Lecroy 93XXC-OM-E07 User Manual

Page 1
&KDQQHO&RQWUROV
These enable selection of displayed traces and adjustment of vertical sensitivity and offset.
75$&(212)) Pressing these buttons either
active
channel models ( channel has its own set of unique, dedicated controls.
&+$11(/6&RXSOLQJ3UREHV
channel. On two-
right
), each
6(/(&7&+$11(/ On four-channel models
right
), these buttons are
( used to attribute all the vertical controls to one channel, independent of whether or not it is the channel displayed. The selected channel number is highlighted in the Acquisition Summ ary field
see previous chapter
(
2))6(7 — vertically positions the active channel.
),1' — automatically adjusts
²
).
Page 2
&+$11(/6&RXSOLQJ3UREHV
92/76',9 — selects the vertical sensitivity factor either in a 1–2–5 sequence or
continuously ( acquisition offset can be chosen from the “SPECIAL MODES” menu.
9$5 — allows the user to determine whether the VOLTS/DIV k nob will
modify the vertical sensitivity in a continuous manner or in disc rete 1–2–5 steps.
The format of the vertic al sens itivity in the Acquisition Summ ar y field (bottom left of screen) shows whether the VOLTS/DIV knob is operating in
see VAR, below
continuous
or
). The effect of gain changes on the
stepping
mode.
&283/,1* — menu-entry button that accesses the “Coupling” m enus (
).
section
see next
²
Page 3
&RXSOLQJ
&RXSOLQJ0HQXV Press for access to selection of:
À Coupling and grounding of each input channel À ECL or TTL gain, offset and coupling preset for the
channel shown
À Bandwidth limiter for all channels À Probe attenuation of each input channel.
&RXSOLQJ
Used to select the input channel’s coupling. If an overload is detected, the instrument will automatically set the channel to the grounded state: the menu can then be reset to the desired coupling.
9GLY2IIVHW
When NORMAL is highlighted, pressing the corresponding menu button sets the offset, Volts/div, and input coupling to display ECL signals. Press the button a second time and the settings for TTL signals are given. And a third time r eturns the s ettings to those used at the last manual setup of the channel.
*OREDO%:/
To turn the bandwidth limit “OFF” or “ON”. The bandwidth can be reduced from 500 MHz or 1 GHz, to either 200 MHz or 25 MHz, or 30 MHz (–3dB), depending on the model ( Bandwidth limiting can be useful in reducing signal and system noise or preventing high-frequency aliasing, reducing — for example — any high-frequency signals that may cause aliasing in single-shot applications.
see
Appendix A).
1RWH
This command is global and affects all input channels.
3UREH$WWHQ
Sets the probe attenuation factor related to the input channel (
following for probe details
²
).
see
Page 4
&+$11(/6&RXSOLQJ3UREHV
3UREHVDQG3UREH&DOLEUDWLRQ
3UREH&DOLEUDWLRQ To calibrate the probe supplied, connect it to one of the input
channels’ BNC connectors. Connect the probe’s grounding alligator clip to the CAL BNC ground and touch the tip to the inner conductor of the CAL BNC. The CAL signal is a 1 kHz square wave, 1 V p-p.
Set the channel coupling to DC 1 MΩ, turn the trace ON and push AUTO SETUP to set up the oscilloscope. If over- or undershoot of the displayed signal occurs, the probe can be adjusted by inserting the small screwdriver, supplied with the probe package, into the trimmer on the probe’s barrel and turning it clockwise or counter- clockwise to achieve the optimal s quare­wave contour.
0RUH2Q&RXSOLQJ In the AC position, signals are coupled capacitively, thus
blocking the input signal’s DC component and limiting the
signal frequencies below 10 Hz. In the DC position, all signal frequency components are allowed
to pass through, and 1 M or 50 Ω may be chosen as the input impedance.
The maximum dissipation into 50 Ω is 0.5 W and inputs will automatically be grounded whenever this is attained. An overload message will be displayed in the Acquisition Summary
Field and “Grounded” will be indicated in the “Coupling” menu. The overload condition is reset by removing the signal from the input and again selecting the 50 Ω input impedance from the menu.
²
Page 5
3UR%XV6\VWHP LeCroy’s ProBus system provides a complete measurement solution
from probe tip to osc illos c ope display. This intelligent interc onnec tion between LeCroy oscilloscopes and a growing range of accessories is achieved via a six-wire bus following Philips’ I
2
C protocol. It provides major benefits over standard BNC or even probe-ring connections:
À Autosensing the probe type, eliminating all the guesswork —
and the errors — from manually setting attenuation or amplification factors, and ensuring proper input coupling.
À Transparent gain and offset control right from the front panel
— particularly useful for FET (
FET menus shown here
) and
current probes.
À Gain and offset correction factors are uploaded from the
ProBus EPROMS on FET probes and automatically compensated to achieve fully calibrated measurements.
&RXSOLQJ
Used to select the input channel’s coupling. If an overload is detected, the instrument will automatically set the channel to the grounded state: the menu can then be reset to the desired coupling.
9GLY2IIVHW
When NORMAL is highlighted, pressing the corresponding menu button sets the offset, Volts/div, and input coupling to display ECL signals. Press the button a second time and the settings for TTL signals are given. And a third time r eturns the s ettings to those used at the last manual setup of the channel.
*OREDO%:/
To turn the bandwidth limit “OFF” or “ON”. The bandwidth can be reduced from 500 MHz or 1 GHz, to either 200 MHz or 25 MHz, or
see
30 MHz (–3dB), depending on the model (
Appendix A). Bandwidth limiting can be useful in reducing signal and system noise or preventing high-frequency aliasing, reducing — for example — any high-frequency signals that may cause aliasing in single-shot applications.
When a FET probe is used, “Probe sensed…”, automatically appears to indicate settings. When other ProBus probes are used, this is redefined.
²
Loading...