Test Kit: - OAKITE ALKALINE TEST KIT (CODE OKT 9205)
Reagents and Equipment:
CodeQuantityDescription
OKT 6130-J100 ml1 N Hydrochloric Acid
OKT 6131-J100 ml0.1 N Hydrochloric Acid
OKT 2246-G25 mlPhenolphthalein
OKT 2207-G25 mlBromcresol Green
OKT 2292-G25 mlMethyl Red
03762Direct Reading Titrator; 0-1 ml
07781Titration Tube marked at 5, 10 and
15 ml, with cap
03541Plastic dropper marked at 1 ml
TK 121
Step 1:Using the 1 ml dropper, pipet the appropriate sample volume of Oakite solution
listed in the table below into the empty titration tube. Add water to this solution
until the volume is at least to the 5 ml mark. For 5 and 10 ml sample volumes,
use the graduations on the titration tube. No water need be added.
Step 2:Add 1 or 2 drops of Phenolphthalein Indicator. Swirl to mix. The color will be
red. Place the cap on the titration tube.
Step 3:Fill the Direct Reading Titrator* to the zero mark with 1N Acid.
Step 4:Insert tip of Titrator into the central hole in the titrating tube cap. While holding
the Titrator assembly in one hand, slowly depress the plunger of the Titrator
with the other hand to dispense the titrating solution. Swirl the solution gently
after each drop has been added. When the red color of the solution disappears
completely, stop any further movement of the plunger. Record the reading on
the Titrator in ml. Call this Titration T. This completes the analysis.
Calculation:T x F = % by vol, g/lor oz/gal of Oakite product.
*NOTE:For complete instructions on how to use the Direct Reading Titrator please
refer to the enclosed instruction sheet.
(con tin ued)
Test Kit procedure 121 continued
Use ConcentrationSample VolumeMultiplying Factor (F)
Oakite Material
2430-90 (4-12)175 (10)
451.5-15 (0.2-2)106.3 (0.83)
623.8-15 (0.5-2)511 (1.4)
7715-60 (2-8)251 (6.8)
9045-90 (6-12)164 (8.5)
947.5-30 (1-4)1013 (1.8)
10015-60 (2-8)224 (3.2)
1450.5-2100.80
16015-45 (2-6)222 (3.0)
1613.8-15 (0.5-2)1015 (2.0)
1627.5-30 (1-4)516 (2.2)
19060-120 (8-16)154 (7.2)
2450.5-2100.95
29530-45 (4-6)242 (5.6)
g/l (oz/gal)
15-90 (2-12)163 (8.3)
% by volin ml% by vol
g/l (oz/gal)
3450.5-2101.7
360 L2-815.7
99045-90 (6-12)154 (7.2)
232730-60 (4-8)242 (5.6)
ALC-1
Alclean NS
Aluminum
Cleaner 164
Anodet
Anoprep
Dynadet
Dynamerse
Enprox 714
Enprox Tow e r
Cleaner
FiSan-500
FiSan-500M
1.9-7.5 (0.25-1)106.6 (0.88)
15-60 (2-8)1030 (4.0)
45-75 (6-10)282 (11)
45-90 (6-12)171 (9.5)
15-60 (2-8)244 (5.9)
22-90 (3-12)242 (5.6)
22-90 (3-12)247 (6.2)
2-816.0
3.8-15 (1.5-2)1015 (2.0)
0.5-251.2
0.5-251.2
FiSan-Bottle
Clean
15-60 (2-8)1020 (2.7)
(con tin ued)
Test Kit Procedure 121 continued
Use ConcentrationSample VolumeMultiplying Factor (F)
Oakite Material
FiSan Brite
FiSan-CQD
Fleetline TC
FM 183
Liqualin
Low Heat
Cleaner 1
Low Heat
Cleaner 3
LRS
Okemclean
Rustripper
Rustripper M
Solventol 530
Solventol 530
Solventol 828
g/l (oz/gal)
% by volin ml% by vol
g/l (oz/gal)
7.5-30 (1-4)224 (3.2)
7.5-30 (1-4)224 (3.2)
60-120 (8-16)172 (9.6)
30-120 (4-16)282 (11)
1-4104.0
22-90266 (8.8)
2-852.8
1-452.5
2.856.5
30-90 (4-12)160 (8.0)
15-45 (2-6)226 (3.5)
7.5-15 (1-2)1015 (1.9)
30-60 (4-8)529 (3.9)
7.5-30 (0.5-2)1010 (1.3)
Solventol 3641
Solventol 3830
Solventol
Actostrip 3929FB
Solventol ES
3899
STC
Stripper M3
Stripper Q5
Stripper Q9
Stripper Q9L
Stripper Q9L
Special
Stripper Q10
Stripper R6
Stripper R8
Super Rustripper
3.8-15 (0.5-2)109.7 (1.3)
3-1017.1
40-60128
45-90 (6-12)248 (6.4)
2-853.0
22-90 (3-12)147 (6.3)
22-90 (3-12)145 (6.0)
22-90 (3-12)152 (7.0)
5-2018.4
5-2018.5
10-40123
30-120 (4-16)148 (6.4)
30-120 (4-16)152 (7.0)
30-120 (4-16)175 (10)
9/1/87
Supersedes 1/22/87Ptd. in U.S.A.
TP/TK 121
·TEST KIT PROCEDURE 122
TITRATION OF OAKITE ALKALINE PRODUCTS WITH N ACID USING
BROMCRESOL GREEN INDICATOR
(THIS PROCEDURE REPLACES PREVIOUS OAKITE PROCEDURES 1 AND 56)
Test Kit: - OAKITE ALKALINE TEST KIT (CODE OKT 9205)
Reagents and Equipment:
CodeQuantityDescription
OKT 6130-J100 ml1 N Hydrochloric Acid
OKT 6131-J100 ml0.1 N Hydrochloric Acid
OKT 2246-G25 mlPhenolphthalein
OKT 2207-G25 mlBromcresol Green
OKT 2292-G25 mlMethyl Red
03762Direct Reading Titrator; 0-1 ml
07781Titration Tube marked at 5, 10 and
15 ml, with cap
03541Plastic dropper marked at 1 ml
TK 122
Step 1:Using the 1 ml dropper, pipet the appropriate sample volume of Oakite solution
listed in the table below into the empty titration tube. Add water to this solution
until the volume is at least to the 5 ml mark. For 5 and 10 ml sample volumes,
use the graduations on the titration tube. No water need be added.
Step 2:Add 1 or 2 drops of Bromcresol Green Indicator. Swirl to mix. The color will be
blue. Place the cap on the titration tube.
Step 3:Fill the Direct Reading Titrator* to the zero mark with 1N Acid.
Step 4:Insert tip of Titrator into the central hole in the titrating tube cap. While holding
the Titrator assembly in one hand, slowly depress the plunger of the Titrator
with the other hand to dispense the titrating solution. Swirl the solution gently
after each drop has been added. When the color of the solution changes from
blue to a distinct yellow, stop any further movement of the plunger. Record the
reading on the Titrator in ml. Call this Titration T. This completes the analysis.
Calculation:T x F = % by vol, g/lor oz/gal of Oakite product.