Lecroy AP034-OM-E-05 User Manual

&DUHDQG0DLQWHQDQFH
CLEANING
!
The exterior of the probe and cable should be cleaned only using a soft cloth moistened with water or isopropyl alcohol. Abrasive agents, strong detergents, or other solvents may damage the probe. Always ensure that the input receptacles are free of debris before connecting accessories.
CALIBRATION INTERVAL
To guarantee accurate performanc e, you should have the probe calibrated every 12 months. Avoid exposing the probe to extreme mechanical shock and excessive bending of the cable because these may alter the calibration.
SERVICE STRATEGY
The AP034 circuits utilize fine pitch surface m ount devices; it is, therefore, impractical to attempt repair in the field. Defective probes must be returned to a LeCr oy service f a cility for diagnos is and exchange. A defective probe under warranty will be replaced with a factory refurbished probe. A probe that is not under warranty can be exchanged for a factory refurbished probe. A modest fee is charged for this s ervice. The defec tive probe mus t be returned in order to receive credit for the probe core.
&DUHDQG0DLQWHQDQFH
TROUBLESHOOTING
If the probe is not operating properly, the problem may be the way in which it is used. Before assuming the probe is defective, perform the following troubleshooting procedures.
A. Trace Off Scale
This is typically caused by improper offset setting, or by an input signal that exceeds the probe differential or common mode range. Perform the following:
1. Remove the input signal from the pr obe, retur n the of f s et to zero, and Autobalance the probe. Does the trace return to approximately the center of the graticule? If not, proceed to step 7.
AP034-OM-E Rev D ISSUED: January 2000 ²
$3$FWLYH3UREH
2. Set the oscilloscope calibrator to output a 100-m V, 1-kHz square wave. Using the flex lead set, connect the probe + input to the calibrator output signal, leave the – input open. Set the oscilloscope to 50 mV/div and 500 µs/div. Is the displayed waveform a 100-mV, 1-kHz square wave with the correct polarity?
3. Repeat step 2 with the – input connected to the c alibrator and the + input left open. In this case, the displayed waveform should be inverted.
4. Connect both the + input and – input to the calibrator output. Is the trace approximately a flat line near zero volts?
5. If steps 1 to 4 give the correct results, the problem is likely a result of the input signal exceeding the differential or common mode range.
6. Connect both the + input and – input to one of the two input signals. If the trace is off sc ale, the input signal is probably exceeding the common m ode range. Repeat with the other input signal.
7. Is a ProBus Power Supply Overload error message displayed? If so, remove all other ProBus ac cessories fr om the oscilloscope. Is the message still displayed? If so, remove the AP034. Is the message still displayed? If so, the oscilloscope should be returned for service.
8. If, after removing the AP034, the Power Supply Overload error message is not displayed, the problem m ay be either the probe or the oscilloscope. Repeat the test with a different ProBus acces sory. If the mess age does not return with a different accessory, the AP034 m ay be def ective and should be returned for service.
B. Incorrect Frequency Response
Possible causes are a defective probe or oscilloscope, poor connections, or poor grounding. Try the following:
1. Verify that the BW limiting of the oscilloscope is off.
² ISSUED: January 2000 AP034-OM-E Rev D
C. DC Errors
&DUHDQG0DLQWHQDQFH
2. Connect the probe to another oscilloscope. If the probe now measures properly, the problem may be in the oscilloscope.
3. If the probe behaves as if it is ac-coupled at high frequency, check for an open input connection.
4. Poor frequency or transient response and AC gain errors may result when one of the two input connections is open.
5. Excessive “ring” and other transient problems can result from exces sive lead length. To test this, shorten the input leads to less than 1 cm. If the transient response changes significantly, the lead parasitics are the cause.
Incorrect DC gain requires recalibration or factory repair. This can be determined by completing the gain checks in the Performance Verification Procedure.
1. Extremely high source resistance will result in DC gain errors. Check the probe accuracy with the oscilloscope calibrator signal.
2. Verify that the probe is not being overdriven into clipping for its current gain setting.
3. Exces sive offset can result from large changes in ambient temperature. Remove the input s ignal from the probe and repeat the Autobalance cycle. With the Of fset set to zero, did the trace return to the center of the graticule?
D. Poor Common Mode Rejection
Use the 1-kHz calibrator signal from the oscilloscope to check common mode rejection. With both the + input and – input connected to the calibrator signal, a f lat line at zero volts should be seen on the graticule.
1. Check the probe with the plug-on attenuator installed and removed. If excessive com mon mode signal appears only when the attenuator is present, the attenuator m ay need to be rematched to the probe. Use the proc edure listed in this section to match the attenuator.
2. If the common mode signal appears when the probe is connected to the test circuit, but not when it is attached to
AP034-OM-E Rev D ISSUED: January 2000 ²
RETURNING A DEFECTIVE PROBE
The procedure for returning a defec tive probe to be exchanged, is as follows:
Contact your local LeCroy sales representative to find out where to return the product. All returned products should be identified by model and serial number. You should describe the defect or failure, and provide your name and contact num ber. In the case of products returned to the factory, a Return Authorization Number (RAN) should be used. T he RAN can be es tablished by contacting your nearest LeCroy office, or the New York Customer Care Center.
Return shipments should be made prepaid. LeCroy cannot accept COD or Collect Return shipments. We recommend air­freighting. It is important that the RAN be clearly shown on the outside of the shipping package for prompt redirection to the appropriate LeCroy department.
$3$FWLYH3UREH
the calibrator, the problem may be caused by large mismatches in the source impedance. T ry connecting both inputs to one of the input signals in the test circuit, then the other. If the comm on mode signal disappears, try probing lower impedance points within the circuit.
1. Contact your local LeCroy sales or service representative to obtain a Return Authorization Number.
2. Remove all accessories, including all removable cables, from the probe.
3. Pack the probe in its case, surrounded by the original packing material (or equivalent) and box.
4. Label the case with a tag containing:
The RAN
Name and address of owner
Instrument model and serial number
Description of the failure mode
5. Package the pr obe case in a cardboard shipping box with adequate padding to avoid damage in transit.
² ISSUED: January 2000 AP034-OM-E Rev D
Loading...
+ 8 hidden pages