TO TAL, CAL CIUM AND
MAG NE SIUM HARD NESS TEST KIT
AUTOMA TIC BURET METHOD
MODEL AB-HCM • CODE 7173-01
QUANTITY CONTENTS CODE
100 Hardness Reagent #6 Tablets 4484-J
100 mL *Sodium Hydroxide Reagent
w/Metal Inhibitors
100 Calcium Hardness Indicator Tablets T-5250-J
2 x 250 mL Hardness Titration Reagent 4257-K
30 mL *Hardness Reagent #5 *4483-G
1 Erlenmeyer Flask, 125 mL, glass 0431
1 Graduated Cylinder, 25 mL, glass 0417
1 Pipet, 1.0 mL, plastic 0354
1 Automatic Buret 0427
*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these
reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by
e-mail, phone or fax.
To order individual reagents of test kit components, use the specified code
number.
Read the Automatic Buret Instruction Manual before proceeding.
PRO CE DURE
TOTAL HARDNESS
1. Fill graduated cylinder (0417) to the 25 mL line with sample water.
Transfer to the clean 125 mL Erlenmeyer flask (0431).
2. Add 10 drops of *Hardness Reagent #5 (4483). Mix.
3. Add one Hardness Reagent #6 Tablet (4484). Mix until tablet
disintegrates. Sample should turn red.
*4259-J
4. Fill the Automatic Buret (0427) with the Hardness Titration Reagent
(4257). While gently swirling the flask, add Hardness Titration
Reagent until red color changes to blue. Record the buret reading to
the nearest 0.05 mL.
5. Multiply buret reading by 10. Record as ppm Total Hardness as
CaCO3.
Total Hardness (ppm CaCO3) = 10 x Buret Reading
CALCIUM HARDNESS
1. Fill graduated cylinder (0417) to the 25 mL line with sample water.
Transfer to the clean 125 mL Erlenmeyer flask (0431).
2. Use the 1.0 mL pipet (0354) to add 2.0 mL (two measures) of
*Sodium Hydroxide Reagent w/Metal Inhibitors (4259). Mix.
3. Add one Calcium Hardness Indicator Tablet (T-5250). Mix until
tablet disintegrates.
4. Fill the Automatic Buret with the Hardness Titration Reagent
(4257). While gently swirling flask, add Hardness Titration Reagent
until red color changes to blue. Record the buret reading to the
nearest 0.05 mL.
5. Multiply buret reading by 10. Record as ppm Calcium Hardness as
CaCO3.
Calcium Hardness (ppm CaCO3) = 10 x Buret Reading
To convert to calcium, multiply result by 0.4. Record as ppm
Calcium.
Calcium (ppm Ca) = 0.4 x Calcium Hardness
MAGNESIUM HARDNESS
1. Subtract Calcium Hardness from Total Hardness. Record as ppm
Magnesium Hardness as CaCO3.
Magnesium Hardness (ppm CaCO3) =
Total Hardness - Calcium Hardness
2. To convert to Magnesium, multiply result by 0.24. Record as ppm
Magnesium.
Magnesium (ppm Mg) = 0.24 x Magnesium Hardness
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Challenges
PO Box 329 • Chestertown • Mary land • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
04.05