Lauda Ecoline RE 220, Ecoline RE 206, Ecoline RE 207, Ecoline RE 212, Ecoline RE 306 Operating Instructions Manual

...
LAUDA DR. R. WOBSER
GMBH & CO. KG
P.O. Box 1251
97912 Lauda-Königshofen
Germany From Series X01 Software version 3.1 3/99 YACE0060
Phone: (+49) (0) 9343/ 503-0
Fax: (+49) (0) 9343/ 503-222
e-mail info@lauda.de
Internet http://www.lauda.de
Operating Instructions
Ecoline
Low-temperature thermostats
RE 204, RE 206, RE 207, RE 212, RE 220
RE 306, RE 307, RE 312, RE 320
I
Safety notes
Before operating the equipment please read carefully all the instructions and safety notes. If you have any questions please phone us!
Follow the instructions on setting up, operation etc. This is the only way to avoid
incorrect operation of the equipment and to ensure full warranty protection.
Transport the equipment with care!
Equipment and its internal parts can be damaged:
by dropping
by shock.
Equipment should only be operated by technically qualified personnel!
Never operate the equipment without the bath liquid!
Do not start up the equipment if
it is damaged or leaking
the supply cable is damaged.
Switch off the equipment and pull out the mains plug for
servicing or repair
the supply cable is damaged!
Drain the bath before moving the equipment!
Have the equipment serviced or repaired by properly qualified personnel only!
The Operating Instructions include additional safety notes which are identified by a triangle with an exclamation mark. Carefully read the instructions and follow them accurately! Disregarding the instructions may have serious consequences, such as damage to the equipment, damage to property or injury to personnel
We reserve the right to make technical alterations!
II
1 BRIEF OPERATING INSTRUCTIONS...........................................................................................1
2 CONTROL AND FUNCTIONAL ELEMENTS.................................................................................3
3 UNIT DESCRIPTION .....................................................................................................................4
3.1 Unit types ............................................................................................................................... 4
3.2 Pumps.....................................................................................................................................4
3.3 Temperature indication, control, and safety circuit ............................................................4
3.4 Programmer (Types RE 3xx only).........................................................................................5
3.5 Materials.................................................................................................................................5
3.6 Refrigeration system .............................................................................................................5
3.7 Serial Interfaces RS 232, RS 485...........................................................................................6
3.7.1 Specification and interface test.........................................................................................................6
3.7.2 General information............................................................................................................................7
3.7.3 Output commands..............................................................................................................................8
3.7.4 Data request from the thermostat.....................................................................................................9
3.7.5 Error messages.................................................................................................................................10
4 UNPACKING ............................................................................................................................... 11
5 PREPARATIONS.........................................................................................................................12
5.1 Assembly and setting up.....................................................................................................12
5.2 Filling and emptying............................................................................................................13
5.3 Bath liquids and hose connections....................................................................................15
5.4 Connection of external circuits...........................................................................................17
6 STARTING UP.............................................................................................................................19
6.1 Connection to the supply....................................................................................................19
6.2 Switching on ........................................................................................................................19
6.3 Setpoint selection
(level 0)
.................................................................................................. 20
6.4 Menu functions ....................................................................................................................21
6.4.1 Refrigeration system
(level 1)
..........................................................................................................21
6.4.2 Pump output......................................................................................................................................22
6.4.3 Setting the setpoint resolution........................................................................................................23
6.4.4 Programmer (Types RE 3xx only)...................................................................................................24
6.4.4.1 Programme example....................................................................................................................24
III
6.4.4.2 Indicating/ altering of programme segments................................................................................25
6.4.4.3 Number of programme running....................................................................................................27
6.4.4.4 Starting of the programmer..........................................................................................................28
6.4.4.5 Inserting/ deleting of programme segments ................................................................................29
6.4.4.6 Holding/continuing the programme..............................................................................................31
6.4.4.7 Terminating the programme ........................................................................................................32
6.4.5 Parameters........................................................................................................................................33
6.4.5.1 Setting the proportional band of the PID-controller......................................................................34
6.4.5.2 Setting the reset time of the PID-controller..................................................................................35
6.4.5.3 Selection of the interface .............................................................................................................35
6.4.5.4 Setting the Baud rate/ transfer rate (serial interface)...................................................................36
6.4.5.5 User calibration............................................................................................................................38
6.5 Warning and safety functions.............................................................................................39
6.5.1 Overtemperature protection and testing........................................................................................39
6.5.2 Low-level protection and testing ....................................................................................................41
6.5.3 Pump motor monitoring...................................................................................................................42
6.5.4 Connection floating contact ”Combination fault” 31 N................................................................42
6.5.5 Other error messages ......................................................................................................................43
7 SAFETY NOTES..........................................................................................................................44
7.1 General safety notes............................................................................................................44
7.2 Other safety notes................................................................................................................44
8 MAINTENANCE...........................................................................................................................46
8.1 Cleaning................................................................................................................................46
8.2 Maintenance and repair.......................................................................................................46
8.2.1 Maintenance of the refrigeration unit.............................................................................................47
8.2.2 Note on repair and disposal ............................................................................................................47
8.3 Ordering spares ...................................................................................................................47
9 TECHNICAL DATA (TO DIN 58966)............................................................................................48
10 ACCESSORIES ......................................................................................................................... 50
11 CIRCUIT DIAGRAM...................................................................................................................51
12 PIPE PLAN ................................................................................................................................55
IV
Explanation of signs
Danger: This sign is used where there may be injury to
personnel if a recommendation is not followed accurately or is disregarded.
Note Here special attention is drawn to some aspect. May
include reference to danger.
Reference: Refers to other information in different Sections.
Brief operating instructions
10.08.99 YACE0060
1
1 Brief operating instructions
This brief instruction shall give you the possibility to operate the unit quickly. For safe operation of the unit it is absolutely necessary to read carefully all the instructions and safety notes!
1. Assemble unit and add items as appropriate (➾ Section 5). Take care of the hose tubing connections (➾ Section 5.1. and 5.4.).
2. Fill the unit with corresponding liquid. (➾ Section 5.3.). The units are designed for operation with non-flammable and flammable liquids to EN 61010-2-010. ! Take care of the level of the bath liquid! (➾ Section 5.2.)
3. Connect the unit only to a socket with a protective earth (PE) connection. Compare the information on the rating label with the supply details.
4.
Using a screwdriver, set the overtemperature cut-out point to a value clearly above ambient temperature (➾ Section 6.5.1.).
5. Switch on at the mains switch
6. Setting of the functions
Brief operating instructions
2
Control and functional elements
10.08.99 YACE0060
3
2 Control and functional elements
Cooling
LED green, Supply ON
Menu functions
Mains switch
Setting of the overtemperature switch-off point
Pump connections
return
Pump connection:
outflow
Bath bridge
Bath cover
Unit description
4
3 Unit description
3.1 Unit types
The type designation of the Ecoline low-temperature thermostats consists of the letter R (identification as low-temperature unit), the control unit E 200 or E 300, and the type of bath and refrigeration system
Example: Control unit E 200 and bath RE 004 produces Thermostat Type RE 204.
Type RE 220 is supplied without bath cover. A bath cover is available as accessory. (➾ Section 10. Accessories).
3.2 Pumps
All units are equipped with a pressure pump with vario drive. The pump has an outflow with a rotatable discharge elbow. An additional outflow provides circulation inside the bath. By turning the setting knob it is possible to choose between both outlets or to divide the flows.
The pump ca n be used up to viscosities of 150 mm²/s however, to get an optimum accuracy of control a viscosity of 30 mm²/s is recommended.
One of five pump output steps can be selected using the operating menu. On small low-temperature thermostats (e.g. RE 204 or RE 306) and with operation as bath thermostat it is advisable to use output step 2. The advantage is a low heat generation while having a uniform circulation, that means, that the thermostat can work without cooling down to just above ambient temperature.
When operating as circulation thermostat with an external circuit it is preferable to use a larger flow rate in order to ensure a small temperature difference, especially at higher temperatures and in conjunction with oil as the bath liquid.
The pump connections of Types RE 3xx are fitted with M 16x1 nipples.
The pump connection outflow can be closed off without causing any damage to the pump.
Pump characteristics
(➾ Section 9. Technical data)
3.3 Temperature indication, control, and safety circuit
The unit is equipped with a 2-line LCD-Display with additional symbols for indicating bath temperature and settings as well as operating states. On RE 3xx-Types the display has back-lightning. The input of the setpoint and other additional adjustments are done using either two or three keys.
Remote operation is possible via an electrically isolated RS 232 interface
Unit description
10.08.99 YACE0060
5
A Pt 100 temperature probe is used for measuring the actual temperature and for control. A second Pt 100 serves as temperature probe for the safety circuit (overtemperature protection) which is independent of the control function
A low-level protection switches off the heating on both poles in order to prevent dry operation of the heater. The pump is switched off electronically. The overtemperature switch-off point is adjusted with a tool on a potentiometer and is always limited to 5 °C above the operating temperature range. A floating contact "Combination fault" is available (➾ Section 6.5.4.).
All settings and fault messages are stored in the memory in case of a supply failure or when the mains switch is set to OFF.
The tubular heater is controlled from a modified PID controller through a triac circuit specially designed to be unaffected by supply variations and interference.
3.4 Programmer (Types RE 3xx only)
Types RE 3xx incorporate a programmer which can be used to run temperature programs with up to 20 temperature-time-segments (➾ Section 6.4.4).
3.5 Materials
All parts which come into contact with the bath liquid are made from high-grade materials appropriate to the operating temperature. These are rust-free stainless steel, PPS plastics and fluoride rubber.
3.6 Refrigeration system
The refrigeration system consists essentially of a hermetically sealed compressor. Heat of condensation and motor heat are dissipated by a fan-cooled finned condenser. Fresh air is drawn in at the front of the unit, warmed air is discharged at the back and to the sides. The ventilation openings must not be restricted in order to ensure proper air circulation.
At working temperatures below approx. 30 °C the refrigeration system operates continuously to remove a certain amount of heat, with the heater acting in opposition to provide automatically controlled heating power.
The compressors are fitted with a temperature monitor which responds both to the compressor temperature and to the motor current. In addition the cooling system is protected against excessive pressure by a pressure monitor. The refrigeration unit is switched on either automatically or manually through the operating menu .(➾ Section 6.4.1.).
Types RE 207, RE 212, RE 220 as well as RE 307, RE 312 and RE 320 are additionally provided with a cooling capacity adjustment which saves energy in many operating situations, reduces the load on the refrigeration system and produces a better temperature control. This feature, too, can be operated either automatically or manually.
Cooling curves
(➾ Section 9. Technical data)
Unit description
6
3.7 Serial Interfaces RS 232, RS 485
3.7.1 Specification and interface test
Computer
Thermostat
Data
9-pin sub-D
socket
25-pin sub-D
socket
9-pin sub-D socket
Data
R x D 2 2 3 3 2 2 T x D
T x D 3 3 2 2 3 3 R x D DTR 4 20 4 DSR Signal Ground
5 5 7 7 5 5 Signal Ground
DSR 6 6 6 DTR RTS 7 4 7 7 CTS CTS 8 5 8 8 RTS
with Hardware Handshake: When connecting the thermostat to the computer please
use a 1:1 cable and not a zero-modem-cable!
without Hardware Handshake: The computer needs an operating mode: ”without
hardware handshake”. In the plug of the thermostat a bridge has to be inserted between Pin 7 and 8.
Thermostat
RS 485 interface
9-pin sub-D-socket
pin Data
1 Data A
5 SG (Signal Ground)
optional
6 Data B
Use shielded connecting cables.
Connect the shielding to the plug case.
Cover unused connectors with protection caps!
The lines are electrically isolated from the remaining electronics.
Unoccupied pins must not be connected!
Unit description
10.08.99 YACE0060
7
The RS 232 Interface can easily be tested with the PC connected, using the MS-Windows operating system. On Windows 3.11 with the programme "Terminal" and on Windows 95 with the programme "Hyper Terminal Terminal”.
The RS 485 bus absolutely needs a bus-termination in form of a terminating network, which ensures a defined unattended time in the high resistance phases. The bus-terminal is defined as follows
Generally this network is integrated on the plug in card of the computer (RS 485).
3.7.2 General information
The interfaces operate with 1 stop bit, no parity bit and 8 data bits
Transfer rate (selectable): 2400, 4800, 9600 (factory setting) or 19200 baud.
RS 232 Interface:
This interface can also be handled by Hardware – Handshake (RTS/CTS) (➾ Section
3.7.1.)
Values can be transferred directly from the computer to the thermostat via the commands: OUT, STOP, START (➾ Section 3.7.3.). After a correct transfer of these commands the thermostat always responds with the message "OK" followed by "CR" and "LF". (Must be read out by the computer like any other response.)
RS 485 Interface:
The unit address has to be added in front of the RS 232 interface commands (possible are unit addresses 000...127 ! the address always has to have 3 digits).
Example:
Transfer the setpoint temperature 30,5°C from the thermostat to the unit address 15 ! ”A015_OUT_SP_00_30.5°C”CR (➾ Section 3.7.3.). The thermostat always responds with the unit address first ! ”A015_OK” CR LF
Values are transferred from the thermostat to the computer using an IN-command. (➾ Section 3.7.4.)
Unit description
8
3.7.3 Output commands
Command
Explanation
OUT_SP_00_XXX.XX Setpoint transfer with up to 3 places before the decimal point and
up to 2 places behind
OUT_SP_01_XXX.XX Pump output step 1, 2, 3, 4 or 5, 0 = STOP
OUT_SP_02_XXX.XX Operating mode of the refrigeration system “0“ = OFF,
"1" = ON, “2“ = ON 50 % (reduced capacity RE 207, 212, 220, 307, 312, 320 only) , "3" = Automatic operation
OUT_PAR_00_XXX.XX Setting of the control parameter Xp for controller (0,5...9,9°C)
OUT_PAR_01_XXX.XX Setting of the control parameter Tn (5...60 s)
START switches the unit on (from stand-by)
STOP switches the unit to stand-by (pump, heating, refrigeration system
off)
OUT_MODE_00_X
(A015_OUT_MODE_00_X --> RS 485)
Operating mode: 0 = RS 232 + keyboard, 1 = only RS 232
(Operating mode : 0 = RS 485 + keyboard/ 1 = only RS 485)
RMP_START Starting the programmer
RMP_PAUSE Stopping the programmer
RMP_CONT Restart the programmer after PAUSE
RMP_STOP Terminating the programme
RMP_RESET Deleting the programme (all segments)
RMP_OUT_00_XXX.XX_XXX.X X
Programmer segments (max. 20) (temperature, time)
RMP_OUT_02_XXX.XX Number of programme running: 0...255,
0= infinite
The blank character " " is also permissible instead of "_".
Thermostat response "OK" or in case of a fault "ERR_X" (RS 485 interface: ”A015_OK” or in case of a fault ”A015_ERR_X”)
EXAMPLE: set setpoint to 20.00°C: " OUT_SP_00_20"CR LF (RS 495 interface: ”A015_OUT_SP_00_20”CRLF)
Unit description
10.08.99 YACE0060
9
Permitted data formats
-XXX.XX -XXX.X -XXX. -XXX XXX.XX XXX.X XXX. XXX
-XX.XX -XX.X -XX. -XX XX.XX XX.X XX. XX
-X.XX -X.X -X. -X X.XX X.X X. X
-.XX -.X .XX .X
3.7.4 Data request from the thermostat
The response from the thermostat is always in the fixed format "XXX.XX" for negative values "-XXX.XX" or " ERR_X" (RS 485 interface: ”A015_XXX.XX”, for negative values ”A015_-XXX.XX” or ”A015_ERR_X”).
Input command Explanation
IN_PV_00 Request to indicate the bath temperature
IN_SP_00 Request to indicate the active setpoint
IN_SP_01 Request to indicate the pump output step
IN_SP_02 Request to indicate the operating mode of the refrigeration
system “0“ = OFF, "1" = ON, “2“ = ON 50 % (reduced capacity RE 207, 212, 220, 307, 312, 320 only) , "3" = Automatic operation
IN_SP_03 Request to indicate the current overtemperature switch-off
point
IN_PAR_00 Request to indicate the current value of Xp
IN_PAR_01 Request to indicate the current value of Tn
TYPE Request to indicate the unit type
VERSION Request to indicate the software version number
STATUS Request to indicate the unit status
0 = OK, - 1 = fault
STAT
Input for fault diagnosis answer XXXXX ! X = 0 no fault, X = 1 fault char. 1 from right = internal fault microcontroller 2 char. 2 from right = internal fault microcontroller 1 char. 3 from right = pump blocked char. 4 from right = low level char. 5 from right = overtemperature
Unit description
10
IN_ERR_00 fault diagnosis microcontroller 1
IN_ERR_01 fault diagnosis microcontroller 2
IN_MODE_00 (A015_IN_MODE_00
-->RS485)
0 = RS 232 + keyboard/ 1 = only RS 232 interface (0 = RS 485 + keyboard/1 = only RS 485 interface)
RMP_IN_00_XX
(1...20)
Request to indicate the programme segments (answer e.g 030.00_010.00 = 30.00 °C, 10 min)
RMP_IN_01 Request to indicate the current segment number, if
programmer is activated
RMP_IN_02 Request to indicate the number of programme running (pre-
set)
RMP_IN_03 Request to indicate the current programme running
➀ ➾Section 6.5.5. Other error messages
3.7.5 Error messages
Message Explanation
ERR_2 incorrect input (e.g. buffer overflow) ERR_3 incorrect command ERR_5 syntax error on value ERR_6 illegal value ERR_28 receive – frame – error (e.g. stop bit missing) ERR_29 Function blocked (at this time)! wait a few ms, then try again ERR_30 Programmer, all segments are occupied
RS 485 interface: Message ”A015_ERR_2”.
Unpacking
10.08.99 YACE0060
11
4 Unpacking
After the unit and accessories have been unpacked they have to be examined for possible transport damage. If there is any damage visible on the unit, the forwarding agent or the post office has to be notified so that the shipment can be examined.
Standard accessories:
Bath cover on all low-temperature thermostats except RE 220
13 mm dia. nipple union nuts (M16x1)
on all types RE 3xx
Closing plugs on all low-temperature thermostats
Warning label
on all low-temperature thermostats
Operating Instructions on all low-temperature thermostats
Preparations
12
5 Preparations
5.1 Assembly and setting up
Place the unit on a flat surface
After transport and before starting up, store it standing in upright position for two hours if possible
Do not cover the ventilation openings at the back of the unit and its lower part. Keep clear distance of at least 40 cm.
Adjust the pump setting knob so that when using the unit as bath thermostat (no external circulation) the flow emerges at the opening for the bath circulation, or link together the pump connections. (See ill. 3).
Operation with external consumer
(circulation
thermostat) (➾Section 5.4.)
Adjustments of the pump outflows
Ill 1
Pump setting knob turned
anticlockwise
Ill 1
Pump setting knob medium
position
Ill 3.
Pump setting knob turned
clockwise
When operating as bath thermostat without external consumer the pump connection outflow has to be closed (use closing plugs) or linked to the return.
At bath temperature above 70°C the label
supplied must be affixed on
the bath in a clearly visible position!
Preparations
10.08.99 YACE0060
13
The unit can be operated safely up to an ambient temperature of 40°C
Depending on the loading of the refrigeration system it may switch off temporarily especially at ambient temperatures above 35°C. Elevated ambient temperatures also result in reduced cooling capacity.
When starting up the refrigeration system after it has been cut of operation for longer time it may take up to 30 min, depending on ambient temperature and unit type, before the nominal cooling capacity is reached.
5.2 Filling and emptying
Filling
Close the drain cock.
Fill baths up to a maximum level of 20 mm below the bath bridge.
Optimum operation at 20-40 mm below the bath bridge.
Operation is possible down to 60 mm below the bath bridge.
The low-level cut-out operates approx. 90 mm below the bath bridge!
When using thermal oils it is necessary to allow for an expansion of approx. 8 %/100 °C.
When operating with an external consumer the total expansion takes place in the bath!
Preparations
14
Emptying
Switch off the thermostat, pull out the mains plug!
b) Drain the bath liquid through the drain cock
!
using tubing.
The units are designed for operation with non-flammable and flammable liquids to EN 61010-2-010! Flammable liquids can be operated up to no more than 25°C below the firepoint (➾ Section 5.3.).
Observe the appropriate regulation when disposing of used thermostating liquid.
When connecting an external consumer take care that the level of the bath liquid does not decrease too much ! fill in bath liquid if necessary.
Do not drain the thermostating liquid when it is hot or very cold (below 0°C)!
Preparations
12.07.00 YACE0060
15
5.3 Bath liquids and hose connections
Bath liquids
LAUDA
Designation
Working
tem-
perature
range
Chemical
Designation
Visco-
sity
(kin)
Viscosity
(kin) at
Temperature
Fire-
point
Ref.No.
Quantity
Former
designation
from °C
to °C
at 20°C mm²/s mm²/s 5 l 10 l 20 l
water +5...+90
deionised
water
-- -- --
Kryo 30
G 100
-30...+90
Mono-
ethylene-
glycol/water
4 50 at -25°C -- LZB 109 LZB 209 LZB 309
Kryo 50 SK Super
Frigor
-50...+95
Silicone oil
6 35 at -50°C > 139 LZB 103 LZB 203 LZB 303
Kryo 20 160 MS -20...+180
Silicone oil
11 28 at -20°C > 230 LZB 116 LZB 216 LZB 316
Ultra 350 330 SCB +30...+200
synthetic
thermal oil
47 28 at +30°C > 240 LZB 107 LZB 207 LZB 307
Therm 230 RDS 50 +60...+230
Silicone oil
44 28 ... +60°C > 362 LZB 117 LZB 217 LZB 317
At higher temperatures ! Evaporation losses ! Use bath covers (➾ Section 10. Accessories).
Distilled water or fully deionised water must only be used with the addition of 0,1g sodium carbonate/l water, otherwise
!
danger of corrosion!
Water content falls after prolonged operation at higher temperatures
!
mixture becomes flammable (flash point 128 °C).
!
Check the mixture ratio with a densiometer.
When selecting bath liquids it should be noted that performance must be expected to worsen
at the lower limit of the operating temperature range due to increasing viscosity. The full operating range should only be utilised if really necessary.
The operating ranges of the bath liquids and tubing represent general data which may be
limited by the operating temperature range of the unit.
Silicone oil causes pronounced swelling of Silicone rubber ! never use Silicone oil with Silicone tubing!
DIN Safety data sheets are available on request
Preparations
16
DIN Safety data sheets are available on request!
Hose connections
a) Elastomer tubing
Tubing type Int. dia.
Ø mm
Temperature
range °C
Application Ref. No.
Perbunan tubing,
uninsulated
9 0 to 120 for all bath
liquids
RKJ 011
Perbunan tubing,
uninsulated
12 -0 to 120 for all bath
liquids
RKJ 012
Perbunan tubing
insulated
12
ext. dia.
35mm approx.
-60 to 120 for all bath liquids
LZS 008
Silicone tubing,
uninsulated
11 -30 to 100 water,
water/glycol
mixture
RKJ 059
Silicone tubing
insulated
11
ext. dia.
35mm approx.
-60 to 100 water,
water/glycol
mixture
LZS 007
Viton 11 -60 to 200 for all bath
liquids
RKJ 091
Silicone oil causes pronounced swelling of Silicone rubber ! never use Silicone oil with Silicone tubing!
Protect tubing with hose clips against slipping off.
b) Metal hoses for Types RE 3xx, rust free stainless steel, with M 16 x 1 union nut 10 mm
int. Dia
Type
Length
cm
Notes Ref. No.
MC 50 50 With single insulation
LZM 040
MC 100 100 "
LZM 041
MC 150 150 "
LZM 042
MC 200 200 "
LZM 043
Pump connection link 20 "
LZM 044
MK 50 50 with foam insulation for low
temperatures
LZM 052
MK 100 100 "
LZM 053
MK 150 150 "
LZM 054
MK 200 200 "
LZM 055
Pump connection link 20 "
LZM 045
Preparations
10.08.99 YACE0060
17
5.4 Connection of external circuits
Operation as circulation thermostat
Connect 11-12 mm int. dia. tubing (for Types RE 3xx use metal hoses) (➾ Section 5.3.) to the pump connector
Pump connection outflow always in front, return connection always at the back.
If the cross-section of the tubing is too small ! temperature drop between bath and external system due to low flow rate. Increase the bath temperature appropriately.
Always ensure the maximum possible flow cross-section in the external circuit!
On Types RE 3xx, when tightening the union nut hold on to the pump nipple with a SW 14 spanner (see ill.)!
When the external consumer is placed at a higher level than the thermostat the pump is stopped and air penetrates into the thermostating circuit the external liquid may drain down into the bath even with a closed system ! danger of flooding the thermostat
Protect tubing with hose clips against slipping off!
When no external circuit is connected to the thermostat, the pump connection outflow must be closed (use closing plugs) or linked to the return!
Preparations
18
Using the setting knob at the pump outflows, divide up the pump flow in accordance to the thermostating task.(➾ Section 5.1)
Position
➀➀➀➀
! maximum flow in the external circuit
(setting knob turned anticlockwise).
Position
➁➁➁➁
!
flow passes through pump outflow and outlet for bath circulation (setting knob in medium position).
Position
➂➂➂➂
! external circuit is closed and the outlet for bath circulation fully open (setting knob turned clockwise).
Operate the setting knob only when the bath contents are near the ambient temperature.
When no tubing is connected, close the pump connection with closing plugs even in position ➂.
Starting up
10.08.99 YACE0060
19
6 Starting up
6.1 Connection to the supply
Compare the supply voltage against the data on the rating label.
Connect the unit only to a grounded mains power socket (PE).
No warranty when the thermostat is connected to a wrong supply!
Without external circuit ensure that the pump pressure outflow is closed or linked to the pump return.
Ensure that the unit is filled in accordance with Section 5.2.!
6.2 Switching on
Using a screwdriver. Set the overtemperature switch-off point to a value clearly above ambient temperature.
Switch on at the mains switch. The green LED for "Supply ON" lights up.
"
""
"
0,25s
A tone sounds for approx. 0,25 s.
The unit self-test starts up. All display segments and symbols light up for approx. 1 s. Then the software version is indicated for approx. 1 s.
Display shows the actual bath temperature (above) (resolution 0,05 °) and of the setpoint. The pump starts up. The values which were active before switching off are used.
If necessary add more bath liquid to replace the amount pumped out to the external circuit.
If the pump does not purge the system immediately. The unit may switch off again although it is filled sufficiently (only when starting up for the first time).
Starting u
p
20
"
""
"

A double signal tone sounds.
The display for LEVEL (low-level) appears.
The fault triangle is flashing
Press the key. If necessary repeat several times.
Also press the key if the unit had switched off under a fault condition.
6.3 Setpoint selection
(level 0)
or
Input the setpoint with one of the keys.
Speeding the setting process by:
a) continuous pressing the keys or
b) pressing one key (holding it down) and shortly pressing
the other key.
Briefly releasing (1 s) the key (s) and again pressing one of the keys moves the cursor one place to the right.
Display flashes 4 s ! the new value is accepted automatically, or
Value is entered immediately with this key.
For safety reasons the setpoint can only be adjusted up to 2 °C above the upper limit of the operating temperature range of the particular unit type
Starting up
10.08.99 YACE0060
21
6.4 Menu functions
Switching from setpoint selection (level 0) to level 1 using the key
or
Within one level it is possible to scroll using the keys.
In principle
, after each setting has been made it is entered
automatically after approx. 4 s or
Settings are entered immediately on operating this key
6.4.1 Refrigeration system
(level 1)
The display shows the current bath temperature, COOL and the set operating mode. To alter the setting, press the key on the left.
The display flashes (approx. 4 s). During this time start to set the operating mode with one of the keys.
or
0 = Refrigeration system OFF 1 = Refrigeration system ON 100 % capacity 2 = Refrigeration ON, 50% capacity (only RE 207, 212, 220, 307, 312, 320) A = automatic operation (➾ Section 3.6.)
Forward with key to "pump output" or
with key back to level 0 (setpoint input).
Starting u
p
22
The refrigeration system can normally work with automatic operation = A. The refrigeration system switches on or off depending on the temperature and operation status.
In special cases the cooling machine can be switched off ! ”0” or adjusted at permanent running ! ”I”.
6.4.2 Pump output
and
1x
For setting the pump output from level 0 press the key combination shown on the left, or
move forward from COOL function with this key
The display shows the current bath temperature, PUMP and the current output step. To alter the setting press the key shown on the left.
Display flashes (approx. 4 s).
or
During this time start to set the required step with one of the keys. 0 = pump stopped, heating off 1 = low pump output 2 , 3 , 4 = medium pump output 5 = maximum pump output
The pump responds immediately (can be heard). (Setting is entered after approx. 4 s ➾ Section 6.4.)
Move forward with key to "Selecting the setpoint resolution" or
back with the key to "COOL - function" (RE 3xx) or to "Setpoint selection" (RE 2xx).
Starting up
10.08.99 YACE0060
23
6.4.3 Setting the setpoint resolution
and
2x
To set the setpoint resolution from level 0 press the key combination on the left
move forward with the key from the PUMP-function.
The current bath temperature, RES and the current setpoint resolution are indicated.
To alter the setting, press the key on the left.
Display flashes (approx. 4 s).
or
During this time, start to set the required resolution with one of the two keys. 0,1 = 0,1 °C setpoint resolution, normal setting 0,01 = 0,01 °C setpoint resolution. During setpoint input only S is indicated instead of SET.
Forward with the key to "parameter level" (RE 2xx only) or to "programmer" (RE 3xx only) or
with key back to "PUMP“.
Starting u
p
24
6.4.4 Programmer (Types RE 3xx only)
and
3x
To view or to set the programmer, starting from level 0 (setpoint input) press the key combination on the left, or
from RES function scroll with this key.
The display shows PGM (programmer). Data for up to 20 programme segments can be input there.
6.4.4.1 Programme example
Segment
1 2 3 4 5 6 7
Temperature 30,0 50,0 70,0 70,0 60,0 30,0
Time 20 20 40 10 30 0
Segment
1 2 3 4 5 6 7
Temperature 30,0 50,0 50 70,0 70,0 60,0 30,0
Time 20 20 20 20 10 30 0
➀ A new segment has been inserted after segment No. 2 (➾ Section 6.4.4.5.)
➁ The time at segment No. 3 has been altered (➾ Section 6.4.4.2.)
Starting up
10.08.99 YACE0060
25
6.4.4.2 Indicating/ altering of programme segments
When PGM appears on the display (therefore proceed as described in 6.4.4.) press the key on the left.
The display shows RUN. Here the programmer can be started (➾ Section 6.4.4.4.)
Forward with key until EDIT appears..
Here the programme segments can e.g. be indicated and altered. To do so press the key.
The display indicates 3 variables: segment number (a), segment end temperature (b) and segment time (c).
EXAMPLE
:
segment number 1, b = 30,0 °C, c = 20 min. The bath liquid has to be heated up or cooled down to 30 °C within 20 min.
When having deleted the programme example the variables on the display show 0. Before altering the segments it is necessary to insert new segments (➾ Section 6.4.5.5.)
or
These keys can be used to scroll through the different variables.
Sequence with the key: : 1 (a) ! 30,0 °C (b)
!
20 (c) ! 50,0 ° (2. Segment), ! 20 (2. Segment).
When the segment number is flashing, pressing the key leads directly to the menu for inserting or deleting segments (➾ Section 6.4.4.5.). Otherwise scroll e.g. from segment time back to segment temperature and segment number.
Starting u
p
26
The variable currently activated flashes quickly (here the segment number).
To alter the required variable start with this key
or
If e.g. the segment number is flashing, all segments can be indicated in sequence by pressing one of the two keys, or
if segment temperature or segment time are flashing, the required temperature or time can be input by pressing the key and then using the other two keys:
Segment temperature.: 2 °C max. above the upper limit of the operating temperature range of the particular thermostat type.
Segment time: 0...255 min.
After having changed the segments move forward with key to END
1. Forward with the key to EDIT..
with the key to END and then
back to level 0 with the key, or
2. with the key back to LOOP (➾ Section 6.4.4.3.) resp. to RUN (➾ Section 6.4.4.4.)
Starting up
10.08.99 YACE0060
27
While the programmer is in operation segments can be altered (including the current segment) and new segments can be inserted. All segments can also be deleted at each time (except the current segment) (➾ Section 6.4.4.5.) BUT: If the new segment time is shorter than the segment time which has already elapsed, the next segment is activated. (➾ Section 6.4.4.1. graph b.)
If a segment time longer than 255 min is required, this time must be distributed over several consecutive segments.
6.4.4.3 Number of programme running
From level 0 proceed as described under 6.4.4. When PGM appears on the display, press the key on the left
The display first shows RUN.
− −
Forward with key until LOOP appears, or
from EDIT with this key to LOOP.
Here the number of programme running can be input. Therefore press the key. The display is flashing for approx. 4 s.
or
During this time start to set the required number of running with one of the two keys. Input possibility: 0...255 ( 0 = infinite).
Then with the key back to RUN (➾ Section 6.4.4.4.), or
with key until END and then
Starting u
p
28
with key back to "setpoint selection". (level 0).
6.4.4.4 Starting of the programmer
From level 0 proceed as described under 6.4.4. When PGM appears on the display, press the key on the left.
The display shows RUN, or
forward with key from EDIT resp. LOOP until RUN appears, then
start the programme with the key on the left. The setpoint level (level 0) is then on display.
During the start the current setpoint is accepted as the starting value.
If the programmer is activated: level 0 shows PGM XXX.XX instead of SET XXX.XX (setpoint temperature), with PGM flashing short
(short off, long on)
.
Setpoint can not be input with 0,01 °C resolution (only possible via RS 232 interface.)
General Rule: Programmer can also be loaded and operated via the RS 232 interface
Starting up
10.08.99 YACE0060
29
6.4.4.5 Inserting/ deleting of programme segments
1. From level 0 proceed as described under 6.4.4. When PGM appears on the display, press the key on the left.
2. The display shows RUN (or STOP if the programmer had been started). Scroll with this key until EDIT appears.
3. The display shows EDIT, press the key.
4. Segment number is flashing, press the key
or
Using the keys, select the segment number
behind
which
the new segment has to be inserted.
EXAMPLE:
Section 6.4.4.1.: A new segment has to be inserted after segment 2. Select segment 2 with the keys.
With the key forward until NEW appears. The segment number ”2” behind which the new segment and the segment temperature are indicated.
Press the key, the display shows NEW as well as the segment number and temperature of the new segment.
Then jump with the key to segment temperature or segment time and start input with the key, then
or
using both keys input the required segment time and segment temperature, then continue as described below (step 5-8).
Starting u
p
30
When a new segment is inserted, all subsequent segments are shifted on by one position.( ➾ example Section 6.4.4.1).
When 20 segments are inserted, the last one will disappear when a new segment will be input.
New segments can also be inserted while the programmer is activated.
or
In order to delete a segment, proceed as above step 1-4.
Using the keys, select the segment number which has to be deleted.
Scroll forward with the key until DEL appears. Next to it the segment number to be deleted is shown.
Press the key, the segment is deleted.
5. Then with the key forward to END
6. with the key to EDIT, then
7. with the key to END and
8. with the key back to level 0.
When a segment is deleted, all subsequent segments move forward by one position.
When the programmer is activated, the currently active segment cannot be deleted.
Starting up
10.08.99 YACE0060
31
− −
To input a segment time longer than 255 min it has to be split between several consecutive segments.
6.4.4.6 Holding/continuing the programme
When the programmer is activated, the programme can at any time be held and be continued again. For this
1. When the programmer is activated, the programme can at any time be held and be continued again. For this
2. The display shows STOP. Scroll forward with the key until PAUSE appears.
The currently running segment is shown after PAUSE, press the key.
The programme is held. Scroll forward to END, then press the key to return to level 0
− −
When the programme is held with PAUSE, the display at level 0 no longer shows SET XXX.XX (set temperature) but PGM XXX.XX, with PGM not flashing.
To continue the programme, proceed as described in steps 1-2 above, but scroll forward until the display shows CONt. The segment number of the programmer during which the programme was held, is shown after CONt.
Press the key, the programme continues, the programmer jumps back to level.
When the programme is continued, the display PGM XXX.XX at level 0 is again flashing.
− −
If there is a fault, the programme is stopped by PAUSE. After the fault has rectified, the system has to be reset with CONt.
Starting u
p
32
6.4.4.7 Terminating the programme
From level 0 proceed as described under 6.4.4. When the display shows PGM, press the key.
The display shows STOP. The current segment is indicated after STOP. Press the key, the programme is terminated immediately
With the key, forward to END, then
with the key back to level 0.
Starting up
10.08.99 YACE0060
33
6.4.5 Parameters
and
4x
Directly from level 0 (setpoint selection) press the key combination on the left (on RE 3xx)
and 3
x
on RE 2xx ,
or
forward with the key from PGM function.
Here it is possible to switch over to level 2. Press the key on the left, continue with Section 6.4.5.1. or
with key to step END.
End of the menu.
Return to level 0 (setpoint selection) with the key on the left, or
with key back to programmer (RE 3xx only) or to "Setting the setpoint resolution" (RE 2xx).
Starting u
p
34
6.4.5.1 Setting the proportional band of the PID-controller
and 4
x
and
3x
Directly from level 0 (setpoint selection), press the key combination on the left (top one on RE 3xx, bottom one RE 2xx), until the PARA function is reached, then
switch to level 2 from PARA (see above) with key on the left.
or
Within this level it is possible to scroll with the keys.
The display shows the current bath temperature Xp and the current setting. To alter the setting press the key on the left. Available settings from 0,5 to 9,9 °C . (➾ Section 6.4.5.2.
Note).
(➾ Section 9. Technical data))
Display flashes (approx. ca. 4 s).
or
During this time start to set the required value with one of the two keys
Forward with the key to "Setting the reset time" or
with the key back to "PARA".
Starting up
10.08.99 YACE0060
35
6.4.5.2 Setting the reset time of the PID-controller
and
4x
and
3x
Directly from level 0 (setpoint selection), press the key combination on the left (top one on RE 3xx, bottom one RE 2xx), until the PARA function is reached, then
and
1x
switch to level 2 and move forward with the keys on the left.
The display shows the current bath temperature, Tn and the current setting. To alter the setting, press the key on the left.
!
Possible adjustment from 5 to 60 s
The display is flashing (approx. 4 s).
or
During this time, start to set the required value with one of the two keys.
Forward with the key to "Selection of the interface" or
with key back to "Setting the proportional band".
− −
The control parameters are pre-set to suit the unit type. Normally no change is required. Some adjustment is necessary only when using Silicone oil and with very stringent demands on short-term stability. If there are control fluctuations, increase the values for Xp and Tn. If the setpoint is not reached ! select smaller values. The derivative time Tv (D-part) is altered automatically through a fixed factor to Tn. (Standard settings of control parameters and pump ➾ Section 9. Technical data)
6.4.5.3 Selection of the interface
and
4x
and
3x
Directly from level 0 (setpoint selection), press the key combination on the left (top one on RE 3xx, bottom one
Starting u
p
36
RE 2xx), until the PARA function is reached, then
and
2x
switch to level 2 and move forward with key on the left
The display shows the current bath temperature and the currently set interface. To alter the setting press the key.
or
Display flashes (approx. 4 s.). During this time start to set the required interface.
Having chosen the RS 485 interface press the key.
The display shows the current bath temperature, ADR and the current unit address (e.g.. 15). Press the key, display flashes (approx. 4 s.)
or
During this time start to set the required unit address. (possible setting: 000...127).
Forward with key to ”Setting the baud rate” or
back with key to ”Selection of the interface” resp. when pressing the key twice you return to ”Setting the reset time...”“
6.4.5.4 Setting the Baud rate/ transfer rate (serial interface)
and
4x
and 3x
Directly from level 0 (setpoint selection), press the key combination on the left (top one on RE 3xx, bottom one RE 2xx), until the PARA function is reached, then
Starting up
10.08.99 YACE0060
37
and
3x
switch to level 2 and move forward with key on the left.
The display shows the current bath temperature, KBD and the current setting. To alter the setting press the key.
Display flashes (approx. 4 s).
or
During this time, start to set the required value with one of the two keys. Settings available: 2400, 4800, 9600 or 19200 baud
Forward with key to "user calibration"
or
Forward with key to "user calibration" or.
Starting u
p
38
6.4.5.5 User calibration
Remove the external consumers and switch the setting knob of the pump to right side. (➾Section 5.4).
A reference thermometer with necessary accuracy is required. Otherwise the factory calibration should not be altered. The reference thermometer has to be inserted far enough and long enough into the bath.
It is not allowed to calibrate to more than ± 3 °C. Multiple calibration to more than ± 3 °C cause internal faults (after 2 min ”E1006” or ”e16”).
− −
The factory calibration will be lost through overwriting ! please work carefully!!!
and
4x
and
3x
Directly from level 0 (setpoint selection), press the key combination on the left (top one on RE 3xx, bottom one RE 2xx), until the PARA-function is reached, then
and
4x
switch to level 2 and move forward with key on the left..
>2,5 s
1. The display shows CAL and the current bath temperature. To carry out a calibration, press the key longer than 2.5 s.
2. The actual value appears
or
3. Input the value indicated on the reference thermometer with one to the two keys.
4. The additive calibration must be entered with the key shown on the left.
5. Forward with key to "END", then
Starting up
10.08.99 YACE0060
39
6. Switch back to level 0 or
7. with key back to "Setting the transfer rate
Example
a) Insert a suitable thermometer into the bath (long enough and far enough).
b) Remove the external consumers and turn the setting knob of the pump outflows to the
right side.
c) Set the setpoint to a temperature where you use to work (e.g. set the setpoint to 45°C
(➾ Section 6.3.)
d) Wait until the actual bath temperature has reached the setpoint temperature of 45°C
and until the indication on the reference thermometer does not change any more.
e) Remove the reference thermometer, which shows e.g. 44,8 °C.
f) Select CAL on the display and go forward as mentioned under point 1-7 (see
above).The actual bath temperature switches from 45°C to 44,8°C and the unit starts to heat up until the actual bath temperature has reached 45°C. (! the reference thermometer should also indicate 45°C).
6.5 Warning and safety functions
6.5.1 Overtemperature protection and testing
− −
The units are designed for operation with non-flammable and flammable liquids to EN 61010-2-010
Set the overtemperature switch-off point. Recommended setting 5°C above required bath temperature.
Not higher than 25 °C below the firepoint of the bath liquid (➾ Section 5.3.).
The actual switch-off point is indicated on the display, e.g. max. 110°C
Starting u
p
40
When the switch-off point is being adjusted by more than 2°C ! display shows MAX and actual overtemperature switch-off point with 1°C resolution for approx. 4 s.
The position of the potentiometer is decisive for the setting. The display is just a help for the setting.
− −
Setting is possible only up to a upper limit of the operating temperature range + 5 °C.
"
""
"

When the bath temperature arises above the overtemperature switch-off point.
1. Double signal tone sounds
2. The display shows the indication for overtemperature (TEMP) the fault triangle is flashing
!
heating is switched off on both poles,
!
pump and refrigeration system are switched off by the
electronics.
Rectify the cause of the fault.
Wait until the bath temperature has cooled down below the switch-off point or set the switch-off point at a higher value. When the display shows TEMP.
Reset with the key
− −
Before the unit is running unattended for longer periods
overtemperature protection
should be tested.
Therefore
Turn the potentiometer slowly anticlockwise. ! The unit must switch off at the bath temperature.
Step 1 - 2 (see above) must follow.
Set the overtemperature switch-off point again above the bath temperature and wait until the indication TEMP appears on the display, then
reset with the key.
Starting up
10.08.99 YACE0060
41
6.5.2 Low-level protection and testing
"
""
"

Double signal tone sounds, if the bath liquid falls so much that the heater is no longer covered with liquid completely
1. The display shows LEVEL (low-level) and the fault triangle is flashing
!
heating is switched off on both poles,
!
pump and refrigeration system are switched off by
the electronics.
2. Top up the bath ➾ Section 5.2 and reset with the key.
If necessary repeat several times in case that the pump does not purge immediately.
Testing at regular intervals by lowering the bath level. Place a hose on the pump connector and pump some of the bath liquid into a suitable container.
− −
Step 1 - 2 must follow.
Bath temperature during this test not below 0°C or higher than 50°C, otherwise danger of burn injuries !
If there is any irregularity when testing the safety devices, switch off the unit immediately and pull out the mains plug !
Have the unit checked by the LAUDA service or the local service organisation!
Starting u
p
42
6.5.3 Pump motor monitoring
In case of pump motor overload or a blockage the heating and the pump are switched off.
"
""
"

Double signal tone sounds.
The display shows PUMP and the fault triangle is flashing
Rectify the cause of the fault, e.g. clean the pump or check the viscosity, then
reset with the key.
If several faults appear simultaneously, they have to be reset individually
6.5.4 Connection floating contact ”Combination fault” 31 N
(Alarm out) 3-pole locking connector 1 = common, 2 = n.c. (make) , 3 = n.c. (break). When the unit is o.k. 1 and 3 are closed (see ill.). Max contact rating: 0.2A 24 V. 3-pin plug (➾ Section 10. Accessories).
Contact operates in case of error at overtemperature protection, low-level protection, pump monitoring or any other error message.
Starting up
10.08.99 YACE0060
43
6.5.5 Other error messages
E 0 etc. is flashing in the bottom line
!
various temperature probe faults.
!
pump fault, proceed as in Section 6.5.3.
If the fault report is repeated ! pull out the mains plug and try whether the motor can be rotated by the fan blade inserting a screwdriver into the ventilation opening at the back of the unit.
Error code 0 ...255 ! microprocessor error.
Error code 1000...1255 ! slave processor error.
Indication can be used for remote diagnosis.
After rectifying the fault, reset with the key.
Safety notes
44
7 Safety notes
7.1 General safety notes
A laboratory thermostat is intended for heating and pumping liquids according to the needs of the user. This leads to hazards by high temperatures, fire, and the general hazards by the use of electrical energy.
The user is largely protected through the application of the appropriate standard specifications..
Additional hazards may arise from the type of material being thermostated, e.g. when going above or below certain temperature levels or through breaking of the container and reaction with the thermostating liquid. It is not possible to cover all possibilities; they remain largely within the responsibility and the judgement of the user.
The unit must only be used as intended and as described in these Operating Instructions. This includes operation by suitably instructed qualified personnel.
The units are not designed for use under medical conditions according to EN 60601-1 or IEC 601-1 !
7.2 Other safety notes
Connect the unit only to grounded mains power (PE).
Parts of the bath cover may reach surface temperatures above 70 °C when operating at higher temperatures. Take care when touching it!
Use suitable hoses ➾ Section 5.3..
Protect tubing with hose clips against slipping off. Prevent kinking of tubing!
Check tubing from time to time for possible material defects!
Heat transfer tubing and other hot parts must not come into contact with the supply cable!
When using the thermostat as circulation thermostat, failure of tubing may lead to leaking of hot liquid and become a danger to personnel and objects.
When no external consumer is connected to the thermostat the pump outflow must be closed (use closing plugs) or linked to the return!
Allow for expansion of the bath oil at elevated temperatures!
Safety notes
10.08.99 YACE0060
45
Depending on the bath liquid used and the mode of operation it is possible for toxic vapours to be produced. Ensure appropriate ventilation!
When changing the bath liquid from water to oil, for temperatures above 100 °C, carefully remove all traces of water, also from tubing and from the external consumer, otherwise ! danger of burns through delayed boiling!
Always pull out the mains plug before cleaning, maintenance or moving the thermostat!
Repairs on the control unit and the refrigeration system must be carried out by properly qualified personnel only
Values for temperature control and indicating accuracy apply under normal conditions according to DIN 58966. High-frequency electromagnetic fields may under special conditions lead to unfavourable values. This does not affect the safety
Maintenance
46
8 Maintenance
8.1 Cleaning
Before cleaning the unit, pull out the mains plug!
The unit can be cleaned with water adding a few drops of detergent (washing up liquid), using a moist cloth
Water must not enter the control unit
Carry out appropriate detoxification if dangerous material has been spilled on or inside the unit.
Method of cleaning and detoxification are decided by the special knowledge of the user. In case of doubt please contact the manufacturer
8.2 Maintenance and repair
Before any maintenance and repair work pull out the mains plug!
Repairs on the control unit must only be carried out by properly qualified personnel!
LAUDA thermostats are largely maintenance-free. If the thermostating liquid becomes dirty it has to be replaced (➾ Section 5.2.).
If a fuse blows (! supply indication not alight) fit only fuses as specified (2 x T 16 A; 1 x T 2,5 A, size 5 x 20 ! fuses are inside the unit).
F3; T 2,5 A
UL 487-1
F1; T 16 A
F2; T 16 A
Maintenance
10.08.99 YACE0060
47
8.2.1 Maintenance of the refrigeration unit
The refrigeration unit operates largely without maintenance. Depending on the ambient dust conditions and the operating time, any dust on the heat exchanger (condenser) must be removed at intervals on 2 weeks or longer. This is done after taking off the front grille. Brush off the condenser and if necessary blow through with compressed air.
8.2.2 Note on repair and disposal
The refrigeration circuit is filled with a CFC-free refrigerant. Type and charging quantity are indicated on the unit. Repair and disposal by a qualified refrigeration engineer only!
If the equipment does have to be returned to the factory, it may only be necessary to dismantle the thermostat unit and return it.
If the equipment has to be returned to the factory, please ensure that it is carefully and properly packed. LAUDA accepts no responsibility for damage due to unsatisfactory packing.
8.3 Ordering spares
When ordering spares please quote instrument type and serial number from the rating label. This avoids queries and supply of incorrect items.
We shall always be happy to deal with queries and to receive suggestions and criticism.
LAUDA DR. R. WOBSER
GMBH & CO.KG
P.O. Box 1251
97912 Lauda-Königshofen
GERMANY
Phone: (+49) (0) 9343/ 503-0
Fax: (+49) (0) 9343/ 503-222
E–mail info@lauda.de
Internet http://www.lauda.de
Technical data
48
9 Technical data (to DIN 58966)
Low-temperature thermostats
RE 204 RE 206 RE 207 RE 212 RE 220
Operating temperature range °C -10 to 200 -20 to 200 -35 to 200 -30 to 200 -30 to 200 Ambient temperature range °C 5...40
(#Section 5.1) Setting resolution °C 0.01 Indication resolution °C 0.05 Indication accuracy ±0.2 °C ➂ additive re-calibration Temperature control ± °C 0.01 0.01 0.02 0.02 0.02 Safety features
FL
Additional functions 2-line LCD-display
autom. compressor control autom. compressor control + cooling adjustments Heater power
230 V; 50/ 60 Hz 115 V; 60 Hz 100 V; 50/60 Hz
kW 2.25
1.3
1.0 Cooling capacity (eff.) @ with ethanol at 20 °C ambient temperature
20°C
0°C
-20°C
-30°C
kW 0.15
0.10
0.05*
----
0.20
0.15
0.05
----
0.30
0.22
0.10
0.06
0.30
0.23
0.13
0.04
0.35
0.25
0.10
0.04 Pump type pressure pump with choice of 5 output steps Max. discharge pressure
bar 0.4
Max. flow rate
l/min 17 Pump connections mm nipples 13 mm dia. Max. bath volume l 3 to 4.5 4 to 6 4 to 6 9 to 12 14 to 20 Bath opening (W x D) mm 130x105 150x130 150x130 200x200 300x350 Bath depth mm 160 160 160 200 160 Usable depth mm 140 140 140 180 140 Height top edge of bath mm 363 396 396 441 441 Overall size (W x D x H) mm 180x320x524 200x400x557 200x400x557 250x400x602 350x530x602 Weight kg 21 24 26 30 34 Power con­sumption
230 V; 50/ 60Hz 115 V; 60 Hz 100 V; 50/60 Hz
kW 2.5
1.4
1.1
2.6
1.4
1.1
2.7
1.4 1,1
2.7
1.4
1.1
2.8
1.4
1.2
Ref. No. 230 V; 50/60 Hz LCK 0865
230 V; 50 Hz LCK 1862 LCK 1868 LCK 1871 LCK 1874 230 V; 60 Hz LCK 2862 LCK 2868 LCK 2871 LCK 2874 115 V; 60 Hz LCK 4862 LCK 4865 LCK 4868 LCK 4871 LCK 4874
100 V; 50/60 Hz LCK 6862 LCK 6865 LCK 6868 LCK 6871 LCK 6874
*@ -10°C ➀ FL: suitable for flammable and non-flammable liquids
➂ ➩
Section 7.2 last item
➁ at pump output step 5
We reserve the right to make technical alterations!
Technical data
10.08.99 YACE0060
49
RE 306 RE 307 RE 312 RE 320
Operating temperature range °C -20...200 -35...200 -30...200 -30...200 Ambient temperature range °C 5...40 ( # Section 5.1.) Setting resolution °C 0.01 Indication resolution °C 0.05 Indication accuracy ±0.2 °C ➂ additive re-calibration Temperature control ±°C 0.01 0.02 0.02 0.02 Safety features
FL
Additional function 2-line LCD-display and back-lightning. programmer
autom.
compressor
control
autom. compressor control + cooling adjustment
Heater power
230 V; 50/ 60 Hz 115 V; 60 Hz 100 V; 50/60 Hz
kW 2.25
1.3
1.0 Cooling capacity (eff.) @ with ethanol at 20°C ambient temperature
20°C
0°C
-20°C
-30°C
kW 0.20
0.15
0.05
----
0.30
0.22
0.10
0.06
0.30
0.23
0.13
0.04
0.35
0.25
0.10
0.04 Pump type pressure pump with choice of 5 output steps Max. discharge pressure
Bar 0.4
Max. flow rate
l/min 17
Pump connections mm nipples 13 mm dia. ( M 16 x 1 )
Max. bath volume l 4 to 6 4 to 6 9 to 12 14 to 20 Bath opening (W x D) mm 150x130 150x130 200x200 300x350 Bath depth mm 160 160 200 160 Usable depth mm 140 140 180 140 Height top edge of bath mm 396 396 441 441 Overall size (W x D x H) mm 200x400x557 200x400x557 250x400x602 350x530x602 Weight kg 24 26 30 34
Power con­sumption
230 V; 50/ 60Hz 115 V; 60 Hz 100 V; 50/60 Hz
kW 2.6
1.4
1.1
2.7
1.4
1.1
2.7
1.4
1.1
2.8
1.4
1.2
Ref. No. 230 V; 50/60
Hz
LCK 0866
230 V; 50 Hz LCK 1869 LCK 1872 LCK 1875 230 V; 60 Hz LCK 2869 LCK 2872 LCK 2875 115 V; 60 Hz LCK 4866 LCK 4869 LCK 4872 LCK 4875
100 V; 50/60 Hz LCK 6866 LCK 6869 LCK 6872 LCK 6875
*@ -10°C ➀ FL suitable for flammable and non-flammable liquids ➁ at pump output step 5 ➂ ! Section 7.2 last item
Units to EU Directive 89/336/EWG (EMC) and 73/ 23/ EWG (low-voltage) with CE-mark.
We reserve the right to make technical alterations!
Technical data
50
Standard settings of control parameters and pump ➾ Section 6.4.5.1. and 6.4.5.2. and 6.4.2.
Unit Type Xp (°C) Tn (s) Pump outflow step
RE 104, RE 204 7,0 20 2 RE 106, RE 206, RE 306 4,0 20 2 RE 107, RE 207, RE 307 3,0 25 3 RE 112, RE 212, RE 312 4,5 30 4 RE 120, RE 220, RE 320 2,0 25 5
Pump characteristics
measured with water
0
0,05
0,1
0,15
0,2
0,25
0,3
0,35
0,4
0,45
024681012141618
l/min
bar
step 5
step 4
step 3
step 2
step 1
Technical data
10.08.99 YACE0060
51
Cooling curve
measured with ethanol
Bath liquid:
water/glycol 1:1
(to -25°C) as bath liquid
Time from graph x 1,7
We reserve the right to make technical alterations!
Accessories
52
10 Accessories
Accessories suitable for Ref. No.
Bath cover two parts RE 220 LCZ 0633 Gable cover RE 220, RE 320 LCZ 011 Rising platform 8 steps RE 206, RE 306, RE 207,
RE 307
LCZ 0646
Rising platform 8 steps RE 212, RE 312 LCZ 0647 Rising platform 8 steps RE 220, RE 320 LCZ 0635 Pump short circuit all RE 3xx models LZM 045 3-pole locking connector (floating contact „Combination fault“ 31 N)
for all models EQS 054
Wintherm Software under Windows for all units LDSE 4001
For further accessories please contact us.
Circuit diagram
10.08.99 YACE0060
53
11 Circuit diagram
Circuit diagram
54
Circuit diagram
12.07.00 YACE0060
55
230V; 50Hz ◆ 230V; 50/60Hz ◆ [230V; 60Hz] at Serialno.: X01
RE 1xx RE 2xx RE 3xx
A 1 Printed c i rcuit board „Mains“ UL 487-1 UL 487-1 UL 487-1 A 2 Printed c i rcuit board „Display“ UL 488-1A UL 488-1B UL 488-1C A 3 Printed c i rcuit board serial interface RS 232/RS 485 ---------- UL 490 UL 490 A 4 Printed c i rcuit board Mains LED-Backlight --------- --------- UL 492 A 5 Printed c ircuit board Display LED-Backlight --------- --------- EAO 015
B 1 Pt100 probe saf ety circuit ETP 057 ETP 057 ETP 057 B 2 Pt100 probe actual value
E 1 Heater 1,5 kW EH 168 --------- ---------­E 2 Heater 2,25 kW ---------- EH 169 EH 169
M 1 Pump motor EM 109 EM 109 EM 109 S 1 Mains switch EST 101 EST 101 EST 101 U 3 SSR (BRT22H) Y 1 output A1 --------- EYI 158 --------­X 1 Mains connec t i on EKN 001 EKN 001 EKN 001
X 2 Lock s crew --------- 2x EQZ 048 2x EQZ 048 X 8 Connection s ocket Cooling (Stakei 2) --------- --------- EQK 004
EQZ 006
X 10 Connection socket Cooling unit (S takei 200) --------- --------- EQD 037
EQZ 006 X 13 Housing 2pol. --------- --------- EQF 067 X 21 Plug strip terminal 12pol. EQF 079 EQF 079 EQF 079 X 23 Line up terminal 2pol. --------- --------- EZK 063
RE 004
F 4 Pressure switch ES 045 ES 045 ---------­M 2 Compressor EMK 176 EMK 176 ---------­M 3 Fan [T 1 Trafo EIT 122 EIT 122 EIT 122]
RE 006
F 4 Pressure switch ES 045 ES 045 ES 045 M 2 Compressor EMV 011 EMV 011 EMV 011 M 3 Fan EML 007 EML 007 EML 007 [T 1 Trafo EIT 122 EIT 122 EIT 122]
RE 007
F 4 Pressure switch ES 045 ES 045 ES 045 M 2 Compressor EMV 011 EMV 011 EMV 011 M 3 Fan EML 042 EML 042 EML 042 U 3 SSR (BRT22H) Y 1 output A1 --------- EYI 158 --------­Y 1 Solenoid valve --------- EVM 079 EVM 079 [T 1 Trafo EIT 122 EIT 122 EIT 122]
RE 012
F 4 Pressure switch ES 045 ES 045 ES 045 M 2 Compressor EMV 011 EMV 011 EMV 011 M 3 Fan EML 042 EML 042 EML 042 U 3 SSR (BRT22H) Y 1 output A1 --------- EYI 158 --------­Y 1 Solenoid valve --------- EVM 079 EVM 079 [T 1 Trafo EIT 122 EIT 122 EIT 122]
RE 020
F 4 Pressure switch ES 045 ES 045 ES 045 M 2 Compressor EMK 146 EMK 146 EMK 146 M 3 Fan U 3 SSR (BRT22H) Y 1 output A1 --------- EYI 158 --------­Y 1 Solenoid valve --------- EVM 079 EVM 079 [T 1 Trafo EIT 122 EIT 122 EIT 122]
Circuit diagram
56
115V; 60Hz ◆ [100V; 50/60Hz] at Serialno.: X01
RE 1xx RE 2xx RE 3xx
A 1 Printed c i rcuit board „Mains“ UL 499 UL 499 UL 499 A 2 Printed c i rcuit board „Display“ UL 488-1A UL 488-1B UL 488-1C A 3 Printed c i rcuit board serial interface RS 232/RS 485 ---------- UL 490 UL 490 A 4 Printed c i rcuit board Mains LED-Backlight --------- --------- UL 492 A 5 Printed c ircuit board Display LED-Backlight --------- --------- EAO 015
B 1 Pt100 probe saf ety circuit ETP 057 ETP 057 ETP 057 B 2 Pt100 probe actual value
E 1 Heater 1,3 kW at 115V
1,0 kW at 100V
EH 171 E H 171 EH 171
M 1 Pump motor EM 109 EM 109 EM 109 S 1 Mains switch EST 101 EST 101 ES T 101 U 3 SSR (BRT22H) Y 1 output A1 --------- EYI 158 --------­X 1 Mains connection EKN 003 EKN 003 EKN 003
X 2 Lock screw --------- 2x EQZ 048 2x EQZ 048 X 8 Connection s ocket Cooling (Stakei 2) --------- --------- EQK 004
EQZ 006
X 10 Connection socket Cooling unit (S takei 200) --------- --------- EQD 037
EQZ 006 X 13 Housing 2pol. --------- --------- EQF 067 X 21 Plug strip terminal 12pol. EQF 079 EQF 079 EQF 079 X 23 Line up terminal 2pol. --------- --------- EZK 063
RE 004
F 4 Pressure switch ES 045 ES 045 ---------­M 2 Compressor EMK 176 EMK 176 ---------­M 3 Fan [T 1 Traf o EIT 122 EIT 122 EIT 122]
RE 006
F 4 Pressure switch ES 045 ES 045 ES 045 M 2 Compressor EMV 012 EMV 012 EMV 012 M 3 Fan EML 033 EML 033 EML 033 [T 1 Traf o EIT 122 EIT 122 EIT 122]
RE 007
F 4 Pressure switch ES 045 ES 045 ES 045 M 2 Compressor EMV 012 EMV 012 EMV 012 M 3 Fan EML 033 EML 033 EML 033 U 3 SSR (BRT22H) Y 1 output A1 --------- EYI 158 --------­Y 1 Solenoid valve --------- EVM 079 EVM 079 [T 1 Traf o EIT 122 EIT 122 EIT 122]
RE 012
F 4 Pressure switch ES 045 ES 045 ES 045 M 2 Compressor EMV 012 EMV 012 EMV 012 M 3 Fan EML 033 EML 033 EML 033 U 3 SSR (BRT22H) Y 1 output A1 --------- EYI 158 --------­Y 1 Solenoid valve --------- EVM 079 EVM 079 [T 1 Traf o EIT 122 EIT 122 EIT 122]
RE 020
F 4 Pressure switch ES 045 ES 045 ES 045 M 2 Compressor EMK 181 EMK 181 EMK 181 M 3 Fan U 3 SSR (BRT22H) Y 1 output A1 --------- EYI 158 --------­Y 1 Solenoid valve --------- EVM 079 EVM 079 [T 1 Traf o EIT 122 EIT 122 EIT 122]
Pipe plan
16.08.99 YACE0060
57
12 Pipe plan
Pipe plan
58
Pipe plan
16.08.99 YACE0060
59
Pipe plan
60
230V; 50Hz ◆ 230V; 50/60Hz ◆ [230V; 60Hz] at Serialno.: X01
Typ RE 1xx RE 2xx RE 3xx
RE 004
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ---------­M 2+M 3+J 5 Cooling unit PL 50 FX NO EMK 173 EMK 173 ---------­J 1 Drier EO 003 EO 003 ---------­J 2 Capillary HKA 114 HKA 114 ---------­J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­[T 1 Trafo EIT 124 EIT 124 ----------]
RE 006
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ES 045 M 2 Compressor AZ 411 EMV 011 EMV 011 EMV 011 M 3 Fan W2S-115-AB07-16 EML 007 EML 007 EML 007 J 1 Drier EO 003 EO 003 EO 003 J 2 Capillary HKA 114 HKA 114 HKA 114 J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­J 5 Condenser EOW 084 EOW 084 EOW 084
RE 007
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ES 045 M 2 Compressor AZ 411 EMV 011 EMV 011 EMV 011 M 3 Fan W2S-130-AA75-(A2) EML 042 EML 042 EML 042 J 1 Filter drier / receiver CNO 432 s EO 040 EO 040 EO 040 J 2 Injection valve TEL 0,5 R404A MOP 3,0 EVE 111 EVE 111 EVE 111 J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­J 5 Condenser EOW 085 EOW 085 EOW 085 Y 1 Solenoid valve 2/2Wege 6mm Löt ---------- EVM 077 EVM 077
Typ: 1028/2
[T 1 Trafo EIT 125 EIT 125 EIT 125]
RE 012
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ES 045 M 2 Compressor AZ 411 EMV 011 EMV 011 EMV 011 M 3 Fan W2S-130-AA75-(A2) EML 042 EML 042 EML 042 J 1 Filter drier / receiver CNO 432 s EO 040 EO 040 EO 040 J 2 Injection valve TEL 0,5 R404A MOP 3,0 EVE 111 EVE 111 EVE 111 J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­Y 1 Solenoid valve 2/2Wege 6mm Löt ---------- EVM 077 EVM 077
Typ: 1028/2
[T 1 Trafo EIT 125 EIT 125 EIT 125]
RE 020
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ES 045 M 2+M 3+J 5 Cooling unit UB 6144 Z EMK 146 EMK 146 EMK 146 J 1 Drier DN 032 s EO 041 EO 041 EO 041 J 2 Injection valve TEL 0,5 R404A MOP 3,6 EVE 112 EVE 112 EVE 112 J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­Y 1 Solenoid valve 2/2Wege 6mm Löt ---------- EVM 077 EVM 077
Typ: 1028/2
[T 1 Trafo EIT 125 EIT 125 EIT 125]
Pipe plan
16.08.99 YACE0060
61
1150V; 60Hz ◆ [100V; 50/60Hz] at Serialno.: X01
Typ RE 1xx RE 2xx RE 3xx
RE 004
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ---------­M 2+M 3+J 5 Cooling unit PL 50 FX NO EMK 176 EMK 176 ----------
J 1 Drier EO 003 EO 003 ---------­J 2 Capillary HKA 114 HKA 114 ---------­J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­[T 1 Trafo EIT 123 EIT 123 ----------]
RE 006
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ES 045 M 2 Compressor AZ 4419 Y-A EMV 012 EMV 012 EMV 012 M 3 Fan W2S-115-AB07-07 EML 015 EML 015 EML 015 J 1 Drier EO 003 EO 003 EO 003 J 2 Capillary HKA 114 HKA 114 HKA 114 J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­J 5 Condenser EOW 084 EOW 084 EOW 084 [T 1 Trafo EIT 122 EIT 122 EIT 122]
RE 007
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ES 045 M 2 Compressor AZ 4419 Y-A EMV 012 EMV 012 EMV 012 M 3 Fan W2E-142-CC13-16 EML 033 EML 033 EML 033 J 1 Filter drier / receiver CNO 432 s EO 040 EO 040 EO 040 J 2 Injection valve TEL 0,5 R404A MOP 3,0 EVE 111 EVE 111 EVE 111 J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­J 5 Condenser EOW 085 EOW 085 EOW 085 Y 1 Solenoid valve 2/2Wege 6mm Löt ---------- EVM 079 EVM 079
Typ: 1028/2
[T 1 Trafo EIT 122 EIT 122 EIT 122]
RE 012
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ES 045 M 2 Compressor AZ 4419 Y-A EMV 012 EMV 012 EMV 012 M 3 Fan W2E-142-CC13-16 EML 033 EML 033 EML 033 J 1 Filter drier / receiver CNO 432 s EO 040 EO 040 EO 040 J 2 Injection valve TEL 0,5 R404A MOP 3,0 EVE 111 EVE 111 EVE 111 J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­Y 1 Solenoid valve 2/2Wege 6mm Löt ---------- EVM 079 EVM 079
Typ: 1028/2
[T 1 Trafo EIT 122 EIT 122 EIT 122]
RE 020
F 4 Pressure switch CC20 24/17 bar ES 045 ES 045 ES 045 M 2+M 3+J 5 Cooling unit UB 6144 Z/2 EMK 181 EMK 181 EMK 181 J 1 Filter drier DN 032 s EO 041 EO 041 EO 041 J 2 Injection valve TEL 0,5 R404A MOP 3,6 EVE 112 EVE 112 EVE 112 J 3 Evaporator ---------- ---------- ---------­J 4 Bath ---------- ---------- ---------­Y 1 Solenoid valve 2/2Wege 6mm Löt ---------- EVM 079 EVM 079
Typ: 1028/2
[T 1 Trafo EIT 122 EIT 122 EIT 122]
BESTÄTIGUNG / CONFIRMATION / CONFIRMATION
Formblatt / Form / Formulaire: Unbedenk.doc Erstellt / published / établi: LSC Änd.-Stand / config-level / Version: 0.1 Datum / date: 30.10.1998
LAUDA DR. R. WOBSER GmbH & Co. KG
Pfarrstraße 41/43 Tel: +49 (0)9343 / 503-0 D - 97922 Lauda-Königshofen Fax: +49 (0)9343 / 503-222 Internet: http://www.lauda.de E-mail: info@lauda.de
UNBEDENK.DOC
An
/
To / A
:
LAUDA Dr. R. Wobser LAUDA Service Center Fax: +49 (0) 9343 - 503-222
Von
/
From / De
:
Firma
/ C
ompany / Entreprise
:
Straße
/ S
treet / Rue
:
Ort
/
City / Ville
:
Tel
.:
Fax
:
Betreiber
/ R
esponsible person / Personne responsable
:
Hiermit bestätigen wir, daß nachfolgend aufgeführtes LAUDA-Gerät (Daten vom Typenschild):
We herewith confirm that the following LAUDA-equipment (see label): Par la présente nous confirmons que l’appareil LAUDA (voir plaque signalétique):
Typ
/
Type / Type
:
Serien-Nr.
/
Serial no. / No. de série:
mit folgendem Medium betrieben wurde
was used with the below mentioned media a été utilisé avec le liquide suivant
Darüber hinaus bestätigen wir, daß das oben aufgeführte Gerät sorgfältig gereinigt wurde, die Anschlüsse verschlossen sind, und sich weder giftige, aggressive, radioaktive noch andere gefährliche Medien in dem Gerät befinden.