What side-effects can the use of LIBAIRATOR® have?
Even whole-day use of the device is safe, but a new device in its first 24 hours of operation produces more salt particles than
usual. Therefore, for the first 24 hours of its operation place the device at least 2 metres from you. After 24 hours of
operation you can place it to the generally recommended distance of 40 to 60 cm. If you do not keep to this rule, you
may experience a stinging feeling, swelling of the nasal mucosa, slight bleeding when you blow your nose, and
headache.
Use from a very close distance (intensive therapy) may cause a mild headache, mild sharp pain in the frontal sinus, the nasal
cavity and the temple. In this case, place the device further away or point it away from yourself.
As a general rule, if you feel that you have inhaled too much salt, place the device further away or point it away from yourself.
Be assured, a salty taste in the mouth and the feeling of salt in the nose is normal. However, be alert to your body's reactions.
In the first few days of using the device, you may experience unusually strong mucus clearance, mild huskiness and coughing.
This phenomenon is attributed to the dislocation of stuck mucus, and should diminish in a few days’ time.
If, before starting the treatment, you suffer from symptoms with a high amount of mucus clearance, restrict intensive use to the
morning and early afternoon hours to avoid strong mucus clearance during the night hours that could cause sleeping
difficulties.
If even after the first few days of use you experience too strong mucus clearance in the morning or you find the eventual
sneezing-like side-effect disturbing, place the device further away.
Checking the device and troubleshooting .
For your personal safety, the device is equipped with internal lighting. Every time you power up the device - under
appropriate lighting conditions - check if the salt in the drum has discoloured, by looking through the slits in the front
(Figure C). If in the illuminated area you see not only shades of white, but also another colour (the colour of the salt
looks different, brownish), it indicates an irreversible, moderately harmful corrosion. You need to unplug the device and
contact a service centre. Also stop using the device if, turned face-down, grains of salt fall out through the grille.
If you do not hear the characteristic “tzitzing” sound of the device or if its vibrations are abnormally strong, it is likely caused by
the clumping of salt grains, caused by not following the directions of the device. The device thus becomes ineffective. In such
a case, leave the device powered up for a few hours so that the salt becomes dry. Shake the device gently from time to time
User manual web EN ● Version: V03 ● 26/02/2018 ● Libairator LIB-111-W/B ● Page 12/16