CHLO RINE DI OX IDE TEST KIT
LOW RANGE METHOD
MODEL LP-64 • CODE 3592
QUANTITY CONTENTS CODE
50 *Chlorine DPD #1 Tablets *6903-H
15 mL Glycine Solution 6811-E
4 Test Tubes, w/caps 1085
1 Distilled Water Ampoule, 5 mL 2748
1 Axial Reader 2071
1 Stirring Rod, plastic 0792
1 Chlorine Dioxide Comparator,
0.02-0.30 ppm
1 Pipet, plastic, dropping 0352
*WARNING: Reagents marked with a * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents
see MSDS CD or www.lamotte.com. To obtain a printed copy, contact LaMotte by
e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code
numbers.
3593
(over)
PRO CE DURE
1. Use the pipet (0352) to add 5 drops of sample water to one of the test
tubes (1085).
2. Add one *Chlorine DPD #1 Tablet (6903) to this tube. Crush
thoroughly with the stirring rod (0792).
3. Fill the other three test tubes to the line with sample water.
4. Place two of these tubes in position as blanks in the axial reader
(2071) following the instructions in the manual. Place the distilled
water ampoule in position in the comparator.
5. Add 8 drops of Glycine Solution (6811) to the third tube full of
sample. Cap and mix.
6. Pour the treated sample from this tube into the tube in step 2
containing the crushed tablet. Cap and mix.
7. Remove cap and place tube in sample position in the axial reader.
8. Following the directions in the axial reader manual, read the test
result in ppm Chlorine Dioxide (ClO2).