Leidos PDE5 Operation and Maintenance Manual

Safety & Security Instruments
A Division of Science Applications International Corporation
OPERATION & MAINTENANCE MANUAL
RADSTAR SYSTEM
MODELS PD-4, PDE-4 AND PDX-4
16701 West Bernardo Drive San Diego, CA 92127 Rev: 010695 (800) 962-1632 or (858) 826-9831 Drawing 201313
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page1
USER'S FCC NOTICE
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of FCC Rules. These limits are designed protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help. Shielded cables and I/O cords must be used for this equipment to comply with the relevant FCC
regulations.
Changes or modifications not expressly approved in writing by Science Applications International Corporation may void the user's authority to operate this equipment.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page2
RadStar
TM
Radio Transmission Dosimetry System
PREFACE
Congratulations! You have made an excellent purchase decision. Science Applications International Corporation's (SAIC's) RadStar™ system will provide improved remote radiation monitoring performance which enables:
New levels of ALARA conformance
Minimum staffing and training requirements
Rapidly deployable monitoring
Cost effective spot monitoring
Outage time reduction
• Improved safety
In addition, SAIC is committed to providing the total support net required for such a system which can be accessed by calling (800)-962-1632. Enough flexibility has been designed into the RadStar™ Radio Dosimetry system to handle a wide range of remote monitoring applications. Although simple networks can be powered up and run without modifying the factory default configuration, setting up the more complex network architectures will usually require some reconfiguration of operating parameters within the PD-4s™ and PDE-4s™ [or simply, PD(E)-4] teledosimeters. For this purpose, SAIC recommends the use of the included PDRC4 application program which allows users to edit control settings within the PD(E)-4 via a PDR-1™ or PDR­1/S™ Reader. After the network has been configured, it can be monitored and controlled by the RadStar™ application program, RS which is also included. Both RS and PDRC4 are compatible with and can be interfaced to user written applications. Both of these application programs are referenced in this manual to provide a context for describing configuration and control operations. A limited amount of descriptive material concerning the use of these programs is contained herein; however, both are provided with on-line Help which will provide additional details concerning active keys, editing procedures, and setup procedures.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page3
1.0 INTRODUCTION
The RadStar™ System is the most advanced radio transmission dosimetry system available, providing detection, radio transmission, and remote display of radiation exposure information.
RadStar™ is unique in its use of BIDIRECTIONAL transmissions for high communication
reliability at low RF power levels, minimizing the potential for electromagnetic interference. The primary application for RadStar™ is remote monitoring of personnel working under severe radiation conditions where accurate radiation dosimetry is difficult. Typical monitoring problems solved by RadStar™ include situations involving
high radiation field gradients,
radiation fields that are difficult to characterize with a high level of confidence, and
potentially high radiation fields that vary with time.
A secondary application for RadStar™ Dosimeters is as local area monitors. Each RadStar™ Dosimeter is equipped with an external power jack, thus allowing it to monitor specific points of interest, unattended, for the duration of an outage. Alternatively, RadStars™ can be temporarily placed to monitor the progress of processes such as chemical decontamination.
From its inception, the RadStar™ concept has been entirely consistent with the spirit of ALARA. RadStar™ allows the Health Physicist and his staff to monitor the progress of sensitive tasks from outside the control area. RadStar™ products, like the PDE-4™ with 5 parallel radiation measurement channels, allow the HP to simultaneously monitor trunk and extremity doses as the radiation worker performs his task. Thus, unplanned doses due to field gradients, hot spots, and time variant conditions can be detected and remediated immediately.
RadStar™ can reduce personnel doses in yet other ways. Accurate, real time monitoring allows
contingencies to be reduced, and in cases where the worker still has allowable dose after completion of a task, he can move on to another job without leaving the control zone, thus avoiding a second "entry dose".
The use of BIDIRECTIONAL transceivers in RadStar™ Dosimeters adds a totally new dimension to radio transmission dosimetry. Now, alarm levels and transmission intervals can be changed "on the fly" as the worker moves from job to job. No longer is the radio energy simply launched into space with the hope that it will be received. RadStar™ communications use a sophisticated error checking protocol which includes automatic retransmission and acknowledgment; if a RadStar™ message packet fails to reach its destination, the sender is notified immediately. Bidirectionality also allows for implementation of the repeater function. Any RadStar™ Dosimeter can be configured to repeat messages that it receives. Strategic placement of a single repeater can double the range of a communication link or bend messages around an otherwise blind corner. In the unlikely event that a network link is broken (e.g., a RadStar™ Dosimeter moves out of range of its primary receiver), the Dosimeter automatically attempts to reroute its transmissions through a different network path.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page4
2.0 RadStar PRODUCT FAMILY Key components of the RadStar™ System include the following products:
PD-4™ The PD-4™ is a mobile Transceiving Dosimeter which uses the same
radiation detection circuitry that was developed for the PD-1™ electronic dosimeter. Over the years, the PD-1™ has proven itself in nuclear power plants to be the most rugged, reliable and accurate electronic dosimeter available. To produce the PD-4™, the PD-1™ circuitry was repackaged in a larger case to provide space for the spread-spectrum radio transceiver. Like the PD-1™, the PD-4™ is an independent dose and dose rate measurement system and does not require communication with a base station to acquire and store radiological data. Thus, if communications are temporarily interrupted, the PD-4™ continues its monitoring functions and then provides up-to­date total dose and dose rate data when the link is re-established.
PDE-4™ PDE-4™ is also mobile and provides the same functionality as the PD-
4™ with the additional capability to simultaneously monitor up to four extremity sensors. For instance, a single PDE-4™ can provide radiological data for the trunk, both hands and both feet, all at the same time.
PDX-4™ The PDX-4™ RadBase is the base station end of the communication
link. It contains the same radio modem that is used in the PD(E)-4. The PDX-4™ receives messages from PD(E)-4s and transfers them to the host PC over an RS-232 serial cable. Similarly, the PDX-4™ accepts commands from the host PC (e.g., change alarm value commands) and transmits them to the designated PD(E)-4. Spread-spectrum radios, operating in or near the GHz range, are extremely robust and provide reliable communications under a surprising variety of adverse conditions. It is not unusual to establish communications up and down elevator shafts, through reinforced concrete walls, and into Faraday cage type screen rooms. However,
SAIC highly recommends that the viability of a particular network
configuration and geometry be demonstrated by test before being put into service.
PDR-1™ The PDR-1™ Dosimeter Reader provides access to the PD(E)-4's
EEPROM storage area where calibration factors, alarm levels, and other system constants are stored. The PDR-1™ communicates with the PD(E)-4 through the same capacitance interface as the PD-1™ and to the host PC through a standard RS-232 serial cable. Since only alarm levels and transmission intervals can be changed via radio commands, the PDR-1™ is required for adjusting calibration factors during
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page5
calibration procedures and for configuring communication parameters and the external sensors of a PDE-4™.
Host PC Customers generally use their own PCs. SAIC recommends the use
of an IBM-compatible 286 or higher end machine.
3.0 QUICK START
3.1 Hardware Connection
While the RadStar™ System offers a high degree of flexibility, it is easy to use in simple network configurations. Let's consider the simplest possible configuration consisting of one
RadStar™ PD(E)-4 Dosimeter and one RadStar™ PDX-4™ RadBase (Fig. 3-1). As shown in
the figure, the PDX-4™ RadBase is connected to the host PC with a 9-pin to 25-pin RS-232 serial cable (included with the PDX-4™). Don't use other cables for this purpose because it is unlikely that their pinouts will be correct. The 25-pin end must be connected to either the COM1 or COM2 serial communication port of the PC. If gender-changers or other adaptors are required to make this connection, they should be of the "straight-through" variety (i.e., don't use null modem components which swap the transmit and receive lines). Next, supply power to the PDX­4™ by connecting the provided 9 V power adapter to a 120 VAC receptacle. Note that the PDX­4™ power adapter is very similar in appearance to the adapters supplied with PD(E)-4s which have opposite output polarity.
CAUTION:
Before applying power to the PDX-4, be sure you are using a 9 V device which applies positive polarity to the center conductor of the connector.
Polarity can be verified by checking the manufacturer's label on the power adapter's case.
IMPORTANT:
The white recessed button on the front of the PDX-4™'s case (just to the right of the RS232 connector) is used to generate a non-maskable interrupt request, not a reset. It should be pressed to put the PDX-4™ into serial communication mode for configuration. A momentary interruption of power is the easiest way to reset the PDX-4™.
The PD(E)-4 Remote may be powered either by a 9 V alkaline battery or by its external power adapter. The unit's power switch should be in the "Off" position prior to making the power source connection.
CAUTION:
Before applying external power to a PD(E)-4, be sure you are using a 9 V device which applies negative polarity to the center conductor of the connector and that the power switch is in the "Off" position.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page6
IMPORTANT:
If the power switch is in the "On" position when a power source is connected, it is probable that contact bounce will cause PD(E)-4 built-in test circuitry to disengage the main voltage regulator from the power source (there will be no powerup chirp and the display will remain blank). If this happens, turn the unit off for a few seconds using the power switch and then try again.
If the power adapter option is chosen for the PD(E)-4, it makes no difference whether or not a battery is present. If the battery option is chosen, be sure the battery is fresh. See section 4.7 for installing a fresh battery. Finally, insure that antennas are installed on both the PDX-4™ and the PD(E)-4. This completes the hardware setup for the trial network.
3.2 Software Installation Now is a good time to install the required software on your PC. The software includes two (2)
applications packages: 1) RS - the RadStar Communications Control Package, and 2) PDRC4 ­which allows for editing of the PD(E)-4 database. First create a separate directory to receive the software. You might name the directory PD4 or RADSTAR. The following DOS commands will do the job:
C:\>md radstar C:\>cd radstar C:\RADSTAR>_
Now use the DOS COPY command to copy the required files from the SAIC distribution diskettes. The necessary files are listed below:
HELP_RAD.1 HELP_RAD.2 HELP_RAD.3 HELP_RAD.4 HELP_RAD.5 RS.EXE SETUP.VAL HLPDRC_4.1 HLPDRC_4.2 HLPDRC_4.3 HLPDRC_4.4 HLPDRC_4.5 HLPDRC_4.6 HLPDRC_4.7 PDRC4.EXE
Start the RadStar™ application by typing RS at the DOS prompt, e.g., C:\RADSTAR>RS_
Pressing the Enter key should produce the RadStar™ title screen which will persist for five seconds (unless some key is pressed) followed by a display similar to the following:
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page7
PDX-4/PD-4 Communication Interface Release date: 05-JAN-1994
Checking COM Port #1... PC Time-Out Error! Checking COM Port #2... PDX-4 found Base-Station serial driver installed at COM2 PDX-4 Serial No: 32486 ROM Version: V2.4 Radio Channel: 4 Sub-Channel: 1
RF Signal Strength (Channel,%Max): (1,37) (2,37) (3,36) (4,34) (5,33) (6,33) (7,39)
--ESC or RETURN for Main Menu--
In this case, a PDX-4™ RadBase was located at COM2 and the selected radio channel and subchannel were 4 and 1, respectively (the factory defaults). The RF signal strength report gives you an idea of the relative amounts of ambient RF energy that is currently present in each of the PDX-4™'s seven radio channels. Smaller numbers are better; if the relative signal strength in the selected radio channel is substantially higher than the other channels, you might want to consider using a different radio channel to set up your network (described later in this manual). If the PDX-4™'s radio channel is changed, a corresponding change must be made to the PD(E)-4 as described below in "In Case of Trouble". If you obtain a "PC Time-Out Error!" for both COM ports, the application was unable to obtain a PDX-4™ response from either COM port. In this event, check cabling between the PDX-4™ and the PC and insure that the PDX-4™ is powered (the red PWR Light Emitting Diode should be illuminated). Then, reboot your PC to eliminate any serial port drivers (e.g., mouse drivers) that might have been previously installed. If the application still fails to locate the PDX-4™, refer to Appendix C (Configuring the PDX-4™ RadBase).
Once the application has located a PDX-4™ connected to one of the COM ports, press Enter (Return,Ret) to obtain the Main Menu. The options available from the Main Menu are:
A. PDX-4 Base Station Controller B. Base Station Monitor C. B-S Monitor (HEX-to-disk) D. Display Help Files E. Base Station Controller Setup F. Exit to DOS
Select Option 'A' (PDX-4™ Base Station Controller ) by pressing the 'A'-key. You will be looking at the RadStar™ message screen with no entries. Now, turn on the PD(E)-4. Within 3 to 4 seconds, it should issue an audible chirp and its first message should be received and written to the screen in ROW #0. The message entry in the "Xmit Int (s)" column tells you how many seconds you must wait for the next message update. Other similarly configured PD(E)-4's may be turned on at this time and their messages will be added to the display as they are attached by
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page8
the Base Station. Note that five display lines are required by a PDE-4™; one line for the internal detector and four lines for the external probes.
3.3 In Case of Trouble
If you're reading this section, events have probably not transpired as described above. Almost invariably, communication problems can be traced to dead batteries or incompatibilities between Base Station and PD(E)-4 configurations. Check the PD(E)-4's display; if it is completely blank turn the unit off for about 10 seconds and then turn if back on. If the display remains blank or if the battery icon is visible, replace the battery (or check external power connections, whichever is appropriate). If the display appears normal (i.e., a dose or dose rate value is being displayed), press the red "CALL" button; the PD(E)-4 should respond immediately by illuminating its "RD" LED for about one second followed by a simultaneous flash of its "AT" LED and an audible chirp. If the letters, "rPr", are visible on the display, the unit has been configured as a Repeater and needs to be reconfigured as a Remote.
3.4 Base Station QUICK START Configuration
Press the ESC key to return to the Main Menu and select Option 'E' (Base Station Controller Setup) by pressing the 'E'-key. A new RF Signal Strength report will be generated and you will then be offered the option to select a new radio channel; press Enter to leave the channel number unchanged. Press Enter again to leave the radio sub-channel unchanged. Press Enter once again to leave the communication group number unchanged. You will now be looking at the following prompt (except, perhaps, for the numerical value):
MAX Units for this Base-Station = 10; New MAX-Units? _
For the purpose of discussing the MAX-Units value, it is convenient to define two terms for use in the context of this manual. The first term is "acquire". PD(E)-4s "acquire" a Base Station in the same sense that a guided missile or torpedo acquires a target. Once a PD(E)-4 acquires a Base Station, it will remain locked onto it for as long as the communication link remains unbroken. However, Base Stations are the masters of the acquisition sequence; that is, they must allow themselves to be acquired. When they do, we say that the Base Station has "attached" the Remote. Using these definitions, MAX-Units is the maximum number of attachments that the Base Station will allow. Enter a value of 10 for MAX-Units (or simply press Enter if the current value is already 10).
The next prompt is
MAX Direct Remotes for this Base-Station = 3; New MAX-Remotes? _
Attachments are termed "direct" when the Base Station attaches directly to a PD(E)-4 Remote. They are "indirect" when the Base Station communicates with the Remote through one or more Repeaters. MAX-Remotes is the maximum number of direct attachments to Remotes that the
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page9
Base Station will allow. Set MAX-Remotes = 10 or press Enter if it already has a value of 10 (note that in QUICK START, we are trying to achieve one Direct Remote attachment).
The next prompt is
MAX Direct Repeaters for this Base-Station = 2; New MAX-Repeaters? _
Just as with Remotes, Repeaters can be attached by Base Stations either directly or indirectly. A Repeater that communicates with a Base Station through one or more additional Repeaters is considered to be indirectly attached by the Base Station. MAX-Repeaters is the number of direct Repeater attachments that the Base Station will allow. Set it to zero for now since we are not currently dealing with any Repeaters. Lastly, you are offered the option of having all data logged into files that will be created in the current working directory. For now, refuse the option by entering 'N'. (See Appendix H for examples of PD-4™ and PD(E)-4 LOG files).
Before returning to the Main Menu, make a note of the radio channel, radio subchannel, and communication group number because the next step will be to verify that the PD(E)-4 is configured with the same values.
NOTE: In order for a PD(E)-4 Remote to communicate with a PDX-4™ RadBase Base Station, both units must have the same radio channel, the same subchannel, and the same Communication (Control) Group number (Section 3.5 descibes how these values are edited).
Press Enter to return to the Main Menu and select Option 'F' to return to DOS.
3.5 PD(E)-4 QUICK START Configuration
The principal purpose of the PDRC4 application is to provide the capability for editing PD(E)-4 EEPROM values. Connect a PDR-1™ to your other COM port and run PDRC4. Press any key to get past the title screen. If the PD(E)-4 unit was originally configured as a Repeater, turn it off, wait 4 or 5 seconds, and turn it back on. Place the PD(E)-4 upside down on the PDR-1™ Reader with its antenna pointed in the direction of the PDR-1s™ "DOSE RESET" button and select Option 'A' (Edit). The resulting screen displays a good portion of the PD(E)-4s EEPROM. These items will be discussed in detail in subsequent portions of this manual; however, our present concern is only that the following items have the following values:
1) Repeater Mode
DISABLED
2) Transmission Mode PNTtoPNT
3) Control Group (same as Comm Group Number of Base Station)
4) Broadcast Address:
FFFFFFFF
5) Radio Channel: (same as Radio Channel of Base Station)
6) Radio Sub-Channel: (same as Radio Sub-Channel of Base Station)
7) Transmit Int (s): 10
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page10
Use the TAB and arrow keys to position the cursor as necessary while making changes. Then press ESC to advance to the second screen and edit the following values as necessary:
8) Lost-Coms to Re-Acq:
3
9) Max Attachments: 10
10) Max Direct Remotes: 10
11) Max Direct Repetrs: 0
12) Auto-'A' Command ENABLED
13) Attach Lower FSNs ENABLED
Now press ESC once to end the editing session and then press Enter to install the edited values in the PD(E)-4's EEPROM.
IMPORTANT:
If any trouble is encountered with data transfers between the PD(E)-4 and the PDR-1™, press down gently (a few ounces) on the back of the PD(E)-4 case to insure proper contact with the Reader. This is particularly important in the case of a PDE-4™ with attached external probes which can unbalance the unit and cause it to tilt on the Reader.
Turn the PD(E)-4 off and restart the RS application. Select Option 'A' from the Main Menu and turn the PD(E)-4 back on. If the proper response is not obtained within a few seconds, contact the factory at (800) -962-1632)
4.0 PD(E)-4 Functional Description
4.1 Radiation Detection
4.1.1 Sensors
The PD-4™ uses the same radiation detector as the PD-1™ and PD-3™, an energy compensated Geiger Mueller Tube (GMT). The GMT provides very large signals compared to solid state devices which eliminates the need for high gain preamplifiers and pulse shaping amplifiers, while providing superior immunity to microphonics and electromagnetic interference. GMTs typically over-respond at photon energies in the range which begins with their detection threshold (30 to 60 keV) and extends up to about 200 keV; hence the need for energy compensation which provides selective photon removal in this energy range. The PDE-4™ is identical to the PD-4™ with respect to the internal GMT; that is, it uses the same energy compensated GMT for its internal radiation detector. In addition, the PDE-4™ supports up to four external GMT Extremity Probes which, as their name implies, can be affixed to various portions of the body (arms, legs, head) to simultaneously monitor local doses and dose rates. The effective center of detection for the internal GMT is indicated by registration marks (+) on the top and back of the PD(E)-4 case.
To minimize size, the Extremity Probe (SAIC Assembly No. 201406) is based on an uncompensated GMT (compensation typically doubles the diameter of a small GMT). Therefore,
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page11
Extremity Probes will overrespond somewhat in the 50 to 200 keV photon energy range. However, after calibration using a standard calibration source such as
137
Cs (661 keV) or 60Co (1173 and 1332 keV), the data provided by an Extremity Probe will be conservative from the personnel safety standpoint. It will either indicate accurately or it will indicate slightly high depending on the relative fraction of total dose that lies below 200 keV. The relative energy response curves for the compensated internal GMT and the uncompensated external GMTs shown in Fig. 4-1 demonstrate that either GMT provides reasonable low energy response (below 200 keV).
CAUTION:
The 4-pin Extremity Probe connectors are of the push/pull variety. Do not twist these connectors while connecting or disconnecting from the PDE-4. Align either the red dots or the black dots (whichever are most visible) to engage. SAIC recommends that the PDE-4 be momentarily powered down while external probes are being connected or disconnected to eliminate the possibilities of false counts being accumulated and component damage from the high voltage inrush current.
PDE-4™ Extremity Probes are selected from production lots based on their small sensitivity variation from the mean sensitivity. After installing the appropriate average calibration factors into the PDE-4™, the External Probes may be connected in any order to the PDE-4™ external connectors with minimum impact on measured responses. SAIC currently uses two different types of miniature GMTs during production of External Probes. These two GMTs have slightly different average gamma sensitivities (15 to 20%) and should not be interchanged with each other. The more sensitive of the two probes can easily be identified by a one inch long red cable sheath at the GMT end of the assembly. Of course, when traceability from data to sensors is mandatory, or for monitoring applications that require maximum accuracy, the sensors cannot be switched. In these cases, separate calibration factors and serial numbers for each probe should be installed in the PDE-4™. However, the SAIC probe selection process will still minimize the uncertainty in cases where probes are inadvertently switched or probe serial numbers are misread.
4.1.2 Rate Linearity
GMTs have a linear rate response over the range of rates typically experienced by radiation workers. Each GMT pulse increases the accumulated dose by the same increment (i.e., by the linear calibration factor, 1.5 µR/count) and no correction is necessary up to a few R/h. At rates higher than a few R/h, the fraction of time required by the GMT to recover from previous Geiger avalanches (the recovery or dead time) becomes significant so dose and dose rate values are multiplied by the factor, 1/(1-Cτ) where C is the measured count rate and τ is the dead time per Geiger avalanche. The purpose of the dead time calibration procedure is to establish the proper value for τ which typically lies in the range from 12 to 25 µs/count for miniature GMTs. With dead time correction the GMT is accurate at any rate up to 500 R/h. Above 500 R/h the unit will read conservatively high. GMT foldover will not occur at less than 10,000 R/h.
The maximum permissible cable length for PDE-4™ External Probes is also a function of the GMT dead time. For simplicity and reliability, the Geiger pulse is transmitted back to the PDE-
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page12
4™ by an open collector driver which leads to the inequality, l < τ/(RC'), where l is the cable length, R is the value of the pullup resistor, and C' is the distributed capacitance of the cable.
With R=10 kΩ and C'100 pF/foot, the relation becomes l < τ feet, where τ is in µs. Thus, for a GMT with τ ≈15 µs, the cable length should be less than 15 feet. Probe assemblies with cables exceeding this length will be subject to foldover at high rates.
4.1.3 Dose Rate Sampling Algorithm
PD(E)-4s are not designed as survey instruments. Nonetheless, they can measure dose rate from background to 500 R/h. The dose rate algorithm runs once each Display Period (one DSP=0.75 s), regardless of the display mode, so rate alarms can be activated even when the PD(E)-4 is not displaying a dose rate.
To estimate dose rates, the PD(E)-4 always attempts to acquire statistically significant data. At lower dose rates, longer Rate Sampling Intervals (RSIs) are used to acquire an accurate rate sample. At very low dose rates (e.g., background rates in the 10 to 20 µR/h range), the units sample for 1024 s (17 minutes) to obtain a rate sample with reasonable statistical accuracy. As the ambient dose rate increases, the PD(E)-4 automatically decreases the length of the RSI to provide faster rate response.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page13
Rate Sampling Intervals (RSI) for
PD(E)-4
50
40
30
20
10
RSI=1024s
64/4 <- RSI Switching Limits (Geiger Pulses/RSI)
RSI=64s
256/16
RSI=4s RSI=1/4s
1024/64
0
0.01 0.1 1 10
Dose Rate (mR/h)
100 1000 10000 100000
Switching Limits. First number is number of counts to
-
cause switch from longer to next shorter sampling inte rval. Second number is number of counts to cause switch from shorter to next longer sampling interval.
Rate Sampling Interval, Time in seconds.
-
Rate Sampling Intervals and Switching Limits
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page14
4.2 Operating Modes
The three basic operational modes of the PD(E)-4 are DOSE, RATE and CONFIRMATION. The operating mode is changed by pressing the "Mode" pushbutton (the black button nearest the front of the case, see Fig. 4-3). The PD(E)-4 acknowledges by momentarily blanking its display and issuing a short chirp from its audio annunciator. An additional operating mode communicates with a PDR-1™ or PDR-1/S™ reader through a serial link to download data and to upload commands and setpoints. This mode is not reachable via the MODE button and can be activated only by a reader in which case "I-O" will appear on the display. On powerup, the PD(E)-4 defaults to DOSE display of Internal GMT #1; successive closures of the MODE button will cause the PD(E)-4 to sequence through the following display modes:
Powerup Default Internal #1 DOSE Internal #1 RATE External #1 DOSE (if present) External #1 RATE " " External #2 DOSE " " External #2 RATE " " External #3 DOSE " " External #3 RATE " " External #4 DOSE " " External #4 RATE " " CONFIRMATION Mode back to Internal #1 DOSE
Display modes for non-functional or missing sensors are skipped.
4.2.1 DOSE Mode
In DOSE mode, the display shows the accumulated dose (in mR or R) since the last reset operation using a three-digit floating point format with autoranging. The dose is recalculated at each DSP (0.75 s) for alarm purposes even when the PD(E)-4 is in some other display mode.
4.2.2 RATE Mode
In RATE mode, the display shows the dose rate value estimated from the most recent RSI in mR/h or R/h using the same floating point format that is used for dose values. The RSI is always reset to the shortest interval (1/4 s) on entering RATE mode; therefore, the PD(E)-4 can always be forced to post a new RATE value by cycling through the display modes until RATE display of the desired sensor is reached. The new rate value will be displayed during the DSP following entry to RATE mode.
Upon entering RATE mode, the PD(E)-4 will display the current dose rate while calculating a new dose rate based on the 1/4 s RSI. After approximately one second, the new rate value will be written to the display. If the ambient dose rate is so low that a statistically significant value was not obtained, the RSI will be increased to 4 s. The PD(E)-4 will continue to increase the length of
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page15
the RSI until it satisfies its internal, statistical criteria or until it reaches the maxi mum RSI length of 1024 s.
During an RSI, the total number of counts accumulated since the beginning of the RSI is tested at each DSP. If this count exceeds the internal maximum count limit for the given RSI, the PD(E)-4 posts an approximation of the count rate based on the count limit and the current RSI length. It then reduces the RSI length and begins acquiring a new rate sample. The following example illustrates how the PD(E)-4 would react to a step increase in dose rate as might occur when a worker moves into a controlled zone (please refer to Fig. 4-2).
SITUATION: A worker is using an SAIC PD(E)-4 which is already in RATE MODE. He has been
working in an area where the background radiation levels are normal so the PD(E)-4 is operating with an RSI of 1024 s. The user then enters an area where the dose rate is 1 R/h. How long will it take for the PD(E)-4 to respond?
RESPONSE: 1 R/h corresponds to approximately 200 cnts/s. It will therefore take at most .32 s to
accumulate the 64 counts required to trigger a decrease of the RSI (see Fig. 4-2) at which time the PD(E)-4 computes and displays a RATE estimate based on _64 cnts/1024 s 320 µR/h. The PD(E)-4 then decreases the RSI to 64 s and begins the 64 s acquisition. The count will exceed 256 in 1.28 s at which time the PD(E)-4 computes and displays a RATE estimate based on _256 cnts/64 s 20 mR/h, decreases the RSI to 4 s and begins the 4 s acquisition. After 4 s, the PD(E)-4 computes a true rate of ≈800 cnts/4 s (within the limits of statistics) and displays 1 R/h. Since the count rate during the 4 s RSI does not exceed 1024, the RSI is maintained at four seconds.
RATE display mode can be totally disabled by setting the appropriate bit in the PD(E)-4's EEPROM in which case successive closures of the MODE button will cause the PD-4™ to advance directly from DOSE mode to CONFIRMATION mode.
4.2.3 Confirmation Mode
Confirmation Mode displays alarm and chirp settings for all active sensors and then generates an LCD (Liquid Crystal Display) test pattern. Upon entering Confirmation mode, a PD-4™ first displays the DOSE alarm value (indicated by a visible DOSE icon) followed by the RATE alarm value (indicated by a visible RATE icon) followed by the CHIRP dose increment value (indicated by a visible Gamma icon). All LCD segments are then turned on simultaneously for the test pattern to allow visual verification that no segments are missing. The user should verify the presence of all seven segments of each of the three digits; the three decimal points; the Dose, Rate, Battery, and Gamma icons; the 'm', 'R', and '/h' icons; and the "EXT" and '1'-'2'-'3'-'4' icons (see Fig. 4-4) mode
. Pressing the MODE button at any time during execution of Confirmation mode will
. Upon completion of the display test, the PD-4™
automatically returns to Dose
immediately return the PD-4™ to Dose mode.
Confirmation mode for a PDE-4™ is the same except that before the display test, the DOSE-RATE-CHIRP display sequence is executed for each external probe that is attached. The applicable sensor is indicated on the LCD by a combination of the EXT icon and the 1-2-3-4 icons.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page16
4.2.4 Chirp Increment
The dose increment chirp is a feature rather than a mode
. Each time the total dose accumulator
increases by a preset increment (the CHIRP dose increment), a brief tone or "chirp" is issued from the audio annunciator
. For example, if the chirp increment were set to 20 mR, then each
time the PD-4™ accumulated 20 mR of dose, no matter what the rate, a chirp would be annunciated. As the unit is moved into higher radiation fields, the time between chirps would decrease alarm setpoints have been exceeded
. This feature gives an audible indication that the field is increasing even if none of the
. Setting the chirp increment to a relatively low value in the
10 to 20 µR range provides a sensitive audible ratemeter. Even in fields as low as 36 mR/h, the PD-4™ would chirp about once each second. Since the PD-4™ can chirp at most once each DSP (0.75 s), chirp rate is not linear with dose rate under high dose rate and low chirp increment conditions
. The Audio Chirp function can be disabled through the PD-4™ EEPROM editing
utility function of the PDRC application program (See Section 5.1).
In a PDE-4™ which is fully equipped with four external probes, chirp requests from all five sensors will be logically "ORed" together. Therefore, if all five sensors are immersed in the same radiation field, the chirp rate will be five times higher than with a PD-4™
. Chirping can be
totally disabled by setting the appropriate bit in the PD(E)-4's EEPROM.
4.2.5 Gamma Icon
The Gamma icon, on the far right of the LCD, is turned on for one DSP each time a photon is detected. As with the PDE-4™ chirp feature, photon events from all PDE-4™ sensors are "ORed" together. Hence, in the case of the PDE-4™, activation of the Gamma icon indicates that a photon was detected either by the internal sensor or by one of the attached external probes. At very low dose rates, the Gamma icon will appear to flash and, at dose rates in excess of 10 mR/h, it will be visible nearly continuously.
4.3 Alarms
PD(E)-4 alarms arise either as the result of a radiation measurement, as the result of a change in communication status, or as the result of a change in the supply voltage.
4.3.1 Dose and Rate Alarms
Radiation-related alarms activate both audible and visible indicators when the alarm value is reached or exceeded
. When the total accumulated dose alarm setpoint is exceeded, the "DOSE"
icon flashes on the LCD and the audio annunciator is activated. Similarly, when the rate alarm setpoint is exceeded, the "RATE" icon flashes and the annunciator is activated. Radiation alarms can be acknowledged (silenced) by pressing the MODE button
. Rate alarms will also
self-acknowledge when the associated sensor is moved into a radiation field below the rate alarm setpoint each has been acknowledged
. Multiple dose and rate alarms will require multiple closures of the MODE button until
. Both DOSE and RATE alarms can be individually and totally
disabled by setting the appropriate bits in the PD(E)-4's EEPROM.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page17
4.3.2 Failed GMT, Failed HVPS and PDE-4 Probe Disconnect Alarms
The PD(E)-4 constantly monitors a number of internal functions and notifies the user by way of an audio alarm when one of these functions is abnormal
. These alarms can all be acknowledged
by pressing the MODE button and include the following:
1) Failed GMT - If no GMT counts are detected from any internal or attached
external radiation sensor within a user-selectable time interval, the Failed GMT alarm is activated. The interval is set in units of 1024 s and would typically be set to a value in the 1 to 5 range.
2) Failed HVPS - If the High Voltage Power Supply is no longer able to maintain the
factory set high voltage on the anodes of the GMTs, the Failed HVPS alarm is activated.
3) Probe Disconnect - If a previously connected PDE-4™ external probe is disconnected
during operation, the Probe Disconnect alarm is activated
. After
acknowledging the alarm, the probe can be reconnected (the PDE­4™ will acknowledge proper connection with a chirp). Note, however, that the recommended procedure for attaching external sensors is to power the PDE-4™ down first. This eliminates the possibility of accumulating false counts as the high voltage connection is made.
4.3.3 Communication Alarms
The Lost Communication Alarm and the Attention/Call alarm can be acknowledged (silenced) by pressing the red CALL button toward the rear of the PD(E)-4. The Attention/Call alarm results from reception of the 'C'- command from the Base Station (or a Repeater) and causes the audio annunciator to be activated. It is intended to initiate a pre-planned procedure such as one of the following:
1) Worker immediately abandons his task and leaves the area.
2) Worker abandons primary task and moves on to secondary task.
3) Worker contacts Base Station operator (using telephone or portable voice radio
link).
Acknowledgment of the Attention/Call alarm not only silences the alarm but also results in the immediate transmission of a standard data message with the Attention/Call-bit of the message's status byte set
. Thus, the Base Station operator can be assured not only that the 'C'-command was
received by the intended PD(E)-4 target but also that it was noticed by the worker.
The Attention/Call command/message sequence can also be reversed. That is, the worker can press his CALL button first to get the attention of the Base Station operator.
The Lost Communication Alarm is activated whenever a Remote's transmission is not acknowledged by its intended recipient
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page18
. The audio alarm will continue until acknowledged by
depressing the CALL button or by re-establishing the communication link. When the transmission link is re-established, the Lost Communication Alarm is re-enabled allowing for additional alarms if the link is subsequently broken again. Typical causes of the Lost Communication Alarm include the following:
1) The Remote temporarily moves out of range of its Base Station (or Repeater) due
either to excessive distance or interfering structures.
2) Communications were already marginal and the worker turns around such that his
body is interposed between his transmitter and the Base Station (or Repeater).
3) The transmitter's RF output stage is powered directly from the battery, but is
limited to 7V. As the battery voltage drops below 7V with use, the range will decrease slightly.
The lost communication alarm can be totally disabled by clearing the appropriate bit in the PD(E)-4's EEPROM from the PDR-1(/S)™ Reader.
4.3.4 LEDS and Audio Alarms (General)
There are three Light Emitting Diodes on the front panel of the PD(E)-4 labeled "AT" (Attention/Call), "RD" (Ready), and "CD" (Carrier Detect). Electrically, the AT-LED is in parallel with the audio alarm and will be illuminated whenever the audio alarm is active which includes the following situations:
1) Powerup reset chirp (also occurs following a PDR-1™ write operation)
2) An unacknowledged dose alarm
3) An unacknowledged rate alarm
4) An unacknowledged CALL from the Base Station
5) An unacknowledged failed sensor alarm
6) A failed high voltage power supply alarm
7) An unacknowledged PDE-4™ external probe disconnect alarm
8) Lost radio communication
9) A dose increment chirp
10) Display mode change
11) Low battery warning (issued only once as a short series of chirps when the low
battery detection threshold is crossed)
The RD-LED is illuminated when the PD(E)-4's microprocessor wakes up the radio modem in preparation for transmission of a data message powered down
. Illumination of the CD-LED indicates that the modem's receiver has detected
. It is extinguished just before the modem is
radio carrier. It will flash at least once per transmission interval when the PD(E)-4's messages are being received and acknowledged by a PDX-4™ RadBase or a Repeater. It may flash several times if multiple PD(E)-4 units are operating in the same vicinity on the same radio channel.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page19
4.3.5 Battery Voltage Thresholds
The PD(E)-4 constantly monitors the terminal voltage of its power source. If the supply voltage dips below 6.53 V (the nominal low battery voltage threshold), the battery display icon begins to blink and a short burst of chirps is emitted from the audio annunciator (no acknowledgment is necessary). If the supply voltage dips below 6.03 V (the nominal shutdown voltage), the PD(E)-4 disconnects power from most of its internal circuitry (to prevent operation at subnormal supply voltage) and crowbars a load resistor across the battery terminals to complete the battery discharge process
. This mechanism prevents the battery voltage from rising under a no-load
condition to a level where the PD(E)-4 would attempt to re-initiate the powerup process.
Crossing of the 6.53 V threshold also starts the "Xmit on Lo-Bat" timer (see Section 5.1, Item 'i')
.
If this timer expires before the 6.03 V threshold is crossed, the battery icon will become solid, indicating that the PD(E)-4 is still operating as a personal dosimeter with no radio communications.
4.4 Display Backlight
The PD(E)-4 includes a display backlight to enhance readability under subdued lighting conditions. It can be activated by pressing the black Backlight button located behind the MODE button.
4.5 External Audio Output Specification
The output from the PD(E)-4's external audio jack (when active) is an AC-coupled, 5 V peak-to­peak square wave of approximately 2.5 kHz. Its output impedance is approximately 3.3 kΩ
resistive.
4.6 Higher Gain Directional Antennas
In some situations, communication range can be increased by replacing the SAIC supplied omni­directional antennas with directional antennas which increase range by decreasing solid angular coverage power radiated by the RadBase in the southerly direction is wasted
. For instance, if all Remotes and Repeaters are located north of their RadBase, then RF
. In this particular case, it
would make sense to employ a directional antenna which concentrates most of its radiated RF power toward the North, thus increasing the RadBase-to-Remote range
. The gain provided by a
directional antenna is equally effective when the RadBase is receiving, so the Remote-to-RadBase range is also increased. It is not difficult to imagine network situations where directional antennas could be advantageously used with Repeaters as well.
The use of directional antennas with Remotes could easily cause some problems since communications would be lost each time the worker turned his blind spot toward the RadBase. However, when Remotes are used as stationary area monitors, the directional antenna can provide a distinct performance improvement.
SAIC PROPRIETARY - Use or disclosure of the information contained on this page is subject to the restrictions
contained on the cover sheet of this document.
Page20
Loading...
+ 46 hidden pages