Leybold vacuum SC 5 D, SC 15 D, SC 30 D Operating Instructions Manual

Contents
Section
Page
General safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 Order Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3 Technical Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3.1 Motor Related Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.3.2 Performance data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.3.3 Outside dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.1 Check the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2 Operating environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.3 Setting up the pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.4 Electrical Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4.1 Single-phase motors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4.2 Three-phase motors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
2.5 Connection to vacuum system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.1 Start up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.2 Pumping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.3 Pumping vapor (air flush operation) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.4 Shut down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.1 Table - Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.2 Leybold Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
EC Conformance Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Figures
The first digit of a reference, (1/2) for example, is the figure number; the second digit is its item number in this figure.
We reserve the right to modify the design and the specified data. The illu­strations are not binding.
2
GA01423_0102 - 11/2004
Safety information
General safety information
Indicates procedures that must be strictly observed to prevent
Warning
hazards to persons.
Indicates procedures that must be strictly observed to prevent damage to, or destruction of the appliance.
The Leybold Oil-free Scoll vacuum pump is designed for safe and efficient operation when used properly and in accordance with this manual. It is the responsibility of the user to carefully read and strictly observe all safety pre­cautions described in this section and throughout the manual. This product must be operated and maintained by trained personnel only. Consult local, state, and national agencies regarding specific requirements and regulati­ons. Address any further safety, operation and/or maintenance questions to your nearest Leybold Vacuum office.
Failure to observe the following precautions could result in serious personal injury.
Never pump toxic, explosive, flarmmable, corrosive gases,
chemicals, solvents or powders.
Flowing substances, explosion or fire can cause bodily injury.
Install in an area without explosive or flammable substances.
Oherwise, it can cause explosion or fire.
Wiring must be done by a qualified electrician. Otherwise,
electric shock may occur.
Prevent short-circuit with ground fault circuit interuptor (GFCI)
of proper capacity.
Failure to do so can result in electric shock or fire.
Ask specialist to perform repairs. If not , it can cause failure,
damage or shorter lifetime. Trained customers may perform maintenance themselves.
Be sure to install a main disconnect switch.
Failure to do so can result in damage, fire, or injury.
Be sure to disconnect the electric source before wiring or
inspection. Failure to do so can result in electric shock or damage by the rotating fan.
Be sure to ground. Failure to ground can result in electric
shock or fire.
Caution
Warning
Cut off electric Source
Be sure to ground
GA01423_0102 - 11/2004
3
Safety information
Do not touch
Avoid moisture
Do not alter
With a thermal
protector
Install in a proper
location
Use designated
temperature
Conduct periodic
maintenance
Maintenance after
pump is cool
Start or stop after
closing isolation
valve
Never put fingers or foreign matter into the pump's inlet or the clearance of cooling fins. If done, it can cause injury.
Install in an area which is not exposed to moisture such as rain or steam. Exposure to moisture can cause eIectric shock.
Never alter the equipment. Modifications can damage the equipment of shorten its lifetime.
Be sure to check Wiring diagram before connecting a power supply. Failure to do so can result in damage.
Pumps with single-phase motors have an internal thermal pro­tector. This motor restarts without warning after the protector trips.
Be sure to switch off the electric power before maintenance or inspection. Failure to do so can result in electric shock or bodily injury by rotating objects.
Install in a proper location.
Install on a level foundation (an inclination of less than 5°)
in a location free of dirt or dust from iron, stone or wood.
in a location free of corrosive gas.
Failure to install in a proper location can result in damage, decrease in performance or shorter lifetime.
Use at ambient temperatures of 5 - 40 °C (during operation). Operating outside of this range can cause damage or shorten lifetime.
Do not operate when the fan cover is removed. Do not operate, when the terminaI cover is removed. Do not operate the single-phase pump when its condenser cover is removed.
Conduct periodic maintenance and inspections. Failure to do so can cause damage or shorter lifetime.
Do the maintenance after the pump has fully cooled down. Performing maintenance on a hot pump can cause burns.
Be sure to close the isolation valve between pump and vacu­um system (chamber) during startup and shutdown.
If this valve is open, debris attached to inside of pump can be drawn back into the vacuum chamber.
4
GA01423_0102 - 11/2004
Description
SC 5 D
Cover
Shock
Cover
Electric
Te rminal Condenser
SC 15 D
Single phase
SC 15 D
Three phase
SC 30 D
Single phase
SC 30 D
Three phase
Always keep warning stickers clean. If they become dirty or lost, replace them with new ones. If you need stickers, contact Leybold.
Fig. 1 Where to attach warning stickers
GA01423_0102 - 11/2004
5
Description
1 Description
1.1 Name of each section
SC 5 D
Fig. 2 Name of each section
6
GA01423_0102 - 11/2004
Description
SC 15 D
SC 30 D
Fig. 3 Connections and controls (SC 15 D/30 D)
GA01423_0102 - 11/2004
7
Description
Fig. 4 Connections and controls (SC 30 D, 3-phase)
SC 30 D
1.2 Order Information
Part No.
SC 5 D, single-phase motor, cable with Schuko plug 133 000 SC 15 D, single-phase motor, cable with Schuko plug 133 001 SC 30 D, single-phase motor, cable with Schuko plug 133 002
SC 15 D, three-phase motor 133 003 SC 30 D, three-phase motor 133 004
SC 5 D, single-phase motor, cable with NEMA plug 133 100 SC 15 D, single-phase motor, cable with NEMA plug 133 101 SC 30 D, single-phase motor, cable with NEMA plug 133 102
8
GA01423_0102 - 11/2004
Description
1.3 Technical Data
Model SC 5 D
50 Hz 60 Hz
SC 15 D
50 Hz 60 Hz
SC 30 D
50 Hz 60 Hz
Displacement (swept volume)
l · min
-1
m
3
· h
-1
cfm
90
5
3.1
108
6.48
3.8
250
15
8.8
300
18
10.6
500
30
17.6
600
36
21.2
Pumping speed
1
l · min
-1
m
3
· h
-1
cfm
80
4.8
2.8
93
6.2
3.6
210
13.6 8
230
16.4
9.6
420
26.3
15.4
500
31.5
18.5
Ultimate pressure Pa
mbar
To rr
5
0.05
0.037
1.6 0.016 0.012
1.0
0.01
0.0075
Leak tightness
3
mbar . l . s
-1
1.0 · 10
-6
1.0 · 10
-4
1.0 · 10
-4
Max. inlet pressure Atmospheric pressure
Ambient operating temperature range
°C 5 - 40 5 - 40 5 - 40
Inlet connection NW 25 25 40
Outlet connection NW 16 16 25
Cooling system Air-cooling
Protection class IP 20 20 20
Class of equipment Class 1
Others With hour meter and air flush
Water vapor handling at 25 °C, humidity 60 %, with opened air flush
4
g/day
l · min
-1
5
9
25
10
25
10
GA01423_0102 - 11/2004
9
Description
1.3.1 Motor Related Data
Motor Related Data SC 5 D
Type Single-Phase Induction / Motor 4P / Totally-Enclosed
Insulation Class B IP 44 Thermal Protector TP 212 - Automatic Reset Type
Power consumption W 150
Voltage V 100 115 200 230
Rated current 50 Hz A 2.6 1.3 1.6
60 Hz A 2.1 2.2 1.1 1.1
Speed at 50 Hz min
-1
1430 1430 1440
60 Hz min
-1
1730 1740 1730 1740
Noise level
2
at 1m dB(A) ≤ 52
with air flush ON (opened) dB(A) ≤ 57
Direction of inlet Vertical
Dimensions (L x W x H) mm 308 x 214 x 225
Weight kg 14
Order information 133 100 133 000
Motor Related Data SC 15 D
Type Single-phase Induction motor 4P
To tally-Enclosed Insulation Class B / IP 44 Thermal Protector TP 212 Automatic Reset Type
3-phase Induction motor5 4P Totally-Enclosed Insulation Class B / IP 44
Power consumption kW 0.4 0.4
Voltage V 100 115 200 230 200 208 230 380 415 460
Rated current 50 Hz
60 Hz
A A
4.8 *
4.8
4.3
2.6
2.4
2.8
2.4
1.6 * * 0.9 1.0
1.9 1.9 1.8 * * -1.0
Speed at 50 Hz
60 Hz
min
-1
min
-1
1440 * 1710 1740
1430 1450 1700 1730
1420 * * 1440 1440 1660 1660 1690 * * * 1720
Noise level
2
at 1m dB(A) ≤ 58
with air flush ON (opened) dB(A) ≤ 66
Direction of inlet Vertical
Dimensions (L x W x H) mm 400 x 252 x 336 370 x 252 x 336
Weight kg 25 23
Order information 133 101 133 001 133 003
10
GA01423_0102 - 11/2004
Description
Motor Related Data SC 30 D
Type Single-phase Induction motor 4P
To tally-Enclosed Insulation Class B / IP 44 Thermal Protector TP 212 Automatic Reset Type
3-phase Induction motor5 4P Totally-Enclosed Insulation Class B / IP 44
Power consumption kW 0.6 0.6
Voltage V 100 115 200 230 200 208 230 380 415 460
Rated current 50 Hz
60 Hz
A A
8.5 * 10 8.6
4.3 3.9
4.8
4.0
2.7 * * 1.57 1.63 -
2.8 2.6 2.5 * * 1.47
Speed at 50 Hz
60 Hz
min
-1
min
-1
1430 * 1660 1720
1430 1450 1690 1730
1460 * * 1470 1470 ­1740 1740 1760 * * 1770
Noise level
2
at 1m
with air flush ON (opened)
dB(A) dB(A)
62 70
60 68
Direction of inlet Horizontal (Laterally fitted)
Dimensions (L x W x H) mm 443 x 328 x 372 (443 x 298 x 397) 372 x 328 x 372 (372 x 298 x 397)
Weight kg 44 38
Order information 133 102 133 002 133 004
*Can not operate at 115V/50 Hz, 208V/50Hz, 230V/50Hz, 460V/50 Hz, 380V/60 Hz, or 415V/60 Hz.
Note 1 : Pumping speed remain the same when air flush is ON (opened) and OFF (closed). Note 2 : Noise level is measured at ultimate pressure in an anechoic room. Note 3 : Leak tightness is measured when pump is stopped and air flush is OFF (closed). Note 4 : Air Flush is OFF (closed) when pump is delivered. Note 5 : Three Phase Motor does not have internal thermal protection. Be sure to install thermal protection like a breaker.
1.3.2 Performance data
Pa
Pumping Speed (l/min)
10
100
1000
0
10
10
1
10
-1
10
2
10
3
10
5
1
Inlet Pressure
10
4
mbar
-2
10
10
-1
10
-3
10
0
10
1
10
3
10
2
Torr
-2
10
10
-1
10
-3
10
0
10
1
10
3
10
2
SC 30 D
60 Hz 50 Hz
SC 15 D
60 Hz 50 Hz
SC 5 D
60 Hz 50 Hz
Fig. 5 Pumping speed characteristics
GA01423_0102 - 11/2004
11
Description
1.3.3 Outside dimensions
SC 5 D
Fig. 6 Dimensional drawing for the SC 5 D
12
GA01423_0102 - 11/2004