Larson Davis SoundTrack LxT, SoundExpert LxT User Manual

LxT Manual
for
SoundTrack LxT
TM
®
& SoundExpert LxT
Larson Davis
for
SoundTrack LxT
& SoundExpert LxT
I770.01 Rev K Supporting Firmware Version 2.301
Copyright
Copyright 2016 by PCB Piezotronics, Inc. This manual is copyrighted, with all rights reserved. The
manual may not be copied in whole or in part for any use without prior written consent of PCB Piezotronics, Inc.
Trademarks
PCB
, SoundTrack LxT and SoundExpert are registered trademarks of PCB Piezotronics, Inc.
Swithcraft
is a registered trademark of Switchcraft, Inc. All other trademarks are property of their
respective owners.
Disclaimer
The following paragraph does not apply in any state or country where such statements are not
agreeable with local law:
Even though PCB Piezotronics, Inc. has reviewed its documentation, PCB Piezotronics, Inc. makes no
warranty or representation, either expressed or implied, with respect to this instrument and documentation, its quality, performance, merchantability, or fitness for a particular purpose. This documentation is subject to change without notice, and should not be construed as a commitment or representation by PCB Piezotronics, Inc.
This publication may contain inaccuracies or typographical errors. PCB Piezotronics, Inc. will
periodically update the material for inclusion in new editions. Changes and improvements to the information described in this manual may be made at any time.
Record of Serial Number and Purchase Date
LxT Model: LxT1 LxT2 Serial Number: ___________ Preamplifier Model: ________ Serial Number: ___________ Microphone Model: _________ Serial Number: ___________
Recycling
PCB Piezotronics, Inc. is an environmentally friendly organization and encourages our customers to
be environmentally conscious. When this product reaches its end of life, please recycle the product through a local recycling center or return the product to:
PCB Piezotronics, Inc. Attn: Recycling Coordinator 1681 West 820 North Provo, Utah, USA 84601-1341
where it will be accepted for disposal.
Warranty
For warranty information, refer to our Terms and Conditions of Sale on our website at
www.larsondavis.com/TermsConditions.aspx.
Table of Contents
Chapter 1 LxT Features 1-1
SoundTrack LxT ................................................................................................. 1-1
SoundExpert
Chapter 2 Overview 2-1
LxT Components ................................................................................................... 2-1
Summary of Displays and Icons ........................................................................... 2-6
Navigating and Selecting ...................................................................................... 2-8
Basic Run Functions ............................................................................................. 2-9
Tab and Setting Displays ...................................................................................... 2-9
Chapter 3 Preparing for First Use 3-1
Unpacking and Inspection ..................................................................................... 3-1
Connecting the Microphone and Preamplifier ...................................................... 3-2
Connecting the Preamplifier to the LxT ............................................................... 3-3
Disconnecting the Preamplifier from the LxT ...................................................... 3-4
Powering the SoundTrack LxT ............................................................................. 3-4
Chapter 4 Basic Measurement Setup 4-1
Measurement Settings Tabs .................................................................................. 4-1
LxT ............................................................................................... 1-5
Chapter 5 Data Display 5-1
Data Labels ........................................................................................................... 5-1
Tabbed Display ..................................................................................................... 5-1
Live Tab ................................................................................................................ 5-3
Overall Tab ........................................................................................................... 5-8
Session Log Tab .................................................................................................. 5-20
View Spectrum Normalized (Optional) .............................................................. 5-21
Chapter 6 Making Measurements 6-1
Preparation ............................................................................................................ 6-1
Positioning the LxT ............................................................................................... 6-1
Performing Measurements .................................................................................... 6-2
Storing Measurements ........................................................................................... 6-7
Data Storage After Improper Shutdown ............................................................... 6-8
Chapter 7 Calibration 7-1
Calibration Overview ............................................................................................ 7-1
Control Panel - Calibrate .......................................................................................7-2
Acoustic Calibration ..............................................................................................7-4
Sensitivity Tab .....................................................................................................7-11
Calibration Without Preamplifier ........................................................................7-13
Certification .........................................................................................................7-14
Chapter 8 Voice Recording 8-1
Launching the Voice Recorder Dialog ..................................................................8-1
Making a Voice Recording ....................................................................................8-2
Playing a Voice Recording ....................................................................................8-3
Storing Voice Recordings ......................................................................................8-3
Chapter 9 Time History 9-1
Parameters Logged ................................................................................................9-1
Time History Setup ................................................................................................9-3
Time History Tab ...................................................................................................9-5
Link to Measurement History Display ..................................................................9-8
Markers ..................................................................................................................9-9
Chapter 10 Measurement History 10-1
Enabling Measurement History ...........................................................................10-1
Continuous and Timer Modes .............................................................................10-2
Timed Stop Mode ................................................................................................10-3
Manual and Stop When Stable Modes .................................................................10-3
Display of Measurement History .........................................................................10-3
Link to Time History .........................................................................................10-12
Chapter 11 Data Explorer 11-1
Control Panel - Data Explorer .............................................................................11-1
Chapter 12 System Properties 12-1
Control Panel - System Properties .......................................................................12-1
Device ..................................................................................................................12-2
Time .....................................................................................................................12-3
Power ...................................................................................................................12-4
Preferences ...........................................................................................................12-8
Localization .......................................................................................................12-15
Displays .............................................................................................................12-17
Options ...............................................................................................................12-21
Chapter 13 Lock/Unlock the LxT 13-1
Control Panel - Lock ........................................................................................... 13-1
Fully Locked ....................................................................................................... 13-4
Locked With Auto-Store ..................................................................................... 13-5
Locked With Manual-Store ................................................................................. 13-7
Calibration When The LxT Is Locked ................................................................ 13-7
Chapter 14 About 14-1
Control Panel - About ......................................................................................... 14-1
About Tab ........................................................................................................... 14-1
Standards ............................................................................................................. 14-2
Options ................................................................................................................ 14-3
User ..................................................................................................................... 14-3
Chapter 15 System Utilities 15-1
System Utilities ................................................................................................... 15-1
File System .......................................................................................................... 15-2
Chapter 16 Parameters Measured 16-1
Basic Sound Level Measurements ...................................................................... 16-1
Sound Exposure Metrics Measured .................................................................... 16-3
Statistical Metrics Measured ............................................................................... 16-4
Community Noise Parameters ............................................................................ 16-4
Exceedance Counters .......................................................................................... 16-5
Miscellaneous Parameters ................................................................................... 16-5
Chapter 17 Memory Utilization 17-1
Out Of Memory Stop .......................................................................................... 17-1
Overall Data ........................................................................................................ 17-1
Session Log ......................................................................................................... 17-1
Measurement History .......................................................................................... 17-1
Time History ....................................................................................................... 17-2
Voice Messages ................................................................................................... 17-2
Chapter 18 Upgrading Firmware and Options 18-1
SLM Utility-G3 ................................................................................................... 18-1
Upgrading LxT Firmware ................................................................................... 18-1
Upgrading LxT Options ...................................................................................... 18-4
Appendix A Technical Specifications A-1
Standards Met by LxT ..........................................................................................A-1
LxT Specifications ................................................................................................ A-2
Electromagnetic Emission ..................................................................................A-15
1/1 and 1/3 Octave Filters ...................................................................................A-16
Position of Instrument and Operator ...................................................................A-19
Frequency Response ...........................................................................................A-20
Microphone Preamplifier Specifications ............................................................ A-43
Vibration Sensitivity ...........................................................................................A-59
Appendix B Testing to IEC61672-1 1
Sections 5, 6, 7 and 9 (except 9.3) ........................................................................ B-1
LxT1 ..................................................................................................................... B-8
LxT2 ................................................................................................................... B-13
Appendix C Glossary and Formulas C-1
CHAPTER

LxT Features

1

SoundTrack LxT

Hardware Features
This chapter describes the features for the SoundTrack
LxT
and SoundExpert LxTsound level meters.
The following sections describe the features for the SoundTrack LxT.
The Larson Davis SoundTrack LxT has the following features:
Precision integrating sound level meter
2 GB unformatted standard data memory
160 X 240 pixel LCD display with backlight and icon­driven user interface
Silent Touch elastomeric keypad
Large dynamic range
Jack for AC/DC output or headset (ACC003)
Preamplifier can drive a 61 m (200 ft.) microphone extension cable (EXC200)
4-AA batteries provide 16 hour operating time
USB 2.0 peripheral connector
Field-upgradeable firmware
Windscreen (WS001)
Basic Measurements
SPL, Leq, Lmax, Lmin, SEL, Lpeak, Lpeak(max)
RMS Detectors: Slow, Fast & Impulse
RMS Frequency Weighting: A, C & Z
Peak Frequency Weighting: A, C & Z
•L
statistics: 6 user-selected values over the range (L
N
through L
2 Sets of hygiene metrics: Lavg, TWA(x), Dose, ProjDose, Lep,d
LxT Manual LxT Features 1-1
) and Histogram tables
99.99
0.01
Basic Operation
•E, E8, E40
SEA peak exposure
2 RMS event counters and 3 Peak event counters
Auto-Store with Auto-Reset
Run Timer and Stop-When-Stable Control
Real-time clock
Start time, elapsed time and paused time
Time stamping for Lmax, Lmin, Lpeak(max) metrics
Session Log
Lock functions
Calibration with calibration history and list of calibrators
Power management
Status bar and About display
Multiple language support
Data files and Data Explorer
Automatic data backup to prevent data loss on power failure
Overall measurement
Available Options
LXT-OB1: Real-time 1/1 Octave Frequency Analysis
LXT-OB3: Real-time 1/3 & 1/1 Octave Frequency Analysis
LXT-LOG: Automatic data logging with intervals from 1 second to 24 hours
LXT-HSLOG: Extends data dogging (LXT-LOG) with intervals down to 100 milliseconds
LXT-ENV: Measurement History Environmental Data Logging
LXT-CN: Community Noise
1-2 SoundTrack LxT LxT Manual
Standard Accessories
LXT-DVA: Digital Voice Annotation (includes headset ACC003)
The LxT is delivered with the standard accessories described below.
One of the following preamplifier/microphone combinations:
- PRMLXT1 preamplifier with a 377B02 micro­phone
- PRMLXT1L preamplifier with a 377B02 micro­phone
- PRMLXT2B preamplifier with a 375B02 micro­phone
- PRMLXT2L preamplifier with a 375B02 micro­phone.
SWW-SLM-UTILG4 SLM Utility-G4 software
SWW-SLM-UTILG3 SLM Utility-G3 software
WS001 Windscreen, 3 1/2 in. diameter
Alkaline Batteries: 4-AA
•Lanyard
Optional Accessories
Equivalent Electrical Impedance Adaptor
An equivalent electrical impedance adapter can be used in place of the microphone when a measurement is being made electrically. The adapter is simply a series capacitor with the same capacitance as the microphone it is replacing. The following adapters are available:
ADP002 6.8pF adaptor for 1/4 in., 7pF microphone (377C01 or 377C10)
ADP005 18pF adaptor for 1/2 in., 18pF microphone (375B02)
ADP090 12pF adaptor for 1/2 in., 12 pF microphone (377B02)
LxT Manual SoundTrack LxT 1-3
Cables
Power Supplies
Software
Accessory Kits
Direct Input Cable or Adaptor
EXCXXX Microphone extension cable, where XXX is the length in feet (XXX = 010, 020, 050, 066, 100 and 200 available)
CBL138 USB Cable
CBL139 AC/DC Output Cable
PSA029 Universal AC Power Adaptor
PSA031 12 Volt DC to USB Power Adaptor
BAT015 External battery powering device for the LxT, holding 4 or 8 D-sized alkaline 1.5 volt batteries to extend run time
SWW-BLAZE-LXT Blaze
SWW-DNA DNA
LXT-ACC including
- LXT-CCS Hard Shell Case
- CAL200 Class I Calibrator
- PSA029 Power Supply
- CBL138 USB Cable
LXT-ACC1 including
- LXT-CCS Hard Shell Case
- CAL150 Class I Calibrator
- PSA029 Power Supply
- CBL138 USB Cable
Other
377C20 1/2” random incidence pre-polarized microphone, 50 mV/Pa, providing performance conforming to Class 1 sound level meter standards
ACC003 Headset with microphone for voice recording/ Playback (included LXT-DVA)
1-4 SoundTrack LxT LxT Manual
LXT-CCS Storage Case
Environmental Protection
EPS2116 Environmental Shroud
EPS2106-2 Environmental Shroud
EPS2108-2 Environmental Shroud
EPS030-LXT Environmental Case with one lead acid battery to be used with an external microphone tripod (the tripod is not included)
Tripods
TRP001 Instrument/Camera Tripod with ADP032 1/2 in. microphone clip. Use with EPS2108-2
TRP002 Microphone Stand with adjustable height and boom angle
TRP003 Support Tripod, heavy duty, can be used with EPS030 or EPS2106-2
Calibrators
CAL150
CAL200
Printer
PRN003 USB Serial Printer
SoundExpert LxT
The SoundExpert LxT sound level meter provides targeted measurement of environmental noise, and is sold in two base models:
LXT1-SE-FF: SoundExpert LxT with free-field microphone (377B02)
• LXT1-SE-RI: SoundExpert LxT with random microphone (377C20).

LxT Manual SoundExpertâ LxT 1-5

Hardware Features
Standard Features
Standard Accessories
The hardware features for the SoundExpert LxT are the
same as those for the SoundTrack LxT. following accessories are included with the SoundExpert LxT:
PSA029 Power Supply
CBL138 USB Cable
The SoundExpert LxT is delivered with the standard features listed below.
LXT-LOG: Automatic data logging with intervals from
1 second to 24 hours
LXT-ENV: Measurement History Environmental Data
Logging
LXT-CN: Community Noise
LXT-OB3: Real-time 1/1 & 1/3 Octave Band Analysis
The SoundExpert LxT is delivered with the standard accessories described below.
CBL138 USB Cable
PSA029 Universal AC Power Adaptor
One of the following preamplifier/microphone
combinations:
Additionally, the
- PRMLXT1 preamplifier with a 377B02 micro­phone
- PRMLXT1L preamplifier with a 377B02 micro­phone
- PRMLXT1 preamplifier with a 377C20 micro­phone
- PRMLXT2L preamplifier with a 375C20 micro­phone.
SWW-SLM-UTILG4 SLM Utility-G4 software
SWW-SLM-UTILG3 SLM Utility-G3 software
WS001 Windscreen, 3 1/2 in. diameter
Alkaline Batteries: 4-AA
1-6 SoundExpertâ LxT LxT Manual
•Lanyard
Optional Accessories
Equivalent Electrical Impedance Adaptor
An equivalent electrical impedance adapter can be used in place of the microphone when a measurement is being made electrically. The adapter is simply a series capacitor with the same capacitance as the microphone it is replacing. The following adapters are available:
ADP090 12pF adaptor for 1/2 in., 12 pF microphone (377B02)
Cables
Direct Input Cable or Adaptor
EXCXXX Microphone extension cable, where XXX is the length in feet (XXX = 010, 020, 050, 066, 100 and 200 available)
CBL139 AC/DC Output Cable
Power Supplies
PSA031 12 Volt DC to USB Power Adaptor
BAT015 External battery powering device for the SoundExpert volt batteries to extend run time
LxT, holding 4 or 8 D-sized alkaline 1.5
Software
SWW-BLAZE-LXT Blaze
SWW-DNA +SWW-DNA-LXT
Other
377C20 1/2” random incidence pre-polarized microphone, 50 mV/Pa, providing performance conforming to Class 1 sound level meter standards
LXT-CCS Storage Case
Environmental Protection
EPS042 LxT1-SE-XX +BAT015 in small hard-shell carrying case
LxT Manual SoundExpertâ LxT 1-7
EPS2106/8-2 Environmental Shroud for outdoor microphone protection
EPS030-LXT Environmental Case with one lead acid battery to be used with an external microphone tripod (the tripod is not included)
Tripods
TRP001 Instrument/Camera Tripod with ADP032 1/2 inch microphone clip. Use with EPS2108-2
Calibrators
CAL200
•CER-LXT1 LxT calibration with report
•CER-MIC Microphone calibration
1-8 SoundExpertâ LxT LxT Manual
CHAPTER
Microphone
Preamplifier
LCD Display
Keypad
Connectors
2

LxT Components

Overview

This chapter provides an overview of the SoundTrack
LxT
sound level meter, including the following sections:
LxT Components
Summary of Displays and Icons
Navigating and Selecting
Basic Run Functions
Tab and Setting Displays
FIGURE 2-1 The LxT
LxT Manual Overview 2-1
The standard LxT shown in FIGURE 2-1 includes the
Hardware Power Switch
USB Interface
AC/DC Output and Headset Jack
Auxiliary Connector
following:
1/2 in. diameter condenser microphone
Backlit graphic 160 x 240 pixel LCD display
13-key soft rubber backlit keypad
AC/DC output, control, USB, and external power connectors (shown in FIGURE 2-2)
True “hand held” instrument with “sure grip” pads
DO NOT use the hardware power switch to turn the LxT OFF. This may cause data to be lost. Press the 0 (ON/OFF) key, then the Off soft key to turn the LxT off.
2-2 LxT Components LxT Manual
FIGURE 2-2 LxT Bottom View
Hardware Power Switch: When set to “O”, the hardware power switch completely powers down the LxT for storage. Set the switch to “|” for instrument operation.
USB Interface: The USB 2.0 full-speed peripheral port is used to control LxTs from PCs and transfer data to PCs using a CBL138 or other USB cables under 5 m in length. The LxT can also be powered via USB interfaces using PSA029 external power supplies.
Display
Measurement
Name
OBA Overload Icon
Input Overload Icon
Run Time
Measurement
Status
Power Indicator
Ta bs
Position
Indicator
Scroll Bar
Stability
Indicator
Left
Softkey
Center
Softkey
Right
Softkey
AC/DC Output and Headset Jack: This jack is used to output analog AC and DC signals or to connect to headsets for recording and playback of voice records
Auxiliary USB Connector: The auxiliary USB connector allows attaching USB storage devices.
The LxT has a 160 x 240 graphic, liquid crystal display that is backlit to provide comfortable viewing in most ambient light situations. Controls are provided for contrast and backlight adjustments.
When the LxT is first turned on, a display similar to FIGURE 2-3 is shown.
LxT Manual LxT Components 2-3
FIGURE 2-3 Data Display Screen
Keypad
Left
Softkey
Center
Softkey
Right
Softkey
RUN/PAUSE
STOP/STORE
UP
RIGHT
DOWN
LEFT
ENTER
RESET
TOOLS
POWER
Softkeys
Hardkeys
The LxT has a 13 button keypad. This section describes the buttons on the keypad.
The three buttons just beneath the display, on the body of the LxT, are called Softkeys, as shown in Figure 2-3. Above each Softkey, on the bottom of the display, is an icon or label indicating the action that takes place when the key is pressed. Softkeys are so named because the action associated with the key can change.
The ten remaining keys below the Softkeys are shown in FIGURE 2-4 and are described in TABLE 2-1.
2-4 LxT Components LxT Manual
FIGURE 2-4 LxT Keys
.
Use the Power, or ON/OFF button to turn the LxT on and off. The hardware power wwitch on the base of the unit must be in the “|” position.
Use the Navigation buttons Up, Down, Left and Right to move to areas on the display, to make selections from multiple options, or to enter alphanumeric characters into data fields.
Use the Enter button to select data, options, or displays or to enter alphanu­meric characters into data fields.
Use the Run/Pause button to initiate and pause measurements, and to continue paused measurements.
Use the Stop/Store button to stop measurements and to store measurements when measurements are stopped.
Use the Reset button to reset measurements.
Use the Tools button to specify settings such as date and time, managing power options and setting personal preferences (i.e. language, decimal and date for­mats, etc.).
TABLE 2-1 Keypad Hardkeys
LxT Manual LxT Components 2-5

Summary of Displays and Icons

Tabs
Data on the LxT is presented in a tabbed format. Move between tabs by using the right and left Softkeys.
Pages
Tabs are divided into pages that logically group the data together (i.e., 1/3 Octave data on the Live tab). Navigate up or down to different pages by using the 8 (Up) and 2 (Down) keys.
Scroll Bar and Position Indicator
The scroll bar represents the entire tab, and the position indicator shows the relative position of the page you are viewing. The position indicator in FIGURE 2-3 shows that the first page on the Live tab is being viewed.
Power Indicator
The icon indicates whether the LxT is being powered by batteries (battery level is also indicated), or by an external power source.
Measurement Filename
The name of the data file, or the measurement filename, is configurable as described on page 4.
Stability Indicator
For certain measurement modes and for calibration, an indication of the stability of the measured signal is presented by the following icon.
Run Time
This is the amount of time the measurement has been running.
Input Overload Icon
When signals from the preamplifier exceed the calibrated input range of the LxT, the Input Overload icon appears.
2-6 Summary of Displays and Icons LxT Manual
Under Range Icon
OBA Overload Icon
While the overload is present, the icon flashes. When the overload is removed, the icon disappears from the display.
If a measurement is running and an overload occurs, the icon shown below flashes during overloads.
When the overload has been removed, the icon is still present (not flashing) to indicate that overloads have occurred during the measurement. Resets clears the icon from the display.
When signals from the preamplifier drop below levels that can be accurately measured, an under range condition exists. When this happens the Under Range icon appears.
As long as the under range condition exists, the icon flashes. When the measured level no longer produces an under range condition, the icon is removed from the display.
When a measured level is in an under range condition, its displayed level appears in gray rather than black.
If inputs to the Octave Band Analyzer (optional firmware LXT-OBA required) become overloaded, the icon shown below appears to indicate overloads.
This icon operates similar to the Input Overload Icon shown in the above section “Input Overload Icon.”
Measurement Status
Reset Icon
The Reset icon indicates that a measurement is in a “reset” state.
Run Pending Icon
The Run Pending icon appears when the PAUSE) key is pressed and the LxT is waiting for filters and detector initialization to complete. The LxT automatically
LxT Manual Summary of Displays and Icons 2-7
9 (RUN/
Run Icon
Pause Icon
Stop Icon
Store Icon
starts the run after the initialization has completed (less than 10 seconds).
The Run icon is animated, moving from left to right to indicate that a measurements is in progress.
The Pause icon indicates that the current measurement has been paused.
The Stop icon is displayed when a measurement has been stopped.
The Store icon indicates that the current measurement has been stored.

Navigating and Selecting

To navigate between tabs on the display, press the right or left Softkeys. To navigate within tabs, use the 4 and 6 keys for moving horizontally on screens. This includes moving the highlight from one property to the next.
The 8 and 2 keys are used for moving vertically on screens. This includes moving the highlight from one property to the next and to move to previous or subsequent tab pages.
These keys are also used for character entry by navigating through lists of characters in text boxes.
The
5 key is typically used for completing selections,
completing actions, or accepting values.
2-8 Navigating and Selecting LxT Manual

Basic Run Functions

The basic measurement run functions are as follows:
Running
•Pausing
Stopping
•Storing
The
9 (RUN/PAUSE) key initiates a run. If a
measurement is running, this key pauses the run. It does not end the run; to end the measurement run, press the
(STOP/STORE) key. Pressing the 9 (RUN/PAUSE) key
when the unit is PAUSED continues the run. This key is only active on a Data View screen.
7
Pressing the when the unit is in STOP mode continues the previous run.
9
(RUN/ PAUSE) key
The 7 (STOP/STORE) key ends a run. Pressing the key a
second time stores the data in a file. This key is only active on a Data View screen.

Tab and Setting Displays

The LxT features and functions are organized into four different types of displays.
Data Display tabs: used to display measured data.
Measurement Settings tabs: used to set the parameters
Control Panel (Tools) Properties: used to set user
Power Control Page: used to check battery power,
Data Display Tabs
When the the Data Display tabs appear.
for a measurement.
preferences, to set non-measurement related parameters, and to implement calibration.
control the contrast and backlight of the display and other features.
0 (ON/OFF) key is pressed to turn on the LxT,
LxT Manual Basic Run Functions 2-9
Measurement Settings Tabs
Opening
From the Data Display tabs, pressing the Center Softkey labeled Menu brings up the menu shown in FIGURE 2-5.
FIGURE 2-5 Menu
Control Panel (Tools) Properties
Power Control Page
Data Display Tabs
For a more detailed description of the Data Display tabs and their associated pages, see Chapter 5 in the section entitled “Basic Data Display” on page 5-1.
Select Settings and press
Closing
Press the Center Softkey to return the Data Display tabs.
The Control Panel is accessed by pressing the key at the lower right of the LxT front panel. To exit from the Control Panel and return to the Data Display tabs, press the Center Softkey labeled Close.
The Power Control Page is opened by pressing the / OFF) key while on Data View tabs. To exit from the Power Control Page, press the Center Softkey labeled Close.
The Data Display tabs include the following: “<$paratext>”
Live: Data is continuously displayed on this tab whether there is a measurement in progress or not.
Overall: The data displayed on this tab represents data measured and averaged beginning from the time the measurement was started by pressing the Run key until the elapsed time indicated above the display. If the Pause or Stop key is pressed, the elapsed time is stopped. However, pressing the Run key continues the overall measurement, as shown by the elapsed time restarting from the time when it had previously been paused or stopped.
5 to open the Settings tabs.
3 (TOOLS)
0 (ON
2-10 Tab and Setting Displays LxT Manual
Measurement Settings Tabs
Session Log: The Session Log is a record of data accumulation actions. A time-stamped record is made for every Run, Pause, Stop or Voice Message action.
Current (optional) used in conjunction with Measurement History. Similar to the Overall tab except that data is based on the most recent run instead of the first run of the measurement.
Measurement History (optional): This tab displays current data measurement times or stops using the Time History measurement feature.
Time History (optional): This tab displays data measured using the Time History measurement feature.
The screen is not wide enough to show all thirteen setup tabs at the same time. Use the Right and Left Softkeys to navigate between tabs and bring them within view.
For a more detailed description of the Measurement Settings tabs and their associated pages, see Chapter 4 in the section entitled "Basic Measurement Setup" on page 4-1.
The Measurement Settings tabs allow for specific settings and include the following:
General: used to create a file name and a measurement description.
SLM: used to setup the parameters for the measurement of sound levels.
OBA (optional): used to setup the real-time octave band frequency analysis.
Dosimeter 1: used to setup the parameters for the measurement of sound exposure and noise dose.
Dosimeter 2: used to setup the parameters for the measurement of sound exposure and noise dose.
Ln: used to define the parameters for the measurement of Ln statistics.
Control: used to setup the mode of measurement timing.
Time History (optional): Permits the automatic logging of a specified number of parameters as a function of time.
Triggers: used to setup the triggers which define noise exceedance events.
LxT Manual Tab and Setting Displays 2-11
Markers (optional): Use in conjunction with time
Day/Night (optional): Defines hours for day, night, and
Control Panel (Tools) Properties
The Control Panel uses icons to represent the different functions available. Pressing the the Control Panel icons.
The position indicator on the scroll bar indicates that there are additional icons not currently visible on the screen.
history measurements, this feature permits the user to annotate portions of a time history record to identify noise sources or make other notes.
evening periods for 24-hour noise monitoring.
3 (TOOLS) key displays
FIGURE 2-6 Control Panel
For more information, see “Control Panel - System Properties” on page 11-1.
2-12 Tab and Setting Displays LxT Manual
To select an icon, navigate to the desired icon and press The functions for icons on the Control Panel are described in subsequent chapters.
5.
Data Explorer
For a detailed description of Data Explorer, see Chapter 12 in the section entitled “Data Explorer” on page 12-1.
For a detailed description of System Properties, see Chapter 11 in the section entitled “System Properties” on page 11-1.
For a detailed description of the Lock feature, see Chapter 13 in the section entitled “Lock/Unlock the LxT” on page 13-1.
The Data Explorer is used to examine data that has been stored following previous measurements. It is also used to manage stored measurements, such as rename or delete files.
System Properties
System Properties tabs are used for general instrument bookkeeping. Functions such as setting the instrument date and time, display contrast adjustment, date format, etc. are located here. These are single page tabs.
The System Property tabs: are as follows:
Device: Enter instrument identification.
Time: Set the date and time.
Power: Set controls that affect power consumption.
Preferences: Set a variety of system parameters such as microphone correction, auto-store, jack function, reset prompting, takt maximal and USB port.
Localization: Set regional characteristics such as language, decimal symbol, data format and units.
Lock
Lock permits the LxT to be configured such that certain keys are locked to prevent unauthorized use or tampering.
Calibrate
For a detailed description of the calibration procedure, see Chapter
Calibrate is used to verify and adjust the calibration of the LxT prior to a measurement.
15 in the section entitled “Calibration” on page 15-1.
For a detailed description of the voice recording feature, see Chapter 8 in the section entitled “Voice
Voice Recorder
A method to allow voice annotation of the data is described in Chapter 8 in the section entitled “Voice Recording” on page 8-1.
Recording” on page 8-1.
LxT Manual Tab and Setting Displays 2-13
About
For a detailed description of the About tabs, see Chapter 14 in the section entitled “About” on page 14-
1.
The About tabs provide the user with information specific to this instrument, such as serial number, options, etc.
The About tabs include the following:
About: shows information such as serial number and firmware revision.
Standards: lists the standards that the LxT meets.
Options: shows the options that are available in this instrument.
User: allows user entered instrument identification.
2-14 Tab and Setting Displays LxT Manual
CHAPTER

Preparing for First Use

3
This chapter outlines the steps to unpack the SoundTrack LxT
Unpacking and Inspection
Connecting the Microphone and Preamplifier
Disconnecting the Preamplifier
Powering the SoundTrack LxT

Unpacking and Inspection

Retain the packaging for safe shipment for calibration service.
Your LxT has been shipped in protective packaging. Please verify that the package contains the items listed below. Report any damage or shortage immediately to PCB Piezotronics, Inc. at 888 258-3222 (U.S. toll free) or (716) 926-8243.
•LxT
PRMLxT Microphone Preamplifier
Microphone
•Lanyard
WS001 3 1/2” Windscreen
4 - AA Alkaline Batteries
SLM Utility-G3 software
SLM Utility-G4 software
and prepare it for first use, including:
Record Serial Numbers of LxT and Components
If you have not already done so, please record the purchase date, model and serial number for your instrument, preamplifier and microphone in the spaces provided at the beginning of this manual. You find the instrument model and serial numbers printed on the label on the back panel of the instrument.
LxT Manual Preparing for First Use 3-1
The microphone model and serial numbers are engraved on the outside of the microphone, as shown in FIGURE 3-1. The preamplifier model and serial numbers are engraved on the preamplifier body as shown in FIGURE 3-2.
FIGURE 3-1 Microphone
You may be asked to provide this information during any future communications with PCB
Piezotronics, Inc.

Connecting the Microphone and Preamplifier

Caution: Take care when handling the preamplifier, as the gold pin is sensitive to electrostatic discharge (ESD).
Carefully place the bottom end of the microphone over the top end of the preamplifier and gently screw the assembly together. The microphone body seats smoothly against the preamplifier body. DO NOT use excessive force.
When removing the microphone, turn while gripping the microphone body, not the grid cap.
FIGURE 3-2 Microphone-Preamplifier
3-2 Connecting the Microphone and Preamplifier LxT Manual

Connecting the Preamplifier to the LxT

The connectors are keyed for correct alignment.
Caution: Do not attempt to unscrew the collar/ring at the top of the LxT body.
Insert the preamplifier into the mating connector on the LxT and rotate the preamplifier until the keyways line up. Press the assemblies together until a small click is heard.
If the LxT is powered when the preamplifier is inserted, a message similar to the one in FIGURE 3-3 appears for several seconds.
FIGURE 3-3 Preamplifier Connected
Press
5 (ENTER) to close the message.
LxT Manual Connecting the Preamplifier to the LxT 3-3

Disconnecting the Preamplifier from the LxT

Preamplifier Release Button
When transporting the LxT, it is recommended that the preamplifier be detached and placed in a secure location in the carrying case.
FIGURE 3-4 Preamplifier Release Button
On the front of the LxT, just below the preamplifier connector, is a small button. Press and hold this button while pulling the microphone/preamplifier assembly out of the LxT, as shown in Figure 3-4.
Powering the SoundTrack LxT
Important Notice Regarding Proper Shutdown
3-4 Disconnecting the Preamplifier from the LxT LxT Manual
The following sections provide power information for the LxT, including the following:
Important Notice Regarding Proper Shutdown
Batteries
Hardware Power Switch
•USB Power
External Power Supply PSA029
•Power Settings
Improperly turning off the power to the LxT may damage the instrument. To properly shut off power, use the on/off button on the front of the meter.
Batteries
If the LxT is being powered externally via a USB cable, do not unplug the cable without ensuring that the batteries in the instrument have adequate charge or properly powering down the LxT first.
The LxT should also be properly shut off prior to changing batteries.
Do not use 3.8 V Lithium batteries; they will blow the fuse.
Battery Status
The LxT is compatible with AA alkaline, nickel-metal hydride (NiMH) batteries and 1.5 volt Lithium batteries.
NOTICE:
• NiMH batteries cannot be charged in the LxT. Do not mix alkaline and NiMH batteries in the LxT.
• Do not mix batteries from different manufacturers.
• Replace all four batteries when installing fresh cells.
• NiMH batteries may not be used in areas requiring Intrinsic
Safety Approval.
Battery voltage and estimated run time are displayed on the Power Control page and the last page of the Live tab. When the LxT is powered by batteries, one of the icons shown in FIGURE 3-5 is shown on the status bar at the top of the screen. The icon shows the state of the battery charge as a full battery when the batteries are fresh, decaying to an empty battery near the end of the battery life. The battery voltage and the battery icon directly reflect the remaining estimated run time as displayed by the instrument.
FIGURE 3-5 Battery Status Icons
When the battery voltage becomes critically low, the empty battery icon begins to flash, indicating that the LxT is about to shut down. When the LxT shuts down, it stops running, saving all data and the instrument state, and then turns off. When the unit is powered on again, either with fresh batteries or an external power supply, the unit returns to the state it was in when it shut down.
LxT Manual Powering the SoundTrack LxTâ 3-5
Hardware Power Switch
DO NOT use the hardware power switch to turn the LxT OFF. This may cause data to be lost. Press the key, then the Off soft key to turn the LxT off.
It is recommended that the batteries be removed from the instrument if it will not be used for a month or longer as the batteries may self­discharge and leak, damaging the instrument.
0
USB Power
The LxT cannot be operated under USB power if the internal batteries are discharged (flat). You can run solely on USB power if you remove the depleted batteries. However, if operated on external power only, with no batteries installed, an interruption of power to the LxT, for any reason, may result in instrument malfunction.
The Hardware Power Switch on the bottom of the LxT, shown in Figure 2-2 disconnects the batteries from the LxT hardware, including the real time clock. This prevents battery drain when the LxT is not in use for an extended period of time ( in the “O” position, the batteries are disconnected.
After installing batteries be sure to move the switch to the "|" position. This applies power to all of the LxT hardware.
The Hardware Power Switch should not be used to turn the LxT on and off. If the Hardware Power Switch is used to turn the LxT off, data may be lost and Flash corruption may result.
The LxT can be powered from batteries or, if available, from the USB host portion of your computer.
The LxT must run on batteries until allowed by the host to run on USB or external power. If the batteries cannot provide sufficient power, the LxT does not power on, even with USB external power. If batteries are installed in the LxT, ensure that they are good so that the LxT can power on.
If the LxT has discharged batteries installed, the batteries should be removed or replaced with fresh batteries in order for the LxT to be USB powered.
To avoid memory corruption when using USB Power or flash drives, follow these precautions:
Always shut down the LxT completely before unplugging USB power connections.
Do not unplug USB drives from the USB port on the LxT while the drive is being copied, or if the LxT is within Data Explorer mode.
> 2 weeks)If the Hardware Power Switch is
External Power Supply PSA029
In addition to running on batteries, or USB power, the LxT can be powered from a PSA029 power supply. When
3-6 Powering the SoundTrack LxTâ LxT Manual
external power is being supplied, the Battery icon is replaced with the icon shown in FIGURE 3-6.
FIGURE 3-6 External Power Icon
The LxT cannot be operated with an external power supply if the internal batteries are discharged (flat). You can run solely on external power if you remove the depleted batteries. However, if operated on external power only, with no batteries installed, an interruption of power to the LxT, for any reason, may result in instrument malfunction.
Power Settings
The PSA029 is designed to work on power systems worldwide.
The LxT must run on batteries until allowed by the host to run on external power. If the batteries cannot provide sufficient power, the LxT does not power on, even with external power. If batteries are installed in the LxT, ensure that they are good so that the LxT can power on.
If the LxT is ON, pressing the 0 (ON/OFF) key brings up
the Power Control Page, as shown in FIGURE 3-7.
FIGURE 3-7 Power Control Page
The first section of this page shows the estimated battery run time (calculated using the voltage of the installed batteries), battery voltage, and the USB power voltage.
LxT Manual Powering the SoundTrack LxTâ 3-7
The Backlight can also be adjusted from the Power Page as described in the section "Power" on page 11-4.
The backlight mode and display contrast are adjusted using the 4, 6, 8 and 2 keys. There are three options for Backlight: Off, Dim, and Bright, which are adjusted using the 4 and 6 keys. The Display Contrast has a range of -9 to 9, which is adjusted using the 8 and 2 keys.
The bottom of the Power Control page displays the LxT temperature that is used to automatically adjust the contrast of the display to compensate for temperature changes.
The Center Softkey provides an exit, escape or cancel function as well as the function displayed above it on the display.
Pressing the Center Softkey, labeled Close, closes the Power Control page.
3-8 Powering the SoundTrack LxTâ LxT Manual
CHAPTER

Basic Measurement Setup

4
This chapter describes how to setup the LxT to perform basic sound level measurements, including the following:
Leq, Lmax, Lmin corresponding to user-selected values of frequency weighting and detector
Lpeak and Lpeak(max) corresponding to a user­selected value of frequency weighting
1/1 and/or 1/3 Octave real-time spectra (LxT-OB3 required)
Six values of Ln based on six user-selected values of the parameter n
Count of the number of times the levels (RMS and Peak) exceeded user-selected threshold values
Sound exposure and sound exposure level data
The LxT can measure many additional sound parameters simultaneously with these basic sound measurements, as described in subsequent chapters.

Measurement Settings Tabs

Accessing the Measurement Settings Tabs
The parameters defining measurements are set from the Measurement Settings tabs.
To access these tabs, press the Center Softkey labeled
Menu, press the 2
5 (ENTER) key. The Measurement Settings tab most
appropriate for the data now appears.
key to select Settings, and press the
LxT Manual Basic Measurement Setup 4-1
Settings In Use Message
If the LxT is not already connected to a computer running Blaze, SLM Utility-G3, or G4 software, ignore this section.
If the Blaze, SLM Utility-G3, or G4 software is already connected to the LxT when an attempt is made to access the Measurement Settings Screen, the display shown in FIGURE 4-1 "Settings In Use By PC Message" appears.
FIGURE 4-1 Settings In Use By PC Message
This message indicates that setup changes made with SLM Utility-G3 or G4 software in this session will be lost if you continue.
To continue and access the Measurement Settings tabs, highlight Ye s and press the Measurement Settings Screen, highlight No and press
5. To cancel the attempt to access
5.
4-2 Measurement Settings Tabs LxT Manual
General Tab
Figure 4-2 shows the General tab.
FIGURE 4-2 General Tab
The SLM Utility-G3 or G4 software can be used to easily enter both the file name and the measurement description.
The General Tab is used to enter file names and measurement descriptions for the measurements being defined. Upon opening, the Default File Name “LxT_Data” may appear in the file name field.
To enter new file names, select the Default File Name text box and press new names and press
5. Use the 4, 6. 8 and 2 keys to enter
5.
LxT Manual Measurement Settings Tabs 4-3
SLM Tab
Frequency Weighting
Integration Method
Figure 4-2 shows the SLM tab.
FIGURE 4-3 SLM Tab
To modify settings on the SLM tab, navigate to either Frequency Weighting, Detector, Peak Weighting, or Integration Method page and press
A, C and Z frequency weightings are provided for the RMS and peak detectors. These are selected separately.
Two Integration methods are available: Linear and Exponential.
5.
Exponential Integration
Exponential integration would typically be selected to provide compatibility with older analog instruments in which measurements are exponential time weighted signals. Exponential detectors tend to hide small events in the long decay of loud impulsive events.
Linear Integration
Linear integration utilizes sampled sound pressure levels to compute RMS levels directly, without an intermediate time weighting.
4-4 Measurement Settings Tabs LxT Manual
Octave Band Analyzer Tab (Optional)
Left Click to Select
The default values for these parameters are as shown in FIGURE 4-4.
OBA Parameter Selection
This tab only appears when the LxT has the optional LxT­OB1 or LxT-OB3 firmware enabled.
FIGURE 4-4 OBA Tab
The OBA parameters are selected as shown in FIGURE 4-5.
LxT Manual Measurement Settings Tabs 4-5
FIGURE 4-5 OBA Parameter Selection
OBA Range Setting
In the Low range, the full scale level is reduced by 30 dB on the display. The default display ranges are as follows:
Normal Range: 20 to 140 dB
•Low Range: -10 to 110 dB
Graph scaling range can be modified by the user, as described in “Graph Scale Adjustment” on page 5-2.
OBA Frequency Weighting
The user can select that the 1/1 and/or 1/3 Octave frequency analysis modules process data from the A, C or Z weighting filters.
OBA Max Spectrum Setting
Two methods can be used to define the maximum spectrum:
At Lmax: using this method, the maximum values for each frequency band are those that are being measured at the instant the overall sound pressure level reached its maximum value during the measurement period.
•Bin Max: using this method, the level measured for each frequency band is the maximum measured during the measurement period. Since the maximum levels for the different frequency bands may have occurred at different times, the ensemble of frequency band maximum levels may represent a spectrum that never existed at any single instant during the measurement.
4-6 Measurement Settings Tabs LxT Manual
Dosimeter 1 and 2 Tabs
Predefined Setups
Figure 4-6 shows the Dosimeter 1 tab.
FIGURE 4-6 Dosimeter 1
The Dosimeter 1 and Dosimeter 2 tabs are provided to permit the evaluation of two independent noise dose data sets. Other than being on separate tabs, they are identical.
In most cases, measurements of this type are setup to conform to specific standards. The LxT permits the user to create such setups in a single step by simply selecting the applicable standard. The standards addressed by the LxT and the corresponding parameters are as shown in Table 4-1 "Predefined Noise Dosimeter Setups".
Standard
Exchange
Rate
Threshold Criterion
Level Hours
OSHA-1 5 90 90 8
OSHA-2 5 80 90 8
ACGIH 3 80 85 8
NIOSH 3 80 85 8
IEC 3 Not Enabled 85 8
Table 4-1 Predefined Noise Dosimeter Setups
LxT Manual Measurement Settings Tabs 4-7
The Name field is already highlighted when the Dosimeter View is opened. If this has been changed, use the the highlight back to the Name field.
8
key to move
The names of the predefined setups can be accessed from the Name field at the top of the display. Press of predefined setups by name as shown in FIGURE 4­7 "Predefined Dose Setups"
.
Parameters Individually Defined
The Dosimeter tab contains two fields (Name and Exchange Rate) and two sections (Threshold and Criterion).
Name Field
If a predefined setup has been selected and any of the preset parameters are modified, the user should consider changing the name of the setup.
Threshold and Criterion
When setting the Threshold value, the Enabled check box must be checked before data can be entered into the numeric field. Use the Left Softkey to highlight the box and press
5.
Auto-Calculate
The Name field is optional, although many users enter text associated with the measurement to be performed, such as specific company Dose standards (“My Dose”) or names of standards not in the list.
To set these parameters, after highlighting the appropriate section, press the numeric value desired, then press
The Criterion Level and Time are set independently. However, Criterion Level and the Time have a linear relationship, so when Auto-Calculate is activated for either, both are automatically set to follow the applicable standard.
5 to show a list
FIGURE 4-7 Predefined Dose Setups
5, use the 4, 6, 8, and 2 keys to enter
5.
4-8 Measurement Settings Tabs LxT Manual
Ln Tab
Figure 4-8 shows the Ln tab.
FIGURE 4-8 Ln Tab
The Ln value is the measured sound level that exceeds n% of the measurement time. For example, a value of L
= 35 dB
90
means that the measured sound level is above 35 dB for 90% of the measurement period. These statistical values are commonly used to describe the characteristics of non-steady sound such as environmental noise.
The LxT can calculate six different Ln values based on user­defined values of n, which can be in the range 00.01 to
99.99%.
LxT Manual Measurement Settings Tabs 4-9
Control Tab
The Control tab, shown in Figure 4-9, is used to set the Run Mode for the measurement to be performed.
FIGURE 4-9 Control Tab
When the optional firmware LXT­ENV has been enabled, Measurement History can be enabled in the Run Control Setup.
The Run Mode on the Control tab can be setup for one of six modes, including the following.
Manual Stop: The measurement is initiated manually by pressing the 9 (Run/Pause) key and is stopped by pressing the 7 (Stop) key.
•Timed Stop: The measurement is initiated manually by pressing the 9 key and is stopped automatically after a user-defined time period.
Stop When Stable: The measurement is initiated manually by pressing the 9 key. The measurement stops when the measured level has remained within a user-defined range and the measurement has run for a user-defined time period.
Continuous: The measurement is initiated manually by pressing the 9 key and is stopped by pressing the 7 key. Measurements are made continuously from start to stop.
Single Block Timer: The measurement is initiated manually by pressing the 9 key and is stopped by pressing the 7 key. Measurements are made only during the time interval defined by the single block timer.
4-10 Measurement Settings Tabs LxT Manual
Daily Timer: The measurement is initiated manually by
pressing the 9 key and is stopped by pressing the 7 key. Measurements are made only during the separate time blocks defined by the setup.
When the Control tab is opened, the Run Mode field is already selected. Press shown in Figure 4-10.
FIGURE 4-10 Run Mode Menu
5 to open the Run Mode Menu, as
Navigate to select the desired mode and press the selection. The screen then appears as one of those shown in Figure 4-11, depending upon the mode selected. For the Timed Stop and Stop When Stable modes, further information must be entered.
Manual, Timed Stop, or Stop When Stable Modes
When the run mode used is Manual, Timed Stop or Stop When Stable, a single measurement runs continuously from
start time to end time. The selection and setting of parameters for these run modes is shown in FIGURE 4-11.
5 to make
FIGURE 4-11 Setup of Manual Stop, Timed Stop, and Stop When Stable Run Modes
LxT Manual Measurement Settings Tabs 4-11
Entering Run Time for the Timed Stop Mode
To enter the Run Time, navigate to select the Time data field and press digit positions in the data field using the 4 and 6 keys, as shown in Figure 4-12.
FIGURE 4-12 Entering Time in Timed Stop
Stop When Stable Mode
The Stop When Stable run mode contains the data fields Delta and Time, as shown in Figure 4-13.
FIGURE 4-13 Stop When Stable Run Mode
5. You can then move left and right to different
Run Mode
Delta Level
The Delta level is the maximum one minute change in overall average level (i.e. L
) allowed for the
Aeq
measurement to be considered stable.
Time
The Time is the duration that the measurement must run before the measurement can stop.
If the time were set to 0, the measurement would run until the stability condition was met.
Once duration is set, the measurement runs for the duration specified and then continues until the stability condition is met.
Run Modes Without Measurement History
All Run Modes include check box options to enable the Measurement History feature. The descriptions presented in
4-12 Measurement Settings Tabs LxT Manual
the following sections are for setups without Measurement History enabled.
When Measurement History is not enabled, the measurement must be manually stored at the conclusion of the duration. Because only a single measurement exists, the data displayed on the Overall and Current tabs is identical.
Continuous
The Continuous run mode is similar to the Manual Stop mode, except that Daily Auto-Store can also be enabled, in which case daily measurement reports for 24-hour time periods are automatically stored. The user specifies a beginning time for such periods. This also presumes that the measurement time period encompasses at least one 24-hour time period as programmed. The parameter Time defines the start time for the 24-hour time period to be used for the report.
Single Block Time or Daily Timer
When the run mode is Single Block Timer or Daily Timer, the single measurement consists of data measured over different blocks of time between the start date and time and the end date and time.
LxT Manual Measurement Settings Tabs 4-13
The selection and setting of appropriate parameters for the
Click to open data fields for blocks 2 and
3.
Continuous, Single Block Timer and Daily Timer run modes is shown in FIGURE 4-14.
FIGURE 4-14 Setup of Continuous, Daily Timer and Single Block Time Run Modes
4-14 Measurement Settings Tabs LxT Manual
For each run mode option, select the corresponding data field and press the
5. To enter the desired values, press the
5 again.
For the Continuous mode, when the Auto-Store checking the box in the Daily tab results in the following: daily measurement reports for 24-hour time periods are automatically stored, beginning at a user-specified time (assuming that the measurement time period encompasses at least one 24-hour time period as programmed).
When the option is checked, a data field opens to define the start time for the 24-hour time period to be used for the report, as shown in FIGURE 4-15.
FIGURE 4-15 Auto-Store Report Start Time
Select the data field and use the navigation keys to specify the start time.
Run Mode with Measurement History
Measurement History requires that the optional firmware LXT-ENV be enabled.
Triggers Tab
Note that the default values for these parameters are as shown in FIGURE 4-16.
To measure and store sequences of measurements using the same setup, either manually or automatically, you can use the Measurement History feature, which is described in detail in "Measurement History".
FIGURE 4-16 shows the Triggers tab.
FIGURE 4-16 Triggers Tab
The Triggers tab is used to define trigger levels that can than be used to detect when the measured sound level (SPL or Peak) exceeds one of these trigger levels.
Day/Night
The Day/Night parameters are used in conjunction with Community Noise Measurements that require the optional firmware LXT-CN to be enabled.
The default values for these parameters are as shown in FIGURE 4-17.
LxT Manual Measurement Settings Tabs 4-15
Among the parameters measured and displayed as part of a basic sound level measurement are the community noise descriptors L
DN
and L
. The Day/Night tab defines the
DEN
times and penalties to be used.
FIGURE 4-17 shows the Day/Night tab.
Ldn10Log
10
1
24
------
10
Li10+10
0000
0700
10
Li10
0700
2200
10
Li10+10
2200
2400
++



=
DEN
L =
 
  
 
 
10
10
10
5
10
10*810*410*12
24
1
lg10
LnightLeveningLday
FIGURE 4-17 Day/Evening/Night Definition
L
DN
The day-night level LDN defined by the following formula:
L
DEN
The day-evening-night level L following formula:
In the default form, the day has twelve hours, the evening has four hours and the night has eight hours, as can be seen in the equation. The default times for these periods are as follows:
4-16 Measurement Settings Tabs LxT Manual
is defined by the
DEN
CNEL
CNEL 10log
10
1
24
------
10
Li10+10
0000
0700
10
Li10
0700
1900
10
Li5+10
1900
2200
10
Li10+10
2200
2400
++ +



=
Day: 07.00 to 19.00
Evening: 19.00 to 23.00
Night: 23.00 to 07.00
Lday, Levening and Lnight are A-weighted long-term average sound levels measured during the day, evening and night, respectively.
To account for the increased impact of environmental noise during the evening and night, penalties are added to the measured level; 5 dB for evening and 10 dB for night, as can be seen in the equation.
The Directive 2002/49/EC of the European Parliament and of the Council of 25 June 2002, relating to the assessment of environmental noise permits member states to shorten the evening period by one or two hours and lengthen the day and/or the night accordingly and also to choose the time for the start of the day.
To accommodate these and other possible modifications, the Day/Night tab permits the user to modify the times for the beginning of the Day, Evening and Night periods and the penalties to be utilized when calculating 24-hour integrated values.
In the state of California, a commonly used community noise descriptor is Community Noise Equivalent Level (CNEL), defined by the following formula:
This is essentially the same as the L values, with the exception that the evening period begins at
22.00 instead of 23.00. Thus, by making this change in the
LxT Manual Measurement Settings Tabs 4-17
L
DEN
settings, the measured value represents CNEL.
using default
DEN
4-18 Measurement Settings Tabs LxT Manual
CHAPTER

Data Display

5

Data Labels

Tabbed Display

This chapter describes how data is displayed for basic sound level measurements.
The LxT sound metrics labels are designated by international standards. For many displayed values, the frequency and time weighting are indicated in the name of the metric. Example: L level measured using the Slow detector. Sound pressure level is often referred to as SPL.
Measured data are displayed using a number of tabs arranged horizontally across the screen, as shown in Figure 5-1. Depending on the firmware options loaded in the LxT, multiple tabs appear in the Data Display, each identified by a title at the top.
is the A-weighted sound pressure
AS
FIGURE 5-1 Tabbed Display
Navigating through Tabs
Not all of the tabs can be seen at one time on the display. To see tabs on the right, use the Right Softkey beneath the display. To navigate to the left, use the Left Softkey beneath the display.
Navigating within Tabs
Each tab contains multiple pages. To navigate between pages, use the 8 or 2 keys.
The 4 and 6 keys can be used move the cursor left and right, respectively, to select a specific filter or record, depending on the data being displayed.
LxT Manual Data Display 5-1
Selecting Options for Settings within Tabs
By using the 4 and 6 keys or the 8 or 2 keys to move between options on tabs or pages, you can select options by highlighting them. Once the desired value or option is selected, press
On some tabs, you can also select check box options by highlighting them and pressing
5 (ENTER).
Specifying Values for Settings within Tabs
Sometimes the 4 and 6 keys or the 8 or 2 keys may be used to increase or decrease the scaling or level of some settings within pages of the tabs, once the setting is selected. In these cases, once the desired value is specified, press
Under Range Condition
When a measured level is in an under range condition, its displayed level appears in gray rather than black.
Graph Scale Adjustment
The default amplitude (dB) settings for the graphic display of sound pressure level versus time and frequency spectra (1/1 and 1/3 octave) are as indicated below:
5.
5.
Level vs. Time Graph: 20 dB to 140 dB
Frequency Spectra, Normal Range: 20 dB to 140 dB
Frequency Spectra, Low Range: -10 dB to 110 dB
To change the scaling of any one of these graphs, press the Menu softkey to obtain the display shown in FIGURE 5-2.
FIGURE 5-2 Menu
5-2 Tabbed Display LxT Manual

Live Tab

Select Adjust Graph and press 5 to obtain the Adjust
Graph menu shown in FIGURE 5-3
FIGURE 5-3 Adjust Graph Menu
Move the 4 and 6 keys to change the baseline level and the 8 or 2 keys to adjust the height (range between the
baseline and the top of the display). The arrow icons indicate which arrow key adjusts which value. Press implement the change.
Once the scaling on one or more graphic displays has changed, it remains that way until changed again or until the defaults settings are restored.
When changing range between normal and low, the scale is automatically adjusted.
This sections describes the pages on the Live tab. The Live tab contains the following pages:
Profile
Digital
1/1 Octave (optional with separate license)
1/3 Octave (optional with separate license)
Triggering
•Power
Upon turning on the LxT, the Live tab is displayed. The measurements displayed on the Live tab are active, real-time measurements. The displayed values are not controlled by the
9 (RUN/ PAUSE) key. This allows you to view the
current SPL without disrupting any measured data. For example, suppose you are making a measurement and an unwanted event takes place, causing you to stop the measurement. With the measurement stopped, you can monitor the actual level on the Live tab to be certain that the residual effects of the unexpected event have finished before beginning a new measurement.
5 to
LxT Manual Live Tab 5-3
Profile Page
Profile of 1 Sec Leq using selected Frequency Weighting from Setup
Current Value of 1 Sec Leq using selected Frequency Weighting from Setup
Current SPL using selected Frequency Weighting and Detector
Value of L
Peak
using selected Frequency
Weighting, Reset Every Second
Date and Time Measurement Began
The Profile page presents recent graphic history of Leq, using the user-selected frequency weighting, calculated for each second. The graph presents the last 120 seconds of the measurement. Figure 5-4shows the Profile page.
FIGURE 5-4 Live Tab, Profile Page
5-4 Live Tab LxT Manual
The time at the bottom of the page is the date and time that the measurement started.
Digital Page
Indication of Current Exceedance of SPL1 Trigger Level
Current SPL using selected Frequency Weighting and Detector, Updated Once per Second
SPL1 Trigger Level
The Digital page displays both the instantaneous sound level and the value of the user-selected SPL1 Trigger Level, as shown in Figure 5-5.
FIGURE 5-5 Live Tab, Digital Page
In addition to displaying the current value of Leq, this page displays check marks to indicate current of the SPL1 trigger level, which are user-defined, as described in page "Triggers Tab” on page 4-15.
LxT Manual Live Tab 5-5
1/1 Octave Page (Optional)
Frequency Weighting and Time Averaging for Spectrum Measurement
Frequency Spectrum Bandwidth
Measured Level at Cursor Position
Broadband Sound Pressure Level
The 1/1 Octave Band Analyzer appears only when the sound level meter is loaded with the optional LxT-OB1 or LxT-OB3 firmware.
FIGURE 5-6 Live Tab, 1/1 Octave Page
The 1/1 Octave page displays bar graphs of sound level in 1/ 1 octave frequency bands, as shown in Figure 5-6. The right most bar on the graph is the selected broadband SPL value (in this instance, L
AS
).
The spectrum frequency weighting is selected independently from that of the sound level measurement, as described in “OBA Frequency Weighting”. The detector is the same as that of the sound level measurement.
5-6 Live Tab LxT Manual
1/3 Octave Page (Optional)
The 1/3 Octave Band Analyzer page appears only when the sound level meter is loaded with the optional LxT-OB3 firmware.
Triggering
For a description of the setup of these trigger values, see “Triggers Tab”.
The 1/3 Octave Page is similar to the 1/1 Octave Page, but the graph and data are presented for 1/3 octave filters, as shown in Figure 5-7.
FIGURE 5-7 Live Page, 1/3 Octave Page
The Triggering page displays the sound level and peak level with their associated trigger points, as shown in Figure 5-8. Exceedances of these trigger points are indicated by check marks appearing to the right of each. Check marks appear only as long as the measured level remains above the trigger point.
FIGURE 5-8 Live Tab, Trigger Page
LxT Manual Live Tab 5-7
Power Page
The Power page displays the current Date and Time, the run time for the measurement, the battery voltage, the calculated run time and the memory usage.
The Memory section indicates the amount of memory available as a percentage and in number of kilobytes. The number of stored data files is also indicated.
Figure 5-9 shows the Power page.
FIGURE 5-9 Live Tab, Power Page

Overall Tab

This section describes the pages of the Overall tab associated with basic sound level measurements. The
Overall tab includes the following pages:
•Profile
•Digital
•Leq
1/1 Octave (optional with separate license)
1/3 Octave (optional with separate license)
Dose 1
Dose 2
•SEL
•SEA
5-8 Overall Tab LxT Manual
Profile (with Overall Leq)
Leq using Frequency Weighting and Detector from Setup
Sound Level for Currently Selected Weighting and Detector
Leq using selected Frequency Weighting and Detector from Setup
SPL1 Trigger Level
Maximum Level using selected Frequency Weighting and Detector
Percent Of Time Measured Level is Above SPL1 Trigger Level
Measurement Duration
Peak Level using selected Frequency Weighting
Percentiles
Exceedances
Overloads
Community Noise
Miscellaneous
Memory
Figure 5-10 shows the Profile page of the Overall tab.
Digital
LxT Manual Overall Tab 5-9
FIGURE 5-10 Overall Tab, Profile Page
Figure 5-11shows the Digital page of the Overall tab.
Leq
L
Peak
using selected
Frequency Weighting
FIGURE 5-11 Overall Page, Digital Page
The Digital page presents a profile of the sound level for the run time of the measurement. The run time for the average calculation is shown at the top of the screen. The graph is updated once per second and the calculation of the average sound level is updated approximately four times per second.
The Leq Page presents the maximum, minimum, and peak sound levels, as shown in Figure 5-12. On this page, you can view the current sound levels and instantaneous peak values to note their effect on the maximum and minimum sound levels, and the maximum peak level.
5-10 Overall Tab LxT Manual
FIGURE 5-12 Overall Tab, Leq Page
The L
Zpeak(max)
is the highest level the peak detector has measured during the run time of the measurement. The date and time of occurrence is recorded with this event.
The L
Zpeak(max)
is also considered the peak hold. Whenever data is reset, this parameter is cleared. To reset data, press the
1 (RESET) key.
The L
is the highest level the RMS detector has
ASmax
measured during the run time of the measurement. The date and time of occurrence is recorded with this event.
The L
is also considered the max hold. Whenever data
ASmax
is reset, this parameter is cleared. To reset data, press the (RESET) key.
The L
is the lowest level the RMS detector has
ASmin
measured during the run time of the measurement. The date and time of occurrence is recorded with this event.
1
1/1 Octave (Optional)
Frequency Spectrum Bandwidth
Lmax Values
Leq Values
Lmin Values
The 1/1 Octave Band Analyzer page appears only when the optional firmware LxT-OB1 or LxT-OB3 has been enabled and this measurement mode has been selected in the setup.
FIGURE 5-13 Overall Tab, 1/1 Octave Page
Leq
Lmin
Lmax
The 1/1 Octave page displays bar graphs of sound level in 1/ 1 octave frequency bands, as shown in Figure 5-13.
L
is the energy average sound level of the frequency band
eq
for the duration of the measurement.
L
is the minimum sound level of the frequency band for
min
the duration of the measurement.
The value of Lmax for each frequency band is the maximum value that occurred in that band during the entire measurement period. Since individual frequency bands may reach their maximum levels at different times, this spectrum might be one that never occurred at any instant during the measurement period.
1/3 Octave Band Analyzer (Optional)
The 1/3 Octave Band Analyzer page appears only when the sound level meter is loaded with the optional LxT-OB3 firmware and this measurement mode has been selected in the setup.
LxT Manual Overall Tab 5-11
The data displayed in the 1/3 Octave page is similar to that displayed for 1/1 octave spectrum measurements, except that it represents 1/3 octave data. Figure 5-14 shows the 1/3 Octave page.
Dosimeter 1 and 2
FIGURE 5-14 Overall Tab, 1/3 Octave Page
There are two separate, but similar, dosimeter data displays in these pages. Figure 5-15 shows the measurement parameters that correspond to the setup named “OSHA-1”.
.
FIGURE 5-15 Overall Tab, Dosimeter 1
& 2 Pages
TWA(8)
The value of TWA(8) (Time Weighted Average for 8 hours) is based on data measured during the run time and calculated for the user-defined Criterion Time, in this case 8 hours. The
5-12 Overall Tab LxT Manual
value of Criterion Time is set by selecting predefined setups, as described in “Predefined Setups” or by entering numerical values, as described in “Threshold and Criterion”.
For example, suppose a measurement was performed over a time period of ten minutes. The value of TWA(8) would be the same as the TWA measured over an eight hour period if there were no other sound exposure other than that which occurred during that ten minute period.
ProjTWA
The ProjTWA (Projected Time Weighted Average) is calculated from data measured during the measurement run time and calculated for the user-defined Criterion Time, in this case 8 hours. Continuing with the example in the previous paragraph, the ProjTWA for that ten minute measurement represents the value of TWA that would be measured if the noise measured during the ten minute period had continued for eight hours.
L
ep,d
The Daily Personal Noise Exposure, L
is calculated from
ep,d
data measured during the run time of the measurement.
DOSE
Dose is based on data measured during the run time calculated for the user-defined Criterion Time and Criterion Level (100% definition). For example, suppose a measurement was performed over a time period of ten minutes. The value of Dose would be the same as the Dose measured over an eight hour period if there were no other sound exposure other than that which occurred during that ten minute period.
ProjDOSE
Projected Dose is based on data measured during the run time and calculated for the user-defined Criterion Time and Criterion Level (100% definition). Continuing with the example in the above paragraph, the Projected Dose for that ten minute measurement represents the value of Dose that would be measured if the noise measured during the ten minute period had continued for eight hours.
The remainder of the display shows the parameters used for the measurement: Frequency Weighting, Exchange Rate, Threshold and Criterion (time and level).
LxT Manual Overall Tab 5-13
SEL (Sound Exposure)
Figure 5-16 shows the SEL page. The SEL page displays Sound Exposure metrics, (in this instance for A-weighted, Slow).
FIGURE 5-16 Overall Tab, SEL Page
L
is the sound exposure level (previously known as
ASE
SEL). The Sound Exposure metrics indicate the actual and extrapolated (8 and 40 hours) exposure accumulated in terms of hours and seconds. These are discussed in “Sound Exposure (SE)” and “Sound Exposure Level (SEL, LE)”.
5-14 Overall Tab LxT Manual
SEA
The SEA parameter is used mainly in the Canadian province of Quebec.
The SEA page is an integration of 1 second peaks that exceeded 120 dB, as shown in Figure 5-17. Both the SEA value and the frequency weighting used for the measurement are displayed. See “SEA” in the Glossary for a detailed description.
FIGURE 5-17 Overall Tab, SEA Page
LxT Manual Overall Tab 5-15
Percentiles
The Percentiles page displays the Ln statistics for the measurement based on the run time, as shown in Figure 5-
18. Also shown are the maximum and minimum sound levels measured. An Ln is the level that was exceeded “n” percent of the time.
FIGURE 5-18 Overall Tab, Percentiles Page
5-16 Overall Tab LxT Manual
Exceedances
The exceedances count and time shown on this page, as well as those exported in data files, are computed according to your current weighting and detector settings. An exceedance begins when the measured level is greater than the specified threshold and ends when the level is less than or equal to -2 dB of the measured threshold. This prevents excessive exceedance counts when the measured level is at or near the threshold.
Exceedances occur when the instantaneous sound levels are greater than set trigger levels. The Exceedances page shows the number of exceedances that have occurred during the measurement and the total duration of exceedances.
Exceedances are shown for two threshold levels of the RMS detector and three for the peak detector, as shown in Figure 5-19.
FIGURE 5-19 Overall Tab, Exceedances Page
LxT Manual Overall Tab 5-17
Overloads
Community Noise
The Overloads page displays any overloads that might have occurred during the measurement, as shown Figure 5-20.
FIGURE 5-20 Overall Tab, Overloads Page
The times intervals associated with the Day, Evening and Night periods are set as described in “Triggers Tab ”.
5-18 Overall Tab LxT Manual
The Community Noise page displays three equivalent levels calculated for the total measurement time, as shown in Figure 5-21.
FIGURE 5-21 Community Noise Page
The Community Noise page appears only when the optional firmware LXT-CN has been loaded on the sound level meter and this measurement mode has been selected in the setup.
C minus A and Impulsivity
Figure 5-22 shows the C minus A level and Impulsivity pages.
FIGURE 5-22 C-A Level and Impulsivity Page
The integrated levels for L
are always calculated using
Aeq
the linear detector, regardless of the value selected in the SLM Setup. The LAIeq value is from the impulse detector.
L
CSeq
and L
are equivalent levels measured using Slow
ASeq
RMS averaging and using frequency weightings of C and A, respectively. The difference between them, L
CSeq - LASeq,
often used as an indicator of the amount of low frequency content in sounds.
It is also used as a parameter for the selection of hearing protectors, since noise fields having large amounts of low frequency sound can require more effective hearing protectors than would otherwise be indicated by the measured L
level alone.
ASeq
The parameter LAFTMS only appears when Takt M ax i ma l Data has been selected on the Preferences tab.
is
LxT Manual Overall Tab 5-19
Memory

Session Log Tab

The Memory page presents the quantity of each type of measurement made and the memory status, as shown in Figure 5-23.
FIGURE 5-23 Overall Tab, Memory Page
The Session Log is a record of sound measurement actions, as shown in Figure 5-24.
Select the icon and press the Enter key to play a recording.
FIGURE 5-24 Session Log Tab
5-20 Session Log Tab LxT Manual
A time-stamped record is made for every Run, Pause, Stop, Voice Message or Marker action. The source responsible for each action is also recorded, which may be any of the following:
Key press
USB command
Run timer compete
Low battery
•Out of memory
Preamplifier disconnect
The icons in the left column of the display indicate the action: Run, Pause, Voice Recording, etc. The date and time of the action is displayed next to the icon.
Each measurement segment (from Run to Stop) is numbered, as is each voice message.
Navigate through the list and expand each item. The number in the upper right corner of the expanded item indicates which item is being viewed out of how many total items are in the list.

View Spectrum Normalized (Optional)

The Live Spectrum continues to change in time following the normalization, whereas the Reference spectrum remains the same.
LxT Manual View Spectrum Normalized (Optional) 5-21
The View Normalized function permits the display of the difference between two spectra by subtracting a user­selected reference spectrum from the measured spectrum. This function can be used with both 1/1 and 1/3 octave spectra, although the measured spectrum and the reference spectrum must have the same bandwidth: 1/1 or 1/3 octave.
View Spectrum Normalization is context sensitive and displays the normalized spectrum for the data from the tab where it was activated, whether Live, Overall, Measurement History, or others.
A and C frequency weighting curves can also be used for references, as described in "Normalizing using Frequency Weighting” on page 5-24.
A standard spectrum displayed on the Live tab appears as shown in FIGURE 5-25.
FIGURE 5-25 Live Spectrum Display
To access the View Normalized display, press the center softkey Menu to display the menu shown in FIGURE 5-26 when viewing 1/1 or 1/3 octave data.
FIGURE 5-26 Menu
5-22 View Spectrum Normalized (Optional) LxT Manual
Select View Normalized. This displays the spectrum in the normalized view, as shown in FIGURE 5-27.
FIGURE 5-27 Normalized Live Spectrum Display
By default the display represents a spectrum normalized to reference spectrum 1. If reference spectrum 1 has not been previously defined, the display shows the actual live SPL spectrum.
Selecting the Spectrum Type
In the live display, it is not necessary to select a spectrum type since only SPL is available.
Selecting the Reference Spectrum
Select the field referenced in Figure 5-27 and press open the menu shown in FIGURE 5-28.
FIGURE 5-28 Spectrum Type Menu
Select from the three spectrum types: Leq, Lmax and Lmin.
LxT Manual View Spectrum Normalized (Optional) 5-23
5 to
Select the Ref: data field and press 5 to open the
Reference Menu, as shown in FIGURE 5-29.
FIGURE 5-29 Live Reference Menu
The items listed in this menu are as follows:
•1
•2
•3
•4
•A
•C
•-A
•-C
The first four items permit the user to define four reference spectra.
The last four items permit the user to use add positive or negative A or C frequency weightings to the displayed spectrum.
Setting a Reference Spectrum
By pressing the Set softkey, the current spectrum is set as the reference spectrum for the selected reference (1, 2, 3 or 4).
Following this procedure, reference spectra 1, 2, 3 and 4 can be defined. Once defined, the user can choose to display the live SPL spectrum normalized to any one of these four reference spectra.
Normalizing using Frequency Weighting
The A and C reference spectrum represent the A and C frequency ratings sampled at the center frequency of the selected filter.
Selecting A or C approximates an A or C weighted spectrum when the original data is unweighted, z weighted. Using -A and -C removes the effects of A or C frequency weighting.
5-24 View Spectrum Normalized (Optional) LxT Manual
FIGURE 5-30 shows an -A reference display for an A­weighted version of a spectrum.
Graph Relative
FIGURE 5-30 Normalized Spectrum with -A
Weighting
Selecting Graph Relative changes the graph to show the difference between the current data and the reference, with the center of the graph being 0dB. This feature can be used to easily determine if a reference has been exceeded.
LxT Manual View Spectrum Normalized (Optional) 5-25
5-26 View Spectrum Normalized (Optional) LxT Manual
CHAPTER

Making Measurements

6

Preparation

Positioning the LxT

Observer Position
This chapter describes how to make and store accurate sound level measurements.
Before making a measurement, make sure of the following:
The instrument hardware has been properly assembled as described in "Preparing for First Use".
The instrument has been calibrated as described in Chapter 7 "Calibration”.
The measurement has been configured as described in "Basic Measurement Setup".
This section describes how to position the LxT for optimal sound level measurements.
In order to avoid the effect of sound reflections from the body of the operator interfering with the measurement, the meter should be located as far as possible from the body. Thus, when actually performing the measurement, the operator should place himself at a distance behind the tripod-mounted meter, or extend the hand-held meter as far from the body as is comfortable.
Microphone Extension Cable
If desired, a microphone extension cable may be placed between the meter and the preamplifier/microphone. No correction is required when using Larson Davis Model EXC<xxx> shielded microphone extension cables in combined lengths up to 200 feet. (The variable <xxx> represents the length in feet of the cable.)
LxT Manual Making Measurements 6-1
Use of a Windscreen
Wind blowing across the microphone generates pressure fluctuations on the microphone diaphragm that can produce errors in the measurement. As a result, when performing measurements in the presence of low level airflows, it is recommended that a windscreen be placed over the microphone. Larson Davis provides the WS001 windscreen, a 3 1/2” diameter ball made of open cell foam which can be placed over the microphone and preamplifier as shown in FIGURE 6-1.
FIGURE 6-1 Position of Windscreen

Performing Measurements

This section describes the steps for performing basic sound level measurements.
Starting the Measurement
The LxT uses a single range for sound level measurements, so there is no need to select a range. as part of making a measurement.
6-2 Performing Measurements LxT Manual
The Live tab displays current acoustic data that is not being recorded or stored. Pressing the 9 causes the LxT to begin storing data, which is displayed on the Overall tab.
(RUN/PAUSE) key
Measurement Range
Under Range
Not Under Range
Overload Indication
Under Range Indication
The measurement ranges in which the LxT meets the standards, which depend upon the selected frequency weighting, are shown in Table A-4, “LxT Performance Specifications,” on page A-4. Measurements which include levels outside this range should not be considered accurate.
When input signals exceed the input range of the LxT, the Input Overload Icon appears at the top of the display.
If a measurement is running and an overload occurs, the icon flashes on and off for as long as the overload condition exists, or one second minimum. When the overload has been removed, the icon remains present (not flashing) to indicate that an overload has occurred during the measurement. A reset clears the icon from the display.
When input signals drop below the level that the LxT can measure within specified tolerances, an under range condition exists. When this happens the Under Range Icon appears.
LxT Manual Performing Measurements 6-3
As long as the under range condition exists, the icon flashes. When the measured level no longer produces an under range condition, the icon is removed from the display.
At any time when a measured parameter is in an under range condition, it’s numeric display appears in gray rather than the usual black, as shown FIGURE 6-2.
FIGURE 6-2 Normal vs. Under Range Data Display
Pausing Measurements
5 Second Back Erase Label
Measurements may be paused and then run again multiple times.
Back Erase
At any time the measurement of overall data can be temporarily suspended by pressing the 9 run clock also pauses. However, instantaneous data continues to be displayed on the Live tab.
Pressing the 9 key one more time starts the measurement again; overall data continues to accumulate. The run clock also begins again from the time indicated when the pause occurred. The overall data is not affected by any acoustic events occurring during the time period that the LxT is paused.
key. Note that the
Back Erase is disabled when Measurement History has been enabled in the LxT setup, as described in Chapter 10 "Measurement History”.
The back erase function permits the user to rapidly delete from the measurement the effects of acoustical events that have occurred during the previous five or ten seconds. The back erase can be implemented when the measurement is paused, as described in the preceding section. When the measurement is paused, the center softkey is labeled Back- 5s, as shown in FIGURE 6-3.
FIGURE 6-3 Five Second Back Erase Label
The Back-5s label does not indicate the state of the instrument but that an action can now be taken to delete the last five second segment.
6-4 Performing Measurements LxT Manual
Press the center softkey to implement a five second back erase.
> Ten Seconds Since Last Stop or Pause
Ten Second Back Erase Label
After pressing the center softkey to implement a five second back erase, if the measurement duration since the last Stop or Pause has been more then ten seconds, the center softkey is then labeled Back-10s, as shown in FIGURE 6-4.
FIGURE 6-4 Ten Second Back Erase Label
The Back-10 label does not indicate the state of the instrument but that an action can now be taken to delete the last ten second segment.
LxT Manual Performing Measurements 6-5
The user can take one of the following actions:
Press the 9 key to accept the five second back erase and continue the measurement.
Press the center softkey to extend the back erase to ten seconds. The center softkey is then labeled Undo, as shown in FIGURE 6-5.
< Ten Seconds Since Last Stop or Pause
Back Erase Undo Indication
After pressing the center softkey, if the measurement duration since the last Stop or Pause has been less than ten seconds, the center softkey is then labeled Undo, as shown in FIGURE 6-5.
FIGURE 6-5 Back Erase Undo Indication
Time History Records
6-6 Performing Measurements LxT Manual
The user can take one of the following actions:
•Press the 9
five second segment removed.
Press the center softkey to implement the Undo action and then press the 9 key to continue the measurement without removing the previous five second time segment.
The time history records from the point that data was restored to the last record are marked as back erase records in the marker field.
key to continue the measurement with the
Resetting Measurements
A measurement is most often reset when a noise event which is not typical of the desired measurement takes place. For example, an aircraft passing overhead when attempting to measure the background noise in a normally quiet area may be cause for resetting.
Starting a New Measurement
Stopping Measurements
The LxT can be stopped when either running or paused.

Storing Measurements

Measurements can only be stored when they have been stopped.
To reset a measurement in progress, press the key. This erases all data previously measured and resets the run time clock to zero. A reset does not reset stored data files.
A reset can be initiated when the LxT is running, paused or stopped. However, it must be stopped for the reset operation to be performed.
The 9
Press the 7 (STOP) key to suspend the overall measurement.
Pressing the 9 measurement which had been stopped.
To store the measurement, press the 7 key one more time. The Save File menu is then displayed, as shown in FIGURE 6-6 "Save File Menu” .
key must be pressed to start a new measurement.
key afterwards continues the overall
1 (RESET)
FIGURE 6-6 Save File Menu
LxT Manual Storing Measurements 6-7
After a file has been successfully stored, the LxT automatically resets when the another measurement.
Overwriting a Saved File
9
key is pressed to begin
The data is stored under the file name defined in the section "General Tab" along with a file number. The file number automatically begins at 000 for the first measurement stored. The file number is also indexed so that whenever a measurement is stored, the file number assigned is the next in sequence following the measurement previously stored.
If you wish to use this data to replace a data file already saved in the LxT, select the box with the title “...” and press
5 (ENTER). This opens a window listing all the data files
already saved in the LxT, as shown in FIGURE 6-7.
FIGURE 6-7 Saved Data Files
Select the file that is to be overwritten. This replaces the file name and number that previously appeared with the one shown on the display, as shown in FIGURE 6-8.
FIGURE 6-8 Overwriting a Saved File
6-8 Storing Measurements LxT Manual
To continue with the overwrite operation, select Ye s and press
5.

Data Storage After Improper Shutdown

When the LxT has been shutdown improperly during a measurement, for example during a power outage, the procedure for handling the data depends upon the setup being used at the time of the measurement. This section describes two cases.
Case 1
If the Run Mode is:
Continuous
or
Single Block Timer
or
Daily
and
Daily Autostore is enabled
Normal Operation
Under normal operation the stored data files are stored with the following name format:
yymmdd00.LD0
where yymmdd is the date the file was stored.
Improper Shutdown
Following an improper shutdown, when the instrument is next turned On, the data is automatically stored using the following name format:
yymmddxx.LD0
where yymmdd is the date the data is stored and xx is a number, beginning at 01, which is automatically incremented for subsequent instances of improper shutdown.
Case 2
This case covers all setups other than those described in Case 1.
LxT Manual Data Storage After Improper Shutdown 6-9
Improper Shutdown
When present, the user must eventually select to store the data, reset the instrument, or take no action, in which case the instrument is automatically reset.
Following an improper shutdown, when the instrument is next turned On:
Step 1 The user is prompted to save the data.
If the user responds by selecting to store the data, the data is stored and the instrument is reset
If there is no user response to the prompt within ten seconds, the instrument is reset
If the user responds by selecting not to store the data, the sequence moves to Step 2.
Step 2 The user is prompted to reset the instrument
If the user responds by selecting to reset the instrument, the instrument is reset.
If there is no user response to the prompt within ten seconds, the instrument is reset.
If the user responds by selecting not to reset the instrument, the sequence moves back to Step 1.
This sequence is diagramed below
6-10 Data Storage After Improper Shutdown LxT Manual
FIGURE 6-9 Improper Shutdown Sequence, Case 2
Instrument Reset
No Response for 10 seconds
No
Reset Instrument prompt
Ye s
No
No Response for 10 seconds
Save Data prompt
Data Saved
Ye s
LxT Manual Data Storage After Improper Shutdown 6-11
6-12 Data Storage After Improper Shutdown LxT Manual
CHAPTER

Calibration

7

Calibration Overview

Sensitivity Determination
Overload/Under Range Conditions
This chapter describes both the purposes and steps for calibrating the SoundTrack LxT LxT.
The primary role of sound level meter calibration is to establish a numerical relationship between the sound level at the diaphragm of the microphone and the voltage measured by the meter so that the sound pressure level can be read directly from the display of the meter in units of dB. The result of a calibration is the determination of the sensitivity of the meter, including microphone and preamplifier, typically in units of dB re 1V/Pa or mV/Pa.
A secondary role of calibration is to determine the sound level that would overload the instrument and the minimum sound level that can be accurately measured, referred to as the under range level. This requires a knowledge of the electrical noise levels of the microphone, preamplifier, and the instrument circuitry.
and SoundExpert
Calibration Stability
The LxT should maintain a stable value of sensitivity over long periods of time. Significant changes in sensitivity, or a pattern of small but regular sensitivity changes, are indicative of problems with the measurement system, calling for laboratory calibration and possibly service. To assist the user in identifying these situations, the LxT provides two notifications:
Calibration History
Data and date/time of the most recent ten calibrations.
LxT Manual Calibration 7-1
Large Change Notification
During calibration, an automatic comparison is made between the sensitivity determined by the calibration and a published value of sensitivity. An on-screen window appears to warn the user when the difference between these two values exceed 3 dB.

Control Panel - Calibrate

To activate the Calibration function, press the 3 (TOOLS)
key and select the Calibrate icon as shown in Figure 7-1.
FIGURE 7-1 Control Panel
Press
5 (ENTER) to open the Calibrate tabs.
7-2 Control Panel - Calibrate LxT Manual
The four Calibration tabs are shown in FIGURE 7-2.
Calibrate Tab
History Tab
Sensitivity Tab Certification Tab
FIGURE 7-2 Calibration Tabs
Calibrate Tab
LxT Manual Control Panel - Calibrate 7-3
The Calibrate tab is used when performing an acoustic calibration, including the selection of the sound level calibrator to be used and the implementation of the calibration procedure.
History Tab
To create a history record for a calibration, it must be saved; calibrations must also be saved to export post-calibration data to software.
Sensitivity Tab
Certification Tab
Exiting from the Calibration Function
The History tab lists the results, along with the date and time, of the ten most recent calibrations performed using the same type of preamplifier as presently connected to the LxT. The preamplifier name appears at the top of the tab (PRMLxT1 in this example). The preamplifier type is read automatically when the instrument is booted up, or following a change in preamplifier. The value of sensitivity in dB re. 1 V/Pa and the variation of the sensitivity determined from that calibration relative to the calibration prior to that,
When performing an acoustic calibration, the Sensitivity tab is used to select the microphone being used.
The Certification tab shows the date of the last certification, the due date for the next certification, and information about the calibration facility. You can also enter your certification interval and certification reminder from this tab.
Press the Close Softkey to exit from any of the calibration tabs to the Control Panel.
dB, are presented for each calibration.

Acoustic Calibration

This is the most commonly used calibration method, and the one required by most national and international standards prior to performing a measurement. A sound level calibrator is used to apply an acoustical signal of a known amplitude and frequency to the microphone. From the voltage level measured by the meter, the sensitivity can be determined. With this technique, it is assumed that the calibrator is functioning correctly; any variation in level results in an improper calibration and an erroneous value of sensitivity. For this reason, the user is advised to compare the newly determined sensitivity with the previous sensitivity to ensure that significant variations have not occurred.
7-4 Acoustic Calibration LxT Manual
Frequency Weighting
Calibrator
Recommended Calibrator
The LxT automatically switches to C frequency weighting and Fast detector response for calibration. This permits 250 Hz and 1000 Hz calibrators to be used. The Fast detector response reduces the stabilization time required before calibration.
After calibration, the LxT returns to the original frequency and time weighting set by the user.
The Calibrator section of the Calibrate tab includes an area to enter information about a calibrator and a list of calibrators. The user may select a calibrator from the list or enter new information about a calibrator.
Table 7-1 'Recommended Calibrators for Use with LxT1 and LxT2' lists the sound level calibrators which Larson Davis recommends for calibrating the LxT1 and LxT2.
When using a 1/4” microphone, the adaptor ADP024, a 1/4” microphone adaptor for the 1/2” opening in the CAL150 and CAL200 calibrators, is also required.
Calibrator Instrument
CAL200 LxT1, LxT2 Class 1 94/114 dB 1 kHz CAL150 LxT2 Class 2 94/114 dB 1 kHz
Table 7-1 Recommended Calibrators for Use with LxT1 and LxT2
Calibrating the LxT1 and 377B02 microphone
The CAL200 provides a nominal pressure level of 94 dB or 114 dB. The exact levels are printed on the Larson Davis calibration sheet that came with the calibrator. When using a free-field microphone, the pressure level at the microphone diaphragm is slightly different. Thus, a free field correction of -0.12 dB (0.03 dB uncertainty at 95% confidence level) should be applied to either of these levels. Pressure and random incidence microphones do not require this correction. If the calibrator and instrument are near room
LxT Manual Acoustic Calibration 7-5
Calibrator
Precision
Output Frequency
temperature (23° C) and near sea level (101.3 kPa) then no other corrections need to be made. For example, if the calibration sheet for the CAL200 indicates 113.98 dB for its level when set to 114 dB then set the Cal Level in the LxT to
113.86 dB and 1000 Hz.
When the microphone and instrument are at a temperature other than near room temperature or static pressures not near sea level, then corrections need to be added for the ambient temperature and the prevailing static pressure. Check the calibration data shipped from Larson Davis with the CAL200 to get these corrections. The corrections can be added to the level obtained in the previous paragraph to get the actual level of the CAL200.
The 377B02 microphone's sensitivity varies with static pressure. If the instrument is calibrated in one environment and moved to another, then the sensitivity changes (after stabilization) depending on the change of temperature and pressure. The coefficient of static pressure is -0.01 dB/kPa. If the system is calibrated at 85 kPa for instance then it is
0.16 dB less sensitive at sea level (101.3 kPa). The
sensitivity of the 377B02 and LxT vary slightly with temperature also. The coefficient of temperature is +0.009 dB/°C. If the system is calibrated at 18° C then it is 0.045 dB more sensitive at 23° C.
The Larson Davis 3" Wind Screen has less than 0.05dB effect on the system response at 1 kHz.
Set the CAL200 level switch to 94 or 114 dB.
Calibrating the LxT2 and 375B02 microphone
The CAL200 and CAL150 provide a nominal pressure level of 94 dB or 114 dB. The exact levels are printed on the Larson Davis calibration sheet that came with the calibrator. When using a free-field microphone, the pressure level at the microphone diaphragm is slightly different. Thus, a free field correction of -0.12 dB (0.03 dB uncertainty at 95% confidence level) should be applied to either of these levels. If the calibrator and instrument are near room temperature (23° C) and near sea level (101.3 kPa) then no other corrections need to be made. For example, if the calibration sheet for the CAL200 or CAL150 indicates 113.98 dB for it's level when set to 114 dB, then set the Cal Level in the LxT to 113.86 dB and 1000 Hz.
7-6 Acoustic Calibration LxT Manual
Environmental Parameter Ranges
When the microphone and instrument are at a temperature other than near room temperature or static pressures not near sea level, then corrections need to be added for the ambient temperature and the prevailing static pressure. Check the calibration data shipped from Larson Davis with the CAL200 or CAL150 to get these corrections. The corrections can be added to the level obtained in the previous paragraph to get the actual level of the CAL200 or CAL150.
The 375B02 microphone's sensitivity varies with static pressure. If the instrument is calibrated in one environment and moved to another, then the sensitivity changes (after stabilization) depending on the change of temperature and pressure. The coefficient of static pressure is -0.01 dB/kPa. If the system is calibrated at 85 kPa for instance then it is
0.16 dB less sensitive at sea level. The sensitivity of the 375A02 and LxT vary slightly with temperature also. The coefficient of temperature is -0.015 dB/°C. If the system is calibrated at 18° C then it is 0.07 dB less sensitive at 23° C.
The Larson Davis 3" Wind Screen has less than 0.05dB effect on the system response at 1 kHz.
For proper calibration, the calibration procedure and the correction values apply over the ranges presented in Table 7-
2.
Instrument Class Parameter Range
Static Pressure 65 kPa to 108 kPa Class 1, LxT1 with 377B02 microphone
Table 7-2 Environmental Parameter Ranges for Calibration
LxT Manual Acoustic Calibration 7-7
Temperature - 10 C to + 50C
Relative Humidity 25 % to 90%, without condensation
9.4 psi to 15.7 psi
14 F to + 122F
from - 10 C to + 39C (14 F to + 102F)
Loading...