Lamotte LOW RANGE PHOSPHATE KIT User Manual

LOW RANGE PHOSPHATE KIT
LOW RANGE COMPARATOR, 0.0-2.0 PPM
CODE 3121-02
QUANTITY CONTENTS CODE
2 x 30 mL *Phosphate Acid Reagent *V-6282-G
5 g Phosphate Reducing Reagent V-6283-C
2 Test Tubes, 10.0 mL, glass, w/cap 0843
1 Pipet, 1.0 mL, plastic 0354
1 Low Range Comparator Viewer 1102
1 Phosphate Low Range Comparator Bar 0.0-2.0 ppm 3122-01
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com. Search for the four digit reagent code number listed on the reagent label, in the contents list or in the test procedures. Omit any letter that follows or precedes the four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel (US, 1-800-255-3924) (International, call collect, 813-248-0585).
To order individual reagents or test kit components, use the specifi ed code number.
NOTES:
This test determines levels of orthophosphates only.
This test should be run on clear samples only. Filter the sample if necessary.
Best results are obtained when solution temperatures are 23-25°C.
Warning! This set contains chemicals that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision.
USE OF THE LOW RANGE COMPARATOR
The Low Range Comparator allows extremely faint colors to be matched to color standards by viewing the reaction down the length of the tube increasing the path length by ten times compared to comparators that view the reaction across the diameter of the tube.
g.1
g.2
g.3
g.4
1. Slide the Low Range Comparator Bar into the Low Range Comparator (fi g.1).
2. Fill a test tube to the 10 mL line with untreated sample water. Insert it in the rear hole on the top of the Low Range Comparator (fi g.2).
3. Fill a test tube with sample water. Follow the test procedure. Remove the cap. Insert the tube in the front hole on the top of the Low Range Comparator (fi g.3).
4. Position the comparator so that light shines down through the test tubes (fi g.4). Tilt the comparator until the color standards and sample are illuminated. Match the color of the reaction to the color standards. Read the result from the Low Range Comparator Bar.
PROCEDURE
1. Slide the Phosphate Low Range Comparator Bar (3122-01) into Low Range Comparator Viewer (1102). Fill one test tube (0843) to the 10 mL line with sample water. Place in Low Range Comparator.
2. Fill test tube (0843) to 10 mL line with sample water.
3. Use 1.0 mL pipet (0354) to add 1.0 mL of *Phosphate Acid Reagent (V-6282). Cap and mix.
4. Use 0.1 g spoon (0699) to add one level measure of *Phosphate Reducing Reagent (V-6283). Cap and mix until dissolved. Wait 5 minutes.
5. Remove cap from test tube. Place tube in Low Range Comparator. Match sample color to a color standard. Record as ppm Orthophosphate.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
PO Box 329 • Chestertown • Maryland • 21620 • USA
Visit us on the web at www.lamotte.com
63121-02 11/19