LaMotte NITROGEN IN SOIL User Manual

NI TRO GEN DI OX IDE IN AIR TEST KIT
L
M
o
tt
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
CODE 7690
QUANTITY CONTENTS CODE
2 x 120 mL *Nitrogen Dioxide #1 Absorbing Solution *7684-J 30 mL *Nitrogen Dioxide Reagent #2 *7685-G 10 g Nitrogen Dioxide Reagent #3 Powder 7688-D 2 Test Tubes, 10 mL, glass, w/caps 0822 1 Spoon, 0.05g, plastic 0696 1 Pipet, dropping, plastic 0352 1 Nitrogen Dioxide in Air Comparator 7689 1 Tubing 23609 1 Pipet 30410 1 Needle 27336-01
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
The sampling period specified in the directions corresponds to concentrations which may be encountered in ambient air conditions. High concentrations of pollutants in the atmosphere will require shorter sampling periods, while low concentrations will require longer sampling periods.
USE OF THE OCTET COM PARA TOR
L
Mott
The Octet Comparator contains eight permanent color standards. A test sample is inserted into the openings in the top of the comparator. The sample can then be compared to four color standards at once, and the value read off the comparator. For optimum color comparison, the comparator should be positioned between the operator and a light source, so that the light enters through the special light-diffusing screen in the back of the comparator. Avoid viewing the comparator against direct sunlight or an irregularly lighted background.
USE OF THE ADAPTER ASSEMBLY
This test kit includes an adapter for restricting the flow of air through the impinging apparatus. The adapter assembly consists of a 27 gauge hypodermic needle (27336-01) which is fitted into a small plastic holder (30410). A piece of plastic tubing (23609) is attached to the plastic holder. By joining one end of the tubing to the intake portion of the impinging apparatus, the flow of air is restricted to 0.2 Lpm. To sample the test atmosphere at this rate for nitrogen dioxide, use the following procedure:
1. Attach adapter to intake of impinging apparatus.
2. Unscrew knob of flowmeter (counterclockwise - 6 complete turns
from closed position of flowmeter).
3. Turn switch to “ON” position. Follow recommended operating procedure for nitrogen dioxide test.
NOTE: The sampling period should not exceed 20 minutes for a single determination of nitrogen dioxide when the adapter is in place as damage might result to the pumping mechanism.