LaMotte LP-1-NMC User Manual

DPD CHLO RINE TEST KIT
L
Mott
L
Mott
MODEL LP-1-NMC •CODE 6901-NMC
QUANTITY CONTENTS CODE
50 *Chlorine DPD #1R Tablets (*6999A) FMC6905A-
6999ABOX 50 Chlorine DPD #3R Tablets (*6905A) 50 *Chlorine DPD #2R Tablets (6904A) FMC6904A-
6906ABOX 50 *Chlorine DPD #4 Instrument Grade Tablets
(*6906A) 3 Test Tubes, plastic, w/caps 0106 1 Chlorine DPD Comparator, 0.1-1.0 ppm 6978
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www.lamotte.com. To obtain a printed copy, contact Lamotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
USE OF THE OCTET COM PARA TOR
The Octet Comparator contains eight permanent color standards. A test sample is inserted into the openings in the top of the comparator. The sample can then be compared to four color standards at once, and the value read off the comparator. For optimum color comparison, the comparator should be positioned between the operator and a light source, so that the light enters through the special light-diffusing screen in the back of the comparator. Avoid viewing the comparator against direct sunlight or an irregularly lighted background.
PRO CE DURE
FREE AVAILABLE CHLORINE
1. Rinse the test tube (0106) with sample water. Fill to 10 mL line with
sample water.
2. Add one *Chlorine DPD #1R Tablet (6999A). Cap and mix until
tablet disintegrates.
3. Immediately insert test tube into Chlorine DPD Comparator (6978).
Match sample color to a color standard. Record as Reading #1. Do not discard this sample if testing for Monochloramine, Dichloramine or Total Residual Chlorine.
NOTE: Reading should be made within one minute of adding reagent.
Reading #1 = ppm Free Available Chlorine
MONOCHLORAMINE
4. Add one Chlorine DPD #2R Tablet (6904A) to sample from Step 3.
Cap and mix until tablet disintegrates.
5. Insert test tube into Chlorine DPD Comparator (6978). Match
sample color to a color standard. Record as Reading #2.
Reading #2 - Reading #1 = ppm Monochloramine
DICHLORAMINE AND TOTAL RESIDUAL CHLORINE
6. Add one *Chlorine DPD #3R Tablet (6905A) to sample from Step 5.
Cap and mix until tablet disintegrates.
7. Insert test tube into Chlorine DPD Comparator (6978). Match
sample color to a color standard. Record as Reading #3.
Reading #3 - Reading #2 = ppm Dichloramine
Reading #3 = ppm Total Residual Chlorine