PHOSPHATE KIT
OCTA-SLIDE 2, 1-10 ppm
CODE 4408-01
QUANTITY CONTENTS CODE
2 x 30 mL *VM Phosphate Reagent *4410-G
30 mL *Reducing Reagent *6405-G
1 Pipet, dropping, plastic 0352
1 Pipet, 1.0 mL, plastic 0354
1 Pipet, 0.5 mL, plastic 0353
2 Test Tubes, 2.5-10.0 mL, plastic, w/caps 0106
1 Phosphate Octa-Slide 2 Bar, 1-10 ppm 4414-01
1 Octa-Slide 2 Viewer 1101
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specifi ed code number.
NOTES: This test determines the levels of orthophosphates only.
USE OF THE OCTA-SLIDE 2 VIEWER
The Octa-Slide 2 Viewer should be held so
non-direct light enters through the back of
the Viewer. Insert the reacted sample into
top of the Viewer. Slide the Octa-Slide 2
the
Bar
into the Viewer and match the color of the
reaction to the color standards.
WARNING! This set contains chemicals
that may be harmful if misued. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision.
PROCEDURE
NOTE: If sample water is cloudy, fi lter before testing.
1.0 - 10.0 ppm PHOSPHATE (PO4)
1. Fill a test tube (0106) to the 5 mL line with sample water.
2. Use the 1.0 mL pipet (0354) to add 1.0 mL of *VM Phosphate Reagent (4410).
Stopper and mix by inverting several times. Wait 5 minutes.
3. Use the plain pipet (0352) to add 3 drops of *Reducing Reagent (6405). Cap and
mix. Color will develop in 10 seconds.
4. Insert the Phosphate Octa-Slide 2 Bar (4414-01) into the Octa-Slide 2 Viewer
(1101).
5. Insert test tube into the Octa-Slide 2 Viewer. Match sample color to a color standard. Record as ppm Phosphate (PO
10 - 100 ppm PHOSPHATE (PO4): 1 to 10 DILUTION
).
4
1. Use the 0.5 mL pipet (0353) to add 0.5 mL of sample water to a test tube (0106).
2. Dilute to 5 mL line with phosphate-free (distilled) water.
3. Use the 1.0 mL pipet to add 1.0 mL of *VM Phosphate Reagent (4410). Cap and
mix by inverting several times. Wait 5 minutes.
4. Use the plain pipet (0352) to add 3 drops of *Reducing Reagent. Cap and mix.
Color will develop within 10 seconds.
5. Insert the Phosphate Octa-Slide 2 Bar (4414-01) into the Octa-Slide 2 Viewer
(1101)
6. Insert test tube into the Octa-Slide 2 Viewer. Match sample color to a color standard. Multiply by 10. Record as ppm Phosphate (PO4).
CAUTION: Do not use soaps or detergents containing phosphates to wash test tubes.