LaMotte MOLYBDENUM KIT User Manual

MOLYBDENUM
1200 COLORIMETER, 0-30 ppm
CODE 3676-01
QUANTITY CONTENTS CODE
60 mL *Mo Buffer *3997-H 2 x 30 mL *Molybdenum Oxidizing Reagent *6485-G
2.5 g *Molybdenum Indicator Powder *6486-S 1 Colorimeter Tubes 10 mL, w/cap, set of 6 0290-6 1 Spoon, 0.05g, plastic 0696 1 Pipets, 1.0 mL, plastic, w/cap 0372 1 Pipet, 1.0 mL 0354 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Molybdenum 26729
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specifi ed code number. NOTE: After use, return *MO Buffer (3997) to zipper top storage bag to reduce
exposure to corrosive reagent.
INTRODUCTION
Molybdenum occurs naturally in the earth’s crust as molybdenite and wolfenite, and is an important element in many biochemical reactions, including nitrogen fi xation. In industrial processes, such as the operation of boilers and cooling towers, molybdenum, in the form of sodium molybdate, is used as an environmentally safe corrosion inhibitor.
MOLYBDENUM PROCEDURE - Thioglycolate Method
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber
MOLYBDENUM
12
Fill theWater Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Rinse and fi ll a colorimeter tube (0290) to the 10mL line with sample water.
34
Use the 1.0 mL pipet (0372) to add1mLof *Mo Buffer (3997). Cap and mix.
Use the 1.0 mL pipet (0354) to add 1mL of *Molybdenum Oxidizing Reagent (6485). Cap and mix.
5
Wipe the tube dry.
7
Push the READ button to turn the meter on. Press the ZERO button and hold it for 2 seconds until is displayed. Release the button to take a blank reading (0.0 ppm).
6
8
Insert the tube into the chamber, being sure to align the index line with the arrow on themeter. Close the lid. This tube is the sample blank.
Use the 0.05 spoon (0696) to add 0.05g of *Molybdenum Indicator Powder (6486).
5
Cap and mix until the powder dissolves. Solution will turn yellow if molybdenum is present.
6
Wipe the tube dry.
7
MOLYBDENUM TEST METHOD SPECIFICATIONS
APPLICATION
Boiler and cooling water
RANGE
0 to 30 ppm Molybdenum
METHOD
Calcium thioglycolate reacts with molybdenum to give a yellow color with an intensity proportional to the amount of molybdenum present.
HANDLING & PRESERVATION
Molybdenum samples may be stored in either plastic or glass containers.
Align the index line with the arrow on the meter, insert tube into chamber. Close the lid. Push the button. Record results as ppm Molybdenum.
NOTE: If the molybdenum reading exceeds 50 ppm, repeat procedure on diluted sample, and multiply the result by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
INTERFERENCES
Nickel levels less than 50 ppm do not interfere; aluminum levels less than 10 ppm do not interfere; chromate, at higher concentrations, interferes due to the intense yellow color. Ferrous iron levels below 50 ppm do not interfere, but low levels of ferric iron will cause a large blank. Highly buffered samples may exceed the capacity of the system possibly producing inaccurate results. Add an extra 1.0 mL of *Mo Buffer (3997) to adjust the pH of the sample to approximately 4.5. Multiply the result in ppm by 1.08.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
®
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside USA) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
63676-01 7/14
Loading...