Lifeloc FC20 Operation Manual

Page 1
Lifeloc Technologies, Inc
FC20 Portable Breath Tester
Operations Manual
Users Manual
Version 3.0
PatentNo.6,596,153
Page 1
Lifeloc Technologies, Inc
FC20 Portable Breath Tester
Operations Manual
Users Manual
Version 3.0
PatentNo.6,596,153
Page 2
Page 2
Page 3
INTRODUCTION Welcome 5
Front View 6 Standard Features 7 PREPARATION Installing Batteries 8 Attaching a Mouthpiece 8 On/Off 9 Observing the Subject 9
OPERATION Taking an Auto Test 10
Breath Flow 11 Alcohol Curve 11 Test Results 12 Removing the Mouthpiece 12 Viewing Previous Test Results 12 Printing Test Results 13
BREATH TESTING MODES
Explained 14 Conducting an Automatic Test 14 Conducting a Manual Test 15 Manually Overriding an Automatic Test 15 Conducting a Passive Test 16 SUMMARY 17 USER SETTINGS Settings Chart 18 Settings Screen 19 Setting the Time 20 Setting the Date 21 Setting the Shutoff Time 22 I.D. Entry Explained 23 Entering a Batch or Subject I.D. 23 Print Settings Explained 26 Setting Number of Copies Printed 26 Print Format 27 Printer Selection 27 Sample Print-Outs 28
Lifeloc Technologies, Inc
Table of Contents
Page 3
INTRODUCTION Welcome 5
Front View 6 Standard Features 7 PREPARATION Installing Batteries 8 Attaching a Mouthpiece 8 On/Off 9 Observing the Subject 9
OPERATION Taking an Auto Test 10
Breath Flow 11 Alcohol Curve 11 Test Results 12 Removing the Mouthpiece 12 Viewing Previous Test Results 12 Printing Test Results 13
BREATH TESTING MODES
Explained 14 Conducting an Automatic Test 14 Conducting a Manual Test 15 Manually Overriding an Automatic Test 15 Conducting a Passive Test 16 SUMMARY 17 USER SETTINGS Settings Chart 18 Settings Screen 19 Setting the Time 20 Setting the Date 21 Setting the Shutoff Time 22 I.D. Entry Explained 23 Entering a Batch or Subject I.D. 23 Print Settings Explained 26 Setting Number of Copies Printed 26 Print Format 27 Printer Selection 27 Sample Print-Outs 28
Lifeloc Technologies, Inc
Table of Contents
Page 4
Lifeloc Technologies, Inc
Table of Contents
USER SETTINGS Display Contrast 29 Test Order Sequence 29 Results Format Explained 30 Setting Results Format (PASS/WARN/FAIL) 31
FACTORY OPTIONS
Password Protection 32 Auto Air Blank 33 Automatic Altitude Compensation 34
CALIBRATION SETTINGS
Calibration Explained 35 Calibration Display Chart 37
Setting the Calibration Standard 38 Setting the Standard Type 39
Setting the Lock-Out Periods 40
WET BATH CALIBRATION
Simulator Set-Up 41 Wet Bath Calibration 42 Wet Bath Cal Check 43
DRY GAS CALIBRATION
Dry Gas Set-Up 44 Dry Gas Calibration 45 Dry Gas Cal Check 46
MAINTENANCE
Fuel Cells 47 Cleaning 47 Batteries 47
COMPUTER CONNECTION
DataTrak 48
ERROR MESSAGES
Chart 49
SERVICE
Factory Limited Warranty 50 Extended Service Plans 50 Service 50
NOTICE
Notice 51
Page 4
Lifeloc Technologies, Inc
Table of Contents
USER SETTINGS Display Contrast 29 Test Order Sequence 29 Results Format Explained 30 Setting Results Format (PASS/WARN/FAIL) 31
FACTORY OPTIONS
Password Protection 32 Auto Air Blank 33 Automatic Altitude Compensation 34
CALIBRATION SETTINGS
Calibration Explained 35 Calibration Display Chart 37
Setting the Calibration Standard 38 Setting the Standard Type 39
Setting the Lock-Out Periods 40
WET BATH CALIBRATION
Simulator Set-Up 41 Wet Bath Calibration 42 Wet Bath Cal Check 43
DRY GAS CALIBRATION
Dry Gas Set-Up 44 Dry Gas Calibration 45 Dry Gas Cal Check 46
MAINTENANCE
Fuel Cells 47 Cleaning 47 Batteries 47
COMPUTER CONNECTION
DataTrak 48
ERROR MESSAGES
Chart 49
SERVICE
Factory Limited Warranty 50 Extended Service Plans 50 Service 50
NOTICE
Notice 51
Page 5
To the FC20 Portable Breath Tester Operator:
WELCOME!
The FC20 breath alcohol tester is manufactured in Den­ver, Colorado, by Lifeloc Technologies, Inc. Lifeloc offers premium quality products combined with exceptional service and technical support.
The FC20 is a state-of-the-art breath alcohol tester that is software based and incorporates unique cutting edge technologies. Because of the advanced FC20 design:
èResults on a positive test register in 10 seconds.
èYou can take another test 30 seconds after a posi-
tive.
èYour FC20 will automatically take the test when it
senses a deep lung sample is delivered.
èYour FC20 will provide an accurate test, or else
explain to you why it cannot, and even provide sug­gestions on how to proceed to complete an accurate test on your subject.
èYour FC20 includes a host of other features: Cali-
bration/Calibration Check lock-out periods, Batch and Subject IDs, numerous printout options, passive testing, password protection (option) and more. All are explained in this manual.
Introduction
Page 5
To the FC20 Portable Breath Tester Operator:
WELCOME!
The FC20 breath alcohol tester is manufactured in Den­ver, Colorado, by Lifeloc Technologies, Inc. Lifeloc offers premium quality products combined with exceptional service and technical support.
The FC20 is a state-of-the-art breath alcohol tester that is software based and incorporates unique cutting edge technologies. Because of the advanced FC20 design:
èResults on a positive test register in 10 seconds.
èYou can take another test 30 seconds after a posi-
tive.
èYour FC20 will automatically take the test when it
senses a deep lung sample is delivered.
èYour FC20 will provide an accurate test, or else
explain to you why it cannot, and even provide sug­gestions on how to proceed to complete an accurate test on your subject.
èYour FC20 includes a host of other features: Cali-
bration/Calibration Check lock-out periods, Batch and Subject IDs, numerous printout options, passive testing, password protection (option) and more. All are explained in this manual.
Introduction
Page 6
FC FRONT VIEW
Introduction
Frontview without mouthpiece.
Frontviewwith mouthpiece aached.
Mouthpiece removaltab.
AUTO TEST
#00013
Page 6
FC FRONT VIEW
Introduction
Frontview without mouthpiece.
Frontviewwith mouthpiece aached.
Mouthpiece removaltab.
AUTO TEST
#00013
Page 7
Introduction
STANDARD FEATURES
Large graphic LCD display, capable of showing numbers, let-
ters, icons, and plain-English text messages
Automatic Calibration, software controlled adjustments, no screwdriver or tools necessary
Auto Test Mode, the easiest, simplest way to take a test of any breath tester, Auto Test is fast, accurate, and virtually “hands­free”
On Board Memory, stores the last 500 tests
Real Time Clock, stores time and date information with test re-
sults as well as calibration and cal check results
Printer Ready, print out any or all results using the Lifeloc AP893 thermal printer
Auto Shut-Off, preserves battery life
Passive Test Mode, check for the presence of alcohol without
using a mouthpiece
Two Printout Options, long or short
Calibration Reminder with Lockout, prevents you from using an
FC20 when it is due for calibration
User Selectable Test Order, allows choice of either Auto Test or Passive Test default mode
I.D. Entry, capable of storing two separate I.D. numbers, one for subject and one for additional uses, operator, precinct, etc.
Fast Simple Operation, while the FC20 contains a host of fea­tures, it is still easy to use
Page 7
Introduction
STANDARD FEATURES
Large graphic LCD display, capable of showing numbers, let-
ters, icons, and plain-English text messages
Automatic Calibration, software controlled adjustments, no screwdriver or tools necessary
Auto Test Mode, the easiest, simplest way to take a test of any breath tester, Auto Test is fast, accurate, and virtually “hands­free”
On Board Memory, stores the last 500 tests
Real Time Clock, stores time and date information with test re-
sults as well as calibration and cal check results
Printer Ready, print out any or all results using the Lifeloc AP893 thermal printer
Auto Shut-Off, preserves battery life
Passive Test Mode, check for the presence of alcohol without
using a mouthpiece
Two Printout Options, long or short
Calibration Reminder with Lockout, prevents you from using an
FC20 when it is due for calibration
User Selectable Test Order, allows choice of either Auto Test or Passive Test default mode
I.D. Entry, capable of storing two separate I.D. numbers, one for subject and one for additional uses, operator, precinct, etc.
Fast Simple Operation, while the FC20 contains a host of fea­tures, it is still easy to use
Page 8
Preparation
INSTALLING BATTERIES
ATTACHING A MOUTHPIECE
Remove the mouthpiece from its wrapper, making sure not to touch the end which the subject will be blowing into.
Attach the mouthpiece to the port on the back of the FC20.
Line up the mouthpiece port over the hole in the back of the mouthpiece. Press in place.
Ensure it is snuggly fit.
Press in and down on the battery door located on the back of the FC20 in the direction of the arrow.
Install the 4 AA Alkaline batteries in the direction of the symbols in the battery case.
Close the case by pushing up on the battery door until it locks shut.
Page 8
Preparation
INSTALLING BATTERIES
ATTACHING A MOUTHPIECE
Remove the mouthpiece from its wrapper, making sure not to touch the end which the subject will be blowing into.
Attach the mouthpiece to the port on the back of the FC20.
Line up the mouthpiece port over the hole in the back of the mouthpiece. Press in place.
Ensure it is snuggly fit.
Press in and down on the battery door located on the back of the FC20 in the direction of the arrow.
Install the 4 AA Alkaline batteries in the direction of the symbols in the battery case.
Close the case by pushing up on the battery door until it locks shut.
Page 9
OBSERVING THE SUBJECT
TURNING THE FC20 ON & OFF
Preparation
Press and hold the Power button on the bottom of the unit until it beeps. The FC20 performs an automatic internal diagnostics check.
To turn the unit off, press and hold the Power button until it beeps. The unit will shut down.
The FC20 provides a highly accurate reading of breath alcohol acquired by sampling deep lung air. However, the reading can be corrupted by residual mouth alcohol.
To prevent mouth alcohol from affecting a test, make certain that the subject is not allowed to put anything in their mouth for the 15 minutes prior to taking a test.
If the subject just took a drink, a 15 minute obser­vation period in which they are not permitted to put anything in their mouth should be observed before testing. This will ensure all residual alcohol from any source is completely dissipated and test results will be valid.
Page 9
OBSERVING THE SUBJECT
TURNING THE FC20 ON & OFF
Preparation
Press and hold the Power button on the bottom of the unit until it beeps. The FC20 performs an automatic internal diagnostics check.
To turn the unit off, press and hold the Power button until it beeps. The unit will shut down.
The FC20 provides a highly accurate reading of breath alcohol acquired by sampling deep lung air. However, the reading can be corrupted by residual mouth alcohol.
To prevent mouth alcohol from affecting a test, make certain that the subject is not allowed to put anything in their mouth for the 15 minutes prior to taking a test.
If the subject just took a drink, a 15 minute obser­vation period in which they are not permitted to put anything in their mouth should be observed before testing. This will ensure all residual alcohol from any source is completely dissipated and test results will be valid.
Page 10
Operation
TAKING AN AUTO TEST
Attach the mouthpiece to the back of the unit.
Verify the display reads “AUTO TEST”.
Instruct subjects to blow into the mouthpiece firmly and steadily for as long as they can. (Not neces­sarily as hard as they can.)
Read the result.
After taking a test the FC20 will display the results in large numbers on the display.
Additionally, the results are stored in memory and available for viewing at a later date.
Press the Function button to return to the test mode.
Page 10
Operation
TAKING AN AUTO TEST
Attach the mouthpiece to the back of the unit.
Verify the display reads “AUTO TEST”.
Instruct subjects to blow into the mouthpiece firmly and steadily for as long as they can. (Not neces­sarily as hard as they can.)
Read the result.
After taking a test the FC20 will display the results in large numbers on the display.
Additionally, the results are stored in memory and available for viewing at a later date.
Press the Function button to return to the test mode.
Page 11
Operation
BREATH FLOW
As the subject blows into the mouthpiece, the FC20 will show a graph of the breath flow on the dis­play.
ALCOHOL CURVE
If the FC20 detects alcohol, the alcohol level is graphed and will be displayed before the result.
Page 11
Operation
BREATH FLOW
As the subject blows into the mouthpiece, the FC20 will show a graph of the breath flow on the dis­play.
ALCOHOL CURVE
If the FC20 detects alcohol, the alcohol level is graphed and will be displayed before the result.
Page 12
Operation
Press the Function button repeatedly until the print­er icon is displayed. The results of the last test will be displayed along with its test number at the lower left of the display.
The FC20 retains up to 500 tests along with all related data in memory.
Press the + or- buttons to view previous tests.
Press the Function button repeatedly to return to the testing mode.
VIEWING PREVIOUS TEST RESULTS
TEST RESULT
.080
BAC
After the alcohol is graphed, the test result is dis­played.
REMOVING THE MOUTHPIECE
Remove the mouthpiece by pushing against the back of the tab located on the exhaust end.
The result will remain on the screen for 90 seconds, or until the Function or Power button is pressed, or the test is printed.
Page 12
Operation
Press the Function button repeatedly until the print­er icon is displayed. The results of the last test will be displayed along with its test number at the lower left of the display.
The FC20 retains up to 500 tests along with all related data in memory.
Press the + or- buttons to view previous tests.
Press the Function button repeatedly to return to the testing mode.
VIEWING PREVIOUS TEST RESULTS
TEST RESULT
.080
BAC
After the alcohol is graphed, the test result is dis­played.
REMOVING THE MOUTHPIECE
Remove the mouthpiece by pushing against the back of the tab located on the exhaust end.
The result will remain on the screen for 90 seconds, or until the Function or Power button is pressed, or the test is printed.
Page 13
Operation
Printing the Current Test
Take a test.
Plug the printer cable into the connector on the side of the FC20.
Press the Execute button, under the printer icon.
Printing a Specific Test
Press the Function button repeatedly until the print­er icon is displayed.
Press the + or – button to select the test number you wish to print.
Press the Execute button, under the printer icon, and the printer will activate.
Printing All Stored Tests
Press the Function button repeatedly until the print­er icon is displayed..
Press the + or – button to select ‘ALL’.
Press the Execute button, under the printer icon, and the printer will activate.
PRINTING TEST RESULTS
Page 13
Operation
Printing the Current Test
Take a test.
Plug the printer cable into the connector on the side of the FC20.
Press the Execute button, under the printer icon.
Printing a Specific Test
Press the Function button repeatedly until the print­er icon is displayed.
Press the + or – button to select the test number you wish to print.
Press the Execute button, under the printer icon, and the printer will activate.
Printing All Stored Tests
Press the Function button repeatedly until the print­er icon is displayed..
Press the + or – button to select ‘ALL’.
Press the Execute button, under the printer icon, and the printer will activate.
PRINTING TEST RESULTS
Page 14
Turn the FC20 on.
Attach a mouthpiece to the back of the unit.
Verify the display reads “AUTO TEST”.
Instruct subjects to blow into the mouthpiece firmly and steadily for as long as they can. (Not neces­sarily as hard as they can.)
Read the result.
CONDUCTING AN AUTOMATIC TEST
Breath Testing Modes
The FC20 is capable of conducting 3 distinct types of breath tests.
èAutomatic Test is the most used and highest accuracy test.
The FC20 monitors the subject’s breath and automatically takes the sample near the end of the breath flow.
èManual Test is normally used only when the subject is una-
ble to provide a sufficient air sample for the automatic test.
èPassive Test is a quick screen to detect alcohol but is not
designed to quantify the results. Passive results are report­ed as POS if alcohol is detected, NEG if alcohol is not de­tected. In this mode, no mouthpiece is required.
EXPLAINED
Page 14
Turn the FC20 on.
Attach a mouthpiece to the back of the unit.
Verify the display reads “AUTO TEST”.
Instruct subjects to blow into the mouthpiece firmly and steadily for as long as they can. (Not neces­sarily as hard as they can.)
Read the result.
CONDUCTING AN AUTOMATIC TEST
Breath Testing Modes
The FC20 is capable of conducting 3 distinct types of breath tests.
èAutomatic Test is the most used and highest accuracy test.
The FC20 monitors the subject’s breath and automatically takes the sample near the end of the breath flow.
èManual Test is normally used only when the subject is una-
ble to provide a sufficient air sample for the automatic test.
èPassive Test is a quick screen to detect alcohol but is not
designed to quantify the results. Passive results are report­ed as POS if alcohol is detected, NEG if alcohol is not de­tected. In this mode, no mouthpiece is required.
EXPLAINED
Page 15
Turn the FC20 on.
Attach a mouthpiece to the back of the unit.
Press the Function button until the display reads “MANUAL TEST”.
Instruct the subject to blow into the mouthpiece firmly and steadily for as long as they can.
When they are near the end of their breath, press the Execute button.
Read the result.
CONDUCTING A MANUAL TEST
Breath Testing Modes
MANUALLY OVERRIDING AN AUTOMATIC TEST
Note: This feature allows the completion of a test
in the occasional instance when the subject may have diminished lung capacity and cannot activate the Auto Test .
Turn the FC20 on.
Attach a mouthpiece to the back of the unit and verify the display reads “AUTO TEST”.
Instruct the subject to blow into the mouthpiece firmly and steadily for as long as they can.
When they are near the end of their breath, press the Execute button.
Read the result.
Page 15
Turn the FC20 on.
Attach a mouthpiece to the back of the unit.
Press the Function button until the display reads “MANUAL TEST”.
Instruct the subject to blow into the mouthpiece firmly and steadily for as long as they can.
When they are near the end of their breath, press the Execute button.
Read the result.
CONDUCTING A MANUAL TEST
Breath Testing Modes
MANUALLY OVERRIDING AN AUTOMATIC TEST
Note: This feature allows the completion of a test
in the occasional instance when the subject may have diminished lung capacity and cannot activate the Auto Test .
Turn the FC20 on.
Attach a mouthpiece to the back of the unit and verify the display reads “AUTO TEST”.
Instruct the subject to blow into the mouthpiece firmly and steadily for as long as they can.
When they are near the end of their breath, press the Execute button.
Read the result.
Page 16
Turn the FC20 on.
Press the Function button until the display reads “PASSIVE TEST”.
To take the test:
Hold the FC20 sample port (orange-colored open­ing labeled Port on the back label of the FC20) about 4 inches from the subjects mouth.
Have the subject blow towards the port.
Press the Execute button while the subject is blow­ing.
Read the result. (“POS” or “NEG”)
CONDUCTING A PASSIVE TEST (No mouthpiece)
Breath Testing Modes
Page 16
Turn the FC20 on.
Press the Function button until the display reads “PASSIVE TEST”.
To take the test:
Hold the FC20 sample port (orange-colored open­ing labeled Port on the back label of the FC20) about 4 inches from the subjects mouth.
Have the subject blow towards the port.
Press the Execute button while the subject is blow­ing.
Read the result. (“POS” or “NEG”)
CONDUCTING A PASSIVE TEST (No mouthpiece)
Breath Testing Modes
Page 17
Summary
OK, was that easy or what!
Now that you have been through the first part of this manual you know everything necessary to begin using your FC20. You are capable of conducting Auto, Manual, and Passive tests with ease… but that’s not the entire story.
The following pages will guide you through the addition­al user-selectable settings available in this state-of-the­art instrument. We highly recommend you look these over and see if they can improve the efficiency and ease-of­use of your FC20 breath alcohol tester by customizing it to your particular situation.
After looking over the settings you will want to proceed to the Calibration and Maintenance section. There, we will take you through step-by-step processes for calibra­tion of your FC20 and make maintenance recommenda­tions, to keep your unit functional for years of trouble­free use.
Finally, in the warranty and service section, you can find the details of your FC20 factory warranty, and infor­mation on service and high quality supplies available through Lifeloc Technologies.
Fast, accurate, and easy to use, the FC20 breath alcohol tester is a premium quality, precision, state-of-the-art breath alcohol tester, designed for professional use.
Page 17
Summary
OK, was that easy or what!
Now that you have been through the first part of this manual you know everything necessary to begin using your FC20. You are capable of conducting Auto, Manual, and Passive tests with ease… but that’s not the entire story.
The following pages will guide you through the addition­al user-selectable settings available in this state-of-the­art instrument. We highly recommend you look these over and see if they can improve the efficiency and ease-of­use of your FC20 breath alcohol tester by customizing it to your particular situation.
After looking over the settings you will want to proceed to the Calibration and Maintenance section. There, we will take you through step-by-step processes for calibra­tion of your FC20 and make maintenance recommenda­tions, to keep your unit functional for years of trouble­free use.
Finally, in the warranty and service section, you can find the details of your FC20 factory warranty, and infor­mation on service and high quality supplies available through Lifeloc Technologies.
Fast, accurate, and easy to use, the FC20 breath alcohol tester is a premium quality, precision, state-of-the-art breath alcohol tester, designed for professional use.
Page 18
User Settings
THE SETTINGS CHART
Time
Date
Auto Airblank (Option)
Shutoff
Time
I.D.
Settings
Print Settings
Display Settings
EXIT
Execute
Function
Function
Function
Function
Function
Function
Function
Number of
Copies
Print
Format
Printer
Exit
Execute
Function
Function
Function
Settings
LCD
Contrast
Test
Order
Results Format
Pass Level
Fail Level
Function
Function
Function
Function
Exit
Function
Execute
Security Setting (Option)
Function
Page 18
User Settings
THE SETTINGS CHART
Time
Date
Auto Airblank (Option)
Shutoff
Time
I.D.
Settings
Print Settings
Display Settings
EXIT
Execute
Function
Function
Function
Function
Function
Function
Function
Number of
Copies
Print
Format
Printer
Exit
Execute
Function
Function
Function
Settings
LCD
Contrast
Test
Order
Results Format
Pass Level
Fail Level
Function
Function
Function
Function
Exit
Function
Execute
Security Setting (Option)
Function
Page 19
User Settings
THE SETTINGS SCREEN
The FC20 has numerous internal settings allowing you to customize the unit to your individual needs. The settings include the following:
1. Time
2. Date
3. Auto Air Blank (If installed see page 33)
4. Auto Shut-Off Time (1 – 15 minutes or “Off”)
5. I.D. Settings
Batch ID
User + Batch ID
6. Password Protection (If installed see page 32)
7. Print Settings
Number of copies printed (1, 2 or 3)
Print Format (Long or Short)
Printer selection (Ap890, DataTrak, Other)
8. Display Settings
LCD Contrast
Test Order
Results Format
Pass & Fail Level Settings
Page 19
User Settings
THE SETTINGS SCREEN
The FC20 has numerous internal settings allowing you to customize the unit to your individual needs. The settings include the following:
1. Time
2. Date
3. Auto Air Blank (If installed see page 33)
4. Auto Shut-Off Time (1 – 15 minutes or “Off”)
5. I.D. Settings
Batch ID
User + Batch ID
6. Password Protection (If installed see page 32)
7. Print Settings
Number of copies printed (1, 2 or 3)
Print Format (Long or Short)
Printer selection (Ap890, DataTrak, Other)
8. Display Settings
LCD Contrast
Test Order
Results Format
Pass & Fail Level Settings
Page 20
User Settings
THE SETTINGS SCREEN
1. CHANGING THE TIME
From the “Settings” display, press the Execute but­ton. The display will now read “Time” with the hours digits flashing.
Press the + or – buttons to change the hour.
Press the Execute button to move to minutes.
Press the + or- buttons to change the minutes.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Press the Function button repeatedly until display reads “SETTINGS”.
Press the Execute button once.
Pressing the Function button repeatedly scrolls you through the FC20 user settings.
Page 20
User Settings
THE SETTINGS SCREEN
1. CHANGING THE TIME
From the “Settings” display, press the Execute but­ton. The display will now read “Time” with the hours digits flashing.
Press the + or – buttons to change the hour.
Press the Execute button to move to minutes.
Press the + or- buttons to change the minutes.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Press the Function button repeatedly until display reads “SETTINGS”.
Press the Execute button once.
Pressing the Function button repeatedly scrolls you through the FC20 user settings.
Page 21
User Settings
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “Date”.
Press the + or – buttons to change the year.
Press the Execute button to move to the month set­ting.
Press the + or – buttons to change the month.
Press the Execute button to move to the day.
Press the + or- button to change the day.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
2. CHANGING THE DATE
Date
09/24/02
Next Digit
Page 21
User Settings
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “Date”.
Press the + or – buttons to change the year.
Press the Execute button to move to the month set­ting.
Press the + or – buttons to change the month.
Press the Execute button to move to the day.
Press the + or- button to change the day.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
2. CHANGING THE DATE
Date
09/24/02
Next Digit
Page 22
3. AUTOMATIC AIR BLANK (OPTION)
Auto Air Blank is a factory option available on the FC20.
If you have this option installed, refer to page 34 of this manual for instructions on set-up and use.
User Settings
4. SETTING THE AUTO SHUTOFF TIME
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “Shutoff Time”.
Press the + or - buttons to adjust the shutoff time between 1-15 minutes or disable auto shutoff by selecting “0”.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Page 22
3. AUTOMATIC AIR BLANK (OPTION)
Auto Air Blank is a factory option available on the FC20.
If you have this option installed, refer to page 34 of this manual for instructions on set-up and use.
User Settings
4. SETTING THE AUTO SHUTOFF TIME
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “Shutoff Time”.
Press the + or - buttons to adjust the shutoff time between 1-15 minutes or disable auto shutoff by selecting “0”.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Page 23
User Settings
5. I.D. ENTRY EXPLAINED
The FC20 allows you to enter two different I.D. numbers and/or names: the Operator (Batch I.D.) and the Subject (Subject I.D.).
The Batch I.D. remains the same once entered, until changed again. If entered, it is stored and printed out with every test result.
The Subject I.D. is unique to each test. If turned on, you will be prompted to enter a Subject I.D. every time you take a test. The Subject I.D. will be stored and printed out with the test result.
5. ENTERING A BATCH I.D.
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “I.D. Setting”.
Press the Execute button to select either “Subject + Batch” or “Batch”, then press the Function button once. The display will read “Batch I.D.”
Press the + and – buttons to enter numbers/letters of the I.D you wish to enter (hold down to scroll quickly).
Press the Execute button to move to the next digit. Up to 16 characters may be entered.
Press the Function button to save the changes and return to the testing mode.
Page 23
User Settings
5. I.D. ENTRY EXPLAINED
The FC20 allows you to enter two different I.D. numbers and/or names: the Operator (Batch I.D.) and the Subject (Subject I.D.).
The Batch I.D. remains the same once entered, until changed again. If entered, it is stored and printed out with every test result.
The Subject I.D. is unique to each test. If turned on, you will be prompted to enter a Subject I.D. every time you take a test. The Subject I.D. will be stored and printed out with the test result.
5. ENTERING A BATCH I.D.
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “I.D. Setting”.
Press the Execute button to select either “Subject + Batch” or “Batch”, then press the Function button once. The display will read “Batch I.D.”
Press the + and – buttons to enter numbers/letters of the I.D you wish to enter (hold down to scroll quickly).
Press the Execute button to move to the next digit. Up to 16 characters may be entered.
Press the Function button to save the changes and return to the testing mode.
Page 24
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “I.D. Setting”.
Press the Execute button to select “Subject + Batch”, then, press the Function button once. The display will read “Batch I.D.”
Press the + and – buttons to enter numbers/letters of the Batch I.D. (hold down to scroll quickly).
Press the Execute button to move to the next digit. Up to 16 characters may be entered.
Press the Function button to save the changes and return to the testing mode.
The FC20 will now prompt you to enter a Subject I.D. using the + and - buttons before every test.
Use the + and – buttons to enter numbers/letters (hold down to scroll quickly).
Press the Execute button to move to the next digit. Up to 16 characters may be entered.
Press the Function button to proceed to take a test. The Subject I.D. will be stored and printed with the next test taken.
To by-pass entering a Subject I.D., simply press the Function button and take a test.
5. ENTERING A SUBJECT I.D.
User Settings
Page 24
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “I.D. Setting”.
Press the Execute button to select “Subject + Batch”, then, press the Function button once. The display will read “Batch I.D.”
Press the + and – buttons to enter numbers/letters of the Batch I.D. (hold down to scroll quickly).
Press the Execute button to move to the next digit. Up to 16 characters may be entered.
Press the Function button to save the changes and return to the testing mode.
The FC20 will now prompt you to enter a Subject I.D. using the + and - buttons before every test.
Use the + and – buttons to enter numbers/letters (hold down to scroll quickly).
Press the Execute button to move to the next digit. Up to 16 characters may be entered.
Press the Function button to proceed to take a test. The Subject I.D. will be stored and printed with the next test taken.
To by-pass entering a Subject I.D., simply press the Function button and take a test.
5. ENTERING A SUBJECT I.D.
User Settings
Page 25
User Settings
6. PASSWORD PROTECTION (OPTION)
The password option is a factory installed option that allows the FC20 administrators to prevent field personnel from performing any change to the FC20 user settings or from performing a calibra­tion or cal check.
For more information regarding the use of this option refer to page 33 of this manual for instruc­tions on set-up and use.
Page 25
User Settings
6. PASSWORD PROTECTION (OPTION)
The password option is a factory installed option that allows the FC20 administrators to prevent field personnel from performing any change to the FC20 user settings or from performing a calibra­tion or cal check.
For more information regarding the use of this option refer to page 33 of this manual for instruc­tions on set-up and use.
Page 26
User Settings
7. NUMBER OF COPIES PRINTED
7. PRINT SETTINGS EXPLAINED
The FC20 settings allow you to customize your printout to your particular needs. You can choose between the following:
Number of Copies, choose to print 1, 2 or 3 cop­ies of a test when you print.
Print Format, choose to include or not to include the calibration and calibration check information with every printout by selecting the long or short format.
Printer, select between the AP890, DataTrak, or Other to print your results.
From the “Print Settings” display, press the Execute button.
Press the Function button until the display reads “Number of Copies”.
Press the + or- buttons to change the number of copies printed to 1, 2 or 3.
Press the Function button to move to the next set­ting or repeatedly until the display reads Exit. Press the Execute button to save the changes and return to the testing mode.
Page 26
User Settings
7. NUMBER OF COPIES PRINTED
7. PRINT SETTINGS EXPLAINED
The FC20 settings allow you to customize your printout to your particular needs. You can choose between the following:
Number of Copies, choose to print 1, 2 or 3 cop­ies of a test when you print.
Print Format, choose to include or not to include the calibration and calibration check information with every printout by selecting the long or short format.
Printer, select between the AP890, DataTrak, or Other to print your results.
From the “Print Settings” display, press the Execute button.
Press the Function button until the display reads “Number of Copies”.
Press the + or- buttons to change the number of copies printed to 1, 2 or 3.
Press the Function button to move to the next set­ting or repeatedly until the display reads Exit. Press the Execute button to save the changes and return to the testing mode.
Page 27
User Settings
7. PRINTER SELECTION
7. PRINT FORMAT
From the “Print Settings” display, press the Execute button.
Press the Function button until the display reads “Print Format”.
Press the Execute button to select “Long” or “Short” printouts.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Printer selection allows you to select what device your FC20 will communicate with.
From the “Print Settings” display, press the Execute button.
Press the Function button until the display reads “Printer”.
Press the Execute button to select between the Life­loc “AP890” printer, “DataTrak” computer based download software or “Other” printer.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Page 27
User Settings
7. PRINTER SELECTION
7. PRINT FORMAT
From the “Print Settings” display, press the Execute button.
Press the Function button until the display reads “Print Format”.
Press the Execute button to select “Long” or “Short” printouts.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Printer selection allows you to select what device your FC20 will communicate with.
From the “Print Settings” display, press the Execute button.
Press the Function button until the display reads “Printer”.
Press the Execute button to select between the Life­loc “AP890” printer, “DataTrak” computer based download software or “Other” printer.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Page 28
-----------------------------­ Lifeloc Technologies, Inc. FC20 v3.00d Serial No. 01080
-----------------------------­Units: BAC
AUTO TEST # 00857
TEST RESULT: .101 Time: 15:26 Date: 12/12/2000
Last Calibrated: Cal Factor: 0.988 Time: 07:47 Date: 12/11/2000
Last Verified: Result: .101 Time: 11:25 Date: 12/12/2000
_____________________ Subject
_____________________ I.D.
_____________________ Operator
7. SAMPLE PRINT-OUTS
-----------------------------­ Lifeloc Technologies, Inc. FC20 v3.00d Serial No. 01080
-----------------------------­Units: BAC
AUTO TEST # 00857
TEST RESULT: .101 Time: 15:26 Date: 12/12/2000
_____________________ Subject
_____________________ I.D.
_____________________ Operator
Long Printout Short Printout
User Settings
Page 28
-----------------------------­ Lifeloc Technologies, Inc. FC20 v3.00d Serial No. 01080
-----------------------------­Units: BAC
AUTO TEST # 00857
TEST RESULT: .101 Time: 15:26 Date: 12/12/2000
Last Calibrated: Cal Factor: 0.988 Time: 07:47 Date: 12/11/2000
Last Verified: Result: .101 Time: 11:25 Date: 12/12/2000
_____________________ Subject
_____________________ I.D.
_____________________ Operator
7. SAMPLE PRINT-OUTS
-----------------------------­ Lifeloc Technologies, Inc. FC20 v3.00d Serial No. 01080
-----------------------------­Units: BAC
AUTO TEST # 00857
TEST RESULT: .101 Time: 15:26 Date: 12/12/2000
_____________________ Subject
_____________________ I.D.
_____________________ Operator
Long Printout Short Printout
User Settings
Page 29
User Settings
8. DISPLAY CONTRAST
From the “Display Settings” screen press the Exe­cute button.
Press the Function button until the display reads “LCD Contrast”.
Press the + or- buttons to set the contrast to the desired level.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
8. TEST ORDER SEQUENCE
From the “Display Settings” screen, press the Exe­cute button.
Press the Function button until the display reads “Test Order”.
Press the Execute button to choose between “Auto” or “Passive” testing as the default when you turn the unit on.
Press the Function button to move to the next set­ting or repeatedly until the display reads Exit. Press the Execute button to save the changes and return to the testing mode.
Test Order 1 Test Order 2
Auto Test (Default) Passive Test (Default) Manual Test Auto Test Passive Test Manual Test
Page 29
User Settings
8. DISPLAY CONTRAST
From the “Display Settings” screen press the Exe­cute button.
Press the Function button until the display reads “LCD Contrast”.
Press the + or- buttons to set the contrast to the desired level.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
8. TEST ORDER SEQUENCE
From the “Display Settings” screen, press the Exe­cute button.
Press the Function button until the display reads “Test Order”.
Press the Execute button to choose between “Auto” or “Passive” testing as the default when you turn the unit on.
Press the Function button to move to the next set­ting or repeatedly until the display reads Exit. Press the Execute button to save the changes and return to the testing mode.
Test Order 1 Test Order 2
Auto Test (Default) Passive Test (Default) Manual Test Auto Test Passive Test Manual Test
Page 30
User Settings
8. RESULTS FORMAT EXPLAINED
The FC20 displays Auto Test and Manual Test results in a 3 digit BAC format. Passive test results are displayed as POS or NEG.
If however, you are simply trying to qualify whether the subject is within “set” limits, you may not need a quantitative result. Es­pecially if the user is not trained in understanding BAC levels (such as teachers at a high-school dance testing juveniles).
In this instance you may choose to have the FC20 display the results as PASS, WARN, or FAIL (PWF mode).
PASS
8. SETTING RESULTS FORMAT (PASS/WARN/FAIL)
From the “Display Settings” screen, press the Exe­cute button.
Press the Function button repeatedly until the dis­play reads “Results Format”.
Press Execute button, to toggle between “Numeric” or “PWF”.
Press the Function button until the display reads “Pass Level”.
Press + or – button to set the BAC Pass level (eg., .005).
FAIL
Page 30
User Settings
8. RESULTS FORMAT EXPLAINED
The FC20 displays Auto Test and Manual Test results in a 3 digit BAC format. Passive test results are displayed as POS or NEG.
If however, you are simply trying to qualify whether the subject is within “set” limits, you may not need a quantitative result. Es­pecially if the user is not trained in understanding BAC levels (such as teachers at a high-school dance testing juveniles).
In this instance you may choose to have the FC20 display the results as PASS, WARN, or FAIL (PWF mode).
PASS
8. SETTING RESULTS FORMAT (PASS/WARN/FAIL)
From the “Display Settings” screen, press the Exe­cute button.
Press the Function button repeatedly until the dis­play reads “Results Format”.
Press Execute button, to toggle between “Numeric” or “PWF”.
Press the Function button until the display reads “Pass Level”.
Press + or – button to set the BAC Pass level (eg., .005).
FAIL
Page 31
User Settings
8. SETTING RESULTS FORMAT (PASS/WARN/FAIL)
Press the Function button till the display reads “Fail Level”.
Press + or – button to set the BAC Fail level (eg., .040).
Press the Function button until the display reads “Exit”. Press the Execute button to save and return to the testing mode.
NOTE: You cannot set the PASS level higher than the FAIL level.
An alcohol reading higher than the PASS level and lower than the FAIL level will read WARN.
Page 31
User Settings
8. SETTING RESULTS FORMAT (PASS/WARN/FAIL)
Press the Function button till the display reads “Fail Level”.
Press + or – button to set the BAC Fail level (eg., .040).
Press the Function button until the display reads “Exit”. Press the Execute button to save and return to the testing mode.
NOTE: You cannot set the PASS level higher than the FAIL level.
An alcohol reading higher than the PASS level and lower than the FAIL level will read WARN.
Page 32
User Settings-Factory Options
PASSWORD PROTECTION EXPLAINED
If installed, this option limits field personnel to basic breath testing tasks such as taking tests and printing results.
Without a password, operators will not be able to change any setting or calibrate/cal check the unit.
Note: Failure to remember your password will require that the FC20 be sent back to Lifeloc to re­initialize the password.
The FC20 has extended features that can be or­dered from the factory. These are not standard on the unit but only available as a special order items.
The available factory options include Password Protection, Auto Air Blank and Automatic Altitude Compensation for use with dry gas standards.
Contact Lifeloc Technologies for further information on FC20 factory options.
FACTORY OPTIONS EXPLAINED
Page 32
User Settings-Factory Options
PASSWORD PROTECTION EXPLAINED
If installed, this option limits field personnel to basic breath testing tasks such as taking tests and printing results.
Without a password, operators will not be able to change any setting or calibrate/cal check the unit.
Note: Failure to remember your password will require that the FC20 be sent back to Lifeloc to re­initialize the password.
The FC20 has extended features that can be or­dered from the factory. These are not standard on the unit but only available as a special order items.
The available factory options include Password Protection, Auto Air Blank and Automatic Altitude Compensation for use with dry gas standards.
Contact Lifeloc Technologies for further information on FC20 factory options.
FACTORY OPTIONS EXPLAINED
Page 33
If installed, this options inserts an automatic air blank function in front of Auto, Manual, and Passive tests.
An automatic air blank is a test of the ambient air. It checks to ensure that no alcohol is present, which could affect the test result.
If Auto Air Blank is enabled, the FC20 will prompt you to take a sample before testing the subject.
Auto Air Blank is normally utilized to satisfy the re­quirements of a specific testing protocol.
AUTO AIR BLANK EXPLAINED
User Settings-Factory Options
From the “Settings” screen, press the Execute but­ton. The display will read “Time”.
Press the Function button until the display reads “SECURITY SETTINGS”.
Press the Execute button. The unit is now ready to accept a password. Use the + or - buttons to en­ter a number or letter.
Use the Execute button to move to the next digit or press the Function button repeatedly until the dis­play reads “EXIT”. Press the Execute button to re­turn to the testing mode.
SETTING PASSWORD PROTECTION
Page 33
If installed, this options inserts an automatic air blank function in front of Auto, Manual, and Passive tests.
An automatic air blank is a test of the ambient air. It checks to ensure that no alcohol is present, which could affect the test result.
If Auto Air Blank is enabled, the FC20 will prompt you to take a sample before testing the subject.
Auto Air Blank is normally utilized to satisfy the re­quirements of a specific testing protocol.
AUTO AIR BLANK EXPLAINED
User Settings-Factory Options
From the “Settings” screen, press the Execute but­ton. The display will read “Time”.
Press the Function button until the display reads “SECURITY SETTINGS”.
Press the Execute button. The unit is now ready to accept a password. Use the + or - buttons to en­ter a number or letter.
Use the Execute button to move to the next digit or press the Function button repeatedly until the dis­play reads “EXIT”. Press the Execute button to re­turn to the testing mode.
SETTING PASSWORD PROTECTION
Page 34
User Settings-Factory Options
SETTING THE AUTO AIR BLANK
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “Auto Air Blank”.
Press the Execute button to turn the air blank “On” or “Off”.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
AUTOMATIC ALTITUDE COMPENSATION
The FC20 may be purchased with a fully integrat­ed automatic barometric pressure sensor installed.
This sensor greatly simplifies calibration with dry gas standards by automatically adjusting for the effects of altitude.
Page 34
User Settings-Factory Options
SETTING THE AUTO AIR BLANK
From the “Settings” display, press the Execute but­ton.
Press the Function button until the display reads “Auto Air Blank”.
Press the Execute button to turn the air blank “On” or “Off”.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
AUTOMATIC ALTITUDE COMPENSATION
The FC20 may be purchased with a fully integrat­ed automatic barometric pressure sensor installed.
This sensor greatly simplifies calibration with dry gas standards by automatically adjusting for the effects of altitude.
Page 35
Calibration of an FC compares its internal settings to a known alcohol standard, thereby providing it with the baseline from which it can accurately cal­culate the subject’s alcohol level.
You can use most available BAC levels of gas or solution to calibrate your FC20, however, most commonly used are: .100, .080 or .040 BAC.
The FC20 must be between 67º and 100º to cali­brate. (19° - 37° C)
Lifeloc recommends you calibrate your FC20:
Once every 12 months, regardless of how many tests you have performed;
OR, after two failed Calibration Checks;
OR, at intervals specified by your Internal Policies,
Quality Assurance Plan, or State Regulations.
A Calibration Check simply verifies the FC20 was calibrated correctly and is within the acceptable accuracy range. (Commonly called “External Cali­bration Check”, “Accuracy Check”, “Cal Check”, or “Verification”).
Lifeloc recommends you check your FC20:
Once every 30 days;
OR, at intervals specified by your Internal Policies,
Quality Assurance Plan, or State Regulations.
Calibration Settings
CALIBRATION EXPLAINED
Page 35
Calibration of an FC compares its internal settings to a known alcohol standard, thereby providing it with the baseline from which it can accurately cal­culate the subject’s alcohol level.
You can use most available BAC levels of gas or solution to calibrate your FC20, however, most commonly used are: .100, .080 or .040 BAC.
The FC20 must be between 67º and 100º to cali­brate. (19° - 37° C)
Lifeloc recommends you calibrate your FC20:
Once every 12 months, regardless of how many tests you have performed;
OR, after two failed Calibration Checks;
OR, at intervals specified by your Internal Policies,
Quality Assurance Plan, or State Regulations.
A Calibration Check simply verifies the FC20 was calibrated correctly and is within the acceptable accuracy range. (Commonly called “External Cali­bration Check”, “Accuracy Check”, “Cal Check”, or “Verification”).
Lifeloc recommends you check your FC20:
Once every 30 days;
OR, at intervals specified by your Internal Policies,
Quality Assurance Plan, or State Regulations.
Calibration Settings
CALIBRATION EXPLAINED
Page 36
Calibration Settings
There are two types of Calibration/Calibration Check methods:
 Wet Bath Simulator
 Dry Gas
You can calibrate and check your FC20 using ei­ther method. However, you must first set your FC20 to recognize which method, or which ‘Standard Type’, you will be using.
Once you choose the standard type, the FC20 will store that information in memory. You do not have to set it again unless you change to a different method of performing a calibration/calibration check.
Dry Gas Calibration requires that you enter the corrected Standard Value before calibration, based on your altitude or elevation. Using the chart on the outside of the canister, multiply the number next to your elevation by the standard. Example: Denver Colorado elevation 5280 Ft. above sea level. Correction factor .820 x .100 = .082 corrected standard.
If you move to a location at a significantly differ­ent altitude, you will have to change the standard in the FC20.
Note: The FC20 comes from the factory set for wet bath simulator and a .100 solution standard.
CALIBRATION EXPLAINED
Page 36
Calibration Settings
There are two types of Calibration/Calibration Check methods:
 Wet Bath Simulator
 Dry Gas
You can calibrate and check your FC20 using ei­ther method. However, you must first set your FC20 to recognize which method, or which ‘Standard Type’, you will be using.
Once you choose the standard type, the FC20 will store that information in memory. You do not have to set it again unless you change to a different method of performing a calibration/calibration check.
Dry Gas Calibration requires that you enter the corrected Standard Value before calibration, based on your altitude or elevation. Using the chart on the outside of the canister, multiply the number next to your elevation by the standard. Example: Denver Colorado elevation 5280 Ft. above sea level. Correction factor .820 x .100 = .082 corrected standard.
If you move to a location at a significantly differ­ent altitude, you will have to change the standard in the FC20.
Note: The FC20 comes from the factory set for wet bath simulator and a .100 solution standard.
CALIBRATION EXPLAINED
Page 37
Calibration Settings
CALIBRATION DISPLAY CHART
XXX CHECK
XXX
CALIBRATE
CAL
SETTINGS
EXIT
Execute
Function
Function
Function
Cal
Standard
Standard
Type
Cal Period
Check Period
Execute
Function
Function
Function
CALIBRATION
EXIT
Function
Execute
Perform a Cal
Check
Perform a
Calibration
Execute
Page 37
Calibration Settings
CALIBRATION DISPLAY CHART
XXX CHECK
XXX
CALIBRATE
CAL
SETTINGS
EXIT
Execute
Function
Function
Function
Cal
Standard
Standard
Type
Cal Period
Check Period
Execute
Function
Function
Function
CALIBRATION
EXIT
Function
Execute
Perform a Cal
Check
Perform a
Calibration
Execute
Page 38
Calibration Settings
SETTING THE CALIBRATION STANDARD
From the “CALIBRATION” display, press the Exe­cute button. The display now reads “XXX CHECK”.
Press Function button until the display reads “CAL SETTINGS“.
Press the Execute button and the display will read “Cal Standard”
Use the + and – buttons to change the number to the BAC level of standard you will be calibrating to (it should be the same as on the bottle of certi­fied solution or your corrected standard if using a dry gas tank).
Press the Function button to move to the next set­ting or repeatedly until the display reads Exit. Press the Execute button to save the changes and return to the testing mode.
Once you set the standard you do not have to set it again unless you change to a different method of performing a calibration/calibration check.
Cal Standard
0.100
Page 38
Calibration Settings
SETTING THE CALIBRATION STANDARD
From the “CALIBRATION” display, press the Exe­cute button. The display now reads “XXX CHECK”.
Press Function button until the display reads “CAL SETTINGS“.
Press the Execute button and the display will read “Cal Standard”
Use the + and – buttons to change the number to the BAC level of standard you will be calibrating to (it should be the same as on the bottle of certi­fied solution or your corrected standard if using a dry gas tank).
Press the Function button to move to the next set­ting or repeatedly until the display reads Exit. Press the Execute button to save the changes and return to the testing mode.
Once you set the standard you do not have to set it again unless you change to a different method of performing a calibration/calibration check.
Cal Standard
0.100
Page 39
From the “CAL SETTINGS” display, press the Exe- cute button.
Press the Function button until the display reads “Standard Type XXX XXX”.
Press the Execute button to choose between “Dry Gas” and “Wet Bath”.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
SETTING THE STANDARD TYPE
Calibration Settings
The calibration/cal check lock-out allows you to prevent usage of your FC20 if it is not calibrated or cal checked in a specified time period.
Beginning 48 hours before the specified lock-out time, the unit will display “WARNING CAL EXPIR­ING”, “WARNING CHECK EXPIRING” or both.
When the time period has elapsed, the unit will display “CAL EXPIRED”, “CHECK EXPIRED” or both and not allow testing until it is calibrated/cal checked.
You can specify time periods up to 999 days be­tween calibrations or cal checks.
CALIBRATION /CAL CHECK LOCK-OUT EXPLAINED
Page 39
From the “CAL SETTINGS” display, press the Exe- cute button.
Press the Function button until the display reads “Standard Type XXX XXX”.
Press the Execute button to choose between “Dry Gas” and “Wet Bath”.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
SETTING THE STANDARD TYPE
Calibration Settings
The calibration/cal check lock-out allows you to prevent usage of your FC20 if it is not calibrated or cal checked in a specified time period.
Beginning 48 hours before the specified lock-out time, the unit will display “WARNING CAL EXPIR­ING”, “WARNING CHECK EXPIRING” or both.
When the time period has elapsed, the unit will display “CAL EXPIRED”, “CHECK EXPIRED” or both and not allow testing until it is calibrated/cal checked.
You can specify time periods up to 999 days be­tween calibrations or cal checks.
CALIBRATION /CAL CHECK LOCK-OUT EXPLAINED
Page 40
Calibration Settings
SETTING THE CALIBRATION CHECK LOCK-OUT
From the “CAL SETTINGS” display, press the Exe­cute button.
Press the Function button until the display reads “Check Period XXX”.
Press the +/- buttons to enter the number of days until you require the next cal check of your FC20. Choosing “0” turns this setting off.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
SETTING THE CALIBRATION LOCK-OUT PERIOD
From the “CAL SETTINGS” display, press the Exe­cute button.
Press the Function button until the display reads “Cal Period XXX”.
Press the +/- buttons to enter the number of days until you require the next calibration of your FC20. Choosing “0” turns this setting off.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Page 40
Calibration Settings
SETTING THE CALIBRATION CHECK LOCK-OUT
From the “CAL SETTINGS” display, press the Exe­cute button.
Press the Function button until the display reads “Check Period XXX”.
Press the +/- buttons to enter the number of days until you require the next cal check of your FC20. Choosing “0” turns this setting off.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
SETTING THE CALIBRATION LOCK-OUT PERIOD
From the “CAL SETTINGS” display, press the Exe­cute button.
Press the Function button until the display reads “Cal Period XXX”.
Press the +/- buttons to enter the number of days until you require the next calibration of your FC20. Choosing “0” turns this setting off.
Press the Function button to move to the next set­ting or repeatedly until the display reads “Exit”. Press the Execute button to save the changes and return to the testing mode.
Page 41
SIMULATOR SET-UP
Wet Bath Calibration
Pour a bottle of alcohol solution into the simulator glass jar, tighten lid securely.
Turn on the simulator.
The simulator automatically heats the solution to 34°C.
Attach a mouthpiece adapter snuggly to the outlet port on the simulator lid.
F
U
N
C
T
I
O
N
E
X
E
C
U
T
E
Page 41
SIMULATOR SET-UP
Wet Bath Calibration
Pour a bottle of alcohol solution into the simulator glass jar, tighten lid securely.
Turn on the simulator.
The simulator automatically heats the solution to 34°C.
Attach a mouthpiece adapter snuggly to the outlet port on the simulator lid.
F
U
N
C
T
I
O
N
E
X
E
C
U
T
E
Page 42
CALIBRATION (Simulator)
Turn the FC20 ON. Attach a mouthpiece to the back of the unit.
Verify the “Cal Standard” is set to the concentra­tion you will be using when you calibrate.
Verify the “Standard Type” is set to “Wet Bath”.
Press Function button until display reads “CALIBRATION”, then press the Execute button.
Press Function button until display reads “WET CALIBRATE”.
Slide the FC mouthpiece over the other end of the adapter.
Blow through the input tube (or use a calibration pump) to create 1/2” bubbles on the surface of the solution.
After 3 seconds of blowing, press the Execute but­ton to take the sample, while continuing to blow for another 3 seconds.
Stop blowing.
If successful, the display will read “CAL COM­PLETE”.
Disconnect the equipment and wait two minutes before conducting a Cal Check.
Wet Bath Calibration
Page 42
CALIBRATION (Simulator)
Turn the FC20 ON. Attach a mouthpiece to the back of the unit.
Verify the “Cal Standard” is set to the concentra­tion you will be using when you calibrate.
Verify the “Standard Type” is set to “Wet Bath”.
Press Function button until display reads “CALIBRATION”, then press the Execute button.
Press Function button until display reads “WET CALIBRATE”.
Slide the FC mouthpiece over the other end of the adapter.
Blow through the input tube (or use a calibration pump) to create 1/2” bubbles on the surface of the solution.
After 3 seconds of blowing, press the Execute but­ton to take the sample, while continuing to blow for another 3 seconds.
Stop blowing.
If successful, the display will read “CAL COM­PLETE”.
Disconnect the equipment and wait two minutes before conducting a Cal Check.
Wet Bath Calibration
Page 43
CALIBRATION CHECK (Wet Check)
Wet Bath Calibration
Turn the FC20 ON. Attach a mouthpiece to the back of the unit.
Verify the Standard Type is set to Wet Bath.
Press Function button until display reads “CALIBRATION”.
Press Execute button. The display will read “WET Check”.
Slide the FC20 mouthpiece over the mouthpiece adapter.
Blow through the input tube (or use a calibration pump) to create 1/2” bubbles on the surface of the solution.
After 3 seconds of blowing, press the Execute but­ton to take the sample, while continuing to blow for another 3 seconds.
Stop blowing.
Read the result. It should be within +/- .005 BAC of the standard used.
Example: a .100 BAC solution should read be­tween .095 and .105 BAC.
Page 43
CALIBRATION CHECK (Wet Check)
Wet Bath Calibration
Turn the FC20 ON. Attach a mouthpiece to the back of the unit.
Verify the Standard Type is set to Wet Bath.
Press Function button until display reads “CALIBRATION”.
Press Execute button. The display will read “WET Check”.
Slide the FC20 mouthpiece over the mouthpiece adapter.
Blow through the input tube (or use a calibration pump) to create 1/2” bubbles on the surface of the solution.
After 3 seconds of blowing, press the Execute but­ton to take the sample, while continuing to blow for another 3 seconds.
Stop blowing.
Read the result. It should be within +/- .005 BAC of the standard used.
Example: a .100 BAC solution should read be­tween .095 and .105 BAC.
Page 44
DRY GAS SET-UP
Dry Gas Calibration
Screw the regulator onto the dry gas tank.
Attach the mouthpiece adapter to the outlet port on the regulator.
Note: Do not store the dry gas tank with the regu­lator attached. The regulator is not designed as the primary sealing mechanism and your dry gas may leak from the tank.
Page 44
DRY GAS SET-UP
Dry Gas Calibration
Screw the regulator onto the dry gas tank.
Attach the mouthpiece adapter to the outlet port on the regulator.
Note: Do not store the dry gas tank with the regu­lator attached. The regulator is not designed as the primary sealing mechanism and your dry gas may leak from the tank.
Page 45
CALIBRATION (Dry Gas)
Dry Gas Calibration
Verify the “Calibration Standard” is set correctly (refer to pages 36 and 39).
Verify the “Standard Type” is set to Dry Gas.
Turn FC20 on. Attach a mouthpiece to the back of the unit.
Press Function button until display reads “CALIBRATION”.
Press Execute button. The display will read “DRY CHECK”.
Press Function button until display reads “DRY CAL­IBRATE”.
Attach the FC20 mouthpiece to the regulator by sliding it over the mouthpiece adapter, ensuring a snug fit.
Press and hold down the regulator button to deliv­er a gas sample.
After the gas flows for 3 seconds, press and re­lease the Execute button to take the sample.
After 3 more seconds release the valve button to discontinue the flow of gas from the tank.
If successful, the display will read “CAL COM­PLETE”.
Disconnect the equipment and wait two minutes before conducting a cal check.
Page 45
CALIBRATION (Dry Gas)
Dry Gas Calibration
Verify the “Calibration Standard” is set correctly (refer to pages 36 and 39).
Verify the “Standard Type” is set to Dry Gas.
Turn FC20 on. Attach a mouthpiece to the back of the unit.
Press Function button until display reads “CALIBRATION”.
Press Execute button. The display will read “DRY CHECK”.
Press Function button until display reads “DRY CAL­IBRATE”.
Attach the FC20 mouthpiece to the regulator by sliding it over the mouthpiece adapter, ensuring a snug fit.
Press and hold down the regulator button to deliv­er a gas sample.
After the gas flows for 3 seconds, press and re­lease the Execute button to take the sample.
After 3 more seconds release the valve button to discontinue the flow of gas from the tank.
If successful, the display will read “CAL COM­PLETE”.
Disconnect the equipment and wait two minutes before conducting a cal check.
Page 46
CALIBRATION CHECK (Dry Check)
Dry Gas Calibration
Verify the Standard Type is set to Dry Gas.
Turn FC20 ON. Attach a mouthpiece to the back of the unit.
Press Function button until display reads “CALIBRATION”.
Press Execute button. The display will read “DRY CHECK”.
Attach the FC20 mouthpiece to the regulator by sliding it over the mouthpiece adapter, ensuring a snug fit.
Press and hold down the regulator button to deliv­er a gas sample.
After the gas flows for 3 seconds, press and re­lease the Execute button to take the sample.
After 3 more seconds release the regulator button to discontinue the flow of gas from the tank.
The display will read a result that should be within +/-.005 BAC of the standard (if .100 BAC or less).
Check your result against the actual tank standard, not the corrected standard.
Page 46
CALIBRATION CHECK (Dry Check)
Dry Gas Calibration
Verify the Standard Type is set to Dry Gas.
Turn FC20 ON. Attach a mouthpiece to the back of the unit.
Press Function button until display reads “CALIBRATION”.
Press Execute button. The display will read “DRY CHECK”.
Attach the FC20 mouthpiece to the regulator by sliding it over the mouthpiece adapter, ensuring a snug fit.
Press and hold down the regulator button to deliv­er a gas sample.
After the gas flows for 3 seconds, press and re­lease the Execute button to take the sample.
After 3 more seconds release the regulator button to discontinue the flow of gas from the tank.
The display will read a result that should be within +/-.005 BAC of the standard (if .100 BAC or less).
Check your result against the actual tank standard, not the corrected standard.
Page 47
Maintenance
FUEL CELLS
Fuel cells are highly durable sensors that are ca­pable of providing accurate breath alcohol results for years. There are however, a few precautions you should take to make certain that these devices perform for the longest period of time possible.
Use the device. Fuel cells like moisture. It is a good idea to take tests periodically to provide needed moisture to the fuel cell, especially in dry climates. You do not need alcohol, just breath.
Avoid cigarette smoke. Make certain no one is permitted to blow cigarette smoke into the unit. This can damage the fuel cell quickly.
CLEANING
Use of a mild disinfectant cleaner and a soft cloth on the outside of the case is recommended period­ically to keep your unit clean. Do not use alcohol to clean the unit.
BATTERIES
The 4 AA batteries in your FC20 should last for about 160 hours of “on” time which can equate to as many as 6000 to 8000 tests. It is recommended you use high-quality alkaline batteries with your unit.
Page 47
Maintenance
FUEL CELLS
Fuel cells are highly durable sensors that are ca­pable of providing accurate breath alcohol results for years. There are however, a few precautions you should take to make certain that these devices perform for the longest period of time possible.
Use the device. Fuel cells like moisture. It is a good idea to take tests periodically to provide needed moisture to the fuel cell, especially in dry climates. You do not need alcohol, just breath.
Avoid cigarette smoke. Make certain no one is permitted to blow cigarette smoke into the unit. This can damage the fuel cell quickly.
CLEANING
Use of a mild disinfectant cleaner and a soft cloth on the outside of the case is recommended period­ically to keep your unit clean. Do not use alcohol to clean the unit.
BATTERIES
The 4 AA batteries in your FC20 should last for about 160 hours of “on” time which can equate to as many as 6000 to 8000 tests. It is recommended you use high-quality alkaline batteries with your unit.
Page 48
Computer Connection
DATATRAK EXPLAINED
DataTrak™ is a software program developed by Lifeloc, exclusively for use with Lifeloc’s breath testing equipment. Ordered separately, DataTrak software enhances the capabilities of the FC20 by allowing it to communicate with a computer and comes complete with CD, cable and instructions.
Data Download, with DataTrak™ you can download any or all the tests stored in the FC20 in a comma delimited text format for easy import into most spreadsheet or datbase programs.
Printout, with DataTrak™ and your computer you can print the results of any test result stored in the FC20.
Remote Troubleshooting, with just a click of a button, DataTrak can download critical operation parameters of the FC20. This information can then be e-mailed or sent to Lifeloc Technical support for analysis.
Connect the FC20 to the serial port on the PC, using the FC20 computer cable supplied with DataTrak software.
From the “PRINTER” screen choose the DataTrak option.
Refer to the DataTrak Operations Manual to estab­lish communication with the FC20.
DATATRAK SET-UP
Page 48
Computer Connection
DATATRAK EXPLAINED
DataTrak™ is a software program developed by Lifeloc, exclusively for use with Lifeloc’s breath testing equipment. Ordered separately, DataTrak software enhances the capabilities of the FC20 by allowing it to communicate with a computer and comes complete with CD, cable and instructions.
Data Download, with DataTrak™ you can download any or all the tests stored in the FC20 in a comma delimited text format for easy import into most spreadsheet or datbase programs.
Printout, with DataTrak™ and your computer you can print the results of any test result stored in the FC20.
Remote Troubleshooting, with just a click of a button, DataTrak can download critical operation parameters of the FC20. This information can then be e-mailed or sent to Lifeloc Technical support for analysis.
Connect the FC20 to the serial port on the PC, using the FC20 computer cable supplied with DataTrak software.
From the “PRINTER” screen choose the DataTrak option.
Refer to the DataTrak Operations Manual to estab­lish communication with the FC20.
DATATRAK SET-UP
Page 49
Error Messages
ERROR MESSAGES COURSE OF ACTION
Flo w Error Retry &
Blow Steadily
Exhalation not complete or unterrupted. Instruct subject to blow steadily as long as they can.
<1.3L Retest or Try
Manual Test
Breath flo w ended before the subje ct ble w 1.3 liters o f breath. Instruct subj ect to try again or use manual test mode.
Temperature When taking a test, unit is outside of
temperature limits for taking a test.
Low Battery Battery vo ltage is too lo w to take a test.
Replace batteries.
Low Li Battery The internal clock battery is low. Contact Lifeloc
support.
Pump Failure or Pump
Reset Needed
The pump needs reset. Restart unit or follow screen instructions. I f this problem persists contact Lifeloc support.
Excessive Sensor Noise External interference (such as RF noise) has
been detected. Try to mo ve to a di fferent location and try again.
Calibration / Cal
Check expiring
The unit is within 48 hours of the Calibration or Cal Check lockout time. Calibrate or Cal Check the unit.
Calibration / Cal
Check e xpired
Calibrate or Cal Check the unit.
Air Blank Failure The unit detected alcohol on an air blank. Move
to a different location, use a new mouthpiece and try agai n.
Page 49
Error Messages
ERROR MESSAGES COURSE OF ACTION
Flo w Error Retry &
Blow Steadily
Exhalation not complete or unterrupted. Instruct subject to blow steadily as long as they can.
<1.3L Retest or Try
Manual Test
Breath flo w ended before the subje ct ble w 1.3 liters o f breath. Instruct subj ect to try again or use manual test mode.
Temperature When taking a test, unit is outside of
temperature limits for taking a test.
Low Battery Battery vo ltage is too lo w to take a test.
Replace batteries.
Low Li Battery The internal clock battery is low. Contact Lifeloc
support.
Pump Failure or Pump
Reset Needed
The pump needs reset. Restart unit or follow screen instructions. I f this problem persists contact Lifeloc support.
Excessive Sensor Noise External interference (such as RF noise) has
been detected. Try to mo ve to a di fferent location and try again.
Calibration / Cal
Check expiring
The unit is within 48 hours of the Calibration or Cal Check lockout time. Calibrate or Cal Check the unit.
Calibration / Cal
Check expired
Calibrate or Cal Check the unit.
Air Blank Failure The unit detected alcohol on an air blank. Move
to a different location, use a new mouthpiece and try agai n.
Page 50
Warranty
The FC20 comes with a one year limited parts and labor warranty, effective on the date of purchase by the end-user.
The Warranty covers:
 Parts and labor on covered repairs  Software updates as applicable  Airfreight back to the customer after the unit is
repaired (U.S. only)
The Warranty does not cover:
 Freight to the Lifeloc factory  Misuse, abuse, negligence or accidents
FACTORY LIMITED WARRANTY ON NEW UNITS
EXTENDED SERVICE PLANS
Extended service plans are available for your FC20. These provide complete coverage for an additional year at a reasonable cost and include free Factory Diagnostic Checks. Call Lifeloc for additional details, or apply online.
SERVICE
If your FC20 should require repairs or mainte­nance Lifeloc is there for you. Just an email or phone call will put you in contact with our technical support personnel.
Lifeloc typically repairs units within 4 working days from the time it is received.
Page 50
Warranty
The FC20 comes with a one year limited parts and labor warranty, effective on the date of purchase by the end-user.
The Warranty covers:
 Parts and labor on covered repairs  Software updates as applicable  Airfreight back to the customer after the unit is
repaired (U.S. only)
The Warranty does not cover:
 Freight to the Lifeloc factory  Misuse, abuse, negligence or accidents
FACTORY LIMITED WARRANTY ON NEW UNITS
EXTENDED SERVICE PLANS
Extended service plans are available for your FC20. These provide complete coverage for an additional year at a reasonable cost and include free Factory Diagnostic Checks. Call Lifeloc for additional details, or apply online.
SERVICE
If your FC20 should require repairs or mainte­nance Lifeloc is there for you. Just an email or phone call will put you in contact with our technical support personnel.
Lifeloc typically repairs units within 4 working days from the time it is received.
Page 51
Notice
The FC20 is a professional device designed to be used by trained operators in conjunction with a specified, periodic maintenance and calibration/ calibration check regimen. Use by untrained opera-
tors or without periodic calibration or calibration checks may result in invalid results or incorrect inter­pretation of results.
DO NOT DRINK AND DRIVE. Lifeloc strongly rec­ommends that no vehicle be operated after alco­hol consumption. Even small quantities of alcohol can result in driving impairment.
Page 51
Notice
The FC20 is a professional device designed to be used by trained operators in conjunction with a specified, periodic maintenance and calibration/ calibration check regimen. Use by untrained opera-
tors or without periodic calibration or calibration checks may result in invalid results or incorrect inter­pretation of results.
DO NOT DRINK AND DRIVE. Lifeloc strongly rec­ommends that no vehicle be operated after alco­hol consumption. Even small quantities of alcohol can result in driving impairment.
Page 52
W E RE ON THE WEB! WWW.LIFELOC.COM
Mailing address 12441 West 49th Avenue
Unit 4
Wheat Ridge, CO 80033
Telephone (303) 431-9500
Toll Free (in US) (800) 722-4872
Facsimile (303) 431-1423
E-mail service@lifeloc.com
sales@lifeloc.com
To receive a pricelist or brochure, or to find your nearest Distributor or Sales Representative, please call the Lifeloc Sales Depart­ment.
LIFELOC SALES
All Lifeloc products,, including mouthpieces, calibration equip­ment, dry gas, alcohol solution, etc., are in stock and are availa­ble from our factory in Denver, Colorado.
Order by fax, phone or online at www.lifeloc.com
Page 52
W E RE ON THE WEB! WWW.LIFELOC.COM
Mailing address 12441 West 49th Avenue
Unit 4
Wheat Ridge, CO 80033
Telephone (303) 431-9500
Toll Free (in US) (800) 722-4872
Facsimile (303) 431-1423
E-mail service@lifeloc.com
sales@lifeloc.com
To receive a pricelist or brochure, or to find your nearest Distributor or Sales Representative, please call the Lifeloc Sales Depart­ment.
LIFELOC SALES
All Lifeloc products,, including mouthpieces, calibration equip­ment, dry gas, alcohol solution, etc., are in stock and are availa­ble from our factory in Denver, Colorado.
Order by fax, phone or online at www.lifeloc.com
Loading...