Lamotte FREE, TOTAL User Manual

ACCEPTED
or
APPROVED
FREE, TOTAL, & COMBINED
CHLORINE KIT
OCTA-SLIDE 2, 0.2-3.0 ppm
CODE 3308-01
QUANTITY CONTENTS CODE
50 50
2 Test Tubes, 2.5-5-10mL, plastic, w/caps 0106
1 Octa-Slide 2 Viewer 1101
1 Chlorine Octa-Slide 2 Bar, 0.2-3.0 ppm 3401-01
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com. Search for the four digit reagent code number listed on the reagent label, in the contents list or in the test procedures. Omit any letter that follows or precedes the four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel: (US, 1-800-255-3924) (International, call collect, 813-248-0585)
To order individual reagents or test kit components, use the specied code number.
Chlorine DPD #1R Tablets (6999A) Chlorine DPD #3R Tablets (6905A)
6905A-6999ABOX
Warning! This set contains chemicals that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision.
USE OF THE OCTA-SLIDE 2 VIEWER
The Octa-Slide 2 Viewer should be held so non-direct light enters through the back of the Viewer. Insert the Octa-Slide 2 Bar into the Viewer. Insert the reacted sample into the top of the Viewer. Match the color of the reaction to the color standards.
EPA ACCEPTED PROCEDURE
EPA Accepted for NPDWR monitoring. For compliance monitoring a Check Standard should be prepared.
CHECK STANDARD PREPARATION: 1 ppm equivalent solution
1. Prepare a stock solution containing 891 mg potassium permanganate diluted to 1 L
with distilled water in a volumetric ask (1000 ppm equivalent solution).
Prepared fresh daily.
2. Dilute 1 mL of this solution to 1000 mL with distilled water in a volumetric ask.
This solution is equivalent to 1 ppm Free Available Chlorine.
PROCEDURES
FREE AVAILABLE CHLORINE
1 2 3
Insert Chlorine Octa-Slide 2 Bar (3401-01) into the Octa-Slide 2 Viewer (1101)
4
Cap and mix until tablet disintegrates.
NOTE: Retain this sample if Total Residual Chlorine and Combined Chlorine are to be tested.
.
5
Fill a test tube (0106) to the 5 mL line with the water sample.
Insert test tube into Octa-Slide 2 Viewer.
Add one Chlorine DPD #1R Tablet (6999A).
Match sample color
6
to a color standard. Record as ppm Free Available Chlorine.
Loading...