LASCAR DPM942 User Manual

DPM 942 4-20mA Loop Powered Indicator
A loop powered indicator giving an accurate, zero corrected indication of 4-20mA loop current. Calibration is by two 20-turn cermet potentiometers which allow sensitive adjustment of the instrument. The meter is housed in a robust carrier, which can be bolted in place or panelmounted using thelow profilebezel and clipsprovided.
19mm (0.75“) Digit Height
Programmable Decimal Points
Low Volt Drop
Simple Screw Terminal Connections
CUSTOM SCALING
For alternative scalingto the standard 0 - 1000, cutLink 8 and short Link 7. Add two resistors R12 and R13 which will set the basic scale of the meter. The highest practical value for the R12 and R13 series combination is 12R. To select your desiredscaling, use thefollowing formula: R12 + R13= Full Scale readingat 20mA
160
Re-calibrate as necessary.
R25R25 R14R14
R13R13 R12R12 R11R11
To p right hand corner of the pcb.Top right hand corner of the pcb.
Standard Meter DPM 942
Specification Min. Typ. Max. Unit
Accuracy (overall error) * 0.05 0.1 %(±1 count) Linearity ±1 count Sample rate 3 samples/sec Operating temperature range 0 50 °C Temperature s 100 ppm/°C
tability
Loop Volt Drop 3.9 4.3 4.8 V
*To ensure maximum accuracy, re-calibrate periodically.
CONNECTOR SOURCING GUIDE
METHOD Screw Terminals - No Connector Required
Stock Number
DIMENSIONS All dimensions in mm (inches)
76.0 (2.99)
72.5 (2.85)
44.0 (1.73)
40.5 (1.59)
51.0 (2.01)
61.5 (2.42)
51.0 (2.01)
24.0 (0.94)
39.0 (1.54)
71.5 (2.81)
24.0 (0.94)
Ø5.7
(0.22)
24.5 (0.96)
Ø3.0
(0.12)
a
a. 1.0 (0.04) b. 2.0 (0.08) c. 6.5 (0.26) d. 11.5 (0.45) e. 13.0 (0.51)
Panel cut-out 72 x 40 (2.83 x 1.57) Maximum panel thickness 3mm (0.12).
b
c
d
e
OFFSET SPAN
6 10
DS
REAR VIEW
1
578
DP2
DP3
DP1
ON BOARD SOLDER LINKS
2
3 4
9
I-
I+
LASCAR ELECTRONICS LTD.
MODULE HOUSE, WHITEPARISH,
TEL: +44 (0)1794 884567 FAX: +44 (0)1794 884616
WILTSHIRE SP5 2SJ UK
E-mail: sales@lascar.co.uk
LASCAR ELECTRONICS INC.
3750 West 26th Street,
TEL: +1 (814) 835 0621 FAX: +1 (814) 838 8141
Erie, PA 16506 USA
E-mail: us-sales@lascarelectronics.com
LASCAR ELECTRONICS (HK) LIMITED
FLAT C, 5/FL., LUCKY FTY. bldg., 63-65 HUNG TO ROAD
KWUN TONG, KOWLOON, HONG KONG
TEL: +852 2797 3219 FAX: +852 2343 6187
E-mail: b4lascar@samsongroup.com.hk
TERMINAL FUNCTIONS
1. I+ Positive currentinput.
2. I- Negative current input.
3. DP3 1.000
4. DP2 10.00 Decimal point inputs.
5. DP1 100.0
6. DS Decimal Pointselect. Connectto required DP inputto displaydecimal point.
CALIBRATION
The meter is suppliedcalibrated toread000 for 4mAloop currentand 1000 for 20mA. To re-calibrate:
1. Apply4mA to I+/I-and adjust'OFFSET' toread000.
2. Apply20mA and adjust'SPAN'to read 1000.
3. Repeatsteps 1 and2 untilthere isnomore adjustment of 'SPAN' and'OFFSET' requiredtogive desired readings.
SAFETY
To comply with the Low Voltage Directive (LVD 93/68/EEC), input voltages to the module’s pins must not exceed 60Vdc. If voltages to the measuring inputs do exceed 60Vdc, then fit scaling resistors externally to the module. The user must ensure that the incorporation of the DPM into the user’s equipment conforms tothe relevant sectionsof BS EN61010 (Safety Requirementsfor Electrical Equipmentfor Measuring, Controland Laboratory Use).
VARIOUS OPERATING MODES
ON-BOARD LINKS: In order to quickly and easily change operating modes for different applications, the meter has several "on-board links". They are designed to be easily opened (cut) or shorted (soldered). The DPM 942 is powered from the 4-20mA signal loop and needs no other power supply. Ensure correct polarity when connecting.
1
4-20mA
Transmitter
I+
I-
I+
2
I-
DP2
DPM 942
DS
PANELFITTING
Normally SHORTED Cut to OPEN
4
Normally OPEN Solder to SHORT
DP2 selected in this example.
6
Specifications liable to change without prior warning DPM 942 Issue 5 August/2002 M.C. Applies to DPM 942/3