Overview 78 The Keypad 79 The Display and Menus 79 Negative Results 81 Tubes and Chambers 81
MAINTENANCE
Cleaning 82 Repairs 82 Meter Disposal 82
GENERAL INFORMATION
PACKAGING AND DELIVERY
Experienced packaging personnel at LaMotte Company assure
adequate protection against normal hazards encountered in
transportation of shipments.
After the product leaves LaMotte Company, all responsibility for safe
delivery is assured by the transportation company. Damage claims
must be fi led immediately with the transportation company to receive
compensation for damaged goods.
GENERAL PRECAUTIONS
READ THE INSTRUCTION MANUAL BEFORE ATTEMPTING TO
SET UP OR OPERATE THE METER. Failure to do so could result in
personal injury or damage to the meter. The meter should not be used
or stored in a wet or corrosive environment. Care should be taken to
prevent water from wet tubes from entering the meter chamber.
NEVER PUT WET TUBES IN THE METER.
SAFETY PRECAUTIONS
Read the label on all reagent containers. Some labels include
precautionary notices and fi rst aid information. Certain reagents are
considered potential health hazards and are designated with a * in the
instruction manual. To view or print a Material Safety Data Sheet (MSDS)
for these reagents go to www.lamotte.com. To obtain a printed copy,
contact LaMotte by e-mail, phone or FAX. Additional information for all
LaMotte reagents is available in the United States, Canada, Puerto Rico,
and the US Virgin Islands from Chem-Tel by calling 1-800-255-3924. For
other areas, call 813-248-0585 collect to contact Chem-Tel’s International
access number. Each reagent can be identifi ed by the four-digit number
listed on the upper left corner of the reagent label, in the contents list
and in the test procedures.
4
LIMITS OF LIABILITY
Under no circumstances shall LaMotte Company be liable for loss of life,
property, profi ts, or other damages incurred through the use or misuse
of their products.
SPECIFICATIONS - LTC3000we/wi
Instrument Type:Turbidity: Nephelometer
Color: Colorimeter
Chlorine: Colorimeter
Standard:Turbidity: EPA 180.1, LTC3000we;
ISO 7027, LTC3000wi
Color: Adapted from Standard Methods 2120 B
Chlorine: Standard Methods 4500-Cl-G
Units of Measure: Turbidity: NTU (Nephelometric Turbidity Units)
(LTC3000we only)
FNU (Formazin Nephelometric Units) (LTC3000wi
only)
ASBC (American Society of Brewing Chemists)
EBC (European Brewery Convention)
Color: Platinum Cobalt Color Units (cu)
Chlorine: Parts Per Million (ppm), Milligrams Per
Liter (mg/L)
Range:Turbidity: 0-4000 NTU, 0-4000 FNU,
0-10,500 ASBC, 0-150 EBC
Color: 0–1000 cu
Chlorine: 0.00–10.00 ppm free and total chlorine
Range Selection: Turbidity: Automatic
Color: Automatic
Chlorine: Automatic
Resolution:
(display)
Turbidity: 0.01 NTU/FNU,
0.00–10.99 NTU/FNU Range
0.1 NTU/FNU, 11.0–109.9 NTU/FNU Range
1 NTU/FNU, 110–4000 NTU/FNU Range
Color: 1 cu, 0-1000 cu Range
Chlorine: 0.01 ppm, 0.00–5.00 ppm Range
0.1 ppm, 5.0–10.00 ppm Range
5
Accuracy:Turbidity: From 0-2.5 NTU/FNU the accuracy
is ±0.05 NTU/FNU. From 2.5-100 NTU/FNU
the accuracy is ±2%. Above 100 NTU/FNU the
accuracy is ±3%.
Color: ±15 cu
Chlorine:
TABLET:
0-1.0 ppm Range: ±0.03 ppm
1.0-3.0 ppm Range: ±0.06 ppm
3.0-6.0 ppm Range: ±0.3 ppm
6.0-10.0 ppm Range: ±2.5 ppm
LIQUID:
0-0.5 ppm Range: ±0.03 ppm
0.6-3.0 ppm Range: ±0.06 ppm
3.0-8.0 ppm Range: ±0.4 ppm
8.0-10.0 ppm Range: ±1.5 ppm
Detection Limit:Turbidity: 0.05 NTU/FNU
Color: 20 cu
Chlorine: 0.03 ppm
Light Source:Turbidity: Tungsten lamp 2300°C ±50 °C,
LTC3000we; IR LED 850 nm ±10 nm, spectral
bandwidth 50 nm, LTC3000wi
Color: 428 ±2 nm UV LED
Chlorine: 525 ±2 nm LED
DetectorTurbidity: Photodiode, centered at 90°, maximum
peak 400-600 nm, LTC3000we
Photodiode, centered at 90°, LTC3000wi
Color: Photodiode
Chlorine: Photodiode
Response Time:<2 seconds
Signal
Averaging:
Sample
Chamber:
Sample:10 mL in capped tube
Display:Graphic Liquid Crystal Display
Software:Auto Shut-off: 5, 10, 30 min, disabled
Temperature:Operation: 0–50 °C; Storage: -40–60 °C
Operation
Humidity Range:
Turbidity
Accepts 25 mm fl at-bottomed test tubes
Calibration: Field adjustable, blank and 1 point
Data Logging: 500 points
Chinese, Japanese (Kana)
0–90 % RH, non-condensing
6
Auto Shut-off:5, 10, 30 min, disabled
†
Power Source
:USB wall adapter, USB computer connection or
lithium ion rechargeable battery 2200 mAH, 3.7V
Battery Life:~380 tests (backlight on) to 1000 tests
(backlight off)
(with signal averaging disabled)
Electrical
Provided on nameplate label
Ratings:
Dimensions:(W x L x H) 8.84 x 19.05 x 6.35 cm; 3.5 x 7.5 x 2.2
inches
Weight:362 g, 13 oz (meter only)
USB Interface:mini B
†
CE Mark: The device complies to the product specifi cations for the Low
Voltage Directive.
STATISTICAL & TECHNICAL DEFINITIONS RELATED
TO PRODUCT SPECIFICATIONS
Method Detection Limit (MDL): “The method detection limit (MDL)
is defi ned as the minimum concentration of a substance that can
be measured and reported with 99% confi dence that the analyte
concentration is greater than zero and is determined from analysis of
a sample in a given matrix containing the analyte.”
1
Note that, “As Dr.
William Horwitz once stated, ‘In almost all cases when dealing with
a limit of detection or limit of determination, the primary purpose of
determining that limit is to stay away from it.’”
2
Accuracy: Accuracy is the nearness of a measurement to the accepted
or true value.3 The accuracy can be expressed as a range, about the
true value, in which a measurement occurs (i.e. ±0.5 ppm). It can also
be expressed as the % recovery of a known amount of analyte in a
determination of the analyte (i.e. 103.5 %).
Resolution: Resolution is the smallest discernible difference between
any two measurements that can be made.
4
For meters this is usually
how many decimal places are displayed. (i.e. 0.01). Note that the
resolution many change with concentration or range. In some cases
the resolution may be less than the smallest interval, if it is possible to
make a reading that falls between calibration marks. A word of caution,
that resolution has very little relationship to accuracy or precision. The
resolution will always be less than the accuracy or precision but it is
not a statistical measure of how well a method of analysis works. The
resolution can be very, very good and the accuracy and precision can
be very bad! This is not a useful measure of the performance of a test
method.
Repeatability: Repeatability is the within-run precision.
5
A run is a
single data set, from set up to clean up. Generally, one run occurs
on one day. However, for meter calibrations, a single calibration is
considered a single run or data set, even though it may take 2 or 3 days.
7
Reproducibility: Reproducibility is the between-run precision.
6
Detection Limit (DL): The detection limit (DL) for the LTC3000we/wi
is defi ned as the minimum value or concentration that can be
determined by the meter, which is greater than zero, independent of
matrix, glassware, and other sample handling sources of error. It is the
detection limit for the optical system of the meter.
1
CFR 40, part 136, appendix B
2
Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and L
Vanatta, American Laboratory, Sept 2003, P. 31.
3
Skoog, D.A., West, D. M., Fundamental of Analytical Chemistry, 2nd ed.,
Holt Rinehart and Winston, Inc, 1969, p. 26.
4
Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and L
Vanatta, American Laboratory, Sept 2003, P. 34.
5
Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s Textbook of
Quantitative Chemical Analysis, 5
th
ed., Longman Scientifi c & Technical,
1989, p. 130.
6
Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s Textbook of Quantitative Chemical Analysis, 5th ed., Longman Scientifi c & Technical,
1989, p. 130
Grade Tablets, 100
Water Sample Bottle, 60 mL06880688
Tubes, 6——
Cable, USB, 3 ft.17201720
USB Wall Adapter17211721
88
Accessories
CodeDescription
1452100 NTU Standard, 60 mL (EPA)
1455100 NTU Standard, 60 mL (ISO)
6197-L*Chlorine DPD #1; Instrument Grade Tablets, 500
6197-M*Chlorine DPD #1; Instrument Grade Tablets, 1,000
6903-L*Chlorine DPD #3; Instrument Grade Tablets, 500
6903-M*Chlorine DPD #3; Instrument Grade Tablets, 1,000
P-6740-G*DPD #1A, Free Chlorine Reagent, 30 mL
P-6740-H*DPD #1A, Free Chlorine Reagent, 60 mL
P-6741-G*DPD #1B, Free Chlorine Reagent, 30 mL
P-6741-H*DPD #1B, Free Chlorine Reagent, 60 mL
P-6743-G*DPD #3, Total Chlorine Reagent, 30 mL
P-6743-H*DPD #3, Total Chlorine Reagent, 60 mL
0290-6Tubes, Code 0290, Set of 6
3-0038Replacement Chamber
6973-HChlorine Standard, 250 ppm, 60 mL
3176-01Chlorine Titration Kit, 0-10 ppm
4140-01Chlorine Secondary Standards, set of 4
6058-HColor Standard, 500 Color Units, 60 mL
4185Turbidity-Free Water Kit
2-2097Filters, 0.1 micron, Pack of 50
1901-CDSMARTLink 3 Software
*WARNING: Reagents marked with an * are considered to be potential
health hazards. To view or print a Material Safety Data Sheet (MSDS)
for these reagents go to www.lamotte.com. To obtain a printed copy,
contact LaMotte by e-mail, phone or fax.
EPA COMPLIANCE
The LTC3000we meter meets or exceeds EPA design specifi cations for
NPDWR and NPDES turbidity monitoring programs as specifi ed by the
USEPA method 180.1.
ISO Compliance
This LTC3000wi meter meets or exceeds ISO design criteria for
quantitative methods of turbidity using optical turbidimeters as specifi ed
by ISO 7027.
99
CE COMPLIANCE
The LTC3000we and LTC3000wi meters have been independently
tested and have earned the European CE Mark of compliance for
electromagnetic compatibility and safety. To view certifi cates of
compliance, go to the LaMotte website at www.lamotte.com.
NOTE: The device complies to the product specifi cations for the Low
Voltage Directive.
WARRANTY
LaMotte Company warrants this instrument to be free of defects in
parts and workmanship for 2 years from the date of shipment. If it
should become necessary to return the instrument for service during or
beyond the warranty period, contact our Technical Service Department
at 1-800-344-3100 for a return authorization number or visit
www.lamotte.com for troubleshooting help. The sender is responsible
for shipping charges, freight, insurance and proper packaging to
prevent damage in transit. This warranty does not apply to defects
resulting from action of the user such as misuse, improper wiring,
operation outside of specifi cation, improper maintenance or repair, or
unauthorized modifi cation. LaMotte Company specifi cally disclaims any
implied warranties or merchantability or fi tness for a specifi c purpose
and will not be liable for any direct, indirect, incidental or consequential
damages. LaMotte Company’s total liability is limited to repair or
replacement of the product. The warranty set forth above is inclusive
and no other warranty, whether written or oral, is expressed or implied.
REGISTER YOUR METER
To register your meter with the LaMotte Service Department, go to
www.lamotte.com and choose SUPPORT on th top navigation bar.
COMPUTER CONNECTION
PC LINK
The LTC3000we/wi may be interfaced with any Windows-based
computer by using the LaMotte SMARTLink 3 Program and USB Cable.
The program will store test information and results in a database.
To transfer data from the meter to a computer, plug the smaller end of
the USB cable (USB mini B connector) into the meter and the larger
end of the USB cable (USB Type A connector) into a USB port on a
computer. The LTC3000we/wi will send the following data: test name,
wavelength, concentration, transmittance, absorbance, sample, blank,
time of test, and date of test.
OUTPUT
USB
COMPUTER CONNECTION
USB Type A, USB mini B, Order Cable Code 1720.
10
BATTERY/AC OPERATION
The LTC3000we/wi may be operated on battery power or using a USB
computer/ wall adapter. If using the meter as a bench top unit, use the
AC wall adapter if possible to extend the battery life.
To charge the battery with the wall adapter, plug the smaller end of the
USB cable (USB mini B connector) into the meter and the larger end of
the USB cable (USB Type A connector) into the wall adapter. Plug the
wall adapter into an AC outlet.
To charge the battery from a computer, plug the smaller end of the USB
cable (USB mini B connector) into the meter and the larger end of the
USB cable (USB Type A connector) into a USB port on the computer..
The battery icon will show no bars and fl ash when the unit fi rst turns on.
Then the indicator will indicate the battery status by showing 0, 1, 2, 3 or
4 bars.
It will take 5 hours to fully charge a low battery. The battery icon will
fl ash when the battery is charging. The battery icon will show four bars
and stop fl ashing when it is fully charged. The charging circuit will
automatically switch to a fl oat charge when the battery is fully charged.
The charger may remain connected. Some computers will NOT supply
power to their USB ports during standby operation. The wall charger will
charge the unit continuously.
The battery icon will show no bars and continuously fl ash if the battery
is getting low but the unit will still operate normally. A “Low Battery”
message on the status bar of the display will replace the time when the
battery voltage is too low for proper operation and accuracy may be
degraded. A “Shutdown Low Batt” message on the display will appear
for a few seconds before the power is switched off when the battery is
too low to operate the unit.
To extend the battery life:
• Shut down the unit with the power switch when not taking
measurements or use the power save option to have the unit
automatically turn off after 5 minutes.
• Store the unit in a cool dry place.
• Fully charge the battery before storing the unit for extended
periods of time.
• Limit backlight use. The unit consumes 3X normal power with
the backlight on.
Set the backlight time option to 10 seconds, or select “Button Control”
and keep the backlight off.
Battery replacement: The lithium-ion battery used in this unit should
last for many years with normal use. When it no longer powers the unit
long enough to meet testing requirements it will need to be replaced.
Lithium-ion batteries that are properly charged and stored do not usually
11
lose all capacity; they just have less capacity after hundreds of charge
cycles. This unit uses a custom battery assembly that is only available
from LaMotte Company. Battery replacement must be performed at a
LaMotte authorized repair facility. Contact LaMotte Company by phone
(1-800-344-3100) or email (tech@lamotte.com) for a return authorization
number.
INTRODUCTION
TURBIDITY
WHAT IS TURBIDITY?
Turbidity is an aggregate property of the solution, which is water in most
cases. Turbidity is not specifi c to the types of particles in the water.
The particles could be suspended or colloidal matter, and they can
be inorganic, organic, or biological. At high concentrations, turbidity
is perceived as cloudiness, haze, or an absence of clarity in the water.
Turbidity is an optical property that results when light passing through
a liquid sample is scattered. The scattering of light results in a change
in the direction of the light passing through the liquid. This is most
often caused when the light strikes particles in solution and is scattered
backward, sideways and forward. If the turbidity is low, much of the light
will continue in the original direction. Light scattered by the particles
allows the particle to be ”seen” or detected in solution, just as sunlight
allows dust particles in the air to be seen.
In the past 10 years, turbidity has become more than just a measure
of water clarity. Because of the emergence of pathogens such as
Cryptosporidium and Giardia, turbidity now holds the key to assuring
proper water fi ltration. In 1998, the EPA published the IESWTR (interim
enhanced surface water treatment rule) mandating turbidities in
combined fi lter effl uent to read at or below 0.3 NTU. By doing so, the
EPA hoped to achieve a 2 log (99%) removal of Cryptosporidium. There
is presently consideration to lower this to 0.1 NTU. The trend has been
to check the calibration of on-line turbidimeters with hand-held fi eld
units. The optical design and low detection limit of the LTC3000we/wi
allows very accurate readings for such calibrations.
The meter also allows the user to choose the units of measure for
expressing turbidity. While nephelometric turbidity unit (NTU) has
been the standard for years, FNU (formazin nephelometric unit) and
FAU (formazin attenuation unit) are now being used in ISO 7027 units.
American Society of Brewing Chemists (ASBC) units and European
Brewery Convention (EBC) units allow the brewing industry to check
process waters.
HOW IS TURBIDITY MEASURED?
Turbidity is measured by detecting and quantifying the scattering of light
in water (solution). Turbidity can be measured in many ways. There are
visual methods and instrumental methods. Visual methods are more
suitable for samples with high turbidity. Instrumental methods can be
used on samples with both high and low levels of turbidity.
12
Two visual methods are the Secchi Disk method and the Jackson
Candle method. The Secchi Disk method is often used in natural waters.
A black and white Secchi Disk is lowered into the water until it can no
longer be seen. It is then raised until it can be seen again. The average
of these two distances is known as the “Secchi Depth”. The Jackson
Candle method uses a long glass tube over a standard candle. Water
is added or removed from the tube until the candle fl ame becomes
indistinct. The depth of the water measured with a calibrated scale is
reported as Jackson Turbidity Units (JTU). The lowest turbidity that
can be determined with this method is about 25 NTU. There are two
common methods for instruments to measure turbidity. Instruments
can measure the attenuation of a light beam passing through a
sample and they can measure the scattered light from a light beam
passing through a sample. In the attenuation method, the intensity
of a light beam passing through a turbid sample is compared with
the intensity passing through a turbidity-free sample at 180° from the
light source. This method is good for highly turbid samples. The most
common instrument for measuring scattered light in a water sample is
a nephelometer. A nephelometer measures light scattered at 90° to the
light beam. Light scattered at other angles may also be measured, but
the 90° angle defi nes a nephelometric measurement. The light source
for nephelometric measurements can be one of two types to meet EPA
or ISO specifi cations. The EPA specifi es a tungsten lamp with a color
temperature of 2,200–3,000 K. The units of measurement for the EPA
method are nephelometric turbidity units (NTU). The ISO specifi es a light
emitting diode (LED) with a wavelength of 860 ± 30 nm and a spectral
bandwidth less than or equal to 60 nm. The units of measurement for the
ISO method are formazin nephelometric units (FNU). The LTC3000we
meets the EPA specifi cation and the LTC3000wi meets the ISO
specifi cation. The nephelometric method is most useful for low turbidity.
The LTC3000we/wi is a nephelometer that is capable of measuring
turbidity by both the attenuation method and the nephelometric method.
It uses a detector placed at 180° to the light source for high turbidity
samples. It uses a detector placed at 90° to the light source for the
nephelometric method for low turbidity samples. The LTC3000we/wi
has a signal averaging option to improve the stability of readings on low
turbidity samples.
The LTC3000we/wi has two different turbidity calibrations, formazin and
Japan Standard. The formazin calibration is the EPA and ISO approved
method of calibrating nephelometers. This calibration can be used with
user prepared formazin standards or commercially purchased formazin
standards. LaMotte Company approved AMCO
TM
standards labeled
for use with the LTC3000we/wi can also be used with the formazin
calibration. Stablcal® standards below 50 NTU should not be used to
calibrate the LTC3000we/wi.
The Japan Standard calibration is a calibration for a Japanese Water
Works standard. It is based on Japanese formulated polystyrene
turbidity standards. This calibration should only be used to meet
13
Japanese Water Works requirements. The Japanese polystyrene
standards can only be purchased in Japan. Formazin, AMCO and
Stablcal® standards cannot be used with this calibration.
TURBIDITY UNITS
Traditionally, turbidimeters designed for use in the United State were
made to the specifi cations of EPA Method 180.1. This method defi ned
the NTU, nephelometric turbidity unit, as a unit to measure turbidity
in the range of 0 – 40 NTU using a nephelometer. According to the
EPA a nephelometer was a turbidimeter that measured turbidity with a
90° detector. Also, if the turbidity was greater than 40 NTU, a dilution
was necessary to bring the sample into the 0 – 40 NTU range. Today,
many turbidimeters have additional detectors which increase the range
of the turbidity measurement, eliminate interferences and generally
improve the performance. Currently, many turbidimeters are capable of
measuring above 40 NTU by using detectors other than a 90° detector.
Even though they may use a 180° detector to measure the light that is
attenuated by high turbidity samples they may continue to report the
results as NTU.
Recently there has been an effort to use the units of turbidity
measurements to indicate which type of detector and light source was
used. For EPA compliant meters, measurements made with a 90° degree
detector and an incandescent white light source are reported as NTU.
When an attenuation measurement is made with a 180°detector, using
the same meter, the results are reported as AU, attenuation units. ISO
Method 7027, which specifi es a 860 nm light source, also uses two
turbidity units. When the 90° degree detector is used, the results are
reported as FNU, formazin nephelometric units. When an attenuation
measurement made with a 180°detector, the results are reported as
FAU, formazin attenuation units. It should be noted that all units are
numerically equivalent if the meters are calibrated to formazin and
that the units only designate which detector was used to make the
measurement. For example, 1 NTU = 1 AU = 1FNU = 1FAU.
14
Acronyms Defi nitionsNotes
NTUNephelometric
Turbidity Units
FNUFormazin
Nephelometric
Units
AUAttenuation
Units
FAUFormazin
Attenuation
Units
ASBCAmerican
Society of
Brewing
Chemists
EBCEuropean
Brewery
Convention
Incandescent white
light between 400 and
600 nm, 90° detection,
LTC3000we
IR LED (usually) 860
nm, bandwidth less than
60 nm, 90° detection,
LTC3000wi
Incandescent white
light between 400 and
600 nm, 180° detection,
LTC3000we
IR LED (usually) 860
nm, bandwidth less than
60 nm, 180° detection,
LTC3000wi
LTC3000we/wiNot
LTC3000we/wiNot
Regulatory
Method
EPA 180.1
ISO 7027
Not
applicable
ISO 7027
applicable
applicable
The LTC3000we and LTC3000wi each use two turbidity units. The
LTC3000we reports the result as NTU when the 90° detector is used and
AU when the 180° detector is used. The LTC3000wi reports the result as
FNU when the 90° detector is used and FAU when the 180° detector is
used.
TAKING TURBIDITY WATER SAMPLES
Clean plastic or glass containers may be used for turbidity samples.
Ideally, samples should be tested soon after collection and at the same
temperature as when collected.
15
CHLORINE
WHAT IS CHLORINE
Chlorine is added to water systems to sanitize the water. There are
various forms of chlorine that are added to water. These can be gas,
liquid (commonly called bleach or sodium hypochlorite), calcium
hypochlorite mixtures, stabilized chlorine products and chlorine
generated from salt. When these forms of chlorine are added, they
react with water to form free chlorine, hypochlorous acid. If free chlorine
reacts with ammonia, it will form various types of combined chlorine
(chloramines). Depending on the chlorine to ammonia ratio, these can
be mono, di or tri chloramines.
Because free chlorine can react with precursors in the water to form
carcinogenic trihalomethanes (THMs), many water systems have
switched to chloramines. In these systems, free chlorine and ammonia
are added together and controlled to form monochloramine. Although
not as active a sanitizer as free chlorine, chloramine is less likely to form
THMs. Since it is a slower sanitizer, the concentration of chloramine
in water is higher than the concentration of free chlorine in water
distribution systems.
The present EPA limit of chlorine in water systems is 4.0 ppm. The
amount of chlorine used to process waste may be higher than this.
Many states also establish limits on the amount of chlorine that can be
discharged into a body of water after waste processing. These usually
are less than 0.1 ppm. The low detection limit of the LTC3000we/
wi makes it ideal for such measurements. Because of its wide range,
the LTC3000we/wi can be used to measure the water used in the
wastewater process, in a distribution system and for many low level
discharge requirements.
HOW IS CHLORINE MEASURED?
The most common methods for measuring chlorine are colorimetric
methods. In colorimetric methods, chlorine reacts with reagents added
to a water sample. The reaction of the chlorine with the reagents
produces a color. The intensity of the color produced is proportional to
the concentration of chlorine in the sample. The intensity of the color
can be measured by visual comparison with a calibrated color chart or
other types of visual color comparators. Visual methods suffer due to the
subjective observations of the person judging the colors.
The LTC3000we/wi uses EPA approved DPD reagents to react with
chlorine. In the absence of iodide, free available chlorine reacts instantly
with DPD to produce a pink color. Subsequent addition of potassium
iodide (DPD 3) causes a reaction with the combined form of chlorine.
The LTC3000we/wi electronically measures the color produced in these
reactions in comparison to a colorless water sample. First it measures
the intensity of a light beam passing through a clear colorless sample,
the blank. Then it measures the intensity of light passing through the
pink reacted sample. The LTC3000we/wi uses the ratio of these two
measurements to calculate the concentration of chlorine and displays
the result. The LTC3000we/wi uses the EPA approved wavelength of
525 nm to make these measurements.
16
TAKING CHLORINE WATER SAMPLES
Chlorine solutions are not stable and should be analyzed immediately.
Samples may be collected in glass. Amber or opaque bottles are
recommended since exposure to sunlight or agitation will decrease
chlorine concentrations. Since agitation will also decrease chlorine
concentrations, it is best to fi ll bottles completely to assure there is no air
space in the container. If sampling from a tap, allow the water to run for
a minute to assure a representative sample.
COLOR
WHAT IS COLOR?
Many different dissolved or suspended materials contribute to the color
of water. These can include industrial wastes, plant materials, metals
and plankton. There are two terms used to defi ne color. If one examines
a water sample straight from a water source, the color of the water is
its apparent color. The color of the water without the contribution of
suspended substances is called true color. True color can decrease after
precipitation and increase in drier weather.
Some bodies of water can change color quickly, depending on the
runoff conditions and plant life around them. Wind can also stir up
substances more in shallower bodies of water causing quick color
change. Major contributors are tannins, hemic acids and inorganic
minerals. Color can be critical, since as the color increases, the amount
of light that penetrates the water decreases, and thus submerged plant
life, that depend on this light for photosynthesis, will decrease.
HOW IS COLOR MEASURED?
Since most natural waters have color that is similar to a solution
of chloroplatinate and cobalt, the APHA specifi es the use of dilute
chloroplatinate/cobalt color standards to defi ne color values. In the
APHA method, the color of a water sample is compared visually to 6
to 9 chloroplatinate/cobalt standards. However, visual methods suffer
due to the subjective observations of the person judging the colors. To
eliminate this source of error, color can be measured electronically with
a spectrophotometer or a colorimeter like the LTC3000we/wi.
The LTC3000we/wi is calibrated with APHA color standards at 428 nm.
The meter electronically measures color in comparison to a colorless
water sample. First it measures the intensity of a light beam passing
through a clear colorless sample, the blank. Then it measures the
intensity of light passing through the colored sample. The LTC3000we/wi
uses the ratio of these two measurements to calculate the color and
displays the result. The results are expressed in APHA color units (cu).
There is no standard wavelength for measuring color, and it is common
for meters to use different wavelengths. Since chloroplatinate/cobalt
standards will have different absorbance values at various wavelengths,
comparing results from the LTC3000we/wi to meters using wavelengths
other than 428 nm is not valid.
Meters using different wavelengths will only give the same reading
when measuring chloroplatinate/cobalt standards since they are
both calibrated to those standards. When measuring natural water,
17
meters using different wavelengths should not be expected to give
the same result because the absorbance spectrum of natural water is
usually not identical to the absorbance spectrum of chloroplatinate/
cobalt standards. The reading that the meter displays is a correlation
between the color of the sample water and the color standards at a fi xed
wavelength. The correlation and reading will change as the wavelength
changes.
TAKING COLOR WATER SAMPLES
Samples should ideally be collected in glass containers. Perform the
analysis soon after sampling since the color may change with time. For
true color determinations, remove turbidity by fi ltration or centrifugation.
SAMPLE DILUTION TECHNIQUES
If a test result is out of the range of the meter, it must be diluted. The
test should then be repeated on the diluted sample. The following table
gives quick reference guidelines for dilutions of various proportions.
Amount of Sample Deionized Water to Bring
Final Volume to 10 mL
10 mL0 mL1
5 mL5 mL2
2.5 mL7.5 mL4
1 mL9 mL10
0.5 mL9.5 mL20
All dilutions are based on a fi nal volume of 10 mL, so several dilutions
will require small volumes of the water sample. Graduated pipets should
be used for all dilutions. If volumetric glassware is not available, dilutions
can be made with the colorimeter tube. Fill the tube to the 10 mL line
with the sample and then transfer it to another container. Add 10 mL
volumes of deionized water to the container and mix. Transfer 10 mL of
the diluted sample to the colorimeter tube and follow the test procedure.
Repeat the dilution and testing procedures until the result falls within
the range of the calibration. Multiply the test result by the dilution factor.
For example, if 10 mL of the sample water is diluted with three 10 mL
volumes of deionized water, the dilution factor is four. The test result of
the diluted sample should be multiplied by four.
Multiplication Factor
18
OPTIONS & SET UP
FACTORY DEFAULT SETTINGS
Settings that have user options have been set at the factory to default
settings.
The factory default settings are:
AveragingDisabled
Turbidity UnitsNTU/FNU
Turbidity CalibrationFormazin
Chlorine Unitsppm
Chlorine CalibrationTablet
Date FormatMM-DD-YYYY
Power Save5 minutes
Backlight10 seconds
LanguageEnglish
AVERAGING
The averaging option allows the user to average multiple readings. This
option will improve the accuracy of samples with readings that may tend
to drift with time. When the two, fi ve or ten measurement option has
been selected the fi nal average is displayed. The averaging option is
available only for turbidity. The default setting is disabled. To change the
setting:
Options/Set Up
1. Press and briefl y hold
to turn the meter on. The
LaMotte logo screen will
appear for about 3 seconds
and the Main Menu will
appear.
2. Press to scroll to
Options.
Main Menu
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
Main Menu
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
19
3. Press
Options.
Options/Set Up
ENTER
to select
Options Menu
Averaging
Turbidity Options
Chlorine Options
Set Clock
12:00:00 001/500
4. Press
ENTER
to select
Averaging.
5. Press or to scroll
to the desired option.
6. Press
ENTER
to save the
selection. The screen will
display Storing... for about
1 second and return to the
Options Menu.
Averaging Menu
Disabled
2 Measurements
5 Measurements
10 Measurements
12:00:00 001/500
Averaging Menu
Disabled
2 Measurements
5 Measurements
10 Measurements
12:00:00 001/500
Options Menu
Averaging
Turbidity Options
Chlorine Options
Set Clock
12:00:00 001/500
NOTE: When the Averaging option is enabled, more time will be
required to display a reading and more power will be used.
20
TURBIDITY
The default units are NTU and FNU and the default calibration curve is
formazin. NTU will be used in this example. To change the settings:
SELECTING TURBIDITY UNITS
1. Press and briefl y hold
to turn the meter on. The
LaMotte logo screen will
appear for about 3 seconds
and the Main Menu will
appear.
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
Main Menu
Options/Set Up
2. Press to scroll to
Options.
3. Press
ENTER
to select
Options. Press to scroll
to Turbidity Options.
4. Press
ENTER
to select
Turbidity Options.
Main Menu
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
Options Menu
Averaging
Turbidity Options
Chlorine Options
Set Clock
12:00:00 001/500
Turbidity Options
Turbidity Units
Turbidity Calibration
12:00:00 001/500
21
5. Press
Turbidity Units.
ENTER
to select
NTU
Set Turbidity Units
ASBC
Options/Set Up
EBC
12:00:00 001/500
Available units are:
NTU (Nephelometric Turbidity Units)( LTC3000we only)
FNU (Formazin Nephelometric Units) (LTC3000wi only)
ASBC (American Society of Brewing Chemists)
EBC (European Brewery Convention)
NOTE: The meter will automatically switch to the attenuation mode
above 600 NTU or FNU. In Attenuation mode, measurements will be
made with the 180° detector only, as indicated by AU (attenuation units)
or FAU (formazin anttenuation units) on the display. AU and FAU are
numberically equivalent to NTU. See page 15 for defi nitions of turbidity
units.
6. Press or to scroll
to the desired units.
7. Press
ENTER
to save the
selection. The screen will
display Storing... for about
1 second and return to the
Turbidity Options menu.
Press
EXIT
to return to a
previous menu.
Set Turbidity Units
NTU
ASBC
EBC
12:00:00 001/500
Turbidity Options
Turbidity Units
Turbidity Calibration
12:00:00 001/500
22
Selecting a Turbidity Calibration Curve
1. Press and briefl y hold
to turn the meter on. The
LaMotte logo screen will
appear for about 3 seconds
and the Main Menu will
appear.
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
Options/Set Up
Main Menu
2. Press to scroll to
Options.
3. Press
ENTER
to select
Options. Press to scroll
to Turbidity Options.
4. Press
ENTER
to select
Turbidity Options.
Main Menu
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
Options Menu
Averaging
Turbidity Options
Chlorine Options
Set Clock
12:00:00 001/500
Turbidity Options
Turbidity Units
Turbidity Calibration
12:00:00 001/500
5. Press to scroll to
TurbidityCalibration.
Turbidity Options
Turbidity Units
Turbidity Calibration
12:00:00 001/500
23
6. Press
Turbidity Calibration.
Options/Set Up
ENTER
to select
Turbidity Calibration
Formazin
Japan Standard
12:00:00 001/500
7. Scroll to the desired
calibration option. Select
a calibration option based
on the composition of the
Formazin
Japan Standard
Turbidity Calibration
standards that will be used to
calibrate the meter.
12:00:00 001/500
NOTE: Stablcal® standards below 50 NTU should not be used to
calibrate the LTC3000we/wi. The diluent has a different refractive index
than traditional formazin standards and will affect the results.
8. Press
ENTER
to save the
selection. The screen will
display Storing... for about
1 second and return to the
Turbidity Units
Turbidity Calibration
Turbidity Options
Turbidity Options menu.
Press
previous menu.
EXIT
to return to a
12:00:00 001/500
24
CHLORINE
The default units are ppm and the default calibration curve is for DPD
Tablet reagents. To change the setting:
SELECTING CHLORINE UNITS
1. Press and briefl y hold
to turn the meter on. The
LaMotte logo screen will
appear for about 3 seconds
and the Main Menu will
appear.
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
Main Menu
Options/Set Up
2. Press to scroll to
Options.
3. Press
Options. Press
ENTER
to select
to scroll
to Chlorine Options.
4. Press
ENTER
to select
Chlorine Options.
Main Menu
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
Options Menu
Averaging
Turbidity Options
Chlorine Options
Set Clock
12:00:00 001/500
Chlorine Options
Chlorine Units
Chlorine Calibration
12:00:00 001/500
25
5. Press
Chlorine Units. Available
units are: ppm (parts per
million) and mg/L (milligrams
Options/Set Up
per liter).
ENTER
to select
Set Chlorine Units
ppm
mg/L
12:00:00 001/500
6. Press or to scroll
to desired units.
7. Press
ENTER
to save
selection. The screen will
display Storing... for about
1 second and return to the
Chlorine Options menu.
Press
EXIT
to return to a
previous menu.
Set Chlorine Units
ppm
mg/L
12:00:00 001/500
Chlorine Options
Chlorine Units
Chlorine Calibrations
12:00:00 001/500
26
Loading...
+ 58 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.