LaMotte LB-DC User Manual

CHLO RINE BLEACH TEST KIT
DROP COUNT METHOD
MODEL LB-DC • CODE 7894
QUANTITY CONTENTS CODE
30 mL *Hypochlorite Reagent A *7939PS-G 30 mL Hypochlorite Reagent B 7940-G 60 mL Hypochlorite Reagent C 7941PS-H 2 Test Tubes, 5-10-15 mL, glass, w/caps 0778 2 Pipets, 0.5 mL, plastic, w/caps 0369 1 Pipet, 0.5 mL, plastic 0353 1 Pipet, plain, glass 0342
To order individual reagents or test kit components, use the specified code number.
PRO CE DURE A: 0-0.1% (0-1000 ppm)
1. Fill test tube (0778) to 5 mL line with sample solution.
2. Use a pipet (0369) to add 0.5 mL of Hypochlorite Reagent B (7940).
Swirl to mix.
3. Use second pipet (0369) to add 0.5 mL of *Hypochlorite Reagent A (7939PS). Swirl to mix. Sample will turn brown.
4. Fill glass pipet (0342) with Hypochlorite Reagent C (7941PS). Hold pipet vertically. While gently swirling tube, add Hypochlorite Reagent C, one drop at a time, until brown color disappears. Count the number of drops added.
5. Calculate result:
Available Chlorine, % = 0.005 x Number of Drops
Available Chlorine, ppm = 50 x Number of Drops
PRO CE DURE B: 0-1.0% (0-10 ppt)
1. Use 0.5 mL pipet (0353) to add 0.5 mL of the sample solution to a test tube (0778). Dilute to 5 mL line with tap water. Cap and mix.
2. Use a pipet (0369) to add 0.5 mL of Hypochlorite Reagent B (7940). Swirl to mix.
3. Use second pipet (0369) to add 0.5 mL of *Hypochlorite Reagent A (7939PS). Swirl to mix. Sample will turn brown.
4. Fill glass pipet (0342) with Hypochlorite Reagent C (7941PS). Hold pipet vertically. While gently swirling tube, add Hypochlorite Reagent C, one drop at a time, until brown color disappears. Count the number of drops added.
5. Calculate result:
Available Chlorine, % = 0.05 x Number of Drops
Available Chlorine, ppt = 0.5 x Number of Drops
PRO CE DURE C: 0-10% (0-100 ppt)
1. Use the 0.5 mL pipet (0353) to add 0.5 mL of the sample solution to a test tube (0778). Dilute to 5 mL line with tap water. Cap and mix.
2. Use the 0.5 mL pipet to transfer 0.5 mL of the diluted sample to second test tube (0778). Dilute to 5 mL line with tap water. Cap and mix.
3. Use a pipet (0369) to add 0.5 mL of Hypochlorite Reagent B (7940). Swirl to mix.
4. Use second pipet (0369) to add 0.5 mL of *Hypochlorite Reagent A (7939PS). Swirl to mix. Sample will turn brown.
5. Fill glass pipet (0342) with Hypochlorite Reagent C (7941PS). Hold pipet vertically. While gently swirling tube, add Hypochlorite Reagent C, one drop at a time, until brown color disappears. Count the number of drops added.
6. Calculate result:
Available Chlorine, % = 0.5 x Number of Drops
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Challenges
PO Box 329 • Chestertown • Mary land • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
10.10
Loading...