AP015 Current Probe Instruction Manual
Specifications
The following specifications are valid for model AP015 current probes after the probe has reached operating
temperature, which is 20 minutes with power applied in a environment with stable ambient temperature. The
probe must be operating within the environmental conditions listed in the General Characteristics section, and
has been calibrated within the past 12 months in a ambient temperature of 23 ±5
Nominal Characteristics
Nominal characteristics describe parameters and attributes which are guaranteed by design, but do not have
associated tolerances.
INTERFACE: PROBUS
COUPLING: AC, GND, DC
MAXIMUM CURRENT, DC±30A (CONTINUOUS)
±50A MAXIMUM (<10S)
SENSITIVITY: 10 MA/DIV – 20 A/DIV
o
C.
OFFSET RANGE ±100A MAXIMUM
MAXIMUM CONDUCTOR SIZE 5 MM
MAXIMUM VOLTAGE, BARE CONDUCTOR 300 V CAT I
150 V CAT II
CABLE LENGTH 2 M
WEIGHT, PROBE ONLY300 G (10.6 OZ.)
Warranted Electrical Characteristics
Warranted characteristics describe parameters which have guaranteed performance. Unless otherwise noted,
tests are provided in the Performance Verification Procedure for all warranted specifications.
DC ACCURACY:±1% OF READING TO 15A (AT 25O C, AFTER DEGAUSS)
OF READING TO 30A (AT 25O C, AFTER DEGAUSS)
±2%
BANDWIDTH DC TO 50MHZ
Typical Electrical Characteristics
Typical characteristics describe parameters which do not have guaranteed performance, however are
representative of the average performance from a sample of several probes. Tests for typical characteristics
are not provided in the Performance Verification Procedure.
WOE-CS-TS02-009v002 1/111
AP015 Current Probe Instruction Manual
RISE TIME: < 7 NS
SLEW RATE> 1.6 A/NS ( SENSITIVITY ≥ 1 A/DIV)
INSERTION IMPEDANCE< 0.06Ω AT 5 MHZ
General Characteristics
TEMPERATURE: 0 – 40 O C OPERATING
RELATIVE HUMIDITY80% MAXIMUM UP TO 31O C
USAGE ENVIRONMENT INDOOR
ALTITUDE 2000 M MAXIMUM
POWER REQUIREMENTS:POWERED FROM OSCILLOSCOPE THROUGH PROBUS INTERFACE. NO RESTRICTION ON
NUMBER OF PROBES USED WITH OSCILLOSCOPE.
WOE-CS-TS02-009v002 2/111
AP015 Current Probe Instruction Manual
PERFORMANCE VERIFICATION
This procedure can be used to verify the warranted characteristics of the AP015 Current Probe.
The recommended calibration interval for the model AP015 Current Probe is one year. The complete
performance verification procedure should be performed as the first step of annual calibration. Test results can
be recorded on a photocopy of the Test Record provided at the end of this manual.
Performance verification can be completed without removing the instrument covers or exposing the user to
hazardous voltages. Adjustment should only be attempted if a parameter measured in the Performance
Verification Procedure is outside of the specification limits.
Adjustment should only be performed by qualified personnel.
Test Equipment Required
The following table lists the test equipment and accessories, or their equivalents, which are required for
performance verification of the AP015 Current Probe.
Because the input and output connector types may vary on different brands and models of test instruments,
additional adapters or cables may be required.
List of Required Equipment
Description Minimum Requirements Test Equipment Examples
Wide Band oscilloscope 1 Minimum 200 MHz bandwidth
Digital Multimeter
(2 required)
Leveled Sine Wave Generator. Relative output level accurate to
DC Power Supply Output adjustable to >6 V at 1 A
Calibration Fixture, 50 Turn Loop
Calibration Fixture, High Speed Shunt
Calibration Fixture ProBus Extension Cable LeCroy PROBUS-CF01
BNC Adapter, T Male to dual Female Pomona 3285
BNC coaxial cable
Banana Plug adapter BNC female to banana plug. Pomona 1269
Patch Cables
(4 reequired)
Preliminary Procedure
ProBus interface equipped
Software Version 7.6.0 or greater
DC: 0.1 % accuracy
5½ digit resolution
with Null capability
(referenced measurements)
± 0.3 dB into 50 Ω
1 – 100 MHz frequency range
Output adjustable to 18 dBm
50 Turn loop in series with 0.5 Ω ± 0.2%
resistor with sense terminals
50 Ω Termination into single turn loop.
VSWR < 1.25:1 DC – 50 MHz
Male-male BNC, 50 Ω, 36”
Male Banana – Male Banana, 12” Pomona B-12-0 (black)
TABLE 1
LeCroy LT322
Agilent Technologies 34401A,.
Fluke 8842A-09,.
Keithley 2001
Tegam SG503 with TM series mainframe and
matching output cable.
A semiautomatic software leveled signal source
calibrated with a power meter may be substituted.
Agilent Technologies E3610A
LeCroy CP015-CF02
LeCroy CP015-CF01
Pomona 5697-36
B-12-2 (red)
1. Connect the AP015 to the channel 1 input of the oscilloscope, and completely close the probe
slider.
WOE-CS-TS02-009v002 3/111
AP015 Current Probe Instruction Manual
2. Turn the oscilloscope on and allow at least 30 minutes warm-up time for the AP015 and test
equipment before performing the Verification Procedure.
3. Turn on the other test equipment and allow these to warm up for the time recommended by the
manufacturer.
4. While the instruments are reaching operating temperature, make a photocopy of the Performance
Verification Test Record (located at the end of this manual), and fill in the necessary data.
Functional Check
The functional check will verify the basic operation of all probe functions. It is recommend that the
Fucntional Check be performed prior the Performance Verification Procedure.
1. Select Channel 1 and enter the
2. Verify that
Probe sensed (AP015) is displayed on the right hand menu.
coupling menu and verify that the Global BWL is set to OFF.
3. Unlock the slider on the probe by pushing the lever towards the probe cable. Release the lever so
the input slider remains closed, but unlocked.
4. Verify that the message:
Warning: AP015 Probe Unlocked… is displayed on the top of the
screen.
5. Lock the probe slider by pushing fully away from the probe cable.
6. Verify that the warning message dissappears.
7. Degauss the probe by pressing the
DEGAUSS button, (located on the coupling menu), twice.
8. Verify that no error message remains displayed on the top of the screen.
WOE-CS-TS02-009v002 4/111