Lighthouse Handheld 2016, Handheld 3016, Handheld 5016 Operating Manual

OPERATING MANUAL
Handheld
PARTICLE COUNTERS
Handheld 2016, 3016, 5016
00
Lighthouse Worldwide Solutions
HANDHELD 2016, 3016, 5016 Airborne Particle Counter Gen F
Operating Manual
Copyright © 2011-2017 by Lighthouse Worldwide Solutions. All rights reserved. No part of this document may be reproduced by any means except as permitted in writing by Lig hthouse Worldwide Solutions.
The information contained herein constitutes valuable trade secrets of Lighthouse Worldwide Solutions. You are not permitted to disclose or allow to be disclosed such information except as permitted in writing by Lig hthouse Worldwide Solutions.
The information contained herein is subject to change without notice. Lighthouse Worldwide Solutions is not responsible for any damages arising out of your use of the LMS program.
HANDHELD 2016™, HANDHELD 3016™, HANDHELD 5016™ and LMS™ are trademarks of Lighthouse Worldwide Solutions.
Microsoft®, Microsoft Windows© and Excel© are trademarks of Microsoft Corporation.
LWS Part Number: 248083400-1 Rev 7
EU DECLARATION OF CONFORMITY
Manufacturer’s Name Lighthouse Worldwide Solutions, Inc.
Manufacturer’s Address: Lighthouse Worldwide Solutions, Inc.
1221 Disk Drive Medford, OR 97501
Declares that the product:
Product Name: Handheld Airborne Particle Counter Model Number(s): Handheld 2016, 3016, 5016
Conforms to the following Product Specifications:
SAFETY
LASER SAFETY
EMC
UL 61010A-1 – UL Standard for Safety Electrical Equipment for Laboratory Use; Part 1: General Requirements. Replaces UL 3101-1
Supplementary information: The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC amended by Directive 93/68/EEC and the EMC Directive 89/336/EEC amended by Directive 93/68/EEC, and carries the CE marking accordingly.
EN61010-1:2001 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use Part I: General Requirements IEC 61010-1:2000
CAN/CSA C22.2 Safety Requirements for Electrical No. 1010.1-1992 Equipment for Measurement, Control, and Laboratory
Use, Part1: General Requirements
IEC 60825-1 Am. 2 Guidance on Laser Products: Conforms to IEC 60601-2-22 FDA 21 CFR Chapter 1 Subchapter J (Laser Notice 50)
EN61326 Electrical Equipment for Measurement, Control and
Laboratory Use EMC Requirements Part 1: General Requirements Includes Amendment A1:1998;IEC 61326:1997 + A1:1998
Fremont, CA, May 15, 2007 William L. Shade – V.P. Engineering
00
Table of Contents
About this Manual
Text Conventions ................................................................................................................ i
Additional Help ................................................................................................................... i
Chapter 1 General Safety
General Safety ............................................................................................................... 1-1
Laser Safety Information ............................................................................................... 1-1
Chapter 2 Introduction
Overview ........................................................................................................................ 2-1
Description ..................................................................................................................... 2-2
HANDHELD Specifications .......................................................................................... 2-3
Chapter 3 Unpacking, Inspecting and Installation
Initial Inspection ............................................................................................................ 3-1
Unpacking ...................................................................................................................... 3-1
Shipping Instructions ..................................................................................................... 3-2
Accessories .................................................................................................................... 3-2
Prepare Instrument for Use ............................................................................................ 3-3
Power Requirements ......................................................................................... 3-3
Install the Battery ................................................................................... 3-3
Connect Power ................................................................................................... 3-6
Battery Removal ............................................................................................... 3-7
Optional Printer Accessory ................................................................................ 3-9
Connecting to an External Computer or Facility Management System .......... 3-10
Chapter 4 Operation
Using the Instrument ...................................................................................................... 4-1
Touch Screen Overview ................................................................................................. 4-2
Menu Map ......................................................................................................... 4-2
MAIN Screen ................................................................................................................. 4-3
248083400-1 Rev 7
t-i
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
LOCATION Selection ....................................................................................... 4-7
Changing Locations ............................................................................... 4-7
Locations in AUTO Mode ..................................................................... 4-7
Zoomed Data View ............................................................................................ 4-8
Viewing Two Columns of Data ........................................................... 4-10
CONFIGURATION Screen ......................................................................................... 4-13
DATA SETUP ............................................................................................................. 4-14
Particle Channels ............................................................................................. 4-14
Analog Channels .............................................................................................. 4-15
SAMPLE .......................................................................................................... 4-18
SETTINGS ....................................................................................................... 4-21
COUNT MODE ................................................................................... 4-21
Geiger Counter Mode .......................................................................... 4-22
PARTICLE Display ............................................................................. 4-23
ALARM ........................................................................................................... 4-23
Alarm Threshold .................................................................................. 4-24
Clear Buffer ..................................................................................................... 4-26
DEVICE SETUP .......................................................................................................... 4-26
CLOCK ............................................................................................................ 4-27
OPTIONS ......................................................................................................... 4-30
CONTRAST ADJUST ........................................................................ 4-30
AUDIBLE BEEP ADJUST ................................................................. 4-30
ALIGN TOUCH SCREEN .................................................................. 4-30
Autostart Mode .................................................................................... 4-33
DiffCuml on Zoom .............................................................................. 4-33
One Channel ........................................................................................ 4-33
ONE SECOND OUTPUT .................................................................. 4-34
Pump Startup ........................................................................................ 4-35
LANGUAGE ....................................................................................... 4-35
COMM ADDRESS .......................................................................................... 4-36
OUTPUT SETUP ............................................................................................ 4-37
SECURITY ...................................................................................................... 4-38
POWER ON PASSWORD .................................................................. 4-39
CONFIGURATION PASSWORD ...................................................... 4-39
SERVICE ......................................................................................................... 4-40
STATUS ...................................................................................................................... 4-40
RECIPE ........................................................................................................................ 4-41
LOCATION ................................................................................................................. 4-44
Data View Buffer Screen ............................................................................................. 4-47
Analog Data ......................................................................................... 4-49
Printing Data View Buffer Report ....................................................................................49
Printing Range ..................................................................................... 4-51
REPORTS .................................................................................................................... 4-53
Setting up Counter to Run Reports .................................................................. 4-62
Report Output Setup ........................................................................................ 4-63
Printing a Report .............................................................................................. 4-63
t-ii
248083400-1 Rev 7
Report Requirements ....................................................................................... 4-64
Fed Std ft3 ............................................................................................ 4-64
ISO 14644-1 ......................................................................................... 4-64
EU GMP 2009 ..................................................................................... 4-64
Sample Printouts of Standard Reports ................................................. 4-65
Power Shutdown Levels .............................................................................................. 4-68
Chapter 5 Setting up the Counter
Using Report Parameters ............................................................................................... 5-1
Basic Requirements ....................................................................................................... 5-1
Fed Std 209E (feet) ............................................................................................ 5-1
ISO 14644-1 ....................................................................................................... 5-2
EU GMP 2009 ................................................................................................... 5-2
Chapter 6 Maintenance Procedures
Safety ............................................................................................................................. 6-1
Calibration ..................................................................................................................... 6-1
Cleaning ......................................................................................................................... 6-1
Purge Test ...................................................................................................................... 6-1
Table of Contents
Appendix A Limited Warranty
Limitation Of Warranties: ............................................................................................. A-1
Warranty Of Repairs After Initial Two (2) Year Warranty: ......................................... A-1
Appendix B HANDHELD Default Settings
Appendix C HANDHELD MODBUS Register Map v1.48
COMM Settings ............................................................................................................. C-1
Supported MODBUS Commands .................................................................................. C-1
Register Map .................................................................................................................. C-2
Sensor Settings Registers ................................................................................... C-2
Device Status ..................................................................................................... C-5
Command Register ........................................................................................................ C-6
Data and Alarm Registers .............................................................................................. C-8
Data and Alarm Enable Registers ...................................................................... C-8
Enable Alarming for a Channel ......................................................................... C-9
Threshold Setup Registers ............................................................................... C-10
Setting the Alarm Threshold Value ................................................................. C-11
Data Registers .............................................................................................................. C-12
Device Status Word (30007 - 30008) .............................................................. C-14
248083400-1 Rev 7
t-iii
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Valid Data in Channels (30073 - 30076) ......................................................... C-15
Data Type Registers ......................................................................................... C-15
Data Units Registers ........................................................................................ C-17
Index
t-iv
248083400-1 Rev 7
00
About this Manual
This manual describes the detailed operation and use of the Lighthouse HANDHELD 2016, 3016 and 5016 Airborne Particle Counters.
Text Conventions
Note:
the sidebar to give extra information regarding a feature or suggestion.
WARNING:
warning appears in a paragraph like this and warns that doing something incorrectly could result in personal injury, damage to the instrument or loss of data.
A note appears in
A
The following typefaces have the following meanings:
italics Represents information not to be typed
or interpreted literally. For example, file represents a file name. Manual titles are also displayed in italics.
boldface Introduces or emphasizes a term.
Courier font Indicates command syntax or text
displayed by the diagnostic terminal.
Bold Courier Indicates commands and information that
you type. You can use uppercase or lowercase letters; in this manual, commands are shown in uppercase.
Helvetica Italics Indicates a comment on a command or
text output.
Hexadecimal numbers are shown with the word “hex” or with a small “h” following the digits. For example:
hex 0D 0Dh
Additional Help
248083400-1 Rev 7
For more information about the Lighthouse HANDHELD 2016, 3016, 5016 Airborne Particle Counter, contact Lighthouse Worldwide Solutions:
Service and Support Tel: 800-945-5905 (USA Toll Free) Tel: 541-770-5905 (Outside of USA) techsupport@golighthouse.com
i
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
ii
248083400-1 Rev 7
00
1 General Safety
General Safety
Laser Safety Information
WARNING:
of controls, adjustments, or performance of procedures other than those specified within this manual may result in exposure to invisible (infrared) radiation that can quickly cause blindness. As a general precaution, avoid any possible exposure to laser radiation by honoring manufacturer seals and warranty stickers .
The use
Warnings and cautions are used throughout this manual. Familiarize yourself with the meaning of a warning before operating the particle counter. All warnings will appear in the left margin of the page next to the subject or step to which it applies. Pay close attention to each warning message. Take extreme care when performing any procedure preceded by or containing a warning.
There are several classifications of Warnings directed as follows:
Laser - pertaining to exposure to visible or invisible laser radiation
Electrostatic - pertaining to electrostatic discharge
This product contains a laser-based sensor that is a Class 1 product (as defined by 21 CFR, Subchapter J of the Health and Safety Act of 1968) when used under normal operation and maintenance. Service procedures on the sensor can result in exposure to invisible radiation. Service should be performed only by factory-authorized personnel.
The particle counter has been evaluated and tested in accordance with EN 610109-1:1993, “Safety Requirements For Electrical Equipment for Measurement, Control, and Laboratory Use” and IEC 825-1:1993, “Safety of Laser Products”. See Figure 1-1.
Figure 1-1 Example of Laser Warning Label
For further technical assistance, contact Lighthouse at 800-945-5905 (USA Toll Free), 541-770-5905 (Outside of USA).
248083400-1 Rev 7
1-1
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
1-2
248083400-1 Rev 7
00
2 Introduction
Overview This operating guide describes how to use the Lighthouse
HANDHELD 2016, 3016 and 5016 Airborne Particle Counters.
The HANDHELD 2016 has up to six particle-size channels starting at
0.2 microns with a flow of 0.1 CFM and a touch screen interface. The HANDHELD 3016 has up to six particle-size channels starting at 0.3 microns and the HANDHELD 5016 has up to six channels starting at
0.5 microns. They are otherwise identical to the 2016 model. A microprocessor controls all instrument functions. Count data is displayed as cumulative or differential count.
The model number signifies the minimum particle size measured by the instrument. The number “2016” indicates a 0.2μm minimum channel size at 0.1 CFM with up to 6 channels. The number “3016” indicates a
0.3 μm minimum channel size at 0.1 CFM with up to 6 channels. The number “5016” represents a 0.5μm minimum channel size at 0.1 CFM with up to six channels.
The instrument uses a laser-diode light source and collection optics for particle detection. Particles scatter light from the laser diode. The collection optics collect and focus the light onto a photo diode that converts the bursts of light into electrical pulses. The pulse height is a measure of particle size. Pulses are counted and their amplitude is measured for particle sizing. Results are displayed as particle counts in the specified size channel.
248083400-1 Rev 7
2-1
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Description Ergonomically designed and lightweight, the Lighthouse HANDHELD
2016, 3016 and 5016 particle counters are the newest and most advanced handheld particle counters on the market. See Figure 2-1.
Figure 2-1 Handheld 3016
The HANDHELD 2016/3016/5016 displays cumulative or differential particle count data as well as Temperature/Relative Humidity data on its easy to read 3.8" (9.25cm) touch screen. A rechargeable battery maximizes the HANDHELD’s uptime. Data is easily downloaded using the LMS XChange software.
The HANDHELD allows you to:
Set the Sample Time
Configure the number of samples taken in a given Location
Sample many different locations
Print reports (with optional, external printer) based on Federal Standard 209E (ft
3
), ISO 146441-1, and EU GMP 2009 (if unit has
0.5 and 5.0 channels).
Save your data for historical data review
Print data tables using the data transfer software included with your
2-2
instrument. Additionally use LMS Express to print graphs, data tables, and standards reports.
248083400-1 Rev 7
HANDHELD Specifications
Introduction
Size Range 0.2 - 10.0 μm
HANDHELD 2016 Channel Thresholds
HANDHELD 3016 Channel Thresholds
HANDHELD 5016 Channel Thresholds
Flow Rate 0.1 CFM (2.83 LPM)
Counting Efficiency 50% (per ISO 21501-4)
Laser Source Laser diode
Zero Count Level <1 count/5 minutes (per ISO 21501-4)
Calibration NIST Traceable
Count Modes Concentration, manual/automatic, beep,
Data Storage Up to 3000 sample records, includes particle
0.2, 0.3, 0.5, 0.7, 1.0, 2.0 μm
0.3, 0.5, 0.7, 1.0, 2.5, 5.0 μm
0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm
Other sizing available for each model; specify at time of order
cumulative/differential
& environmental data, plus location and time
Communication Modes RS232 or RS485 via RJ-45 to PC or printer
Supporting Software LMS XChange Data Transfer Software, LMS
Express/RT/RT Plus; LMSNet
Environmental Sensors Temperature/Relative Humidity Probe:
0-150°F (-17.8 to 65.6°C) ±1.8°F @ 77°F, 0­100% ± 5% @ 33%
Touch Screen Display 3.8" (9.25 cm), 320x240
Printer External thermal printer (optional)
Reports FS-209E (ft) , FS-209E (m), ISO 14644-1, EU
GMP (0.5 and 5.0μ channels must be active)
Key Software Features Historical data review, password protection
Enclosure High impact injection molded plastic
Sample Output Internally filtered to HEPA standards
(>99.97% @ 0.3μm)
Vacuum Source Internal pump, flow controlled
Table 2-1 Specifications
248083400-1 Rev 7
2-3
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Power Unit: +12VDC; AC/DC Adapter: 100-240V,
50-60Hz
Battery Li-Ion, removable and rechargeable
Dimensions 8.75"(L) x 5.0"(W) x 2.5"(H) [22.23 x 12.7 x
6.35 cm]
Weight 2.2 lb (1kg)
Operating Temp/RH 50°F to 104°F (10°C to 40°C) / 20% to 95%
non-condensing
Storage Temp/RH 14°F to 122°F (-10°C to 50°C) / Up to 98%
non-condensing
Table 2-1 Specifications
The manufacturer recommends that your Lighthouse instrument be calibrated annually by a Certified Lighthouse Service Provider to ensure that it continues to perform within specification.
2-4
248083400-1 Rev 7
00
3 Unpacking, Inspecting
and Installation
Initial Inspection
The instrument is thoroughly inspected and tested at the factory and is ready for use upon receipt.
Unpacking It is presumed that when the shipment was received, the following took
place:
1. The shipping container was inspected for damage.
2. If the container was damaged, the shipper was notified
immediately.
3. The instrument was carefully inspected for broken parts,
scratches, dents and other damage before use, container appeared undamaged, and
4. Any damages were reported to Lighthouse Technical Support at
800-945-5905 (USA Toll Free) or 541-770-5905 (Outside of USA) before proceeding.
Verify the contents of the package against the Shipping List. If anything appears to be missing, please contact your sales representative at Lighthouse Worldwide Solutions immediately at 510-438-0500 (Sales).
even if the
To maintain your warranty, keep the undamaged shipping container and all packing material for reshipment of the instrument for annual calibration. Order replacement containers and packing materials from Lighthouse directly, or from a Lighthouse-authorized distributor.
248083400-1 Rev 7
3-1
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Shipping Instructions
WARNING:
instrument is damaged during a return shipment due to inadequate user packing, the warranty may be voided and may result in additional repairs being billed to the customer.
If the
Should it become necessary to return the unit to the factory for any reason, contact Lighthouse Customer Service or visit our website, www.golighthouse.com/rma, and obtain a Return Merchandise Authorization (RMA) number. Reference this number on all shipping documentation and purchase orders. After receipt of the RMA number, follow the shipping instructions below:
1. Use the original container, nozzle caps and packing materials
whenever possible. Remove any instrument battery and package it to ship separately - refer to www.golighthouse.com/rma for detailed instructions. Remove attachments, such as TRH or Iso­kinetic probes, and package to prevent physical and ESD damage.
2. If the original container and packing materials are not available, wrap the unit in “bubble pack”, surround with shock-absorbent material and place in a double-wall carton - the instrument should not rattle around when the carton is vigorously shaken. If the instrument is damaged during shipment due to inadequate user packing, the warranty may be voided and may result in additional repairs being billed to customer. You may contact Lighthouse to purchase a replacement shipping container and nozzle caps.
WARNING:
ship the instrument with the battery installed.
Do not
3. Seal container or carton securely. Mark “FRAGILE” and write the Return Merchandise Authorization (RMA) number on any unmarked corner.
4. Return the instrument to the address provided by your Lighthouse representative or the RMA website.
Accessories You may order several optional accessories to tailor the unit to your
needs.
External Battery Charger with AC and car adapters
Spare Li-Ion Battery (removable and rechargeable)
Carrying Case
Isokinetic Sample Probe
6 ft. Tubing (for extending Isokinetic sampling input)
Thermal Printer with cable (AC or battery operated)
Validation Documentation
3-2
248083400-1 Rev 7
Unpacking, Inspecting and Installation
LMS Express software (standard), an analysis tool that allows the user to:
1. Manually download data from the instrument
2. Save data for historical review
3. Have advanced reporting with standard reports and much more.
LMS Express RT software (optional), an analysis tool that allows the user to perform the following:
1. Download data from the instrument
2. Collect data real time
3. Save data for historical review
4. Have advanced reporting with standard reports
...and much more.
Please contact Lighthouse Worldwide Solutions at 800-945-5905 (USA Toll Free) or 541-770-5905 (Outside of USA) for details.
Prepare Instrument for Use
Power Requirements
The power adapter input requirement is 100-240VAC, 50-60Hz, 1.25 Amps. Its output is +12VDC, 3A. A power cord and power adapter are included with your HANDHELD 2016, 3016 or 5016.
To protect the instrument from voltage spikes, Lighthouse recommends using protected power . Using an uninterruptible power supply (UPS) when the HANDHELD is kept in a stationary location will help prevent damage to the instrument or loss of data in the event of a power outage.
Install the Battery
The HANDHELD comes with a standard rechargeable battery. An optional external battery charger is available; otherwise, the battery recharges in the unit when the unit is plugged into AC power.
Install the battery as illustrated in the following instructions.
1. Make sure that the power switch is in the OFF position and the
instrument’s power adapter is not connected.
248083400-1 Rev 7
3-3
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
2. Open the battery compartment by sliding the battery compartment
door on the bottom of the unit to the left to unlock it. When the door is unlocked, open the door. The hinge is on the right side of the door. See Figure 3-1.
Figure 3-1 Open the Battery Compartment Door
3. Pull the battery release ribbon out of the battery compartment as
shown in Figure 3-2.
3-4
Figure 3-2 Pull Battery Release Ribbon Out
4. Hold the battery connector side toward the HANDHELD. Place
the battery into the compartment, on top of the release ribbon. Push the battery fully into the compartment until it connects and is entirely inside the battery compartment. See Figure 3-3.
248083400-1 Rev 7
Unpacking, Inspecting and Installation
Figure 3-3 Insert Battery into Battery Compartment
5. Tuck the battery release ribbon over the battery. See Figure 3-4.
Figure 3-4 Tuck Battery Release Ribbon
6. Close the door so that it is flush with the bottom of the
HANDHELD as shown in Figure 3-5.
Figure 3-5 Close the Battery Compartment Door
7. Slide the battery compartment door to the right until it clicks and
latches. See Figure 3-6.
248083400-1 Rev 7
3-5
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Figure 3-6 Secure the Battery Compartment Door
Connect Power
Connect the power adapter as illustrated in the following steps:
1. Insert the round DC power plug into its receptacle on the left side
of the HANDHELD. See Figure 3-7.
3-6
Figure 3-7 DC Power Plug
2. Push the connector in until fully seated as shown in Figure 3-8.
Figure 3-8 Inserting DC Power Plug.
248083400-1 Rev 7
Unpacking, Inspecting and Installation
3. Plug the adapter’s AC power cord into the AC power source. The
battery will begin and continue to charge until it is fully charged or DC power is removed.
WARNING:
ship the instrument with the battery installed.
Do not
Battery Removal
1. To remove the battery, open the door to the battery compartment
by sliding the door to the left to unlatch it. See Figure 3-9.
Figure 3-9 Unlatch Battery Compartment Door
248083400-1 Rev 7
3-7
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
2. Swing the battery compartment door open as shown in Figure 3-
10.
Figure 3-10 Open Battery Compartment Door
Note: Remove the
battery if the instrument will be stored for a month or longer. Leaving the battery in during storage will drain the battery to the point where it will not recharge.
3. Hold the instrument securely then pull the battery release ribbon
outward to disconnect the battery and slide the battery out. Remove the battery. See Figure 3-11.
Figure 3-11 Remove Battery
3-8
248083400-1 Rev 7
Unpacking, Inspecting and Installation
Optional Printer Accessory
An optional external thermal printer can be used with the HANDHELD to print reports directly from the instrument.
Connect the external printer as illustrated in the printer Read Me First, document #248083378-1, which is included on the Operating Manual CD.
Figure 3-12 HANDHELD External Printer
248083400-1 Rev 7
3-9
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Connecting to an External Computer or Facility Management System
The HANDHELD has the ability to be connected to the Lighthouse Monitoring System (LMS), LMS Express or LMS XChange Data Transfer Software to download its data.
By transferring the instrument’s data to the Lighthouse Monitoring System (LMS) or a PC running LMS Express, historical data can be stored externally for future reviewing and trending. LMS XChange software can export data to a *.csv file for historical review.
Note:
can be used for the printer or to connect HANDHELD to an external computer. It cannot be used to connect to both at the same time.
The DATA port
Note: LMS Express
allows you to upload data from your HANDHELD; LMS Express RT allows you to also collect data in real time.
1. To connect to the HANDHELD, insert the RJ45 end of the
HANDHELD data communications cable to the DATA port as shown in Figure 3-13.
Figure 3-13 Connecting RJ45 Data Communications Cable
2. Connect the other end of the data communication cable to your
Lighthouse Monitoring System or PC running LMS Express or LMS XChange software.
Please refer to the LMS XChange, LMS Express or LMSNet manual for additional information.
3-10
248083400-1 Rev 7
00
4 Operation
This chapter describes how to use the HANDHELD 2016, 3016 and 5016 Airborne Particle Counters.
Using the Instrument
WARNING:
attempt to sample reactive gases (such as hydrogen or oxygen) with this instrument. Reactive gases create an explosion hazard in the instrument.
Sampling any gas under pressure can damage the instrument and void the warranty.
Sampling any gas that is not the same density as ambient air can result in inaccurate data.
Contact Lighthouse for more information.
Do not
The HANDHELD comes with a charged battery and is ready for use. To start using the instrument, proceed as follows:
1. Insert the battery included with the shipment. (See Chapter 2 for
instructions.)
2. Position the instrument in the environment to be measured.
3. Remove the protective cap from the inlet tube. To use the
provided isokinetic probe, install it by connecting to inlet tube on the top of the instrument. NOTE: Do not discard the protective
cap. It should be placed on the inlet tube any time the instrument is moved outside the environment being measured.
4. The included Temperature/Relative Humidity probe can be
attached to the provided receptacle to read environmental data.
5. Set on/off switch found on the left side of the unit to ON.
6. The Start Up screen displays on the LCD.
WARNING:
allow water, solvents, or other liquids enter the instrument via the inlet tube - the instrument will be damaged and void the warranty.
Do NOT operate the instrument with the inlet tube capped or plugged ­the internal pump will be damaged and void the warranty.
248083400-1 Rev 7
Do NOT
7. The MAIN screen appears.
8. On the touch screen, press the START button to start the
9. “STARTING” will display when the pump is initially turned on.
instrument.
4-1
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
10. When the HANDHELD starts counting, “COUNTING” appears
on the display. Particle counts are displayed according to the size of each particle.
Note:
requires a minimum of 5 seconds to ramp up to full flow. Refer to “Pump Startup” on page 35.
The pump motor
Touch Screen Overview
11. If the instrument is in AUTO mode with cycles and a hold time,
“HOLDING” will display after each cycle and “FINISHED” will display when all the cycles are complete.
12. Press the “STOP” button to stop the instrument before the cycles
are complete.
The HANDHELD incorporates a unique touch screen interface to control and configure the instrument.
This interface allows the user to easily view and configure the instrument to specific needs and applications. See Figure 4-1.
Menu Map
4-2
Figure 4-1 Menu Map
248083400-1 Rev 7
MAIN Screen
Operation
Note: The Screen Shots
shown in this chapter were taken from the 3016 model. The 2016 and 5016 screens are identical except for the channel sizes.
The MAIN screen gives the user a single snapshot view of the status of the instrument. The instrument can be powered by a power supply or from a removable battery. When a battery is used, the battery indicator will show the level of the battery charge remaining. See Figure 4-2 & Figure 4-3.
Figure 4-2 MAIN Screen - Battery Operation
When the AC/DC indicator symbol is displayed, it indicates that the instrument is getting its power from an AC source.
248083400-1 Rev 7
Figure 4-3 MAIN Screen - AC Operation
4-3
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
The MAIN screen displays the following options and information.
LOCATION: Displays the location that is currently being measured. Up to 200 alphanumeric locations can be configured.
LOCATION SELECT button: Allows user to change location before sampling.
RECIPE button: Allows the user to view, load and unload recipes that have been configured and saved in the recipe data base.
PRINT LAST RECORD: Prints the last recorded sample using the current configuration to determine the type of data printed (i.e. cumulative vs. differential, raw vs. normalized, ft
The print configuration is set through the PRINT SETUP button in the Configuration screen. For more details about printing, see the Configuration section later in this chapter.
3
vs. m3).
Note:
in the battery indicator while the instrument is counting, the pump will stop automatically to prevent the battery from discharging completely.
If the “X” appears
Date/Time: Displays the current date and time.
Battery Indicator: Indicates that the instrument is being powered by a rechargeable battery. The amount of battery life left is denoted by the fill inside the battery icon. When the battery is low, the words “BATT LOW!” will appear on the screen and the unit will beep continually until it is plugged into its AC power cord for recharging. See Figure 4-4.
Figure 4-4 Battery Indicator Levels
AC Indicator: Indicates that the instrument is being powered with its AC power cord. If the battery is installed, the HANDHELD will charge the battery when the instrument is plugged in.
Flow Status: When the instrument is sampling, the Flow Indicator will display sufficient or insufficient flow.
WARNING: If the Air Flow is insufficient, turn the instrument off and
contact Lighthouse Worldwide Solutions tech support.
4-4
248083400-1 Rev 7
Operation
Service Indicator: Indicates that the instrument may be in need of service. If wrench displays, please contact your authorized Lighthouse Service Provider for assistance or send an e-mail to techsupport@golighthouse.com.
μ: The mu symbol indicates the particle sizes, in micrometers, configured for the instrument. These sizes are pre-configured at the factory. To order an instrument with different sizes, please contact a sales representative at Lighthouse ((800) 945-5905 Toll Free).
DATA DISPLAY: This indicates whether the counts are being displayed in Differential (Diff) mode or in Cumulative (Cuml) mode and when the counts are normalized to ft
3
or m3.
Analog Data: Gives a snapshot view of the enabled analog channels. (Analog channels are enabled by default.)
•MODE: Displays the current mode selected; possible modes are AUTO, MANUAL, CONCEN (Concentration) and BEEP.
Note:
greater than 1 minute, the pump will stop during that time. At the end of the hold time, the pump will restart.
If Hold time is
•CYCLES: Indicates the number of times that the count will be taken at a given location in Auto mode. “1/ 3" indicates that the count is the first of three samples to be recorded at this location. The maximum number of cycles is 999. When set to 0, the unit will run in Auto mode continuously until the STOP button is pressed.
SAMPLE: The Sample Time (hh:mm:ss) is the duration of one counting cycle. The Sample Time will count down on the MAIN screen when the instrument is in AUTO or Manual mode so you can see how much time is remaining in the sample period. In Concentration mode, the Sample Time will count up to 6 seconds per cycle.
•HOLD: Displays the hold time in between cycles. The maximum hold time is 99 hours, 59 minutes, 59 seconds.
•RECS: This displays the current number of records stored in the instrument and the total number of records that can be stored. The data buffer is a circular buffer. The HANDHELD can store up to 3000 records. An asterisk (*) will appear in front of counts when the buffer wraps.
248083400-1 Rev 7
4-5
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Note: There is a one
second startup during which the pump is accelerating to full power. During this time, the word “STARTING” will display.
•START/STOP: Press START button on the screen to start counting. When running, the instrument will display “COUNTING” in the lower right portion of the screen as shown in Figure 4-5.
Figure 4-5 Main Screen, Counting Mode
Press the STOP button to stop counting; the word “STOPPED” will display.
4-6
248083400-1 Rev 7
Operation
LOCATION Selection
Changing Locations
The location number for the environment to be measured can be changed by pressing the LOCATION button at the top of the MAIN screen. The Select Location screen opens as shown in Figure 4-6.
Figure 4-6 Location Select Screen
The blue highlight indicates which location is currently selected.
Use the UP and DOWN arrows to select a location. The single arrows will move the cursor up and down by a single line. Locations can also be selected by touching the location name on the screen.
Recipes associated with a specific location may also be loaded when that location is selected.
The AUTO INC button, when activated will allow the user to advance to the next location once the current cycle is completed.
Press the MAIN button to return to the MAIN screen. Whichever location is currently selected will be the location displayed on the MAIN screen.
Locations in AUTO Mode
When the instrument is in Automatic Mode and the START button is pressed, the instrument will start counting particles automatically according to the SAMPLE time, HOLD time and number of cycles that are configured.
248083400-1 Rev 7
4-7
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Zoomed Data View
Press anywhere in the Particle Data area to display the Zoomed view. See Figure 4-7.
4-8
Figure 4-7 Zoomed Data View
248083400-1 Rev 7
Operation
In Figure 4-8, the following functions can be performed using the toggle buttons on the right side bar:
START or STOP counting
Display Particle data or Analog data
Display Cumulative / Differential data
Display Raw / Normalized particle data
3
Display ft
/ m3 when configured
PRINT the last record
Figure 4-8 Zoomed Data View Buttons
When the instrument is STOPPED or HOLDING, press anywhere in the white data area to return back to the MAIN screen.
248083400-1 Rev 7
4-9
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Viewing Two Columns of Data
The unit can display both Differential and Cumulative data at the same time on the Zoomed View screen. To enable this feature, go to the CONFIG: Options screen as shown in Figure 4-9.
Figure 4-9 CONFIG: Options Button
Figure 4-10 CONFIG: Options Screen
4-10
248083400-1 Rev 7
Operation
START/STOP counting
Display Particle data or Analog data
PRINT the last record
Display Raw / Normalized particle data
Display ft
3
/ m3 when configured
Select the button “DiffCuml on Zoom” as shown in Figure 4-10; this will display both differential and cumulative data on the Zoomed View screen as shown in Figure 4-11.
Figure 4-11 Differential and Cumulative Data in Zoomed View
The buttons available when touching the data area of the screen after selecting DiffCuml on Zoom are shown in Figure 4-12. Note the selections are different but their functions are the same as on dual display screen.
248083400-1 Rev 7
Figure 4-12 Diff / Cuml Screen Button Actions
4-11
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
To return the instrument to the default single-column zoom view, return to the MAIN Screen view, press CONFIG, Options and deselect the button “DiffCuml on Zoom” button. See Figure 4-13.
Figure 4-13 DiffCuml on Zoom Disabled
Select the MAIN button to return to the MAIN Screen view or press BACK to return to the previous level. Here, the user can choose what data type is to be displayed in one column on the Zoomed View screen.
Go to the CONFIG: Settings screen and select Differential or Cumulative for the data type. SeeFigure 4-14 .
4-12
Figure 4-14 Configuration: Setting Screen
248083400-1 Rev 7
Operation
Press the MAIN button, then tap in the data area to go into the Zoomed View. See Figure 4-15.
Figure 4-15 Zoomed View with Only Cumulative Data
Only one column of data is displayed.
CONFIGURATION Screen
Press CONFIG on the MAIN screen to display the Configuration screen as shown inFigure 4-16 .
Figure 4-16 Configuration Screen
Data Setup includes buttons to enable/disable particle channels, configure analog temperature readings, set sample record parameters, sample settings, thresholds, enable/disable alarms, and clear the data buffer.
Device Setup includes buttons to configure the instrument’s date and time, set the LCD contrast, adjust the instrument’s beep volume, enable the instrument to AutoStart, set it to display only one channel of data, enable 1 Second Output, set the instrument’s communication address, enabling print functions, enable password restrictions, and/or (with proper authorization) adjust service settings.
248083400-1 Rev 7
4-13
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
DATA SETUP Figure 4-17 displays the Data Setup option buttons. These options
allow the user to configure the parameters for data collection during a sample.
Figure 4-17 Data Setup Options
Particle Channels
The instrument’s particle channels can be enabled or disabled on the Particle Channel. A checkmark is displayed next to each enabled channel as shown in Figure 4-18.
Note: The data buffer is
cleared when channels are enabled or disabled. This ensures that all of the data in the buffer will have the same number of channels.
4-14
The checkmark button next to each channel size is a toggle. Press it
Figure 4-18 Particle Channel Configuration
once to disable a channel and a red “X” will display, indicating that the channel is disabled. Press the button again and it will change back to the checkmark, enabling the channel.
248083400-1 Rev 7
Operation
Pressing MAIN or BACK will prompt to clear all collected data as shown in Figure 4-19.
Figure 4-19 Clear Data Screen
From this screen, press CANCEL to cancel changes and return to the Configuration screen.
When channels are disabled, they are removed from the MAIN screen display, from the reports, and from the printouts.
The channel size for disabled channel(s) remain in the Data screen, however the instrument does not record data for disabled channels.
Analog Channels
A Temperature and Relative Humidity probe can be connected to the HANDHELD via a connector on the top of the unit.
Users can select if the HANDHELD displays analog temperature data as Fahrenheit or Celsius by pressing the ANALOG button on the Configuration screen. The Select Temperature Units window displays.
There are two buttons on the Select Temperature Units window. The button displayed shows the units that are currently selected: F for Fahrenheit and C for Celsius.
248083400-1 Rev 7
4-15
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
To change to Celsius, press the C button. To change to Fahrenheit, press the F button. See Figure 4-20 & Figure 4-21.
Figure 4-20 Analog Channels - Select Celsius Units
4-16
Figure 4-21 Analog Channel - Select Fahrenheit Units
248083400-1 Rev 7
Operation
Temperature and relative humidity values appear on the MAIN screen with the units you selected as shown in Figure 4-22.
Figure 4-22 Analog Data on MAIN Screen
When the unit is not counting or when it is holding, you can display the instrument’s current Analog data by pressing press anywhere in the Particle Data area to bring up the Zoomed Data View, then press the Analog toggle button to display Analog data. See Figure 4-23.
Figure 4-23 Zoomed Data View: Analog Data
248083400-1 Rev 7
4-17
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
SAMPLE
Configure the Sample Time and the number of samples to be collected on the Sample screen as shown in Figure 4-24.
Note:
immediately when START is pressed. If a Start Delay is required, the Delay should not be set to less than 5 seconds or readings may be unreliable.
The pump starts
Figure 4-24 Sample Timing Configuration Screen
•CYCLES: The number cycles is set to determine how many times the instrument samples the air in a single location. This is used only in AUTO mode. The range is 0 - 999. When Cycles is set to 0, the instrument will continue running samples indefinitely until the STOP button is pressed.
Select the CYCLES button; enter the number of desired cycles using the numeric keypad on the right. Press ERASE to erase a number, if needed. Press ENTER to set the Cycles.
DE LAY: The Initial Start Delay (hh:mm:ss) is the time between pressing the START button and the unit actually starting counting.
The Initial Start Delay gives the operator time to exit the area under test so that the measurement is taken under a controlled condition. The maximum delay time is 99 hours, 59 minutes and 59 seconds.
Select the DELAY button; enter the initial delay time in hours, minutes and seconds using the numeric keypad on the right.
4-18
After the value is entered, press ENTER.
248083400-1 Rev 7
Operation
Note: If the Hold time is
set to 00:00:00 in Auto Mode, the instrument will run the samples according to the sample time and the # of cycles, but with no hold time in between cycles.
Note:
greater than one minute, the pump will shut off during the specified hold time.
Note:
Sample Time is 23:59:59.
If the Hold time is
The maximum
The maximum hold time is 99 hours, 59 minutes and 59 seconds. This field will count down to indicate how much time is left for the Hold period.
Select the HOLD button; enter the time in hours, minutes and seconds using the numeric keypad on the right. Press ERASE to erase a number, if needed. Press ENTER to set the Hold Time.
HOLD: The Hold Time (hh:mm:ss) is the time between count cycles when the instrument is not counting particles.
SAMPLE: The Sample Time (hh:mm:ss) is the duration of one counting cycle. The Sample Time will count down on the MAIN screen when the instrument is in Auto or Manual mode to indicate how much time is remaining in the Sample.
Select the SAMPLE button; enter the time in hours, minutes and seconds using the numeric keypad on the right. Press ERASE to erase a number, if needed. Press ENTER to set the Sample time.
Note:
volume is ft sample volume is 0.1 ft
When the particle
3
the minimum
SAMPLE VOLUME: Instead of selecting a specific Sample Time,
3
.
the instrument can be set to measure a specific Sample Volume in cubic feet (ft
3
), cubic meters (m3) or liters (l). When this is set, the corresponding Sample Time will automatically be set. See Figure 4-25.
248083400-1 Rev 7
4-19
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Note: If the particle
volume is liters or m sample volume is displayed in liters. If the particle volume is cubic feet (ft volume is displayed in cubic feet (ft
3
3
3
), the sample
).
,
4-20
Figure 4-25 Changing Sample Volume unit of measure
Press BACK to return to the Configuration screen or press MAIN to return to the MAIN screen.
248083400-1 Rev 7
Operation
SETTINGS
The instrument can be configured to count in different modes and formats. See Figure 4-26.
Note:
for Auto, Manual, and Beep modes count down and the sample times for Concentration mode count up.
The sample time
Figure 4-26 Sample Settings Screen
COUNT MODE
The following modes are available: Auto, Manual, Beep and Concentration.
AUTO - When the instrument is in Automatic Mode and the START button is pressed, the instrument will start counting particles automatically according to the Sample Time, Hold Time and the number of Cycles that are configured. If Cycles are set to 0, the instrument will continue indefinitely in Auto Mode until the STOP button is pressed.
MANU (Manual Mode) - When the instrument is in Manual Mode, it will start counting when START is pressed and stop at the end of one programmed Sample Time.
CONC (Concentration Mode) - When the instrument is in Concentration mode, it gives a calculated value of the concentration of particles in a volume of air, measured and displayed on the MAIN screen in either counts per cubic foot or per cubic meter.
Counting starts when the START button is pressed and it will continue until the STOP button is pressed. The sample time for Concentration mode is six seconds. As the sample time on the MAIN screen counts from one to six, the particle counts are updated continuously.
248083400-1 Rev 7
Concentration data will be recorded and can be viewed in the Data screen and on the printouts.
4-21
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Note: BEEP mode only
works with CUMULATIVE data and for Sample Times greater or equal to 6 seconds.
Note:
works with CUMULATIVE data.
Note:
set to one count, the beeps may not be for every single count.
BEEP mode only
If BEEP mode is
BEEP - In this mode, the instrument is pre-configured to beep according to the alarm threshold set in Alarm Configuration and the instrument’s sample time if the instrument is set to collect cumulative data. Counting starts one second after the START button is pressed and will continue until the STOP button is pressed.
If no channel is set for alarming and BEEP mode is selected, the smallest channel size will be automatically selected and its alarm threshold will be used to trigger the alarm.
If alarming is enabled on more than one channel size, only the smallest channel size will be used to trigger the alarm in BEEP mode. Larger channel sizes will be ignored.
The data will be recorded based on the set sample time and hold time and can be viewed in the View Buffer and on the printouts. There will be no indication on the record, however, that the data was saved while the instrument was in BEEP mode.
Geiger Counter Mode
While in BEEP mode the instrument can be set to beep 1 to 4 beeps per second depending on the number of particles the instrument has counted per second.
To set up the instrument for Geiger Counter Mode:
1. Enable BEEP Mode.
2. Set the smallest particle channel to 1.
In Geiger Counter Mode the instrument will beep from 1 beep up to 4 beeps per second if it counts from 1 particle up to 4 particles per second. The maximum beeps per second the instrument can emit are 4 per second even if the instrument counts 100 particles per second.
The value set in the smallest particle channel’s alarm is the trigger for a single beep per second. If the smallest particle channel is set to 2, the instrument will beep once every second for every 2 particles the instrument counts per second.
4-22
248083400-1 Rev 7
Operation
PARTICLE Display
Data on the instrument can be displayed in Differential (DIFF) or Cumulative (CUML) counts.
For example, the cumulative count for the 1μm channel is the sum of that channel’s count + 2.5μm count + the 5μm count.
The differential count for the 1μm channel is the number of particles between 1μm and 2.5μm.
The data displays on the MAIN screen according to whichever is selected (DIFF or CUML).
The data format is either Raw (RAW) or Normalized (NORM). Raw data displays the actual number of particles counted. Normalized data shows particle concentrations calculated from the raw data based on the settings chosen (ft
3
or m3).
Volume of Air = Sample time (minutes) x FlowRate (CFM)
Normalized Data = Number of Particles/Volume of Air
Press BACK to return to the Configuration screen or press MAIN to return to the MAIN screen.
ALARM
The user can enable alarming on specific channels as illustrated in Figure 4-27.
Threshold Button
248083400-1 Rev 7
Figure 4-27 Particle Alarm Configuration, 2 Channels Enabled
4-23
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Note: Alarming is only
applicable for AUTO and MANUAL mode. It applies only to Raw particle counts even if the instrument is displaying Normalized data.
If the instrument is set to Differential data, the alarm threshold will apply to the differential counts.
If the instrument is set to display Cumulative data, the alarm threshold will apply to the cumulative counts.
To enable the alarming for any channel, press the “X” next to that channel. When a checkmark is displayed, that channel is enabled for alarming. Press the checkmark to disable the alarming for that channel.
Alarm Threshold
Press the threshold button next to the enabled channel in order to set the alarm threshold for that channel. See Figure 4-28.
Note: To receive alarms,
the Sample Time must be greater than 1 second.
Figure 4-28 Configure Alarm Threshold
Enter the desired alarm threshold for the selected particle channel in number of particles, then press ENTER. The threshold value will be updated on the Particle Alarm screen as shown in Figure 4-29.
4-24
248083400-1 Rev 7
Note:
triggered per sample record. At the end of the sample time, the alarms reset.
Alarms are
Operation
Press BACK to return to the Configuration screen or press MAIN to return to the MAIN screen.
Figure 4-29 2 channels Enabled for Alarming
Note:
enabled on two channels, if the user presses the Alarm Acknowledge button when the first channel goes into alarm, the alarm will not sound if the second channel’s threshold is reached within the sample period.
If alarms are
When a particle channel that is enabled for alarming goes into alarm, the selection cursor (>) and the channel size are highlighted in red as shown in Figure 4-30.
Figure 4-30 Channel in Alarm
When the instrument begins to beep in response to the Alarm settings, silence the beep by pressing anywhere on the MAIN screen. After acknowledging the alarm, the count will reset when the next sample cycle begins.
248083400-1 Rev 7
4-25
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Clear Buffer
Press the Clear Buffer button to clear the instrument’s data buffer. See Figure 4-31.
DEVICE SETUP
Figure 4-31 Clear Buffer Screen
Press OK to clear the data or press Cancel to exit screen without clearing the data.
The Device Setup buttons allow the user to set and adjust how the instrument operates when running samples. Specific functions can be enabled and disabled for display options, data output and security. See Figure 4-32.
Figure 4-32 Device Setup Options
4-26
248083400-1 Rev 7
Operation
CLOCK
Use the Clock screen to set the instrument’s date and time. See Figure 4-33.
Figure 4-33 Date & Time Configuration Screen
Set the instrument’s Date by entering values for the desired month, day and year and then press the ENTER button.
248083400-1 Rev 7
4-27
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Note:
MONTH 1ST (M / D/
Y) is the default date format.
Change the Date’s format by pressing the M / D/ Y button to display the date as month-first. Alternatively pressing the D / M/ Y button displays the Date as day-first and the Y / M/ D button displays the date year­first. See Figure 4-34 and Figure 4-35.
Figure 4-34 Date Option: Day First
Figure 4-35 Date Option: Year First
4-28
248083400-1 Rev 7
Set the instrument’s Time by pressing the TIME button. See Figure 4-36.
Figure 4-36 Configuring TIME
Operation
Enter the desired Time in hours, minutes and seconds then press ENTER to save the new time.
Press BACK to return to the Configuration screen or press MAIN to return to the MAIN screen.
248083400-1 Rev 7
4-29
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
OPTIONS
Several optional configuration settings are found on the OPTIONS screen. See Figure 4-37.
Figure 4-37 Options Configuration Screen
CONTRAST ADJUST
The contrast/brightness of the LCD screen can be adjusted by pressing the first set of UP and DOWN arrows.
AUDIBLE BEEP ADJUST
The audio level of the BEEP can be adjusted by pressing the second set of UP and DOWN arrows.
ALIGN TOUCH SCREEN
The ALIGN button allows you to re-align the touch screen so the locations that you touch on the screen correspond to the expected button or function.
Press the ALIGN button.
4-30
248083400-1 Rev 7
Operation
WARNING: Be
careful to touch the screen at the specified locations only. If touching the screen elsewhere during this process, the screen will be incorrectly aligned.
The screen in Figure 4-38 appears. Touch anywhere on the screen to continue calibration.
Figure 4-38 Alignment Screen 1
Note:
Using a PDA Stylus provides higher accuracy to the touch screen interface.
At the next screen, touch the arrow displayed in the upper right corner. This needs to be done a total of three times before continuing to the next alignment. See Figure 4-39.
Figure 4-39 Alignment Screen 2
248083400-1 Rev 7
4-31
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
The third screen displays the arrow in the upper left corner. The arrow will have to be touched three times to ensure calibration as Figure 4-40 illustrates this.
Figure 4-40 Alignment Screen 3
The fourth alignment screen shown in Figure 4-41, displays the arrow in the at the bottom center of the screen. Touch the arrow three times to finish the alignment process.
4-32
Figure 4-41 Final Alignment Screen
248083400-1 Rev 7
Operation
Note: The screen
alignment function can also be accessed by touching the screen when powering the instrument on.
Once the final alignment is complete, the following screen is displayed. See Figure 4-42. Touch anywhere in the screen to return to the OPTIONs screen.
Figure 4-42 Alignment Saved Screen
Autostart Mode
When Autostart mode is enabled, set the Delay time to at least 5 seconds.
If Autostart mode is enabled, when the instrument is powered on (or regains power after a power outage), the instrument will immediately begin sampling based its configured mode, delay, start, and hold times.
DiffCuml on Zoom
If DiffCuml on Zoom is enabled, the counter’s display will show two columns of data simultaneously - Differential and Cumulative.
One Channel
If One Channel is enabled, only the first channel will be displayed on the MAIN screen. Data will continue to be recorded, printed and downloaded for all channels.
248083400-1 Rev 7
4-33
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Note: If the first channel
is disabled and One Channel mode is enabled, the display and zoomed data view will be blank.
One Channel mode only affects how data is displayed on the MAIN screen. Refer to Figure 4-43.
Figure 4-43 Main Screen - One Channel Option Enabled
Note:
Output disables Alarming.
One Second
Note: Changing any
setting (mode, sample time, hold time, etc) disables the One Second Output.
ONE SECOND OUTPUT
If One Second Output is enabled, the instrument’s settings reset to the following.
Mode: AUTO
Cycles: Zero
Sample Time: One second
Hold Time: Zero seconds
Cumulative/Differential: Cumulative (CUML)
Raw/Normalized: Raw
When counting, the MAIN Screen will update continuously and data will not be recorded to the data buffer.
4-34
248083400-1 Rev 7
Operation
Pump Startup
This is ON by default and sets a three-second pump ramp up time at the beginning of the first sample, or at the beginning of all samples if the HOLD time is greater than one minute, to stabilize the laser and air flow. This setting should be left in the default mode unless special applications, such as “surface scans”, require it to be disabled.
LANGUAGE
Pressing the LANGUAGE button allows the user to change the operating language displaying the screen shown in Figure 4-44.
248083400-1 Rev 7
Figure 4-44 Operating Language Selection Screen
Press the desired language button then BACK or MAIN. The default is English.
4-35
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
COMM ADDRESS
When the HANDHELD is connected to a data collection system or daisy chained at the end of a chain of RS-485 instruments, the instrument’s COMM address is used to identify it.
LMS XChange or LMS Express/RT will search for the instrument by the COMM Address specified on the Communication screen shown below. COMM addresses range from 1 to 63.
For RS-485 communications, each device on a multi-port chain must have a unique address.
4-36
Figure 4-45 COMM Address Configuration screen
Set the COMM Address by using the numeric keypad to select the address; press ERASE to erase a number, if needed. Press ENTER to accept the value. See Figure 4-45.
Press BACK to return to the Configuration screen or press MAIN to return to the MAIN screen.
248083400-1 Rev 7
Operation
OUTPUT SETUP
The Print configuration has several options for printing the data that the user sets on the Printer Setup screen shown in Figure 4-46.
Note:
both Differential and Cumulative printing at the same time. For normalized values with more than 8 digits, only the whole number will be printed.
You can select
Figure 4-46 Printer Setup Screen
Output on Sample / Output on Alarm - When Output on Sample is enabled, a single record will print at the end of every
sample.
When Output on Alarm is enabled, a single record printout will print at the end of any sample that experiences an alarm condition.
Model Name - When enabled, the Model name of the instrument will print in the header of all printouts.
Serial Number - When enabled, the Serial Number of the instrument will print in the header of all printouts.
Separator - When enabled, a line separator will print after the Model Name and Serial Number in the header of all printouts.
Differential / Cumulative Toggle - This toggle option specifies how the data will appear on the printouts.
Press BACK to return to the Configuration screen or press MAIN to return to the MAIN screen.
248083400-1 Rev 7
4-37
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
SECURITY
To restrict access to the instrument and/or configuring the instrument, the HANDHELD has two different password levels. See Figure 4-47.
Figure 4-47 Security Password Configuration Screen
To restrict who can operate the instrument, enable the POWER ON password. When the POWER ON password is enabled, in order to use the instrument the user will be required to enter the correct password each time the instrument is turned on.
To restrict who can configure the instrument, enable the Configuration (CFG) password. When the Configuration password is enabled, the user will be required to enter the correct password before they can access the Configuration screen.
To set the password for the POWER ON, select the POWER ON button, then type in a password using the numeric keypad. Press the ERASE button to delete the last character, if needed.
Press ENTER to save.
To set the Configuration (CFG) password, select the CFG button, then type in a password using the numeric keypad.
Press the ERASE button to delete the last character, if needed.
Press ENTER to save the changes.
Press the “X” button to enable either or both passwords.
Press BACK to return to the Configuration screen or press MAIN to return to the MAIN screen.
4-38
248083400-1 Rev 7
POWER ON PASSWORD
Operation
WARNING:
record the unit’s passwords in a safe place. If the password is lost or forgotten, contact Lighthouse technical support for assistance. The unit may have to be returned to the factory to reset the password.
Be sure to
To require that a password must be entered before the instrument can be used, enable the POWER ON password. See Figure 4-48.
Figure 4-48 POWER ON Password Access Screen
When the POWER ON password is enabled, you will see a password access screen just after you turn the unit on. The instrument will remain locked until the correct password is entered.
CONFIGURATION PASSWORD
The Configuration password prevents unauthorized access to the Configuration screen. See Figure 4-49.
Figure 4-49 CONFIGURATION Password Access Screen
248083400-1 Rev 7
4-39
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
SERVICE
This section of the Configuration screen is reserved for Lighthouse Authorized Service Providers only. The correct service password must be entered to access this area.
Three options on the lower CFG screen, STATUS, RECIPE and LOC display the instrument’s current status, configure user recipes and locations respectfully.
STATUS Touching the STATUS button displays the instrument programmed
version of the various firmware modules as shown in Figure 4-50. This information is useful when contacting Lighthouse Technical Support personnel.
4-40
Figure 4-50 HANDHELD 3016 STATUS Screen
248083400-1 Rev 7
Operation
RECIPE Selecting the RECIPE button displays the Recipe setup screen. See
Figure 4-51.
Note:
selected is already assigned to another recipe or if there are no available free locations, the user will not be able to add a new recipe. The “Add” button will not be displayed.
If the location
Figure 4-51 Recipe Setup Screen
The Recipe feature allows the user to save instrument settings for sampling and reports in a database that can store up to 50 recipes.
Selecting the ADD button displays the RECIPE text screen. See Figure 4-52. The recipe can be named using up to 12 characters.
248083400-1 Rev 7
Figure 4-52 Recipe Name Screen
4-41
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Pressing the ENTER button will add the recipe to the database and display the recipe CONFIG screen as shown in Figure 4-53. Each option allows the user to configure the instrument to the current recipe.
Figure 4-53 Recipe Configuration
Press the BACK button to save the settings and return to the RECIPE screen.
Pressing the EDIT button on the main RECIPE screen will allow the user to change settings on the highlighted recipe.
The LOAD and UNLOAD buttons add or remove the highlighted recipe as the instrument’s current operating configuration.
The DELETE button will delete any highlighted recipes from the database.
4-42
248083400-1 Rev 7
Operation
The VIEW button displays the current settings for the highlighted recipe. See Figure 4-54.
Figure 4-54 Recipe - Channel Settings
Press the NEXT button to view the analog and output settings. See Figure 4-55.
248083400-1 Rev 7
Figure 4-55 Recipe - Analog & Output Settings
4-43
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
LOCATION The HANDHELD allows up to 200 different locations and associated
alphanumeric labels.
Press the LOC button on the Configuration screen to display the Select Location screen as shown in Figure 4-56.
Figure 4-56 LOCATION Setup Screen
On the Select Location screen, the Location name can be selected using the UP and DOWN arrows or by touching the location name on the screen.
The following options can be enabled to configure locations for the HANDHELD.
When enabled, the Auto Increment option will automatically select the next location after completing the programmed number of cycles. The default position is “off” when the unit is powered up.
The unit can be configured for a maximum of 200 Locations. Each location name can be configured with a maximum of 8 characters per name.
Recipes associated with a specific location are denoted by an asterisk (*) to the right of the location name.
4-44
248083400-1 Rev 7
Operation
When selecting a location with a recipe, the user is provided with the option of loading the recipe as shown in Figure 4-57. If the user chooses not to load the recipe, the location can still be used with the current settings.
Figure 4-57 Location - Load Recipe
If a recipe is loaded and a location is selected that is not associated with a recipe, the user is given the option to unload the current recipe. If the user chooses to keep the loaded recipe, the location will adopt the current recipe settings.
To create a name for a location, press the EDIT button to display the Edit screen. See Figure 4-58.
248083400-1 Rev 7
Figure 4-58 LOCATION Edit Screen
4-45
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Type in the name of the location using the alphanumeric and underscore keys.
Press the ERASE button to erase the last character typed, if needed.
Press ENTER.
Use the Arrow keys to go to the next location to be edited and press the EDIT button. Continue in the same way to edit as many Locations as desired.
Press the BACK button to return to the Configuration screen or the MAIN button to return to the MAIN screen.
Activating the AUTO INC button allows the user to select the next location once the number of cycles is completed.
The user is prompted to move to the next location after the programmed number of cycle. See Figure 4-59.
The Auto Increment mode will select the next corresponding location and continue to sample using the current configuration. If the current location is using a recipe, the user will be prompted to unload the recipe before moving to the next location.
4-46
Figure 4-59 Auto Increment Prompt
248083400-1 Rev 7
Operation
Data View Buffer Screen
Data that is stored on the instrument is viewed in the Data Screen. When the buffer has filled to its limit of 3000 records, the newest records will overwrite the first records stored. The word “Records” on the MAIN screen will have an asterisk (*) next to it to indicate that the buffer has wrapped. When you go to the Data screen after the data has wrapped, the first record viewed will be the most current record, which will not necessarily be record #1.
If the Particle Volume is set to ft display and print in ft
3
. If the Particle Volume is set to m3 or Liters (L),
3
, then the concentration data will
then the concentration data will display and print in m3.
Press the DATA button on the MAIN screen to display the Data screen.
Depending on how the HANDHELD is configured, the data will display in either raw data mode, or be normalized to ft
3
or m3. See
Figure 4-60 & Figure 4-61.
248083400-1 Rev 7
Figure 4-60 Data Screen - RAW Data
4-47
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Scroll through the data using the UP and DOWN arrow buttons.
The single arrow moves one record at a time.
Note:
has wrapped, the record that is first displayed in the Data screen is the first record in the data buffer . This may not be Rec#1.
If you press the single down arrow button once, the instrument will display the last record in the data buffer.
When the buffer
Figure 4-61 Data Screen - Normalized data in ft
3
The Data screen displays the following information for each data record.
Rec# - Identifies which record is currently viewed.
*Recs - Displays how many records are currently stored in the HANDHELD’s buffer. When there is an asterisk (*) next to the word “Recs”, this means that the 3000 record data buffer has wrapped. The basic concept is First In, First Out (FIFO).
Size - Lists the channel sizes configured on the instrument.
Diff#, Diff/ft3, Diff/m3 - Indicates that each channel size’s data is displayed in differential mode. If a channel was disabled, then there is a blank space in is column; # indicates raw counts; ft
3
and m
3
indicate normalized counts.
Cuml#, Cuml/ft3, Cuml/m3 - Indicates that each channel size’s data is displayed in cumulative mode. If a channel was disabled, there will be a blank space in this column. # indicates raw counts while ft
3
and m3 indicate normalized counts.
Loc - Indicates the location at which the data record was recorded. The Location name listed is the alphanumeric label that was saved.
Smpl - Indicates the Sample Time (HH:MM:SS) at which the data
Date - Indicates the instrument date on which the data was
4-48
record was sampled.
recorded. The date will display in the format selected in the configuration (MM/DD/YYYY, DD/MM/YYYY or YYYY/MM/ DD).
248083400-1 Rev 7
Operation
Time - Indicates the instrument time at which the data was recorded (HH:MM:SS).
Instr - Indicates the state of the instrument at the time the data was recorded. Instrument states include GOOD or SRVC.
If SRVC appears, printouts will say “Service Required” and the sensor may need cleaning. Please contact Lighthouse Technical Support at 800-945-5905 (USA Toll Free), 541-770-5905 (Outside of USA).
Flow - Indicates the flow state of the instrument at the time the data record was recorded. Flow is recorded as OK or ALRM. If the flow was in alarm, it will print as Flow: Alert on the printouts.
Alarm - Indicates (NONE or YES) if the data record exceeded any of the alarm thresholds of any of the channels that were enabled for alarms. If the data record exceeded alarm thresholds, printouts show “Alarm: Yes”.
Laser - Indicates the status of the laser at the time the data record was recorded; possible laser states include OK or SRVC. If the laser needs to be serviced, it will print as Laser: Service.
If laser state is SRVC, please contact Lighthouse Technical Support at 800-945-5905 (USA Toll Free), 541-770-5905 (Outside of USA).
Printing Data View Buffer Report
Analog Data
If the Temperature/Relative Humidity probe is attached during data recording, the analog data will be listed in the Data screen as A1 (Temperature) and A2 (Relative Humidity).
The instrument stores up to 3000 records in its data buffer. The Buffer Report prints all or a range of the records in the buffer in the format specified by the selections in the Configuration.
At the end of the Buffer Report, the following summary information is printed.
Data displayed in Cumulative or Differential mode
Raw (“Particle Counts”) or Normalized (“Particles/ft
3
” or
“Particles/m3”)
# of samples in the report
For each channel size:
Maximum Value
248083400-1 Rev 7
Minimum Value
Average
Standard Deviation
4-49
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
**HANDHELD** Serial #: 30804001
----------------------­Location: 29 08-02-03, 15:34:46 Sample Time: 00:01:00 Flow: 0.1 cfm Laser: OK
74.7 DegF 52.2 %RH Particle Counts: Size Diff Cumul
0.3 29118 32543
0.5 2572 3425
1.0 797 853
3.0 29 56
5.0 24 27
**HANDHELD** Serial #: 30804001
----------------------­Location: 29 08-02-03, 15:34:46 Sample Time: 00:01:00 Flow: 0.1 cfm Laser: OK
74.7 DegF 52.2 %RH Particles/ Ft^3: Size Diff Cumul
0.3 2222.0 2350.0
0.5 115.0 128.0
1.0 13.0 13.0
3.0 0.0 0.0
5.0 0.0 0.0
This indicates
whether the data
viewed is in RAW
counts or
Normalized counts.
The UCL calculation will be printed for each channel size if the number of locations in the range is 2 through 9. The calculation will only print if the Format settings (Setting screen) are set to CUML and NORM. Any other format settings or locations less than 2 and greater than 9 will print N/A for each channel.
View the same data in different formats just by changing the Particle Display options on the Sample Setting screen.
The Data Summary can be viewed as:
Raw Counts, Differential Data
Raw Counts, Cumulative Data
Normalized Counts, Differential Data, Particles/ft
Normalized Counts, Differential Data, Particles/m
Normalized Counts, Cumulative Data, Particles/ft
Normalized Counts, Cumulative Data, Particles/m
3
3
3
3
PRINT RECORD
When you push the PRINT RECORD button, whatever record is currently displayed in the Data screen will print as a single sample record. The record will print according to the settings in the Print Configuration. See Figure 4-62.
Figure 4-62 Examples of Output from Auto Printout of Sample
4-50
248083400-1 Rev 7
Operation
Printing Range
To print a range of data from the buffer, scroll to the first data record to be included in the range.
Press the RANGE button. The following screen will display as shown in Figure 4-63.
Note: Printing ALL
records may take some time if the buffer is full.
Figure 4-63 Print Range Screen
Enter the number of records to print and press the ENTER button.
Press ALL to print all records in the buffer.
Press CANCEL to cancel printing.
Press BACK to return to the View Buffer screen without printing the report.
248083400-1 Rev 7
4-51
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
***** SUMMARY ***** **HANDHELD 3016** Serial #: 041004001
----­Location: LOC002 06-20-2013, 15:34:30 Sample Time: 00:01:00 Flow: 0.1 cfm Laser: OK TEMP: 66.77 F RH: 35.71 % Particle/ Ft^3: Size Cumul
0.3 4260.0
0.5 810.0
1.0 580.0
3.0 360.0
5.0 350.0
10.0 310.0
Location: LOC003
Particle/ Ft^3: Summary (Cuml):
----­# of samples =2 Size Max
0.3 4260.0
0.5 810.0
1.0 580.0
3.0 360.0
5.0 350.0
10.0 310.0
Size Min
0.3 3860.0
0.5 750.0
1.0 360.0
3.0 240.0
5.0 180.0
10.0 150.0
Size 95% UCL
0.3 5322.0
0.5 969.3
1.0 1164.1
3.0 678.6
5.0 801.4
10.0 734.8
4-52
Figure 4-64 Sample Buffer Report
248083400-1 Rev 7
Operation
REPORTS The HANDHELD can print the following Cleanroom Classification to
Standards reports:
Note:
The EU GMP 2009 report is only valid if the instrument is configured with the 0.5 and 5.0 micron channel sizes.
Note:
If the user sets the counter for a minimum volume, and takes the exact number of minimum locations and samples, the user could get a “Cannot Classify: too few locations” error if the minimum volume rounds the sample time down so that not enough samples at the minimum volume are taken.
In this case, add one or two seconds to the sample time so that the user is not at the exact minimum volume.
Federal Standard ft
3
ISO 14644-1 EU GMP 2009
Reports General Settings
Enable the Reports feature by pressing the red “X” (Default position) located at the top left of the screen. The button will display a check mark indicating the feature is enabled. See Figure 4-65.
Depending on the reports requirements, the number of cycles and
The number of required locations will be automatically assigned to
The display format will be changed to Normalized (normalized to
248083400-1 Rev 7
Figure 4-65 Typical Reports Screen
minimum sample time will be calculated when the reports mode is enabled.
the report. The user may change locations via the reports SETUP screen.
ft3 for FED-209 and m3 for ISO and EU GMP). The user may change the display mode from the reports SETTINGS screen.
4-53
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
When the reports mode is enabled, the currently selected report type and minimum required samples will be displayed on the MAIN screen as shown in Figure 4-66.
Figure 4-66 Reports Data
The instrument keeps track of and will display on the MAIN screen the number of samples taken for the report.
If Auto Increment is enabled, the user will be prompted to move the particle counter to the next assigned location after the completed number of cycles.
If Printing is enabled and Auto Increment is enabled, the report will be printed after the required number of samples are completed.
After the required number of samples have been completed, the instrument will prompt the user to disable Reports Mode as shown in Figure 4-67.
4-54
Figure 4-67 Disable Reports Prompt
248083400-1 Rev 7
Operation
When one of the report options buttons is pressed, the instrument uses the current Room Area, Class Level, Air Flow and Room Status parameters to determine the minimum number of locations, samples and volume per channel in order to classify a cleanroom with that standard.
Use these displayed minimum values to configure the instrument so that the minimum requirements needed to classify the cleanroom will be met.
In order to set parameters for a report, follow these steps:
Select the desired report.
Press the SETUP button; the Report Setup screen displays as shown in Figure 4-68.
248083400-1 Rev 7
Figure 4-68 Report Setup Screen
4-55
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Press the CLASS button and the Select Class screen displays as shown in Figure 4-69.
Figure 4-69 Select Class Level
Press the displayed values to select the desired Class Level. The selected level will be highlighted in blue.
Class Levels vary based on which standard has been selected in the PRINT REPORT field. Below are the available Class Levels, per standard:
FED_209E ft
3
- 1, 10, 100, 1000, 10000, 100000
ISO14644-1 - 1, 2, 3, 4, 4.8, 5, 6, 7, 8, 9
EU GMP - A, B, C, D (Rev 2009)
Changing the Class Level will also change the minimum number of locations, minimum number of samples and the minimum sample volume per channel size required to satisfy the requirements of the chosen report.
Press the BACK button.
4-56
248083400-1 Rev 7
Operation
Note: If the entered Area
values needs more locations than available, a “Not enough locations” message will be displayed and the user will be asked to enter a lesser area.
Press the AREA button to display screen shown in Figure 4-70.
Figure 4-70 Reports Area Setup Screen
Select the appropriate unit of measure (ft2 or m2) and type in the area using the keypad. Press ENTER to display the value on the screen.
Note:
The minimum area
that can be entered is 6 ft
2
.
1 m
The maximum area that can be entered depends on the number of locations available on the unit.
With all 200 locations available, the Max area that can be entered is limited to a 5 digit value for ft2 and a 4 digit value for m2.
2
or
Press the BACK button.
Press the STATUS button to display screen shown in Figure 4-71.
Figure 4-71 Reports Room Status Scree
Select appropriate room status (As Built, At Rest, In Operation).
Press the BACK button.
248083400-1 Rev 7
4-57
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Press the FLOW button to display screen shown in Figure 4-72.
Figure 4-72 Reports Air Flow Screen
Note:
locations cannot be saved for reports if Reports are disabled and if there are not enough locations to allocate to the report. The "LOC" and "SAVE buttons will not be displayed in these cases.
Recipes and
Choose the appropriate type of air flow (Unidirectional, Non­Unidirectional), then press the BACK button to return to the SETUP screen.
The SETUP screen shown in Figure 4-73 will display the values that were configured in the display fields.
Use this information to set up the instrument to meet or exceed the minimum requirements.
4-58
Figure 4-73 Configure the Instrument
248083400-1 Rev 7
Operation
Select the PARTICLE button to enable or disable particle channels as shown in Figure 4-74.
Figure 4-74 Reports Particle Screen
Press the BACK button.
Select the SAMPLE button to configure the number of Cycles, Delay time, Hold time, Sample time and Volume. See Figure 4-75.
Figure 4-75 Reports Sample Screen
Press the BACK button.
248083400-1 Rev 7
4-59
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Select the SETTING button to format the instrument for CUML/ DIFF, RAW/NORM, ft
Figure 4-76 Reports Settings Screen
3/m3
as shown in Figure 4-76.
Note: If the location is
already assigned to another recipe of if there are no available free locations, the user will not be able to add a new recipe. The "ADD" button will not be displayed.
Press the BACK button.
Select the LOC button to select, add or delete locations or enable Auto Inc as shown in Figure 4-77.
Figure 4-77 Reports Location Screen
Press the BACK button.
4-60
248083400-1 Rev 7
Operation
Select BACK to return to the main REPORTs screen or select SAVE to save the configuration in the Recipe database. See Figure 4-78.
Figure 4-78 Reports Setup Screen
When "REPORTS" are enabled, the "LOC" and "SAVE" buttons are displayed only when thre are at least the minimum number of locations needed for the recipe area available. See Figure 4-79.
The next section contains an example of setting up the instrument to run a report.
248083400-1 Rev 7
Figure 4-79 Setup Screen No LOC or SAVE Buttons
4-61
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Setting up Counter to Run Reports
For this example, the FED STD FT3 report will be used. See Figure 4-80.
Note:
The Minimum Volumes are measured in the standard measurement required by each specific report.
The number of digits allowed for area to calculate volume is limited to a 5 digit value for ft2 and a 4 digit value for m2.
Figure 4-80 FED STD FT3 Report Screen
Set the instrument for the Fed Std ft3 report, 100 ft2, Class 100, unidirectional air flow, operational room status.
This will generate the following minimum requirements:
Min Locations: 4
Min Samples: 5
Table 4-1 Table - Minimum Volume per Channel Size
Ch Size (μ)Min Vol (ft3)
0.3 0.100
0.5 0.200
1.0 0.917
3.0 10.309
5.0 31.696
10.0 145.985
Based on the minimum requirements displayed, configure the instrument as follows:
4-62
248083400-1 Rev 7
Note:
The Sample time automatically changes to 1:28 (1 minute, 28 seconds).
Note:
This is all that is needed to meet the minimum requirements to classify the report.
Increase any of the configuration parameters in order to meet additional reporting needs.
Operation
Press MAIN to return to the MAIN screen.
Press CFG to enter the Configuration screen.
Press SAMPLE to set the Sample time.
Press the VOLUME button to set the instrument’s sample volume.
Type in 146 and press ENTER. This sets the instrument for the channel size with the largest minimum volume requirement.
Press the CYCLES button.
Set the CYCLES value for 1 and press ENTER.
Press the MAIN button.
Press the Location Select button at the top of the screen.
Select the location where the first measurement will be taken.
Press START. Let the instrument record 1 sample at the first location.
After the instrument has recorded the sample, change the Location using the Location Select button.
Then move the instrument to the second location and press START.
After the instrument has recorded the second sample, continue to change the Location setting before recording at the next location, then move it to the third and forth locations, taking at least one sample at each location. At one location take 2 samples.
Note: Report Output can
also be configured as a new or previously saved recipe, which can be set through the Reports option or the Recipe option.
After the instrument finishes recording at the last location, and the instrument has a total of at least 5 samples, the instrument is ready to print the report.
Report Output Setup
The Report output can be directed to an external thermal printer. To configure the destination, press CFG then OUTPUT. In the Output Setup screen, verify that the Output on Sample is enabled with a green check mark.
Printing a Report
To print a report, the user can choose to either clear the data before starting the test or run the samples and then select the range of data to include in the report.
Once the user has collected the necessary samples, press REPORTS from the MAIN screen.
Press the RANGE button on the REPORTS screen
Using the ARROW buttons, scroll to the first record in the range of data for the report.
Press the RANGE button on the Data View Screen.
Type in the number of records to be included in the range.
Press ENTER.
248083400-1 Rev 7
4-63
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
The display will return to the main REPORTS screen and start printing the report.
Note:
If CANCEL is pressed, the report may not stop printing right away. It will continue until there is a break in the report processing.
To stop printing, press the CANCEL button.
Report Requirements
Fed Std ft
3
This report requires a minimum of 2 locations be measured, at least 1 sample be taken per location and at least 5 samples total be taken in each cleanroom at a minimum sample volume of 0.1 cubic feet.
The 95% Upper Confidence Limit (UCL) will be calculated by the HANDHELD if the number of sampled locations is equal to or less than 9.
Air flow (Unidirectional vs. Non-unidirectional) will have an effect on the minimum locations and minimum samples required. In general, the Non-unidirectional option requires more of both locations and samples.
Note:
The EU GMP 2009 report is only valid if the instrument is configured with the 0.5 and 5.0 micron channel sizes.
For the Fed Std ft
3
report, the classification for a particle size is based on its samples taken from all locations in the cleanroom. Each particle size is given one of the following results:
a classification rating which will be one of the possible classifications (1, 10, 100, 1000, 10000, 100000)
“Unable to Classify” if it fails one of the above requirements
an “Out of Range” if the calculated classification at that particle count exceeds 100,000
ISO 14644-1
The ISO14644-1 standard requires a minimum of 60 second Sample Time for each sample recorded.
EU GMP 2009
The EU GMP Standard Report requires that the unit has the channel sizes 0.5μ and 5.0μ as these are the only channels that apply to this standard.
The EU GMP Report has different concentration limits for cleanroom status “At Rest” versus “In Operation”. When the cleanroom is “In Operation”, Class D is not available. The standard is not defined for Class D in an operational cleanroom.
4-64
248083400-1 Rev 7
Sample Printouts of Standard Reports
************* Fed Standard 209E_Ft^3 **HANDHELD 3016** Serial #: 130602021 Targeted Class: 10000 Room Area: 10.00ft^2 Room Status: Operational Air Flow: Unidirect Min Loc: 2 Min Samples/Room: 5 Recipe: TEST
-----
06/20/2013, 09:09:08
-----
Particle Size: 0.3 Cumulative, p/ft^3 Vol Req: 0.100 ft^3 Concen Limit: 30800
----­Loc Samples AvgConcen LOC001 5 6114.0 LOC002 5 4572.0
----­Mean: 5343.0 StdDev: 1090.4 StdError: 771.0 95% UCL: 10208.0 Min Vol: 0.100 ft^3
-----
Particle Size: 0.5 Cumulative, p/ft^3 Vol Req: 0.100 ft^3 Concen Limit: 10000
-----
Loc Samples AvgConcen
Figure 4-81, Figure 4-82 and Figure 4-83 are examples of a Federal Standard Ft3, ISO 14644-1 and a EU GMP 2009 reports.
Note:
loaded, the name of the recipe will be printed on the report.
If a recipe is
Operation
Figure 4-81 Sample Federal Standard 209E (ft3) report
248083400-1 Rev 7
4-65
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
************* ISO 14644-1 **HANDHELD 3016** Serial#: 130602001 Targeted Class: 8 Room Area: 1.49 m^2 Room Status: Operational Air Flow: Unidirect Min Locations: 2 Min Samples/Room: 2 Recipe: TEST
----­06/19/2004, 18:10:17
-----
Particle Size: 0.5 Cumulative, p/m^3 Vol Req: 2.000 L Concen Limit: 3520000
----­Loc Samples AvgConcen LOC003 2 13949.3 LOC004 2 11477.3
----­Mean: 12713.3 StdDev: 1748.0 StdError: 1236.0 95% UCL: 20500.2 Min Vol: 2.832 L
-----
Particle Size: 1.0 Cumulative, p/m^3 Vol Req: 2.000 L Concen Limit: 832000
----­Loc Samples AvgConcen LOC003 2 7062.9 LOC004 2 7769.2
----­Mean: 7416.1 StdDev: 499.4
Note: If a recipe is
loaded, the name of the recipe will be printed on the report.
4-66
Figure 4-82 Sample ISO 14644-1 report
248083400-1 Rev 7
Note: If a recipe is
loaded, the name of the recipe will be printed on the report.
Operation
************ EU GMP 2009 **HANDHELD 3016** Serial #: 130602021 Targeted Class: C Room Area: 1.49 m^2 Room Status: Operational Air Flow: Unidirect Min Locations: 2 Min Samples/Room: 2 Recipe: TEST
-----
06/21/2013, 09:12:09
-----
Particle Size: 0.5 Cumulative, p/m^3 Vol Req: 2.000 L Concen Limit: 3520000
----­Loc Samples AvgConcen LOC001 2 8872.8 LOC002 2 23228.2
----­Mean: 16050.5 StdError: 10150.8 StdDev: 7177.7 95% UCL: 61270.1 Min Vol: 56.634 L
-----
248083400-1 Rev 7
Particle Size: 5.0 Cumulative, p/m^3 Vol Req: 2.000 L Concen Limit: 29000
----­Loc Samples AvgConcen LOC001 2 194.2 LOC002 2 317.8
Figure 4-83 Sample EU GMP 2009 report
4-67
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
Power Shutdown Levels
Note: It is not
recommended to allow the battery to discharge completely.
When the HANDHELD is powered from its rechargeable battery, a Power Shutdown feature protects the battery from discharging completely. A complete discharge could damage the battery.
The battery levels are displayed in Table 4-2.
Table 4-2 Levels of Battery Life
Battery
Icon
Description
FULL
75%
50%
25%
Empty; at this level, the instrument will begin to beep and will display “BATT LOW!” on the main screen. It is recommended to connect the AC cord and plug it into an outlet at this level.
Pump Shutdown; if the instrument is currently on and sampling the pump will turn off and the instrument will shut down.
After the pump shutdown level and before the battery reaches its critical low point, the instrument will shut down all power. To prevent loss of power, attach the AC cord and plug into an outlet as soon as the BATT LOW! message appears on the screen.
4-68
248083400-1 Rev 7
00
5 Setting up the Counter
Using Report Parameters
Note:
reports, the optional printer must be attached to the HANDHELD.
In order to review
Basic Requirements
Besides printing reports, the Report Setup screen can be used to display each report’s minimum requirements - such as how many data records are required and from how many different locations data must be recorded to meet a given class for a given cleanroom area.
Once the user knows these requirements, the user can configure the HANDHELD so that the data collected will fulfill the report’s requirements.
For each report type, as the user changes the Room Area, the Class Level, and sometimes the Air Flow or the Room Status, the values for the Minimum Locations, Minimum Samples and Minimum Volume may change and will be calculated for the requirements for each report type.
These values are indicated on the screen so that the user can configure the instrument appropriately to ensure the samples recorded meet the requirements for the report selected.
Each of the reports have some basic requirements that, if not met, cause the report to FAIL classifying the cleanroom immediately.
Fed Std 209E (feet)
This report requires a minimum of 2 locations measured.
At least 1 sample per location must be taken.
At least 5 samples in total must be taken in the cleanroom.
The minimum sample volume must be at least 0.1 cubic feet.
The 95% Upper Confidence Limit (UCL) must be calculated if the number of sampled locations is less than or equal to 9.
Air flow (Unidirectional vs. Non-unidirectional) will have an effect on the minimum locations and minimum samples required. In general, the Non-unidirectional option requires more of both locations and samples.
248083400-1 Rev 7
5-1
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
For the Fed Std 209E (ft) report, the classification for a particle size is based on its samples taken from all locations in the cleanroom. Each particle size is given one of the following results:
A classification rating which will be one of the possible classifications (1, 10, 100, 1000, 10000, 100000).
“Unable to Classify” if it fails one of the above requirements.
An “Out of Range” if the calculated classification at that particle count exceeds 100,000.
ISO 14644-1
The minimum number of locations is 1.
At least 1 sample must be taken per location.
At least 3 samples, total, must be taken in the cleanroom. The user can take a different number of samples at different locations.
The minimum sample volume is 2 liters.
This report has a minimum Sample Time of 1 minute. If the Sample Time is less than that, the entire room would fail the classification.
The 95% Upper Confidence Limit (UCL) must be calculated if the number of sampled locations is less than or equal to 9.
EU GMP 2009
The EU GMP 2009 only applies to 2 channel sizes: 0.5μ and 5.0μ.
There are only 4 Classes: A, B, C, D, and 2 cleanroom states: At Rest and Operational.
Class D is not applicable when the cleanroom is Operational.
If one of these channel sizes is not present, the report cannot classify the room.
The standard does not define a minimum number of sample locations or a minimum number of samples. In order to take a measurement, at least 1 location and 1 sample are required by the counter.
The standard does not require a minimum sample volume. The volume of air must be shown in liters (L).
5-2
248083400-1 Rev 7
Setting up the Counter
The classification for a particle size is determined based on the mean of all average concentrations taken from all locations in the cleanroom. If the mean exceeds all limits, then the classification is “Out of Range”.
The counter will only print channels 0.5 and 5.0 in this report. If either channel is not available on the HANDHELD, then this report will not be valid.
248083400-1 Rev 7
5-3
Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
5-4
248083400-1 Rev 7
Loading...