LASCAR EL-USB-2-LCD User Manual

EL-USB-2-LCD Humidity, Temperature and Dew Point Data Logger with LCD Display
This standalone data logger measures and stores up to 16,379 relative humidity and 16,379 temperature readings over 0 to 100%RH and -35 to +80°C (-31 to +176°F) measurement ranges. The user can easily set up the logger and view downloaded data by plugging the module into a PC's USB port and using the supplied software. Relative humidity, temperature and dew point (the temperature at which water vapour present in the air begins to condense) data can then be graphed, printed and exported to other applications. The high contrast LCD can show a variety of temperature and humidity information. At the touch of a button, the user can cycle between the current temperature and humidity, along with the maximum and minimum stored values for temperature and humidity. The data logger is supplied complete with a long-life lithium battery, which can typically allow logging for up to 1 year.
FEATURES
0 to +100%RH measurement range
-35 to +80°C (-31 to +176°F) measurement range
Dew point indication via Windows control software
USB interface for set-up and data download
User-programmable alarm thresholds for %RH & T
Status indication via red and green LEDs
High contrast LCD, with two and a half digit
temperature and humidity display function
Immediate, delayed and push-to-start logging
Supplied complete with replaceable internal lithium
battery and Windows control software
WINDOWS CONTROL SOFTWARE
Easy to install and use, the control software runs under Windows 2000, XP (Home and Professional Editions) and Vista (32-bit). It allows the user to set up and download data from any EL-USB product. The latest version of the control software may be downloaded from www.lascarelectronics.com
DISPLAY FUNCTIONS
Six different functions are available, which can be cycled through using the built-in push button. The most recent logged value, maximum logged value and minimum logged value can be displayed separately for humidity and temperature. In addition, logging and alarm status is shown using two high intensity LEDs.
DATA LOGGER SETUP VARIABLES ORDERING INFORMATION
Logger name
°C, °F
Logging rate (10s, 1m, 5m, 30m, 1hr, 6hr, 12hr)
High and low alarms
Immediate, delayed and push-to-start logging
Display off, on for 30 seconds after button press , or
Standard Data Logger EL-USB-2-LCD
(Data Logger, Software on CD and Battery)
Replacement Battery BAT 3V6
permanently on
Data rollover (Allows unlimited logging periods by
overwriting the oldest data when the memory is full)
LASCAR ELECTRONICS LTD. LASCAR ELECTRONICS INC. LASCAR ELECTRONICS (HK) LIMITED MODULE HOUSE 4258 West 12th Street Unit Nos. 6-8 , 19/F., Futura Plaza WHITEPARISH Erie 111-113 How Ming Street WILTSHIRE SP5 2SJ PA 16505 Kwun Tong, Kowloon UK USA Hong Kong TEL: +44 (1794) 884567 TEL: +1 (814) 835 0621 TEL: +852 2389 6502 FAX: +44 (1794) 884616 FAX: +1 (814) 838 8141 FAX: +852 2389 6535 E-mail: sales@lascar.co.uk E-mail: us-sales@lascarelectronics.com E-mail: saleshk@lascar.com.hk
Stock Number
Specifications liable to change without prior warning EL-USB-2-LCD Issue 1c 06/2008 N.J-H Applies to EL-USB-2-LCD
www.lascarelectronics.com Page 1 of 5
EL-USB-2-LCD Humidity, Temperature and Dew Point Data Logger with LCD Display
SPECIFICATIONS
Specification Min. Typ. Max. Unit
Relative Humidity Measurement range 0 100 %RH
Repeatability (short term) ±0.1 %RH
Accuracy (overall error) ±3.0* ±6.0 %RH
Internal resolution 0.5 %RH
Long term stability 0.5 %RH/Yr
Temperature Measurement range -35 (-31) +80 (176) °C (°F)
Repeatability ±0.1 (0.2) °C (°F)
Accuracy (overall error) ±0.5 (1) ±2 (±4) °C (°F)
Internal resolution 0.5 (1) °C (°F)
Dew Point Accuracy (overall error) ±1.1 (±2)**
Logging rate every 10s every 12hr -
Operating temperature range † -35 (-31) +80 (176) °C (°F)
1/2AA 3.6V Lithium Battery Life*** 1 Year
* This specifies the overall error in the logged readings for relative humidity measurements between 20 and 80%RH. ** This specifies the overall error in the calculated dew point for relative humidity measurements between 40 and 100%RH at 25°C. *** Depending on sample rate, ambient temperature and use of LCD display. † At temperatures below -20°C (-4°F) the LCD will exhibit slower response times of approximately 10 seconds.
°C (°F)
LED AND LCD INDICATORS
The EL-USB-2-LCD features a high contrast LCD and two bi-colour LEDs. The LCD shows logged temperature and humidity values using seven segment numbers, along with annunciators. The LCD can also show information regarding the logging status.
Specifications liable to change without prior warning EL-USB-2-LCD Issue 1c 06/2008 N.J-H Applies to EL-USB-2-LCD