LaMotte P-54 User Manual

Page 1
CODE 1929
MULTI-RANGE CONDUCTIVITY METER
OPERATOR’S MANUAL
LaMOTTE COM PANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Mary land • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
SM
61918 · 9/03
Page 2
INTRODUCTION
The DA-1 Multi-Range Conductivity Meter is a portable, battery-operated instrument with solid state circuitry. The meter accurately measures a wide range of conductivity with a single probe. A conductivity reading measures the capacity of a solution to conduct an electric current and is directly related to the total ionic concentration of dissolved substances in the solution.
The instrument’s capabilities range from checking the purity of distilled water to use as a salinometer for monitoring the conductivity of sea water. Other applications include checking air and water pollution, evaluating electroplating baths, measuring soluble salts in soils, and monitoring industrial and municipal water treatment operations.
This easy-to-use instrument has a rugged 3½" taut-band meter movement with a mirrored scale that eliminates reading errors due to parallax. Other features include a battery check circuit and manual temperature compensation from 0°C to 100°C.
The conductivity meter is equipped with a range selector which monitors conductivities from 0-100,000 micromhos/cm. A single conductivity probe on a three foot cable is used to measure all four ranges.
REPLACING THE BATTERY
When meter does not indicate greater than 80 micromhos in the “BAT.CK” position, replace the battery. The battery is located under the meter panel of the instrument. Carefully remove the panel by unscrewing the four screws on the front of the unit and lift the unit out. Make sure battery terminal contacts are aligned with the correct terminals. Replace meter panel to instrument housing.
INSTRUMENT GUARANTEE
This instrument and probe, except for the battery, are guaranteed to be free from defects in material and workmanship for six months from original purchase date. If within that time the instrument is found to be defective, it will be repaired without charge (except for transportation expense to LaMotte Company, Chestertown, Maryland 21620).
This guarantee is void under the following circumstances: operator’s negligence, improper application, and unauthorized servicing.
LIMITS OF LIABILITY
Under no circumstances shall LaMotte Company be liable for loss of life, property, profits or other damages incurred through the use or misuse of their products.
PACKAGING AND DELIVERY
Experienced packaging personnel at LaMotte Company assure the adequate protection of our products against normal hazards encountered during shipment. After the product leaves the manufacturer, all responsiblity for its safe delivery is assured by the transportation company. Damage claims must be filed immediately with the transportation company to receive compensation for damaged goods.
Should it be necessary to return the instrument for repair or servicing, pack instrument carefully in suitable container with adequate packing material. Contact the Technical Services Department for a Return Authorization Number Attach a letter to the shipping carton describing the kind of trouble experienced. This valuable information will enable the service department to make the required repairs more efficiently.
2
11
Page 3
PRO CE DURE
1. Follow steps 1-3 in General Operating Procedure.
2. Set the range selector at “X100" and switch the control knob to the
”ON" position. Measure the temperature of the 6,668 micromhos/cm solution (Solution #1) and set the “TEMP” knob accordingly. Sufficiently immerse the probe in the solution to at least cover the vent holes. Adjust the “STD” knob until the meter reading corresponds to the solution value.
3. Remove the 4 screws from the meter panel and lift unit from the base thereby exposing the internal circuitry.
4. Switch the control knob to the “CAL” position and adjust to a full scale reading of 100 micromhos/cm by using the internal trim pot labeled “full scale” located on the PC board where the switches are mounted.
5. Switch the control knob to the “ON” position and the range selector to “X1000.” The “TEMP,” “STD,” and internal calibration pot should not be changed for the X1000 range. Sufficiently immerse the probe in the 58,640 micromhos/cm solution (Solution #2) to cover the vent holes. Adjust the trim pot labeled “X1000" located on the meter PC board until the meter reading corresponds to the solution value.
6. The instrument is now calibrated for all ranges.
The following prepared conductivity solutions are available from LaMotte Company. Use the specified code number to order.
STAN DARD CON DUC TIV ITY SO LU TIONS
DESCRIPTION SIZE CODE
0.0005M KCl, 74 micromhos/cm 500 mL 6416-L
0.005M KCl, 718 micromhos/cm 500 mL 6417-L
0.01M KCl, 1413 micromhos/cm 500 mL 6354-L
0.05M KCl, 6668 micromhos 500 mL 6418-L
0.5M KCl, 58640 micromhos 500 mL 6419-L
SPECIFICATIONS FOR CONDUCTIVITY METER
METER
Range: Multi-Range
Ac cu racy: Pre ci sion: Power: Di men sions: Weight:
FUNC TION CON TROLS
Op er a tion Switch:
Stan dard ize: Tem per a ture Com pen sa tor:
Range Se lec tor:
CON DUC TIV ITY PROBE
Range: Size: Weight: Ca ble Length: Probe Con stant: El e ments:
CARRYING CASE
Di men sions: Weight:
Low: 0-100 in 2 micromhos/cm in cre ments Me dium: 0-1000 in 20 micromhos/cm in cre ments Stan dard: 0-10,000 in 200 micromhos/cm in cre ments High: 0-100,000 in 2000 micromhos/cm in cre ments ± 2% full scale read ing (all ranges) ± 0.5% full scale read ing (all ranges) Bat tery - sin gle 9 volt tran sis tor 3 ¾”W x 2 5/8”D x 6 ¼”H
1.7 lbs.
ON - Turns in stru ment on for read ing OFF - Cuts off en ergy sup ply BAT. CK. - To check bat tery re serve CAL. (Cal i bra tion) - To cal i brate in stru ment Cal i brates in stru ment by reg u lat ing volt age po ten tial 0° to 100°C in 5°C in cre ments - ad justs in stru ment
to tem per a ture of sam ple be ing tested X1 - con duc tiv ity read ing times 1 X10 - con duc tiv ity read ing times 10 X100 - con duc tiv ity read ing times 100 X1000 - con duc tiv ity read ing times 1000
0-100,000 micromhos/cm 6" x ½" 2 oz. 3 ft.
5.0/cm Car bon
10”W x 6”D x 6 ¼”H With me ter, 3.5 lbs.
10
3
Page 4
NOTES ON THE USE OF THE CONDUCTIVITY PROBE
The sensing elements of the conductivity probe consist of two carbon elements rigidly supported in a rugged plastic material.
These elements have a very precise parallel spacing which determines the probe cell constant. To keep this spacing and to assure accurate results, it is essential that the probe is properly maintained and kept clean.
1. When conductivity measurements are made, the probe must be immersed in at least ½" of solution. The probe may be completely immersed within the sample without adversely affecting the results.
2. There is a tendency for the conductivity probe to entrap air bubbles which may result in erroneous readings. After the probe is immersed in the test solution, tap or “jiggle” the probe until air bubbles are removed.
3. The conductivity probe must be kept from contacting the bottom or sides of the container which holds the sample. Contact with surfaces other than the solution might result in inaccurate readings since the current flow may be altered between the probe elements.
4. After the completion of the conductivity measurement, rinse inside and outside of probe thoroughly with water, preferably distilled water. Allow probe to drain completely before storing. The probe may be stored dry indefinitely without any damage.
GENERAL OPERATING PROCEDURE
1. Plug probe into BNC receptacle (top right corner of panel).
2. With control knob in the “OFF” position, meter pointer must
balance at zero. To adjust the zero position, use meter-zero adjusting screw (“X” slot screw on front of meter face). The mirror-faced scale insures accurate readings by eliminating errors due to parallax problems which would otherwise occur when the indicating needle and the scale are not properly aligned by the reader’s eye.
3. Switch the control knob to the “BAT. CK” position. Meter should indicate greater than 80 micromhos/cm. If not, replace battery. (See instructions for battery replacement, page 8.)
4. Set the “TEMP” knob at 25°C, the Range Selector at “X100,” and switch the control knob to the “CAL” position. The meter should indicate exactly 100 micromhos/cm. If not, carefully adjust by using the “STD” knob.
Con duc tiv ity Sa lin ity
0°C 5°C 10°C 15°C 20°C 25°C 30°C
23,600 27,300 31,100 35,100 39,200 43,500 47,800 28 24,400 28,100 32,100 36,200 40,400 44,800 49,400 29 25,200 29,000 33,100 37,300 41,700 46,200 50,900 30 26,000 30,000 34,100 38,500 43,000 47,600 52,400 31
26,800 30,900 35,100 39,600 44,200 49,000 53,900 32
27,500 31,700 36,100 40,700 45,400 50,300 55,400 33 28,300 32,600 37,100 41,800 46,700 51,700 56,800 34 29,100 33,500 38,100 42,900 47,900 53,000 58,300 35 29,700 34,200 39,000 44,000 49,100 54,400 59,800 36 30,500 35,100 40,000 45,100 50,300 55,700 61,300 37 31,200 36,000 41,000 46,200 51,500 57,100 62,800 38 32,000 36,800 41,900 47,200 52,700 58,400 64,200 39 32,700 37,700 42,900 48,300 53,900 59,700 65,700 40
Data derived from the equation of P.K. Weyl, Limnology and Oceanography; 9,75 (1964).
0
00
CALIBRATION
Although each conductivity meter is precalibrated before it leaves the manufacturer, it may become necessary to check the calibration to determine if the instrument is responding to the specifications, especially if the conductivity probe is different from the original one supplied with the unit. This procedure should only be attempted by qualified personnel. Please read the instructions thoroughly before starting.
Two accurately-prepared standard conductivity solutions are required for the calibration procedure*. These solutions may be prepared with reagent grade potassium chloride (KCl) in the following manner:
Solution #1: 0.05M KCl: In a 1.0 liter volumetric flask, dissolve 3.725g anhydrous KCl in deionized/distilled water and dilute to 1.0 liter mark. At 25°C this solution has a conductivity of 6,668 micromhos/cm.
Solution #2: 0.5M KCl: In a 1.0 liter volumetric flask dissolve 37.25g anhydrous KCl in deionized-distilled water and dilute to 1.0 liter mark. At 25°C this solution has a conductivity of 58,640 micromhos/cm.
4
9
Page 5
CONVERSION TABLE FOR CHANGING CONDUCTIVITY* INTO SALINITY
If the temperature compensation knob was set to the sample temperature when the conductivity reading was recorded, use the 25°C column to receive the results as ppt salinity.
If temperature compensation was not used when you recorded the conductivity reading (the temperature knob was set to 25°C), use the column that reflects the actual sample temperature to receive results as ppt salinity.
CONDUCTIVITY SALINITY
0°C 5°C 10°C 15°C 20°C 25°C 30°C
1,200 1,400 1,500 1,700 2,000 2,200 2,400 1 2,220 2,500 2,900 3,300 3,700 4,100 4,500 2 3,200 3,700 4,200 4,700 5,300 5,900 6,500 3 4,100 4,700 5,400 6,100 6,900 7,600 8,400 4 5,000 5,800 6,600 7,500 8,400 9,300 10,300 5 5,900 6,800 7,900 8,800 9,900 11,000 12,100 6 6,700 7,800 8,900 10,100 11,300 12,600 13,900 7 7,600 8,800 10,100 11,400 12,800 14,200 15,700 8 8,500 9,800 11,200 12,700 14,200 15,800 17,400 9
9,300 10,800 12,300 13,900 15,600 17,300 19,100 10 10,200 11,800 13,400 15,200 17,000 18,900 20,800 11 11,000 12,800 14,500 16,400 18,900 20,400 22,500 12 11,900 13,700 15,600 17,600 19,700 21,900 24,100 13 12,600 14,600 16,700 18,900 21,100 23,400 25,800 14 13,400 15,600 17,800 20,100 22,400 24,900 27,400 15 14,200 16,400 18,800 21,200 23,800 26,400 29,100 16 15,000 17,400 19,800 22,400 25,100 27,800 30,700 17 15,800 18,300 20,900 23,600 26,400 29,300 32,300 18 16,600 14,200 21,900 24,800 27,700 30,700 33,900 19 17,400 20,100 23,000 25,900 29,000 32,200 35,500 20 18,200 21,100 24,000 27,100 30,300 33,600 37,000 21 19,000 22,000 25,100 28,300 31,600 35,000 38,600 22 19,800 22,900 26,100 29,400 32,900 36,500 40,100 23 20,600 23,800 27,100 30,600 34,200 37,900 41,700 24 21,400 24,700 28,100 31,700 35,400 39,300 43,200 25 22,100 25,500 29,100 32,800 36,700 40,700 44,800 26 22,800 26,400 30,100 33,900 37,900 42,100 46,300 27
*Conductivity values are given in micromhos/cm.
8
0
00
5. Measure the temperature of the sample solution, set the “TEMP” knob at the mark indicating the temperature. (Note the markings are in centigrade.) Temperature has a pronounced effect upon conductivity, and must be monitored accurately.
6. Immerse the probe into at least ½" of solution. Make sure all bubbles are removed from the probe.
7. Switch the control knob to the “ON ” position and the range selector to either “X1,” “X10,” “X100,” or “X1000.” Read the conductivity directly from the meter face. For the best accuracy, select the range which gives readings as close to mid-range of the scale as possible. Multiply the reading from the meter face by the appropriate factor. The probe should not be in contact with any surface other than the solution being tested.
8. After completing the conductivity measurement, switch control knob to the “OFF” position. Rinse the probe thoroughly with water, preferably distilled water, and drain.
9. Reference solutions which have known conductivities may be prepared by consulting the latest edition of Standard Methods for the Examination of Water and Wastewater, American Public Health Association, 1015 East Eighteenth Street, NW, Washington, DC
20036. A series of accurately prepared reference solutions in both the high range and low range allow the operator to check on the manipulative procedure as well as the verification of results.
DETERMINATION OF CONDUCTIVITY & TDS IN BOILER WATER
In boiler water where high concentrations of hydroxides may be evident, it is necessary to neutralize the sample solution with gallic acid or lactic acid. Otherwise high conductivities will be noted. For a 100 mL sample add 0.4 grams of gallic acid or lactic acid. This will neutralize approximately 1300 ppm hydroxide alkalinity. If a small amount of gallic acid or lactic acid does not dissolve within the sample solution, the conductivity reading will not be affected. On some highly alkaline waters higher concentrations of gallic acid or lactic acid may be required.
The determination of conductivity for boiler water is made by following the General Operating Procedure given on page 4. To convert the conductivity reading to parts per million Total Dissolved Solids (ppm TDS), multiply the reading by 0.7.
EX AM PLE:
Conductivity reading is 3500 micromhos/cm
3500 x 0.7 = 2450 ppm TDS
5
Page 6
DETERMINATION OF SOIL CONDUCTIVITY & TOTAL DISSOLVED SALTS
An approximation of the Total Dissolved Salts in soil is obtained by making a soil extraction with distilled water and measuring the conductivity of the soil extract. For purposes of comparison, all conductivity values are recorded at 25°C.
NOTE: For determination of Soluble Salts in greenhouse soils, see page 7.
PRO CE DURE
1. Fill a 50 mL beaker with the soil to be tested. Tap it lightly to eliminate any trapped air. Strike the excess soil off the surface.
2. Empty the contents of the beaker into a 250 mL widemouth flask and add 100 mL of distilled water.
3. Stopper and shake vigorously. Allow to stand for thirty minutes. During the thirty minutes waiting period the sample should be shaken vigorously three or four times.
4. Filter the contents of the flask and collect filtrate in suitable conductivity chamber.
5. Take conductivity reading according to method given for General Operating Procedure.
6. To convert to Total Dissolved Salts use the following formula:
micromhos/cm @ 25°C x 0.7 = ppm Total Dissolved Salts
If the Soluble Salts as determined above are greater than 1000 ppm, the chlorides and sulfates should be determined in order to learn whether the Soluble Salts are chlorides or sulfates.
Usually in calcareous soils, the sulfates represent gypsum and cause little injury to plants, while high concentrations of chlorides might result in damage to the plants. Greenhouse and sensitive plants may be damaged if the Souble Salts are over 500 ppm of chlorides. Most other plants will grow well in soils containing less than 1000 ppm chlorides.
DETERMINATION OF SOLUBLE SALTS FOR GREENHOUSE SOILS
The meter is specially designed to read conductivity in the limited volume of soil extract obtained through the soil extraction procedure employed in the LaMotte Model GS-01 Greenhouse Outfit. Use the following formula to convert the conductivity reading to parts per million Soluble Salts:
Conductivity Reading (micromhos/cm) x 0.7 = ppm Soluble Salts
PRO CE DURE
1. With control knob in the “OFF” position, meter pointer must balance at zero. To adjust the zero position, use meter-zero adjusting screw (“X” slot screw on front of meter face). The mirror-faced scale insures accurate readings by eliminating errors due to parallax problems which would otherwise occur when the indicating needle and the scale are not properly aligned by the reader’s eye.
2. Switch the control knob to the “BAT CK” position. Meter should indicate greater than 80 micromhos/cm, if not then replace the battery. (See instructions for battery replacement.)
3. Set the “TEMP” knob at 25°C, the Range Selector at “X100,” and switch the control knob to the “CAL” position. The meter should indicate exactly 100 micromhos/cm, if not then carefully adjust by using the “STD” knob.
4. Measure the temperature of the sample solution, then set the “TEMP” knob at the mark indicating the temperature. (Note the markings are in centigrade.) Temperature has a pronounced effect upon conductivity, and must be monitored accurately.
5. Immerse the probe into soil extract (in special reading chamber) to at least cover the slots at bottom of the probe. Make sure all bubbles are removed from the probe.
6. Switch the control knob to the “ON” position and the range selector to either “X1,” “X10,” “X100,” or “X1000.” Read the conductivity directly from the meter face. For the best accuracy, select the range which gives readings as close to mid-range of the scale as possible. Multiply the reading from the meter face by the appropriate factor. The probe should not be in contact with any surface other than the solution under test.
7. After completing the conductivity measurement, switch control knob to the “OFF” position. Rinse probe thoroughly with water, preferably distilled water, and drain.
8. Do not discard soil extract if it is to be used for other chemical tests.
6
7
Loading...