LaMotte WaterLink SpinTouch 3580 Quick Start Manual

FILLING: SINGLE-USE DISK
TESTING
QUICK START GUIDE
BEFORE FIRST USE
¬ Fully charge battery. ¬ Ensure your Spin Touch has the latest fi rmware:
®
at
www.waterlinkconnect.com/Downloads/Application.
2. Plug meter into the computer with the provided USB cable and launch WaterLink Connect desktop application from the Start Menu.
3. Any fi rmware updates will automatically be applied to the meter.
1. Fill the disk using slow, even pressure. The sample water will fi ll the spaces between the baffl es in a counterclockwise order. Each space will fi ll from the bottom to the top. Sample water should be added until the sample water in the fourth chamber fi lls to the top of the chamber slightly past the embossed fi ll line. It is OK to fi ll slightly past the fi ll line.
Correct Fill
(Fill line)
2. Do not overfi ll the disk. If the disk is overfi lled, sample water will fl ow out of the overfl ow hole in the center of the disk. The disk is not leaking. Dry the disk and run the test.
Over Filled
(Dry disk before placing in meter)
3. Do not under fi ll the disk. If the disk is under fi lled, the reagent chambers will not fi ll entirely and results will be inaccurate.
Under Filled
4. Wet disks should be dried thoroughly with a lint free wipe. The disk should be handled by the edges.
5. Disks should be fi lled and used within 10 minutes. They cannot be fi lled ahead of time.
1. Press
and hold until the meter turns
on.
2. Tap
. Select a sanitizer type. Tap
to confi rm.
3. Tap
. Select a disk series (found on
disk packaging). Tap
to confi rm.
4. Remove disk from packaging.
5. Use syringe (1189) to fi ll disk with pool or spa water sample.
6. Insert disk. Cover disk with Universal Disk Cover. Close lid.
7. Tap
to start test. Tap to stop the test.
8. Results will be displayed.
9. Select an option.
- Tap
to return to the Test Screen.
- Tap highlighted
to save test results
to test log if Auto Save is not enabled.
- Tap the hightlighted
to send the results
to the enabled Mobile Bluetooth Printer.
- Tap highlighted
to send results to
Bluetooth enabled device.
10. Press and hold
for 2 seconds to turn
meter off .
NOTE: For water samples over 100 °F (38 °C) subtract 0.3 from pH result or, for the most accurate result, wait until water sample is below 90 °F (32 °C) to test.
1. Fill any unused section of the disk using slow, even pressure. Add the sample water until the Blank Chamber is full and the Over Flow Chamber is partially fi lled.
Correct Fill
Blank Chamber
Over Flow Chamber
2. Do not overfi ll the disk. If the disk is overfi lled, sample water will fl ow out of the vent hole in the center of the disk. The disk is not leaking. Dry the disk and run the test.
Over Filled
(Dry disk before placing in meter)
Blank Chamber
Over Flow Chamber
3. Do not under fi ll the disk. If the disk is under fi lled, the blank chamber will not be full. The reagent chambers will not fi ll entirely and results will be inaccurate.
Under Filled
Blank Chamber
4. Do not introduce air bubbles into the disk. The reagent chambers will not fi ll entirely and results will be inaccruate. Remove the water sample with the syringe and refi ll.
Air Bubble
5. Wet disks should be dried thoroughly with a lint free wipe. Handle the disk by the edges. Between tests store the disk on a fl at surface. Do not agitate a partially fi lled disk to avoid leaking and contamination. Partially used disks should be discarded at the end of the day.
FILLING: MULTIPLE-USE DISK
Compatible with iPhone 6 Plus, iPhone 6, iPhone 5s, iPhone 5c, iPhone 5, iPad Air 2, iPad Air, iPad mini 3, iPad mini 2, iPad mini, and iPad (4th generation).
Compatible with Android devices.
DOS & DON’TS
¬ Hold syringe vertically when fi lling disks. ¬ Do not fi ll disk while in the meter. Fill disk on clean, dry surface. ¬ The disk should not contain any large air bubbles. Air bubbles will result in erroneous results. ¬ Only the Universal Disk Cover (Code 1719) can be used with the WaterLink
®
Spin Touch™. ¬ Empty syringe of previous sample before fi lling with next sample. ¬ Remove fi lled disk from meter after testing.
BLUETOOTH CONNECTION
The WaterLink® Spin Touch™ is capable of connecting to a Bluetooth enabled device, such as a phone or tablet, and the Mobile Bluetooth Printer. The Spin Touch will only connect to and print from the LaMotte Bluetooth Printer [Code 5-0066]. Other Bluetooth printers are not supported.
Whenever the
button appears on the touch screen the Spin Touch is capable of connecting
to a Bluetooth enabled phone or tablet. The
will become highlighted when connected and
be dim when not connected. Whenever the
button appears on the touch screen the Spin
Touch is capable of connecting to the Mobile Bluetooth Printer. The
will be highlighted when connected and dim when not connected. The Spin Touch cannot connect to both a phone or tablet and the printer at one time.
For fi lling and troubleshooting tips go to:
www.waterlinkspintouch.com
See the WaterLink® Spin Touch™ manual for additional information.
softwaresupport@lamotte.com
410-778-3100 | 800-344-3100
www.lamotte.com
Disk Patent No. 8,734,734
FCl Patent No. 8,987,000 TCl Patent No. 8,993,337
3580-QG 11.15.17
Loading...