LASCAR DPM200 User Manual

DPM 200 3½ Digit LCD Module
The DPM 200 uses advanced components and construction techniques to provide an unrivalled combination of high performance, elegant appearance andlow cost. The module usesa bandgapreference for extratemperature stability. Forsingle railuse, the DPM200S features a builtin negativerail generator,enabling the meter tomeasure a signal referencedto itsown power supply0V.
15mm (0.6“) Digit Height
Auto-zero
Auto-polarity
200mV d.c. Full Scale Reading (F.S.R.)
LOW BATT
°C °F
µMVA
pH
KgM
%
Single Rail Version
Autoranging Outputs
SCALING
Two resistorsRa andRb maybe fittedin orderto alterthe full scale reading(F.S.R.) ofthe meter - seetable. The meter willneed re-calibration.
Required F.S.R. Ra Rb
2V Note 910k 100k 20V Note 1M 10k 200V Note 1M 1k 2kV Note 1M 100R 200 A 0R 1k
µ
2mA 0R 100R 20mA 0R 10R 200mA 0R 1R
NOTE
Ensure that Link10 is open iffitting Ra.
Standard Meter DPM 200 Single Rail Version DPM 200S
Specification Min. Typ. Max. Unit
Accuracy (overall error) * 0.05 0.1 % (±1 count) Linearity ±1 count Sample rate 3 samples/sec Operating temperature range 0 50 °C Temperature s 30 ppm/°C Supply DPM 200 7.5 9 14
voltage (V+ to V-) DPM 200S 3.5 5 6.5 Supply current DPM 200 150
Input leakage reading (Vin = 0V) 1 10 pA
* To ensure maximum accuracy, re-calibrate periodically.
CONNECTOR SOURCING GUIDE
tability
1
7A
DPM 200S 500
Stock Number
V
Aµ
METHOD
Solder Solder wires or standard 0.1" square pin header to PCB
SAFETY
Edge Connector LASCAR EC 24 DS
To comply with the Low Voltage Directive (LVD 93/68/EEC), input voltages to the module’s pins must not exceed 60Vdc. If voltages to the measuring inputs do exceed 60Vdc, then fit scaling resistors externally to the module. The user must ensure that the incorporation of the DPM into the user ’s equipment conforms tothe relevantsections ofBS EN61010 (Safety Requirementsfor Electrical Equipmentfor Measuring, Controland Laboratory Use).
13.0
DIMENSIONS All dimensions in mm (inches)
72.0 (2.83)
2.00
(0.08)
(0.51)
51.5 (2.03)
36.0 (1.42)
46.75 (1.84)
45.0 (1.77)
18.5 (0.73)
Display in TEST mode
Panel cut out 68 x 33 (2.68 x 1.30)
2.5 (0.09)
2.8 (0.11)
H
41
67.5 (2.66)
62.5 (2.46)
40
g
µ
V
%
ph
KEY
TEST
XB3
XDP
LMP+
LMP-
A
22
k
A
m
M
°F
°C
REAR
VIEW
CLK
XG3
-5V
V+
REF+
REF BG
REF-
2.54
(0.10)
E3
DP1
DP2
BPABPOL
~
DP3
REF HI
Ø2.5
23
(0.1)
:
-
BP
27.5 (1.08)
V-
REF LO
COM
IN LO
IN HI
1
7.06
(0.28)
6.10
(0.24)
A
41
912
6
13
10
Ra
Rb
1
ON BOARD SOLDER LINKS
SCALING RESISTORS
14
REAR VIEW
11427
15 3
8
22
PANELFITTING
Locate the meter by passing it through the front of the panel cut-out and, using the four screws provided, fix the plastic spring clips to the rear of the meter.
LASCAR ELECTRONICS LTD.
MODULE HOUSE, WHITEPARISH,
TEL: +44 (0)1794 884567 FAX: +44 (0)1794 884616
WILTSHIRE SP5 2SJ UK
E-mail: sales@lascar.co.uk
LASCAR ELECTRONICS INC.
3750 West 26th Street,
TEL: +1 (814) 835 0621 FAX: +1 (814) 838 8141
Erie, PA 16506 USA
E-mail: us-sales@lascarelectronics.com
LASCAR ELECTRONICS (HK) LIMITED
FLAT C, 5/FL., LUCKY FTY. bldg., 63-65 HUNG TO ROAD
KWUN TONG, KOWLOON, HONG KONG
TEL: +852 2797 3219 FAX: +852 2343 6187
E-mail: b4lascar@samsongroup.com.hk
PIN FUNCTIONS
1. V- Negative power supplyconnection.
2. IN HI Positivemeasuring input. Analogue inputs mustbe nocloser than1V toeither the positive ornegativesupply. The negative
3. IN LO Negativemeasuring input. supply of theDPM 200S is generatedinternally and mirrors the positive supplyvoltage.
4. COM Groundforthe analogue sectionof theA/D converter. It isactively heldat approximately 2.8V belowV+ andmust notbe allowed
5. REF LO Negativeinput for referencevoltage.
to sink excessive current(>100 A) by, for instance,connecting toa highervoltage.
6. REF HI Positiveinput forreference voltage(connected via Link 1toREF+).
7. POL Drivefor negative polarity (-)annunciator. Internally connectedvia Link11 to the annunciator.
8, 10, 15,19 Outputs for auto-ranging applications.
9. BP LCD backplane drivewaveform.
11. REF+ Positive outputfrom internalreference.
12. REF- Negative output frominternalreference.
13. REF BG Output from bandgapreference (1.22V).
14. V+ Positivepower supplyconnection.
16. CLOCK Clock output maybe usedfor systems timing orasan input tooverride the internaloscillator andcontrol the sample rate.
17. -5V Output from negative railgenerator circuit. This is aninversion ofV+ (DPM200S only).
18. XDP Connectto required annunciators/DPs (seenote). This isan inversionof BP.
20. TEST Connect toV+ todisplay segments asillustrated. It should notbe operated for morethana few secondsas theDC Voltage applied to theLCD may'burn'the display. This pinis normallyat 5V belowV+ andis theground for the digitalsection ofthe meter. It canbe usedto powerexternal logic up toamaximum of 1mA.
21. LMP- N.C.
22. LMP+ N.C.
A. KEY Polarisingslot for edgeconnector.
41. H N.C.
SPECIAL NOTE - ANNUNCIATORS.
The DPM annunciators(DPs, °C, etc.)can bedisplayed,either by connectingto XDP or by applying asolder link to theXDPpads located underthe annunciatorinput pads. However,as these annunciatorsare normally'floating', under certainconditions they may appearwhen notwanted. To suppressunwanted annunciators,apply a solder link to thebackplane pads located in-betweentheannunciator input pads. If the annunciatorsare beingswitched, connectthem viaa 1M resistor to the BP (Pin 9), theannunciators willthen operatenormally when connectedto XDP. Ensure that anannunciator isnot connecteddirectly tothe XDP and BPat the same time. Note - ifsuppressing (-)annunciator bydirect connectionto BP, Link11 MUST be cutfirst.
µ
ANNUNCIATOR
XDP
RIGHT
ANNUNCIATOR
BP
XDP
WRONG
ANNUNCIATOR
XDP
ANNUNCIATOR
BP
Link either BP or XDP
to annunciator.
Do not link BP & XDP
to the annunciator
at the same time
XDP
VARIOUS OPERATING MODES
ON-BOARD LINKS: In order to quickly and easily change operating modes for different applications the meter has several on-board links. They are designed to be easily opened (cut) or shorted (soldered). Do not connect more than one meter to the same power supply if the meters cannot use the same signal ground.
Normally SHORTED Cut to OPEN
Taking any input beyond the supply rails will damage the meter.
V+
14
V+
2
IN HI
±200mV
+
3
IN LO
-
V-
1
V-
Check Links 2,3&4areSHORTED.
Measuring a floating voltage source of 200mV full scale
V+
14
V+
2
IN HI
R
3
IN LO
0.2
R=
FSR
I
Check Links 2,3&4areSHORTED.
V-
1
V-
Measuring current. Supply MUST
be isolated.
±200mV
0V
Check Links 3 & 4 are SHORTED.
Split supply operation (DPM 200).
Check Link 1 is OPEN.
Measuring the ratio of two voltages. Reading = 1000 V /V
50mV< V <200mV V <2V .
12
+
-
1V min.
V
V2
2
V+
14
V+
2
3
IN HI
IN LO
14V max 7V min
V-
1
V-
±200mV
Check Links 3 & 4 are SHORTED.
2
IN HI
+
3
-
IN LO
Measuring a single ended input referenced to supply (DPM 200S).
V+
14
V+
2
+
IN HI
1
3
-
IN LO
6
+
REF HI
5
REF LO
-
V-
1
V-
U/RANGE
O/RANGE
V+
TEST (GND)
Generating basic Auto-ranging outputs.
12
Specifications liable to change without prior warning DPM 200 Issue 5 August/2002 M.C. Applies to DPM 200/7
14
V+
V-
1
Normally OPEN Solder to SHORT
V+
V-
8
AB
9
BP
10
E3
15
XG3
19
XB3