MANGANESE KIT
LOW RANGE COMPARATOR, 0.05-1.0 ppm
CODE 3588-02
QUANTITY CONTENTS CODE
60 mL *Hardness Buffer Reagent *4255-H 30 mL *Manganese Indicator Reagent *3956-G 100 mL *Chloroform Reagent *6203-J 15 mL *Metal Inhibitor *2785-E 2 Test Tubes, 6.5 & 10.0 mL, glass w/caps 0893 1 Pipet, 1.0 mL, plastic, w/cap 0372 1 Pipet, 0.5 mL, plastic, w/cap 0369 1 Pipet, 1.0 mL, plastic 0354 1
Low Range Manganese Octa-Slide 2 Bar, 0.05-1.0 ppm
1 Octa-Slide 2 Viewer 1101
3594-01
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specifi ed code number.
USE OF THE OCTA-SLIDE 2 VIEWER
The Octa-Slide 2 Viewer should be held so non-direct light enters through the back of the Viewer. Insert the reacted sample into the top of the Viewer. Slide the Octa-Slide 2 Bar into the Viewer and match the color of the reaction to the color standards.
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
NOTE: The following metals interfere above these levels: Zinc 2 ppm Nickel 4 ppm Copper 5 ppm The following metals produce a purple-brown reaction above these levels: Iron 5 ppm Cobalt 25 ppm If metals above these levels are encountered, the Cyanide Inhibitor Package (LaMotte
Code 7104-01) must be used.
PROCEDURE
1. Fill a test tube (0893) to the lower line with sample water.
2. Use 1.0 mL pipet (0372) to add 1.0 mL of *Hardness Buffer Reagent (4255).
3. Add 2 drops of *Metal Inhibitor (2785). Wait 1 minute.
4. Use 0.5 mL pipet (0369) to add 0.5 mL of *Manganese Indicator Reagent (3956). Cap and mix. Wait 2 minutes.
5. Use 1.0 mL pipet (0354) to add 2.0 mL (two measures) of *Chloroform Reagent (6203). Cap and mix vigorously for 30 seconds. Allow to stand until the two layers separate.
6. Insert the Low Range Manganese Octa-Slide 2 Bar (3594-01) into the Octa-Slide 2 Viewer (1101).
7. Insert the test tube into Octa-Slide 2 Viewer (1101).
8. Match the color of the lower liquid layer with the standards. It may be necessary to raise the tube to compare against the upper end of the bar. Record as ppm Manganese.
9. Wash the tubes with soap and water after testing to prevent staining.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
63588-02 5/12
®