Leica Geosystems NA2, NAK2 User Manual

Page 1
User Manual
English Version 2.0
Leica NA2/NAK2
Page 2
NA2/NAK2-2.0.0en
2
This manual contains important safety directions as well as instructions for setting up the product and operating it. Refer to "Safety Directions" for further information. Read carefully through the User Manual before setting up the product.
Congratulations on your purchase of a NA2/NAK2 automatic level.
Automatic level
Automatic level
Page 3
NA2/NAK2-2.0.0en
3
The model and the serial number of your product are indicated on the housing. Enter the model and serial number in your manual and always refer to this information when you need to contact your agency or Leica Geosystems authorized service workshop.
Type: Serial no.:
Product identification
Product identification
Page 4
NA2/NAK2-2.0.0en
4
The symbols used in this manual have the following meanings:
DANGER:
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING:
Indicates a potentially hazardous situation or an unintended use which, if not avoided, could result in death or serious injury.
CAUTION:
Indicates a potentially hazardous situation or an unintended use which, if not avoided, may result in minor or moderate injury and/or appreciable material, financial and environ­mental damage.
Important paragraphs which must be adhered to in practice as they enable the product to be used in a technically correct and efficient manner.
Symbols
Symbols used in this manual
Page 5
NA2/NAK2-2.0.0en
5
Introduction ......................................................................... 7
Measurement preparation ...............................................10
Measuring .......................................................................... 14
Checking and adjusting ................................................... 20
Care and Storage .............................................................. 22
Safety Directions .............................................................. 25
Accessories....................................................................... 32
Technical Data .................................................................. 34
Index................................................................................... 35
View of chapters
View of chapters
Page 6
NA2/NAK2-2.0.0en
6
Contents
Introduction ................................................ 7
Special features .......................................... 7
Important parts ........................................... 8
Technical terms and abbreviations ............ 9
Measurement preparation ......................10
Case contents ........................................... 10
Setting up the tripod .................................. 11
Levelling up ............................................... 12
Focusing telescope .................................. 13
Centring .................................................... 13
Measuring ................................................. 14
Height reading .......................................... 14
Distance measuring .................................. 15
Angle measuring ....................................... 15
Line levelling ............................................. 16
Area levelling ............................................ 17
Levelling total station measuring ............. 18
Levelled stakeout ...................................... 18
Precise levelling ........................................ 19
Checking and adjusting .......................... 20
Tripod ........................................................ 20
Circular level ............................................. 20
Checking and adjusting of the
line-of-sight ............................................... 21
Care and Storage ..................................... 22
Transport................................................... 22
In the field ................................................ 22
Inside vehicle ........................................... 23
Shipping ................................................... 23
Storage ..................................................... 23
Cleaning .................................................... 24
Safety Directions ..................................... 25
Intended use of instrument ....................... 25
Permitted uses ......................................... 25
Adverse uses ........................................... 25
Limits of use ............................................. 26
Responsibilities ......................................... 27
Hazards of use ......................................... 28
Accessories .............................................. 32
Technical Data ......................................... 34
Index.......................................................... 35
Contents
Page 7
NA2/NAK2-2.0.0en
Introduction
7
Introduction
Special features
NA-01
The NA2/NAK2 universal automatic level meets all precision, convenience and reliability requirements. It‘s innovative technology will make daily surveying jobs easier.
The instrument is ideally suited for all levelling applications and at all orders of accuracy.
The instantaneous check facility with the push-button control promotes confidence.
• Easy operation, quick to learn!
• Robust and reliable.
• High setting accuracy.
• Endless drive.
• Telescope with excellently­corrected optics for bright, high-contrast images.
• Enables angle measurements with internal glass circle (NAK2 model).
• Optional parallel-plate micrometer for precise fine levelling.
Page 8
NA2/NAK2-2.0.0en
Introduction
8
Important parts
1 Eyepiece of scale
microscope (NAK2 only)
2 Telescope eyepiece
3 Bayonet ring, locks
eyepiece in place
4 Open sight
5 Prism for viewing circular
level
6 Rapid/fine focussing knob
7 Objective housing
8 Push-button for compen-
sator control
9 Footscrew
10 Milled rim, for Hz-circle
setting (NAK2 only)
11 Circular level
12 Base plate
13 Endless horizontal drive
(both sides)
11
6
912
4 5
8 10
21
NA-02
3 7
13
Page 9
NA2/NAK2-2.0.0en
Introduction
9
Technical terms and abbreviations
Backsight/Foresight/ Intermediate sight
For determining the height
difference (H) between the
ground points A and B the back sight (R) is measured first followed by the forward sight (V). Additional points relating to A are measured as intermediate sight (S).
Plumb line
By centring the circular level the instrument is almost level. A small instrument tilt remains (the vertical axis tilt).
Compensator
The compensator in the instrument is responsible for compensating the vertical axis tilt enabling an exactly horizontal aiming. The push button gives the pendulum a slight tap to check the functioning of the compensator.
NA-05
NA--06
NA-04
A
B
R
V/S
(∆H)
Plumb line
Vertical axis tilt
Plumb line
Line-of-sight
Page 10
NA2/NAK2-2.0.0en
Measurement preparation
10
G
e
b
r
a
u
c
h
s
a
n
w
e
is
u
n
g
N
A
7
0
0
k
d
k
j
o
d
k
o
l
m
d
l
k
o
m
öm
l
k
o
k
1 Level
2 User Manual
3 Allen key/Adjusting pins
4 Protective cover & sun
shade
Remove NA2/NAK2 level from the case and check that all components are present.
Case contents
Measurement preparation
NA-35
2
1
4
3
Page 11
NA2/NAK2-2.0.0en
Measurement preparation
11
NA-07
NA-08
When setting up the tripod pay attention to a horizontal position of the tripod plate.
Large inclinations of the tripod must be corrected with the footscrews of the tribrach.
Setting up the tripod
1. Loosen screws of tripod legs, pull out to required length and tighten screws.
2. In order to guarantee a firm foothold sufficiently press the tripod legs into the ground.
When pressing the legs into the ground note that the force must be applied along the legs.
Careful handling of tripod
• Check all screws and bolts for correct fit.
• During transport always use the cover supplied.
Scratches and other damages can result in poor fit and measuring inaccuracies.
• Use the tripod only for surveying jobs.
1.
1.
1.
2.
2.
2.
NA-09
NA-10
NA-11
Page 12
NA2/NAK2-2.0.0en
Measurement preparation
12
Levelling up
1. Place level onto tripod head. Tighten central fixing screw of tripod.
2. Turn footscrews of tribrach into centre position.
3. Centre circular level by turning the foot screws.
Centring the circular level
1. Turn foot screws A and B simultaneously in opposite directions until bubble is in the centre (on the imaginary "T").
2. Turn foot screw C until bubble is centred.
NA-12
NA-13
NA-14
B
A
C
Page 13
NA2/NAK2-2.0.0en
Measurement preparation
13
Focusing telescope
1. Aim telescope against a bright background (e.g. white paper).
2. Turn eyepiece until reticule is sharp-focused and black. Now the eyepiece is adjusted to your eye.
3. Aim telescope on staff using the coarse aiming device.
4. Turn focusing knob until image of staff is sharply focused. If the eye is moved up and down behind the eyepiece the image of the staff and the reticule should not move relative to each other.
For centring over a ground point:
1. Attach plumb bob.
2. Loosen central fixing screw slightly and shift instrument parallel on tripod until the plummet is exactly over the point.
3. Tighten central fixing screw.
Centring
NA-15
NA-16
NA-17
26
25
27
Page 14
NA2/NAK2-2.0.0en
Measuring
14
1
0
Measuring
Before starting field work
or after longer periods of storage/transport of your equipment check the field adjustment parameters specified in this User Manual.
Reduce possible
vibrations by holding the tripod legs.
If the optical parts of your
instrument are dirty of fogged, your measurements can be affected. Keep clean all optical parts of your instrument and follow the cleaning instructions specified in the User Manual.
Before starting work, let
the instrument adjust to the ambient temperature (approximately 2 minutes per ºC of temperature difference).
1. Setup instrument, level and sharp-focus the reticule.
2. Setup level staff vertically (refer also to Instruction Manual of staff).
3. Roughly aim on staff using the coarse aiming device.
4. Sharp-focus using the focusing knob.
Height reading
5. Fine-aim on staff using the endless drives.
6. Check if circular level is centred (view level prism).
7. Press the button to verify that the compensator is functio­ning.
8. Read off height H at the horizontal reticule hair. Example above: H=1.143
d
NA-25
NA-18
H
H
11
12
10
Page 15
NA2/NAK2-2.0.0en
Measuring
15
Result:
Distance d = 100 x L
Carry out steps 1 to 8 according to height reading.
Reading:
Upper distance line: 1.216 m Lower distance line: 1.068 m
Difference L: 0.148 m Distance d: 14.8 m
Distance measuring
Circle reading 314°42’
The NAK2 is equipped with a horizontal circle. The graduation is 1° or 1 gon. To measure an angle bring the vertical hair of the reticule to the centre of the staff. By turning the milled ring, the circle can be set to “0” or any desired reading.
Angle measuring
Circle reading 392.66°
The circle is viewed through the scale microscope and sharply focussed by turning the microscope eyepiece. The Hz-angle is read from the graduation line which protrudes into the scale. Each scale line represents 10’. The minutes are read from the top scale line to the circle graduation line.
NA-23
NA-26
NA-27
11
12
10
L
316
315
314
313
312
394
393
392
391
390
Page 16
NA2/NAK2-2.0.0en
Measuring
16
Project:
Height difference (H) between
point A and B.
Select instrument station
and staff location by pacing off such that approx. the same target distances result (d1
d2; approx. 40 to 50m).
Procedure:
1. Setup instrument at I1.
Line levelling
2. Setup level staff vertically at point A.
3. Aim at staff and read off and take down height (backsight R).
4. Setup level at the changepoint 2, aim at staff and read off and take down height (foresight V).
5. Setup level at I2, aim at staff at the changepoint 2 and read backsight and take down.
6. Carry out a foresight at changepoint 3.
7. Continue in the same way until height at point B is measured.
Result:
∆∆
∆∆
H = sum backsight - sum
foresight
Example of the booking:
NA-28
A
B
2
3
RV
RV
RV
I
1
I
2
I
3
d
1
d
2
H
Point
No.
Back-
sight R
Fore-
sight V
Height
A +2.502 650.100
2 +0.911 -1.803
3 +3.103 -1.930
B -0.981 651.902
Sum +6.516 -4.714
H=
+1.802
Page 17
NA2/NAK2-2.0.0en
Measuring
17
Project:
Height difference of several reference points.
The required accuracy is
usually not very high with such measurements. Nevertheless, from time to time read the staff on a stable intermediate point (reading must remain the same).
Area levelling
Procedure:
1. Set up instrument centrally between the desired points. The instrument telescope may not be below the highest measured intermediate point.
2. Set up staff vertically at reference point A.
3. Aim at staff and read and take down height (=backsight to known point).
4. Set up staff vertically at point 1.
5. Aim at staff and read and take down height (=measuring intermediate point, intermediate sight)
6. Repeat steps 4 and 5 for additional intermediate points.
Height =
Height of station point
+ backsight (A)
- intermediate sight
NA-29
R
S1
A
1
2
3
4
S2
S3
S4
7. The height of individual points are:
Example of booking:
=Instrument horizon
Point
No.
Inte rm .
sight
Height
A 592.00
R1 +2.20
594.20
S1 -1.80 592.40
S2 -1.90 592.30
S3 -2.50 591.70
S4 -2.30 591.90
Page 18
NA2/NAK2-2.0.0en
Measuring
18
Wanted:
Position of several ground points.
The levelling total station measuring is normally
carried out during area levelling.
Levelling total station measuring
Procedure:
1. Sequence of measurements is the same as with area levelling. However, beside the height read also the staff section L (see chapter "Distance measuring") and the Hz-angle.
2. Transfer measured value into the map - points are determined by position and height.
Levelled stakeout
The stakeout is the counterpart to the levelling total station measuring - map points are set out in the field.
Procedure:
1. Set up instrument at a known point, centre and level up.
2. Focus instrument and aim on known orientation point.
3. Orient horizontal circle (Hz­direction).
4. Move staff to stakeout point on the basis of known values (distance and Hz-angle, height) and stakeout point
NA-36
NA-37
R
A
1
2
Backsight to known point Measuring ground points
A
Page 19
NA2/NAK2-2.0.0en
Measuring
19
Project:
Height difference of very high accuracy.
Use the GPM3 parallel plate micrometer and
invar levelling staffs.
In sloping terrain, the
lower part of the staff should not be used, as refraction close to the ground can cause errors.
In sunny weather an
umbrella should be used to shelter the instrument to ensure that the level is protected from the glare of the sun.
Precise levelling
Procedure:
1. Sequence of measurements is the same as with line levelling. However, the best observing length is approx. 25m.
2. When reading the staff, turn the micrometer knob until a graduation line is centred between the wedge-shaped hairs of the reticle.
3. Centimetres are read from the staff and millimetres from the micrometer scale.
4. Leica invar staffs have two sets of graduations. These are read alternatively between foresight and backsight. This provides two independent results and serves as a check.
Example above:
Staff reading = 77 cm GPM3 reading = 0.556 cm Height = 77.556 cm
NA-43
Page 20
NA2/NAK2-2.0.0en
Checking and adjusting
20
12
Checking and adjusting
1. Level up instrument.
2. Turn instrument by 180°.
3. If bubble of level is outside the circle then it should be adjusted (see point 4).
NA-31
Circular level
Connection of individual elements must always be tight.
1. Tighten the Allen screws (2) moderately (if available).
2. Tighten the articulated joints on the tripod head (1) just enough to keep the tripod legs open when you lift it off the ground.
Tripod
NA-30
4. Correct the half error using an adjusting pin and repeat steps 2 and 3 until the bubble of level is in the centre in any telescope direction.
NA-32
/2
1.
2.
3.
Page 21
NA2/NAK2-2.0.0en
Checking and adjusting
21
Checking and adjusting of the line-of-sight
Checking (see figure):
1. Choose a flat terrain between 45 and 60 m ling and divide into three equal sections d.
2. Set up a staff at points A & B.
3. Set up the level at point 1, centre the bubble and press the compensator button.
4. Readings A1 and B1 are taken to the staffs.
5. Set up the level at point 2.
6. Readings A2 and B2 are taken to the staffs.
7. Find nominal reading A2; Reading A1 - B1 + B2
When the difference
nominal-/actual- reading is more than 2 mm the line of sight must be adjusted.
1. Screw off the protective cover.
2. Turn the adjusting screw until
the horizontal hair gives the nominal reading A2 on staff A. The last turn of the adjusting screw should be clockwise.
3. Screw on the protective cover.
4. Repeat the checking
procedure.
NA-45
NA-44
dddAB
A2 B2
A1
B1
1
2
8. Compare the nominal-/actual­reading for A2.
Parallel plate micrometer
When using the parallel plate micrometer to perform precise levelling, follow the same check procedure using invar staffs and micrometer reading. When adjusting, the micrometer is set to the correct value and the adjusting screw is turned until the horizontal hair is on the staff graduation.
Page 22
NA2/NAK2-2.0.0en
Care and Storage
22
When transporting or
shipping the equipment always use the original Leica Geosystems packaging (transport case and shipping cardboard).
After a longer period of
storage or transport of your product always check the field adjustment parameters indicated in this manual before using the product.
Care and Storage
Transport
In the field
When transporting the equipment in the field, always make sure that you
• either carry the product in its original transport container,
• carry the tripod with its legs splayed across your shoulder, keeping the attached product upright.
NA-19
NA-39
Page 23
NA2/NAK2-2.0.0en
Care and Storage
23
Never carry the product loose in a road vehicle, as it can be affected by shock and vibration. Always carry the product in its transport container and secure it.
Storage
Respect the temperature
limits when storing the equipment, particularly in summer if the equipment is inside a vehicle. Refer to "Technical Data" for information about temperature limits.
If the equipment is to be
stored for a long time, remove the alkaline batteries from the GEB63 battery box in order to avoid the danger of leakage.
Inside vehicle Shipping
When transporting the product by rail, air or sea, always use the complete original Leica Geosystems packaging, transport container and cardboard box, or its equivalent, to protect against shock and vibration.
NA-22
NA-21
NA-20
+158° +70°
°F °C
+32°
-40° -40°
Page 24
NA2/NAK2-2.0.0en
Care and Storage
24
Cleaning
Objective, eyepiece:
• Blow dust off lenses and prisms
• Never touch the glass with your fingers
• Use only a clean, soft, lint­free cloth for cleaning. If necessary, moisten the cloth with water or pure alcohol.
Storage, continued
If the instrument becomes wet, leave it
unpacked. Wipe down, clean,
and dry the instrument (at not more than 40 °C/ 104°F), transport case, foam inserts, and accessories. Pack up the equipment only when it is completely dry.
When using the instrument in the field always close the transport case.
Do not use other liquids; these may attack the polymer components.
NA-40
NA-24
Page 25
NA2/NAK2-2.0.0en
Safety Directions
25
Safety Directions
The following directions should enable the person responsible for the product, and the person who actually uses the equipment, to anticipate and avoid operational hazards.
The person responsible for the product directions and adhere to them.
Permitted uses
The level is intended to the following applications:
• Line and area levellings
• Height readings
• Precision levelling with attachable parallel-plate micrometer
• Optical distance measuring with stadia readings
• Angle measurements and staking out with horizontal circle (only NAK2 model)
Intended use of instrument
Adverse uses
• Use of the product without instruction
• Use outside of the intended limits
• Disabling safety systems
• Removal of hazard notices
• Opening the product using tools, for example screwdriver, unless this is specifically permitted for certain functions.
• Modification or conversion of the product
• Use after misappropriation
• Use of products with obviously recognizable damages or defects
Page 26
NA2/NAK2-2.0.0en
Safety Directions
26
• Use with accessories from other manufacturers without the prior explicit approval of Leica Geosystems
• Aiming directly into the sun
• Inadequate safeguards at the surveying site, for example when measuring on roads
Limits of use
Environment:
Suitable for use in an atmosphere appropriate for permanent human habitation: not suitable for use in aggressi­ve or explosive environments. Use in rain is permissible for limited periods (splash-water proof).
DANGER:
Local safety authorities
and safety experts must be contacted before working in hazardous explosive areas, or in close proximity to electrical installations or similar situations by the person in charge of the product.
Adverse uses, contd.
WARNING:
Adverse use can lead to
injury, malfunction and damage. It is the task of the person responsible for the equipment to inform the user about hazards and how to counteract them. The product is not to be operated until the user has been instructed on how to work with it.
Page 27
NA2/NAK2-2.0.0en
Safety Directions
27
Responsibilities
Manufacturer of the product:
Leica Geosystems AG, CH­9435 Heerbrugg, hereinafter referred to as Leica Geosystems, is responsible for supplying the product, including the user manual and original accessories, in a completely safe condition.
Manufacturers of non Leica Geosystems accessories:
The manufacturers of non Leica Geosystems accessories for the product are responsible for developing, implementing and communicating safety concepts for their products, and are also responsible for the effectiveness of those safety concepts in combination with the Leica Geosystems product.
Person in charge of the product:
The person in charge of the product has the following duties:
• To understand the safety instructions on the product and the instructions in the user manual.
• To be familiar with local regulations relating to safety and accident prevention.
• To inform Leica Geosystems immediately if the product and the application becomes unsafe.
WARNING:
The person responsible
for the product must ensure that it is used in accordance with the instructions. This person is also accountable for the training and the deployment of personnel who use the product and for the safety of the equipment in use.
Page 28
NA2/NAK2-2.0.0en
Safety Directions
28
Hazards of use
WARNING:
The absence of instruc-
tion, or the inadequate imparting of instruction, can lead to incorrect or adverse use, and can give rise to accidents with far-reaching human, material, financial and environ­mental consequences.
Precautions:
All users must follow the safety directions given by the manufac­turer and the directions of the person responsible for the product.
CAUTION:
Watch out for erroneous
measurement results if the product has been dropped or has been misused, modified, stored for long periods or transported.
Precautions:
Periodically carry out test measurements and perform the field adjustments indicated in the user manual, particularly after the product has been subjected to abnormal use and before and after important measurements.
DANGER:
Because of the risk of
electrocution, it is very dangerous to use poles and extensions in the vicinity of electrical installations such as power cables or electrical railways.
Precautions:
Keep at a safe distance from electrical installations. If it is essential to work in this environment, first contact the safety authorities responsible for the electrical installations and follow their instructions.
Page 29
NA2/NAK2-2.0.0en
Safety Directions
29
WARNING:
By surveying during a thunderstorm you are at
risk from lightning.
Precautions:
Do not carry out field surveys during thunderstorms.
Hazards of use, continued
WARNING:
Inadequate securing of
the surveying site can lead to dangerous situations, for example in traffic, on building sites, and at industrial installations.
Precautions:
Always ensure that the survey site is adequately secured. Adhere to the regulations governing safety and accident prevention and road traffic.
CAUTION:
Be careful when pointing
the product towards the sun, because the telescope functions as a magnifying glass and can injure your eyes and/or cause damage inside the product.
Precautions:
Do not point the product directly at the sun.
Page 30
NA2/NAK2-2.0.0en
Safety Directions
30
CAUTION:
If the accessories used
with the product are not properly secured and the product is subjected to mecha­nical shock, for example blows or falling, the product may be damaged or people may sustain injury.
Hazards of use, continued
Precautions:
When setting-up the product, make sure that the accessories, for example tripod, staff, staff brace are correctly adapted, fitted, secured, and locked in position. Avoid subjecting the product to mechanical stress. Never position the product on the tripod baseplate without securely tightening the central fixing screw. If the screw is loosened always remove the product immediately from the tripod.
CAUTION:
When using a vertical
staff supported by one brace there is always the danger of falling, for example by wind gusts and therefore danger of damage to equipment and danger of personal injuries.
Precautions:
Never leave a vertical staff supported by a brace unsupervised (person at the staff).
Page 31
NA2/NAK2-2.0.0en
Safety Directions
31
WARNING:
If the product is improperly disposed of,
the following can happen:
• If polymer parts are burnt, poisonous gas are produced which may impair health.
• If batteries are damaged or are heated strongly, they can explode and cause poisoning, burning, corrosion or environmental contamination.
• By disposing of the product irresponsibly you may enable unauthorized persons to use it in contravention of the regulations, exposing themselves and third parties to the risk of severe injury and rendering the environment liable to contamination.
Hazards of use, continued
Precautions:
Dispose of the product appropriately in accordance with the regulations in force in your country. Always prevent access to the product by unauthorized personnel.
Page 32
NA2/NAK2-2.0.0en
32
Accessories
GPM3 Parallel-Plate Micrometer
(Art. No. 356 121)
1 Micrometer knob 2 Open sight 3 Housing of parallel plate 4 Knob for locking GPM3 in
position
For precise levelling the GPM3 is fitted onto the NA2 telescope objective and locked in position with the knob (4). By turning the micrometer knob the parallel-sided glass plate is tilted, thereby causing the line of sight to be raised or lowered. The range of displacement is 10mm. On the scale, seen in the GPM3's eyepiece, the amount of displacement is read directly to
0.1mm and estimated to 0.01mm.
NA-46
Accessories
4
213
Page 33
NA2/NAK2-2.0.0en
33
Accessories, continued
GOA2 Autocollimation eyepiece
(Art. No. 199899) Release the bayonet ring to replace eyepieces. With the autocollimation eyepiece the NA2 can be used for alignment purposes in the laboratory and industry.
GFZ3 Diagonal eyepiece
(Art. No. 734514) The diagonal eyepiece can be fitted for observing from any side when working in confined spaces.
FOK73 Eyepiece 40x
(Art. No. 346475) For precise levelling with the parallel plate micrometer, the FOK73 eyepiece giving 40x magnification may be preferred.
GVO10 Auxiliary lens
(Art. No. 335958) When short focussing distances are required, the auxiliary lens can be fitted to the objective. Measuring range is 1.8 to 0.9m.
GEB60 Eyepiece lamp with cable
(Art. No. 394785) With the eyepiece lamp the NA2 can be used as a collimator for horizontal sighting.
GEB62 Plug-in lamp with cable
(Art. No. 394787) This plug-in lamp can be used with the GOA2 autocollimation eyepiece.
GEB63 Battery box
(Art. No. 394792) Power supply for autocollimation eyepiece and eyepiece lamp.
Accessories
Page 34
NA2/NAK2-2.0.0en
34
Technical Data
Accuracy:
• Standard deviation for 1km double levelling, depending on staffs and procedure up to 0.7 mm
• With GPM3 parallel-plate micrometer (optional) 0.3 mm
Telescope:
• Vertical image
• Magnification,
standard eyepiece 32 x FOK 73 eyepiece 40 x
• Clear objective diameter 45 mm
• Field of view at 100 m 2.2 m
• Shortest target distance from instrument axis 1.6 m
Distance measurement:
• Multiplication factor 100
• Additive constant 0
Compensator:
• Working range ± 30‘
• Setting accuracy (stand. dev.) 0.3“
Technical Data
Circular level:
• Sensitivity 8‘ / 2 mm
Glass circle of NAK2:
• Graduation 360º or 400 gon
• Graduation diameter 70 mm
• Graduation interval 1º or 1 gon
• Optical scale interval 10‘ or 10
c
• Reading by estimation to 1‘ or 1
c
Temperature range:
• Storage -40ºC to 70ºC (-40ºF to 158ºF)
• Operating -20ºC to 50ºC (-4ºF to 122ºF)
GPM3 parallel-plate micrometer:
(optional accessory)
• Range 10 mm
• Interval 0.1 mm
• Estimation 0.01 mm
Page 35
NA2/NAK2-2.0.0en
35
Index
Index
A
Accident prevention ..................................... 29
Angle measuring .......................................... 15
Area levelling ............................................... 17
B
Booking ................................................. 16, 17
Bubble ................................................... 12, 20
C
Central fixing screw ..................................... 12
Centring ........................................................ 13
Circular level .................................. 12, 14, 20
Cleaning ....................................................... 24
Compensator ................................................. 9
D
Distance line ................................................ 15
E
Electrical installations .................................. 28
Environment ................................................. 26
Eyepiece ............................................... 13, 33
F
Focusing ............................................... 13, 33
H
Height difference ............................. 9, 16, 17
Horizontal circle ........................................... 15
Hz-angle ....................................................... 18
Hz-circle ....................................................... 15
L
Line levelling ................................................ 16
Line-of-sight ................................................. 21
O
Optical sight ................................................. 13
P
Packaging .................................................... 23
Plumb bob .................................................... 13
Plummet ....................................................... 13
Precise levelling ..................................... 19, 32
Page 36
NA2/NAK2-2.0.0en
36
Index, continued
Index
R
Rearrangement point ................................... 16
Reticule ................................................. 13, 14
S
Safe distance ............................................... 28
Sharp-focus ................................................. 14
Shipping ................................................ 22, 23
Stakeout ....................................................... 18
Storage .................................................. 22, 24
Storing .......................................................... 23
T
Target distances .......................................... 16
Total station measuring ............................... 18
Transport ...................................................... 22
Transport case ...................................... 22, 24
Tripod ........................................................... 20
V
Vibrations ..................................................... 14
W
Wet instrument ............................................ 24
Page 37
Page 38
Leica Geosystems AG, Heerbrugg, Switzerland, has been certified as being equipped with a quality system which meets the International Standards of Quality Management and Quality Systems (ISO standard
9001) and Environmental Management Systems (ISO standard
14001).
Total Quality Management­Our commitment to total customer satisfaction.
Ask your local Leica Geosystems agent for more information about our TQM program.
Printed in Switzerland - Copyright Leica Geosystems AG, Heerbrugg, Switzerland 2004
Original text
664941-2.0.0en
Leica Geosystems AG
CH-9435 Heerbrugg
(Switzerland)
Phone +41 71 727 31 31
Fax +41 71 727 46 73
www.leica-geosystems.com
Loading...