CHLO RINE TEST KIT
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
ORTHO-TOLIDINE METHOD
MODEL P-30 • CODE 4437
QUANTITY CONTENTS CODE
30 mL *Chlorine Reagent O-Tolidine *4100-G
2 Test Tubes, 5.0 mL, glass, w/caps 0230
1 Chlorine O-T Comparator, 0.2-2.0 ppm
(Code 4437 only)
1 Chlorine O-T Comparator, 0.1-1.0 ppm
(Code 7146 only)
*WARNING: Reagents marked with an * are considered to be potential
health hazards. To view or print a Material Safety Data Sheet (MSDS) for
these reagents go to www.lamotte.com. To obtain a printed copy, contact
LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code
number.
USE OF THE OCTET COM PARA TOR
The Octet Comparator contains eight
permanent color standards. A test sample is
inserted into the openings in the top of the
comparator. The sample can then be compared
to four color standards at once, and the value
read off the comparator. For optimum color
comparison, the comparator should be
positioned between the operator and a light
source, so that the light enters through the
special light-diffusing screen in the back of the
comparator. Avoid viewing the comparator
against direct sunlight or an irregularly lighted
background.
4438
7147
PRO CE DURE
TO TAL RE SID UAL CHLO RINE
1. Rinse test tube (0230) with sample water. Fill to 5 mL line with sample
water.
2. Add 8 drops of *Chlorine Reagent O-Tolidine (4100). Cap and mix. Wait
3 minutes.
3. Insert into the Chlorine O-T Comparator (4438 or 7147). Match sample
color to a color standard. Record as ppm total residual chlorine.
4. If test sample color exceeds the highest color standard, repeat the test on a
sample diluted 1:1 with chlorine-free water. Multiply the comparator values
by 2. Record result as ppm Total Residual Chlorine.
FREE AVAIL ABLE CHLO RINE
1. Chill sample water to approximately 34°F. Rinse test tube (0230) with
sample water. Fill to 5 mL line with sample water.
2. Add 8 drops of *Chlorine Reagent O-Tolidine (4100). Cap and mix. Do
not allow the test sample to stand after reagent has been added.
3. Immediately insert into the Chlorine O-T Comparator (4438 or 7147).
Match sample color to a color standard. Read the test result within 5
seconds. Record as ppm Free Available Chlorine.
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Chal lenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®