LaMotte P-70 User Manual

SUL FIDE IN WA TER TEST KIT
L
M
o
tt
L
Mott
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
OCTA-SLIDE METHOD
MODEL SL-P70 • CODE 3322
QUANTITY CONTENTS CODE
30 mL *Sulfide Reagent A *4458-G
15 mL *Sulfide Reagent B *4459-E
60 mL Sulfide Reagent C 4460-H
1 Pipet, 1.0 mL, plastic 0354
1 Octa-Slide Viewer 1100
1 Sulfide Octa-Slide Bar, 0.2-20 ppm 3414
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
USE OF THE OCTA-SLIDE VIEWER
The Octa-Slide Viewer should be held so non-direct light enters through the back of the viewer. With sample tube inserted at top, slide the Octa-Slide bar through the viewer and match with color standard.
PRO CE DURE
NOTE: The sample must be collected with a minimum of aeration and should be analyzed promptly.
1. Fill a test tube (0106) to the 5 mL line with the sample.
2. Add 15 drops of *Sulfide Reagent A (4458). Cap and mix. CAUTION: *Sulfide Reagent A is a strong acid solution and must
be handled with great care.
3. Add 3 drops of *Sulfide Reagent B (4459). Cap and mix. Wait for 1 minute.
4. Use the 1.0 mL pipet (0354) to add 1.0 mL of Sulfide Reagent C (4460). Cap and mix. A blue color indicates Sulfide.
5. Insert the test tube into the top of the Octa-Slide Viewer (1100). Slide the Sulfide Octa-Slide Bar (3414) into the Viewer. Match sample color to a color standard. Record result as ppm Sulfide.