Level 1 H-1200 115V, H-1200 230V User manual

l
Level 1® H-1200 Fast Flow Fluid Warmer
With: H-31, Version B, Air Detector/Clamp
< H-1200 115V
®
< H-1200 230V
6 2
s
l
Level 1® H-1200 Fast Flow Fluid Warmer
With: H-31, Version B, Air Detector/Clamp
<
H-1200 115 V
<
H-1200 230 V
OPERATOR’S MANUAL
®
P/N 40-6917-51A
s
SECTION X • Title
ii
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
General Information
SECTION X • Title
General Information
Level 1® H-1200 Fast Flow Fluid Warmer
Part Number: 40-6917-51A
This revision supercedes all previous revisions.
Under copyright laws, this manual may not be reproduced in any form, in whole, or in part, without prior written permission of Smiths Medical ASD, Inc. (Smiths Medical).
Smiths Medical and Level 1 design marks and Level 1 are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U. S. Patent and Trademark oce and certain other countries. All other names and marks mentioned are the tradenames, trademarks, or service marks of their respective owners.
Every eort has been made to ensure that the information in this manual is accurate and details provided are correct at the time of printing. The company, however, reserves the right to improve the equipment shown.
Mention of third-party products is for informational purposes only and constitutes neither an endorsement nor a recommendation. Smiths Medical assumes no responsibility with regard to the performance or use of these products.
For further information, please call your local Smiths Medical distributor or Smiths Medical direct at 1 800 258 5361 or +1 614 210 7300.
©2014 Smiths Medical.
All rights reserved.
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
iii
Contents
Contents
1 About this Manual 1 2 Description 3 3 Indications for Use 4 4 Important Safety Information 5
CONTRAINDICATIONS 5 WARNINGS 5 CAUTIONS 8
5 Out of the Box – Assembly 9
Step 1 – Verify Components of the Level 1® H-1200 Fast Flow Fluid Warmer 10 Step 2 – Assemble I.V. Pole to the Warming Unit 11 Step 3 – Install the Pressure Chambers 12 Step 4 – Attach the I.V. Bag Hanger 13 Step 5 – Disinfect the Recirculating Solution Reservoir 13 Step 6 – Preliminary Preparation 14 Step 7 – Connect the Pneumatic Tubing 14 Step 8 – Install the Level 1® H-31, Version B, Air-Detector/Clamp 15 Step 9 – Perform Electrical Safety Tests 17
6 Principle of Operation 19
Fluid Warming 19 Pressurized Fluid Delivery 19 Air Detection/Clamping 19
7 Controls and Displays 20
Fluid Warmer Power and Alarm Test Panel 21 Fluid Warmer Display Panel 22 Reservoir Level Display 22 Pressure Chamber Control Panel 24 Interlocks 25
8 Operation 27
Modes of Operation 28
OFF Mode 28 ON/Automatic Operation Mode for Fluid Warmer 28 Alarm Test Mode 28 Over Temperature Test Mode 29 Temperature Display 29 Check Disposables Mode 29 Add Recirculating Solution Mode 30 Over Temperature Alarm Mode 30 Power ON Test for Air Detector/Clamp 31 Automatic Operation Air Detector/Clamp 31 Check Tubing Mode 31 Air Detected/Clamped Mode 32 Pressure Display 32 Pressurized Mode 32 Unpressurized Mode 32
iv
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
9 Operating Instructions 33
WARNINGS 34
9.1 Set Up for Use 34
A – Install the Disposable Administration Set 34 B – Prime the Disposable Administration Set 35 C – Prime the Patient Line 37 D – Test the Audible and Visual Alarms 37 E – Test the Air Detector/Clamp 38
9.2 Use of the Fluid Warmer 39
Step 1—Load the Pressure Chambers 40 Step 2—Pressurize the Pressure Chambers 40 Step 3—Make Patient Connection 40 Step 4—Replace the Gas Vent/Filter Assembly 40 Step 5—Change the Fluid Bag 41
9.3 Replace the Gas Vent/Filter Assembly 41
9.4 Activated Alarms 42
9.5 After Use 44
10 Troubleshooting 45
General Troubleshooting Guide 45 Slow Flow Rate Troubleshooting 46
11 Testing 47
Add Recirculating Solution Alarm 47 Check Disposables Alarm 47 Over Temperature Test 48 Fluid Warmer Alarm Signal Test 48 Performance Testing 48
Cold Start Test 48 Calibration Test 49 Calibration Test with DSTA-40 49 Proper Calibration of Recirculating Solution Temperature 49
Periodic Electrical Testing 50
Earth Leakage 50 Ground Continuity 50
12 Maintenance 51
Maintenance Performed Prior to Every Use 51
Clean the Exterior 51 General Inspection 52
Maintenance Performed Every 30 Days 52
Lubricate O-Ring Seals 52 Change Recirculating Solution with Distilled Water 53
Maintenance Performed Every 12 Months 53
Disinfect the Recirculating Solution Reservoir 53 Change Recirculating Solution with a
0.3% Hydrogen Peroxide/Distilled Water Solution 53 Change O-Rings 54 Clean Fan Filter 54 Inspect Air Detector/Clamp 54 Inspect the Reservoir and Float Switch Assembly 54 Testing Fluid Warmer Operation 54
Maintenance and Calibration Log 55
Scheduled Maintenance and Calibration Checklist 55
Contents
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
v
Contents
13 Limited Warranty 56 14 Service 58
Non-Warranty Work 58 Additional Documentation 58 Disposal Information 58 Service Contacts 59
15 Specications 60
System Specications 61
Standard Compliance 61 Physical 61 Environmental 61 Thermal 61 Electrical 61
Electromagnetic Environment Recommendations 61 Disposable Administration Set Specications 61
DI-50 61 D/DI-60HL 61 D/DI-70 62 D/DI-100 62 D/DI-300 62
16 Symbols 63 Index 66
vi
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
SECTION 1
SECTION 1 • About this Manual
1 About this Manual
About this Manual
WARNING: These instructions contain important information for safe use of the product. Read the entire operator’s manual, including Warnings and Cautions, before using this product. Failure to properly follow warnings, cautions and instructions could result in death or serious injury to the patient.
This Operator’s Manual describes the set-up, use, and maintenance of:
Level 1® H-1200 Fast Flow Fluid Warmer
Level 1® H-1000 Fast Flow Fluid Warmer with H-2 Pressure
Chambers (H-1025)
Level 1® H-1000 Fast Flow Fluid Warmer with H-2 Pressure
Chambers and H-31, Version B, Air Detector/Clamp
The manual is intended for use by individuals trained in the healthcare and biomedical professions.
This operator’s manual is also for users of the H-1000 Fast Flow Fluid Warmer. All references to the “Level 1® H-1200 Fast Flow Fluid Warmer” apply to the H-1000 Fast Flow Fluid Warmer except where indicated by the symbols dened in the following table.
Symbol Description
Appears in the margin to identify information that only applies to the H-1200 Fluid Warmer and the H-1000 Fluid Warmer equipped with the H-31, Version B, Air Detector/Clamp
Appears in the margin to identify information that only applies to the H-1000 Fluid Warmer (not equipped with the H-31, Version B, Air Detector/Clamp)
Conventions Used in this Manual
This manual uses the following text and text conventions:
Convention Description
CONTRAINDICATION
WARNING
CAUTION
A Contraindication statement alerts the user to conditions when the device should not be used.
A Warning statement alerts the user to conditions that may cause death or serious injury to the patient or user.
A Caution statement alerts the user to conditions that may cause malfunction, failure, or damage to the device.
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
1
SECTION 1 • About this Manual
This manual is organized into the following sections:
2 and 3 Description and Indications for Use
These sections provide the purpose and indications for use of the Level 1® H-1200 Fast Flow Fluid Warmer.
4 Important Safety Information
Lists the Contraindications, Warnings, and Cautions associated with the use of the Level 1® H-1200 Fast Flow Fluid Warmer.
5 Out of the Box—Assembly
Guides the user through the installation of the Level 1® H-1200 Fast Flow Fluid Warmer and the Level 1® H-31, Version B, Air Detector/Clamp.
6 Principle of Operation
Provides a functional description of the Level 1® H-1200 Fast Flow Fluid Warmer.
7 Controls and Displays
Provides a description of the function and purpose of the controls, displays, and indicators for the Level 1® H-1200 Fast Flow Fluid Warmer.
8 Operation
Describes Operation, Indicator, and Alarm modes of the Level 1® H-1200 Fast Flow Fluid Warmer.
9 Operating Instructions
Describes the Set Up, Use, and Alarm modes of the Level 1® H-1200 Fast Flow Fluid Warmer.
10 Troubleshooting
Contains information on troubleshooting the Level 1® H-1200 Fast Flow Fluid Warmer. This section also details troubleshooting slow uid ow rates.
11 Testing
Describes Operational, Performance, and Electrical Tests that are used to verify the proper operation of the Level 1® H-1200 Fast Flow Fluid Warmer.
12 Maintenance
Regular maintenance procedures for every use, 30-day, and 12-month intervals are covered in this section.
13 Limited Warranty
Describes the Limited Warranty and its provisions.
14 Service
Explains Non-Warranty Work as well as listing Service Contacts.
15 Specications
Provides physical, environmental, and electrical specications of the Level 1® H-1200 Fast Flow Fluid Warmer.
16 Symbols
Lists the symbols and their denitions used with the Level 1® H-1200 Fast
Flow Fluid Warmer.
2
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
SECTION 2
SECTION 2 • Description
2 Description
Description
®
The Level 1 uid warmer with pressure chambers, air detection, and automatic clamping capability. I.V. uid and/or blood products are warmed through the use of a sealed heat exchanger through which a recirculating solution ows. Pressure Chambers apply pressurization and deliver the uids at a fast ow rate. This non-invasive method employs single-use, disposable administration sets that include a Gas Vent/Filter Assembly and Heat Exchanger.
The Air Detector/Clamp monitors for the presence of air in the disposable Gas Vent/Filter Assembly. When air is detected in the Gas Vent/Filter Assembly, the Air Detector/Clamp closes o the patient line and alerts operators to the presence of air with audible and visual alarms. An ultrasonic signal continually passes through the uid lled Gas Vent/Filter Assembly. As a bolus of air displaces the uid in the Gas Vent/Filter Assembly, the ultrasonic signal is broken and the clamp closes, stopping the air before it enters the patient line. Audible and visual alarms are activated, notifying the user that the uid ow has stopped. Clearing the bolus of air and restoring the uid ow are quickly accomplished without disconnecting from the patient.
H-1200 Fast Flow Fluid Warmer (Fluid Warmer) is an I.V.
Disposable Administration Sets
The installation, set up, and replacement of Level 1® Fast Flow I.V. Disposable Administration Sets (Disposable Sets) follows a four­step sequence that corresponds to numbered blocks on the device. Disposable Sets available for use on the Level 1® H-1200 Fast Flow Fluid Warmer are listed below.
DI-50
D-60 / DI-60HL
D-70 / DI-70
D-100 / DI-100
D-300 / DI-300
D-series Disposable Sets are for use in the U.S.A. DI-series Disposable Sets are for use in markets outside of the U.S.A.
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
3
SECTION 3 • Indications for Use
SECTION 3
Indications for Use
The Level 1® H-1200 Fast Flow Fluid Warmer (Fluid Warmer) provides a rapid ow of warmed uids, such as crystalloid or blood product, including red blood cells, as volume replacement for patients suering from blood loss due to trauma or surgery.
The Fluid Warmer provides fast ow of warmed uid to re-warm patients during surgery by trained medical personnel.
3 Indications for Use
4
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
SECTION 4 • Important Safety Information
SECTION 4
4 Important Safety Information
CONTRAINDICATIONS
WARNINGS
Important Safety Information
This section covers information for prescribers and guidelines for safe
®
use of the Level 1
Not for use in warming platelets, cryo-precipitates, or granulocyte suspensions.
Death or serious injury may occur to the patient or user if these warnings are not followed:
H-1200 Fast Flow Fluid Warmer (Fluid Warmer).
CONTRAINDICATIONS
WARNINGS
These instructions contain important information for safe use
of the product. Read the entire operator’s manual, including Warnings and Cautions, before using this product. Failure to properly follow warnings, cautions, and instructions could result in death or serious injury to the patient or user.
Remove all air from the uid bags before spiking and the uid
lines before connecting to the patient. Failure to do so can result in infusion of air into the patient.
Replace Gas Vent/Filter Assembly every three hours, or when
the lter becomes clogged, or when air is slowly vented. Failure to do so will result in a reduction of ow rate. This may result in inadequate patient treatment
The replacement Gas Vent/Filter Assembly must be fully
primed before continuing infusion. Failure to do so may allow air to be infused into the patient.
Do not use the Fluid Warmer in high-energy elds such as:
MRI, X-RAY, portable and mobile RF communications equipment, and other such devices. The Fluid Warmer may act as a projectile in a strong magnetic eld, cause image artifacts, or not function as intended.
Do not bend the heat exchanger. Bending may damage the
heat exchanger allowing communication between the recirculating solution and I.V. uid path, resulting in the I.V. delivery of inappropriate uids.
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
5
SECTION 4 • Important Safety Information
WARNINGS [continued]
Blood and blood products could contain pathogenic
organisms. Failure to follow institutional policy and procedures for biomedical-hazardous materials could lead to exposure to harmful pathogens.
When injecting medications into the uid path, do not inject
through the triple lumen tubing of the D/DI-60HL Disposable Set. This may allow communication between the recirculating solution and I.V. uid path.
Exposed conductor on MAINS power cord can cause an
electrocution hazard. Remove device from service if MAINS power cord has exposed wires.
Do not re-use partially full uid bags. Fluid bags that have
been partially drained, un-spiked, and then reinstalled may contain air, which if used can result in infusion of air into the patient. Use only new uid bags from which the air has been removed.
Activation of the Over Temperature warning signal indicates
that warming has stopped and immediate operator intervention is required. Failure to clear the over temperature condition or to take the device out of service may result in death or serious injury to the patient.
The Fluid Warmer is not for use with irrigating tubing, which
may not t into the clamp slot of the Air Detector/Clamp causing diminished ow or a failure to stop ow.
The Fluid Warmer is for use only with Smiths Medical
supplied or approved parts, accessories, and D or DI series Disposable Sets. The device may not function as intended with the use of unapproved parts, accessories, or Disposable Sets.
Grounding reliability can only be achieved when MAINS
power cords are connected to a properly grounded receptacle. Risk of electrical shock exists if the equipment is not connected to a properly grounded receptacle.
Use of a bedside leukocyte reduction lter may cause a
sudden precipitous drop in blood pressure resulting in respiratory distress, facial ushing, abdominal pain and nausea, and loss of consciousness. Immediately stop transfusion, and follow institution’s protocol for treatment of transfusion reactions.
Do not operate the Fluid Warmer in the presence of a ammable
anesthetic mixture with air, oxygen, or nitrous oxide. The risk of
6
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
SECTION 4 • Important Safety Information
WARNINGS [continued]
explosion exists if the Fluid Warmer is operated in a potentially explosive environment.
No user-serviceable parts. All service must be performed by
Smiths Medical or competent personnel.
Disposable Sets are supplied with a sterile uid path which may
be compromised if the caps are not in place. Do not use Disposable Set if Luer and spike caps are not securely in place, or if Luer connections are not secure as the uid path may not be sterile and may cause death or serious injury to the patient.
Disposable Sets are for single use only. To reduce the risk of
cross contamination, do not reuse Disposable Sets.
If uid exits the Patient Line of the D/DI-60HL Disposable Set,
replace the Disposable Set.
Do not use auto-transfusion bags. Auto-transfusion bags
may contain air that can result in infusion of air into the patient.
Set-up, priming, and use require aseptic technique. Failure to
use aseptic technique may result in death or serious injury to the patient.
Do not use the Fluid Warmer if equipment or Disposable
Set malfunction is evident.
No modication of this equipment is allowed.
WARNINGS for the Air Detector/Clamp
The tubing must be properly placed in the Clamp Slot of the
Air Detector/Clamp. Failure to ensure that the tubing is correctly positioned in the Clamp Slot may result in failure to stop air infusion.
Activation of the Air Detector/Clamp alarm during infusion
indicates that uid ow has stopped and that immediate operator intervention is required to restore uid ow. Failure to reinstate ow (after purging any air or foam) may result in death or serious injury to the patient.
Do not turn OFF the Fluid Warmer when the Air Detector alarm
is active. If the Fluid Warmer is powered OFF in an active alarm state, the Air Detector/Clamp will open and the Air Detector will become disabled. This could allow any air within the patient line to be delivered to the patient resulting in serious injury or death.
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
7
SECTION 4 • Important Safety Information
WARNINGS for the Air Detector/Clamp [continued]
The functional test for the Air Detector/Clamp accessory
must be performed before each use. If any visual indicator does not illuminate or the audible signal does not sound, do not use the Fluid Warmer. Remove the device from service immediately. Fully functional visual and audible alarm systems are essential for the safe use of the Air Detector/Clamp.
CAUTIONS
Malfunction, failure, or damage to the device may occur if these cautions are not followed:
Never use organic solvents (e.g., acetone), strong acids, or
bases to clean any portion of the Fluid Warmer.
Do not place the Fluid Warmer directly under a faucet or use a
faucet sprayer to rinse. Never spray cleaning or other uids into openings on the Fluid Warmer or into the external connectors.
CAUTIONS
When loading uid bags into Pressure Chambers, choose a
hanging hook that allows the bag port to hang freely in the indented slot at the bottom of the chamber door. If bag ports are positioned above this slot, diminished ow could occur.
Federal law (U.S.A.) restricts this device to sale by or on the
order of a physician.
Medical devices require specic material characteristics to
perform as intended. These characteristics have been veried for single use only. Any attempt to re-process the device for subsequent re-use may adversely aect the integrity of the device or lead to deterioration in performance.
8
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
SECTION 5
SECTION 5 • Out of the Box—Assembly
5 Out of the Box – Assembly
Out of the Box—Assembly
This device must be assembled and tested by authorized Smiths Medical personnel, an authorized distributor of Smiths Medical, or competent personnel prior to placing the device into service.
The following steps describe how to assemble and do preliminary set up of the Level 1® H-1200 Fast Flow Fluid Warmer (Fluid Warmer).
Refer to Step 8 if you need to install the Level 1® H-31, Version B, Air Detector/Clamp to the Level 1® H-1000 Fast Flow Fluid Warmer.
Step 1 Verify components of the Fluid Warmer
Step 2 Assemble I.V. Pole to Warming Unit
Step 3 Install Pressure Chambers
Step 4 Attach the I.V. Bag Hanger
Step 5 Disinfect the Recirculating Solution Reservoir
Step 6 Preliminary Preparation
Step 7 Connect the Pneumatic Tubing
Step 8 Install the Level 1® H-31, Version B,
Air Detector/Clamp
Step 9 Perform Electrical Safety Tests
Read through the instructions completely prior to setting up the device.
Note: After unpacking the system, recycle packaging material according to hospital policy for recyclable materials.
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
9
SECTION 5 • Out of the Box—Assembly
Step 1 – Verify Components of the Level 1® H-1200 Fast Flow Fluid Warmer
Note: The Level 1® H-31, Version B, Air Detector/Clamp is shipped as a
separate accessory only for installation on an existing Level 1® H-1000 Fast Flow Fluid Warmer.
Check the contents of all packaging to verify that the following components are present. If any parts are missing or damaged, do not use the Fluid Warmer. Do not substitute parts not supplied by Smiths Medical. Contact Smiths Medical for replacement parts. Below is a listing of the component parts for the Level 1® H-1200 Fast Flow Fluid Warmer.
Components Checklist
in order of appearance in illustration
Qty Component
2 Pressure Chambers / Contents: (2) “U” brackets (4) Thumbscrews
1 I.V. Bag Hanger 1 Fluid Warming Unit
1 H-31, Version B, Air
Detector/Clamp 1 Accessory Pack
(3) Pan-head screws (3) Power Cord Clips (3) Flat-head screws
1 Operator’s Manual 1 Quick Reference Guide 1 I.V. Pole with Flanking Brackets 1 Accessory Pack / Contents:
(2) Plastic “J” Clamps (1) Y Connector (1) Black Tubing (1) Hex Wrench
1 O-Ring Kit / Contents: (1) Silicone (2) O-Rings (1) Hex Wrench (1) Instructions for Use
3 Bolts 3 Washers 1 I.V. Pole Base
10
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
SECTION 5 • Out of the Box—Assembly
Step 2 – Assemble I.V. Pole to the Warming Unit
There are three steps involved in assembling the I.V. Pole to the Warming Unit. The steps are: 1. Assemble the I.V. Pole to the Base,
2. Close the Drain Valve, and 3. Attach the Warming Unit to the Flanking Brackets. Each step is detailed in a short procedure.
2.1 Assemble the I.V. Pole to the Base
1 Locate the I.V. Pole Base (a).
2 Locate the dark-gray extruded I.V. Pole (b) with Flanking
Brackets.
b
d
e
3 Place the I.V. Pole Base upright on its wheels, (c) and lock the
wheels to prevent movement during set up.
4 Locate three bolts (d) and washers for the
pole base.
5 Align the three holes (e) in the I.V. Pole with the three screw
holes on the pole base.
6 Slide the I.V. Pole down over the pole base, (f) keeping holes
f
aligned.
7 Guide three bolts and washers through the holes (g) at the base
of the pole and tighten.
g
c
a
2.2 Close the Drain Valve
h
Turn valve, located on the bottom of the device, perpendicular to stem (h) of the Warming as shown.
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
11
SECTION 5 • Out of the Box—Assembly
2.3
Attach the Warming Unit to the Flanking Brackets
1 Align the eight hex screws on the back of the WarmingUnit
with the eight keyhole notches on the anking bracket.
2 Slide screw heads down into keyhole notches.
3 Tighten all eight hex screws with the supplied hexwrench
and secure in place.
Attach the Quick Reference Guide
2.4
to the Fluid Warmer by sliding the ring on the Quick Reference Guide over one of the poles on the Flanking Bracket.
Step 3 – Install the Pressure Chambers
1 Locate the two Pressure Chambers.
2 Locate the U-brackets and thumbscrews supplied with
the Pressure Chambers.
3 Attach the U-brackets with thumbscrews to the back of
thePressure Chambers, as shown. Keep thumbscrews and brackets loose.
4 Slide one Pressure Chamber with attached U-bracket over
the top of each anking pole.
5 Align the U-bracket slightly below the top of the anking
pole. Tighten the thumbscrews securely.
12
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
MIN
MAX
1
2
SECTION 5 • Out of the Box—Assembly
Step 4 – Attach the I.V. Bag Hanger
1 Slide the I.V. Bag Hanger on top of the I.V. Pole.
2 Align with tabs.
3 Press down and snap into place.
Step 5 – Disinfect the Recirculating Solution Reservoir
1 Remove the ll-port plug (a) on the reservoir.
2
c
R
b
1
a
MAX
2 Prepare a 0.3% hydrogen peroxide/distilled water solution
for the reservoir. Mix 140 ml of 3% hydrogen peroxide solution and 1,260 ml of distilled water.
3 Fill the reservoir with 1.4 liters of 0.3% hydrogen peroxide/
distilled water solution.
4 Replace the ll-port plug.
5 Insert a Disposable Set into the Fluid Warmer.
6 Insert the power cord into a properly grounded receptacle.
7 Turn the Fluid Warmer ON. Let the solution circulate for a
30-minute disinfection period.
8 Turn the Fluid Warmer OFF.
9 Empty the reservoir.
10 Remove the Disposable Set and discard according to
established hospital procedures.
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
13
MAX
1
2
SECTION 5 • Out of the Box—Assembly
Step 6 – Preliminary Preparation
1 Remove the ll-port plug (a) on the front of the warming unit
and ll the reservoir to the maximum level with 1.4 liters of one of the following solutions:
0.3% Hydrogen Peroxide/Distilled Water Solution
Mix 140 ml of 3% hydrogen peroxide and 1,260 ml of distilled water.
Note: If this option is selected, the maintenance requirement to change the recirculating solution is once every 12 months. Always use a 0.3% hydrogen peroxide/distilled water solution when relling the reservoir.
Distilled Water
Note: If this option is selected, the maintenance requirement to change the recirculating solution is once every 30 days.
2 Replace the ll port plug.
3 Lubricate O-Rings in #1 Block (b) and #2 Block (c). Place a small
amount of silicone lubricant, provided in the supplied O-Ring Kit, on a cotton swab and apply all around the inside of each O-Ring.
c
2
R
b
a
1
Step 7 – Connect the Pneumatic Tubing
1 Locate the Accessory Pack with the black pneumatic tubing (a),
two “J” clamps (b), and one “Y” Connector (c).
a
b
c
14
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
SECTION 5 • Out of the Box—Assembly
2 Locate the orange protective plug in the red ring connector (d)
located on the back of the device. Remove the plug by depressing the red plastic ring as you pull the plug out of the connector.
f
e
g
3
Take the pneumatic tubing and press one end of the tubing rmly into the ring connector
4
Take the “Y” connector and press the other end of the pneumatic tubing into the bottom of the “Y” connector, as shown
(d)
until it can go no further.
(e)
, until it can
go no further.
5
Press the pneumatic tubing from the Pressure Chamber into place
(f)
on the top of the “Y” connector until it can go no further. Repeat
this procedure with the pneumatic tubing from the other Pressure
h
Chamber.
6
Remove the protective backing sheet on one “J” clamp, exposing the adhesive side.
7
d
Carefully position the “J” clamp and press the adhesive side against the gray I.V. Pole in the approximate locations
(g)(h)
shown.
F1
Press down rmly to secure in place.
Repeat this procedure for the other “J” clamp.
8
Press the pneumatic tubing into place on the “J” clamps.
Step 8 – Install the Level 1® H-31, Version B, Air-Detector/Clamp
Before installing the Air detector, determine if the Fluid Warmer was manufactured before 2004. If it is, inspect the “F1” fuse to conrm that it is the correct value and type. The year of manufacture can be determined by the serial number label which is located on the bottom right side of the unit, or for older units on the inside. If the rst four characters of the serial number are numbers, then the numbers represent the year of manufacture (e.g. 20040100 would be manufactured in 2004). If the rst character of the serial number is the letter “J”, then the year of manufacture is prior to 2004 (the letter “S” indicates the year 2007 or later). For units manufactured prior to 2004, perform the following.
1 Disconnect the power.
Two Fuse Conguration
F1
Four Fuse Conguration
FAST FLOW FLUID WARMER
2 Remove the 18 screws from the rear of the Fluid Warmer and remove the back panel.
3 Locate the “F1” fuse on the electronic
assembly. See gures to locate the “F1” fuse for either a two fuse conguration or a four fuse conguration.
|
Operator’s ManualH-1200
15
SECTION 5 • Out of the Box—Assembly
For 100 - 120V Fluid Warmers,
Remove the fuse and verify that T6.3AL250V (time lag,
6.3 amp, 250V fuse), is printed on one of the silver end caps, if not replace the fuse with a new fuse marked T6.3AL250V.
For 220 - 240V Fluid Warmers, Remove the fuse and verify that T3.15AL250V (time lag,
3.15 amp, 250V fuse), is printed on one of the silver end caps, if not replace the fuse with a new fuse marked T3.15AL250V.
4 Replace the back panel and insert all but 6 of the screws (a)
as indicated in Step 1 below. Then proceed to install the Air Detector/Clamp.
a
d
a
a
e
a
To install the Air Detector/Clamp,
1 Remove 6 screws (a) from the rear of the Fluid Warmer,
lower left side, from locations shown in gure.
2 Plug in the power cord (b) of the Level 1® H-31, Version B,
Air Detector/Clamp into the auxiliary MAINS outlet located on the bottom of the Fluid Warmer.
Note: If you turn the Fluid Warmer o and the Air Detector/Clamp does not turn o, contact Smiths Medical or your local Smiths Medical distributor.
3 Loosen the three screws holding the mounting bracket to the
rear of the Air Detector/Clamp (do not remove).
4 Align the slot on the Air Detector/Clamp with #3 Block on the
Fluid Warmer (c). Fit in place over the block.
Note: Do not bend the mounting bracket while holding the Air Detector/Clamp against the uid Warmer.
5 Carefully align the mounting bracket on the rear of the
Air Detector/Clamp with three screw holes, shown (d). Ensure that the gasket seal on the Air Detector/Clamp is ush with the Fluid Warmer. Insert three at-head screws and tighten, securing the Air Detector/Clamp to the Fluid Warmer. Tighten the three screws on the mounting bracket on the Air Detector/Clamp.
e
a
e
a
b
c
6 Locate three power cord clips (e) included with the
Air Detector/Clamp and snap onto power cord. Align clips with the three screw holes as shown, insert screws, and tighten.
7 Install and prime a Level 1® Fast Flow I.V. Disposable
Administration Set, and test the audible and visual alarms and Air Detection/Clamp as described in Section 9 Operating Instructions.
16
FAST FLOW FLUID WARMER
|
Operator’s ManualH-1200
Loading...
+ 53 hidden pages