This booklet provides step-by-step detailed instructions for the Code 3633-04 and
Code 3634-04 test kits. It is important to review these instructions thoroughly
attempting to perform the tests by the short-form instructions contained in the case lid.
To order individual reagents or test kit components, use the specifi ed code number.
before
Testing Hints / Reagent Care ........................................................ 4
Tightly close all reagent
containers immediately
after use. Be sure not to
interchange caps and
pipets from different
containers.
Wipe up any reagent chemical
34
spills, liquid or powder, as
soon as they occur.
Refer to label and
accompanying MSDS at
www.lamotte.com
for proper reagent
disposal.
Avoid prolonged exposure
of equipment and reagents
to direct sunlight. Protect
reagents and
components
from extreme
heat and cold.
Use care when dispensing or
handling all reagents. Some
reagents also may cause
permanent stains if spilled.
4
ANALYTICAL TECHNIQUE
1. Clean glassware is a must for accurate results.
Thoroughly rinse test tubes before and after
each test. Caps and stoppers should also be
cleaned after each use.
2. Use test tube caps, not your fi ngers, to cover test
tubes and fl asks during shaking or mixing.
3. When adding sample to calibrated test tube, be
sure vial is fi lled to the appropriate mark. The
bottom of the liquid (meniscus) should be level
with the desired mark. (Figure 1)
4. When dispensing reagents from bottles fi lled
with dropper plug and cap, be sure to hold
bottle vertically and gently squeeze to dispense
the appropriate number of uniform drops.
(Figure 2)
5. For those reagents to be added with the screwcap
pipet assemblies
enclosed, remove polyseal cap
on bottle and replace with the screwcap pipet.
NOTE: Place the polyseal caps back on the
reagent bottles for longer periods of storage. Be
sure that both pipet assemblies and polyseal caps
are thoroughly cleaned before placing on bottles
to avoid contamination.
6. When dispensing reagents from pipets, hold
pipet vertically to assure uniform drop size. This
is extremely important when performing drop
count titrations. (Figure 3)
7. To fi ll pipets, squeeze rubber bulb and immerse
into reagent. Release bulb to fi ll. (Figure 4)
8. To accurately dispense powdered reagents with
spoon, tap spoon on edge of reagent container
to remove excess reagent. (Figure 5)
9. When performing tests that include Octa-Slide
Comparators, the comparator should be
positioned between the operator and non-direct
sunlight. This allows the light to enter through
the
light-diffusing screen at the back of the
comparator for optimum
color comparison.
Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
5
GENERAL SAFETY PRECAUTIONS
12
Store the test kit in a
cool dry area.
34
Read the labels on all reagent
bottles. Note warnings and
fi rst aid information.
Reagents marked with a * on
instructions are considered
possible health hazards.
Avoid contact between reagent
chemicals and skin, eyes, nose,
and mouth.Wear safety glasses when
65
Read all instructions and note
precautions before performing
the test procedure.
Read all Material
Safety Data Sheets
Material
Safety
(MSDS)
Data
at www.lamotte.com.
Sheet
Keep all
equipment
and reagent
chemicals
out of the
reach of
young children.
performing test procedures.
7
6
In the event of an accident or suspected poisoning, immediately call
the Poison Center phone number in the front of your local telephone
directory or call a physician. Additional information for all LaMotte
reagents is available in the United States, Canada, Puerto Rico, and
the US Virgin Islands from Chem-Tel by calling 1-800-255-3924.
For other areas, call 813-248-0585 collect to contact Chem-Tel’s
International access number. Each reagent can be identifi ed by
the four digit number listed on the upper left corner of the
reagent label, in the contents list and in the test procedures.
TEST METHODS
This test kit uses two basic analytical procedures common to fi eld test kits. A brief
explanation of each follows:
COLORIMETRIC: OCTA-SLIDE 2 VIEWER
In a visual colorimetric test, a sample is treated with reagent(s) to produce a color
reaction, generally in proportion to the amount of test factor present. The sample color
is then compared against color standards representing known concentrations of the
factor being tested over a specifi c range.
Hold the Octa-Slide 2
123
Viewer so that non direct light enters
through the back of
the viewer.
Match sample color
4
to color standard.
Record results.
The calibrated test tubes (0106) included in this kit may be used to perform dilutions
for the Ammonia Nitrogen and Nitrite Nitrogen tests. Distilled or deionized water is
needed to perform dilutions.
The following table provides a quick reference guide for dilutions of various
proportions. Once the dilution is prepared, use this diluted sample to perform the test,
and multiply the result by the dilution factor to obtain the actual concentration.
Insert Octa-Slide 2
Bar into the
Octa-Slide 2
Viewer (1101).
Note:
If sample color is between two standards, the midpoint
is taken as the result.
If the sample is darker than the highest standard, a
dilution may be performed on a fresh sample, and the
test repeated to bring the concentration within range.
DILUTIONS
Insert test tube
containing reacted
sample into the
Octa-Slide 2
Viewer (1101).
Sample SizeDistilled Water to
Bring to 10 mL
5.0 mL5.0 mL2
2.5 mL7.5 mL4
Dilution Factor
7
TITRIMETRIC: DIRECT READING TITRATOR
In a titrimetric method, titrating solution (or titrant) is added to a treated sample
until a color change occurs. The volume of titrant required to reach this endpoint is
proportional to the concentration of the factor being tested. Direct Reading Titrators
provide results directly in the appropriate concentration for the test - no counting of
drops, no calculations.
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
PlungerBarrelAdapter Tip
The Titrator consists of a plastic barrel, a plastic plunger, and a plastic adapter tip. The
adapter tip reduces the size of the drops that are dispensed, increasing the precision of
the test results. DO NOT remove the plunger or adapter tip from the Titrator.
12 3
Fill the test
tube to the
appropriate
line with
sample water.
To fi ll Titrator, slowly
4
withdraw the plunger
until the bottom of the
plunger is
opposite
the zero
mark on
the scale.
6
Add reagents
as specifi ed
in the instructions
for the individual
test method. Cap
with the special test
tube cap. Mix by
swirling gently.
I If small air
Note:
bubbles appear in
the barrel, expel them by
partially fi lling the barrel
and pumping the
solution back into
reagent container.
titration
the
Repeat
until bubble disappears.
7
Insert the Titrator
into the center hole
of the test tube
cap. While gently
swirling tube, add
titrating solution
one drop at a time
until the desired
color changes occur.
Read the test
result directly
from the scale
where the large
ring on the Titrator
meets the Titrator barrel.
Follow individual
test instructions.
Depress the
Titrator plunger
to expel air. Insert
Titrator into the
plastic fi tting of
the titrating
solution bottle
and invert.
5
Turn the bottle
right-side-up
and remove the
Titrator.
8
When testing is
complete, discard
titrating solution in
Titrator. Do not return
the titrant to the
reagent bottle. Rinse
Titrator and titration
tube thoroughly. Do
not remove the plunger
or the adapter tip from
the Titrator.
8
TEST PROCEDURES
INTRODUCTION
Proper control of water quality is an essential part of successful aquaculture operation.
Immediate test results provided by on-site water analysis equipment can confi rm a healthy
environment, or give early warning signals for required treatment.
1. Develop a routine testing schedule.
2. Keep records! Historical data is extremely important if treatments are required. Note
environmental conditions, fi sh activity, feeding habits, etc.
3. Observe fi sh to note any particular behavior or feeding rates, as this may be a sign
of stress.
4. Stable characteristics, such as alkalinity and hardness, do not have to be tested as
frequently as ones that fl uctuate, such as ammonia nitrogen, nitrite nitrogen, pH,
dissolved oxygen and temperature. Keep in mind that these factors fl uctuate throughout
the day and in some cases are interdependent.
5. Be alert to sudden changes in one factor, as it may be a clue to perform further analysis.
9
Loading...
+ 19 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.