This manual serves to explain the use of the Dissolved Oxygen hand-held meter
DO 6.
This manual functions in two ways: first as a step by step guide to help you
operate the meter; second, it serves as a handy reference guide. This manual is
written to cover as many anticipated applications of the DO 6 hand-held
meter as possible. The information presented in this manual is subject to
change without notice.
14.1.1 General Information ..................... 48
14.1.2 Measurement Units ..................... 49
14.1.3 What is Being Measured ................... 49
14.1.4 Air Calibration........................ 50
14.1.5 Applications ......................... 51
15 Limited Warranty .........................51
16 Return of Items ..........................52
4
1 INTRODUCTION
CAL
TA
Low Battery annunciator
Calibration Mode
Annunciator
14-segment Liquid Crystal Display
Percentage
(Percentage Saturation
Mode)
parts per million
milligram per litre
(Concentration Mode)
ATC annunciator
Temperature Mode
mg/L
ppm
%
+
-
8.8.8.8
Thank you for purchasing the DO 6 Dissolved Oxygen palm-top meter. This
economy microprocessor-based hand-held meter has a large custom LCD
(Liquid Crystal Display) for clear and easy reading.
The DO 6 offers the measurement of dissolved oxygen in the percentage
saturation mode and in the concentration mode. Temperature measurement is
also available in degrees Celsius. The meter ensures accurate measurement of
the Dissolved Oxygen values through its temperature, barometric pressure and
salinity compensation features.
Your meter includes a dissolved oxygen probe, refill solution, a rubber boot, 4
alkaline “AAA” batteries, and instruction manual.
Please read this manual thoroughly before operating your meter.
To order other accessories and standard solutions, please refer to the Section
on Accessories for more information.
2 DISPLAY & KEYPAD FUNCTIONS
2.1 Display
The meter has a large custom LCD that consists of 14-segment 4 digit display
and operation annunciators for Percentage Saturation measurement mode (%),
Concentration measurement mode (mg/L or ppm) and Temperature
measurement mode in degrees Celsius (T). Other annunciators include “A”
(when the ATC function is activated), “CAL” (when meter is in calibration
mode) and battery diagram for low battery condition. See Figure 1.
Figure 1, Active LCD display for DO6
5
2.2 Keypad
ON
OFF
CAL
HOLD
ENTER
MODE
The DO 6 meter has 6 keys on its splash-proof keypad; ON/OFF,
HOLD/ENTER CAL, MODE,
p and q keys. Some buttons have several
functions depending on its mode of operation.
ON/OFF – Powers on and shuts off the meter. The meter will start in
the measurement mode it was in when last switched off.
CAL – Activates the calibration mode for Percentage Saturation,
Concentration and Temperature calibrations.
Deactivates calibration or setup without confirming a value in
calibration mode.
HOLD -Activates/Deactivates freezing of the measured reading while
in measurement mode.
ENTER - Confirms the calibration values in Calibration mode and
the selections in the SETUP menu.
pq – Sets the calibration values during the Concentration and
Temperature calibration.
Scrolls through each SETUP and its sub group menu.
Set offset adjustments and configuration settings.
MODE - Selects the measurement option between DO Percentage
Saturation measurement; DO Concentration measurement and
Temperature measurement.
When pressed together with ON/OFF key during power on, SETUP
mode is selected. This menu allows meter customization with
preferences such as activating ATC selection, DO concentration
measurement unit selection, viewing of the last calibration data,
viewing of the electrode properties, selecting the auto power off,
resetting calibration data or meter settings back to factory default,
setting the offset adjustments, setting barometric pressure in Hg or PA
for barometric pressure compensation and setting of the salinity value
for salinity compensation.
6
3 PREPARATION
3.1Inserting & Removing Rubber Boot
1. To remove meter from rubber boot, push out from the bottom edges of
meter until it is completely out of boot. Ensure that the cables of Dissolved
Oxygen electrode or temperature probe are not connected
2.. To insert meter into rubber boot, slide in from the top of meter before
pushing the bottom edges of meter down to set it into position. Lift up the
stand at the back of meter for bench top applications if necessary.
Figure 2: Inserting or removing the rubber boot
7
3.2 Inserting the Batteries
3.0
%
+
-
The battery compartment is found at the back of instrument as shown in
Figure 3. To open the battery compartment first remove protective rubber
boot/stand then:
1. Push in the direction of arrow and lift up the cover.
2. Note the polarity of battery before inserting into position.
3. After replacement, place cover back and press down until it locks tight.
Figure 3: Inserting the batteries
3.3 Battery Replacement
A low battery annunciator in the LCD alerts you when battery power is
running low. See Figure 4. Replace with the same type as recommended by the
manufacturer.
Figure 4: Low Battery Condition
Caution: Power off the meter when changing battery.
8
3.4 Dissolved Oxygen Probe Information
16 mm
12 mm
Detachable
Pre-membraned Cap
(Probe Sensing Area)
Figure 5: 12 mm Galvanic DO probe
The DO 6 hand-held meter is supplied with
a Dissolved Oxygen Probe that works on
galvanic principle, that is, it does not require
any polarising voltage from your meter.
Rather it will generate a millivolt signal
proportional to the amount of oxygen in the
solution.
Your DO Probe has a twin cable, one with a
BNC connector for the DO measurement
input, and the other with a phono jack plug
for the temperature measurement input. Its
sensing area consists of a cathode, anode, and
an inner electrolyte which is separated from
your sample solution by an oxygen permeable
membrane pre-membraned to the probe’s
detachable cap. Reconditioning of your
probe is made easy with the introduction of
this specially designed detachable
pre-membraned cap.
This light weight probe has a built-in
temperature sensor for Automatic
Temperature Compensation (ATC) with
12mm diameter epoxy body housing and 16
mm diameter Noryl detachable cap. Its
compact sensing area reduces air entrapment
which makes it easy to obtain fast, accurate
and stable readings. Simply stir the probe in
the solution being measured. Shaking will aid to remove bubbles if needed
before taking a reading. Also, always ensure to remove water drops from the
membrane when calibrating in air.
Proper use of probe is essential to ensure that the optimum measurement is
taken in a short time. Always immerse the probe beyond the pre-membraned
cap. The minimum water flow rate is 2 inch/second across membrane. The
temperature range for use is within 0 to 50 °C.
Always ensure that the probe’s membrane is protected against any scratches or
dents. The whole pre-membraned cap has to be replaced if the
membranedamaged. It is also important to keep the membrane clean so as to
produce an optimum and accurate measurement.
See Section 8 - "Probe Care and Maintenance" for more information.
9
3.5 Connecting the Probe to Meter
BNC Connector for
DO probe
Phono jack for
Temperatureprobe
1. Insert the BNC connector from the electrode to the BNC connector socket
on the meter accordingly and rotate connector clockwise until it locks. See
Figure 6 below.
2. To remove, simply rotate the connector in counter-clockwise direction
until it unlocks, and slide the connector off the socket.
3. Plug the phono jack of temperature sensor into the phono socket of the
meter as shown in Figure 6 below.
Caution: Do not pull or force the probe cord or the probe wires might
disconnect.
Note: Keep connectors clean. Do not touch connector with soiled hands.
Figure 6: DO 6 probe to meter connections
10
3.6 Switching the Meter On
CAL 13 CAL 24
TAv g
A
1.00
10.00
D06
mg/L
NTU
mg/L
ppm
ppm
%pH
%
%
+
-
8.8.8.8
ON
OFF
When switching the meter on, it will go through a series of display, eventually
showing the revision number of the meter before going to the measurement
mode. The mode of measurement which the meter displays will be similar to
the last measurement mode to
which the meter is switched off.
Press ON/OFF key to power up
your meter.
1. All LCD segments will light up
for 1.5 seconds.
2. The display will then
momentarily show the meter’s
identification “DO 6” with the
annunciators for percentage
saturation mode and
concentration mode.
3. The display continues to switch
to display meter’s revision
number for 1.5 seconds before
finally showing the
measurement mode.
4. Either percentage saturation
measurement mode,
concentration measurement
mode or temperature
measurement mode will be
displayed following the previous
measurement mode to which
the meter is switched off.
By simply pressing the MODE key, you can select the measurement mode of
percentage saturation measurement, concentration measurement or
temperature measurement.
The customized annunciator shall indicate the selection of the measurement
mode.
Figure 8: Measurement mode selection using the mode key
12
4 CALIBRATION
4.1 Important Information on Meter Calibration
Your meter has three measurement modes namely the DO % saturation
measurement mode, the DO mg/L (ppm) concentration mode and the
Temperature measurement mode.
Since the mg/L (ppm) concentration value of dissolved oxygen varies with
temperature, barometric pressure and salinity, the meter with its electrode has
to be calibrated with consideration to these factors to ensure a proper
measurement. Therefore, it is very important to set the proper temperature,
barometric pressure and salinity values in the meter prior to any calibration or
measurement process. Since the dissolved oxygen measurements in %
saturation will linearly affect the mg/L (ppm) concentration measurement, it is
important that the meter has to be first calibrated in the % Saturation mode
and temperature mode before commencing the mg/L (ppm) concentration
mode calibration.
In % saturation calibration mode, the dissolved oxygen calibration values will
vary according to the barometric pressure correction set in the meter.
Therefore it is necessary to set the barometric pressure first before proceeding
to % Saturation calibration. The following table is an example which explains
the acceptance calibration values when % Saturation mode calibration is
attempted under different absolute value and under two different barometric
pressures.
% Saturation factory
default value
less than 10%0%0%
10.1% to 49.9%Err.1 (error 1)Err.1 (error 1)
50% to 200%100%92.1%
Acceptance Calibration
Value (760mmHg)
Acceptance Calibration
Value (760mmHg)
In the mg/L (ppm) concentration calibration mode, barometric pressure,
salinity and temperature of the calibration sample needs to be set in the meter
prior to calibration. Since the pressure has already been set in the % saturation
calibration which is done earlier and assuming that the calibration is done on
the same barometric pressure and if you are using the temperature sensor in
automatic temperature compensation mode, you only need to input the known
salinity of the calibration sample to the meter. The dissolved oxygen mg/L
(ppm) concentration value can then be calibrated to the known sample
concentration calibration value. Note that the concentration calibration
13
window is +/- 40% of the factory default value. However the minimum point of
calibration is 2 mg/L.
When you recalibrate your meter in % saturation mode, old calibrations in %
saturation will be replaced while the % saturation offset adjustment will be
erased. However, recalibration in the mg/L (ppm) concentration mode will
only replace the old calibration in the concentration mode and do not affect
the calibration in the percentage saturation mode.
To completely recalibrate your meter, or when you use a replacement
electrode, it is best to clear all calibration data. To erase all the old calibration
data completely, see Section 7.6 (rSt.5) Reset to Factory Default.
4.2 Preparing the Meter for Calibration
Before starting calibration, make sure you are in the correct measurement
mode and in the correct calibration sequence. The temperature and the %
Saturation calibration must be done first before attempting to do the mg/L
(ppm) Concentration calibration.
In % Saturation, the meter is able to perform either a one point calibration or
a 2 point calibration. For one point calibration, it is recommended that you
perform a 100% Saturation calibration in saturated air. If you opt for 2 point
calibration, you can calibrate for 100% Saturation in saturated air and 0%
Saturation using a zero oxygen solution. The meter will take several minutes to
reach 0% Saturation value after submersion into the zero solution.
Rinse the probe well in the de-ionized (DI) water or rinse solution and wipe
the probe carefully taking care of the membrane.
Calibrate the meter in all the modes to ensure the highest accuracy throughout
the DO measurement range. In % Saturation calibration, should there be a
calibration failure for 0% Saturation slope calibration; the meter may have
exceeded the limit of 10% of the factory calibrated absolute value. The sensor
may have to be re- conditioned. Also, note that the DO 6 meter will not
perform 0% or 100% Saturation calibration for absolute value ranging from
10.1% to 49.9%. The meter also will not perform the mg/L (ppm)
Concentration calibration for an absolute value of less than 2.00 mg/L (ppm)
or calibration which is out of its window tolerance of +/- 40% from the factory
default value. An error message of “Err.1” will be indicated and will return the
meter to the measurement mode.
All new calibration values will automatically override the existing data. It is
recommended to calibrate the meter periodically and or if it is suspected to be
inaccurate.
14
Always rinse the probe with either DI water or rinse solution before and after
CAL
HOLD
ENTER
co
24.3
TA
25.0
TA
“CO” will display for
1.5 seconds
24.3
TA
CAL
25.0
TA
CAL
Factory default
Temperature
Measurement
Figure 9: Temperature Calibration Sequence
each calibration/sample measurement. When calibrating in air, make sure that
any water droplets from the probe’s membrane are removed.
For details please refer to Section 8 - PROBE CARE AND MAINTENANCE.
Note: The DO 6 factory calibrated default value is in respect to 760mm Hg or
101.3 kPa barometric pressure (sea level). To set the barometric pressure to
different value, see Section 5.4.1 Pressure Setting Adjustment.
4.3 Temperature Calibration
DO mg/L (ppm) Concentration
measurement is dependent on the
temperature of the sample to be
measured. Most users will choose
to have the DO 6 automatically
compensate for temperature using
the temperature sensor that is
built into the DO probe.
However, it is possible to
manually enter a known
temperature to the meter
(manual temperature
compensation). The built-in
temperature sensor of the DO 6
probe is factory calibrated.
However, the temperature can be
re-calibrated if needed and is
recommended when a
replacement probe is used.
Before calibration, ensure that
the meter displays annunciator
“A” to indicate that it is in
automatic temperature
compensation mode. Please see
Set Up menu on page 36 for its
activation procedure.
1. Switch the meter on. Press
MODE to select temperature
mode. The display should
show the annunciator “T” for
temperature mode and“A”
for automatic temperature compensation mode.
15
2. Dip the probe to a solution of known temperature (i.e. a temperature
CAL
HOLD
ENTER
co
83.0
A
CAL
100.0
A
“CO” will display for
1.5 seconds
A value with respect to the
factory default calibration
will be displayed
100.0
85.2
CAL
Calibration point will be
displayed for 1.5 seconds
Figure 10: 100% Saturation Calibration
%
%
%
%
bath).Allow some time for the reading to stabilize.
3. Press the CAL key. The CAL indicator will appear and blinks above the
display. The temperature value displayed is with respect to the factory
default calibration.
4. Press the p andq keys to adjust the meter reading to the correct
temperature value (i.e. the temperature of the temperature bath)
5. Press the HOLD/ENTER key to confirm temperature calibration and
return to measurement mode.
NOTE: To exit from Temperature Calibration mode without confirming the
calibration, press CAL instead of HOLD/ENTER. Temperature calibration is
restricted to ± 5°C from the factory
default temperature measurement
displayed during calibration.
4.4 Dissolved Oxygen Calibration in
% Saturation
You can calibrate this meter quickly and
easily in air. The exact calibration value
depends on barometric pressure. The
meter is set to a factory default of 760 mm
Hg, which results in a calibration value of
100% Saturation in air.
NOTE: If the barometric pressure setting
has been changed from 760 mm Hg, the
calibration value in air will automatically
adjust to a value other than 100%. The
adjusted value will be correct for the new
barometric pressure setting.
See Section 5.4.1 Pressure Setting
Adjustment on page 26 to change the
pressure setting.
4.4.1 To calibrate 100% Saturation:
1. Rinse the probe well with DI water or
rinse solution. Do not touch the
membrane.
2. Press the MODE key to select the %
Saturation mode.
16
3. Hold the probe in the air with the sensor facing downwards. Wait for the
CAL
HOLD
ENTER
co
3.0
A
CAL
0.0
A
“CO” will display for
1.5 seconds
A value with respect to the
factory default calibration
will be displayed
0.0
5.2
CAL
Calibration point will be
displayed for 1.5 seconds
Figure 1: 0% Saturation Calibration
%
%
%
%
reading to stabilize.
4. Press the CAL key. The display will show the intended percentage
calibration point (100%) with a CAL indicator at the top corner for 1.5
seconds. The CAL indicator will then blink and the display will show a
value which is with respect to the factory default calibration. When
calibration is attempted on a factory calibrated absolute value measurement
of 10.1% to 49.9%, the display will show the absolute value measurement
for 1.5 seconds before displaying Err.1 and goes to the measurement mode.
5. Press the HOLD/ENTER key to confirm the calibration. The meter
displays “CO” for 1.5 seconds and automatically calibrates to 100.0% air
saturation and returns to measurement
mode.
Note: You can offset your % DO
calibration. See Section 7.7 (OFS.6) %
Saturation Offset Adjustment for
directions.
6. Press the HOLD/ENTER key to confirm
the calibration. The meter displays
“CO” for 1.5 seconds and automatically
calibrates to 0.0% Saturation and
returns to measurement mode.
4.4.2 To calibrate 0% Saturation
1. Rinse the probe well with DI water or
2. Press the MODE key to select the %
3. Dip the probe into the zero oxygen
4. Press the CAL key. The display will
Note: When calibration is attempted on a
factory calibrated absolute value
rinse solution.
Saturation mode, if necessary.
solution. Wait for the reading to
stabilize.
show the intended percentage
calibration point (0%) with a CAL
indicator at the top corner for 1.5
seconds. The CAL indicator will then
blink and the display will show a value
which is with respect to the factory
default calibration.
17
measurement of 10.1% to 49.9%, the
CAL
HOLD
ENTER
co
8.1 6
A
8.24
A
mg/L
“CO” will display for
1.5 seconds
9.00
mg/L
mg/L
8.24
mg/L
Factory default
Temperature
Measurement
Figure 9: Temperature Calibration Sequence
CAL
CAL
display will show the absolute value
measurement for 1.5 seconds before
displaying Err.1 and goes to the
measurement mode.
5. Press the HOLD/ENTER key to
confirm the calibration. The meter
displays “CO” for 1.5 seconds and
automatically calibrates to 0.0%
Saturation and returns to
measurement mode.
4.5 Dissolved Oxygen Calibration in
mg/L or ppm concentration mode :
Calibrating the meter to 100% Saturation
will also calibrate the concentration mode
at the value in mg/L corresponding to
100% Saturation. This should produce
acceptable results in most applications.
This meter also lets you make a
calibration adjustment in mg/L (ppm)
Concentration mode without affecting
your % Saturation calibration.
To select between mg/L and ppm units,
see page 38
mode"
.
4.5.1 To calibrate in mg/L (ppm)
Concentration mode:
1. Calibrate 100% Saturation as per
Section 4.4.
2. Rinse the probe well with DI water or
rinse solution. Wipe the outside of the
probe carefully (do not touch the
membrane).
3. Dip the probe into a sample of known
oxygen concentration. Wait for the
reading to stabilize.
4. Press the MODE key to select the
mg/L (ppm) Concentration mode.
, "Selection of mg/L or ppm
18
5. Press the CAL key. The display will show a value which is with
83.0
%
A
Automatic Temperature
Compensation activated
Figure 8: Automatic Temperature Compensation
respect to the factory default calibration with a CAL indicator
blinking at the corner of the display.
6. Press the p andq keys to adjust the reading to the known oxygen
concentration of the sample.
7. Press the HOLD/ENTER key to confirm the calibration. The meter
will display “CO” for 1.5 seconds and automatically calibrates to the
value you have entered and returns to the measurement mode.
NOTES: The concentration calibration window is +/- 40% of the factory
default value. However, the minimum point of calibration is 2 mg/L. In
an event where the default measured value is less than 2 mg/L and a
calibration is attempted, the display will show “2.00” and then error
message “Err.1” will be displayed momentarily before returning to
measurement mode.
You can change the barometric pressure value and salinity value in the
mg/L (ppm) Set Up menu (see Section and for and ). Please refer to
page if you are intending to use manual temperature compensation.
5 MEASUREMENT
The DO 6 meter is capable of taking measurements with automatic or
manual temperature compensation.
5.1 With Automatic Temperature Compensation (ATC)
For Automatic temperature compensation, make sure the phono jack of
the probe (see Figure 6 in Section 3.5) is securely inserted. The ATC
annunciator “A” will be blinking if the ATC probe is disconnected.
Activate the “A.ATC” mode in the set up menu, if necessary. See page
36.
19
5.2 Manual Temperature Compensation
CAL
HOLD
ENTER
co
25.0
T
25.5
T
“CO” will display for
1.5 seconds
25.0
T
CAL
25.5
mg/L
CAL
Last set manual
temperature setting
Figure 3: Manual Temperature
Compensation value selection sequence.
For manual temperature
compensation, simply deactivate the
“A.ATC” mode as explained in the
set up menu on page . The
automatic temperature
compensation annunciator “A” will
not be displayed to represent that
the meter is in manual temperature
compensation mode.
To use manual temperature
compensation, you need to enter the
temperature value of your process
into the meter. You can select any
temperature between 0 and 50 °C.
Default value is 25 °C.
1. Switch on the meter and if
2. Press CAL key to start
3. The “CAL” indicator will start
4. Check the temperature of your
5. Press the HOLD/ENTER key to
necessary, press MODE key to
select temperature measurement
mode.
temperature calibration process.
blinking and the display will
show the last manual set
temperature value.
sample using an accurate
reference thermometer. Wait for
the thermometer value to
stabilize. Press the p or q key
of your meter to manually set its
temperature value according to
the value of the reference
thermometer used.
confirm the manual temperature
setting made. The “CO”
indicator will appear for 1.5
seconds before the display
returns to the measurement mode.
20
5.3 Taking Measurements
70.5
5.80
24.3
MODE
MODE
A
A
TA
mg/L
%
Figure 4: Measurement mode
During measurement, care must be taken
to ensure that the probe’s membrane
surface does not touch anything. Stir
around the solution to prevent any air
bubbles from being trapped on the
membrane before taking a reading.
IMPORTANT: Since the DO probe
consumes oxygen from the sample, the
sample must constantly flow past the
membrane to achieve more accurate
readings. You can use a stirrer to
accomplish this.
5.3.1 To take measurements:
1. Rinse the probe well with DI water or
rinse solution.
2. Select the appropriate measurement
mode. Press the MODE key to toggle
between modes:
a.Percentage Saturation (%)
b.Concentration (mg/L) or (ppm)
c.Temperature (T)
3.Dip the probe into the sample. Stir
the solution gently to homogenize
the sample. Make sure that the
sample is continuously flowing past
the membrane sensor.
4.Take your measurements once your meter reading has stabilized.
Note: “Ur”/”Or” will be displayed if the instruments exceeds the specified
measuring range. See Section for range specification.
To select between mg/L and ppm units, see Set Up menu on page 38.
5.4 Taking Pressure/Salinity Compensated DO Measurements
In taking pressure and salinity compensated DO measurements, the pressure
and salinity values of your measurements need to be adjusted from the Set Up
menu. The pressure setting adjustment feature is available when Set Up menu
is derived from the % Saturation mode, mg/L (ppm) Concentration mode or
21
Temperature mode. Salinity setting feature, on the other hand, will only be
available when Set Up menu is derived from the mg/L (ppm) Concentration
mode. See Section - Advanced Setup Overview for more information.
During measurement, the DO meter will automatically compensate for salinity
and pressure based on the values entered in the Set Up Menu. The meter is
factory set at 760 mm Hg (101.3 kPa) barometric pressure setting and a salinity
factor of 0.0 ppt.
5.4.1Pressure Setting Adjustment
(See Figure 16 for flow chart explanation)
1. Switch off your meter in mg/L (ppm) Concentration mode.
2. While pressing and hold the MODE key, switch on the meter by pressing
the ON/OFF key.
3. Once the ON/OFF key is released, the meter will enter the Set Up menu
by displaying “SEt.P”. By releasing the MODE key, the display will show
the configuration menu “COF.1”. See section – Advanced Setup
Functions
4. Press the p and q keys to scroll through the Set Up main group menu till
the display shows “DPr.7”.
5. Press the HOLD/ENTER key and the meter will enter the barometric
pressure units selection menu with the display showing one of two; either
“A.HG” for mm Hg or “A.PA” for kilo Pascal, depending on what the
meter’s last set up was.
6. Use the p and q keys to toggle between the two units of pressure
measurement and press the HOLD/ENTER key to confirm selection.
7. Use the p and q keys to set the pressure value and press the
HOLD/ENTER to confirm the setting. A confirmation indicator “CO” will
be displayed for 1.5 seconds before the meter shows the next display of
Salinity setting “b.SAL”. If the Set Up menu is derived from other than the
mg/L (ppm) Concentration mode, the meter will return to the Set Up
main group menu “DPr.7”.
8. Press CAL key to return to measurement mode, or continue to make a
salinity setting adjustment.
.
22
5.4.2 Salinity Setting Adjustment
.sal
HOLD
ENTER
HOLD
ENTER
0.0
45.2
co
“CO” will display
for 1.5 seconds
Display will return back to “DPr.7”
subgroup menu
Figure 5: Salinity Setting Adjustment Sequence
NOTE: This mode is available
only from the mg/L (ppm)
Concentration Set Up menu.
1. Repeat steps 1 to 5 as in the
steps of Pressure Setting
Adjustment.
2. Press the HOLD/ENTER key
twice till the meter displays the
Salinity setting mode “b.SAL”.
3. Press the HOLD/ENTER key
to enter the Salinity setting
mode.
4. Use the p and q keys to enter
5. Press HOLD/ENTER key to
6. Press CAL key to return to the
the salinity of your solution in
parts per thousand (ppt).
Factory default is 0.0. Salinity
values up to 50 ppt can be
entered.
confirm the value. A
confirmation indicator “CO”
will be displayed for 1.5
seconds before returning to the
main group menu “DPr.7”.
measurement mode.
23
HOLD
ENTER
HOLD
ENTER
HOLD
ENTER
HOLD
ENTER
.7
.
.
760
HOLD
ENTER
co
750
101.3
co
105. 2
“CO” will display
for 1.5 seconds
“CO” will display
for 1.5 seconds
Display will continue to
Salinity setting menu
“b.SAL” if Set Up menu is
entered from mg/L (ppm)
mode. Otherwise display will
return back to “DPr.7”
Display will continue to
Salinity setting menu
“b.SAL” if Set Up menu is
entered from mg/L (ppm)
mode. Otherwise display will
return back to “DPr.7”
Figure 16: Pressure Setting Adjustment Sequence
24
6 HOLD Function
70.5
A
Annunciator Blinks when
meter is in “HOLD” mode
Figure 68: Unit in HOLD mode
%
This feature lets you freeze the display for a delayed observation. HOLD can be
used in any measurement mode.
1. To hold a measurement, press the HOLD/ENTER key while in
measurement mode. The %, mg/L, ppm or T annunciator in the display
will start blinking to indicate that the measurement is being held.
2. To release the held value, press the HOLD/ENTER again. The
annunciator will stop blinking to indicate that the held value has been
released.
3. Continue to take measurements.
Note: This meter has a selectable auto off feature which when activated will
shut the meter off automatically after 20 minutes of non use.
If the meter is shut off either automatically or manually, the HOLD value will
be lost.
7 ADVANCED SETUP FUNCTIONS
7.1 Advanced Setup Overview
The advanced setup mode lets you customize your meter’s preferences and
defaults. There are two Set Up menus: - one which is derived from the %
Saturation or Temperature mode and the other is derived from the mg/L (ppm)
Concentration mode.
7.1.1 To enter the % Saturation or Temperature Set Up menu:
1. Make sure that the meter is switched-off while in the % Saturation or
Temperature measurement mode.
25
2. With the MODE key pressed, power the unit on by pressing the ON key.
COF.1
ELE.3
CA L.2
Release the ON key before releasing the MODE key.
3. “SEt.P” indicator will appear momentarily and “COF.1” will appear next.
4. You are now in the % Saturation or Temperature Set Up menu.
7.1.2To enter the mg/L (ppm) Concentration Set Up menu:
1. Switch off the meter in the mg/L (ppm) Concentration measurement
mode.
2. Repeat steps 2 to 3 as above and you will be in the mg/L (ppm)
Concentration Set Up menu.
The following table combines the explanation for the features available in
both Set Up menus. Figures 19 and 20 give an overall view of the Set Up
sequence for the % Saturation or Temperature Set Up menu and the mg/L
(ppm) Concentration Set Up menu.
SEt.P - Enter Set Up Menu
COF.1 – Enter Configuration Menu
A.ATC –Selection for automatic temperatur
b.DO – Selection of mg/L or ppm of the concentration mode
measurement. (Only available in Concentration Set Up menu)
CAL.2 – Viewing the latest Calibration data
‘View only’ parameter for the latest calibration data of the DO
% Saturation calibration and the DO mg/L (ppm)
Concentration calibration. Calibration data will be viewed
according to its respective Set Up menu.
ELE.3 – Viewing the Electrode properties
FACT – Viewing the Slope Factor
OFS – Viewing the % Saturation offset adjustment. (Only
available in % Saturation Set Up menu)
HI.mV – Viewing the mV value for 100% Saturation
LO.mV – Viewing the mV value for 0% Saturation
AtO.4 – Selection of automatic power off (20 minutes from the
last key press)
26
rSt.5 – Enter Reset Menu
OF5.6
p
Figure 79: Overview of DO 6% Saturation or Temperature Setup Menu
Switch off from % Saturation
or Temperature measurement
mode. Press and hold MODE
key and then switch on.
SEt.P will display for
1 second
MODE
ON
OFF
COF.1
CA L.2
ELE. 3
OF5.6
p
A.CAL – Selection of Calibration Reset (Clear only Calibration
and its Data back to factory default condition)
b.USR – Selection of User Reset (Clear all user setting back to
factory default condition)
OFS.6 - % Saturation Offset Adjustment (Only available in %
Saturation Set Up menu)
DPr.7 – Enter Barometric Pressure Selection Menu
A.HG – Selection of Barometric Pressure Setting Adjustment in
mmHg
A.PA - Selection of Barometric Pressure Setting Adjustment in
Kilo Pascal
b.SAL – Enter the Salinity Setting Adjustment Menu (Only
available in mg/L (ppm) Concentration Set Up menu)
27
Figure 19: Overview of DO6 Concentration Set up Menu
Switch off from Concentration
measurement mode. Press
and hold MODE key and then
switch on.
SEt.P will display for
1 second
MODE
ON
OFF
COF.1
CA L.2
ELE. 3
p
Figure 20: Selection of Automatic or
Manual Temperature
Compensation
COF.1
A.AT c
yes
HOLD
ENTER
HOLD
ENTER
7.2 (COF.1) Configuration
Menu
In this menu, the meter lets you
configure the selection of
automatic temperature
compensation (A) and the
selection of mg/L or ppm mode for
the mg/L (ppm) Concentration
measurement mode. (Second
feature is only available in mg/L
(ppm) Concentration Set Up
menu)
7.2.1Selection of Automatic
Temperature Compensation
From "COF".1" of any Set Up
menu.
28
1.1.Press the HOLD/ENTER key for the display to show the Automatic
Figure 21: Selection of mg/L or ppm
mode
mg/L
ppm
Temperature Compensation selection menu “A.ATC”.
2. Press the HOLD/ENTER key to enter the ATC selection menu.
3. Use the p and q keys to either activate or deactivate the ATC feature.
a.YES – activates the ATC
b.NO – deactivates the ATC and activates the Manual temperature
compensation
4.Press the HOLD/ENTER key to confirm your selection.
Note: The meter will either return to the main group menu “COF.1” if %
Saturation or Temperature Set Up menu is used or will continue to the mg/L
or ppm mode configuration if mg/L (ppm) Concentration Set Up menu is used.
7.2.2 Selection of mg/L or ppm
mode
From the “COF.1” of the mg/L (ppm)
Concentration Set Up menu,
1. Press the HOLD/ENTER key thrice
to enter the mg/L or (ppm) mode
selection menu “b.DO”.
2. Use the ? and ? keys to select the
desired mode of measurement.
3. Press the HOLD/ENTER key to
confirm the selection made and to
return to the subgroup menu.
4. Press the CAL key to return to the
measurement mode.
29
7.3 (CAL.2) Calibration Data
10.00
8.24
CAL
CAL
Figure 22: Viewing the last calibration
data for each measurement range.
CA L.2
HOLD
ENTER
Display will show the latest % saturation
Calibration if set up menu is selected from the
DO%Saturation Mode
Display will show the latest mg/L or ppm
calibration if set up menu is selected from the
DO mg/L (ppm) Concentration Mode
You can view the latest
calibration data for the %
Saturation and the mg/L (ppm)
Concentration calibrations. The
calibration data from each mode
can be viewed in the Set Up
menu respectively.
7.3.1Viewing the %
Saturation calibration dat
From the % Saturation or
Temperature Set Up menu,
1. Press the p and q keys to
scroll through the main
group to enter the viewing
menu “CAL.2” for the latest
calibration data.
2. Press HOLD/ENTER key to
view the latest % Saturation
calibration data.
To exit to the main group
menu, press the
HOLD/ENTER key or the
CAL key
3. To exit to the main group
menu, press the HOLD/ENTER key or the CAL key.
4. Press the CAL key to return to the measurement mode.
7.3.2. Viewing the mg/L (ppm) Concentration calibration data
From the mg/L (ppm) Concentration Set Up mode,
1.. Repeat process 1 to 4 above to view the latest mg/L (ppm) Concentration
calibration data.
at steps 1-5 until all points have been calibrated.
7.4 (ELE.3) Electrode Properties
This menu features data of the electrode properties for diagnostic purposes.
The “view only” parameters such as the electrode Slope Factor, % Saturation
Offset setting, 100% Saturation mV value and 0% Saturation mV value are
30
very useful in determining the life efficiency of the electrode. These electrode
properties can be examined through its data in % Saturation and mg/L (ppm)
Concentration which is available in their Set Up menu respectively.
7.4.1Viewing the electrode Slope Factor
This parameter lets you view and gives an indication of the probe’s efficiency.
The value displayed is the ratio of the actual value produced by the probe to
the theoretical value. The higher the number, the lower the mV output from
the probe. The ratio displays from 0.5 to 1.999.
From the % Saturation or Temperature Set Up menu,
1. Enter the Set Up menu & scroll by using the p and q keys until the
meter displays the Electrode Properties menu “ ELE.3”
2. Press the HOLD/ENTER key to enter the menu and view the Slope Factor
of the electrode..
3. The display will momentarily shows “FACT” before displaying the slope
factor value.
7.4.2Viewing the % Saturation Offset (Only available in % Saturation
or Temperature Set Up menu):
This parameter shows you the amount of the % Saturation Offset entered in
the other parameter “OFS” (refer to Section for instructions).
From Step 3 above,
4. Press the HOLD/ENTER key to enter the % Saturation Offset viewing
menu
5. The meter will display “OFS” momentarily before displaying the last offset
adjustment made.
7.4.3Viewing the 0% Saturation mV value
This parameter show the sensor's mV output corresponding to 100%
Saturation. From Step 5 above,
6. Press the HOLD/ENTER key. The display will show "HI.mV" momentarily
before displaying the sensor's mV output with respect to 100% Saturation.
7.4.4Viewing the 0% Saturation mV value
This parameter lets you view the sensor’s mV output corresponding to 0%
Saturation. From step 6 above.
31
7. Press the HOLD/ENTER key. The display will show “LO.mV”
HOLD
ENTER
HOLD
ENTER
HOLD
ENTER
HOLD
ENTER
“LO.mV” will display
for 1.5 seconds
“FACT”displays
for 1.5 seconds
“OFS” displays
for 1.5 seconds
“HI.mV” displays
for 1.5 seconds
Viewing of the offset
percentage adjustment will be
available only if set up menu is
selected from the DO%
Saturation of the Temperature
Mode. Otherwise, this menu
will be skipped.
Figure 23: Viewing electrode data sequence
%
momentarily before displaying the sensor’s mV output value with respect to
0% Saturation.
8. Press the HOLD/ENTER key again to exit to the main group menu. Press
CAL to return to the measurement mode.
Note: To view the electrode properties in the mg/L (ppm) Concentration
mode, repeat the whole steps above using the mg/L (ppm) Concentration Set
Up menu. Viewing of the % Saturation Offset will be skipped in this menu.
32
7.5 (AtO.4) Automatic Off
HOLD
ENTER
Figure 24: Selection of Automatic
Off function
This feature is useful for batteries energy
conservation. The meter automatically
shuts off 20 minutes from the last key
press.
From any Set Up Menu,
1. Scroll the Set Up main group menu
by using the ? and ? keys until the
meter displays the Automatic Off
menu “AtO.4”.
2. Press the HOLD/ENTER key to enter
the menu.
3. Use the p and q keys to activate or
deactivate the automatic off selection.
a.YES – activates Automatic Off
b.NO – deactivates Automatic Off
4. Press the HOLD/ENTER key to
confirm the selection.
5. Press the CAL key to exit the Set Up menu to return to the measurement
mode.
7.6 (rSt.5) Reset to Factory Default
This mode lets you reset all parameters to factory default settings. There are
two levels of reset:
1. Calibration Reset: - Reset only the calibration values. This clears all
previous calibrated values and reset to factory default. See Section for
factory default settings.
2. User Reset: - Clears all data including calibration and other customized
setup functions and reset it to factory default. See Section 11 for factory
default settings.
33
7.6.1 Calibration Reset
HOLD
ENTER
HOLD
ENTER
Figure 25: Calibration reset
selection
Calibration
Reset
From any Set Up menu:
1. Scroll the Set Up main group menu by
using the p and q keys until the meter
displays the Reset to Factory Default
menu “rSt.5”.
2. Press the HOLD/ENTER key to enter
the menu.
3. Press the HOLD/ENTER key to enter 4.
Use the p and q keys to activate or
deactivate the Calibration Reset
selection.
immediately return to measurement
mode. Otherwise it will return to the
main group menu. Press CAL to return
to the measurement mode.
Note: When you reset your meter from %
Saturation Set Up menu, both % Saturation
and mg/L (ppm) Concentration calibrations
are reset to factory default. However, calibration reset in the mg/L (ppm)
Concentration Set Up menu will only reset the calibration in the
concentration mode and do not affect the calibration in the percentage
saturation mode.
.
34
7.6.2 User Reset
HOLD
ENTER
Figure 26: User Reset Selection
User
Reset
This program clears all data
including calibration and
other customized setup
functions and reset it to
factory default.
You can skip Calibration
Reset and proceed straight to
User Reset
From “rSt.5” menu,
1. Press HOLD/ENTER key
three times until meter
displays the User Reset
menu “b.USR”
2. Press HOLD/ENTER key
to enter the menu.
3. Use the p and q keys to
activate or deactivate the
User Reset selection.
a.YES – activates User
Reset
b.NO – deactivates
User Reset
4. Press HOLD/ENTER key
to confirm the selection
made.
5. If selection is ‘YES’, meter would immediately return to measurement
mode. Otherwise it will return to the main group menu. Press CAL to
return to the measurement mode.
35
7.7 (OFS.6) % Saturation Offset Adjustment
HOLD
ENTER
HOLD
ENTER
Figure 27: % Saturation Offset
Adjustment
Measured reading based on last
percentage saturation calibration
is displayed momentarily
HOLD
ENTER
“CO” will display for
1.5 seconds
Display will return to “OFS.6” subgroup menu
%
%
%
This is a useful feature that allows you
to offset meter’s value when cross
referenced with another DO meter.
That way, it can be standardized
without you having to perform manual
calculation. Your DO 6 meter allows %
Saturation Offset adjustment within +/-
10.0% offset and its adjusted offset
value can be viewed in the Electrode
Properties menu “ELE.3”.
From the % Saturation measurement
mode,
1. Dip the DO electrode in the sample
solution and allow the reading to
stabilize.
2. Check the reading of another DO
meter being used as a reference.
This reference meter should have its
probe immersed in the same sample
solution and at the same depth.
3. Switch off your meter and enter the
% Saturation Set Up menu.
4. Scroll the Set Up main group menu
by using the p and q keys until
the meter displays the % Saturation
Offset Adjustment menu “OFS.6”.
5. Press the HOLD/ENTER key to
enter the menu. The display will
momentarily show the measured
reading based on the last calibration
before displaying the last offset
adjusted value.
6. Use the p and q keys to enter the new value.
7. Press the HOLD/ENTER key to confirm the offset adjustment. “CO” will
be displayed for 1.5 seconds before the display returns to the subgroup
menu.
8. Press CAL key to exit the Set UP menu to return to the measurement
mode.
Note: When a user calibration is done, the offset gets reset to zero.
36
7.8 (DPr.7) Dissolved Oxygen Parameters
This Set Up menu allows you to set the barometric pressure and the salinity
value of the sample to be measured. You are given the option to use mmHg or
kilo Pascal barometric pressure units. The Salinity value to be entered will be
based on ppt and is available in mg/L (ppm) Concentration Set Up menu.
The Set Up menu for barometric pressure correction is “A.Hg” (for mmHg)
and “A.PA” (for kilo Pascal) whereas the Salinity menu is indicated as
“b.SAL”.
7.8.1Pressure Setting Adjustment
1. Switch off your meter in mg/L (ppm) Concentration mode.
2. While pressing and hold the MODE key, switch on the meter by pressing
the ON/OFF key.
3. Once the ON/OFF key is released, the meter will enter the Set Up menu
by displaying “SEt.P”. By releasing the MODE key, the display will show
the configuration menu “COF.1”.
4. Press the p and q keys to scroll through the Set Up main group menu till
the display shows “DPr.7”.
5. Press the HOLD/ENTER key and the meter will enter the barometric
pressure unit selection menu with the display showing either the “A.HG”
mode or “A.PA” mode.
6. Use the p and qkeys to toggle between the barometric pressure units and
press the HOLD/ENTER key to confirm selection.
7. Use the p and q keys to set the pressure value and press the
HOLD/ENTER to confirm the setting. A confirmation indicator “CO” will
be displayed for 1.5 seconds before the meter shows the next display of
Salinity setting “b.SAL”. If the Set Up menu is derived from other than the
mg/L (ppm) Concentration mode, the meter will return to the Set Up
subgroup menu “DPr.7”.
8. Press CAL key to return to measurement mode, or continue to make a
salinity setting adjustment.
37
HOLD
ENTER
Figure 28: Pressure Setting Adjustment Sequence
HOLD
ENTER
HOLD
ENTER
HOLD
ENTER
“CO” will display for
1.5 seconds
HOLD
ENTER
HOLD
ENTER
HOLD
ENTER
“CO” will display for
1.5 seconds
Display will continue to
Salinity setting menu
“b.SAL” if Set Up menu is
entered from mg/l (ppm)
mode. Otherwise display will
return back to “DPr.7”
Display will continue to
Salinity setting menu
“b.SAL” if Set Up menu is
entered from mg/l (ppm)
mode. Otherwise display will
return back to “DPr.7”
38
7.8.2Salinity Setting Adjustment
.sal
HOLD
ENTER
HOLD
ENTER
0.0
45.2
co
“CO” will display
for 1.5 seconds
Display will return back to “DPr.7”
subgroup menu
Figure29: Salinity Setting Adjustment Sequence
Note: This mode is available only from
the mg/L (ppm) Concentration Set Up
menu.
1. Repeat steps 1 to 5 as in the steps
of Pressure Setting Adjustment.
2. Press the HOLD/ENTER key twice
till the meter displays the Salinity
setting mode “b.SAL”.
3. Press the HOLD/ENTER key to
enter the Salinity setting mode.
4. Use the p and q keys to enter the
salinity of your solution in ppt.
Note: The salinity value is in ppt (parts
per thousand)
5. Press HOLD/ENTER key to
confirm the value. A confirmation
indicator “CO” will be displayed for
1.5 seconds before returning to the
subgroup menu “DPr.7”.
6. Press CAL key to return to the
measurement mode.
8 PROBE CARE AND MAINTENANCE
The DO 6 probe is a galvanic measuring element which produces an output
proportional to the oxygen present in the medium in which it is placed. The
galvanic probe design lets you take measurements immediately – without the
typical 15 minute wait of other dissolved oxygen probes.
The probe consists of two parts:
· An upper part consisting of the anode, a cathode, and the dual cable.
· A lower part consisting of a pre-membraned cap, and electrolyte solution.
Oxygen diffuses through the membrane onto the cathode, where it is
consumed. This process produces an electrical current which flows through the
cable to the meter. The electric current produced is proportional to the oxygen
that passes through the membrane and the layer of electrolyte. This makes it
39
possible to measure the partial pressure of oxygen in the sample at a given
temperature.
Since the DO in the sample is consumed by the cathode it is essential that a
new sample must flow past the membrane of the probe to prevent the
occurrence of false readings. The probe uses very little oxygen for its
measurement. This enables it to function correctly with liquid movement as
low as 2 inch/sec across membrane.
The permeability of the membrane to oxygen varies greatly with temperature.
Therefore compensation is needed for this variation. The DO6 probe comes
with an in-built Temperature Compensation for the membrane variation.
8.1 Probe Care
Proper care and maintenance will help you receive the maximum probe life
and ensure more accurate readings.
Since any deposits on the membrane surface act as a barrier to oxygen diffusing
through the membrane, the membrane must be cleaned at regular intervals to
assure maximum reliability.
After using the probe, rinse the probe with clean water and wipe it with a soft
cloth or paper to avoid any hardening of deposits if necessary. If growth
develops on the probe, use a disinfecting chemical to clean.
NOTE: Although the membrane is strong and not easily damaged, wipe it
gently while cleaning it. If the membrane is damaged or torn, the probe will no
longer function.
There are no special probe storage requirements.
8.2 Pre-Membraned Caps Replacement
Replacement of the pre-membraned cap is required only when you cannot
calibrate the probe, or if the membrane is damaged.
Typical membrane damages are punctures or wrinkles caused during
measurements or cleaning.
To order replacement probe components or a replacement probe, see Section
“Accessories” on page 48.
8.3 To replace the probe pre-membraned cap.
1. Unscrew counter clockwise the pre-membraned cap from the probe sensing
tip and discard.
2. Wash the probe under running water.
3. Mount the nozzle tip onto the syringe provided. Fill the syringe with the
refill solution through the tip of the plastic bottle.
40
4. Hold the probe upside down. Insert the nozzle tip into one of the 4 holes
surrounding the silver cathode. Inject the fill solution into the probe body
until solution leaks out from the fill hole (approximately 5 mL).
5. Replace pre-membraned cap by tightening clockwise until hand tightened.
6. Allow at least 1 hour for the electrode to equilibrate before usage.
Figure 30: probe pre membraned cap replacement process
8.4 Electrolyte Solution
The electrolyte solution in your probe’s cap will deplete on usage and will need
to be replaced periodically.
Your DO probe comes with replacement electrolyte solution. The replacement
electrolyte comes premixed and ready to use. To order more electrolyte
solution, see “Accessories” section.
41
9 TROUBLE-SHOOTING GUIDE
ProblemCauseSolution
No
display
when
turned on
a) Batteries not in place
b) Batteries not in
correct polarity (+
and – position).
c) Weak batteries
a. Remove rubber
b) Reinsert batteries
c) Replace batteries.
Unstable
readings
a) Insufficient
electrolyte in probe.
a) Fill probe with
b) Air bubbles trapped
around the probe
b) Stir or tap probe to
c) Dirty or damaged
probe
d) Probe not deep
c) Clean the probe and
d) Make sure sample
enough in sample
e) External noise
pickup or induction
e) Move or switch off
caused by nearby
electric motor.
f) Replace probe.
f) Broken probe.
Slow
response
a) Dirty / Oily probe.
b) Temperature is
a) Clean probe. See
changing.
b) Allow time for
Not
respondin
gtokey
press
a) HOLD operation
indicated by flashing
display.
b) Damaged key pad
a) Cancel HOLD mode.
b) Return to dealer.
c) Reset all internal
c) Internal program
error.
boot/stand. Check
that batteries are in
place and making
good contact.
with correct polarity.
electrolyte & replace
pre-membraned cap
remove bubbles.
re-calibrate
entirely covers the
probe sensors.
interfering motor.
“Probe Care &
Maintenance”.
temperature to
stabilize
programs by
reinserting batteries.
42
10 ERROR MESSAGES
+
-
LCD DisplayIndicatesCauseSolution
Low Battery
indicator
Low battery
level
Need new batteries or
battery connection is
bad.
Clean battery contacts.
Replace batteries with
fresh ones, noting
polarity.
"Err 1"
in %
Saturation
Mode
% Saturation
Calibration
error
Calibration is
performed when
factory calibrated
absolute value is out
of calibration range –
10.1% to 49.9%
Check the value of the
calibration solution. If
zero calibration is done,
make sure the limit of
10% is not exceeded.
If message persist,
recondition your probe.
"Err 1" in mg/L
(ppm)
Concentration
Mode
“UR”/”OR”
with blinking
“A”
annunciator in
Temperature
Mode
“——“with
blinking “A”
annunciator in
% Saturation
and
Concentration
Mode
Concentration
Calibration
error
ATC probe
error, Under
Range, Over
Range
ATC probe
error
Calibration is
performed when the
calibration solution
factory absolute
measurement is below
2.00 mg/L or ppm
ATC probe is
removed or broken
while the meter has
the ATC feature
activated.
ATC probe is
removed or broken
while the meter has
the ATC feature
activated
43
Check the calibrating
solution to be above 2
mg/L or ppm.
Check that the correct
temperature and
salinity setting has been
set prior calibration.
Check the probe’s
temperature input
phono jack connection
to the meter. Ensure
probe is not broken or
punctured.
Check the probe’s
temperature input
phono jack connection
to the meter. Ensure
probe is not broken or
punctured.
11 FACTORY DEFAULT SETTINGS
Setting
Menu
COF.1Unit ConfigurationParameter Title
A.ATCSelection of
b.DOSelection of mg/L or
CAL.2Viewing the latest
ELE.3Viewing the Electrode
FACTView the Slope Factor Viewing OnlyViewing Only 1.000
OFSViewing the %
HI.mVViewing the mV value
LO.mVViewing the mV value
AtO.4Selection of automatic
rSt.5Reset to factory
A.CALSelection of
b.USRSelection of User
FunctionParameter
Opting Keys
Only
Automatic or Manual
Temperature
Compenstion
ppm of the
concentration mode
calibration data
properties
Saturation offset
adjustment
for 100% Saturation
for 0% Saturation
power off
defaults
Calibration Reset
Reset
p and qYES; NOYES
p and qmg/L or ppmmg/L
Viewing OnlyViewing Only ----
Parameter Title
Only
Viewing OnlyViewing Only 0.0%
Viewing OnlyViewing Only 50 mV
Viewing OnlyViewing Only 0 mV
p and qYES; NO
Parameter Title
Only
p and qNO, YESNO
p and qNO, YES0.0%
Parameter
Optons
Parameter
Title Only
Parameter
Title Only
Parameter
Title Only
Default
Setting
NO
44
Setting
Menu
FunctionParameter
Opting Keys
Parameter
Options
Default
Setting
OFS.6Offset Adjustment (%
Saturation
DPr.7Barometric Pressure
Selection Menu
A.HGSelection of
Barometric Pressure
Setting Adjustment in
mmHg
A.PASelection of
Barometric Pressure
Setting Adjustment in
Kilo Pascal
b.SALSalinity Setting
Adjustment
Setting the Manual
Temperature
Compensation
p and q+/-10.0 of
Parameter Title
Only
p and q500 to 1499
p and q66.6 to 199.9
p and q0.0 to 50.0 ppt 0.0 ppt
CAL, p and q 0.0 to 50.0 °C 25.0 °C
measured
reading
Parameter
Title Only
mmHg
kPA
0.0%
760 mm Hg
101.3 kPA
Note: The HOLD/ENTER key is used to confirm every parameter opted.
45
12 SPECIFICATIONS
% Saturation Mode
Range
Resolution
Relative accuracy
mg/L (ppm)
Concentration Mode
Range
Resolution
Relative accuracy
Temperature
Range
Resolution
Relative accuracy
Salinity Correction
Range
Resolution
Method
Barometric Pressure
Correction (mm Hg)
Range
Resolution
Method
0.00 – 200.0 %
0.1 %
± 1.5% of Full Scale
0.00 – 20.00 mg/L or ppm
0.01 mg/L; 0.01 ppm
± 1.5% of Full Scale
-5.0 – 105.0 °C
0.1 °C
± 0.5 °C
0.0 – 50.0 ppt
0.1 ppt
Automatic correction after manual
input
500 to 1499 mm Hg or 66.6 to 199.9
kPA
1mmHgor0.1kPA
Automatic correction after manual
input
Automatic Temperature
0.0 to 50.0°C
Compensation
Manual Temperature
0.0 to 50.0°C
Compensation
Probe (DO / Temp)Galvanic / Thermistor
Probe DiameterBody 12 mm, Cap 16 mm
Response Time60 seconds to achieve 95% of the
reading
46
% Saturation Calibration
Points
100% in saturated air or
air-saturated water.
0% in zero oxygen solution
% Saturation Calibration
Limits
Factory calibrated absolute value of
10.0% and below for 0% point &
50% to 200% for 100% point.
Concentration Calibration
Window
+/- 40% from the factory default
measurement value. Minimum
reading allowed is 2.00 mg/L (ppm).
Temperature Calibration
Window
Offset Adjustments (%
Saturation)
+/- 5°C from factory default
measurement
+/- 10.0 of reading in Saturation
mode
HOLD functionYes
Auto-Off functionSelectable Auto Off function. (20
minutes after last key press)
DisplayCustoms Single 4 Digit LCD
InputsBNC for DO & 2.5 mm Phono for
temperature
Operating Range0 to 50 °C
Power Requirements4 AAA-sized batteries (included)
Battery Life> 700 hours (Alkaline Batteries)
DimensionsMeter: 14 cm (L)x7cm(W)x3.5
cm (H)
Probe: 115 mm (L) x 12 mm (Dia)
with 3-ft cable.
Probe’s membrane housing: 16 mm
(Dia)
47
13 ACCESSORIES
Code
DO 6 meter and probe kit5-0107
DO probe26094-65
Replacement caps (2) with
pre-installed membrane and 10-mL
electrolyte
26094-66
14 ADDITIONAL INFORMATION
14.1 Dissolved Oxygen
14.1.1 General Information
Dissolved Oxygen (DO) refers to the volume of oxygen that is contained in
water. There are two main sources of DO in water: from atmosphere and
photosynthesis. Waves and tumbling water mix air into the water where
oxygen readily dissolves until saturation occurs. Oxygen is also produced by
aquatic plants and algae as a by-product of photosynthesis.
The amount of DO that can be held by water depends on 3 factors: water
temperature, salinity, and atmospheric pressure.
1. Amount of DO increases with decreasing temperature (colder water holds
more oxygen);
2. Amount of DO increases with decreasing salinity (freshwater holds more
oxygen than saltwater does);
3. Amount of DO decreases with decreasing atmospheric pressure (amount of
DO absorbed in water decreases as altitude increases).
48
Solubility of oxygen in water contact with water saturated air at
standard atmospheric pressure
Solubility mg/L
16
14
12
10
8
6
4
2
0
0 102030405060
Temperature °C
14.1.2 Measurement Units
One measure of DO in water is parts per million (ppm) which is the number of
oxygen molecules (O
the % Saturation is another way to analyze DO levels. % Saturation is the
measured DO level divided by the greatest amount of oxygen that the water
could hold under various temperature and atmospheric pressure conditions
multiplied by 100.
14.1.3 What Is Being Measured?
DO probes respond to the partial pressure of oxygen in liquid or gas being
2) per million total molecules in a sample. Calculating
measured – they measure the “pressure” of oxygen rather than concentration.
All of the oxygen entering the probe is consumed at the cathode where it is
49
electrochemically reduced to hydroxyl ions producing an electrical current
within the probe:
O
2 +2H2O+4e
-
à 4OH
–
Since all oxygen entering the probe is chemically consumed, the partial
pressure of oxygen in the electrolyte is zero. Therefore, a partial pressure
gradient exists across the membrane and the rate at which oxygen enters the
probe is a function of the partial pressure of oxygen in the gas or in liquid being
measured.
When a probe is placed in air saturated water, the current it produces will not
be affected by the temperature or salinity of the water. The DO concentration
in the water, however, will vary with temperature and salinity. Because it is
convenient to report DO concentration in mg/L or ppm, it is necessary to adjust
for temperature and salinity of the water to get correct readings in these units.
If DO were to be reported in terms of partial pressure or % Saturation, then
temperature and/or salinity compensation for oxygen solubility would not be
necessary. Most probes are temperature compensated – i.e. they convert the
“partial pressure measurement” to mg/L of DO at whatever temperature the
water happens to be at for a given salinity and barometric pressure.
14.1.4 Air Calibration
Understanding the principle of air calibration is easy, once you know that it is
partial pressure that the probe is responding to. When the probe is in air, it is
measuring the partial pressure of oxygen in air. If water is air saturated, then
the partial pressure of oxygen in the water will be the same as it is in air.
Therefore, all you need to know is the temperature of the air in which the
probe is placed. By consulting solubility tables for oxygen at the particular
barometric pressure and salinity of the water being measured, the
corresponding concentration (mg/L or ppm) can be found for air saturated
water at the air calibration temperature, and the meter can be set accordingly.
Because most meters are temperature compensated, they will still give correct
readings in mg/L even though the actual water temperature may be different to
the air calibration temperature.
Note: The closer the air calibration temperature to the water temperature, the
more accurate is the calibration.
50
14.1.5 Applications
Oxygen is essential for fish, invertebrate, plant, and aerobic bacteria
respiration. DO levels below 3 ppm are stressful to most aquatic organisms.
Levels below 2 or 1 ppm will not support fish. Fish growth and activity usually
require 5 to 6 ppm of DO, an important consideration for aquaculture industry.
Low DO indicates a demand on the oxygen of the system. Natural organic
material such as leaves accumulate in the stream and create an oxygen demand
as it is decomposed. Organic materials from human activities also create an
oxygen demand in the system. Micro-organisms consume oxygen as they
decompose sewage, urban and agricultural run-off, and discharge from
food-processing plants, meat-packing plants and diaries. There is an optimum
DO level for this process and if DO level falls too low, the micro-organisms die
and the decomposition ceases. If DO level is too high, more power is used
than necessary for aeration and the process becomes costly.
In boiler water application, presence of oxygen in the water will increase
corrosion and helps build up boiler scale that inhibits heat transfer. In such
instance it is critical to keep DO concentration to a minimum.
Some pollutants such as acid mine drainage produce direct chemical demands
on oxygen in the water. DO is consumed in the oxidation-reduction reactions
of introduced chemical compounds such as nitrate (NO
1+
(NH
4
), sulfate (SO42-), and sulfite (SO32-) and ferrous (Fe2+) and ferric (Fe3+)
1-
3
) and ammonia
ions.
These are important consideration for water and wastewater treatment
industry.
15 LIMITED WARRANTY
LaMotte Company warrants this instrument to be free of defects in parts and
workmanship for 3 years from the date of shipmen and this probe to be free of
defects in parts and workmanship for 6 months from the date of shipment. If it
should become necessary to return the instrument for service during or beyond
the warranty period, contact our Technical Service Department at
1-800-344-3100 for a return authorization number or visit www.lamotte.com
for troubleshooting help. The sender is responsible for shipping charges,
freight, insurance and proper packaging to prevent damage in transit. This
warranty does not apply to defects resulting from action of the user such as
misuse, improper wiring, operation outside of specification, improper
maintenance or repair, or unauthorized modification. LaMotte Company
specifically disclaims any implied warranties or merchantability or fitness for a
specific purpose and will not be liable for any direct, indirect, incidental or
consequential damages. LaMotte Company’s total liability is limited to repair
51
or replacement of the product. The warranty set forth above is inclusive and
no other warranty, whether written or oral, is expressed or implied.
16 RETURN OF ITEMS
Should it be necessary to return the instrument for repair or servicing, pack
instrument carefully in suitable container with adequate packing material. A
return authorization number must be obtained from LaMotte Company by
calling 1-800-344-3100. Attach a letter with the authorization number to the
shipping carton which describes the kind of trouble experienced. This valuable
information will enable the service department to make the required repairs
more efficiently.
52
LaMOTTE COMPANY
Helping People Solve Analytical Challenges®
802 Washington Ave. • PO Box 329 • Chestertown • Maryland • 21620
USA800-344-3100 • 410-778-3100 • Fax 410-778-6394
Visit us on the web at www.lamotte.com
5-0107 - 03/10
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.