Operating instruction
DigiCal Digital Thermometer
DCS 2
DigiCal DCS 2
YAME0017/ 09.10.03 Safety notes 3
Safety notes:
Before operating the equiqment please read carefully all the instructions and safet y notes. If you have any questions please phone u s!
Follow the instruc tions on setting up, operation etc. T his is the only way to avoid
incorrect operat ion of the equipment and to ensure full warranty protection.
Transport the equi pm ent with care!
Equipment and ist int er nal par ts can be damaged:
by dropping
by shock.
Equipment should only be operated by technically qual ified personnel!
Do not start up the equipment if
it is damaged or leaki ng
the supply cable is damaged.
Switch off the equipment and pull out the mains plug for:
servicing or repai r
before moving the equipment!
Have the equipment servi c ed or r epaired by properly qualifi ed per sonnel only!
The Operating Instructions include additional saf ety notes which are identified by a triangle with an exclamat io n mark. Carefully read the instructi on s and follo w them accurately! Disregarding the instructions may h ave serious consequences, such as damage to the equipment, damage to property or injury to personnel!
We reserve the right to make technical alterations!
DigiCal DCS 2
4 Contents YAME0017/09.10.03
Content
1.
Control and functional elements.............................................................................................6
2. Handling / Introduction............................................................................................................ 8
2.1 General adv ic e...................................................................................................................................8
2.2 Operation........................................................................................................................................... 8
2.3 Switching on and off........................................................................................................................... 8
2.4 Menu.................................................................................................................................................. 9
2.4.1 Measuring unit switc hing °C and °F -
Tlhr
............................................................................. 10
2.4.2 Differenc e temper ature –
Khl1
.............................................................................................. 10
2.4.3 Calibration function -
B@K
........................................................................................................ 11
2.4.3.1 Encoded Calibr ation -
mO0
............................................................................................... 12
2.4.3.2 Calibrati on by physic al standar d references -
mO1
............................................................ 13
2.4.3.3 DIN-Ccalibration DIN IEC 60751 - [
mEE
.......................................................................... 13
2.4.4 Printer port activ ation / deactivation -
OpDl
............................................................................ 14
2.4.5 Selection / Channel activation -
Bgl
..................................................................................... 15
2.4.6 Analogue output -
c@B
............................................................................................................ 16
2.5 Recalling the memory data (HOLD MAX MIN AVE).......................................................................... 17
2.6 Measuring rat e ( FAST-Modus) ......................................................................................................... 17
2.7 AUTO-OFF-Function........................................................................................................................ 17
3. Power supply..........................................................................................................................18
4. Error codes............................................................................................................................. 18
5. Technical data........................................................................................................................19
6. RS232 interface protocol....................................................................................................... 20
7. Connector layout.................................................................................................................... 21
7.1 Connector layout Pt100 4-wire.......................................................................................................... 21
7.2 RS232 PC-adapter c able (Connector layout) .................................................................................... 21
8. Data of probe-calibration.......................................................................................................22
9. Accessories.................................................................................................................... ........ 23
10. Ordering spares......................................................................................................................23
DigiCal DCS 2
YAME0017/ 09.10.03 Explanation of si gns 5
Explanation of signs:
Danger: This symbol is used where there may be injury to
personnel through incorrect handling.
Note: Here spezial attent ion is drawn to some aspect. May
include referenc e to danger.
þ
Reference: Refers to additional information in other secti ons.
DigiCal DCS 2
6 Control and functional elements YAME0017/09.10.03
1. Control a nd fu nc ti on al elements
DCS 2 front view
Measuring port for channel 2
Temperature-M easuring probe channel 1 (firml y connected)
Control panel (Key s) ON/OFF - switch Righthand
adjustable base
DigiCal DCS 2
YAME0017/ 09.10.03 Handling / Introduction 7
DCS 2 rear view
Analogue outl put 0-1V for channel 1 (red= plus)
Analogue output 0- 1V for channel 2 (red= plus)
RS 232 Interface (9 pins)
Power supply input
Temperature measuring probe channel 1 (firmly connected)
DigiCal DCS 2
8 Handling / Introduction YAME0017/09.10.03
2. Handling / Introduction
2.1 General advice
For cleaning the instr ument please do not use abrasive cleaner but a dry or wet pi ece of cloth.
Please store the measuring instrument in a dry and clean place.
Avoid any force like shocks or pressure to the instrument.
Do not use force to connect the pr obe or i nterface plugs in. The interface plug differs
from the probe.
þ
If no sensor is connected to t he instrument while switching on
mODl
will be
indicated on the displ ay (þ chapter 4 Error codes )
2.2 Operation
Please kindly chec k if the prov ided mains plug unit is working properly with y our power suppl y ,. The standard equipment is for 230V; 50/60Hz power supply . If your mains supply a diff er ent voltage, you need the universal (all) mains unit ( 90- 260V; 50/60Hz). Please connect first the measuring devic e with t he suitable mains plug unit (see picture on the back side of the device) and then connect the mains plug with the mains supply socket. (
þ
3 Power supply). The measuring probe 1 i s firmly c onnec ted to CH 1. If you would like to connect a second measuring probe: Plug the
second probe into the socket CH 2 of t he measuring device. The connection socket s on the housi ng ar e marked accordingly a CH 2, RS 232 or Analogue Out.
2.3 Switching on and off
By operating the ON/OFF- k ey the instrument will be switched on or off. After switching on the instrument indicates a full segment test for 1,5 sec. Afterwards the instrument shows the calibration-code and the adjusted measurement category for
channel 1 (e.g.
[
mEE
for standard distinguishing line and P for Pt100).
In the following the c alibr ation data of the second channel (if available) will be shown.
Fig 1: Example of channel info rmation after switching on:
1
15E4 O
EEDa
Channel-1 calibr ation based on the calibration code of the probe. .
DigiCal DCS 2
YAME0017/ 09.10.03 Handling / Introduction 9
Now the thermomet er start s operating in measurement mode indicating the actual measurement value. All instrum ents display the measurement v alue of channel 1 on the top display line (large display line) . Underneath the trending bargraph is shown. The second channel of our 2-channel instruments would be displayed on the bottom line (small display line).
2.4 Menu
The adjustments of the instruments such as like the tem per ature unit, calibration of probes, deactivati on of channels and others are resulti ng from the structur e of the menu. You will reach into the main menu by pressing [ENTER/MENUE]. With the up and down keys
[
] you can choose your required menu i tem.
Push [ESC] to be back in the measuring mode.
up and down keys [
]
key [ESC]
key [ENTER/MENUE]
Menu structure
Press the key
accordingly to pass through the menu in the displayed direction. Using the key it
can be started from the opposite direction at
c@B
.
Tlhr
for channel 1+2
Khl1
B@K
for channel 1+2
OpDl
Bgl
for channel 1+2
c@B
for channel 1+2
9B
T1-T2
mEE
mEE
mEE
c@Ha
9E
mO!
ml
ml
c@HD
mO1
DigiCal DCS 2
10 Handli ng / I ntroduc tion YAME0017/09.10.03
2.4.1 Measuring unit switching °C and °F -
Tlhr
Tlhr
= Measuring unit
Measuring unit temper ature (°C=Celsius, °F=Fahrenheit). To change the measuring unit press [ENTER/MENUE]. Use the up and down keys [
] to select
Tlhr
. Press again [ENTER/MENUE]. Of the left corner of the display appears a small 1, which
indicates the selected channel. For changing the channel use the up and down keys [
].
Press [ENTER/MENUE] to confirm. On the right cor ner of the display appears
°
B
or °
E
. Use the up
and down keys [
] again to adjust the requested measuring unit and press [ENTER/MENUE] to
confirm. Press [ESC] to get back to the measuring m ode.
Display Keys Function
Tlhr
ENTER/MENUE
1
Tlhr
2
Tlhr
up and down keys [
]
select the channel 1 or 2
1
Tlhr
ENTER/MENUE
1
Tlhr
°
B
1
Tlhr
°
E
up and down keys [
]
select the measuri ng unit
[ B] or [°
E
]
ENTER/MENUE To keep the amended value ESC To escape without retention
2.4.2 Temperature Difference–
Khl1
To display the difference between the two temperat ur e channel s pu sh [ENTER/MENUE]. Use the up and down keys [
]to select
Khl1
. Push again [ENTER/MENUE]. On the right corner of the
display appears a T
1–T2 . Use the up and down keys [ ]to adjust the requested selec tion.
Push [ENTER/MENUE] to confirm. Push [ESC] to be back in the measuring mode.
Both channel have to be activated for showing the difference t em per ature.
DigiCal DCS 2
YAME0017/ 09.10.03 Handling / Introduction 11
2.4.3 Calibration function -
B@K
Despite high quali ty manufacturing techniques, eac h probe i s slightly different from specified standards. To eliminate inaccuracies caused by exchanging or ageing of probes, the instrument off er easy calibration functions which guarantee that the system accuracy is always as good as if the instrument was specifically cali br ated t o the individual probes in our laboratory .
The instruments offer three calibration options:
[
mO0
]: LAUDA-factory calibration, encoded calibration wit h optimum accuracy.
Probe and electr onic s are being calibrated at the same time. Theref or e the accuracy reaches 0,03°C for temperature range –100°C to +150°C. This calibration is stored as
mO0
calibration code.
Please note that the acc ur ac y will be 0,05°C in case another probe is connected that was calibrat ed without the electronics. (This probe data can be read as a 2x4 digi ts code on the handle of the LAUDA DigiCal probes and result s from the 2-point-com pari son met hod
mO0
.
[
mO1
]: Calibration by physical standard references with the best possibl e system
accuracy for a users calibr ation (calibrati on with comparative standards: 1-point, 2-point or 3-point-comparison possible) .
[
mEE
]: Standard characteri stic curve wi th a minimum system accuracy
(e.g. Pt100-resistanc e ac c or ding DIN IEC 60751).
B@K
= calibration
Push [ENTER/MENUE] to calibrate the instrument with sensor. Use the up and down keys [
]
to select
B@K
. Push again [ENTER/MENUE]. On the left corner the display appears a small 1, whic h
indicates the selected channel. For changing the channel use the up and down keys [
]. Push
[ENTER/MENUE] to confirm.. Use the up and down keys [
] to select the requested cal ibration option.
Display Keys Function
B@K
ENTER/MENUE
1
B@K
2
B@
up and down keys [ ]
select the channel 1 or 2
1
B@
ENTER/MENUE
1
B@
mO0 mO1
[
mEE
up and down keys [ ]
select the calibration mode
ENTER/MENUE To keep the amended value ESC To escape without retention
DigiCal DCS 2
12 Handli ng / I ntroduc tion YAME0017/09.10.03
2.4.3.1 Encoded Calibrati on -
mO0
This calibrating function is only needed in case you would add another probe to the second measuring channel .
Use the up and down keys [
] to select
mO0
. Press [ENTER/MENUE] to confirm the requested setting. On the bott om of the display appears a small 1, followed by a four- digit number (Hex-Code /0..F) . T he number can be modified with the up key [
]. For stepping t o the next number
use the down key [
]. If the requested num ber is com plete then press [ENTER/MENUE] to confirm. Now on the bottom of the display appears a small 2 and after this number a second four-digit number is displayed. For c hanging the number please follow the manual as befor e.
Press [ESC] to get back to the measuring m ode.
Important!
After confirming
mO0
by pressing [ENTER/MENUE] the functi on
mO0
(encoded calibration) is
activated, ev en though you leave the menu by pressing the [ESC] key
Display-i ndic ation with active calibr ation code
mO0
:
The
B@K
– segment and the small 1 on the left indicates to the user
that
mO0
is activated.
DigiCal DCS 2
YAME0017/ 09.10.03 Handling / Introduction 13
2.4.3.2 Calibration by physi cal stand ard references -
mO1
Use the up and down keys [ ] to select
mO1
. Press [ENTER/MENUE] to confirm. On the bottom
of the display appears
0 O
. To select between a 1-Point [
0 O
], a 2-Point [
1 O
] or 3-Point [
2 O
]-
calibration use the up and down keys [ ].
Example of a 1-Point calibrat ion: Press [ENTER/MENUE] to confirm. On the display appears
5m
. After the displayed measuring value is
stabile press [ENTER/MENUE]. App. 2 seconds later in the first di spl ay line appears
O 0
(measuring
value 1); in the second displ ay l ine appears
cO
. (standing for decimal point).
Use the up and down keys [
] to select the number of digits after decimal point:
cS
= two decimal digits
cO-
= one decimal digit (decimal point is jumping one digit to the right)
Press [ENTER/MENUE] to confirm. On the second li ne of the di spl ay appear s
Rh
_. Use the up and
down keys [
]to select the requested si gn:
Rh
_ = reference standard shows a negative measuring value (below 0.00C°)
Rh _
= reference standard shows a positive measuring value (above 0.00C°)
Press [ENTER/MENUE] to confirm. On the second li ne of the di spl ay appear s
Ec /
. Use the up and
down keys [
]to select the requested r ange:
Ec /
= below 100,00°C
Ec 0
= above 100,00°C
Press [ENTER/MENUE] to confirm. Now the display shows
//-//
. Please enter here the
temperature of your reference. To modify the value use the up k ey [
]. For stepping to the next
number use the down key [
].
If the requested value is complete then press [ENTER/MENUE]. Press [ESC] to get back to the measuring m ode.
Important!
A break of the physical calibration
mO1
cannot be done by the [ESC]-button. Switching off the
instrument can only do a break of the physi cal calibration.
Display-indi cat io n wit h act ive calibration code
mO1
:
The
B@K
– segment and the small 2 on the left indicates to the user
that
mO1
is activated.
2.4.3.3 DIN IEC-Calibrati on DIN IEC 60751 -
[
mEE
Use the up and down keys [
] to select [
mEE
. Pre ss [ENTER/MENUE] to confirm.
Press [ESC] to be back in the measuring mode.
DigiCal DCS 2
14 Handli ng / I ntroduc tion YAME0017/09.10.03
2.4.4 Printer port activation / deactivation -
OpDl
OpDl
= Printe r en a ble
Press [ENTER/MENUE] to activate or deactiv ate t he printer port. Use the up and down keys [
]
to select
OpDl
and press [ENTER/MENUE]. Use the up and down keys [ ] to activate ml or
deactivate
mEE
the requested adjustment and press [ENTER/MENUE] to confirm.
Press [ESC] to get back to the measuring m ode.
Display Keys Function
OpDl
ENTER/MENUE
OpDl
ml mEE
up and down keys [ ]
activati on or deactivation
ENTER/MENUE To keep the amended value ESC To escape without retention
DigiCal DCS 2
YAME0017/ 09.10.03 Handling / Introduction 15
2.4.5 Selection / Channel activation -
Bgl
To activate or deactivate a measuring channel press [ENTER/MENUE]. Use the up and down keys [
] to select
BglK
and press [ENTER/MENUE]. On the left corner of the displ ay appears a small 1, which indicates the selected channel . For changing the channel use the up and down keys [
]. Press [ENTER/MENUE] to confirm. The second display line shows ml for activation or
mEE
for deactivati on. Use the up and down keys [
] again to activate ml or deactivate
mEE
the
requested measuring channel and press [ENTER/MENUE] to con firm. Press [ESC] to get back to the measuring m ode.
In case that one c hannel would not be connected to a probe, the main display would indicate
mODl
in the relevant line.
There must be at l east one ac tive channel!
Display Keys Function
Bgl
ENTER/MENUE
1
Bgl
2
Bgl
up and down keys [
]
select the channel 1 or 2
1
BglK
ENTER/MENUE
1
Bgl
ml mEE
up and down keys [ ]
activati on or deactivation
ENTER/MENUE To keep the amended value ESC To escape without retention
DigiCal DCS 2
16 Handli ng / I ntroduc tion YAME0017/09.10.03
2.4.6 Analogue output scaling-
c@B
The DCS 2-Measuring instrument contains an analogue output (0-1Volt) for each channel. To get the optimal resolution and accuracy it is possible t o scale the analogue output by limiting the measuring range.
Press [ENTER/MENUE] to change the measuring range (default_ 0..100). Use the up and down keys [
] to select
c@B
. Pre ss [ENTER/MENUE] to confirm.
On the left corner of the di spl ay appear s a sm all 1, which indicates the selected channel. For changing the channel use the up and down keys [
] .
Press again [ENTER/MENUE] to confirm. On the display appears
c@0a
. To change between measuring r ange upper limit and measuring range lower limit use the up and down keys [
] .
c@0a
= Lower limit (defaul t 00.00)
c@0D
= Upper limit (default 100.00)
Press [ENTER/MENUE] to confirm. Use the up and down keys [ ] to select the number of digits after decimal point:
cO-
= two decimal digits
cO -
= one decimal digit (decimal point moves one digit to the right)
Press [ENTER/MENUE] to confirm. On the second li ne of the di spl ay appear s
Rh^0
. Use the up
and down keys [
] to select the requested range:
Rh^
= reference standard shows a negative measuring value (below 0,00C°)
Rh^0
= reference standard shows a positive measuring value (above 0,00C°)
Press [ENTER/MENUE] to confirm. On the second line of the display appears
Ec /
. Use the up
and down keys [
] to select the requested range:
Ec /
= below 100,00C°
Ec 0
= above 100,00C°
Press [ENTER/MENUE] to confirm. On the display appears
//-//
. For changing the value use
the up key [
] . For stepping to the next num ber use t he down key [ ]. If the requested value is
complete then press [ENTER/MENUE] to confirm. Press [ESC] to return to measuring mode. For changing the upper measuring range please repeat the above procedure.
The adjustable m easuring range limits are corresponding to the selected measuring size.
DCS 2-menu Scale Meas. range °C Analogue signal Lower limit dA1b 00.00 00,00°C 0,00Volt Upper limit dA1E 10.00 10,00°C 1,000Volt
DigiCal DCS 2
YAME0017/ 09.10.03 Handling / Introduction 17
2.5 Recalling the memory data (HOLD MAX MIN AVE)
After pressing fir st time the key [HOLD MAX MIN AVE] the actual value will be hold on the top display line ( big displ ay ) . pr essing again the key [HOLD MAX MIN AVE] the saved maximum-, minimum and average v alue will be displayed in the bottom display line (small display).
Note for 2-channel instruments: First the MAX-MIN-AVE-values of channel 1 will be displayed – after wards the values of the channel 2 will be display ed. Is there only one probe on a 2-channel instrument we suggest to deactiv ate t he c hannel without probe. (
þ
2.4.5 Selection / Channel activation - Chnl)
During the recall of the memory data the extremes (MAX MIN) and the average value (AVE) will not be calculated or carried on.
Clearing the memory (M AX MIN AVE)
Press [CLEAR] key once to erase the stored maximum, minimum and average from mem ory. On the display appears
Bp
.
After erasing the memory the instrument automatically switches back to measuring mode indicating the actual measured value again.
2.6 Measuring rate (FAST-Mode)
Press [FAST/ ] key one to change the measuring rate. Now the measuring rate is app. 4 measurement s per second. P ress [FAST/
] key again and the instrument is back in the standard
mode (1 measurement per second).
In the fast mode the battery c onsumption is three times higher than in the standard mode.
2.7 AUTO-OFF-Function
c@mE
= Disable Auto-off
D@mE
= Enable Auto-off
Press [ESC/AUTO-OFF] key again. On the display appears
c@mE
Now the Auto-Off-function is
deactivated. Press [ESC/AUTO-OFF] key once. On the display appears
D@mE
Now the instrument switches off
automatic ally aft er app. 30 mi nutes.
After switching off and on the instrument, the Aut o-Off function is automatically deactivated.
DigiCal DCS 2
18 Power supply / Error codes YAME0017/09.10.03
3. Power supply
For the power supply of the i nstrument a 9VDC source is required:
230V; 50/60Hz, Ref. No. EBE 042
90-260V; 50-60Hz, Ref. No. EBE 043
4. Error codes
By displaying t he following error codes the instrument support the operation of the i nstrument.
Dppmp
Description
mODl
no probe or wrong probe is connected
rmGh
too high exceeding of the m easuring range
rmKm
too low underflow of the measuring range
DigiCal DCS 2
YAME0017/ 09.10.03 Technical dat a 19
5. Technical data
Type: DigiCal DCS 2 Probe Ports 1 / 2
Pt100
Measuring range
-200...+450°C
Accuracy Pt 100 temperature probe with manufacturer c alibration
-200...–100°C: 0.05°C
-100...+150°C: 0.03°C +150...+200°C: 0. 05°C
>200°C: 0.1%
Resolution
-200 bis +200°C with 0.01°C resolution +200 bis +400°C with 0.1°C resolution
Outputs
RS232 Interfac e ( serial Printer can be connected)
2 analogue outputs 0-1 V (11- bit scalable)
Connectors
DIN 8-pole
Allowab le ope r a t i ng temp.
0°C ... +40°C
Display
2-line LCD
Housing
plastic (ABS)
Dimensions
200 x 80 x 200 mm (BxHxL)
Weight
950 g
Power supply
Power Supply 230V; 50/ 60Hz or optional 90-260V; 50/60Hz
DCS 2 Ref. No. LMD 018 (230V)
LMD 919 (90-260V)
DigiCal DCS 2
20 RS 232 interface protocol YAME0017/09. 10.03
6. RS232 interface protocol
During the data transmission, the availability via PC of the the tension levels for RTS (=+12V ) and DTR (= -12V) must be assured by a softwares regul ation code.
Parameter: Baudrate 2400 Baud Datenbits 8 Stopbits 2 Parität None
To start the data transmission you have to send the following charact er s. The following index shows you which value can be requested over the interface.
DCS 2 Measuring value 1 + 2 autom. recognize FC (hex) Enabling the keyboar d 0 (hex)
By sending the command FC(hex ) you will get the complete data, t he instr ument lock automatic ally the key board. To release the keyboard you have to send the comm and 0(hex).
The datas would be sent in the f ollowing format.
Datentyp = String
he length of the string depends on the instrument (a two-channel or one- c hannel instrument).
measuring valu e 1 unit measuring valu e 2 unit
LF CR
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
Example: Turbo Pascal Source-Code for initialisatio n o f th e RS232 interf ace
PROCEDURE V24oeffnen(com1,com2:Boolean); { initialisation of RS 232 } VAR wert : Byte;
BEGIN
IF com1 THEN BEGIN { initialisation of COM 1 } PORT [$3FB]:=128; { activate baud rate } PORT [$3F8]:= 48; { baud rate - LSB, 2400 baud } PORT [$3F9]:= 0; { baud rate - MSB } PORT [$3FB]:= 7; { 8 Bits, No Parity, 2 Stop } PORT [$3FC]:= 2; { DTR = 0, RTS = 1 } Wert := PORT [$3F8] { receiving register empty } END;
END
DigiCal DCS 2
YAME0017/ 09.10.03 Connector layout 21
7. Connector layout
7.1 Connector layout Pt100 4-wire
Measuring channel 1 / 2
7.2 RS232 PC-adapter cable (Connector layout)
Measuring channel 1 / 2
DigiCal DCS 2
22 Data of probe-calibration YAME0017/09.10.03
8. Data of probe-calibration
Our quality-probes are t he m ain reason for the precision and qual ity of your measuring. To facili tate the exchange of a probe without a losi ng precision, our probes were measured in our company and added with a code that describes the characteristic line of t he probe. The probe is labelled obvi ously with the relevant code.
If you change the probe you have to enter the code into the measuring instrument. . We recommend to write down both val ues i n order to avoid any problems (especially if the probe is positioned at a place difficult t o access, or if there are several external probes i n use):
Ref.-no: Appellation: Serial number: Calibrati on data:
Ref.-no: Appellation: Serial number: Calibrati on data:
Ref.-no: Appellation: Serial number: Calibrati on data:
Ref.-no: Appellation: Serial number: Calibrati on data:
Ref.-no: Appellation: Serial number: Calibrati on data:
DigiCal DCS 2
YAME0017/ 09.10.03 Accessories / Or dering spares 23
9. Accessories
Accessories Cat.-no.
Pt 100 sheathed resistance probe with 4-wire connection, c lass B 1/3 DIN,
-200 ... +450°C. The cali br ation values of this probe are encoded in the LAUDA calibration code. Probes with a 3mm diameter tubing made of stainless steel 300mm long, length of the sensor at the end of the tube < 35mm, plasti c handle made of ABS
ETP 065
Factory calibr ation certificate for the measurement chain with calibrati on points at 0° and 150°C
LMZ 052
DKD calibration c ertificate for the measurement c hain at 2 selec table calibration temperatures between –40° and +300°C
LMZ 053
PC adapter cable (RS232) with a sub-D-plug for the DCS 2 and a 9-pin sub-D­coupler for the PC
EKS 037
Windows PC Data-Recordi ng- S oftware: WinTherm Plus on CD-ROM LSDM2002 Plug In Mains Supply f or 90-260VAC, 50-60Hz, including plug adapt or s for Euro-plug,
USA, UK, Japan
EBE 043
10. Ordering spares
To avoid questions, del ay and delivery of incorrect it ems, pl ease advise always your instruments type and serial number stat ed on the manufacturers label.
The capital lett er of the seri al number (first position) shows the year of manufacture, e.g. B = 2003 We always would appreciate dealing with your queries and comments and receiving your suggestions,
experience and criticism.
LAUDA DR. R. WOBSER
GMBH & CO.KG
Postfach 1251
97912 Lauda-Königshofen
Tel: 09343/503-0
Fax: 09343/503-222
E-mail in fo @ la uda.de
Internet http://www.lauda.de
BESTÄTIGUNG / CONFIRMATION / CONFIRMATION
Formblatt / Form / Formulaire: Unbedenk.doc Erstellt / published / établi: LSC Änd.-Stand / config-level / Version: 0.1 Datum / date: 30.10.1998
LAUDA DR. R. WOBSER GmbH & Co. KG
Pfarrstraße 41/43 Tel: +49 (0)9343 / 503-0 D - 97922 Lauda-Königshofen Fax: +49 (0)9343 / 503-222 Internet: http://www.lauda.de E-mail: info@lauda.de
UNBEDENK.DOC
An
/
To / A
:
LAUDA Dr. R. Wobser LAUDA Service Center Fax: +49 (0) 9343 - 503-222
Von
/
From / De
:
Firma
/ C
ompany / Entreprise
:
Straße
/ S
treet / Rue
:
Ort
/
City / Ville
:
Tel
.:
Fax
:
Betreiber
/ R
esponsible person / Personne responsable
:
Hiermit bestätigen wir, daß nachfolgend aufgeführtes LAUDA-Gerät (Daten vom Typenschild):
We herewith confirm that the following LAUDA-equipment (see label): Par la présente nous confirmons que l’appareil LAUDA (voir plaque signalétique):
Typ
/
Type / Type
:
Serien-Nr.
/
Serial no. / No. de série:
mit folgendem Medium betrieben wurde
was used with the below mentioned media a été utilisé avec le liquide suivant
Darüber hinaus bestätigen wir, daß das oben aufgeführte Gerät sorgfältig gereinigt wurde, die Anschlüsse verschlossen sind, und sich weder giftige, aggressive, radioaktive noch andere gefährliche Medien in dem Gerät befinden.
Additionally we confirm that the above mentioned equipment has been cleaned, that all connectors are closed and that there are no poisonous, aggressive, radioactive or other dangerous media inside the equipment.
D’autre part, nous confirmons que l’appareil mentionné ci-dessus a été nettoyé correctement, que les tubulures sont fermées et qu’il n’y a aucun produit toxique, agressif, radioactif ou autre produit nocif ou dangeureux dans la cuve.
Stempel
Seal / Cachet.
Datum
Date / Date
Betreiber
Responsible person / Personne responsable
Loading...