CHLO RINE TEST KIT
DUAL RANGE
MODEL SL-16 • CODE 3314
QUANTITY CONTENTS CODE
100 *Chlorine DPD #1R Tablets *6999A-J
100 *Chlorine DPD #3R Tablets *6905A-J
2 Test Tubes 2.5-5.0-10.0 mL, plastic, w/caps 0106
1 Octa-Slide Viewer 1100
1 Chlorine Octa-Slide Bar, 0.1-1.0 ppm 3405
1 Chlorine Octa-Slide Bar, 1.0-6.0 ppm 3404
*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents
go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,
phone or fax.
To order individual reagents or test kit components, use the specified code
number.
This kit is for use in testing water for Free Available Chlorine, Total
Residual Chlorine and Combined Chlorine.
USE OF THE OCTA-SLIDE VIEWER
The Octa-Slide Viewer should be held so
non-direct light enters through the back of the
comparator. With sample tube inserted at top,
slide the Octa-Slide bar through the viewer and
match with color standard.
EPA AC CEPTED PRO CE DURE
EPA accepted for NIPDWR monitoring. For compliance monitoring a
Check Standard should be prepared.
CHECK STAN DARD PREP A RA TION: 1 ppm Equiv a lent So lu tion
1. Dissolve 891 mg of potassium permanganate in 1000 mL of distilled
water in a volumetric flask (1000 ppm equivalent solution).
2. Dilute 1 mL of this solution to 1000 mL with distilled water in a
volumetric flask. This solution is equivalent to 1 ppm Free Available
Chlorine.
PRO CE DURE
FREE AVAIL ABLE CHLO RINE
1. Fill a test tube (0106) to the 5 mL line with sample water.
2. Add one *Chlorine #1R Tablet (6999A). Cap and mix until tablet
disintegrates. A pink to red color indicates the presence of Chlorine.
3. Insert test tube into the top of the Octa-Slide Viewer (1100). Slide
the 0.1 to 1.0 ppm Chlorine Octa-Slide Bar (3405) into the
Octa-Slide Viewer. Match sample color to a color standard. Record as
ppm Free Available Chlorine.
4. If the sample is darker than the 1.0 standard, remove 0.1 to 1.0
Octa-Slide Bar, replace it with the 1.0 to 6.0 Octa-Slide Bar (3404).
Match sample color to a color standard. Record as ppm Free
Available Chlorine.
NOTE: Save sample for Total Residual and Combined Chlorine
tests.
TO TAL RE SID UAL & COM BINED CHLO RINE
5. Add one *Chlorine DPD #3R Tablet (6905A) to the sample from
Step 4. Cap tube and mix until tablet disintegrates.
6. Insert test tube into the Octa-Slide Viewer (1100). Match sample
color to a color standard. Record as ppm Total Residual Chlorine.
7. Subtract Free Available Chlorine from Total Residual Chlorine.
Record as ppm Combined Chlorine.
Combined Chlorine =
Total Residual Chlorine – Free Available Chlorine
NOTE: Thoroughly clean and rinse test tubes after each test.
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Chal lenges
PO Box 329 • Chestertown • Mary land • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
SM
63314 - 5/11