Leadshine 307P Operating Instructions Manual

Page 1
INTRODUCTION
This instrument is a portable 3½ digit, compact-sized digital thermometer designed to use external K-type thermocouple as temperature sensor. Temperature indication follows National Bureau of Standards and IEC 584 temperature/voltage tables for K-type thermocouples. One K-type thermocouple is sup­plied with the thermometer.
SAFETY INFORMATION
It is recommended that you read the safety and op­eration instructions before using the thermometer. Designed to meet IEC-1010-1, CE-EMC.
WARNING
TO AVOID ELECTRICAL SHOCK, DO NOT USE THIS INSTRUMENT WHEN VOLTAGES AT THE MEASUREMENT SURFACE EXCEED 24V AC OR 60V DC.
WARNING
TO AVOID DAMAGE OR BURNS, DO NOT MAKE TEMPERATURE MEASUREMENTS IN MICRO­WAVE OVENS.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To prolong lead life, avoid sharp bends in the leads, especially near the connector. The symbol on the instrument indicates that the operator must refer to an explanation in this manual.
SPECIFICATIONS
ELECTRICAL
Temperature Scale:
Celsius or Fahrenheit user-selectable
Measurement Range:
-50°C to 1300°C, ( -58°F to 2000°F )
Resolution: 1°C or 1°F, 0.1°C or 0.1°F Accuracy:
Accuracy is specified for operating temperatures over the range of 18°C to 28°C ( 64°F to 82°F ) , for 1 year,not including thermocouple error. ±( 2°C ) -50°C to 0°C ±( 0.3% rdg + 1°C ) 0°C to 600°C ±( 0.5% rdg + 1°C ) 600°C to 1300°C ±( 4°F ) -58°F to 32°F ±( 0.3% rdg + 2°F ) 32°F to 1100°F ±( 0.5% rdg + 2°F ) 1100°F to 2000°F
Temperature Coefficient:
0.1x( specified accuracy ) per °C . ( 0°C to 18°C , 28°C to 50°C ).
Input Protection:
60V dc or 24V rms ac maximum input voltage on any combina­tion of input pins.
Reading Rate:
2.5 times per second.
Input Connector:
Accepts standard miniature therm ocouple connectors ( flat blades spaced 7.9mm , center to center ) .
ENVIRONMENTAL
Ambient Operation Range:
0°C to 50°C, (32°F to 122°F)
Storage Temperature:
-20°C to 60°C, (-4°F to 140°F)
Relative Humidity:
0% to 80% (0°C to 35°C) (32°F to 95°F) 0% to 70% (35°C to 50°C) (95°F to 122°F)
GENERAL
Display:
3½ digit liquid crystal display (LCD) with maxi­mum reading of 1999
Battery:
Standard 9V battery (NEDA 1604,IEC 6F22)
Battery Life:
100 hours typical with carbon zinc battery 200 hours typical with alkaline battery
Dimensions: 184mm (H) x 62mm (W) x35mm (D) Weight: 10.6oz (300g) including holster Supplied Probe:
4-foot type “K” thermocouple bead probe (teflon tape insulated). Maximum insulation temperature 260°C (500°F). Probe accuracy ±2.2°C or ±0.75% of reading (whichever is greater) from 0° to 800°C.
OPERATING INSTRUCTIONS
DIGITAL THERMOMETER
T1 T2
T1-T2
ON OFF
C
°1°
0.1°/F
°
/
MIN
MAX/
HOLD
THERMOMETER
OFFSET
T1 T2
Page 2
OPERATING INSTRUCTIONS
Selecting the Temperature Scale
Readings are displayed in either degrees Celsius (°C) or degrees Fahrenheit (°F). When the thermometer is turned on, it is set to the temperature scale that was in use when the thermometer was last turned off. To
change the temperature scale, press the °C or °F key.
Single-Thermocouple Temperature Measurement
The thermometer displays the temperature of the thermocouple that is connected to the selected input. Press the T2 key to display the temperature of the thermocouple connected to the T2 input. Press the T1 key to display the temperature of the themocouple connected to the T1 input. The input selection cur­sor indicates which input is selected.
Differential Temperature Measurement
Differential temperature measurement is selected by pressing the T1-T2 key. This causes the thermom­eter to display the temperature difference between the two thermocouples ( the temperature of thermo­couple T1 minus the temperature of thermocouple T2 ). The selection is indicated by the input selec­tion cursor.
Selecting the Display Resolution
The thermometer allows two choices of resolution:
High resolution: 0.1°C or 0.1°F Low resolution: 1.0°C or 1.0°F
To select the alternate display resolution, press the corresponding “1°” or “0.1°” key.
The 0.1° resolution is applicable for temperature measurements below 200°C or 200°F.
HOLD Mode
Pressing the HOLD key to enter the Data Hold mode, the “HOLD” annunciator is displayed. When HOLD mode is selected, the thermometer holds the present readings and stops all further measurements. Pressing the HOLD key again cancels HOLD mode, causing the thermometer to resume taking measurements.
MIN/MAX Mode
Press MIN / MAX once to begin recording MIN and MAX. Press MIN / MAX to select MIN or MAX. Hold down for 2 seconds to exit MIN / MAX function.
OFFSET ADJUSTMENT
The OFFSET control is set at the factory to allow for the variations found in standard thermocouples. By adjusting the OFFSET control, you can optimize measurement accuracy for a particular thermocouple at a particular temperature.
Adjusting for T1 or T2 Accurate Measurements
1.Connect the thermocouple to the T1 or T2 input connector and turn the thermometer ON , then press the 0.1° key to select the high display resol­ution. ( If using T2 , press the T2 key ).
2.Place the thermocouple in a known , stable tem­perature environment at or near the temperature you wish to measure , and allow the readings to stabilize.
3.Slowly adjust the OFFSET control so that the ther-
mometer reading matches the temperature of the known environment. Leave sufficient time be­tween adjustments to allow for measurement lag.
4.The calibration of the thermometer-thermocouple combination is now optimized for measurements near the temperature measured in step 2.
Resetting the OFFSET Control
To return the OFFSET control to the factory setting without having to recalibrate the thermometer, per­form the following procedure:
1.Connect a thermocouple that is in good working order to the input that is to be adjusted.
2.Place the thermocouple in an ice-water bath and allow the readings to stabilize.
3.Slowly adjust the OFFSET control until the ther­mometer reads 0°C ( 32°F )
Probe Detector
1.The red LED will be ON when no K-type therm­ocouple probe is inserted into the TEMP input of the meter, and will be OFF after K-type thermo­cople probe is inserted. If the red LED stays ON when thermocouple probe is attached, check the thermocouple probe which might be damaged.
2.When T1 button is pressed, the T1 red LED will be ON, and T2 red LED will be OFF. When T2 button is pressed, the T2 red LED will be ON, and T1 red LED will be OFF. When T1-T2 button is pressed,both T1 and T2 red LEDs will be ON. This is to advise the user to insert the thermoco­uple probes into the correct T1 and/or T2 TEMP input jack where the red LED is ON.
OPERATOR MAINTENANCE
WARNING
TO AVOID POSSIBLE ELECTRICAL SHOCK, DISCONNECT THE THERMOCOOUPLE CONNECTORS FROM THE THERMOMETER BEFORE REMOVING THE COVER.
Battery Replacement
Power is supplied by a standard 9 volt battery (NEDA 1604,IEC 6F22). The “ ” appears on the LCD dis­play when replacement is needed. To replace the battery, remove the screw from the battery cover of the meter. Remove the battery and replace with a new equivalent 9 volt battery.
P/N: 7000-1700
Loading...