LaMotte OXYGEN 3103 User Manual

WA TER SAM PLING BOT TLE
L
Mott
CENTER PLUG
INNER CHAMBER
WEIGHT
COLLECTING BOTTLE
INLET TUBING
BOTTOM BRIDLE
SNAP CLAMP
TOP BRIDLE
FOR DIS SOLVED OX Y GEN
CODE 1054-DO
QUANTITY CONTENTS CODE
1 Water Sampling Body 1054 1 Water Sample Collecting Bottle, 60mL, glass 0688-DO 1 Sounding Weight 1068
IN TRO DUC TION
This device is designed for use in the field and is a simplified water sampler. The sample is collected in a removable inner bottle which will overflow five times to insure a representative sample. Samples may be taken at a controlled depth by using a calibrated line. Attaching the two pound weight to the bottom of the sampling device insures rapid descent and minimizes the amount of drift due to currents. More weight should be attached to the sampling device in strong currents.
It is necessary to maintain a position directly over the Water Sampling Bottle when lowering it so that it remains in an upright position. This permits the displacement of all of the air in the sampler so that it will fill completely.
It should be noted that the bubbles of air displaced from the sampler will be observed downstream.
PRO CE DURE
1.
2.
3.
5.
6.
Remove the plastic center plug with inlet tubing attached.
Insert the collecting bottle (0688-DO), with the cap removed, into the inner chamber of the cylinder.
4.
Attach the two pound weight (1068) to the bottom bridle of the sampler by the snap clamp.
Replace the plastic center plug. Make sure the inlet tubing is in the collecting bottle.
Attach the snap clamp on the calibrated line to the bridle on top of the sampler.
Quickly lower the water sampler to the desired depth and leave until full. This can be determined when the bubbles from the displaced air in the sampler cease to appear. This usually takes 3 to 5 minutes.