LaMotte MF-3-0035 User Manual

MAN GA NESE TEST KIT
L
Mott
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
LOW RANGE
MODEL LMN • CODE 3588-01
QUANTITY CONTENTS CODE
60 mL *Hardness Buffer Reagent *4255-H 30 mL *Manganese Indicator Reagent *3956-G 100 mL *Chloroform Reagent *6203-J 15 mL *Metal Inhibitor *2785-E 2 Test Tubes, 6.5 & 10.0 mL, glass w/caps 0893 1 Pipet, 1.0 mL, plastic, w/cap 0372 1 Pipet, 0.5 mL, plastic, w/cap 0369 1 Pipet, 1.0 mL, plastic 0354 1 Low Range Manganese Comparator, 0.05-1.0 ppm 3594
To order individual reagents or test kit components, use the specified code number.
USE OF THE OCTET COM PARA TOR
The Octet Comparator contains eight permanent color standards. A test sample is inserted into the openings in the top of the comparator. The sample can then be compared to four color standards at once, and the value read off the comparator. For optimum color comparison, the comparator should be positioned between the operator and a light source, so that the light enters through the special light-diffusing screen in the back of the comparator. Avoid viewing the comparator against direct sunlight or an irregularly lighted background.
NOTE: The following metals interfere above these levels:
Zinc 2 ppm Nickel 4 ppm Copper 5 ppm
The following metals produce a purple-brown reaction above these levels:
Iron 5 ppm Cobalt 25 ppm
If metals above these levels are encountered, the Cyanide Inhibitor Package (LaMotte Code 7104) must be used.
PRO CE DURE
1. Fill a test tube (0893) to the lower line with sample water.
2. Use 1.0 mL pipet (0372) to add 1.0 mL of *Hardness Buffer Reagent
(4255).
3. Add 2 drops of *Metal Inhibitor (2785). Wait 1 minute.
4. Use 0.5 mL pipet (0369) to add 0.5 mL of *Manganese Indicator
Reagent (3956). Cap and mix. Wait 2 minutes.
5. Use 1.0 mL pipet (0354) to add 2.0 mL (two measures) of *Chloroform Reagent (6203). Cap and mix vigorously for 30 seconds. Allow to stand until the two layers separate.
6. Insert the test tube into the Low Range Manganese Comparator (3594). Match the color of the lower liquid layer with the standards. It may be necessary to raise the tube to compare against the upper set of standards. Record as ppm Manganese.
7. Wash the tubes with soap and water after testing to prevent staining.
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Challenges
PO Box 329 • Chestertown • Mary land • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®