Threshold 2-4
Calibration 2-4
Communication Modes 2-4
Counting Efficiency 2-4
Data Storage 2-4
Dimensions 2-4
Display 2-4
Enclosure 2-4
Environmental Sensors 2-4
Flow Rate 2-4
Laser Source 2-4
Operating Temp/RH 2-4
Power Requirement 2-4
Power Supply Input 2-4 Size Range 2-4 Storage Temp/RH 2-4 Supporting Software 2-4 Vacuum Source 2-4 Weight 2-4 Zero Count Level 2-4
REMOTE 5014P
4 Channel Threshold 2-5 6 Channel Threshold 2-5 Calibration 2-5 Communication Modes 2-5 Counting Efficiency 2-5 Data Storage 2-5 Dimensions 2-5 Display 2-5 Enclosure 2-5 Environmental Sensors 2-5 Flow Rate 2-5 Operating Temp/RH 2-5 Power Requirement 2-5 Power Supply Input 2-5 Size Range 2-5 Specifications 2-5 Storage Temp/RH 2-5 Supporting Software 2-5 Vacuum Source 2-5
Weight 2-5 Remote Control 4-9 REMOTE Display 3-19 REMOTE Display Connector 3-5 Remote Display Connector 3-8 RJ-45 Ethernet Connector 3-5 Rotating data buffer 3-19 RS 4-5 RS-232 Communications 4-6 RS-232/485 IN 3-5 RS-232/485 IN Port 4-5 RS-485 Communications 4-5 RS-485 OUT Port 4-6 RS485 topology 3-5, 4-5 Running the Instrument 4-9
S
Safety 1-1, 5-1
248083408-1 Rev 1 I-3
Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
Laser safety information 1-1 Save 4-9 Saving the Settings 3-18 Sensor Settings Registers A-2 Set adapter’s IP address 3-16 Setting DIP Switches 4-2 Setting the Alarm Threshold Value A-12 Setting the Real Time Clock 4-7 Shipping instructions 3-2 Straight-through Cat5 3-12 Straight-through Cat5 Setup 3-13 Subnet 3-12 Supported MODBUS Commands A-1 Supporting Software 2-3
T
TCP/IP 3-12 Telnet 3-12 Telnet 1 3-15 Telnet 9999 3-16 Telnet Config Password Screen 3-18 Telnet Programming 3-14 Text conventions 1-i Threshold Setup Registers A-11
Setting the Alarm Threshold Value A-12
W
Warning
Infrared Radiation 1-1 Warranty C-1 Warranty Of Repairs After Initial Two (2) Year
Warranty C-1
Z
Zero Count Check B-1 Zero Count Level 2-5 Zero Count Test B-1
U
Understanding LEDs 3-3 USB / RS232 Communications 3-8 USB A to B cable 3-9 USB cable 3-9 USB Connector 3-5 USB driver install 3-8 USB driver installation notes 3-8 USB port assignment 3-10
V
Valid Data in Channels (30073-30076) A-15
I-4 248083408-1 Rev 1