FILMING AMINE
MODEL FMN • CODE 7605
QUANTITY CONTENTS CODE
2 x 100 mL *Chloroform Reagent *6203-J
100 mL *Amine Indicator Reagent *7607-J
15 mL *Sulfuric Acid, 2.0 N *6290-E
4 Test Tubes, 5 mL, glass, w/caps 0230
1 Funnel, plastic 0459
1 Pipet, 1 mL, plastic 0354
1 Beaker, plastic, 100 mL 0896
1 Bottle, Filming Amine, 75 mL, w/cork 0895
1 Test Tube Holder 1113
1 Filming Amine Comparator, 1.0 - 10.0 ppm 7606
*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
USE OF THE OCTET COM PARA TOR
The Octet Comparator contains eight permanent color standards. A test
sample is inserted into the openings in the top of
the comparator. The sample can then be
compared to four color standards at once, and
the value read off the comparator. For optimum
color comparison, the comparator should be
positioned between the operator and a light
source, so that the light enters through the
special light-diffusing screen in the back of the
comparator. Avoid viewing the comparator
against direct sunlight or an irregularly lighted
background.
PRO CE DURE
1. Fill the large bottle (0895) to the 75 mL line with sample water.
2. Use the 1 mL pipet (0354) to add 3 mL of *Amine Indicator Reagent
(7607). Cork and mix.
3. Fill a test tube (0230) to 5 mL line with *Chloroform Reagent
(6203). Add to the bottle containing sample and indicator. Cork and
shake. If the bottom chloroform layer is blue, add *Sulfuric Acid,
2.0 N (6290) one drop at a time, until the bottom layer is yellow. Do
not add more than 5 drops of *Sulfuric Acid, 2.0 N (6290).
4. Be sure that the bottle is tightly stoppered. Shake vigorously for 3
minutes. Allow to stand until the chloroform and water layers
separate.
5. Separate the layers as follows:
Insert the funnel (0459) into a test tube (0230). Use the test tube
holder to hold the test tube over the beaker (0896). Pour the
contents of the bottle through the funnel. The water will overflow
between the funnel and test tube into the beaker. The chloroform
layer is heavier than the water and will remain in the test tube and is
used for the color comparison. Discard the water.
6. Dry the outside of the test tube. Insert into Filming Amine
Comparator (7606). Match sample color to a color standard. Record
as ppm filming amine.
NOTE: If the sample contains over 10 ppm amine, the solution will
be darker than the highest standard in the comparator. Dilute the
solution with an equal volume of *Chloroform (6203). Multiply the
comparator reading by 2 to obtain the result in ppm filming amine.
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Challenges
PO Box 329 • Chestertown • Mary land • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
7/10