Liebert PS096A User Manual

Process Fluid Chiller Designed for Marconi Medical
Keeping Business in Business
Eclipse, Polaris, PD100/150, PD200, Apollo
• Dedicated, equipment-matched cooling supplies the proper temperature and water flow year round.
• Factory-tested, piped, wired and charged for hassle-free startup and operation.
• Clean, non-ferrous cooling loop, required for sensitive medical equipment.
• Local service and parts for quick and knowledgeable maintenance and service.
• Optional remote monitoring systems keep you informed of chiller operation.
• UL/CSA certified.
• ISO 9001 certified company.
Eclipse/Polaris Specifications
Fluid Specified Glycol
Max Heat Dissipation—BTUH (kW) 75,000 (22)
Fluid Temperature Range—°F (°C) 50-55 (10-13)
Fluid Flow Rate—GPM (l/m) 15.0 (57.0)
Fluid Pressure Drop—PSI (kPa) 65.0 (448.0)
Process Fluid Chiller Specifications
Model Number PS096A PS096A PS096A
Fluid Circulated Water Ethylene Glycol Propylene Glycol
Net Cooling Capacity—BTUH (kW) 83,400 (24.4) 79,000 (23.1) 75,600 (22.2)
Leaving Fluid Temperature—°F (°C) 50 (10) 50 (10) 50 (10)
Fluid Flow Rate—GPM (l/m) 15 (57) 15 (57) 15 (57)
Eclipse/ Polaris Pressure Drop—PSI (kPa) 65 (448) 65 (448) 65 (448)
Chiller Pressure Drop—PSI (kPa) 7.6 (52.4) 10.2 (70.3) 12.0 (82.7)
Pump Total Head—PSI (kPa) 86.5 (596.4) 86.5 (596.4) 86.5 (596.4)
Head Available for Piping Circuit—PSI (kPa) 13.9 (95.8) 11.3 (77.9) 9.5 (65.5)
Pump HP (kW) 2 (1.5) 2 (1.5) 2 (1.5)
Process Fluid Chiller Designed for Marconi Medical Eclipse, Polaris, PD100/150, PD200, Apollo 1
D
B
A
C
Model No.
PS096A 77 (1956) 40 (1016) 38.5 (978) 5.5 (140) 750 (340) 460, 3-phase; 208/230, 3-phase
PS096A 77 (1956) 40 (1016) 38.5 (978) 5.5 (140) 750 (340) 460, 3-phase; 208/230, 3-phase
A - Width
Inches (mm)
Standard Features
• Indoor/outdoor cabinet
• Hermetic compressor
• Refrigerant dehydrator
• Expansion valve
• Refrigerant receiver
• Stainless steel fluid reservoir expansion tank
• Pressure relief valve
• Hot gas bypass valve capacity control
• Stainless steel pump
Optional Features
• Refrigerant gauges and reservoir heater
B - Height
Inches (mm)
C - Depth
Inches (mm)
D
Weight
Lbs. (kg)
Voltage
(60 Hz)
Accessories
• Emergency water bypass
• Enviro-source emergency bypass
• Water level control and alarm system
•Fluid pressure gauge
•Check valve
• Dial thermometer
• Relief valve
• Circuit Setter
Required Options
• Multistage, high-head pump
• Part #154650G1—Wallbox with flow meter, filter, thermometers, flow switch and ball valves
Note: Contact Liebert Application Engineering for gradient upgrades or Medical Marconi systems without heat exchangers.
1050 Dearborn Drive P.O. Box 29186 Columbus, OH 43229 Phone: 800.877.9222 Fax: 614.841.6022
Liebert Web Site—http://www.liebert.com
SL-16552 (07/00)
Process Fluid Chiller Designed for Marconi Medical Eclipse, Polaris, PD100/150, PD200, Apollo 2
Via Leonardo Da Vinci 8 Zona Industriale Tognana 35028 Piove Di Sacco Italy Phone: 39.049.9719.111 Fax: 39.049.5841257
23/F Allied Kajima Bldg. 138 Gloucester Road Wanchai Hong Kong Phone: 852.2.572.2201 Fax: 852.2.572.1182
©2000 Liebert Corporation. All rights reserved throughout the world. Specifications subject to change without notice. ®Liebert and the Liebert logo are registered trademarks of Liebert Corporation. All names referred to are trademarks or regisered trademarks of their respective owners.