Four Channels
50k and 200k Point Records
DOS Compatible Floppy Disk, PCMCIA portable hard drive and
Memory Card Options
Glitch, Window, Qualified, Interval, Dropout and Video Triggers
8-
bit vertical resolution, 11 with ERES option
Fully Programmable via GPIB and RS-232
Automatic PASS/FAIL testing
Persistence, XY and Roll Modes
Advanced Signal Processing
Internal High Resolution Graphics Printer Option
The 9304A and 9304AM are general purpose 200 MHz four channel digital
oscilloscopes. They capture single-shot events at up to 100 MS/s, and repetitive
signals at 10 GS/s. Record lengths up to 200k points pr
resolution, and allow fast digitizing of long
segmented, for minimum dead time between acquisitions.
Live waveforms on the main timebase may be viewed simultaneously with up to
3 expansions, show
highlights on the main trace.
The LeCroy ProBus intelligent probe system allows automatic sensing of the
probe type. For LeCroy's active FET probes it also provides variable offset at the
probe tip. Of
SMART Trigger modes like Glitch, Window and Dropout allow you to capture
precisely the events of interest.
A comprehensive range of signal processing functions, on live or stored
waveforms, a
data.
The 9304A and 9304AM feature the proven user-interface of LeCroy's portable
scope family. A bright, high-resolution 9" CRT allows optimum waveform viewing
on a high resolution 810 x 696 p
are arranged around the graticules - they never overwrite the waveforms. Each of
the main control functions has a dedicated single knob, to keep the scope's
performance at your fingertips.
DOS compatible fl
options store waveforms and test setups, and make transferring data to a PC
easier than ever. Hardcopies can be made on GPIB, RSprinters or plotters. An optional internal high resol
available.
Optional packages provide extensive Waveform Processing including FFT and
Enhanced Resolution to 11 bits.
-C
ovide excellent horizontal
-duration events. Memories can be
ing all of the signal detail. Expansions are shown as
fset and coupling are controlled from the scope's front panel.
llows waveform manipulation without destroying the underlying
ixel screen under any conditions. Menus and text
oppy disk, PCMCIA portable hard drive and memory card
232-C or Centronics
ution graphics printer is also
ACQUISITION SYSTEM
Bandwidth (
@ 50 Ohms DC to 200 MHz
@ 1 M Ohm DC: DC to 200 MHz typical at t
No. of Channels:
No. of Digitizers:
Maximum Sample Rate:
Acquisition memories, per channel:
9304A 50k
9304AM 200k
Sensitivity:
Scale factors:
Offset Range:
2.0 - 9.9 mV/div: ± 120 mV
10 -
199 mV/div: ± 1.2 V
0.2 - 5.0 V/div: ± 24 V
DC Accuracy:
Vertical Resolution:
Bandwidth Limiter:
Input Coupling:
Input Impedance:
Max Input:
1 M Ohm: 250 V (DC+peak AC@ 10 kHz)
50 Ohms: ± 5 V DC (500 mW) or
5 V RMS
-3 dB)
4
4
100 MS/s simultaneously on each channel.
2 mV/div to 5 V/div, fully variable.
A wide choice of probe attenuation factors are selectable.
< or equal to ± 2% full scale (8 divisions) at 0 V offset.
8 bits.
30 MHz.
AC, DC, GND.
1 M Ohm//15 pF or 50 Ohm ± 1%.
he probe tip.
TIME BASE SYSTEM
Timebases: Main and up to 4 Zoom Traces.
Time/Div Range:
Clock Accuracy:
Interpolator Resolution:
Roll Mode:
For > 50k points: 10 s to 1,000 s/div.
External Clock:
0 to 100% of full scale (adjustable in 1% div increments).
Slope).
ECL, TTL or zero
Post-trigger delay:
Holdoff by time:
Holdoff by events:
Internal Trigger Sensitivity Range:
± 5 div.
EXT Trigger Max. Input:
1 M Ohm//15 pF: 250 V (DC+peak AC < or equal to 10 kHz)
50 Ohms ± 1%: ± 5 V DC (500 mW) or 5 V RMS
EXT Trigger Range: ±0.5V (±5V with Ext/10).
Trigger Timing:
SMART TRIGGER TYPES
Signal Width:
Signal Interval:
20s.
Dropout:
25ns to 20s.
State/Edge Qualified:
has occurred on another source. The delay b
as a number of events on the trigger channel or as a time interval.
TV:
SECAM, NTSC or non-standard video.
Trigger if the input signal drops out for longer than a time-out from
Allows selection of both line (up to 1500) and field number (up to 8) for PAL,
0 to 10,000 divisions (adjustable in 0.1 div increments).
10 ns to 20 s.
0 to 99,999,999 events.
Trigger Date and Time are listed in the Memory Status Menu.
Trigger on width between two limits selectable from 2.5ns to 20s.
Trigger on interval be-tween two
Trigger on any source only if a given state (or transition)
limits selectable from 10ns to
etween these events can be defined
ACQUISITION MODES
Random Inter
ms/div.
Single shot:
Sequence:
Number of segments available:
9304A 2-200
9304AM 2
Dead Time between segments:
< or equal to 150 us
DISPLAY
Waveform style:
highlighted as dots. Vectors may be switched off.
CRT:
Resolution:
Modes:
Real-time Clock:
Graticules: Internally generated; separate intensity control for grids and
waveforms.
Grids:
Formats: YT, XY, and both together.
12.5 x 17.5 cm (9" diagonal) raster.
Normal, X-Y, Variable or Infinite Persistence.
1, 2 or 4 grids.
leaved Sampling (RIS):
for transient and repetitive signals from 50 ns/div.
Stores multiple events in segmented acquisition memories.
-500
Vectors connect the individual sample points, which are
810 x 696 points.
Date, hours, minutes, seconds.
for repetitive signals from 1 ns/div to 5
Vertical Zoom:
sensitivity).
Maximum Horizontal Zoom Factors:
9304A 1,000x
9304AM 5,000x
INTERNAL MEMORY
Waveform Memory: Up to four 16-bit Memories (M1, M2, M3, M4).
Processing Memory:
(A,B,C,D).
Setup Memory:
portable hard drive or Floppy Disks may also be used for high-capacity waveform
and setup storage.
WAVEFORM PROCESSING
Up to four processing functions may be performed simultaneously. Functions
available are: Add, Subtract, Multiply, Divide, Negate, Identity, Sin(x)/x and
Summation Averaging.
Average:
Up to 106 averages are possible with Option WP01.
Envelope*:
ERES*:
Sampled data is always available, even when trace is turned off. Any of the
above modes can be invoked without destroying the data.
FFT*:
* Envelope and ERES modes are provided in Math
WP02.
Summed a
Low-Pass digital filter provides up to 11 bits vertical resolution.
Spectral Analysis with four windowing functions and FFT averaging.
Up to 5x Vertical Expansion (50x with averaging, up to 40 mV
Up to four 16
Four non-volatile memories. Optional Memory Cards, PCMCIA
veraging of up to 1,000 waveforms in the basic instrument.
Max, Min, or Max and Min values of from 1 to 106 waveforms.
-bit Waveform Processing Memories
Package WP01, FFT is in
AUTOMATIC MEASUREMENTS
The following Parametric measurements are available, together with statistics of
their Average, Highest, Lowest values and Standard Deviation:
amplitude
area
base
cmean
cmedian
crms
f80-
delta t at level (t=0,abs)overshoot +
delta t at level (t-0%)
duty cycle
falltime
20%
f@level (abs)
overshoot
peak to peak
period
risetime
r20-80%
-
csdev
cycles
delay
delta delay
delta t at level (abs)median
delta t at level (%)
Parameters are calculated as defined by ANSI/IEEE Std
Pulse Measurement and Analysis by Objective Techniques". In addition, Rise
and Fall times may be measured at 10 % and 90% levels, or 20% and 80%
levels, or any other user-specified levels.
delta delay provides time between midpoint t
propagation delay measurements.
delta t at level allows the same measurement to be made at any specified level.
Two cursors are used to define the region over which these parameters are
calculated.
Relative Time:
±0.05% full scale for unexpanded traces; up to 10 % of the sampling interval for
expanded traces. The corresponding frequency value is also displayed.
Relative Voltage:
of fullscale in single-grid mode.
Absolute Time:
voltage with respect to ground.
Pass/Fail
selectable thresholds. Waveform Limit Testing is performed using templates
which may be defined inside the instrument.
Two
A cross hair marker measures time relative to the trigger, and
testing allow up to five of the listed
f@level (%)
frequency
maximum
mean
minimum
cursors provide time measurements with resolution of
Two horizontal bars measure volt
top
ransition of two sources, for making
r@level (abs)
r@level (%)
RMS
std dev
width
parameters to be tested against
181-1977 "Standard on
age differences up to ±0.2%
AUTOSETUP
Pressing Autosetup sets timebase, trigger and sensitivity to
of repetitive signals. (Amplitude 2mV to 40V; frequency above 50Hz; Duty cycle
greater than 0.1%).
Autosetup Time:
Vertical Find:
PROBES
Model:
One PP002 (10:1,
The 9304A and 9304AM are fully compatible with LeCroy's range of FET Probes,
which may be purchased separately.
Probe calibration:
Approximately 2 seconds.
Automatically sets sensitivity and offset.
10 M Ohm // 14 pF) probe supplied per channel.
Max 1 V into 1 M Ohm,
display a wide range
500 mV into 50 Ohms , frequency and amplitude program
wave selectable, rise and fall time 1 ns typical. Alternatively, the Calibrator output
can provide a trigger output or a PASS/FAIL test output.
INTERFACING
Remote Control:
possible by GPIB and RS-232-C.
RS-232-C Port:
or printer/plotter connection.
GPIB Port: (IEEE-488.1) Configurable as talker/listener for computer control and
fast data transfer.
Comma
Centronics Port: Optional hardcopy parallel interface included with floppy disk
and graphics printer options.
Hardcopy:
control. TIFF a
programs. The following printers and plotters can be used to make hardcopies:
HP DeskJet (color or B&W), HP ThinkJet, QuietJet, LaserJet, PaintJet and
EPSON printers.
HP 7400 and 7500 series, or HPGL compatible plotters. An optional internal high
resolution graphics printer is also available, see page 53.
nd Language complies with requirements of IEEE-488.2.
Of all front
Asynchronous up to 19,200 baud for computer/terminal control
Screen dumps are activated by a front-panel button or via remote
nd BMP formats are available for importing to Desktop Publishing
-panel controls, as well as all internal functions is
mable, pulse or square
GENERAL
Auto-calibration ensures specified DC and timing accuracy.
Temperature:
5° to 40° C (41° to 104° F) rated 0° to 50° C (32° to
Humidity:
Shock & Vibration:
Meets MIL-STD
MIL
-T-
Power: 90-250 V AC, 45-66 Hz, 150 W.
Battery Backup:
Dimensions: (HWD) 8.5"x14.5"x16.25", 210mm x 370mm x 410mm.
Weight:
18kg (40lbs) shipping.
< 80%.
28800C.
12.5kg (27.5lbs) net,
-810C modified to LeCroy design specifications and
Front-panel settings maintained for two years.
122° F) operating.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.