Laurel Electronics LTM CTR User Manual

LTM “CTR” SERIES
TRANSMITTERS
PULSE, AC or PROCESS TOTALIZER INPUT,
SERIAL DATA OUTPUT
OWNERS MANUAL
LAUREL Electronics Inc.
3183-G Airway Ave, Costa Mesa, CA, 92626, USA Tel: (714) 434-6131 Fax: (714) 434-3766 Website: www.laurels.com
1. ORDERING GUIDE, LTM “CTR” SERIES TRANSMITTERS
Configure a model number in this format: LTM600VF1, CBL04
LTA Transmitter with serial I/O, Modbus or
Custom ASCII protocol. Includes 4-20 mA, 0-20 mA or 0-10V isolated analog output, isolated 5V, 10V or 15V transducer excita­tion output, and dual solid state relays. Default jumpered for full duplex RS485 and RS232, and 10V excitation.
Main Board
6............................... Standard pulse or AC input
8............................... Extended pulse of AC input
Please see notes for “Extended”
Power
0............................................. 95-240 Vac ±10%
1....................................12-30 Vac or 10-48 Vdc
Relay Output
0................................................................. None
2....................................... Dual solid state relays
Input Type
FR .................. .. Dual Channel Pulse or AC Input
Standard main board
Frequency (2 channels), rate (2 channels), total (2 channels), period, stopwatch, time interval.
Extended main board
Above plus rate and total simultaneously, linear­ized inputs using up to 180 points, arithmetic functions applied to channels A & B (A+B, A-B, A*B, A/B, A/B-1), phase angle, duty cycle, up/ down counting, batch control.
VF1 ............................... 4-20 mA Process Input
VF2 ................................. 0-1 mA Process Input
VF3 ................................. .. 0-10V Process Input
VF1 .............................................. Custom Input
Standard main board
Rate or total from analog process signals. Selectable square root extraction for use with differential pressure flow meters.
Extended main board
Above plus rate and total simultaneously (analog totalizer), custom curve linearization using up to 180 points, batch control, and time based on rate.
QD.................................................. .. Quadrature
Standard main board
Position or length from encoders. Accepts diffe­rential or single-ended inputs: 1x, 2x or 4x, plus zero index.
Extended main board
Above plus bidirectional rate (rate and position or length are not simultaneous).
ACCESSORIES
CBL04 ...... RS232 cable, 3-pin connector on
transmitter end, DB9 connector on computer end.
CBL02 ...... USB to DB9 adapter cable.
-
- 2 -
2. TABLE OF CONTENTS
1. ORDERING GUIDE, “CTR” SERIES TRANSMITTERS ........................................................ 2
2. TABLE OF CONTENTS ....................................................................................................... 3
3. INTRODUCTION, “CTR” SERIES TRANSMITTERS ............................................................ 4
4. RECEIVING & UNPACKING YOUR TRANSMITTER ........................................................... 5
5. SAFETY CONSIDERATIONS .............................................................................................. 5
6. TRANSMITTER FIELD WIRING ......................................................................................... 6
7. PROGRAMMING YOUR TRANSMITTER............................................................................ 8
8. OPENING YOUR TRANSMITTER CASE ............................................................................. 16
9. DUAL CHANNEL PULSE OR AC INPUT SIGNAL CONDITIONER BOARD........................... 17
10. V-TO-F CONVERTER SIGNAL CONDITIONER BOARD ...................................................... 21
11. QUADRATURE SIGNAL CONDITIONER BOARD ................................................................ 22
12. MAIN BOARD JUMPER SETTINGS ................................................................................... 24
13. DUAL RELAY OPERATION ............................................................................................... 25
14. INPUT SIGNAL FILTERING................................................................................................ 26
15. TRANSMITTER CALIBRATION .......................................................................................... 26
16. CUSTOM CURVE LINEARIZATION .................................................................................... 27
17. MODBUS PROTOCOL TRANSMITTER COMMUNICATIONS .............................................. 28
18. CUSTOM ASCII PROTOCOL TRANSMITTER COMMUNICATIONS .................................... 33
19. “CTR” SERIES TRANSMITTER SPECIFICATIONS ............................................................. 38
18. WARRANTY ...................................................................................................................... 40
-
- 3 -
3. INTRODUCTION, “CTR” SERIES TRANSMITTERS
This manual covers “CTR” Series DIN rail transmitters with an isolated serial data output and a pulse, AC or process signal input (converted to a frequency by the unit’s voltage-to-frequency converter). The transmitters duplicate the signal conditioning and signal processing features of their 1/8 DIN panel-mounted counter / timer counterparts for exceptional accuracy at high read rate. A wide range of counter / timer functions are accommodated by three signal conditioners:
Dual-channel pulse input signal conditioner (for frequency, rate, total, arithmetic combina­tions of two channels, stopwatch, timer, phase angle, duty cycle).
Voltage-to-frequency signal conditioner (for rate or total from 4-20 mA, 0-1 mA or 0-10V process signals).
Quadrature signal conditioner (for position or rate from quadrature encoder signals).
Serial communications are standard.
The transmitter serial port
is default jumpered for RS232 or
full-duplex RS485 (same jumper settings). Half-duplex RS485 is also selectable either via internal or external jumpers. Three serial protocols are software selectable: Modbus RTU, Modbus ASCII and Custom ASCII. Modbus allows devices by different manufacturers to be addressed on the same data line. The simpler Custom ASCII protocol is recommended when there are no devices by other manufacturers on the data line.
An isolated 4-20 mA, 0-20 mA or 0-10V output is standard.
This output is generated by an
ultra­linear 16-bit digital to analog converter. It is scaled in software and tracks an internal linearized digital reading.
An isolated transducer excitation output is standard. Three output levels are jumper selectable: 5V at 100 mA, 10V at 120 mA, or 24V at 50 mA.
Dual solid state relays are standard. 130 mA at 140 Vac or 180 Vdc.
Isolation to 250V rms is provided for power, signal input, analog output, relay outputs, and com-
munications. Isolation adds safety and avoids possible ground loops. The transducer excitation output is isolated to 50V from signal ground.
Internal jumpers are used to select the signal range, communication type, and excitation level. The transmitter configuration is specified by the model number on the transmitter label. A user can reconfigure the transmitter by opening the case and moving jumpers.
Transmitter scaling is via serial connection to a PC using MS Windows based Instrument Setup Software, which can be downloaded at no charge. The required transmitter-to-PC interface cable is available for purchase.
-
- 4 -
4. RECEIVING & UNPACKING YOUR TRANSMITTER
Your transmitter was carefully tested and inspected prior to shipment. Should the transmitter be damaged in shipment, notify the freight carrier immediately. In the event the transmitter is not configured as ordered or is inoperable, return it to the place of purchase for repair or replacement. Please include a detailed description of the problem.
5. SAFETY CONSIDERATIONS
Warning: Use of this transmitter in a manner other than specified may impair the protection
of the device and subject the user to a hazard. Visually inspect the unit for signs of damage. If the unit is damaged, do not attempt to operate.
Caution:
This unit may be powered with AC (mains) from 95-240 Vac ±10% with the worldwide voltage
power supply option, or 12-30 Vac or 10-48 Vdc with the low voltage power supply option. Verify that the proper power option is installed for the power to be used. This transmitter has no AC (mains) switch. It will be in operation as soon as power is applied.
The 95-240 Vac ±10% mains connector (P1 Pins 1-3) is colored Green to differentiate it from
other input and output connectors. The 12-30 Vac or 10-48 Vdc mains connector is colored Black.
To avoid dangers of electrocution and/or short circuit, do not attempt to open the case while
the unit is under power. However, signal wiring changes external to the case can be made safely while the unit is under power.
To prevent electrical or fire hazard, do not expose the transmitter to excessive moisture.
Do not operate the transmitter in the presence of flammable gases or fumes. Such an
environment constitutes an explosion hazard.
Secure the transmitter to a 35 mm DIN rail.
Symbols used:
Caution (refer to accompanying documents)
Caution, risk of electric shock.
Earth (ground) terminal.
Both direct and alternating current.
Equipment protected throughout by double insulation or reinforced insulation.
Operating environment:
Transmitter Class II (double insulated) equipment designed for use in Pollution degree 2.
-
- 5 -
See manuals
for different
signal types
Control input 2
Control input 1
GND
Analog out - 1
Analog out + 2
AL2 1
AL2 2 AL1 3 AL1 4
6. TRANSMITTER FIELD WIRING
1 2 3 4 5 6
1 2 3
P6 Signal
input & excitation output
P5 Control inputs 1 & 2
P4 Analog output
P3 Solid state relays
Signal
conditioner
board
P2 Serial
data I/O
P1 Power
input
RS485
6 N/C 5 ARX 4 ATX 3 GND 2 BRX 1 BTX
RS232
TX RX NC GND GND N/C
3 Power GND 2 AC neutral or -DC 1 AC high or +DC
6 TX 5 RX 4 NC 3 GND 2 BRX 1 N/C
Transmitter
RS232 wiring
6 N/C 5 ARX / ATX 4 N/C 3 GND 2 BRX / BTX 1 N/C
Transmitter
RS485 wiring, half duplex
with internal jumpers.
5
9
4
8
3
7
2
6
1
DB9 connector
to PC
(rear view)
ATX / ARX
GND BTX / BRX
Master
-
- 6 -
6 N/C 5 ARX 4 ATX 3 GND 2 BRX 1 BTX
ATX ARX GND BTX BRX
Transmitter Master
RS485 wiring, full duplex-
6 N/C 5 ARX
ATX / ARX
4 ATX 3 GND 2 BRX
GND BTX / BRX
1 BTX
Transmitter
Master
RS485 wiring, half duplex
with external jumpers.
P6 - SIGNAL INPUT DETAIL
digital logic, magnetic pickups, or AC inputs to 250V .
PROCESS / TOTALIZER SIGNAL INPUT
DC & Externally Powered Process
Excitation return 1
+ Excitation 2
- Signal input 3
+ Signal input 4
2-Wire Process Transmitter
Excitation return 1
+ Excitation 2
- Signal input 3
+ Signal input 4
QUADRATURE SIGNAL INPUTS
Differential or Complementary Inputs
A Input 1 A Input 2 B Input 3
B Input 4
Z Input (+Excitation) 5
Z Input (Exc. return) 6
Single Ended Inputs
A Input 1
+Excitation 2
B Input 3
Excitation return 4
Z Input 5
Signal Ground 6
-DC +DC
+
-
A Channel In
B Channel In
Zero Index In or Excitation Output
A Channel In + Power Output
B Channel In
- Power Output
Zero Index In Sensor common
DUAL CHANNEL PULSE SIGNAL INPUT..
Single Powered Sensor Input
Excitation return 1
+ Excitation 2
+B Signal Input 3
Signal Ground 4
+A Signal Input 5
Signal Ground 6
Power to sensor
+
Sensor 1
-
Two Powered Sensor Inputs
Excitation return 1
+ Excitation 2
+B Signal Input 3
Signal Ground 4
+A Signal Input 5
Signal Ground 6
Power to sensor
+
Sensor 2
-
+
Sensor 1
-
Active and Passive Inputs
Excitation return 1
+ Excitation 2
+B Signal Input 3
Signal Ground 4
+A Signal Input 5
Signal Ground 6
Inputs can be proximity switches, contact closures,
Warning: Dual-channel signal grounds 4 & 6 are connected internally.
MOUNTING FOR COOLING
P4 - ANALOG OUTPUT DETAIL
Driving a Load with 4-20 mA
Analog return 1
4-20 mA output 2
500 Ohms max
Driving a Load with 0-10V
Analog GND 1
0-10V output 2
5 kOhms min
-
- 7 -
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
Mount transmitters with ventilation holes at top and bottom. Leave minimum of 6 mm (1/4") between transmitters, or force air with a fan.
7. PROGRAMMING YOUR TRANSMITTER
OVERVIEW
Our transmitters are easily programmed using a PC with an RS232 port and Instrument Setup (IS) software, which provides a graphical user interface. The software allows uploading, editing,
downloading and saving of setup data, execution of commands under computer control, listing, plotting and graphing of data, and computer prompted calibration.
USING IS SOFTWARE
Use a 3-wire RS232 cable (P/N CBL04) to connect your transmitter to the COM port of your PC. Download the file
Install Instrument Setup Software Start => Programs => IS2 => Instrument Setup
created. Following a brief splash screen, the
ISx_x_x.exe
from our website and double-click on the file name. Click on
and follow the prompts. To launch IS software, press on
or on the desktop icon that you may have
Communications Setup
screen will appear.
RS232 cable with rear view of DB9 connector to PC
-
- 8 -
In the Communications Setup screen, select the factory default setting. Select
RS-232 (USB)
button. This will take you to the
Transmitter LTA, LTM, LTS
Establish Communications
Custom ASCII
as the Device Type. Then click on the
as the protocol, as this is the
screen.
-
- 9 -
In the
You will be able to change your protocol and baud rate later under the tab. Click on on the
From the
from your counter transmitter. Click on easily edit the setup file using pull-down menus and other selection tools. You can download (or put) your edited file into the transmitter by clicking on your setup file to disk by clicking on disk by click on
The best way to learn IS software is to experiment with it.
For detailed help on any data entry field under any tab, select that field and press on the F1
key.
Establish Communications
Establish
Main Menu
Main Menu
File
, and the two fields at the bottom of the screen should turn green. Click
button.
, click on
=>
Open
Counter
.
screen, select your Com Port and
=>
File
Get Setup
View
=>
=>
Save Setup
to retrieve (or get) the existing setup data
Setup
to bring up screens which allow you to
Counter
and retrieve a previously saved file from
=>
9600
as the Baud Rate.
Communication
Put Setup
. You can save
setup
-
- 10 -
Click on
transmitter, then on screen to set up
Counter
=>
Signal Input, Display
Get Setup
View
=>
to retrieve the current setup information from your counter
Setup
, which will take you to the
, and
Control Inputs
.
Input+Display
tab. Use this
Click on the
methods: 1) direct entry of Scale and Offset, and 2) the Coordinates of 2 Points method, where (Low In, Low Read) and (High In, High Read) data points are entered numerically.
Reminder: for detailed help on any data entry field under any tab, select that field and press on the F1 key.
Scaling
tab to scale your transmitter. You will be given the choice of two scaling
-
- 11 -
Click on the
Filter
tab to set to set up filtering for your readings.
Click on the
Relay Alarms
tab to set up your transmitter’s two solid state relays, which are
standard.
Reminder: for detailed help on any data entry field under any tab, select that field and press on the F1 key.
-
- 12 -
Loading...
+ 28 hidden pages